Micro-current Wrinkle Reduction Facial service, model EP-400
Applicant
Li-Tek Electronic Technology Corporation
Product Code
NFO · Neurology
Decision Date
Feb 3, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device is intended for facial stimulation by electrode heads for cosmetic use. The device is also intended for the treatment of periorbital wrinkles with red Light Emitting Diode (LED) head. It is for over-the-counter use.
Device Story
Device (Model EP-400) provides facial stimulation via two interchangeable electrode heads and periorbital wrinkle treatment via a red LED head. Operated by consumer (OTC); device uses internal rechargeable lithium battery. Micro-current stimulation delivered to cheeks or under-eye area; red light therapy applied to periorbital region. Device includes microprocessor control for output regulation, automatic overload/no-load trips, and automatic shut-off. User monitors status via LED indicators. Benefits include cosmetic facial toning and wrinkle reduction.
Console: ABS plastic. Power: 1200mAh internal lithium rechargeable battery. Output: Pulsed biphasic waveform (micro-current) and 650nm (+/- 5nm) red LED light. Connectivity: USB charging. Software: Microprocessor-controlled with automatic shut-off and overload protection. Biocompatibility: ISO 10993-5, ISO 10993-10 compliant.
Indications for Use
Indicated for facial stimulation for cosmetic use and treatment of periorbital wrinkles in over-the-counter users.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Predicate Devices
Beautiful Image Model 900 Facial Toning Device (K130045)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is an abstract symbol that resembles a stylized human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 3, 2017
Li-Tek Electronic Technology Corporation % Jet Li Regulation manager Guangzhou LETA Testing Technology Co., Ltd 6F, No.1 TianTai Road, Science City, LuoGang District Guangzhou, CN Guangdong
Re: K162106
Trade/Device Name: Micro-current Wrinkle Reduction Facial Service, Model EP-400 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NFO Dated: December 31, 2016 Received: January 9, 2017
Dear Jet Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{2}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Hoffmann -S
for
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162106
#### Device Name
Micro-current Wrinkle Reduction Facial service, Model EP-400
Indications for Use (Describe)
The device is intended for facial stimulation by electrode heads for cosmetic use. The device is also intended for the treatment of periorbital wrinkles with red Light Emitting Diode (LED) head. It is for over-the-counter use.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary K162106
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
#### 510k preparation date: 2017-01-25
#### 1. Submitter's Information
510(k) Owner's Name: Li-Tek Electronics Technologies Establishment Registration Number: Address: No.8~13,the industrial park of Jinshagang Shixia village, Dalangtown, Dongguan city, Guangdong, China
Phone: 0769-83117755 Fax: 0769-83117759 Contact Person: Barry Yuan (Quality Director) E-mail: quality5@li-tek.com
#### Application Correspondent:
Company: Guangzhou LETA Testing Technology Co., Ltd. Address: 6F, No.1 TianTai road Science City, LuoGang District, GuangZhou City, China Contact Person: Mr. Jet Li Tile: Regulation Manager Tel: +86-20-22325619 Email: med-jl@foxmail.com
#### 2. Subject Device Information
Type of 510(k) submission: Traditional Common Name: Laser surqical instrument for use in general and plastic surgery and in dermatology, Transcutaneous electrical nerve stimulator for pain relief.
Trade Name: Micro-current Wrinkle Reduction Facial service, model EP-400
Classification Name: Light Based Over The Counter Wrinkle Reduction; Stimulator,
{5}------------------------------------------------
Transcutaneous Electrical, Aesthetic Purposes Review Panel: General & Plastic Surgery, Neurology
Product Code: OHS, NFO
Requlation Number: 878.4810, 882.5890
Regulation Class: 2
## 3. Predicate Device Information
| Sponsor | Biosonic<br>Technologies,<br>LIC. | EVERYWAY<br>MEDICAL<br>INSTRUMENT<br>CO., LTD. | Carol Cole<br>Company | Nutra Luxe MD,<br>LLC |
|----------------------|------------------------------------------------------|------------------------------------------------|-----------------------|-----------------------|
| Device<br>Name | Beautiful Image<br>Model 900 Facial<br>Toning Device | MT-200 Facial<br>MENS | NOFACE® Plus | Nutra Light Red |
| 510(k)<br>Number | K130045 | K142794 | K103472 | K141308 |
| Product<br>Code | NFO | NFO | NFO | OHS |
| Regulation<br>Number | 882.5890 | 882.5890 | 882.5890 | 878.4810 |
| Regulation<br>Class | 2 | 2 | 2 | 2 |
#### 4. Device Description
The Device is equipped with three replaceable treatment heads; two treatment heads are electrode heads for micro current output to stimulate facial skin (Bigger one, and smaller one), the other treatment head is LED lamp holder.
