K222786 · Q'Apel Medical, Inc. · NRY · Aug 25, 2023 · Cardiovascular
Device Facts
Record ID
K222786
Device Name
072 Aspiration System
Applicant
Q'Apel Medical, Inc.
Product Code
NRY · Cardiovascular
Decision Date
Aug 25, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
As part of the 072 Aspiration System, the 072 Aspiration Catheter with a compatible suction pump is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment. As part of the 072 Aspiration System, the Aspiration Tubing is intended to connect the 072 Aspiration Catheter to a compatible suction pump.
Device Story
072 Aspiration System comprises a single-lumen, variable-stiffness aspiration catheter, aspiration tubing, and accessories (delivery tool, rotating hemostasis valve, flow switch). Used in neurointerventional procedures to revascularize intracranial vessels by removing thrombi. Catheter features hydrophilic coating, radiopaque markers, and luer hub. Operated by physicians in clinical settings; connects to external compatible suction pump via aspiration tubing. Flow switch allows user to control fluid flow/vacuum. Device facilitates clot retrieval; benefits patient by restoring blood flow in acute ischemic stroke. Bench and animal testing confirm mechanical integrity, aspiration performance, and biocompatibility.
Clinical Evidence
No clinical study conducted. Substantial equivalence supported by bench testing (mechanical integrity, flow rates, simulated use in tortuous anatomical models, biocompatibility per ISO 10993) and a GLP-compliant porcine animal study comparing clot aspiration and wedge assessment against predicate/reference devices.
Indicated for patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (internal carotid, middle cerebral M1/M2, basilar, vertebral arteries) within 8 hours of symptom onset. Candidates include patients ineligible for or who failed IV t-PA therapy.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Penumbra System® [JET™ 7 Reperfusion Catheter with MAX Delivery Device (JET™ 7MAX)] (K191946)
Reference Devices
CERENOVUS Large Bore Catheter; CERENOVUS Aspiration Tubing Set (K193380)
{0}------------------------------------------------
August 25, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
Q'Apel Medical, Inc. Kim Ky Manager, Regulatory Affairs 4245 Technology Drive Fremont, California 94538
#### Re: K222786
Trade/Device Name: 072 Aspiration System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: August 17, 2023 Received: August 17, 2023
#### Dear Kim Ky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
# Naira Muradyan -S
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222786
Device Name 072 Aspiration System
#### Indications for Use (Describe)
#### 072 Aspiration Catheter:
As part of the 072 Aspiration System, the 072 Aspiration Catheter with a compatible suction pump is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA therapy are candidates for treatment.
Aspiration Tubing:
As part of the 072 Aspiration System, the Aspiration Tubing is intended to connect the 072 Aspiration Catheter to a compatible suction pump.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--------------------------|
| <input type="checkbox"/> | <input type="checkbox"/> |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "Q'apel" in blue font. The font is sans-serif and the letters are connected. The apostrophe is placed between the "Q" and the "a". The word is centered and takes up most of the image.
#### 510(k) SUMMARY As required by 21 CFR 807.92
| Applicant:<br>Submitter's Name:<br>Address: | Q'Apel Medical, Inc.<br>4245 Technology Drive<br>Fremont, CA 94538 |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Telephone: | 510-738-6255 |
| Contact Person:<br>Title:<br>Telephone: | Kim Ky<br>Manager, Regulatory Affairs<br>510-828-4757 |
| Date Prepared: | August 25, 2023 |
| Device Name /Common name: | 072 Aspiration System |
| Classification: | Class II |
| Product Code(s): | NRY |
| Regulation Number(s): | 21 CFR 870.1250 |
| Classification Name: | Percutaneous, Catheter |
| Predicate Device: | Penumbra System® [JETTM 7 Reperfusion Catheter with MAX Delivery<br>Device (JETTM 7MAX)] (K191946) |
| Reference Devices: | CERENOVUS Large Bore Catheter; CERENOVUS Aspiration Tubing Set<br>(K193380)<br>Penumbra System® (Reperfusion Catheter REDTM 72) (K211654) |
#### Indications for Use: 072 Aspiration Catheter
As part of the 072 Aspiration System, the 072 Aspiration Catheter with a compatible suction pump is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Q'apel. The logo is written in a sans-serif font and is a bright blue color. The logo is simple and modern.
