The SOFIA 88 Catheter is indicated for general intravascular use, including the neuro and peripheral vasculature. The SOFIA 88 Catheter can be used to facilitate introduction of diagnostic agents or therapeutic devices. The SOFIA 88 Catheter is not intended for use in coronary arteries.
Device Story
SOFIA 88 Catheter is a non-tapered, single-lumen, flexible catheter with coil and braid reinforcement; features 60 cm distal-shaft hydrophilic coating for navigation; includes radiopaque marker for fluoroscopic visualization. Used by clinicians in neuro/peripheral vascular procedures to facilitate delivery of diagnostic/therapeutic agents. Device is inserted into vasculature; clinician monitors position via fluoroscopy. Provides a conduit for interventional devices; benefits patient by enabling minimally invasive access to target vascular sites. Single-use, sterile, EtO-sterilized.
Clinical Evidence
No human clinical data. Bench testing included dimensional verification, lubricity/durability, simulated use in tortuous models, burst pressure, leakage, tensile strength, flexural fatigue, particulate testing, kink resistance, torque strength, radiodetectability, and corrosion resistance. GLP-compliant porcine animal study (n=not specified) compared tracking and tip stability to predicate; results showed equivalent performance with no adverse histological findings (hemorrhage, necrosis, edema). Biocompatibility testing (cytotoxicity, irritation, sensitization, systemic toxicity, pyrogenicity, hemocompatibility) confirmed safety.
Indicated for general intravascular use in neuro and peripheral vasculature to facilitate introduction of diagnostic agents or therapeutic devices. Not for use in coronary arteries.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
September 12, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
MicroVention, Inc. Alick Tan, Ph.D. Senior Regulatory Affairs Specialist 35 Enterprise Aliso Viejo, California 92656
### Re: K214024
Trade/Device Name: SOFIA 88 Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY, QJP Dated: August 12, 2022 Received: August 12, 2022
### Dear Alick Tan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K214024
Device Name SOFIA 88 Catheter
Indications for Use (Describe)
The SOFIA 88 Catheter is indicated for general intravascular use, including the neuro and peripheral vasculature. The SOFIA 88 Catheter can be used to facilitate introduction of diagnostic agents or therapeutic devices. The SOFIA 88 Catheter is not intended for use in coronary arteries.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span style="padding-right: 5px;"> <input checked="true" type="checkbox"/> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span style="padding-right: 5px;"> <input type="checkbox"/> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for MicroVention TERUMO. The logo features a stylized, abstract graphic on the left, resembling a series of concentric arcs in shades of blue and gray. To the right of the graphic is the word "MicroVention" in a blue sans-serif font, with "TERUMO" in a smaller, bolder blue font directly underneath.
### K214024 510(k) Summary
| 510(k) Owner | MicroVention, Inc.<br>35 Enterprise<br>Aliso Viejo, CA 92656<br>Establishment Registration No: 3013556777 |
|--------------------------|------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Alick Tan, PhD<br>Senior Regulatory Affairs Specialist<br>Telephone: 949-615-2603<br>Email: alick.tan@microvention.com |
| Date summary<br>Prepared | September 12, 2022 |
| Trade Name | SOFIA 88 Catheter |
| Common Name | Catheter, Percutaneous, Neurovasculature |
| Classification | Class II, DQY, QJP |
| Regulation | 21 CFR 870.1250 |
| Predicate Device | Sofia 6F PLUS/Distal Access Catheters (K150366) |
| Reference Device | TracStar™ Large Distal Platform; ZOOM™ 88 Large Distal<br>Platform; ZOOM™ 88-T Large Distal Platform (K203764) |
### Device Description
The SOFIA 88 Catheter is a non-tapered, single-lumen, flexible catheter equipped with coil and braid reinforcement. The distal seqment is designed to facilitate vessel selection with 60 cm of distal-shaft hydrophilic coating for navigation through the vasculature. The radiopaque marker is located at the distal end of the catheter for visualization under fluoroscopy.