The device is intended for facial stimulation by electrode heads for cosmetic use. The device is also intended for the treatment of periorbital wrinkles with red Light Emitting Diode (LED) head. It is for over-the-counter use.
For two micro current treatment heads of instrument, they can be exchanged for different skin part to use. The bigger one is for face tissue on cheeks and the smaller one is for facial tissue below eye region. It should be used about 10 minutes/2 to 3 times per week. The red is for red-light caring (about 3 minutes/2 to 3 times per week.)The product is supplied with an internal lithium rechargeable battery; It can be charged by 5V voltage of battery charger connected by standard USB cable. During battery charging, the LED lights start flash circling until charging is completed.
{6}------------------------------------------------
There are two kinds of electrode heads, one bigger electrode head is for cheek stimulation; and smaller electrode head is for stimulation on under eyes area. There are no electronic component inside of either heads. It is only for conduct stimulation current.
## 5. Intended Use / Indications for Use
The device is intended for facial stimulation by electrode heads for cosmetic use. The device is also intended for the treatment of periorbital wrinkles with red Light Emitting Diode (LED) head. It is for over-the-counter use.
## 6. Test Summary
EP-400 has been evaluated the safety and performance by lab bench testing as following:
- � Electrical safety test according to IEC 60601-1 and IEC 60601-2-10 standards
- � Electromagnetic compatibility test according to IEC 60601-1-2 standard
- � Photobiological safety of lamps and lamp systems according to IEC62471
- � Software verification and validation test according to the requirements of the FDA "Guidance for Pre Market Submissions and for Software Contained in Medical Devices"
- � Waveform test report to verify the output specifications of the device according to IEC 60601-2-10 and Guidance for Powered Muscle Stimulator.
## 7. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, materials, mode of operation, and intended use of EP-400 is substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety and effectiveness.
| Elements<br>of<br>Comparison | Subject<br>Device | Predicate Device | Predicate Device | Predicate Device | Predicate<br>Device | Remark | |
|-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|----|
| Basic Unit Characteristics | | | | | | | |
| Device Name<br>and Model | Micro-current<br>Wrinkle<br>Reduction<br>Facial Service,<br>model EP-400 | Beautiful<br>Image<br>Model 900<br>Facial<br>Toning Device | MT-200<br>MENS | NOFACE® Plus | Nutra Light Red | -- | |
| 510<br>(K)<br>Number | K162106 | K130045 | K142794 | K103472 | K141308 | -- | |
| Product Code | NFO, OHS | NFO | NFO | NFO | OHS | -- | |
| Regulation<br>Number | 882.5890<br>878.4810 | 882.5890 | 882.5890 | 882.5890 | 878.4810 | -- | |
| Elements<br>of<br>Comparison | Subject<br>Device | Predicate Device | Predicate Device | Predicate Device | Predicate<br>Device | Remark | |
| | | | | | | | |
| Intended Use | The device is<br>intended for<br>facial<br>stimulation by<br>electrode<br>heads for<br>cosmetic use.