# Aspiration Tubing
As part of the 072 Aspiration System, the Aspiration Tubing is intended to connect the 072 Aspiration Catheter to a compatible suction pump.
#### Device Description:
The 072 Aspiration Catheter is a single lumen, variable stiffness catheter. The catheter has a hydrophilic coating to reduce friction during use. The catheter includes radiopaque markers on the distal end for angiographic visualization and a Luer hub on the proximal end allowing attachments for flushing and aspiration. The Delivery Tool is an optional accessory for use with the 072 Aspiration Catheter and should be removed prior to aspiration. The 072 Aspiration Catheter, Delivery Tool. Rotating Hemostasis Valve, and Flow Switch are included in the package. The Aspiration Tubing is provided in a separate package.
For the aspiration source, the 072 Aspiration Catheter is used in conjunction with a compatible suction pump with prespecified performance parameters that is connected using the Aspiration Tubing.
#### Comparison of Technological characteristics with the Predicate Device:
O'Apel Medical, Inc. has demonstrated that the 072 Aspiration System is substantially equivalent to the predicate device, Penumbra System® [JET™ 7 Reperfusion Catheter with MAX Delivery Device (JET™ 7MAX)] cleared under K191946, based on the same or similar materials, similar design, and the same fundamental operating principles. A comparison of the subject device with the predicate and reference devices is summarized in the table below.
| Comparison of the Subject, Predicate, and Reference Devices | | | | | |
|-------------------------------------------------------------|-----------------------------|-------------------------------|-------------------------------------------------------------------------------|--------------------------------------------------------|------------------------------------------------|
| Feature | Subject Device<br>(K222786) | Predicate Device<br>(K191946) | Reference Device<br>1 (K193380) | Reference Device 2<br>(K211654) | Rationale for<br>difference (if<br>applicable) |
| Device Name | 072 Aspiration<br>System | Penumbra System<br>(JET 7MAX) | CERENOVUS<br>Large Bore<br>Catheter;<br>CERENOVUS<br>Aspiration Tubing<br>Set | Penumbra<br>System<br>(Reperfusion<br>Catheter RED 72) | N/A |
| 510(k) Number | K222786 | K191946 | K193380 | K211654 | N/A |
| Classification | Class II | Class II | Class II | Class II | Same |
| Product Code | NRY | NRY | NRY | NRY | Same |
| Classification<br>Name | 21CFR 870.1250 | 21CFR 870.1250 | 21 CFR 870.1250 | 21 CFR 870.1250 | Same |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Q'apel. The logo is written in a sans-serif font and is blue. The Q is stylized with a curved tail. The rest of the letters are lowercase.
