K191830 · Electrocore, Inc. · PKR · Mar 26, 2020 · Neurology
Device Facts
Record ID
K191830
Device Name
gammaCore Sapphire
Applicant
Electrocore, Inc.
Product Code
PKR · Neurology
Decision Date
Mar 26, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5892
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The gammaCore Sapphire Non-invasive Vagus Nerve Stimulator is intended to provide non-invasive vagus nerve stimulation (nVNS) on the side of the neck. The gammaCore Sapphire device is indicated for: - Adjunctive use for the preventive treatment of cluster headache in adult patients. - The acute treatment of pain associated with episodic cluster headache in adult patients. - The acute treatment of pain associated with migraine headache in adult patients. - The preventive treatment of migraine headache in adult patients.
Device Story
Handheld, rechargeable, portable nVNS device; applies mild electrical stimulation to vagus nerve via neck application. User controls signal amplitude (0-40) via slide switch. Device delivers 120-second stimulation cycles; automatic shut-off after cycle. Dosing limited to fixed number per 24-hour period. Refill/reload via RFID cards programmed by healthcare provider prescription; cards encoded using gammaCore Dispensing and Ordering Terminal (DOT) application on tablet via NFC. Bluetooth feature enables manufacturer-only diagnostics (usage/dosing history). Healthcare provider prescribes therapy; patient self-administers stimulation. Output provides non-invasive neuromodulation to potentially reduce headache frequency/pain. Device includes charging station/case.
Clinical Evidence
Evidence from two randomized, sham-controlled, double-blind trials: PREMIUM (n=332) and EVENT (n=59). PREMIUM (episodic migraine): mITT analysis showed significant reduction in migraine days (therapeutic gain 0.74; P=0.043). EVENT (chronic migraine): primary endpoint (headache days) showed -1.4 reduction (nVNS) vs -0.2 (sham). Both studies demonstrated favorable safety/tolerability profiles with no serious device-related adverse events.
Technological Characteristics
Materials: Stainless steel, ABS-PC, SignaGel electrode gel. Energy: 3-V LiFePo4 battery. Waveform: Sinusoidal, symmetrical biphasic 5000-Hz pulses at 25 Hz. Output: 30 V peak, 60 mA peak. Impedance: 450-550 ohms. Connectivity: NFC (RFID refill), Bluetooth (manufacturer diagnostics). Sterilization: Not specified. Software: Embedded firmware for stimulation control and dose management.
Indications for Use
Indicated for adult patients for: adjunctive preventive treatment of cluster headache; acute treatment of pain associated with episodic cluster headache; acute treatment of pain associated with migraine headache; and preventive treatment of migraine headache.
Regulatory Classification
Identification
An external vagal nerve stimulator for headache is a prescription device used to apply an electrical current to a patient's vagus nerve through electrodes placed on the skin for the treatment of headache.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The technical parameters of the device, including waveform, output modes, maximum output voltage and current (with 500, 2,000, and 10,000 ohm loads), pulse duration, frequency, net charge (µC) per pulse, maximum phase charge at 500 ohms, maximum current density (mA/cm
2 , r.m.s.), maximum average current (mA), maximum average power density (W/cm2 ), and the type of impedance monitoring system shall be fully characterized through non-clinical performance testing.(2) Software verification, validation, and hazard analysis shall be performed.
(3) Biocompatibility evaluation of the patient-contacting components of the device shall be performed.
(4) The device shall be tested for electrical, thermal, and mechanical safety, and for electromagnetic compatibility (EMC).