#### Indications for Use
The SOFIA 88 Catheter is indicated for general intravascular use, including the neuro and peripheral vasculature. The SOFIA 88 Catheter can be used to facilitate introduction of diagnostic agents or therapeutic devices. The SOFIA 88 Catheter is not intended for use in coronary arteries.
#### Comparison of Indications for Use and Technological Characteristics
Substantial equivalence of the SOFIA 88 Catheter to the predicate Sofia 6F PLUS/Distal Access Catheters (SOFIA 6F DAC) and to the reference device TracStar Large Distal Platform; ZOOM 88 Large Distal Platform; ZOOM 88-T Large Distal Platform (ZOOM 88) was established through an evaluation of the intended use, indications for use, principles of operation, device design, materials of construction, and an assessment of usability, safety, and effectiveness via bench and animal studies.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for MicroVention TERUMO. The logo features a stylized, abstract graphic to the left, resembling a circular or orbital design with varying shades of gray and blue. To the right of the graphic, the word "MicroVention" is written in a combination of blue and gray lettering, with "Micro" in blue and "Vention" in gray. Below "MicroVention", the word "TERUMO" is written in blue, in a slightly smaller and bolder font.
The data presented in this submission demonstrates the technological similarity and equivalency of the SOFIA 88 Catheter compared with the predicate device SOFIA 6F DAC (K150366) and the reference device ZOOM 88 (K203764).
The devices:
- . have the same intended use,
- . use the same principle of operation,
- incorporate the same basic design,
- use similar construction and material, and ●
- are ethylene oxide (EtO)-sterilized and packaged using the same processes. ●
### Device Comparison Table
| | ZOOM 88 (K203764) | SOFIA 6F DAC<br>(K150366) | SOFIA 88 Catheter |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Reference Device | Predicate Device | Subject Device |
| Indications for Use | The ZOOM 88 and<br>ZOOM 88-T Large<br>Distal Platform are<br>indicated for the<br>introduction of<br>interventional devices<br>into the peripheral,<br>coronary, and neuro<br>vasculature. | The SOFIA<br>PLUS/Distal Access<br>Catheters are<br>indicated for general<br>intravascular use,<br>including the neuro<br>and peripheral<br>vasculature. It can be<br>used to facilitate<br>introduction of<br>diagnostic and<br>therapeutic agents. It<br>is not intended for use<br>in coronary arteries. | The SOFIA 88<br>Catheter is indicated<br>for general<br>intravascular use,<br>including the neuro<br>and peripheral<br>vasculature. The<br>SOFIA 88 Catheter<br>can be used to<br>facilitate introduction<br>of diagnostic agents or<br>therapeutic devices.<br>The SOFIA 88<br>Catheter is not<br>intended for use in<br>coronary arteries. |
| Device<br>Classification | Class II<br>QJP, DQY<br>21 CFR 870.1250 | Class II<br>DQY, DQO<br>21 CFR 870.1250<br>21 CFR 870.1200 | Class II<br>DQY, QJP<br>21 CFR 870.1250 |
| ZOOM 88 (K203764) | SOFIA 6F DAC<br>(K150366) | SOFIA 88 Catheter | |
| Reference Device | Predicate Device | Subject Device | |
| Catheter Body | Commonly used<br>medical grade plastics<br>& metals with<br>hydrophilic coating | Outer layer of<br>polyurethane<br>elastomer, polyether<br>block amide, and<br>polyamide; inner layer<br>of metal braid/coil,<br>PTFE, and polyolefin<br>elastomer | Outer layer of<br>polyurethane<br>elastomer (different<br>polyurethane<br>formulation), polyether<br>block amide, and<br>polyamide; inner layer<br>of metal braid/coil,<br>PTFE, and polyolefin<br>elastomer |
| Marker | - | Platinum/Iridium | Same as SOFIA 6F<br>DAC (K150366) |
| Hub | - | Nylon | Same as SOFIA 6F<br>DAC (K150366) |
| Strain Relief | - | Polyurethane | Same as SOFIA 6F<br>DAC (K150366) |
| Introducer | None | Pebax® | Fluorinated ethylene<br>propylene |
| Shaping Mandrel | None | Stainless steel | None |
