The P-POD Helmet is a cranial orthosis device intended for medical purposes to apply static or gentle pressure to prominent regions of an infant's cranium to improve cranial symmetry or shape. The device is intended to treat infants from four to eighteen months of age with moderate to severe non-synostotic positional plagiocephaly, including plagiocephalic-, brachycephalic-, scaphocephalic-shaped heads.
Device Story
P-POD Plagiocephaly Orthosis Device is a cranial orthosis for infants (4-18 months) with non-synostotic positional plagiocephaly. Device consists of a standardized hard outer shell (polypropylene) with an internal inflatable bladder. Clinical practitioner uses modeling putty to fill flattened regions of the infant's head; a pre-measured liquid solution is poured into the bladder via a top port. Exothermic reaction cures the liquid into a solid foam, creating a custom negative mold of the infant's skull. Device is worn by the infant to promote skull re-contouring through growth. Used in physician offices; enables immediate treatment (same-day) compared to traditional 2-4 week scanning/molding processes. Output is a custom-fitted helmet; clinical benefit is improved cranial symmetry. Effectiveness monitored by practitioner.
Clinical Evidence
Bench testing only. Included biocompatibility testing (ISO 10993-5 cytotoxicity, ISO 10993-10 sensitization and irritation) showing non-toxic, non-sensitizing, and non-irritating results. Human factors studies validated the on-site casting process, confirming naive users can produce the helmet and infants tolerate the treatment.
Technological Characteristics
Cranial orthosis; hard polypropylene outer shell (USP Class VI); polyurethane bladder/liner with polyurethane foam. Dimensions: standardized (2 sizes). Energy: exothermic chemical reaction for foam curing. Connectivity: none. Sterilization: N/A. Software: none.
Indications for Use
Indicated for infants 4-18 months old with moderate to severe non-synostotic positional plagiocephaly (plagiocephalic, brachycephalic, or scaphocephalic head shapes) to improve cranial symmetry or shape via static/gentle pressure.
Regulatory Classification
Identification
A cranial orthosis is a device that is intended for medical purposes to apply pressure to prominent regions of an infant's cranium in order to improve cranial symmetry and/or shape in infants from 3 to 18 months of age, with moderate to severe nonsynostotic positional plagiocephaly, including infants with plagiocephalic-, brachycephalic-, and scaphocephalic-shaped heads.
Special Controls
*Classification.* Class II (special controls) (prescription use in accordance with § 801.109 of this chapter, biocompatibility testing, and labeling (contraindications, warnings, precautions, adverse events, instructions for physicians and parents)).
In addition to the general controls of the act, the Dynamic Orthotic Cranioplasty - DOC™ Band is subject to the following special controls in order to provide reasonable assurance of the safety and effectiveness: (1) The sale, distribution and use of this device are restricted to prescription use in accordance with 21 CFR 801.109. (2) The labeling must include (a) contraindications for the use of the device on infants with synostosis or with hydrocephalus; (b) warnings indicating the need: (i) to evaluate head circumference measurements and neurological status at intervals appropriate to the infant’s age and rate of head growth, and to describe steps that should be taken in order to reduce the potential for restriction of cranial growth and possible impairment of brain growth and development; (ii) to evaluate the skin at frequent intervals, e.g., every three to four hours, and to describe steps that should be taken if skin irritation or breakdown occurs; (c) precautions indicating the need: (i) to additionally treat torticollis, if the positional plagiocephaly is associated with torticollis; (ii) to evaluate device fit and to describe the steps that should be taken in order to reduce the potential for restriction of cranial growth, possible impairment of brain growth and development and skin irritation and/or breakdown; (iii) to evaluate the structural integrity of the device and to describe the steps that should be taken to reduce the potential for the device to slip out of place and cause asphyxiation or trauma to the eyes or skin; (d) adverse events, i.e., skin irritation and breakdown that have occurred with the use of this device; (e) clinician’s instructions for casting the infant, for fitting the device, and for care and use of the device; and (f) parents’ instructions for care and use of the device. (3) The materials must be assessed for biocompatibility with testing appropriate for long term direct skin contact.