(5) The labeling must include:
(i) Instructions for proper use of the device, including placement of the device on the patient; and
(ii) Instructions on care and cleaning of the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 26, 2020
Electrocore, Inc. Mike Romaniw VP, Quality Assurance & Regulatory Affairs 150 Allen Road, Suite 201 Basking Ridge, New Jersey 07920
Re: K191830
Trade/Device Name: gammaCore Sapphire Regulation Number: 21 CFR 882.5892 Regulation Name: External vagal nerve stimulator for headache Regulatory Class: Class II Product Code: PKR. OAK Dated: February 25, 2020 Received: February 27, 2020
Dear Mike Romaniw:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Vivek Pinto, PhD Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191830
Device Name gammaCore Sapphire
#### Indications for Use (Describe)
The gammaCore Sapphire Non-invasive Vagus Nerve Stimulator is intended to provide non-invasive vagus nerve stimulation (nVNS) on the side of the neck. The gammaCore Sapphire device is indicated for:
- · Adjunctive use for the preventive treatment of cluster headache in adult patients.
- · The acute treatment of pain associated with episodic cluster headache in adult patients.
- · The acute treatment of pain associated with migraine headache in adult patients.
- · The preventive treatment of migraine headache in adult patients.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(K) SUMMARY 5
The following information is provided as required by 21 CFR §807.87 for the electroCore gammaCore Sapphire 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the information upon which the substantial equivalence determination is based.
| Applicant: | electroCore, Inc. |
|------------|---------------------------------|
| | 150 Allen Road, Suite 201 |
| | Basking Ridge, New Jersey 07920 |
| | Ph: 973-290-0097 |
| | Fax: 973-290-9171 |
## Establishment Registration Number: 3009060963
| Contact: | Mike Romaniw |
|--------------------|-------------------------------------------|
| | VP, Quality Assurance & Regulatory Affair |
| | Office: 973-355-6702 |
| | Fax: 973-290-9171 |
| | Mike.Romaniw@electroCore.com |
| Alternate Contact: | Eric J. Liebler |
| | SVP, Neurology |
| | Mobile: 908-938-9780 |
| | Eric.Liebler@electroCore.com |
| Alternate Contact: | Deborah Lavoie Grayeski |
| | Sr. Project Manager |
| | M Squared Associates, Inc. |
| | Office: 347-954-0418 |
| | Fax: 703-562-9797 |
| | dgrayeski@MSquaredAssociates.com |
{4}------------------------------------------------
| Alternate Contact: | Janice M. Hogan, Esq. |
|------------------------|----------------------------------------------|
| | Partner |
| | Hogan Lovells US LLP |
| | Direct: 267-675-4611 |
| | Janice.Hogan@HoganLovells.com |
| Date of Submission: | March 24, 2020 |
| Proprietary Name: | gammaCore Sapphire |
| Common Name: | External vagal nerve stimulator for headache |
| Classification Status: | Class II |
| Product Codes: | PKR, QAK |
| Predicate Device: | gammaCore Sapphire, K182369 |
Indications for Use: The gammaCore Sapphire Non-invasive Vagus Nerve Stimulator is intended to provide non-invasive vagus nerve stimulation (nVNS) on the side of the neck. The gammaCore Sapphire device is indicated for:
- Adjunctive use for the preventive treatment of cluster headache in adult patients. .
- The acute treatment of pain associated with episodic cluster headache in adult patients. .
- . The acute treatment of pain associated with migraine headache in adult patients.
- The preventive treatment of migraine headache in adult patients.
Device Description: The gammaCore Sapphire (gammaCore) is a multiuse, handheld, rechargeable, portable device consisting of a rechargeable battery and signal-generating and amplifying electronics, with a slide control switch for user/operator control of the signal amplitude (relative range, 0-40 continuous).
The gammaCore Sapphire:
- . Includes a charging station incorporated into the "clam shell" storage case connected to a power adapter for charging of the device as necessary by the end user.
- Provides visible (light and display) and audible (beep) feedback regarding device and . stimulation status.
{5}------------------------------------------------
- Allows for multiple stimulations or doses; each stimulation or dose lasts 120 seconds, . after which the device automatically turns off unless turned off earlier by the user/operator.
Note: One dose is defined as one stimulation cycle lasting 120 seconds (2 minutes).
- Delivers up to a fixed number of doses within a 24-hour period; once the maximum daily . number of doses has been reached, the device will not deliver any more doses until the following 24-hour period.