| Indications for Use | 072 Aspiration Catheter | Penumbra Reperfusion Catheters and Separators | The CERENOVUS Large Bore Catheter, with the CERENOVUS Aspiration Tubing Set and NOUVAG Vacuson 60 aspiration pump (or equivalent aspiration pump) | Penumbra Reperfusion Catheters and Separators | Similar, the differences do not raise new questions regarding safety and effectiveness. |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| | As part of the 072 Aspiration System, the 072 Aspiration Catheter with a compatible suction pump is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment. | As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. | is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA are candidates for treatment. The CERENOVUS Aspiration Tubing Set is intended to connect the CERENOVUS Large Bore Catheter to the canister of the NOUVAG Vacuson 60 Aspiration Pump | As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. | |
| | Aspiration Tubing As part of the 072 Aspiration System, the Aspiration Tubing is intended to connect the 072 Aspiration Catheter to a compatible suction pump. | Penumbra 3D Revascularization Device As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA | | Penumbra 3D Revascularization Device As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom | |
| | | | | | |
| | therapy are<br>candidates for<br>treatment.<br><br>Penumbra Aspiration<br>Tubing<br>As part of the<br>Penumbra System,<br>the Penumbra Sterile<br>Aspiration Tubing is<br>indicated to connect<br>the Penumbra<br>Reperfusion<br>Catheters to the<br>Penumbra Aspiration<br>Pump.<br>Penumbra Aspiration<br>Pump<br>The Penumbra<br>Aspiration Pump is<br>indicated as a<br>vacuum source for<br>Penumbra Aspiration<br>Systems. | (or equivalent<br>vacuum pump)<br>and to allow the<br>user to control<br>the fluid flow. | onset. Patients who<br>are ineligible for<br>intravenous tissue<br>plasminogen<br>activator (IV t-PA) or<br>who fail IV t-PA<br>therapy are<br>candidates for<br>treatment.<br><br>Penumbra<br>Aspiration Tubing<br>As part of the<br>Penumbra System,<br>the Penumbra<br>Sterile Aspiration<br>Tubing is indicated<br>to connect the<br>Penumbra<br>Reperfusion<br>Catheters to the<br>Penumbra<br>Aspiration Pump.<br>Penumbra<br>Aspiration Pump<br>The Penumbra<br>Aspiration Pump is<br>indicated as a<br>vacuum source for<br>Penumbra<br>Aspiration Systems. | | |
| | | | | | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "Q'apel" in blue font. The font is sans-serif and the letters are connected. The apostrophe is slightly raised above the "Q". The background is white.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Q'apel. The logo is written in a sans-serif font and is blue. The Q has a unique design, with a curved tail that extends upwards. The rest of the letters are standard sans-serif letters.
| Comparison of the Subject, Predicate, and Reference Devices | | | | | |
|-------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Feature | Subject Device<br>(K222786) | Predicate Device<br>(K191946) | Reference Device<br>1 (K193380) | Reference Device 2<br>(K211654) | Rationale for<br>difference (if<br>applicable) |
| Aspiration<br>Catheter | 072 Aspiration<br>Catheter | Penumbra JET™ 7 | CERENOVUS<br>Large Bore<br>Catheter | Reperfusion<br>Catheter RED 72 | N/A |
| Length | 132 cm | 115, 120, 125, 127,<br>132 cm | 125-135 cm | 115, 120, 125, 127,<br>132 cm | Same |
| ID | 0.072" | 0.072" Min | 0.071" | 0.072" | Same as<br>K191946 and<br>K211654 |
| Distal OD | 0.0855" | 0.085" Max | 0.081" | 0.085" | Same as<br>K191946 and<br>K211654 |
| Proximal OD | 0.0855" | 0.085" Max | 0.0825" | 0.085" | Same as<br>K191946 and<br>K211654 |
| Catheter<br>Coating | Hydrophilic | Hydrophilic<br>(proprietary) | Hydrophilic | Hydrophilic<br>(proprietary) | Same |