{0}------------------------------------------------
#### APR 1 8 2014 510(k) Summary of Safety and Effectiveness
In accordance with 21 CFR 807.87 (h) and 21 CRF 807.92, the 510(k) summary for the P-POD Plagiocephaly Orthosis Device is provided below.
| Date | 4/15/2014 |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer/Distributor/Sponsor | Lorica Scientific LLC<br>750 Old Ludlow Ave.<br>Cincinnati, OH 45220<br>Phone 440-315-7830<br>Fax 513-221-2905 |
| 510(k) Contact | Secure BioMed Evaluations<br>Linda Braddon, Ph.D.<br>7828 Hickory Flat Highway<br>Suite 120<br>Woodstock, GA 30188<br>770-837-2681 (direct)<br>855-MED-DEV1 (office)<br>LGB@SecureBME.com |
| Trade Name | Plagiocephaly Orthosis Device |
| Common Name | Cranial Orthosis |
| Code Classification | MVA 21 CFR 882.5970 : Class II |
| Predicate Devices | K072566 Hanger Cranial Band™<br>K021918 Clarren Helmet |
#### Device Description
The Lorica Scientific LLC P-POD Plagiocephaly Helmet is a Class II cranial orthosis intended for medical purposes to apply static or gentle pressure to prominent regions of an infant's cranium to improve cranial symmetry or shape. The device is intended to treat infants from four to eighteen months of age with moderate to severe non-synostotic positional plagiocephaly, including plagiocephalic-, brachycephalic-, scaphocephalic-shaped heads. The P-POD helmet is similar to the predicate devices in such that it is an orthosis designed for each patient from a cast of the infant's head. Each orthosis is composed on an outer shell with a layer of foam and a side strap for securing the orthosis. Additionally, the helmet has a top vent and side opening as typically seen with the predicates. Optimum fit and alignment is insured and monitored by the clinical practitioner.
The Lorica Scientific LLC P-POD Plagiocephaly Helmet differs from the predicate devices in that the device is made in the physician's office via a simplified casting process that can be performed in approximately 30 minutes. The predicate devices rely on a 2 to 4 week process of sending the child for a 3-D scan and then using casting and molding processes to create a customized helmet.
The P-POD standardized (2 sizes) helmet consists of a hard outer shell with an inflatable bladder lining the inside. Modeling putty is used to fill in the flattened portion of the infants head to form the desired
{1}------------------------------------------------
symmetrical shape and the helmet is placed on the infant. The physician then mixes a pre-measured solution in an easy-to-use, pre-measured pouch and pours the mixed solution into the bladder of the helmet through a specially designed filling port located at the top of the hemical solution cures, there is an exothermic foaming process from a liquid to solid foam which expands to fill the empty space in the bladder thus customizing the helmet to the shape making a negative copy of the infant's skull. Since the putty is used to fill in the undesired negative regions of the skull deformity, once the foam is completely cured and putty removed, the resulting helmet provides an ideal cast to help promote proper skull re-contouring. As with all other similar cranial orthosis devices, as the infant wears the helmet, the head grows into the shape formed by the foam, thereby correcting the deformity.
#### Intended Use
The P-POD Helmet is a cranial orthosis device intended for medical purposes to apply static or gentle pressure to prominent regions of an infant's cranium to improve cranial symmetry or shape. The device is intended to treat infants from four to eighteen months of age with moderate to severe non-synostotic positional plagiocephaly, including plagiocephalic-, brachycephalic-, scaphocephalic-shaped heads.
#### Technological Characteristics
The P-POD helmet is essentially the same as the predicate devices in such that it is an orthosis designed for each patient from a cast of the infant's head. Each orthosis is composed on an outer shell with a layer of foam and a strap for securing the orthosis. It has a top vent and side opening. Optimum fit and alignment is insured and monitored by the clinical practitioner.
