The Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities, such as arteriovenous malformations and arteriovenous fistulae. The Axium Detachable Coils are also indicated for arterial and venous embolizations in the peripheral vasculature.
Device Story
Axium Detachable Coil System; platinum/tungsten embolization coil attached to composite implant delivery pusher; includes radiopaque positioning marker; used with hand-held Axium I.D. (Instant Detacher) for coil release. Device used in neurovascular/peripheral procedures; operated by physicians (interventionalists). Coil navigated through microcatheter to target site (aneurysm/malformation); activated via I.D. to detach coil from pusher. Provides mechanical occlusion of vascular abnormalities; promotes embolization. Line extension adds smaller coil diameters (1mm) and replaces PET shrink tubing indicators with laser marks on pusher. Benefits include expanded size options for varied vascular anatomy; maintains safety/effectiveness profile of predicate.
Clinical Evidence
No clinical or animal testing performed. Substantial equivalence established via non-clinical bench testing, including visual/dimensional inspection, coil deformation, force transfer, friction, fatigue, knotting, tensile strength, and physician usability testing in tortuous benchtop models.
Indicated for endovascular embolization of intracranial aneurysms, arteriovenous malformations, arteriovenous fistulae, and arterial/venous embolizations in peripheral vasculature.
Regulatory Classification
Identification
A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of a human figure, represented by three overlapping profiles facing to the right. The profiles are rendered in black and are arranged to create a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 28, 2015
Micro Therapeutics, Inc. d/b/a ev3 Neurovascular Ms. Patricia Casing Regulatory Affairs Specialist 9775 Toledo Way Irvine, California 92618
Re: K151447
Trade/Device Name: Axium™ Detachable Coil System Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: Class II Product Code: HCG and KRD Dated: June 30, 2015 Received: Julv 1, 2015
Dear Ms. Casing:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena நி/Δ
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K151447
#### Device Name Axium Detachable Coil System
#### Indications for Use (Describe)
The Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities, such as arteriovenous malformations and arteriovenous fistulae. The Axium Detachable Coils are also indicated for arterial and venous embolizations in the peripheral vasculature.
Type of Use (Select one or both, as applicable)
| <span style="font-size:12px">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size:12px">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
| 510(k) Owner: | Micro Therapeutics, Inc. d/b/a ev3 Neurovascular<br>9775 Toledo Way<br>Irvine, CA 92618<br>Establishment Registration No. 2029214 | |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Contact Person: | Patricia Casing<br>Regulatory Affairs Specialist<br>Telephone: (949) 680 - 1307<br>E-mail: patricia.casing@covidien.com | |
| Date Summary<br>Prepared: | June 30, 2015 | |
| Trade Name of<br>Device: | Axium™ Detachable Coil System | |
| Common Name of<br>Device: | Neurovascular Embolization Device | |
| Classification of<br>Device: | 21 CFR 882.5950 - Class II | |
| Product Code | HCG and KRD | |
| Predicate Device: | Axium™ Detachable Coil System:<br>K133310, cleared January 10, 2014<br>●<br>K081465, cleared August 19, 2008<br>●<br>K060747, cleared April 24, 2007<br>● | |
| Performance Data: | The following bench tests were performed to support the addition of<br>new sizes to the Axium coil line and to establish substantial<br>equivalence to predicate Axium™ Detachable Coil System : | |
| | Dimensions - Coil, Sheath, Implant Delivery Pusher<br>●<br>Visual Inspections - Coil, Sheath, Implant Delivery<br>●<br>Pusher<br>Friction<br>●<br>Force Transfer<br>●<br>Coil Deformation<br>●<br>Fatigue and Knotting<br>●<br>Coil Tensile – Polypropylene<br>●<br>Coil Tensile - Implant/Weld (Implant Coil to Coil<br>●<br>Shell Weld)<br>Labeling verification<br>●<br>The following testing was adopted from existing test data for currently<br>cleared Axium™ coil sizes: | |
| | Torque Response<br>●<br>Coil Tensile - Assembly<br>● | |
- Pusher Dimensions
- Marker Radiopacity ●
- Tip Buckling
- . Detachment Zone Stiffness
{5}------------------------------------------------
- Kink Resistance ●
- Hypotube and Weld Tensile Strength ●
- Pusher Elongation ●
- MRI Compatibility
Sterilization, biocompatibility and aging data were also adopted from the predicate device as there is no change to the manufacturing process or packaging for the addition of these new SKUs.
In addition, no clinical or animal testing was performed as the subject device has a similar safety profile and a similar effectiveness profile when compared to the predicate device.
In addition, there is no change in the indications for use or the fundamental scientific technology of the device that would require more extensive testing.