| Catheter size | 8F | 6F | 8F, Same as ZOOM<br>88 (K203764) |
| ID | 0.088 in (2.2 mm) | 0.070 in (1.8 mm) | 0.0875 in (2.22 mm)<br>min |
| OD | 0.110 in (2.79 mm)<br>max | 0.0825 in (2.10 mm) | Proximal: 0.108 in<br>(2.74 mm) max |
| | | | Distal: 0.102 in (2.59<br>mm) max |
| Effective Length | 80–110 cm | 115, 125, 131, 135 cm | 115 cm |
| Coating | Hydrophilic coating | Hydrophilic coating | Same as SOFIA 6F<br>DAC (K150366) |
| Tip Configuration | Beveled distal edge,<br>soft, flexible,<br>atraumatic tip | Straight | Same as SOFIA 6F<br>DAC (K150366) |
| Guidewire<br>Compatibility | ≤ 0.038 in | 0.035 in | ≤ 0.038 in |
| Accessories | Rotating hemostasis<br>valve (RHV) | Introducer sheath and<br>shaping mandrel | Introducer sheath |
| | ZOOM 88 (K203764) | SOFIA 6F DAC<br>(K150366) | SOFIA 88 Catheter |
| | Reference Device | Predicate Device | Subject Device |
| Condition<br>Supplied | Sterile and single use | Sterile and single use | Same |
| Sterilization<br>Method | EtO | EtO | Same |
| Packaging<br>Configuration | The catheters are<br>placed in a protective<br>polyethylene tube,<br>mounted with<br>accessory RHV onto a<br>polyethylene<br>packaging card,<br>placed into a pouch,<br>sealed and labeled.<br>The sealed pouch and<br>IFU are placed in a<br>labeled shelf carton<br>box. | Catheter placed into a<br>HDPE dispenser tube.<br>Dispenser tube,<br>introducer and<br>shaping mandrel<br>placed on a<br>polyethylene<br>packaging card that is<br>inserted into a Tyvek<br>pouch. Pouch and IFU<br>placed in bleached<br>sulfate carton box. | Catheter placed into a<br>HDPE dispenser tube.<br>Dispenser tube and<br>introducer placed on a<br>polyethylene<br>packaging card that is<br>inserted into a Tyvek<br>pouch. Pouch and IFU<br>placed in bleached<br>sulfate carton box. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for MicroVention, a company that specializes in neurovascular products. The logo features a stylized graphic of interconnected blue and gray circles on the left. To the right of the graphic is the company name "MicroVention" in blue, with the registered trademark symbol. Below the company name is the word "TERUMO" in blue, which is the parent company of MicroVention.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for MicroVention TERUMO. The logo features a circular graphic on the left side, composed of overlapping curved lines in shades of gray and blue, creating a sense of motion or rotation. To the right of the graphic, the word "MicroVention" is written in a blue sans-serif font, with a small registered trademark symbol next to it. Below "MicroVention", the word "TERUMO" is written in a larger, bolder blue sans-serif font.
## Performance Testing
Performance testing was conducted to demonstrate the substantial equivalence of the SOFIA 88 Catheter to the predicate device.
The following testing data were provided to evaluate the performance of the SOFIA 88 Catheter.
| Test | Test Method Summary | Results |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Dimensional<br>Verification | Dimensional specifications were verified by<br>measuring the catheter length, proximal and distal<br>outer diameters, and distal inner diameter. | Pass |
| Coating Lubricity<br>and Durability | The hydrophilic coating was evaluated for frictional<br>force and durability. | Pass |
| Simulated Use | Performance was evaluated under simulated use in<br>a tortuous anatomical model to assess preparation,<br>introduction, tracking, and support of the device. | Pass |
| Dynamic and Static<br>Burst Pressure | Device was subjected to rated burst pressures to<br>test catheter integrity. | Pass |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for MicroVention, a company that is part of Terumo. The logo features a stylized graphic of three curved lines that resemble a swirl or a vortex. To the right of the graphic is the word "MicroVention" in a blue sans-serif font. Below "MicroVention" is the word "TERUMO" in a smaller, bolder, blue sans-serif font.
