ALTMS Magnetic Stimulation Therapy System, Blossom TMS Therapy System
Applicant
Remed Co., Ltd.
Product Code
OBP · Neurology
Decision Date
Apr 6, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5805
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ALTMS Magnetic Stimulation Therapy System (also Blossom TMS indicated for the treatment of Major Depressive Disorder in adult patients, who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.
Device Story
Computerized, electromechanical system delivering non-invasive, repetitive transcranial magnetic stimulation (rTMS); generates brief, rapidly alternating magnetic fields to induce electrical currents in cerebral cortex. Used in inpatient/outpatient settings (clinics, hospitals, offices) under physician supervision. System includes hardware, software, and standing arm for coil positioning. Clinician sets treatment parameters, determines motor threshold, and positions coil. Output affects cortical neurons; intended to treat Major Depressive Disorder. Modification allows variable inter-train intervals (11-26s), reducing treatment time to 18.8 minutes.
Clinical Evidence
Bench testing only. Electrical safety and performance testing conducted per IEC 60601, IEC 61000, and IEC 62366. Biocompatibility testing performed per ISO 10993-1. No new clinical data presented; equivalence supported by design verification and validation.
Technological Characteristics
Biphasic figure-8 coil; flat spiral winding (AIW 2.0 x 4.0mm wire, 12 turns/wing x 2). Max surface temperature 41°C. Peak magnetic field strength 8.1kT/s at 2cm. Connectivity: integrated system. Standards: IEC 60601, IEC 61000, IEC 62366, ISO 10993-1.
Indications for Use
Indicated for adult patients with Major Depressive Disorder who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.
Regulatory Classification
Identification
A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.
Predicate Devices
ALTMS Magnetic Stimulation Therapy System (K202537)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 6, 2022
REMED Co., Ltd Soojung Moon CEO Allura Medical Solutions Inc. 5485 Rathdrum Way Antioch, California 94531
Re: K220625
Trade/Device Name: ALTMS Magnetic Stimulation Therapy System, Blossom TMS Therapy System Regulation Number: 21 CFR 882.5805 Regulation Name: Repetitive Transcranial Magnetic Stimulation System Regulatory Class: Class II Product Code: OBP Dated: March 2, 2022 Received: March 3, 2022
Dear Soojung Moon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Pamela Scott Assistant Director Neuromodulation Psychiatry Devices DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220625
Device Name
ALTMS Magnetic Stimulation Therapy System (also Blossom TMS Therapy System)
Indications for Use (Describe)
ALTMS Magnetic Stimulation Therapy System (also Blossom TMS indicated for the treatment of Major Depressive Disorder in adult patients, who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
Dver-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
#### l. Submitter Information
| Company: | REMED Co., Ltd. |
|---------------------|-----------------------------------------------------------------------------------------------------|
| Address: | #301-#303 Migun Techno World II, 187, Techno 2-Ro,<br>Yuseong-gu, Daejeon, 34025, Republic of Korea |
| Tel / Fax: | +82-42-934-5560 / +82-42-934-5562 |
| Contact Person: | S.J. Moon, Official correspondent of REMED Co., Ltd.<br>+1 (407) 242 9793 / sjmoon@allurameds.com |
| Date of submission: | Date of submission: March 2nd, 2022 |
## II. Device Information
| Proprietary Name(s): | ALTMS Magnetic Stimulation Therapy System<br>(Blossom TMS Therapy System) |
|-----------------------|---------------------------------------------------------------------------|
| Common Name: | Repetitive Transcranial Magnetic Stimulator |
| Classification Name: | Transcranial Magnetic Stimulator |
| Product Code: | OBP |
| Device Class: | Class II |
| Regulation Number: | 882.5805 |
| Classification Panel: | Neurology |
## III. Predicate Information
- · ALTMS Magnetic Stimulation Therapy System (REDMED Co., Ltd. K202537) - Primary predicate device
- · Neurosoft TMS (TeleEMG, LLC. K173441) - Reference device
## IV. Device Description
The ALTMS Magnetic Stimulation Therapy System (also Blossom TMS Therapy System) is a computerized, electromechanical medical device that generates and delivers non-invasive, magnetic stimulation using brief duration rapidly alternating, or pulsed, magnetic fields to induce electrical currents directed at spatially discrete regions of the cerebral cortex. This method of cortical stimulation by application of brief magnetic pulses to the head is known as Transcranial Magnetic Stimulation.
The ALTMS Magnetic Stimulation Therapy System (also Blossom TMS Therapy System) is substantially equivalent to the predicate device, our own ALTMS Magnetic Stimulation Therapy System (K202537) and Neurosoft TMS (K173441). The intended use and indications for use of the proposed device are identical to the predicate device. Both devices are clinically operated based on the same course of
{4}------------------------------------------------
the device use including the TMS system set up, patient preparation process, and determination of patient's motor threshold, coil position, and treatment administration at the predetermined treatment stimulation parameters.