<br>The device is<br>also intended<br>for the<br>treatment of<br>periorbital<br>wrinkles with<br>red Light<br>Emitting Diode<br>(LED) head. It<br>is for over-the-<br>counter use. | Biosonic<br>Technologies<br>Model 900 Facial<br>Toning Device is<br>intended for facial<br>stimulation and is<br>indicated for<br>prescription<br>cosmetic use. The<br>anatomical site for<br>application of the<br>Model 900 is the<br>face. | The<br>Everyway<br>Facial<br>MENS,<br>model: MT-200 is<br>intended for facial<br>stimulation<br>and<br>indicated for over-<br>the-counter<br>cosmetic use. | The<br>NOFACE®<br>Plus Facial Toning<br>Device is intended<br>for<br>facial<br>stimulation and is<br>indicated for over-<br>the-counter<br>cosmetic use. (21<br>CFR 801 Subpart<br>C). The anatomical<br>site for application<br>of the NOFACE®<br>Plus is the face. | The Nutra Light<br>Red is a non-<br>invasive LED<br>light device is<br>intended/indicate<br>d for over- the -<br>counter use for<br>the treatment of<br>periorbital<br>wrinkles, and<br>rhytides. | SE<br>Note 1 | |
| Apply parts | For micro<br>current<br>stimulation:<br>Face;<br>For Red light:<br>periorbital | Face | Face | Face | periorbital | SE | |
| Power<br>Sources | 1200mAh<br>lithium battery | One 6V battery | 9-Volt battery | 4 rechargeable<br>batteries | Internal NI-MH<br>rechargeable<br>battery | SE<br>Note 1 | |
| Method of Line<br>Current<br>Isolation | Battery Supply<br>N/A | N/A | Type BF | N/A | N/A | SE<br>Note 1 | |
| For Micro current facial stimulation function | | | | | | | |
| Number of<br>Modes for<br>Micro current<br>stimulation | 1 | 1 | 3 (8Hz, 9Hz, 10Hz) | 1 | N/A | SE<br>Note 1 | |
| Number of<br>Channels for<br>Micro current<br>stimulation | 1 | 1 | 1 | 1 | N/A | SE | |
| Elements of<br>Comparison | Subject<br>Device | Predicate Device | Predicate Device | Predicate Device | Predicate Device | Remark | |
| -Synchronous<br>or Alternating | Alternating | N/A | Alternating | Alternating | N/A | SE Note 1 | |
| Regulated Current<br>or Regulated<br>Voltage | Regulated<br>Voltage | Both | 0-3mA (load 4kΩ) | Regulated Voltage | N/A | SE Note 1 | |
| Software/Firmware/Micropro<br>cessor control | Yes | Yes | Yes | Yes | Yes | SE | |
| Automatic<br>Overload Trip | Yes | Yes | Yes | Not required due to<br>circuit design | N/A | SE | |
| Automatic No-<br>load Trip | Yes. | Yes | Yes | Yes | N/A | SE | |
| Automatic<br>Shut Off | Yes. | Yes | Yes | Yes | N/A | SE | |
| Patient<br>Override<br>Control | Yes | Yes | N/A | Yes | N/A | SE | |
| On/<br>Off<br>Status | Yes | Yes | Yes | Yes | N/A | SE | |
| Indicator<br>or<br>Display | Low<br>Battery | Yes | Yes | Yes | Yes | N/A | SE |
| | Voltage/<br>Current<br>Level | Yes | Yes | Yes | Yes | N/A | SE |
| Timer Range | 10 minutes | None | 20, 40 minutes and<br>Continuous | 21 minutes | N/A | SE Note 1 | |
| Housing<br>Materials and<br>Construction | Console: ABS<br>plastic | Thermoplastic | ABS | Thermo Plastic | medical grade<br>biocompatibility<br>plastics via<br>injection molding | SE Note 1 | |
| Output Specification | | | | | | | |
| Elements<br>of<br>Comparison | Subject<br>Device | Predicate Device | Predicate Device | Predicate Device | Predicate<br>Device | Remark | |
| Waveform | Pulsed Biphasic | Biphasic | Biphasic | Pulsed Monoilhasic | N/A | SE<br>Note 2 | |
| Shape | Rectangular | Rectangular | Rectangular | Modulated Square | N/A | SE<br>Note 2 | |
| Maximum<br>Output<br>Voltage<br>(+/-<br>10%) | 1.23V @ 500Ω<br>3.64V @ 2kΩ<br>10.9V @ 10kΩ | 0.347V @ 500Ω<br>1.242V @ 2kΩ<br>5.780V @ 10kΩ | 1.78V @500Ω<br>6.64V @2KΩ<br>16.2V @10KΩ | 137mV @ 500Ω<br>769mV @ 2kΩ<br>3.82V @ 10kΩ | N/A | SE<br>Note 2 | |
| Maximum<br>Current<br>Density | 2.46mA @<br>500Ω<br>1.82mA @ 2kΩ<br>1.09mA @<br>10kΩ | 0.647mA @ 500Ω<br>0.625mA@ 2kΩ<br>0.584mA@ 10kΩ | 3.56mA @500Ω<br>3.32mA @2KΩ<br>1.62mA @10KΩ | 274μΑ @ 500Ω<br>387μA @ 2kΩ<br>382μΑ @ 10kΩ | N/A | SE<br>Note 2 | |
| Frequency<br>range | 59.3Hz | 0.62 1 - 308.6Hz | 8Hz, 9Hz, 10Hz | 8.40 Hz | N/A | SE<br>Note 2 | |
| Pulse<br>width<br>range | 4ms | 3.24-1610ms | N/A | 119ms | N/A | SE<br>Note 2 | |
| Pulse<br>duration | 4ms | 3.24-1610ms | ~49-63ms | N/A | N/A | SE<br>Note 2 | |
