K232542 · MicroVention, Inc. · QJP · May 16, 2024 · Cardiovascular
Device Facts
Record ID
K232542
Device Name
Wedge XL Delivery Catheter
Applicant
MicroVention, Inc.
Product Code
QJP · Cardiovascular
Decision Date
May 16, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The Wedge XL Delivery Catheter is intended to assist in the delivery of interventional devices, such as distal access catheters, in the neurovasculature.
Device Story
Single lumen catheter; introduced over steerable guidewire to access small, tortuous neurovasculature. Semi-rigid proximal section transitions to flexible distal tip; distal segment features larger diameter for navigation stability. Radiopaque markers at distal end for fluoroscopic visualization; hydrophilic coating on distal 110 cm for lubricity. Luer fitting on hub for accessory attachment. Used by physicians in clinical settings for neurointerventional procedures. Facilitates delivery of distal access catheters; enables navigation through complex vascular anatomy. Benefits patient by providing stable platform for interventional device placement.
Clinical Evidence
No human clinical data. Bench testing included physical attributes, coating lubricity/durability, simulated use, burst pressure, leak testing, flexural fatigue, and kink resistance. Biocompatibility testing (ISO 10993) confirmed non-cytotoxic, non-sensitizing, non-irritant, non-pyrogenic, and thromboresistant properties. GLP-compliant porcine animal study (n=not specified) compared subject device to predicate, evaluating tracking performance and vascular response at 3-day and 30-day time points; results showed comparable histological impact and safety profiles.
Technological Characteristics
Single lumen catheter; hydrophilic polymer coating; radiopaque markers. Sterilized via ethylene oxide. Luer-connector per ISO 80369-7. Materials evaluated for biocompatibility per ISO 10993. No electronic components or software.
Indications for Use
Indicated for use in the neurovasculature to assist in the delivery of interventional devices, such as distal access catheters. Prescription use only.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
May 16, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is the Department of Health & Human Services logo, which features a stylized caduceus symbol. To the right of this is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
MicroVention, Inc. Alick Tan, Ph.D. Associate Principal, Regulatory Affairs 35 Enterprise Aliso Viejo, California 92656
#### Re: K232542
Trade/Device Name: Wedge XL Delivery Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP Dated: April 15, 2024 Received: April 15, 2024
#### Dear Alick Tan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Naira Muradyan -S
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232542
Device Name Wedge XL Delivery Catheter
Indications for Use (Describe)
The Wedge XL Delivery Catheter is intended to assist in the delivery of interventional devices, such as distal access catheters, in the neurovasculature.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) Owner | MicroVention, Inc.<br>35 Enterprise<br>Aliso Viejo, CA 92656<br>Establishment Registration No: 3013556777 |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Alick Tan, PhD<br>Associate Principal, Regulatory Affairs<br>Telephone: 949-615-2603<br>Email: alick.tan@microvention.com |
| Date Summary<br>Prepared | May 14, 2024 |
| Trade Name | Wedge XL Delivery Catheter |
| Common Name | Percutaneous Catheter |
| Classification | Class II, QJP |
| Predicate Device | Wedge Microcatheter (K172014) |
| Reference Device | Delivery Catheter of the Route 92 Medical 088 Access System<br>(K200121) |
## Device Description
The Wedge XL Delivery Catheter is a single lumen catheter designed to be introduced over a steerable quidewire to access small, tortuous vasculature. The semi-rigid proximal section transitions to a flexible distal tip to facilitate advancement through vessels. Radiopaque markers at the distal end facilitate fluoroscopic visualization. A larger diameter distal segment helps provide stability for navigation. The distal 110 cm outer surface of the Wedge XL Delivery Catheter is coated with a hydrophilic polymer to increase lubricity. A luer fitting on the Wedge XL Delivery Catheter hub is used for the attachment of accessories.
#### Indications for Use
The Wedge XL Delivery Catheter is intended to assist in the delivery of interventional devices, such as distal access catheters, in the neurovasculature.
# Comparison to the Predicate Device
Substantial equivalence of the Wedge XL Delivery Catheter to the predicate device Wedge Microcatheter and to the reference device Delivery Catheter of the Route 92 Medical 088 Access System was established through an evaluation of the intended use, indications for use, principles of operation, device design, materials of construction, and an assessment of usability, safety, and effectiveness via bench and animal studies.