| Test | Test Method Summary | Results |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Air Leakage | Device was tested for air leakage requirements. | Pass |
| Liquid Leakage | Device was tested for liquid leakage requirements. | Pass |
| Tensile Strength | Device was tensile tested to failure, and recorded<br>the force at breakage. | Pass |
| Flexural Fatigue | Device was subjected to flexural fatigue from<br>repeated bending during simulated use testing and<br>flexural fatigue from repeated hoop stress on the<br>catheter from pressure testing. | Pass |
| Particulate Testing | The number and size of particulates generated<br>during simulated use in a tortuous anatomical model<br>were measured and results were compared with the<br>predicate device. | Pass |
| Kink Resistance | Device was evaluated for kink resistance after<br>simulated use testing. | Pass |
| Torque Strength | Device was evaluated for torque strength by<br>measuring the number of catheter rotations until<br>failure after tracking through a tortuous anatomical<br>model. | Pass |
| Radio Detectability | Device visibility was evaluated under fluoroscopy. | Pass |
| Corrosion Resistance | Device corrosion was evaluated after immersion in<br>saline. | Pass |
| Small-bore Connector | The luer connector was evaluated for dimensional<br>and performance requirements per reference<br>standard. | Pass |
| Stiffness Profile | The catheter stiffness profile along proximal and<br>distal sections was compared to the reference<br>device. | Pass |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for MicroVention TERUMO. The logo features a stylized, abstract graphic to the left, resembling interconnected circles or orbits in shades of gray and blue. To the right of the graphic is the word "MicroVention" in a blue sans-serif font, with "TERUMO" in a bolder, blue font positioned directly below it.
# Biocompatibility Results
| Test | Test Method Summary | | Results |
|----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|
| | Extract(s) & Test<br>Systems | Extract Conditions | |
| Cytotoxicity<br>(ISO MEM Elution<br>Test) | L-929 mouse<br>fibroblast cells<br>prepared using MEM<br>Maintenance Growth<br>Media with 5% FBS | 6.0 cm²/mL (exposed<br>surface area to<br>extraction medium<br>volume ratio),<br>extracted at 37 °C for<br>24 hrs | Non-cytotoxic |
| Irritation Reactivity<br>(ISO Intracutaneous<br>Reactivity Test) | Normal saline and<br>sesame seed oil<br>(SSO) (tested<br>separately)<br>New Zealand White<br>Rabbits | 6.0 cm²/mL (exposed<br>surface area to<br>extraction medium<br>volume ratio),<br>extracted at 50 °C for<br>72 hrs | Non-irritant |
| Maximization<br>(ISO Guinea Pig<br>Maximization Test) | Normal saline and<br>sesame seed oil<br>(SSO) (tested<br>separately)<br>Guinea Pig | 6.0 cm²/mL (exposed<br>surface area to<br>extraction medium<br>volume ratio),<br>extracted at 50 °C for<br>72 hrs | Non-sensitizing |
| Systemic Toxicity<br>(ISO Acute<br>Systemic Toxicity<br>Test) | Normal saline and<br>sesame seed oil<br>(SSO) (tested<br>separately)<br>Albino outbred strain<br>(ND4) mice | 6.0 cm2/mL (exposed<br>surface area to<br>extraction medium<br>volume ratio),<br>extracted at 50 °C for<br>72 hrs | Non-acute<br>systemically toxic |
| Pyrogenicity<br>(ISO/USP Material<br>Mediated<br>Pyrogencity Test) | Normal saline<br>New Zealand White<br>Rabbits | 6.0 cm²/mL (exposed<br>surface area to<br>extraction medium<br>volume ratio),<br>extracted at 50 °C for<br>72 hrs | Non-pyrogenic |
| Test | Test Method Summary | | Results |
| | Extract(s) & Test Systems | Extract Conditions | |
| Hemocompatibility<br>In-Vitro Blood Loop<br>Assay<br>(ISO In-vitro Blood<br>Loop Assay with<br>Comparison Article) | A loop system<br>circulated with freshly<br>drawn sheep blood | The test article is<br>exposed to the<br>circulating blood that<br>fills to the capacity of<br>the loop at 37 °C for<br>4 hrs | Thromboresistant |