| Net Charge | 9.04μC@ 500Ω | 0 uC @5000 | N/A | N/A | N/A | SE<br>Note 2 | |
| Maximum<br>Phase<br>Charge | 16.4µC@ 500Ω | 190µC@ 500Ω | 157.12uC<br>(8Hz/500Ω) | 23.06µC@ 500Ω | N/A | SE<br>Note 2 | |
| Maximum<br>Current<br>Density | Bigger electrode<br>head:<br>0.044 mA/cm²@<br>500Ω<br>Smaller<br>electrode head:<br>0.42 mA/cm²@<br>500Ω | 1.486mA/cm²@<br>500Ω | 0.26mA/cm²<br>(8Hz/500Ω) | 0.419<br>mA/cm²@<br>500Ω | N/A | SE<br>Note 2 | |
| Maximum<br>Power<br>Density | Bigger<br>electrode head<br>0.014 mW/<br>cm²@ 500Ω<br>Smaller<br>electrode head<br>0.139 mW/<br>cm²@ 500Ω | 366 µW/cm² @<br>500Ω | 1.33mW/cm²<br>(8Hz/10kΩ) | 3.22 µW/cm² @<br>500Ω | - N/A | SE<br>Note 2 | |
| ON time | Constant | 10-30s | Constant | Constant | N/A | SE | |
| OFF time | None | 1-6s | None | None | N/A | SE | |
| Elements of<br>Comparison | Subject<br>Device | Predicate Device | Predicate Device | Predicate Device | Predicate<br>Device | Remark | |
| Contraction and<br>Relaxation time | Adjustable, due to different modes. | Adjustable, due to different modes. | -- | -- | N/A | SE | |
| For LED red light irradiation function | | | | | | | |
| LED<br>wavelength | 650+/-5nm | -- | -- | -- | 650 +/- 5nm | SE | |
| LED Power<br>Density | 80mW/cm² | -- | -- | -- | 80mW/cm² | SE | |
| Additional Features | | | | | | | |
| Environment<br>for operating | Temperature: 5 ~ 40° C | + 50 to 1040 F<br>(+10 to +40° C) | -- | -- | -- | SE | |
| Environment<br>for storage | Temperature: - 25 ~70° C<br>Humidity: 10 ~90% RH | - 29 to + 1670 F (- 34 to +76° C)<br>0 to 95% - non-condensing | -- | -- | -- | SE | |
| Biocompatibility | All user directly contacting materials are compliance with ISO10993-5, ISO10993-10 and IEC 62471:2006 requirements. | All user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements. | All user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements. | All user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements. | All user directly contacting materials are compliance with ISO10993-5 , ISO10993-10 and IEC 62471:2006 requirements. | SE | |
| Electrical<br>Safety | Comply with IEC 60601-1 and IEC 60601-2-10 | Comply with IEC 60601-1 and IEC 60601-2-10 | Comply with IEC 60601-1 and IEC 60601-2-10 | Comply with IEC 60601-1 and IEC 60601-2-10 | Comply with IEC 60601-1 and IEC 60601-2-10 | SE | |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | SE | |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
## Comparison in Detail(s):
## Note 1:
Although the Power Sources, Method of Line Current Isolation, Synchronous or Alternating, Regulated Current or Regulated Voltage, Timer Range, Console weight, Housing Materials and
{11}------------------------------------------------
Construction are a little different from the predicate devices, they are all compliant with requirements of IEC 60601-1, IEC 60601-1-2 and Guidance for Powered Muscle Stimulator. So the differences of the function specifications do not raise any safety or effectiveness issue.
#### Note 2:
Although the Waveform. Shape, maximum output voltage, maximum current density, frequency range, pulse width range, pulse duration, net charge, maximum phase charge, maximum current density, maximum power density, on time, off time, contraction time of subject device are a little different from the predicate devices, they are all compliant with the requirements of IEC 60601-1, IEC 60601-2-10, and Guidance for Powered Muscle Stimulator. So the differences of function specification do not raise any safety or effectiveness issue.
#### Conclusion:
The subject device "Micro-current Wrinkle Reduction Facial Service, model EP-400"…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.