- Indicates on the display the number of remaining doses available in a 24-hour period. .
The device will be provided to the patient/user with an initial 10-, 31-, or 93-day radio-frequency identification (RFID) card on the basis of the healthcare provider's prescription. Additional (refill/reload) cards will be provided in response to a user/patient request based on a prescription from his or her healthcare provider. The refill/reload RFID cards will be programmed using the gammaCore Dispensing and Ordering Terminal (gammaCore DOT) by electroCore or its authorized agent. This is a specialized application for dispensing the device therapy.
When a 10-. 31-. or 93-day refill/reload card is requested by a patient/user (in accordance with a prescription from a healthcare provider) for a unique device serial number, an RFID card is encoded with the appropriate dosage according to the prescription. The encoded RFID card is matched to a specific gammaCore device serial number residing in a database maintained by electroCore. The gammaCore DOT application, running on a Sony Xperia® tablet, uses a proprietary encoding algorithm to encrypt the therapy days and doses per day on the refill/reload RFID card using near field communication (NFC) protocols.
The encoding algorithm is based on a seed-value pair of numbers specific to a device ID (unique device serial number) that is registered in the gammaCore DOT database. The gammaCore DOT application ensures that only legitimate seed values allow refilling/reloading of the device through validation of the prescription and seed values in the gammaCore DOT database using the unique device and patient IDs.
{6}------------------------------------------------
The encoded refill/reload RFID card is then provided to the user/patient who requested the refill/reload of the device, along with 1 to 6 additional tubes of conductive gel (the number of conductive gel tubes provided is based on the 10-, 31-, or 93-day refill/reload being provided). On receipt of the RFID card, the user/patient refills/reloads his or her gammaCore device by placing the RFID card across the face of the device (with the device turned on). The device will display "rd" and the "refill" icon as the device reads the RFID card. The device will signal (beeping twice) when it has been loaded with the programmed doses. The device will now be ready for use as treatment. The RFID card can be used for only one refill/reload; upon completion of the device refill/reload, the card can be thrown away.
In addition, a Bluetooth® feature will be enabled to facilitate diagnostics of any devices returned by patients/users to the manufacturer, to allow determination of the number of days the device was used and/or the number of doses, as well as any days/doses remaining on the device. The Bluetooth feature will not be accessible to the patient/user; it is accessible only to the device manufacturer.
## Summary of Technological Characteristics:
There are no changes to the technological characteristics of the gammaCore Sapphire for this expanded indication.
## Summary of Non-clinical Testing:
There are no changes to the technological characteristics of the gammaCore Sapphire for this expanded indication; no additional non-clinical or performance testing is required.
# Summary of Clinical Data:
Clinical data demonstrating the safety and effectiveness of the gammaCore Sapphire for the prophylactic/preventive treatment of migraine headache were collected from the PREMIUM study and the EVENT study. The PREMIUM study was a prospective, randomized, doubleblind, sham-controlled, multicenter study in patients with episodic migraine conducted at 22 European sites from June 1, 2015, to November 21, 2017. It consisted of a 4-week run-in period of no study treatment, which was followed by a 12-week double-blind phase of randomly
{7}------------------------------------------------
assigned preventive treatment with nVNS or sham and a 24-week open-label phase during which all participants received nVNS therapy. Post hoc analysis of the modified intent-to-treat (mITT) population, which included subjects with ≥67% adherence per month, showed significant differences between groups in favor of nVNS for the study's primary endpoint, mean reduction in the number of migraine days per month (therapeutic gain, 0.74; P=0.043), as well as for headache days per month (therapeutic gain, 0.86; P=0.045) and acute medication days per month (therapeutic gain, 0.80; P=0.039). Response to treatment with nVNS during the double-blind period was maintained with further nVNS therapy during the open-label period. The sham device used in the PREMIUM study was recently proven to have activated the vagus nerve. This likely mitigated the therapeutic gain measured in the PREMIUM study. Table 3 summarizes the key characteristics of the PREMIUM study.
