The SOFIA EX Intracranial Support Catheter is indicated for general intravascular use, including the neuro and peripheral vasculature. The SOFIA EX Intracranial Support Catheter can be used to facilitate introduction of diagnostic agents or therapeutic devices. The SOFIA EX Intracranial Support Catheter is not intended for use in coronary arteries.
Device Story
Single-lumen, flexible intracranial support catheter; coil and braid reinforced; distal shaft hydrophilic coating for navigation; radiopaque platinum/iridium marker at distal end for fluoroscopic visualization. Used in neuro and peripheral vasculature; facilitates introduction of diagnostic agents or therapeutic devices. Operated by physicians in clinical settings. Device provides a pathway for interventional procedures; supports navigation through tortuous anatomy. Benefits patient by enabling access to target vasculature for diagnostic or therapeutic intervention.
Clinical Evidence
Bench testing only. Evaluated simulated use in tortuous anatomical models, kink resistance, and particulate generation. Results met all acceptance criteria and were comparable to the predicate device. No animal or clinical studies conducted.
Indicated for general intravascular use in neuro and peripheral vasculature to facilitate introduction of diagnostic agents or therapeutic devices. Not for use in coronary arteries.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
September 14, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
MicroVention, Inc. Miranda Beach Regulatory Affairs Specialist 35 Enterprise Aliso Viejo, California 92656
Re: K230775
Trade/Device Name: SOFIA EX Intracranial Support Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP, DQY Dated: August 15, 2023 Received: August 16, 2023
# Dear Miranda Beach:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Naira Muradyan -S
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230775
Device Name SOFIA EX Intracranial Support Catheter
#### Indications for Use (Describe)
The SOFIA EX Intracranial Support Catheter is indicated for general intravascular use, including the neuro and peripheral vasculature. The SOFIA EX Intracranial Support Catheter can be used to facilitate introduction of diagnostic agents or therapeutic devices. The SOFIA EX Intracranial Support Catheter is not intended for use in coronary arteries.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for MicroVention TERUMO. The logo features a stylized graphic of overlapping curved lines in shades of gray and blue on the left. To the right of the graphic is the text "MicroVention" in a blue sans-serif font, with "TERUMO" in a smaller, bolder blue font underneath.
# 510(k) Summary K230775
| 510(k) Owner | MicroVention, Inc.<br>35 Enterprise<br>Aliso Viejo, CA 92656<br>Establishment Registration No: 3013556777 |
|---------------------|-----------------------------------------------------------------------------------------------------------|
| Contact Person | Miranda Beach<br>Regulatory Affairs Specialist<br>Email: Miranda.Beach@microvention.com |
| Date of Preparation | September 12, 2023 |
| Trade Name | SOFIATM EX Intracranial Support Catheter |
| Common Name | Catheter, Percutaneous, Neurovasculature |
| Classification | Class II, QJP, DQY |
| Regulation | 21 CFR 870.1250 |
| Predicate Device | SOFIA EX Intracranial Support Catheter (K182602) |
| Reference Device | Benchmark Intracranial Access System (K212838) |
# Device Description
The SOFIA™ EX Intracranial Support Catheter is a single-lumen, flexible catheter equipped with coil and braid reinforcement. The distal segment is designed to facilitate vessel selection with 55-65cm of distal shaft hydrophilic coating for navigation through the vasculature. The radiopaque marker is located at the distal end of the catheter for visualization under fluoroscopy. The catheter is placed in a dispenser tube (HDPE) and is placed on a packaging card (polyethylene) that is provided in a sterile barrier Tyvek pouch and placed in a carton box.
# Indications for Use
The SOFIA EX Intracranial Support Catheter is indicated for general intravascular use, including the neuro and peripheral vasculature. The SOFIA EX Intracranial Support Catheter can be used to facilitate introduction of diagnostic agents or therapeutic devices. The SOFIA EX Intracranial Support Catheter is not intended for use in coronary arteries.
#### Comparison of Indications for Use and Technological Characteristics
The subject device is identical in its indication for use, technology, principle of operation, materials, sterilization method, and performance to the predicate device SOFIA™ EX Intracranial Support Catheter (K182602). The only difference between the subject and predicate devices are additional instructions added to the instructions for use (IFU) for the subject device
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for MicroVention TERUMO. The logo consists of a circular graphic on the left, with the text "MicroVention" in blue to the right of the graphic. Below "MicroVention" is the word "TERUMO", also in blue. The circular graphic is made up of curved lines in shades of gray and blue.
related to radial access use. A comparison of the technological characteristics of the subject device and the predicate device is summarized in Table 1 below.