{4}------------------------------------------------
The Wedge XL Delivery Catheter and the predicate device Wedge Microcatheter (K172014)
- have similar intended use,
- . use the same principle of operation,
- incorporate the same basic design, .
- use similar construction and materials, and
- are ethylene oxide (EtO)-sterilized and packaged using the same processes. •
## Device Comparison Table
| | Delivery Catheter of<br>the Route 92 Medical<br>088 Access System<br>(K200121) | Wedge<br>Microcatheter<br>(K172014) | Wedge XL Delivery<br>Catheter<br>(K232542) |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Reference Device | Predicate Device | Subject Device |
| Indications for<br>Use | The Route 92 Medical<br>088 Access System,<br>110 cm, is indicated<br>for use with<br>compatible guide<br>catheters in facilitating<br>the insertion and<br>guidance of<br>microcatheters into a<br>selected blood vessel<br>in the neurovascular<br>system. | The Wedge<br>Microcatheter is<br>intended for general<br>intravascular use,<br>including the<br>peripheral, coronary<br>and neuro vasculature<br>for the infusion of<br>diagnostic agents,<br>such as contrast<br>media, and to assist in<br>the delivery of<br>interventional devices,<br>such as the SOFIA 6F<br>Catheter, in the<br>neurovasculature. | The Wedge XL<br>Delivery Catheter is<br>intended to assist in<br>the delivery of<br>interventional devices,<br>such as distal access<br>catheters, in the<br>neurovasculature. |
| Device<br>Classification | Class II<br>QJP<br>21 CFR 870.1250 | Class II<br>DQY<br>21 CFR 870.1250 | Class II<br>QJP<br>21 CFR 870.1250 |
| Inner Diameter (ID) | 0.019 in. | 0.021 in. | Same as predicate<br>(K172014) |
| Outer Diameter<br>(OD) | Distal 0.080 in.<br>Proximal 0.062 in. | 0.028 - 0.034 in. | Distal Tip 0.056 in.<br>Proximal 0.065 in. |
{5}------------------------------------------------
| | Delivery Catheter of<br>the Route 92 Medical<br>088 Access System<br>(K200121) | Wedge<br>Microcatheter<br>(K172014) | Wedge XL Delivery<br>Catheter<br>(K232542) |
|-----------------------------|--------------------------------------------------------------------------------|-------------------------------------|--------------------------------------------|
| | Reference Device | Predicate Device | Subject Device |
| Shapeable Tip | No | Shapeable | No |
| Distal<br>Segment/Bulb | Yes | Yes | Yes |
| Bulb OD | 0.080 in. | 0.064 – 0.068 in. | 0.083 in. |
| Effective/Working<br>Length | 151 cm | 158 – 160 cm | 160 cm |
| Coating | Hydrophilic coating | Hydrophilic coating | Same |
| Method of Supply | Sterile and single use | Sterile and single use | Same |
| Method of<br>Sterilization | Ethylene oxide | Ethylene oxide | Same |
# Performance Testing - Bench
The following bench performance testing was performed to support the design of the Wedge XL Delivery Catheter and demonstrate that it performs as intended. All tests passed the pre-determined acceptance criteria.
- Physical Attributes ●
- Surface Contamination ●
- Coating Lubricity and Durability ●
- Simulated Use ●
- Preparation/Ease of Assembly o
- Introducer Sheath Insertion O
- Tracking With/Without Guidewire O
- Delivery Catheter/Guidewire Lock Up O
- Catheter Ovalization O
- Lubricity/Durability of Hydrophilic Coating o
- Guide Catheter Tracking over Wedge XL Delivery Catheter o
- Particle Testing o
- Dynamic Burst Pressure ●
- Freedom from Air Leakage ●
- Freedom from Liquid Leakage ●
- Static Burst Pressure ●
- Force at Break ●
- Catheter Flexural Fatigue ●
- Catheter Particle Testing ●
- Kink Resistance
- . Torque Strength
- Luer-Connector Dimensions and Performance Testing per ISO 80369-7 ●
{6}------------------------------------------------
Radiopacity testing and corrosion resistance testing were previously conducted on a test article that was equivalent to the Wedge XL Delivery Catheter in all aspects relevant to the testing performed, therefore it was deemed unnecessary to repeat the testing for Wedge XL Delivery Catheter.