| Test device | gammaCore R-30 (10015-00302) | |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Study design | Randomized, sham-controlled, double-blind trial | |
| n | ITT<br>332 | mITT<br>278 |
| Sites | 22 | |
| Trial duration (per<br>subject) | 10 months (1-month screening, 3-month randomization phase, and 6-month open-label<br>phase) | |
| Treatment paradigm | Each treatment consisted of 2 consecutive 2-minute stimulations administered on the<br>same side of the neck 3 times per day: (1) upon waking and (2, 3) 6 to 8 hours after the<br>first and second daily treatments | |
| Primary efficacy<br>endpoint | Mean reduction in the number of migraine days from the 4-week run-in period<br>(baseline) to the last 4 weeks of the 12-week double-blind period | |
| Secondary efficacy<br>endpoints | Reductions in number of headache daysSubjects with ≥50% response rate | |
| Primary safety<br>endpoint | Assessment of onset, type, severity, and frequency of adverse events (anticipated and<br>unanticipated), including determination of device relatedness | |
| Key inclusion criteria | 1. Between the ages of 18 and 75 years<br>2. Previously diagnosed as suffering from migraine with or without aura in<br>accordance with the International Classification of Headache Disorders-3 Beta<br>criteria<br>3. Experienced between 5 and 12 headache days per month (over the last 4 months)<br>with at least 2 of the migraines lasting more than 4 hours<br>4. No use of any migraine prevention treatments and/or medications (including<br>Botox® [onabotulinumtoxinA] injections)<br>5. Agreed to refrain from changing the type or dosage of any prophylactic<br>medications for indications other than chronic migraine that in the opinion of the<br>clinician might interfere with the study objectives (eg, antidepressants,<br>anticonvulsants, beta-adrenergic blockers) | |
Table 3. Key Characteristics of the PREMIUM Study
{8}------------------------------------------------
| Key exclusion<br>criteria | 1. A history of aneurysm, intracranial hemorrhage, brain tumors, or significant head<br>trauma | | | | | | |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|--------------|----------------------------------------------------------------------|-------------------------|--------------------------------|--------------------------------|
| | 2. Known or suspected severe atherosclerotic cardiovascular disease, severe carotid<br>artery disease (eg, bruits or history of transient ischemic attack or cerebrovascular<br>accident), congestive heart failure, known severe coronary artery disease, or recent<br>myocardial infarction | | | | | | |
| | 3. Uncontrolled high blood pressure | | | | | | |
| | 4. Current implantation of an electrical and/or neurostimulator device, including but<br>not limited to a cardiac pacemaker or defibrillator, vagal neurostimulator, deep<br>brain stimulator, spinal stimulator, bone growth stimulator, or cochlear implant | | | | | | |
| | 5. Current implantation of metal cervical spine hardware or a metallic implant near<br>the gammaCore stimulation site | | | | | | |
| Subject baseline<br>demographics | Device | Population | Sex, No. (%) | | Age, y,<br>mean<br>(SD) | Headache<br>days, mean<br>(SD) | Migraine<br>days,<br>mean (SD) |
| | nVNS | ITT | 142 (86.1) | 23<br>(13.9) | 43.5<br>(11.1) | 8.9 (2.6) | 7.9 (2.2) |
| | | mITT | 119 (86) | 19 (14) | 43.9<br>(11) | 8.9 (2.6) | 8.1 (2.1) |
| | Sham | ITT | 138 (82.6) | 29<br>(17.4) | 41.4<br>(12.3) | 9.1 (2.6) | 8.1 (2.0) |
| | | mITT | 115 (82) | 25 (18) | 41.8<br>(12) | 9.0 (2.6) | 8.1 (2.1) |
| Primary safety<br>endpoint results | Preventive nVNS therapy was safe and well tolerated | | | | | | |
| | Across all study periods, the most common adverse device effects were rash, pain,<br>erythema, discomfort at the application site, and dizziness | | | | | | |
| | No serious adverse device effects were reported during the study | | | | | | |
| Primary efficacy<br>endpoint results | | | | Mean change<br>from baseline in<br>no. of migraine<br>days per month | | Therapeutic<br>gain | <i>P</i> value |
| Population | | n<br>(nVNS/Sham) | nVNS | Sham | | | |
| | | ITT | -2.3 | -1.8 | -0.47 | 0.15 | |
| | | mITT | -2.3 | -1.5 | -0.74 | 0.04 | |
| Control for<br>medication use | The use of migraine prevention medication was not allowed | | | | | | |
| | Medications were allowed for the acute relief of migraine symptoms | | | | | | |
Diener HC, Goadsby PJ, Ashina M, et al. Non-invasive vagus nerve stimulation (nVNS) for the preventive treatment of episodic migraine: the multicentre, double-blind, randomised, sham-controlled PREMIUM trial. Cephalalgia. 2019;39(12):1475-1487.
