BINAXNOW INFLUENZA A & B TEST

K041049 · Binax, Inc. · PSZ · Aug 10, 2004 · Microbiology

Device Facts

Record IDK041049
Device NameBINAXNOW INFLUENZA A & B TEST
ApplicantBinax, Inc.
Product CodePSZ · Microbiology
Decision DateAug 10, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3328
Device ClassClass 2
AttributesPediatric

Intended Use

The BinaxNOW® Influenza A & B Test is an in vitro immunochromatographic assay for the qualitative detection of influenza A and B nucleoprotein antigens in nasopharyngeal swab and nasal wash/aspirate specimens. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. Negative test results should be confirmed by culture.

Device Story

The BinaxNOW® Influenza A & B Test is an immunochromatographic membrane assay used in clinical settings to detect influenza A and B nucleoprotein antigens. Input consists of nasopharyngeal swabs or nasal wash/aspirate specimens. The sample is eluted into a solution and applied to a test strip housed in a book-shaped hinged cardboard device. The strip contains immobilized monoclonal antibodies that capture viral antigens; a blue control line confirms assay validity. Results are interpreted visually by a clinician or technician at 15 minutes based on the presence of pink-to-purple sample lines. The test provides rapid differential diagnosis, aiding clinical decision-making for patient management. Visibly bloody samples are contraindicated due to potential interference.

Clinical Evidence

Clinical performance was evaluated in a multi-site prospective study and retrospective frozen sample analysis. The study included 306 retrospective samples for Flu A and 303 for Flu B, representing diverse patient demographics (64% pediatric, 36% adult). Compared to individual predicate tests, the device demonstrated 100% sensitivity and 96% specificity for Influenza A, and 93% sensitivity and 97% specificity for Influenza B. Reproducibility studies across three sites showed 97% agreement. Analytical sensitivity (LOD) was established using inactivated viral strains. No clinical data regarding patient outcomes was provided; performance is based on diagnostic accuracy against predicate devices and viral culture.

Technological Characteristics

Immunochromatographic membrane assay; utilizes monoclonal antibodies immobilized on a membrane support. Form factor is a book-shaped hinged cardboard test device. No external energy source required. Standalone diagnostic test. Materials include membrane support, reagents, and pads. Sterilization method not specified. Software: None.

Indications for Use

Indicated for the qualitative detection of influenza A and B nucleoprotein antigens in nasopharyngeal swab and nasal wash/aspirate specimens to aid in the rapid differential diagnosis of influenza A and B viral infections in symptomatic patients. Negative results should be confirmed by culture.