| Hemocompatibility<br>Hemolysis Assay<br>(ISO ASTM<br>Hemolysis Assay) | PBS – Phosphate<br>Buffered Saline<br><br>Blood - Three New<br>Zealand White<br>Rabbits | 6.0 cm²/mL (exposed<br>surface area to<br>extraction medium<br>volume ratio),<br>extracted at 50 °C for<br>72 hrs | Non-hemolytic |
| Hemocompatibility<br>Complement<br>Activation Assay<br>(ISO<br>Hemocompatibility:<br>Complement<br>Activation Assay<br>(SC5b-9)) | Normal Human<br>Serum (NHS) | 6.0 cm²/mL (exposed<br>surface area to<br>extraction medium<br>volume ratio),<br>extracted at 37 °C for<br>60 min | Non-activator of<br>complement system |
| Hemocompatibility<br>Partial<br>Thromboplastin<br>(ISO<br>Hemocompatibility:<br>Partial<br>Thromboplastin<br>Time (PTT) Assay) | Human Plasma<br>Freshly Drawn<br>Human Plasma | 6.0 cm²/mL (exposed<br>surface area to<br>extraction medium<br>volume ratio),<br>extracted at 37 °C for<br>15 min | No effect on the PTT |
| Test | Test Method Summary | | Results |
| | Extract(s) & Test Systems | Extract Conditions | |
| Hemocompatibility<br>Heparinized Blood<br>Platelet and<br>Leukocyte | Human Blood | 12 cm²/mL (exposed surface area to extraction medium volume ratio), extracted at 37 °C for 60 min | Pass |
| | Freshly Drawn<br>Human Blood | | |
| (Hemocompatibility:<br>Heparinized Blood<br>Platelet and<br>Leukocyte Count<br>Assay) | | | |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the logo for MicroVention, a Terumo company. The logo features a stylized, abstract graphic to the left of the text "MicroVention". Below "MicroVention" is the word "TERUMO", indicating that MicroVention is associated with Terumo Corporation. The text is in a blue font.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for MicroVention TERUMO. The word "MicroVention" is written in a combination of blue and gray letters, with "Micro" in blue and "Vention" in gray. To the left of the word is a circular graphic made of curved lines in shades of gray and blue. Below "MicroVention" is the word "TERUMO" in blue, in a smaller, bolder font.
# Animal Study
Acute animal testing was conducted in accordance with FDA Good Laboratory Practice (GLP) Regulation (21 CFR Part 58) comparing the SOFIA 88 Catheter to the SOFIA 6F DAC. The testing was intended to assess preclinical safety and efficacy for catheter tracking and tip stability in a porcine model. The porcine model was chosen since the vessel sizes of the pig model allow for insertion and navigation of standard-sized devices used in humans; porcine vessel diameters are comparable with human vasculature. The tracking results demonstrated that the SOFIA 88 Catheter and SOFIA 6F DAC devices performed equally. Catheter tip stability was found to be comparable for both devices. No dissection/perforation was noted for the test-article- and controltreated vessels. Similar degrees of vasospasm were noted in the vessels instrumented with both devices. During postmortem examination there were no remarkable gross findings for any of the vessel samples for both the test and control devices. There was no hemorrhage, necrosis, or edema in test article-treated and control article-treated arterial sections. Overall, the histologic findings were consistent with routine catheterization procedures, which are commonly observed after interventional procedures in porcine safety models. The results of the present study did not raise any safety issues with either the test SOFIA 88 Catheter or control SOFIA 6F DAC. The devices are deemed equivalent.
## Conclusion
MicroVention concludes through a review of the benchtop testing, non-clinical animal study assessments, the comparison of the device classification, indications for use, operating principles, technological characteristics, sterility, and biocompatibility testing that the SOFIA 88 Catheter is substantially equivalent to the predicate SOFIA 6F DAC. Any differences between the subject device and the predicate device do not raise different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.