MODIFCATION TO: NUVASIVE NEUROVISION EMG ENDOTRACHEAL TUBE
K094054 · Nuvasive, Inc. · GWF · May 14, 2010 · Neurology
Device Facts
Record ID
K094054
Device Name
MODIFCATION TO: NUVASIVE NEUROVISION EMG ENDOTRACHEAL TUBE
Applicant
Nuvasive, Inc.
Product Code
GWF · Neurology
Decision Date
May 14, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.1870
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NeuroVision® EMG Endotracheal Tube is intended for use with any compatible monitoring system during surgical procedures for continuous EMG neurological monitoring and status assessment of the nerves supplying the laryngeal musculature as well as for providing an open airway for patient ventilation.
Device Story
Endotracheal tube with integrated electrodes for electromyographic (EMG) monitoring; used during surgery to monitor recurrent laryngeal nerve (RLN) function. Device features 4 laryngeal surface electrodes (conductive silver ink) on a flexible PVC tube with an inflatable low-pressure cuff. Connects to compatible external EMG monitoring systems. Operated by clinicians in surgical settings to provide airway management and real-time nerve status assessment. Output displayed on external monitor; assists surgeons in identifying and protecting nerves during procedures. Benefits include continuous nerve monitoring and patient ventilation.
Clinical Evidence
Bench testing only. Testing included biocompatibility (cytotoxicity, intracutaneous, sensitization per ISO 10993), functional bench testing (ISO 5361, inflation valve functionality, leak test, electrode resistance, system integration, bending test), stability testing, and sterilization/packaging validation. Results demonstrate the device meets established specifications.
Technological Characteristics
Flexible PVC endotracheal tube with inflatable low-pressure cuff; 4 laryngeal surface electrodes using conductive silver ink. Dimensions vary. Sterile, single-use. Connectivity via integration with external EMG monitoring systems. No reinforcing material.
Indications for Use
Indicated for patients undergoing surgical procedures requiring continuous EMG neurological monitoring and status assessment of nerves supplying laryngeal musculature, while simultaneously providing an open airway for ventilation.
Regulatory Classification
Identification
An evoked response electrical stimulator is a device used to apply an electrical stimulus to a patient by means of skin electrodes for the purpose of measuring the evoked response.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Nuvasive. The logo consists of the word "NUVASIVE" in bold, sans-serif font. Below the word "NUVASIVE" is the phrase "Creative Spine Technology" in a smaller, italicized font. To the left of the word "NUVASIVE" is a stylized graphic.
サイ
# 510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:
#### A. Submitted by:
Sheila Bruschi Senior Regulatory Affairs Associate NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 320-4515 Fax: (858) 320-4615 Date Prepared: May 13, 2010
MAY 1 4 2010
#### B. Device Name
| Trade or Proprietary Name: | NuVasive® NeuroVision® EMG Endotracheal Tube |
|----------------------------|-------------------------------------------------------------------------|
| Common or Usual Name: | Endotracheal Tube with Electromyography (EMG)<br>monitoring Electrodes. |
| Classification Name: | Tracheal Tube<br>Inflatable Cuff<br>Surgical nerve stimulator/locator |
| Device Class: | Class II |
| Classification: | §874.1820, §882.1870, §868.5730 and §868.5750 |
| Product Code: | ETN, GWF |
#### C. Predicate Devices
The subject NeuroVision® EMG Endotracheal Tube is substantially equivalent to the following predicate devices currently distributed commercially in the U.S .:
- . K090298 - NuVasive NeuroVision EMG Endotracheal Tube
- K032491 ECOM™ CV4 Endotracheal Cardiac Output Monitor .
- K050162 Contact™ EMG Rotatable Endotracheal Tube .
#### D. Device Description
The NuVasive® NeuroVision® EMG Endotracheal (ET) Tube is an endotracheal tube with integrated electrodes for electromyographic (EMG) monitoring during surgery. The ET tube is made of a flexible PVC material with an inflatable low pressure cuff. The NeuroVision EMG ET Tube is provided as a sterile, single use disposable accessory that connects to a compatible EMG monitor to provide an open airway for patient ventilation during EMG neuromonitoring of the Recurrent Laryngeal Nerve (RLN).