Abbreviations: ITT, intent-to-treat; mITT, modified intent-to-treat; nVNS, non-invasive vagus nerve stimulator; SD, standard deviation.
The EVENT study, a prospective, randomized, sham-controlled, multicenter pilot study for the prevention of chronic migraine, was conducted at 6 sites in the United States from October 2012 to April 2014. It consisted of a 4-week run-in phase of no study treatment, an 8-week doubleblind phase of randomly assigned preventive treatment with nVNS or sham, and a 6-month openlabel phase of nVNS treatment for all participants. The primary efficacy measure, mean change from baseline to month 2 in the number of headache days per month, was – 1.4 in the nVNS group and -0.2 in the sham group. Continued nVNS use led to significant reductions from
{9}------------------------------------------------
baseline in the number of headache days per month during the open-label period. Among subjects who completed the study, the proportion who had a ≥50% reduction from baseline in the number of headache days per month increased with greater duration of treatment. Such findings suggest that the benefits of nVNS therapy may accrue over time. Table 4 summarizes the key characteristics of the EVENT study.
| Test device | gammaCore-150 (10007-00305) | | | |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Study design | Randomized, sham-controlled, double-blind trial | | | |
| n | ITT | PP | | |
| | 59 | 49 | | |
| Sites | 6 | | | |
| Trial duration (per subject) | 9 months (1-month screening, 2-month randomization phase, and 6-month open-label phase) | | | |
| Treatment paradigm | Each treatment consisted of two 2-minute self-administered stimulations delivered 5 to 10 minutes apart to the right side of the neck at 3 prespecified times every day: (1) within 1 hour of awakening; (2) 6 to 8 hours after the first treatment; and (3) 6 to 8 hours after the second treatment | | | |
| Primary efficacy endpoint | Change in the mean number of headache days at the end of month 2 of the randomized phase compared with baseline | | | |
| Secondary efficacy endpoint | Severity of each headache day | | | |
| Primary safety endpoint | Assessment of the onset, type, severity, and frequency of adverse events (anticipated and unanticipated), including the determination of device relatedness | | | |
| Key inclusion criteria | 1. | Between the ages of 18 and 65 years | | |
| | 2. | Previously diagnosed as suffering from migraine with or without aura in accordance with the International Classification of Headache Disorders-2 criteria | | |
| | 3. | Experienced at least 15 headache days per month (over the last 3 months) | | |
| | 4. | No use of any migraine prevention treatments and/or medications (including Botox® [onabotulinumtoxinA] injections) | | |
| | 5. | Agreement to refrain from changing the type or dosage of any prophylactic medications for indications other than chronic migraine that in the opinion of the clinician might interfere with the study (eg, antidepressants, anticonvulsants, beta-adrenergic blockers) | | |
| Key exclusion criteria | 1. | A history of aneurysm, intracranial hemorrhage, brain tumors, or significant head trauma | | |
| | 2. | Known or suspected severe atherosclerotic cardiovascular disease, severe carotid artery disease (eg, bruits or history of transient ischemic attack or cerebrovascular accident), congestive heart failure, known severe coronary artery disease, or recent myocardial infarction | | |