| Tip<br>Configuration | Straight | Straight | Non-shapeable<br>tip | Straight | Same as<br>K191946 and<br>K211654 |
| Coating Length | 28cm +/- 4cm | 30 cm | 30 cm | 30 cm | Similar, the<br>differences do<br>not raise new<br>questions<br>regarding<br>safety and<br>effectiveness. |
| <b>Materials:</b> | | | | | |
| Hub | Polycarbonate | Grilamid (TR55-LX) | Polyamide (Nylon) | Grilamid (TR55-LX) | Similar, the<br>differences do<br>not raise new<br>questions<br>regarding<br>safety and<br>effectiveness. |
| Liner | PTFE Liner | PTFE Liner | PTFE Liner | PTFE Liner | Same |
| Catheter<br>Shaft | | | | | |
| Extrusions | Nylon<br>Pebax<br>Polyurethane | Polyurethane<br>Polyether Block<br>Amide<br>Nylon 12 | Pebax<br>Urethane<br>Nylon | Polyurethane<br>Polyether Block<br>Amide<br>Nylon 12 | Similar, the<br>differences do<br>not raise new<br>questions of<br>safety and<br>effectiveness. |
| Marker Band | Metal Platinum<br>(90%) / Iridium<br>(10%), Tantalum | Platinum (90%) /Iridium<br>(10%) | Metal Platinum<br>(90%) / Iridium<br>(10%) | Platinum (90%)<br>/Iridium (10%) | Similar, the<br>differences do<br>not raise new<br>questions of<br>safety and<br>effectiveness. |
| Reinforced<br>Shaft | Stainless Steel (SS),<br>Coil | NiTi wire, SS wire | Stainless Steel,<br>Nitinol, braid | NiTi wire, SS wire | Similar, the<br>differences do<br>not raise new<br>questions of<br>safety and<br>effectiveness. |
| Strain Relief | Polyolefin | Polyolefin | Polyamide | Polyolefin | Same as<br>K191946 and<br>K211654 |
| Accessories | | | | | |
| Peelable<br>Sheath | N/A | PTFE | N/A | PTFE | N/A |
| Hemostasis<br>Valve | Silicone O-ring,<br>Polycarbonate,<br>EPDM | Polycarbonate,<br>silicone O-ring | Hemostasis<br>Valve with Side<br>Port Extension<br>Tubing | Polycarbonate,<br>silicone O-ring | Similar to<br>K191946,<br>differences do<br>not raise new<br>questions of<br>safety and<br>effectiveness. |
| Flow Switch | Silicone,<br>Polycarbonate,<br>Acetal, ABS | N/A | N/A | N/A | The difference<br>does not raise<br>new questions<br>of safety and<br>effectiveness. |
| Shaping<br>Mandrel | N/A | Stainless Steel | N/A | Stainless Steel | N/A |
| Delivery Tool | Materials:<br>Pebax, Tecoflex<br>93A, Tungsten 75%<br>Filler<br>Dimensions:<br>ID: 0.026 in<br>OD: 0.042 in<br>Extension Length:<br>10 cm | Materials:<br>Nylon 12,<br>Copolyester,<br>Polyolefin,<br>Polyurethane,<br>Polyether Block<br>Amide, PTFE,<br>Platinum/Tungsten,<br>Hydrophilic Coating<br>Dimensions:<br>ID: 0.018 in<br>OD: 0.071 in<br>Extension Length:<br>1.5 cm | N/A | N/A | Similar, the<br>differences do<br>not raise new<br>questions of<br>safety and<br>effectiveness. |
| Aspiration<br>Tubing | Dimensions:<br>ID: 0.110 in<br>Tubing Length: 95 in | Dimensions:<br>Tubing ID: 0.110 in ±<br>0.005 in<br>Tubing Length: 112.0<br>in ± 7.0 in | Dimensions:<br>ID: 0.110 in<br>Tubing Length:<br>112 in | Dimensions:<br>Tubing ID: 0.110 in ±<br>0.005 in<br>Tubing Length:<br>112.0 in ± 7.0 in | Similar, the<br>differences do<br>not raise new<br>questions of<br>safety and<br>effectiveness. |
| Flow Control<br>Mechanism | Flow Control Switch | Flow Control Switch | Flow Control<br>Switch | Flow Control Switch | Same as<br>K191946,<br>K193380,<br>K211654 |
| Packaging<br>Material | | | | | |
| | | | | | |
| Pouch | Tyvek<br>(polyethylene), Mylar<br>(Polyester) | Polyester /<br>Polyethylene/Tyvek | Polyethylene<br>Hoop and<br>Mounting Card,<br>Pouch, Carton | Polyester /<br>Polyethylene/Tyvek | Similar to<br>K191946,<br>the |
| Packaging<br>Tray (Kit<br>Configuration) | PETG (Polyethylene<br>terephthalate glycol) | Polyethylene<br>terephthalate,<br>Polystyrene | | Polyethylene<br>terephthalate,<br>Polystyrene | differences do<br>not raise new<br>questions of<br>safety and |
| Display Carton | SBS Paperboard, | SBS Paperboard | | SBS Paperboard | effectiveness. |