The Lorica Scientific LLC P-POD Plagiocephaly Helmet differs from the predicate devices in that the device is made in the physician's office via a simplified casting process that can be performed in approximately 30 minutes.
#### Non-Clinical Performance Testing Conclusion
Non-clinical performance testing included biocompatibility with the following results:
| Biocompatibility Tests | Results | Conclusions |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|------------------------------|
| ISO Cytotoxicity MEM Elution<br>According to ISO 10993-5 Biological evaluation of<br>medical devices: Part 5 Tests for In vitro Cytotoxicity | Cell culture treated with test sample<br>exhibited no reactivity (Grade O) | Non-toxic |
| Guinea Pig Maximization<br>According to ISO 10993-10 Biological evaluation of<br>medical devices: Part 10 Tests for irritation and<br>delayed hypersensitivity | Albino guinea pigs treated with test<br>sample did not elicit a sensitization<br>response (Grade O) | No sensitization<br>reaction |
| Intracutaneous Irritation Reactivity<br>According to ISO 10993-10 Biological evaluation of<br>medical devices: Part 10 Tests for irritation and<br>delayed hypersensitivity | Rabbits treated with test samples<br>exhibited no irritation (Grade 0) | Non-irritating |
{2}------------------------------------------------
- Evaluation of molding process accuracy ●
- . Foam stiffness
Human factors studies were performed to ensure the following:
- . Naive users can be trained via a video and reading the IFU to produce a helmet and י appropriately accept or reject a helmet based on defined criteria
- . The infants in the intended treatment range of 4 to 18 months can tolerate the treatment
## Substantial Equivalence Summary (Conclusion)
The Lorica Scientific LLC P-POD Helmet is very similar to cranial orthosis devices that are legally commercially available. A comparison between the P-POD Helmet and the predicate devices is shown in the following table.
| Trait | P-POD<br>Plagiocephaly Orthosis<br>Device | Clarren | Hanger Cranial<br>BandTM | Evaluation of<br>Differences |
|----------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | TBD | K021918 | K072566 | N/A |
| Product<br>Classification | Class II<br>882.5970<br>MVA | Class II<br>882.5970<br>MVA | Class II<br>882.5970<br>MVA | Same |
| Use | Prescription Use<br>Part 21 CFR 801 Subpart D | Prescription Use<br>Part 21 CFR 801<br>Subpart D | Prescription Use<br>Part 21 CFR 801<br>Subpart D | Same |
| Intended<br>Population | 4 to 18 months | 3 to 18 months | 3 to 18 months | No risk for change;<br>P-POD more<br>conservative |
| Intended Use | See section 12.2.1 | See section 12.2.1 | See section 12.2.1 | Same |
| Product Design | Cranial orthosis made to<br>individual's specifications | Cranial orthosis<br>made to individual's<br>specifications | Cranial orthosis<br>made to<br>individual's<br>specifications | Same |
| Biocompatible<br>Components | Yes | Yes | Yes | Same |
| Materials:<br>Outer Shell | Polypropylene<br>USP Class VI certified | Polypropylene<br>customized to<br>individual | Polypropylene or<br>Polypropylene-<br>Polyethylene<br>Copolymer | Same or Equivalent |
| Materials:<br>Bladder / Liner | Polyurethane liner filled<br>with polyurethane foam | Polyurethane | Polyethylene foam | Same or Equivalent |
| Trait | P-POD<br>Plagiocephaly Orthosis<br>Device | Clarren | Hanger Cranial<br>BandTM | Evaluation of<br>Differences |
| Helmet<br>Production | Casting<br>Manufactured by Physician<br>in doctor's office on infant | Computer scan,<br>Casting<br>Manufactured by<br>Orthotist | Computer scan,<br>Casting<br>Manufactured by<br>Orthotist | No New Risk<br>Differences in<br>manufacturing<br>helmets will not<br>affect quality of<br>final product.<br>Human factors<br>studies for P-POD<br>helmet show<br>physicians can make<br>the helmet and<br>determine<br>adequacy of helmet<br>for child. |