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for NUVASIVE. The logo consists of the word "NUVASIVE" in a bold, sans-serif font. Below the word "NUVASIVE" is the phrase "Creative Spine Technology" in a smaller, italicized font. To the left of the word "NUVASIVE" is a stylized graphic.
#### E. Intended Use
The NeuroVision® EMG Endotracheal Tube is intended for use with any compatible monitoring system during surgical procedures for continuous EMG neurological monitoring and status assessment of the nerves supplying the laryngeal musculature as well as for providing an open airway for patient ventilation.
#### F. Technological Characteristics
As was established in this submission, the subject NeuroVision® EMG Endotracheal Tube is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have the same technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, function, packaging, and sterilization. See the table below for a detailed comparison:
| | Subject Device | Predicate Devices | | |
|--------------------------------|--------------------------------------------|---------------------------------------------------|-----------------------------------------------------------------|-------------------------------------------------------------|
| | NeuroVision<br>EMG<br>Endotracheal<br>Tube | NeuroVision EMG<br>Endotracheal Tube<br>(K090298) | ECOM CV4<br>Endotracheal<br>Cardiac Output<br>Monitor (K032491) | Contact™ EMG<br>Rotatable<br>Endotracheal Tube<br>(K050162) |
| Laryngeal Surface<br>Electrode | YES | YES | YES | YES |
| Endolaryngeal<br>Location | YES | YES | YES | YES |
| Number of<br>Electrodes | 4 | 4 | unknown | 4 |
| Electrode<br>Surface Material | Conductive<br>Silver Ink | Stainless Steel Wire | Conductive<br>Silver Ink | Stainless Steel Wire |
| Tube & Cuff<br>Materials | PVC | Silicone | PVC | Silicone |
| Reinforcing<br>Material | None | Stainless Steel Wire | None | Stainless Steel Wire |
| Tube Dimensions | Various<br>Dimensions | Various Dimensions | unknown | Various Dimensions |
| Sterilization &<br>Packaging | Sterile, single use<br>only | Sterile, single use<br>only | Sterile, single use<br>only | Sterile, single use<br>only |
--
{2}------------------------------------------------
#### G. Performance Data
Nonclinical testing was performed to demonstrate that the subject NeuroVision® EMG Endotracheal Tube is substantially equivalent to other predicate devices. The following testing was performed:
- . Biocompatibility testing per ISO 10993-1 requirements, including:
- · cytotoxicity testing (per ISO 10993-5)
- · intracutaneous testing (per ISO 10993-10)
- · sensitization testing (per ISO 10993-10)
- Functional bench testing, including: ●
- · Functional testing per ISO 5361
- inflation valve functionality
- leak test
- electrode resistance test
- system integration test
- " bending test
- . Stability Testing
- . Sterilization & Packaging Validations
The results of these studies showed that the modified device meets the same specifications and criteria as the predecessor predicate device, and the device was therefore found to be substantially equivalent.
#### H. Conclusions
Based on the indications for use, technological characteristics, performance testing, and comparison to predicate devices, the subject NeuroVision® EMG Endotracheal Tube has been shown to be substantially equivalent to legally marketed predicate devices, and safe and effective for its intended use.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a serif font. The text is centered on the image and is the only element present.
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling an emblem. The emblem is a stylized representation of a bird or abstract design, possibly symbolizing health and well-being.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
NuVasive, Inc. c/o Ms. Sheila Bruschi Senior Regulatory Affairs Associate 7475 Lusk Blvd. San Diego, CA 92121
MAY. 1 4, 2010
Re: K094054
Trade/Device Name: NuVasive® NeuroVision EMG Endotracheal Tube Regulation Number: 21 CFR 882.1870 Regulation Name: Evoked Response Electrical Stimulator Regulatory Class: Class II Product Code: GWF, ETN Dated: April 12, 2010 Received: April 13, 2010
Dear Ms. Bruschi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm11.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Qula. Kumm, wo fr
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
| 510(k) Number (if known): | K |
|---------------------------|---|
|---------------------------|---|
Device Name:
Indications For Use:
The NeuroVision EMG Endotracheal Tube is intended for use with any compatible monitoring system during surgical procedures for continuous EMG neurological monitoring and status assessment of the nerves supplying the laryngeal musculature as well as for providing an open airway for patient ventilation.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
co61
.
Page 1 of 1
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K094054
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.