### Biocompatibility Evaluation
| Test | Test Summary | Results |
|----------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|----------------------|
| Cytotoxicity<br>(ISO MEM Elution<br>Test) | The test article extract exhibited slight reactivity<br>(grade 1). | Non-cytotoxic |
| Sensitization (ISO<br>Guinea Pig<br>Maximization Test) | No sensitization was observed in the tested<br>animals. | Non-sensitizer |
| Irritation Reactivity<br>(ISO Intracutaneous<br>Reactivity Test) | No irritation was observed in the tested<br>animals. | Non-irritant |
| Material Mediated<br>Pyrogenicity<br>(ISO/USP Material<br>Mediated Pyrogenicity<br>Test) | No single animal showed a temperature rise of<br>0.5 °C or more above its baseline temperature. | Non-pyrogenic |
| Systemic Toxicity (ISO<br>Acute Systemic<br>Toxicity Test) | No weight loss, mortality, or evidence of<br>systemic toxicity in the tested animals. | No systemic toxicity |
| Hemocompatibility<br>Hemolysis Assay -<br>Extract and Direct<br>Methods<br>(ASTM<br>Hemolysis Assay) | The test article demonstrated 1.2% hemolysis<br>in direct contact and 0.0% hemolysis in<br>indirect contact. | Non-hemolytic |
| Hemocompatibility<br>Complement<br>Activation Assay<br>(ISO<br>Hemocompatibility:<br>Complement<br>Activation Assay<br>(SC5b-9)) | The human serum exposed to the test article<br>for 60 minutes was found not to activate<br>SC5b-9. | Non-activator |
{7}------------------------------------------------
| Test | Test Summary | Results |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|------------------|
| Hemocompatibility<br>Partial Thromboplastin<br>Time (PTT) Test<br>(ASTM<br>Hemocompatibility:<br>Partial Thromboplastin<br>Time (PTT) Assay) | The average clotting time of the test article<br>showed no significant difference from the<br>control. | Non-activator |
| Hemocompatibility<br>Heparinized Blood<br>Platelet and Leukocyte<br>(ASTM<br>Hemocompatibility:<br>Heparinized Blood<br>Platelet and<br>Leukocyte Count<br>Assay) | The average leukocyte and platelet counts of<br>the test article showed no significant<br>difference from the control. | Non-activator |
| Hemocompatibility<br>In-Vitro Blood Loop<br>Assay<br>(ISO In-vitro Blood<br>Loop Assay with<br>Comparison Article) | The worst-case sample thrombus observed<br>on the test article showed minimal thrombus<br>formation. | Thromboresistant |
# Animal Study
Acute, subacute, and chronic animal testing was conducted in accordance with Good Laboratory Practice (GLP) Regulation (21 CFR Part 58) comparing the Wedge XL Delivery Catheter to the predicate Wedge Microcatheter (K172014). The study was designed to assess the safety and performance of catheter tracking in a porcine model. The porcine model was chosen because the vessel sizes allow the insertion and navigation of standard-sized catheters used in humans. Animals were survived to subacute (3 days) and chronic (30 days) time points after procedures to evaluate the vascular response to device use. The study results demonstrated that use of the subject Wedge XL Delivery Catheter with the SOFIA 88 Catheter (K214024) and Wedge Microcatheter with the SOFIA 6F Catheter (K172014, K150366) performed equally in all animal testing. During postmortem examination there were no remarkable gross findings in any of the vessels evaluated with either test or control devices. There was a comparable histological impact of device use on arterial sections in which test and control devices were used. Artery sections subjected to the stiffest segment of the Wedge XL Delivery Catheter exhibited similar tissue responses as to control devices. Overall, the histologic findings were consistent with routine catheterization procedures, which are commonly observed after interventional procedures in porcine models. The results of the animal study did not raise any safety concerns with either the test Wedge XL Delivery Catheter or control Wedge Microcatheter. The devices are deemed equivalent.
{8}------------------------------------------------
# Conclusion
MicroVention concludes based upon the comparison of the device intended use, indications for use, operating principles, technological characteristics, and the review of the performance bench testing, animal study assessments, sterility, and biocompatibility evaluation that the Wedge XL Delivery Catheter is substantially equivalent to the predicate Wedge Microcatheter and performs as intended. Any differences between the subject device and the predicate device do not raise new or different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.