Regulatory Classification

Identification

An influenza virus antigen detection test system is a device intended for the qualitative detection of influenza viral antigens directly from clinical specimens in patients with signs and symptoms of respiratory infection. The test aids in the diagnosis of influenza infection and provides epidemiological information on influenza. Due to the propensity of the virus to mutate, new strains emerge over time which may potentially affect the performance of these devices. Because influenza is highly contagious and may lead to an acute respiratory tract infection causing severe illness and even death, the accuracy of these devices has serious public health implications.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The device's sensitivity and specificity performance characteristics or positive percent agreement and negative percent agreement, for each specimen type claimed in the intended use of the device, must meet one of the following two minimum clinical performance criteria: (i) For devices evaluated as compared to an FDA-cleared nucleic acid based-test or other currently appropriate and FDA accepted comparator method other than correctly performed viral culture method: (A) The positive percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent. (B) The negative percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent. (ii) For devices evaluated as compared to correctly performed viral culture method as the comparator method: (A) The sensitivity estimate for the device when testing for influenza A must be at the point estimate of at least 90 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 80 percent. The sensitivity estimate for the device when testing for influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent. (B) The specificity estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent. (2) When performing testing to demonstrate the device meets the requirements in paragraph (b)(1) of this section, a currently appropriate and FDA accepted comparator method must be used to establish assay performance in clinical studies. (3) Annual analytical reactivity testing of the device must be performed with contemporary influenza strains. This annual analytical reactivity testing must meet the following criteria: (i) The appropriate strains to be tested will be identified by FDA in consultation with the Centers for Disease Control and Prevention (CDC) and sourced from CDC or an FDA-designated source. If the annual strains are not available from CDC, FDA will identify an alternative source for obtaining the requisite strains. (ii) The testing must be conducted according to a standardized protocol considered and determined by FDA to be acceptable and appropriate. (iii) By July 31 of each calendar year, the results of the last 3 years of annual analytical reactivity testing must be included as part of the device's labeling. If a device has not been on the market long enough for 3 years of annual analytical reactivity testing to have been conducted since the device received marketing authorization from FDA, then the results of every annual analytical reactivity testing since the device received marketing authorization from FDA must be included. The results must be presented as part of the device's labeling in a tabular format, which includes the detailed information for each virus tested as described in the certificate of authentication, either by: (A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where the analytical reactivity testing data can be found; or (B) In the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access. (4) If one of the actions listed at section 564(b)(1)(A)-(D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services (HHS) determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain: (i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate. The procedure and location of testing may depend on the nature of the emerging virus. (ii) Within 60 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by: (A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or (B) In a section of the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # AUG 1 0 2004 # 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in This Summary of or or or or or or of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________ # SUBMITTER Binax, Inc. 217 Read Street Portland, Maine 04103 (207) 772-3988 (Office) (207) 871-5751 (FAX) # CONTACT PERSON Pamela S. Angell pangell@binax.com (email) # DATE PREPARED April 21, 2004 TRADE NAME BinaxNOW® Influenza A & B Test #### COMMON NAME NOW® Flu A/B Test, NOW® Influenza A/B, NOW® Influenza A & B, Binax NOW® Influenza A & B, Binax NOW® Influenza A/B ## CLASSIFICATION NAME Antigen, CF (including CF Controls), Influenza Virus A, B, C (per 21 CFR 866.3330) ## PREDICATE DEVICES Binax