| Kit Packaging<br>Configuration | Tray/Retainer/Lid<br>/Accessory<br>Pouch/Pouch/Box | Tray/Retainer/Lid/<br>Aspiration<br>Tubing/Accessory<br>Pouch/Pouch/Box | | Tray/Retainer/Lid/<br>Aspiration<br>Tubing/Accessory<br>Pouch/Pouch/Box | |
| Shelf Life | 6 months | 36 months | 1 year | 36 months | The differences<br>do not raise<br>new questions<br>of safety and<br>effectiveness. |
| Sterilization<br>Method | Ethylene oxide | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | Same |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Q'apel. The logo is written in a sans-serif font and is blue. The Q has a tail that extends to the left, and the apostrophe is connected to the 'a'.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "Q'apel" in a sans-serif font. The word is written in blue. The apostrophe is placed between the "Q" and the "a". The letters are all lowercase except for the "Q".
{10}------------------------------------------------
# Performance Testing – Bench:
The necessary testing was identified based on design, risk analysis, and the intended use of the 072 Aspiration System to verify the device performs as intended and to demonstrate that it is substantially equivalent to the predicate device. The following performance data have been provided, supporting the substantial equivalence determination. All testing was conducted per Q'Apel Medical, Inc. Design Control procedures. The bench testing included the following tests:
| Test | Test Summary | Result |
|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|
| Visual Surface<br>Requirement | Confirm that the device<br>meets the visual surface<br>requirements. | Pass<br>All samples met the<br>predetermined acceptance<br>criteria |
| Packaging Visual<br>Inspection | Confirm that the packaging<br>meets the visual inspection. | Pass<br>All samples met the<br>predetermined acceptance<br>criteria |
| Dimensional/Visual<br>Inspection | Device dimensions were<br>measured to confirm<br>conformance to the<br>specifications. | Pass<br>All samples met the<br>predetermined acceptance<br>criteria |
| Liquid Leakage Under<br>Pressure | Verify that the catheter joint<br>strength meets the<br>freedom from leakage<br>(liquid leakage during<br>pressurization)<br>requirements of ISO<br>10555-1:2013, Annex C. | Pass<br>All samples met the<br>predetermined acceptance<br>criteria |
| Hub Aspiration Air<br>Leakage | Confirm that the device<br>passes the hub aspiration<br>air leakage test of ISO<br>10555-1:2013, Annex D. | Pass<br>All samples met the<br>predetermined acceptance<br>criteria |
| Simulated Use | Simulated use testing with<br>accessories in an<br>anatomical model which<br>simulated the tortuosity of<br>the neurovasculature.<br>Devices were delivered<br>through the tortuous model<br>to evaluate effectiveness of<br>the device at retrieval of soft<br>and firm clots and<br>mechanical integrity after<br>multiple passes. Testing<br>performed with reference<br>device (K211564) for<br>comparison | Pass<br>All samples met the<br>predetermined acceptance<br>criteria |
| Test | Test Summary | Result |
| Flex Fatigue | Meets minimum value per specification for multiple passes in the simulated use model. | Pass<br>All samples met the predetermined acceptance criteria |
| Track and Advance Force | Test specimens were tested for track and advance force. | Pass<br>All samples met the predetermined acceptance criteria |
| Tip Deflection | Test specimens were tested for tip deflection forces and compared to predicate catheters. | Pass<br>All samples met the predetermined acceptance criteria |
| Torque | Determine the number of revolutions to failure of the Catheter in simulated anatomy. | Pass<br>All samples met the predetermined acceptance criteria |
| Tip Elongation and Compression | Test specimens were tested for tip elongation and compression. | Pass<br>All samples met the predetermined acceptance criteria |