| | SOFIA™ EX Intracranial<br>Support Catheter (K182602) | SOFIA™ EX Intracranial<br>Support Catheter |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Predicate Device | Subject Device |
| Indications for Use | The SOFIA EX Catheter is<br>indicated for general<br>intravascular use, including the<br>neuro and peripheral vasculature.<br>The SOFIA EX Catheter can be<br>used to facilitate introduction of<br>diagnostic agents or therapeutic<br>devices. The SOFIA EX Catheter<br>is not intended for use in<br>coronary arteries. | The SOFIA EX Intracranial<br>Support Catheter is indicated for<br>general intravascular use,<br>including the neuro and<br>peripheral vasculature. The<br>SOFIA EX Intracranial Support<br>Catheter can be used to facilitate<br>introduction of diagnostic agents<br>or therapeutic devices. The<br>SOFIA EX Intracranial Support<br>Catheter is not intended for use in<br>coronary arteries. |
| Device Classification | Class II<br>DQY<br>21 CFR 870.1250 | Class II<br>QJP, DQY<br>21 CFR 870.1250 |
| Catheter Body<br>Materials | Outer layer of polyolefin<br>elastomer, polyurethane<br>elastomer (Pellethane), polyether<br>block amide (Pebax) and<br>polyamide (Grilamid); inner layer<br>of stainless-steel braid, Nitinol<br>coil, and PTFE. | Same |
| Marker | Platinum/Iridium | Same |
| Hub | Nylon | Same |
| Strain Relief | Polyurethane | Same |
| Introducer | Pebax | Same |
| Catheter Size | 5F | Same |
| Inner Diameter (ID) | 0.058 in. | Same |
| | SOFIA™ EX Intracranial<br>Support Catheter (K182602) | SOFIA™ EX Intracranial<br>Support Catheter |
| | Predicate Device | Subject Device |
| Outer Diameter (OD) | 0.071 in. | Same |
| Effective Length | 105 cm, 115 cm. | 115 cm. |
| Coating | Hydrophilic Coating | Same |
| Tip Configuration | Straight | Same |
| Guidewire<br>Compatibility | $\leq$ 0.038 in. OD | Same |
| Accessories | Introducer Sheath | Same |
| Method of Supply | Sterile and Single Use | Same |
| Sterilization Method | Ethylene Oxide | Same |
| Packaging<br>Configuration | Catheter placed into a HDPE<br>dispenser tube. Dispenser tube<br>and introducer placed on a<br>polyethylene packaging card that<br>is inserted into a Tyvek pouch.<br>Pouch and IFU placed in<br>bleached sulfate carton box. | Same |
# Table 1: Technological Characteristics Comparison
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for MicroVention TERUMO. The logo features a circular graphic on the left side, followed by the text "MicroVention" in a blue, sans-serif font. Below "MicroVention" is the word "TERUMO" in a smaller, bold, blue font. The overall design is clean and professional.
# Performance Testing
The following bench testing data were provided to evaluate the performance and support the substantial equivalence of the SOFIA™ EX Intracranial Support Catheter.
| Test | Test Method Summary | Results |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|
| Simulated Use and<br>Physician Simulated Use | The simulated use of the<br>device is evaluated in a<br>tortuous anatomical benchtop<br>model to assess tracking and<br>hydrophilic coating lubricity<br>of the device.<br><br>Physician simulated use is<br>conducted with subject the | Device met acceptance<br>criteria. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for MicroVention TERUMO. The MicroVention part of the logo is in blue, with the word TERUMO in a smaller, bolder font below it. To the left of the text is a graphic of three overlapping circles in shades of gray and blue.
| Kink Resistance | Device is subjected to<br>bending experienced in<br>tortuous anatomy. | Device met acceptance<br>criteria. |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Particulate Testing | The number and size of<br>particulates generated during<br>simulated use in a tortuous<br>anatomical model were<br>measured and results were<br>compared with the predicate<br>device. | Number of particulates<br>generated are comparable to<br>the predicate device. |
# Biocompatibility
The subject device, SOFIA™ EX Intracranial Support Catheter is categorized as a limited exposure (≤ 24 hours), externally communicating device with circulating blood contact in accordance with ISO 10993-1. The design and manufacturing of the subject device, SOFIA™ EX Intracranial Support Catheter use identical materials, processing, and sterilization method as the predicate device, SOFIA™ EX Intracranial Support Catheter (K182602) for which MicroVention has already successfully conducted biocompatibility testing per ISO 10993-1. Therefore, no additional biocompatibility testing is required.
# Performance Data - Animal, Clinical
No animal or clinical studies were conducted because bench testing was determined sufficient for verification and validation purposes.
# Conclusions
The subject SOFIA™ EX Intracranial Support Catheter is substantially equivalent to the predicate device SOFIA™ EX Intracranial Support Catheter. The subject device has the same indications for use as the predicate device. The 510(k) Summary demonstrates that the subject device is substantially equivalent to the predicate device regarding operating principle, design, fundamental technology, and the device performs as intended. The changes to the instructions for use do not raise new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.