NOW® Flu A Test, K021649 Binax NOW® Flu B Test; K021646 FLU OIA A/B Test Kit; K023556 ## DEVICE DESCRIPTION: The BinaxNOW® Influenza A & B Test is an immunochromatographic membrane assay that uses highly sensitive monoclonal antibodies to detect influenza type A & B nucleoprotein antigens in nasopharyngeal specimens. These antibodies and a control antibody are immobilized onto a membrane support as three distinct lines and combined with other reagents/pads to construct a test strip. This test strip is mounted inside a cardboard, book-shaped hinged test device. Swab specimens require a sample preparation step, in which the sample is eluted off the swab into elution solution or transport media. Nasal wash/aspirate samples require no oreparation. Sample is added to the top of the test strip and the test device is closed. Test results are interpreted at 15 minutes based on the presence or absence of pink-to-purple colored Sample Lines. The blue Control Line turns pink in a valid assay. {1}------------------------------------------------ #### INTENDED USE The BinaxNOW® Influenza A & B Test is an in vitro immunochromatographic assay for The BinaxNOVP "Influenza A and B nucleoprotein antigens in nasopharyngeal the qualitative detection of influenza A and B nucleoprotein antigens in nasophary the qualitative delection of influenza A and B hasto provid in the rapid differential Swab and flashfaspirate opedination will infections. Negative test results should be confirmed by culture. # TECHNOLOGICAL CHARACTERISTICS The BinaxNOW® Influenza A & B Test and the two predicate Binax NOW® Flu Tests rne DinaxNOW - Innlaoize Annochromatographic technology while the Thermo Biostar tuse identival librar low minner. All tests are rapid immunoassays that employ specific test is an option infridhodoud). All this phase to capture and visualize influenza nucleoprotein antigens. #### PERFORMANCE SUMMARY ## Analytical and Clinical Comparison to Predicates Allarytoar and Olinkour Companies of the individual NOW® Flu A and NOW® Flu B Tests vs. rne oiltion performance were originally established in a multi-center prospective conventional ouring the 2002 flu season at physician offices and clinics located in stady of national auring and States. The analytical sensitivity and specificity of the BinaxNOW® Influenza A & B Test are equivalent to that of the individual NOW® Flu the Binaxive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . clinical specimens and inactivated viral standards. #### Clinical Sample Comparison: A total of 191 nasal wash and 182 nasopharyngeal (NP) swab specimens were evaluated as part of a 2002 multi-site prospective clinical study comparing the original Binax NOW® Flu A and Flu B Tests to viral culture. In this study, sensitivity to influenza A was 82% (95% Cl of 69-90%) for nasal washes and 78% (95% Cl of 62-88%) for NP swabs. Sensitivity to influenza B was 71% (95% CI of 56-83%) for nasal washes and 58% for NP swabs (95% CI of 42-73%). Specificity for washes and swabs ranged from 92% to 97% with 95% CI of 86% to 99%. Performance of the BinaxNOW® Influenza A & B Test was compared to the current NOW® Flu A Test on 306 retrospective frozen clinical samples and to the NOW® Flu B Test on 303 retrospective frozen clinical samples. All clinical samples were collected from symptomatic patients at multiple physician offices, clinics and hospitals located in the Southern, Northeastern and Midwestern regions of the United States and from one hospital in Sweden. Fifty-three percent (53%) of the population tested was male, 47% female, 64% pediatric (< 18 years) and 36% adult (≥ 18 years). Nasal wash/aspirate specimens comprised approximately 57% of the samples tested, while NP swabs represented 42%. No differences in test performance were observed based on patient age and gender or based on sample type tested. The BinaxNOW® Influenza A & B Test was 100% sensitive and 96% specific for detection of influenza A vs. the NOW® Flu A Test and 93% sensitive and 97% specific for detection of influenza B vs. the NOW® Flu B Test. Test performance by {2}------------------------------------------------ virus type (A vs. B), by sample type (NP swab vs. wash/aspirate), and overall, including 95% confidence intervals, is detailed in the following tables. | Sample | +/+ | +/- | % Sens | 95% CI | -/- | +/- | % Spec | 95% CI | |---------------|-----|-----|--------|---------|-----|-----|--------|---------| | NP Swab | 30 | 0 | 100% | 89-100% | 96 | 1 | 99% | 94-100% | | Wash/Aspirate | 47 | 0 | 100% | 93-100% | 123 | 9 | 93% | 88-96% | | Overall | 77 | 0 | 100% | 95-100% | 219 | 10 | 96% | 88-96% | | BinaxNOW® Influenza A & B Test vs. NOW® Flu B Test for Detection of Influenza B | | |---------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | and the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the consistence