| Peak Tensile | The tensile strength of the catheter sections and bonds was tested after simulated use. | Pass<br>All samples met the predetermined acceptance criteria |
| Particulates, Coating Integrity | The integrity of the hydrophilic coating was evaluated after multiple insertion and withdrawal cycles. Particulates were measured during simulated use and compared to the reference device. | Pass<br>All samples met the predetermined acceptance criteria |
| Flow Rate | Determine the flow rate through a catheter, based on ISO 10555-1, Annex E. | Pass<br>All samples met the predetermined acceptance criteria |
| Aspiration Flow Rate | Determine the aspiration flow rate through the aspiration catheter when the catheter is connected to a constant vacuum source. | Pass<br>All samples met the predetermined acceptance criteria |
| Kink Resistance | Test specimen segments were formed into a defined bend diameter to evaluate kink resistance. | Pass<br>All samples met the predetermined acceptance criteria |
| Test | Test Summary | Result |
| Corrosion Resistance | No visible corrosion immediately after Corrosion Testing procedure based on ISO 10555-1, Annex A. | Pass<br>All samples met the<br>predetermined acceptance<br>criteria |
| Radiopacity | The marker band is fluoroscopically visible. | Pass<br>All samples met the<br>predetermined acceptance<br>criteria |
| Burst Pressure-Static | Tested per ISO 10555-1:2013, Annex F. Testing conducted after simulated use. | Pass<br>All samples met the<br>predetermined acceptance<br>criteria |
| Burst Pressure-Dynamic | Minimum value per specification. | Pass<br>All samples met the<br>predetermined acceptance<br>criteria |
| Connectors for<br>Intravascular or<br>Hypodermic<br>Applications | Hubs were tested per ISO 80369-7. | Pass<br>All samples met the<br>predetermined acceptance<br>criteria |
| Lumen Collapse | Aspiration Catheter samples were tested for lumen patency under maximum applied vacuum pressures. | Pass<br>All samples met the<br>predetermined acceptance<br>criteria |
| Manual Syringe Injection<br>Peak Pressure (psi) | Measure peak pressure during manual injection of contrast media with a syringe. | Pass<br>All samples met the<br>predetermined acceptance<br>criteria |
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the word "Q'apel" in blue font. The font is a sans-serif font, and the letters are all lowercase except for the "Q". The word is slightly italicized, and the apostrophe is placed above the "a". The background is white.
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the word "Q'apel" in blue font. The font is sans-serif and the letters are bold. The apostrophe is placed between the "Q" and the "a". The word is slightly tilted to the right.
Q'Apel also confirmed that the Aspiration Tubing meets all design and performance requirements through the following bench testing:
| Test | Test Summary | Result |
|----------------------------------|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|
| Dimensional/Visual<br>Inspection | Confirm that the Aspiration<br>Tubing meets all dimensional<br>and visual specifications. | Pass<br>All samples met the<br>predetermined<br>acceptance criteria |
| Tensile Strength | Confirm that the Aspiration<br>Tubing meets the existing<br>tensile strength specifications. | Pass<br>All samples met the<br>predetermined<br>acceptance criteria |
| Simulated Use Test | Confirm that the Aspiration<br>Tubing passes testing specified<br>in the simulated use test<br>protocol. | Pass<br>All samples met the<br>predetermined<br>acceptance criteria |
{13}------------------------------------------------
Image /page/13/Picture/1 description: The image shows the word "Q'apel" in blue font. The font is a sans-serif font, and the letters are all lowercase except for the "Q". The word is slightly italicized, and the apostrophe is placed between the "Q" and the "a". The background is white.