The ALTMS Magnetic Stimulation Therapy System (also Blossom TMS Therapy System) is a non-invasive tool for the stimulation of cortical neurons for the treatment of Major Depressive Disorder in audit patients, who have failed to achieve satisfactory improvement from antidepressant medication in the current episode. The ALTMS Magnetic Stimulation Therapy System (also Blossom TMS Therapy System) is used for patient treatment by prescription only under the supervision of a licensed physician. The proposed device can be used in both inpatient and outpatient settings, including physicians' offices, hospitals, and clinics. The ALTMS Magnetic Stimulation Therapy System (also Blossom TMS Therapy System) is an integrated system consisting of a combination of hardware, software, and accessories. The Standing Arm, an accessory device included in the ALTMS Magnetic Stimulation Therapy System (also Blossom TMS Therapy System) provides the coil positioning function.
## V. Indication for Use
ALTMS Magnetic Stimulation Therapy System (also Blossom TMS Therapy System) is indicated for the treatment of Major Depressive Disorder in adult patients, who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.
# VI. Comparison of Indications for Use Statements and Technological Characteristics
The ALTMS Magnetic Stimulation Therapy System (also Blossom TMS Therapy System) is substantially equivalent to the predicate devices (our own ALTMS Magnetic Stimulation Therapy System and Neurosoft TMS). The ALTMS Magnetic Stimulation Therapy System (also Blossom TMS Therapy System) and the predicate devices have identical intended use / indication for use, and technological characteristics. The principles of operation, the output stimulation parameters and the materials are equivalent to the predicates. The modification to the device allows a range of inter-train intervals from 11 to 26 seconds, rather than the fixed 26 second duration, which will allow a reduction in treatment time from 37.5 minutes to a minimum of 18.8 minutes.
The proposed, ALTMS Magnetic Stimulation Therapy System (also Blossom TMS Therapy System) and predicate device, our own ALTMS Magnetic Stimulation Therapy System (K202537) and Neurosoft TMS (K173441) share the same intended
{5}------------------------------------------------
use and indications for use. The actual substance and essential scopes of the intended use and indications for use between the ALTMS Magnetic Stimulation Therapy System (also Blossom TMS Therapy System) and predicate device are substantially equivalent.
Few minor differences between the proposed and predicate devices are primarily identified through dimensional differences in modest levels, which are not considered to raise different questions of safety and effectiveness compared to the predicate device. The favorable results of the design verification and electrical safety testing of the ALTMS Magnetic Stimulation Therapy System (also Blossom TMS Therapy System) have demonstrated that the proposed device has met the predetermined design requirements and the FDA's recognized standards, including the electrical safety and performance standards and EMC standards as the predicate device. The design and function of the ALTMS Magnetic Stimulation Therapy System (also Blossom TMS Therapy System) do not involve a new or different technological principle that would be considered to raise different questions in safety and effectiveness compared to the predicate device. The principle of operation, critical function and clinical use and application are the same as the predicate device, our own ALTMS Maqnetic Stimulation Therapy System (K202537) and Neurosoft TMS (K173441).
## VII. Safety & Performance
Electrical safety and performance testing have been successfully carried out in accordance with IEC 60601, IEC 61000, and IEC 62366. Additionally, biocompatibility test also has been successfully fulfilled following ISO 10993-1. All the results already demonstrated through the primary predicate device, ALTMS Magnetic Stimulation Therapy System (K202537), and there is no change of the proposed device from the primary predicate device. The minor change to indicate the treatment protocols on the IFU has been evaluated by the risk analysis procedure. Therefore, the proposed device, ALTMS Magnetic Stimulation Therapy System(also Blossom TMS Therapy System), is as safe and effective as the reference device, which meet user's needs, and the Instruction For Use (IFU) support the safe and effective use of the proposed device.