of the con | | FLU B SENSITIVITY | | | | | |-------------------|-----|-----|--------|--------| | Sample | +/+ | -/+ | % Sens | 95% CI | | NP Swab | 2 | 0 | 100% | 29-99% | | Wash/Aspirate | 12 | 1 | 92% | 66-98% | | Overall | 14 | 1 | 93% | 70-98% | | FLU B SPECIFICITY | | | | | | |-------------------|-----|----------------------------------------------------------------------------------------------------------------------------------------------|---------|--|--| | | +/- | % Spec | 95% C | | | | 126 | | යිති බවට ගිහිටි මිතිවර්තිය හිටි බවට පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි මිතිබු මිත්‍රියා හිටි පිහිටි මිතිබු මිත්‍රියා හිත්‍යයි. | 96-100% | | | | 152 | ರಾ | 94% | 90-97% | | | | 278 | 10 | 97% | 94-98% | | | # Analytical Sensitivity Comparison: The BinaxNOW® test limit of detection (LOD), defined as the concentration of influenza virus that produces positive BinaxNOW® test results approximately 95% of the time, was identified by evaluating different concentrations of inactivated Flu A/Beijing and inactivated Flu B/Harbin in the BinaxNOW® test. Twelve (12) different operators each interpreted 2 devices run at each concentration for a total of 24 determinations per level. The following results identify a concentration of 1.03 x 102 ng/ml as the LOD for Flu A/Beijing and 6.05 x 10' na/mi for Flu B/Harbin. | Influenza A/Beijing | | | Influenza B/Harbin | | | |-------------------------------|------------|------------|-------------------------------|------------|------------| | Concentration (ng/ml) | # Detected | % Detected | Concentration (ng/ml) | # Detected | % Detected | | $1.03 \times 10^2$ (LOD) | 23/24 | 96 | $6.05 \times 10^1$ (LOD) | 23/24 | 96 | | $5.60 \times 10^1$ (Cutoff) | * | 50 | $2.42 \times 10^1$ (Cutoff) | 11/24 | 46 | | $3.27 \times 10^1$ (High Neg) | 4/24 | 17 | $1.51 \times 10^1$ (High Neg) | 6/24 | 25 | | True Negative | 0/24 | 0 | True Negative | 0/24 | 0 | *Linear regression was used to calculate a line equation, which was then used to project the cutoff concentration of Flu A/Beijing. To demonstrate comparable analytical sensitivity of the BinaxNOW® Influenza A & B Test and the individual NOW® Flu A and Flu B Tests, the Flu A and B cutoff levels identified above were evaluated in the NOW® Flu A and Flu B Tests. The A/Beijing cutoff sample detected 50% of the time in the BinaxNOW® Influenza A & B Test was also detected 50% (12/24) of the time in the NOW® Flu A Test when tested by six (6) operators interpreting a total of 24 devices. Likewise, the B/Harbin cutoff sample detected 46% of the time in the BinaxNOW® Influenza A & B Test was detected 10% (4/40) of the time in the NOW® Flu B Test when tested by ten (10) operators interpreting 40 devices. {3}------------------------------------------------ These data demonstrate that the analytical sensitivity of the BinaxNOW® Influenza These data demonorate than that of the individual NOW® Flu A and B Tests. #### Reactivity Testing The seven (7) live influenza A strains and five (5) live influenza B strains listed below The seven (7) Inve mindenza A Strains and The (6) The Tat concentrations ranging from lested positive in the BinaxNOW - militenza strains causing infection in humans 10 -10 GEIDsomm. All contain the conserved nucleoproteins targeted by the BinaxNOW® test.1 | Influenza Strain | ATCC # | |--------------------------|--------| | Flu A/WS/33 | VR-825 | | Flu A/NWS/33 | VR-219 | | Flu A/Hong Kong/8/68 | VR-544 | | Flu A/Aichi/2/68 | VR-547 | | Flu A/New Jersey/8/76 | VR-897 | | Flu A/Mal/302/54 | VR-98 | | Flu A/Port Chalmers/1/73 | VR-810 | | Flu B/Lee/40 | VR-101 | | Flu B/Brigit | VR-786 | | Flu B/Russia/69 | VR-790 | | Flu B/Hong Kong/5/72 | VR-791 | | Flu B/R75 | VR-789 | #### Analytic Specificity (Cross-Reactivity) Allariti Specificity (Sross Reasting) To determine the analytical specificity of the BinaxNOW® Influenza A & B Test, 36 commensal and pathogenic microorganisms (27 bacteria, 8 viruses and 1 yeast) that may be present in the nasal cavity or nasopharynx were tested. All of the following may be probons were negative when tested at concentrations ranging from 104 to 108 TCIDsoml (viruses), 10 to 10° organisms/ml (bacteria) and 10° organisms/ml (yeast). | Bacteria | Viruses | Yeast | |------------------------------------|-----------------------------------|------------------| | Acinetobacter | Adenovirus | Candida albicans | | Bordetella pertussis | Coronavirus | | | Enterococcus faecalis | Coxsackie B4 | | | Escherichia coli | Cytomegalovirus (CMV) | | | Gardnerella vaginalis | Parainfluenza 1 | | | Haemophilus influenzae | Parainfluenza 2 | | | Klebsiella pneumoniae | Parainfluenza 3 | | | Lactobacillus casei | Respiratory Syncytial Virus (RSV) | | | Legionella pneumophila | | | | Listeria monocytogenes | | | | Moraxella catarrhalis | | | | Neisseria gonorhoeae | | | | Neisseria