K211055 · Shenzhen DJ Medical Equipment Co., Ltd. · NFO · Nov 5, 2021 · Neurology
Device Facts
Record ID
K211055
Device Name
Microcurrent device (Model: HBR2-1)
Applicant
Shenzhen DJ Medical Equipment Co., Ltd.
Product Code
NFO · Neurology
Decision Date
Nov 5, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device uses microcurrent electrical to stimulate facial tissues for aesthetic purposes.
Device Story
Hand-held, reusable microcurrent stimulator for home aesthetic use. Device consists of main unit and wireless charging station. Internal lithium battery powers MCU-controlled circuit generating low-frequency microcurrent. User adjusts output intensity via buttons (10 levels). Device applies current to facial skin. Safety features include automatic overload trip, no-load trip, and automatic shut-off. User operates device to achieve cosmetic effects. Output parameters and safety profile are comparable to predicate devices.
Clinical Evidence
No clinical data required. Substantial equivalence supported by bench testing, electrical safety (IEC 60601-1, IEC 60601-2-10, IEC 60601-1-11), EMC (IEC 60601-1-2), and biocompatibility (ISO 10993-5, ISO 10993-10) testing.
Indicated for aesthetic facial tissue stimulation in home settings for adult users.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Micro-current Wrinkle Removing Facial Service model EP-400 (K162106)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 5, 2021
Shenzhen DJ Medical Equipment Co., Ltd. % Jet Li Regulation Manager Guangzhou KEDA Biology Technology Co., Ltd. 6F, No.1 TianTai road, Science City, LuoGang District Guangzhou, Guangdong China
Re: K211055
Trade/Device Name: Microcurrent device (Model: HBR2-1) Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NFO Dated: October 1, 2021 Received: October 7, 2021
Dear Jet Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211055
Device Name Microcurrent device (Model:HBR2-1)
Indications for Use (Describe)
The device uses microcurrent electrical to stimulate facial tissues for aesthetic purposes.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
## 510k number: K211055
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
#### Submitter's Information 1.
- Manufacturer Name: Shenzhen DJ Medical Equipment Co., Ltd.
- Address: Room 801, Building A, FirstFlag Science & Technology Park, No. 26 Baili Road, Nanwan Street., Longgang District, Shenzhen, Guangdong, China
- Tel.: +86-755-8557 3525
- Fax: +86-0871-67263826
- Contact Person: Feiyang Huang
- E-mail: info@dimiot.com
## Application Correspondent
- Guangzhou KEDA Biological Tech Co., Ltd.
- Address: 6F, No.1 TianTairoad, Science City, LuoGang District, GuangZhou City, China ●
- Contact Person: Mr. Jet Li ●
- Title: Regulation Manager ●
- Tel: +86-18588874857 ●
- Email: med-il@foxmail.com ●
## 2. Subject DeviceInformation
Type of 510(k) submission: Traditional Common Name: Stimulator, Nerve, Transcutaneous, Over-The-Counter Trade Name: Microcurrent device (Model: HBR2-1) Classification Name: Stimulator, Nerve, Transcutaneous Review Panel: Neurology, Physical Medicine Product Code: NFO Regulation Number: 882.5890 Regulation Class: 2
{5}------------------------------------------------
#### 3. Predicate Device Information
| Sponsor | TAMA Research<br>Corporation | Li-Tek Electronics<br>Technologies |
|-------------------|------------------------------|------------------------------------------------------------------|
| Device Name | TAMA BEMS Device | Micro-current Wrinkle<br>Removing Facial Service<br>model EP-400 |
| 510(k) Number | K173093 | K162106 |
| Product Code | NFO | NFO |
| Regulation Number | 882.5890 | 882.5890 |
| Regulation Class | 2 | 2 |
## 4. Device Description
The Microcurrent device (Model: HBR2-1) is a hand-held, non-sterile, reusable device designed to achieve the cosmetic effect. The device consist of main unit and wireless charging station with undetachable USB cable. The device is supplied by internal rechargeable lithium battery. It can be recharged by external charger through the charging station.
The device design with Micro current output stimulation function. The MCU-controlled microcurrent circuit generates a low-frequency microcurrent which will applied in the facial skin to do facial stimulation. In the microcurrent stimulation mode, there is ten levels of output intensity which would be adjusted by the button in the device.
The device is only intended for home use.
## 5. Intended Use / Indications for Use
The device uses microcurrent electrical to stimulate facial tissues for aesthetic purposes.
#### 6. Performance Summary
#### The device conforms to the following standards:
- IEC 60601-1: 2012 Medical electrical equipment Part 1: General requirements for basic safety and essentialperformance
{6}------------------------------------------------
- IEC 60601-1-2:2014, Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral standard: Electromaqnetic Compatibility - Requirements and Tests.
- IEC 60601-1-11:2015 Medical Electrical Equipment - Part 1-11: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment
- IEC 60601-2-10: 2016, Medical electrical equipment - Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators
- ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Test for in vitro cytotoxicity.
- ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Test for irritation and skin sensitization.
## Software Verification and Validation Testing
Software verification and validation testing were conducted and its documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was determined to beof "moderate" level of concern.
## Usability Testing
Usability testing was conducted to demonstrate that the device and its labeling can meet the following requirements:
1) the lay user can self-select themselves as being appropriate users of this device by the external box labeling,
2) the lay user can apply the treatment safely and correctly according to the instructions for use, and
3) the lay user can understand all indications, contraindications, warnings and precautions, and be able to identify whether they are within any contraindicated group; and be able to understand the user manual.
## Clinical Testing
Clinical data were not required in this submission to support a finding of substantial equivalence.