| Aspiration Tubing<br>Resistance to Collapse<br>and Leakage at<br>Maximum Aspiration<br>Pressures | Confirm that the Aspiration<br>Tubing resistance to collapse at<br>maximum aspiration pressure<br>testing is as specified in the test<br>protocol. Confirm that the<br>Aspiration Tubing does not show<br>signs of leakage at maximum<br>aspiration pressure. | Pass<br>All samples met the<br>predetermined<br>acceptance criteria |
|--------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|
| Flow Switch<br>Functionality Testing | The Flow Control Switch<br>completely and immediately<br>stops fluid flow after a specified<br>number of ON/OFF cycles. | Pass<br>All samples met the<br>predetermined<br>acceptance criteria |
#### Performance Testing - Animal:
An animal study was conducted to evaluate the safety, effectiveness, and usability of the 072 Aspiration System in comparison to cleared control devices, the CERENOVUS Large Bore Catheter (K193380) and the Penumbra System (Reperfusion Catheter RED 72) (K211654), in a porcine model according to Good Laboratory Practices (GLP) per 21 CFR Part 58. Sub-chronic and chronic time points were assessed. The study included aspiration of experimental clots and a worst-case wedge assessment. Clot aspiration and wedge assessment were comparable between the test and control catheters, and both were shown to be safe in porcine vessels via angiography and vessel histology.
#### Performance Data - Clinical:
No clinical study was conducted as bench and animal testing was determined sufficient for verification and validation purposes and to support the substantial equivalence of the 072 Aspiration System.
#### Sterilization:
The 072 Aspiration System, which includes the Aspiration Tubing, is sterilized using a validated 100% Ethylene Oxide sterilization process to ensure sterility assurance level (SAL) of 10 6 in accordance with AAMI TIR 28:2016 "Product Adoption and Process Equivalence for Ethylene Oxide Sterilization" and per the requirements of ISO 11135:2014 "Sterilization of health-care products – Ethylene oxide – Requirements for the development, validation and routine control of a sterilization process for medical devices." The 072 Aspiration System, accessories, and Aspiration Tubing are for single use only.
#### Shelf-Life Testing:
The 072 Aspiration System, which includes the Aspiration Tubing, has a shelf life of six months based on the successful completion of stability testing. Shelf-life testing was performed using standard test methods and acceptance criteria. All samples were exposed to standard transportation conditioning and distribution before aging. Results of testing on the subject device met the established acceptance criteria.
{14}------------------------------------------------
### Biocompatibility Testing:
The 072 Aspiration System was assessed for biocompatibility in accordance with ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation of Testing within a Risk Management Process," and the FDA Guidance for Industry and FDA Staff, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process"," (issued September 4, 2020).
The 072 Aspiration Catheter and the Delivery Tool are considered externally communicating medical devices with direct contact with circulating blood for a limited (≤ 24 hours) duration.
The Rotating Hemostasis Valve (RHV), Flow Switch, and Hub are considered external communicating devices contacting blood indirectly for a limited (< 24 hours) duration.
The Aspiration Tubing is considered to have contact with intact skin for a limited (≤ 24 hours) duration.
Biocompatibility testing on the 072 Aspiration System (072 Aspiration Catheter and Delivery Tool) included:
| Test Name | Acceptance Criteria | Conclusion |
|---------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| Cytotoxicity - MEM Elution<br>ISO 10993-5:2009 | The achievement of a numerical<br>grade greater than 2 is considered<br>a cytotoxic effect. | Non-cytotoxic |
| Sensitization<br>Magnusson- Kligman Method<br>ISO 10993-10:2010 | Magnusson and Kligman grades<br>of 1 or greater in the test group<br>generally indicate sensitization,<br>provided grades of less than 1 are<br>seen in control animals. If grades<br>of 1 or greater are noted in control<br>animals, then the reactions of test<br>animals which exceed the most<br>severe reaction in control animals<br>are presumed to be due to<br>sensitization.…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.