| | 3. | Uncontrolled high blood pressure | | |
| | 4. | Current implantation of an electrical and/or neurostimulator device, including but not limited to cardiac pacemaker or defibrillator, vagal neurostimulator, deep brain stimulator, spinal stimulator, bone growth stimulator, or cochlear implant | | |
| | 5. | Current implantation of metal cervical spine hardware or a metallic implant near the gammaCore stimulation site | | |
| Table 4. Kev Characteristics of the EVENT Study | | |
|-------------------------------------------------|--|---|
| | | े |
{10}------------------------------------------------
| Subject baseline<br>demographics | Device | Population | Sex, No. (%) | | Age, y,<br>mean<br>(SD) | Headache days, mean<br>(SD) | |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|------------------|------------------------------------------------------------------------------|-------------------------|-----------------------------|----------------|
| | | ITT | Female | Male | 40.5<br>(14.2) | 20.8 (5.0) | |
| | nVNS | PP | 22 (85) | 4 (15) | 42.0<br>(14.4) | 22.7 (6.0) | |
| | | ITT | 27 (93) | 2 (7) | 38.8<br>(11.1) | 22.3 (4.9) | |
| | Sham | PP | 22 (96) | 1 (4) | 39.3<br>(11.1) | 25.2 (5.0) | |
| Primary safety<br>endpoint results | nVNS had a benign tolerability profile that was generally similar to that of the sham treatmentMost adverse events were mild or moderate and transientThe most commonly reported adverse events were upper respiratory tract infections and gastrointestinal symptoms | | | | | | |
| Primary efficacy<br>endpoint results | Population | | n<br>(nVNS/Sham) | Mean change from<br>baseline in no. of<br>migraine days per<br>month<br>nVNS | Sham | Therapeutic<br>gain | <i>P</i> value |
| | ITT | | 30/29 | -1.4 | -0.2 | -1.2 | 0.56 |
| | PP | | 26/23 | -2.0 | -0.1 | -1.9 | 0.44 |
| Control for<br>medication use | The use of migraine prevention medication was not allowedMedications were allowed for the acute relief of migraine symptoms | | | | | | |
Silberstein SD, Calhoun AH, Lipton RB, et al. Chronic migraine headache prevention with noninvasive vagus nerve stimulation: the EVENT study. Neurology. 2016;87(5):529-538.
Abbreviations: ITT, intent-to-treat; nVNS, non-invasive vagus nerve stimulator; PP, per-protocol; SD, standard deviation.
Consistent with results from previous trials. nVNS was shown to be safe and well tolerated in both the PREMIUM and EVENT studies. No serious device-related adverse events (AEs) were reported, and no new safety concerns for nVNS use in the intended population were apparent.
# Substantial Equivalence Discussion:
gammaCore Sapphire technology is identical to the device technology cleared under K182369. There have been no changes in the technological characteristics or intended use of the gammaCore Sapphire in the proposed Indications for Use statement. The instructions for the device associated with the dosing for the prevention of migraine headache differ from those for preventive or acute treatment of cluster headache and those for the acute treatment of migraine headache; however, for all indications, use of more than 24 stimulations per day has not been formally evaluated and continues to be listed as a precaution in the labeling.
{11}------------------------------------------------
# Summary:
Table 5 establishes the substantial equivalence of the subject device to that of the predicate device.