{7}------------------------------------------------
# 7. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, materials, mode of operation, and intended use of the Microcurrent device (Model: HBR2-1) is substantially equivalent to the predicate devices quoted above.
The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
{8}------------------------------------------------
| Elements of<br>Comparison | Subject Device<br>(K211055) | Primary Predicate Device<br>(K173093) | Reference device<br>(K162106) |
|-----------------------------------------------|--------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name and<br>Model | Microcurrent device<br>Model: HBR2-1 | TAMA BEMS Device | Micro-current Wrinkle<br>Removing Facial Service,<br>model EP-400 |
| Intended Use &<br>Indications for<br>Use | The device uses<br>microcurrent electrical to<br>stimulate facial tissues for<br>aesthetic purposes. | The TAMA BEMS Device uses<br>microcurrent electrical to<br>stimulate facial tissues for<br>aesthetic purposes. | The device is intended for<br>facial stimulation by<br>electrode heads for<br>cosmetic use. The device<br>is also intended for the<br>treatment of periorbital<br>wrinkles with red Light<br>Emitting Diode (LED)<br>head. It is for over-the-<br>counter use. |
| Prescription Status | Over-the-Counter Use (OTC)<br>Note: Usability study provided<br>to demonstrate OTC use | Prescription Use (Rx) Only | Over-the-Counter Use<br>(OTC) |
| Number of<br>outputs mode | 1 | 4 | 1 |
| Number of<br>channels | 1 | 1 | 1 |
| Output Intensity<br>Level | 10 | -- | -- |
| Stimulated<br>muscles | Face | Face | For micro current<br>stimulation: Face;<br>For Red light: periorbital |
| Maximum output<br>voltage | 0.54V@500Ω<br>2.24V @ 2kΩ<br>10.8V @ 10kΩ | ±0.400 @500Ω<br>±1.600 @2 k Ω<br>±8.00 @10k Ω | 1.23V @ 500Ω<br>3.64V @ 2kΩ<br>10.9V @ 10kΩ |
| Maximum Output<br>Current | 1.08mA @ 500Ω<br>1.12mA@2kΩ<br>1.08mA@10kΩ | ±0.800 @500 Ω<br>±0.800 @2 k Ω<br>±0.800 @10k Ω | 2.46mA @ 500Ω<br>1.82mA @ 2kΩ<br>1.09mA @ 10kΩ |
| Frequency range<br>(Hz) | 8.34Hz | Frequency (Hz) [or Rate (pps)]<br>0.045 - 2560 | 59.3Hz |
| Pulse width | 60ms | Pulse Duration (msec) 52.6 - 2400 | 4ms |
| Contraction and<br>Relaxation time | Adjustable, due to different modes. | Adjustable, due to different modes. | Adjustable, due to different modes. |
| Net Charge, (μC)<br>@ 500 Ω | 0μC @500Ω | 0μC @500Ω | 9.04μC@ 500Ω |
| Maximum Current<br>Density (mA/cm²<br>@ 500Ω) | 0.0325 mA/cm² @500Ω | 1.591mA/cm2@500Ω | Bigger electrode head:<br>0.044 mA/cm²@ 500Ω<br>Smaller electrode head:<br>0.42 mA/cm²@ 500Ω |
| Maximum pow er<br>Density | 106uW/cm² @500Ω | 318uW/cm2 @500Ω | Bigger electrode head<br>0.014 mW/ cm²@ 500Ω<br>Smaller electrode head<br>0.139 mW/ cm²@ 500Ω |
| Timer Range | 10 minutes | 0-30 minutes | 10 minutes |
| Additional Features | | | |
| Power | Adaptor Output: DC 5V,<br>0.3A | Internal 4.3V<br>rechargeable Lithium<br>Polymer battery | 1200mAh lithium battery |
| indication display | pow er on, pow er off,<br>operation mode. | On/Off Display Status,<br>Low Battery,<br>Voltage/Current Level | On/Off Status, Low<br>Battery, Voltage/Current<br>Level |
| Dimension | About 180 mm*40mm | 3x5x0.7(in.) | --; |
| Weight | 80g | 9.5 ounces | --; |
| Housing Materials | ABS | Anodized aluminum 6061 | ABS plastic |
| Microprocessor<br>control | Yes | Yes | Yes |
| Automatic<br>Overload trip | Yes | Yes | Yes |
| Automatic no-load<br>trip | Yes | Yes | Yes |
| Automatic shut-off | Yes | Yes | Yes |
| User override<br>control | Yes | Yes | Yes |
| Environment for<br>operating | Temperature: +18° C ~<br>+40° C<br>Relative humidity: 35 % ~<br>75 % | --; | Temperature: 5 ~ 40° C |
| Environment for<br>storage | Temperature: -10° C ~ +55°<br>C<br>Relative humidity: 20 % ~<br>80 % | --; | Temperature: -25 ~70° C<br>Humidity: 10 ~90% RH |
| Biocompatibility | All user directly contacting<br>materials are compliance<br>with ISO10993-5 and<br>ISO10993-10 requirements. | All user directly contacting<br>materials are compliance with<br>ISO10993-5 and ISO10993-10<br>requirements. | All user directly contacting<br>materials are compliance<br>with ISO10993-5,<br>ISO10993-10 and IEC<br>62471:2006 requirements. |
| Electrical Safety | Comply with IEC 60601-1<br>and IEC 60601-2-10;<br>IEC60601-2-57<br>IEC60601-1-11 | Comply with IEC 60601-1 and<br>IEC 60601-2-10 | Comply with IEC 60601-1<br>and IEC 60601-2-10 |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 |
{9}------------------------------------------------
{10}------------------------------------------------
# Final Conclusion:
The subject device "Microcurrent device (Model: HBR2-1)" is Substantial Equivalence to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.