| Foam Stiffness | Durometer A<br>$45 \pm 4.61$<br>Min: 39<br>Max: 54 | Durometer A<br>$64 \pm 6.71$<br>Min: 55<br>Max: 74 | Durometer A<br>$50 \pm 9.36$<br>Min: 39<br>Max: 65 | Equivalent |
| Foam<br>Thickness | 0.1875 inches minimum | 0.1875 inches | Not measured | Same |
| Daily Wearing<br>Time | 23 hours | 23 hours | 23 hours | Same |
| Daily Care | Cleaning daily with water<br>and isopropyl alcohol | Cleaning daily with<br>water and isopropyl<br>alcohol | Cleaning daily with<br>water and<br>isopropyl alcohol | Same |
| Time from<br>initial<br>evaluation to<br>application of<br>treatment | Same day;<br>Helmet is made onsite at<br>the physician's office. As<br>soon as a clinical need is<br>determined, the treatment<br>can start immediately | Typically 2 to 4<br>weeks delay from<br>diagnosis to<br>beginning of<br>treatment | Typically 2 to 4<br>weeks delay from<br>diagnosis to<br>beginning of<br>treatment | P-POD allows the<br>immediate<br>treatment of a<br>diagnosed condition<br>whereas predicates<br>delay treatment for<br>weeks |
| Adverse<br>Effects | Device may cause skin<br>irritations or breakdown | Device may cause<br>skin irritations or<br>breakdown | Device may cause<br>skin irritations or<br>breakdown | Same |
| Caregiver<br>Instructions<br>for Use | Wear and care guide<br>provided to caregiver | Wear and care guide<br>provided to<br>caregiver | Wear and care<br>guide provided to<br>caregiver | Same |
| Discontinuanc<br>e of Device<br>Use | When infant outgrows the<br>cranial helmet or orthosis<br>is discontinued for any<br>reason | When infant<br>outgrows the cranial<br>helmet or orthosis is<br>discontinued for any<br>reason | When infant<br>outgrows the<br>cranial helmet or<br>orthosis is<br>discontinued for<br>any reason | Same |
{3}------------------------------------------------
This submission demonstrates the equivalency in the indication for use, device classification, product code, environment of use, and the equivalency of the principles of operation. The Lorica Scientific LLC P-POD
{4}------------------------------------------------
Plagiocephaly Orthosis Device and the predicates underwent non-clinical evaluation which confirmed device equivalency. Additionally, the change in the helmet process was validated in a two part human factors study which showed the P-POD device production process can be successfully produced by clinicians and is well tolerated by the intended infant population.
·
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract representation of human figures or a bird in flight.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 18, 2014
P-POD c/o Linda Braddon, Ph.D. Secure BioMed Evaluation 7828 Hickory Flat Highway, Suite 120 Woodstock, Georgia 30188
Re: K133397
Trade/Device Name: P-POD Plagiocephaly Orthosis Device Regulation Number: 21 CFR 882.5970 Regulation Name: Plagiocephaly Orthosis Device Regulatory Class: Class II Product Code: MVA Dated: March 17, 2014 Received: March 20, 2014
Dear Dr. Braddon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{6}------------------------------------------------
Page 2 - Linda Braddon, Ph.D.
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Felipe Aquel -S
Carlos L. Peña, Ph.D, M.S. for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K133397
#### Device Name
P-POD Plagiocephaly Orthosis Device
#### Indications for Use (Describe)
The P-POD Plagiocephaly Orthosis Device is intended for medical purposes to apply static or gentle presions of an infant's cranium to improve cranial symmery or shape. The device is infants from four to eighteen months of age with moderate to severe non-synostolic plagiscephalic-, brachycephalic-, brachycephalic-, scaphocephalic-shaped heads.
Type of Use (Select one or both, as applicable)
[ Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Date: 2014.04.18 10:59:53 -04'00'
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
{8}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.