- Conclusion: The thirty-five (35) new Axium™ Detachable Coil System size offerings are substantially equivalent to the currently cleared Axium™ Detachable Coil System based on the successful completion of nonclinical testing; a thorough assessment of existing test data; as well as identical materials of coil implant construction, principles of operation, packaging and indications for use.
### Device Description:
The Axium™ Detachable Coil System consists of a platinum/tungsten embolization coil attached to a composite implant delivery pusher with a radiopaque positioning marker and a hand-held Axium™ I.D. (Instant Detacher) which, when activated, detaches the coil from the delivery pusher tip. The Axium™ I.D. (Instant Detacher) is sold separately.
This submission expands the size offerings of the Axium™ Detachable Coil (subject device) by adding thirty-five (35) new model numbers to the currently cleared Axium product portfolio (predicate device). These new model numbers are a line extension of the currently sold predicate device and range in size from 1mm diameter x 1cm length to 3.5mm diameter x 10cm length. All thirtv-five (35) new SKUs have a smaller primary wire diameter than the predicate device. This line extension includes six (6) coils having a 1mm diameter which is outside of the currently cleared Axium size range of 1.5mm diameter x 1cm length to 25mm diameter x 50cm length.
The second modification features a change to the Implant Delivery Pusher by replacing the PET material of the Positive Load Indicator (PLI) and Hypotube Break Indicator (BI) with a laser mark.
These modifications to the currently cleared predicate device were made solely to accommodate the new smaller implant coil size of the subject line extension. All other aspects of the subject device (materials of construction of the implant coil, principles of operation, packaging, labeling and indications for use) are identical to the predicate device.
{6}------------------------------------------------
### Indications for Use:
The Axium™ Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities, such as arteriovenous malformations and arteriovenous fistulae. The Axium™ Detachable Coils are also indicated for arterial and venous embolizations in the peripheral vasculature.
### Device Comparison:
The table below provides a comparison of the technological characteristics of the subject Axium device line extension and the currently cleared predicate Axium device line.
| Characteristics | Predicate Device | Subject Device | Rationale for<br>Difference (if<br>applicable) |
|-------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | Axium™ (K133310, (K081465,<br>K060747) | Axium Line Extension | N/A |
| | The Axium Detachable Coil<br>System is indicated for the<br>endovascular embolization of<br>intracranial aneurysms and other<br>neurovascular abnormalities,<br>such as arteriovenous<br>malformations and arteriovenous<br>fistulae. The Axium Detachable<br>Coils are also indicated for<br>arterial and venous<br>embolizations in the peripheral<br>vasculature. | Same | |
| Method of Supply | Stored within dispenser coil,<br>Tyvek pouch, & shipping carton | Same | N/A |
| Sterilization Method | Ethylene Oxide | Same | N/A |
| Device Size Range<br>(Coil Diameter x Coil<br>Length) | 1.5mm x 1cm – 25mm x 50cm | 1mm x 1cm - 3.5mm x<br>10cm | Difference in size is<br>due to the smaller<br>diameter primary<br>wire. Bench testing<br>has demonstrated that<br>the smaller<br>dimensions do not<br>affect the safety and<br>effectiveness of the<br>device. |
| Materials<br>(Implant Coil) | | | |
| Characteristics | Predicate Device | Subject Device | Rationale for<br>Difference (if<br>applicable) |
| Implant Coil Wire | Axium™ (K133310, (K081465,<br>K060747) | Axium Line Extension | N/A |
| | Pt (92%)/W (8%) | Same | |
| Characteristics | Predicate Device | Subject Device | Rationale for Difference (if applicable) |
| Stretch Resistant Member | Axium™ (K133310, (K081465, K060747) | Axium Line Extension | N/A |
| Coil Shell | Pt (92%)/ W (8%) | Same | N/A |
| Detach Subassembly | SS 316LVM | Same | N/A |
| Materials<br>(Implant Delivery Pusher) | | | |
| Characteristics | Predicate Device | Subject Device | Rationale for Difference (if applicable) |
| Unibody | Axium™ (K133310, K081465, K060747) | Axium Line Extension | N/A |
| Outer Jacket | SS 304 | Same | N/A |
| Marker Coil | PTFE | Same | N/A |
| Lumen Stop | Pt (92%)/ W (8%) | Same | N/A |
| Inner Liner | SS 304 | Same | N/A |
| Coupler Tube | PTFE | Same | N/A |
| Actuator Interface | SS 304 | Same | N/A |
| Release Wire | SS 304 | Same | N/A |
| Retainer Ring | SS 304 | Same | N/A |
| Break Indicator | SS 304 | Same | N/A |
| Positive Load Indicator | PET shrink tubing | Laser mark | Bench testing confirmed that the change in indicator design does not affect the safety and effectiveness of the device |
| | PET shrink tubing | Laser mark | Bench testing confirmed that the change in indicator design does not affect the safety and effectiveness of the device |
| Materials<br>(Introducer sheath) | | | |
| Introducer sheath | Polypropylene/HDPE | Same | N/A |
{7}------------------------------------------------
### Sterilization and Shelf Life
The subject device was adopted into the EO sterilization cycle originally cleared under K060747 for the predicate device. The manufacturing process and packaging are identical to the predicate device. The smaller dimensions and the replacement of the PET shrink
{8}------------------------------------------------
tubing with laser marking for the PLI and BI indicators on the Implant Delivery Pusher result in a reduction of mass/material.