## Substantial Equivalence Comparison
| DESCRIPTIVE<br>INVORMATION | Proposed Device:<br>ALTMS Magnetic<br>Stimulation Therapy | Predicate Device:<br>ALTMS Magnetic<br>Stimulation Therapy<br>System (K202537) | Predicate Device:<br>Neurosoft TMS<br>(K173441) |
|--------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | System (also Blossom<br>TMS Therapy System) | | |
| Pulse Shape | The output waveform<br>produced by the biphasic<br>figure 8 coil was<br>measured using a<br>calibrated search coil<br>connected to an<br>oscilloscope. As a result,<br>the waveform could be<br>quantified in three<br>directions as well as in<br>time. | The output waveform<br>produced by the biphasic<br>figure 8 coil was<br>measured using a<br>calibrated search coil<br>connected to an<br>oscilloscope. As a result,<br>the waveform could be<br>quantified in three<br>directions as well as in<br>time. | The output waveform<br>produced by the biphasic<br>figure 8 coil was<br>measured using a<br>calibrated search coil<br>connected to an<br>oscilloscope. As a result,<br>the waveform could be<br>quantified in three<br>directions as well as in<br>time. |
| Pulse Train<br>Duration Range<br>(sec) | 0.1~1 Hz: 1~1800sec<br>2~30 Hz: 1~20sec | 0.1~1 Hz: 1~1800sec<br>2~30 Hz: 1~20sec | 0.5~100 (sec) |
| Inter-train interval<br>range | 1~120sec | 1~120sec | 0~300 (sec) |
| Pulse Width | 430 µs | 430 µs | 280µS |
| Pulse Amplitude | 2.76V Pk-Pk at peaking<br>coil | 2.76V Pk-Pk at peaking<br>coil | Unknown |
| Spatial<br>distribution of the<br>output level | Refer to the sponsor's<br>test report for the Spatial<br>distribution of the output<br>level by the proposed<br>device ALTMS Magnetic<br>Stimulation Therapy<br>System. | Refer to the sponsor's<br>test report for the Spatial<br>distribution of the output<br>level by the proposed<br>device ALTMS Magnetic<br>Stimulation Therapy<br>System. | Unknown |
| Linearity of the<br>output level | Refer to the sponsor's<br>test report for the linearity<br>of the output level by the<br>proposed device ALTMS<br>Magnetic Stimulation<br>Therapy System. | Refer to the sponsor's<br>test report for the linearity<br>of the output level by the<br>proposed device ALTMS<br>Magnetic Stimulation<br>Therapy System. | Unknown |
| Magnetic field<br>intensity | 120% of the MT | 120% of the MT | 120% of the MT |
| Number of trains | 75 | 75 | 75 |
| Magnetic pulses<br>per session | 3000 | 3000 | 3000 |
| Treatment<br>session duration | 18.8 - 37.5 min | 37.5 min | 18.8 min - 37.5 min |
| Sessions per<br>week | 5 | 5 | 5 |
| Area if the brain<br>to be stimulated | Frontal Cortex area | Frontal Cortex area | Frontal Cortex area |
| Applicator<br>configuration and<br>core material | Biphasic Figure 8 Coil | Biphasic Figure 8 Coil | Figure 8 Coil |
| Coil Parameters | Flat spiral winding,<br>AIW 2.0 x 4.0mm wire,<br>12 turns/wing x 2 | Flat spiral winding,<br>AIW 2.0 x 4.0mm wire,<br>12 turns/wing x 2 | FEC-02-100C<br>Inner diameter:<br>47x50mm1<br>Outer diameter:<br>97x100mm1<br>Area: 184725mm²<br>Average Inductance:<br>10μH<br>Flat spiral winding<br>N: 16 turns (2 layers x 2<br>wings)<br>AFEC-02-100C<br>Inner diameter:<br>36x51mm1<br>Outer diameter:<br>84x106mm1<br>Area: 184725mm²<br>Average Inductance:<br>10μH<br>Flat spiral winding<br>N: 16 turns (2 layers x 2<br>wings)<br>FEC-02-100<br>Inner diameter:<br>47x50mm1<br>Outer diameter:<br>97x100mm1<br>Area: 91560mm²<br>Average Inductance:<br>10μH<br>Flat spiral winding<br>N: 16 turns (2 layers x 2<br>wings)<br>AFEC-02-100<br>Inner diameter:<br>36x51mm1 |
| | | | Outer diameter:<br>84x106mm¹<br>Area: 87200mm²<br>Average Inductance:<br>10μH<br>Flat spiral winding<br>N: 16 turns (2 layers x 2 wings) |
| Maximum trains<br>per session | 120 | 120 | 4800 |
| Temperature on<br>Surface at<br>Maximum Output | 41°C | 41°C | 41°C |
| Magnetic Field:<br>Peak Magnetic<br>Field Strength at<br>2cm (in dB/dt) | 8.1kT/s | 8.1kT/s | Unknown |
| Machine Output<br>Stimulation<br>Parameters, with<br>Amplitude in SMT<br>Units | 0.3-1.9 SMT | 0.3-1.9 SMT | FEC-02-100-C: 0 - 1.89<br>AFEC-02-100-C: 0 - 2.38<br>FEC-02-100: 0 - 1.92<br>AFEC-02-100: 0 - 2.33 |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
## VIII. Conclusion
In conclusion, the critical aspects of the ALTMS magnetic Stimulation Therapy System (also Blossom TMS Therapy System) are the same as the predicate devices, Neurosoft TMS (K173441) and to our own ALTMS Magnetic Stimulation Therapy System (K202537) including the intended use, indications for use, operation principle. The comprehensive evaluations of the ALTMS magnetic Stimulation Therapy System (also Blossom TMS Therapy System) along with the aforementioned design verification and validation testing assessments provide assurance that the ALTMS magnetic Stimulation Therapy System (also Blossom TMS Therapy System) has met the predetermined design requirements per the FDA's recognized consensus standards or manufacturer's standards applied to the predicate devices in the same manner. The comparative and comprehensive assessments all point to the conclusion that the ALTMS magnetic Stimulation Therapy System (also Blossom TMS Therapy System) is substantially equivalent to the predicate, Neurosoft TMS and to our own ALTMS Magnetic Stimulation Therapy System.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.