{12}------------------------------------------------
| | gammaCore Sapphire<br>(Subject Device) | gammaCore Sapphire<br>(Predicate Device) | Substantial Equivalence |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | TBD | K182369 | |
| Intended use | The gammaCore Sapphire is a device that<br>provides non-invasive vagus nerve stimulation<br>(nVNS) when applied to the side of the neck.<br>This is a mild electrical stimulation of the<br>vagus nerve, which runs through the neck and<br>carries information to the central nervous<br>system. Each stimulation with gammaCore<br>lasts 2 minutes. The patient controls the<br>stimulation strength. | The gammaCore Sapphire is a device that<br>provides nVNS when applied to the side of<br>the neck. This is a mild electrical stimulation<br>of the vagus nerve, which runs through the<br>neck and carries information to the central<br>nervous system. Each stimulation with<br>gammaCore lasts 2 minutes. The patient<br>controls the stimulation strength. | No change in intended use |
| Indications for use | The gammaCore Sapphire is indicated for:<br>• Adjunctive use for the preventive treatment<br>of cluster headache (CH) in adult patients<br>• The acute treatment of pain associated with<br>episodic cluster headache (eCH) in adult<br>patients<br>• The acute treatment of pain associated with<br>migraine headache in adult patients<br>• The preventive treatment of migraine<br>headache in adult patients | The gammaCore Sapphire is indicated for:<br>• Adjunctive use for the<br>preventive treatment of CH in adult<br>patients<br>• The acute treatment of pain associated with<br>eCH in adult patients<br>• The acute treatment of pain associated with<br>migraine headache in adult patients | The expansion of the indication<br>does not alter the intended<br>therapeutic effect or otherwise<br>create a new intended use, as<br>explained previously. Supported by<br>clinical data in this submission. |
| Rx vs OTC | Prescription use only | Prescription use only | No change |
| | gammaCore Sapphire<br>(Subject Device) | gammaCore Sapphire<br>(Predicate Device) | Substantial Equivalence |
| Treatment recommendation | Preventive treatment of migraine:<br>120-second stimulation cycle,<br>2 consecutive stimulations on either side of the neck as follows:<br>• First daily treatment: within 1 hour of waking<br>• Second daily treatment: 4-6 hours after the first daily treatment<br>• Third daily treatment: within 1 hour before going to sleep<br><br>Preventive treatment of CH:<br>120-second stimulation cycle,<br>3 consecutive stimulations on either side of the neck as follows:<br>• First daily treatment: within 1 hour of waking<br>• Second daily treatment: 7-10 hours after the first daily treatment | Preventive treatment of CH:<br>120-second stimulation cycle,<br>3 consecutive stimulations on either side of the neck as follows:<br>• First daily treatment: within 1 hour of waking<br>• Second daily treatment: 7-10 hours after the first daily treatment<br><br>Acute treatment of eCH:<br>120-second stimulation cycle,<br>3 consecutive stimulations up to 8 times a day<br><br>Acute treatment of migraine:<br>120-second stimulation cycle,<br>2 bilateral stimulations up to 3 times a day | Change in treatment protocol to reflect different forms of primary headache |
| | Acute treatment of eCH:<br>120-second stimulation cycle,<br>3 consecutive stimulations up to 8 times a day | | |
| | Acute treatment of migraine:<br>120-second stimulation cycle,<br>2 bilateral stimulations up to 3 times a day | | |
| | | | |
| Patient-contacting materials | SS, ABS-PC, SignaGel® electrode gel | SS, ABS-PC, SignaGel electrode gel | No change in materials |
| Electrical classification | UL 60601-1 Class III Type BF Applied Part | UL 60601-1 Class III Type BF Applied Part | No change in classification |
| Waveform/frequency | Sinusoidal wave, symmetrical biphasic 5000-Hz pulses at a rate of 25 Hz | Sinusoidal wave, symmetrical biphasic 5000-Hz pulses at a rate of 25 Hz | No change in waveform or frequency |
| | gammaCore Sapphire<br>(Subject Device) | gammaCore Sapphire<br>(Predicate Device) | Substantial Equivalence |
| Maximum output | 30 V (peak), 60 mA (peak) | 30 V (peak), 60 mA(peak) | No change in outputs |
| Load impedance | 450-550 ohms | 450-550 ohms | No change in impedance |
| Power supply | 3-V LiFePo4 battery…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.