EO residual testing for the subject device was adopted, as the materials of construction of the implant coil and the manufacturing process are identical to the currently cleared predicate device. In addition, the smaller dimensions and the replacement of the PET shrink tubing with laser marking for the PLI and BI indicators on the Implant Delivery Pusher result in a reduction of mass/material.
Aging and packaging studies for the predicate device have established the product and packaging remain functional and maintain sterility for 3 years. Aging studies for packaging integrity (per ASTM F2096-11) and conditioning (ASTM-4169, ISTA-2A) and device functionality were adopted from the predicate device and met all acceptance criteria. The materials of implant coil construction, manufacturing process, and packaging of the new smaller sized models of the subject device are identical to the predicate device.
### Biocompatibility
Biocompatibility data for the subject device has been adopted from the predicate device. Physicochemical and cytotoxicity testing were performed on the laser marks of the Implant Delivery Pusher for reference only. Passing results provide confirmation that the laser marks do not impact the safety of the device and add justification for adopting previous biocompatibility studies conducted for the Axium coil predicate device.
#### Performance Data – Bench
A summary of the non-clinical bench testing performed for the subject device is presented in the table below:
| Test | Test Method | Conclusion |
|--------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|
| Visual and Dimensional<br>Inspection:<br>• Coil<br>• Sheath<br>• Implant Delivery Pusher | Dimensions (ID, OD, length) were<br>measured and key characteristic<br>were inspected of the implant coil,<br>sheath and coupler tube | All devices met<br>acceptance criteria. |
| Softness and Conformability<br>• Coil Deformation | The first loop of the coil is advanced<br>until it exits the sheath and is<br>compressed to a deformation<br>distance that is a percentage of the<br>coil's loop OD. The peak force is<br>recorded. | All devices met<br>acceptance criteria. |
| Ease of Deliverability<br>• Force Transfer<br>• Friction Testing<br>• Fatigue and Knotting | • Peak delivery force was<br>measured through a<br>representative tortuous<br>anatomical model.<br>• The proximal end of the device<br>is advanced until the distal force | All devices met<br>acceptance criteria. |
{9}------------------------------------------------
| Test | Test Method | Conclusion |
|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|
| | exceeds a specified value. The<br>force transfer is calculated.<br>• The device is placed inside the<br>microcatheter and advanced<br>until the coil is deployed<br>completely inside the aneurysm<br>model. The device is retracted<br>and the cycle is repeated. After<br>the required number of cycles<br>the device is inspected. | |
| Detachment<br>• Coil Tensile –<br>Polypropylene<br>• Coil Tensile –<br>Implant/Weld (Implant<br>Coil to Coil Shell Weld) | • The coil is stretched until the<br>stretch resistant member breaks.<br>The peak force result was<br>recorded<br>• The primary wire of the coil is<br>stretched until the coil-coil shell<br>weld breaks. The peak force<br>result was recorded | All devices met<br>acceptance criteria. |
| Labeling Verification | Text and format of drawings were<br>visually compared to labeling and<br>packaging product specifications. | All devices met<br>acceptance criteria. |
| Physician Usability Testing | The device was navigated through a<br>tortuous benchtop model and<br>deployed into a simulated silicone<br>aneurysm in order to assess coil<br>softness, conformability, ease of<br>delivery and biomechanical<br>stability | All test results met the<br>acceptance criteria. |
### Performance Data – Animal
No animal study was performed as there is no change to the indications for use or the fundamental scientific technology for the new model numbers. Substantial equivalence of the subject device has been established to the predicate device through the results of the bench testing that was performed as well as the testing that was adopted from the predicate device.
### Performance Testing - Clinical
No clinical study was performed as there is no change to the indications for use or the fundamental scientific technology for the new model numbers. Substantial equivalence of the subject device has been established to the predicate device through the results of the bench testing that was performed as well as the testing that was adopted from the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.