meningitidis | | | | Neisseria sicca | | | | Neisseria subflava | | | | Proteus vulgaris | | | | Pseudomonas aeruginosa | | | | Serratia marcescens | | | | Staphylococcus aureus | | | | Staphylococcus aureus | | | | (Cowan protein A producing strain) | | | | Staphylococcus epidermidis | | | | Streptococcus, Group A | | | | Streptococcus, Group B | | | | Streptococcus, Group C | | | | Streptococcus, Group F | | | | Streptococcus mutans | | | | Streptococcus pneumoniae | | | {4}------------------------------------------------ #### Interfering Substances Interientig oubstances, naturally present in respiratory specimens or that may be The lollowing oubotarioo, haturaly cavity or nasopharynx, were evaluated in the armicially influenza A & B Test at the concentrations listed and were found not to Diffect test performance. Whole blood (1%) did not interfere with the interpretation of anest test performance. but did interfere with the interpretation of Flu A LOD negative samples. Therefore, visibly bloody samples may not be appropriate for use in this test. | Substance | Concentration | |-------------------------|---------------| | 1 OTC mouthwash | 20% | | 3 OTC nasal sprays | 15% | | 3 OTC throat drops | 15% | | 2 OTC throat sprays | 20% | | 4-acetamidophenol | 10 mg/ml | | Acetylsalicylic acid | 15 mg/ml | | Albuterol | 20 mg/ml | | Chlorpheniramine | 5 mg/ml | | Dextromethorphan | 10 mg/ml | | Diphenhydramine | 5 mg/ml | | Guaiacol glycerol ether | 20 mg/ml | | Oxymetazoline | 0.05% | | Phenylephrine | 50 mg/ml | | Phenylpropanolamine | 20 mg/ml | | Rebetol | 500 ng/ml | | Relenza | 20 mg/ml | | Rimantadine | 500 ng/ml | | Synagis | 0.1 mg/ml | | Tamiflu | 50 mg/ml | #### Transport Media The following transport media were tested in the BinaxNOW® Influenza A & B Test as negative samples (no virus present) and after inoculation with the LOD levels of Inflyenza A & B. Media did not impact BinaxNOW® test performance, with the media alone testing negative in the NOW® test and media inoculated with LOD Influenza A & B testing positive on the appropriate test line in BinaxNOW® test. Amies Media Hank's Balanced Salt Solution M4 Media M4-RT Media M5 Media Stuart's Media Saline #### Reproducibility A blind study of the BinaxNOW® Influenza A & B Test was conducted at 3 separate sites using panels of blind coded specimens containing negative, low positive, and moderate positive samples. Participants tested each sample multiple times on 3 different days. There was 97% (242/250) agreement with expected test results, with no significant differences within run (replicates tested by one operator), between run (3 different days) or between sites (3 sites). Signed Karen Hickey Date Date April 21, 2004 Karen Hickey VP Regulatory & Clinical Affairs {5}------------------------------------------------ 1) Dowdle, W.R., Kendal, A.P., and Noble, G.R. 1980. Influenza Virus, p 836-884. Manual of Clinical Microbiology, 3rd edition, In Lennette, et. Al (ed.). American So {6}------------------------------------------------ Food and Drug Administration 2098 Gaither Road Rockville MD 20850 AUG 1 0 2004 Ms. Pamela Angell Regulatory Manager Binax, Inc 217 Read Street Portland, ME 04103 K041049 Re: Trade/Device Name: BinaxNOW Influenza A & B Regulation Number: 21 CFR 866.3330 Regulation Name: Influenza Virus Serological Reagents Regulatory Class: Class I Product Code: GNX Dated: April 21, 2004 Received: April 22, 2004 Dear Ms. Angell: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your bection 510(x) premained is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the caciosate) to regally manation program and of the Medical Device American Comments, or to commerce prof to May 20, 1970, the encentialic and the Federal Food, Drug, devices mat have been reciasmed in acce aapsoval of a premarket approval application (PMA). and Cosmetic Acr (110) market the device, subject to the general controls provisions of the Act. The Y bu may, merciole, market the act include requirements for annual registration, listing of general controls provisions of the free labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (500 a0070) and existing major regulations affecting your device can may be subject to such additional controls. Existing major regulation withi may be subject to such additional ocharely into and to 895. In addition, FDA be found in Thic 21, Oods of Peach one one ming your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualled of a backannal requirements of the Act that FDA has made a determination that your device complies with other requirements of the Act that FDA has made a decertifications administered by other Federal agencies. You must or any rederal statutes and regulations and limited to: registration and listing (21 comply with an the Ace 3 requirements, more of 809); and good manufacturing practice CFR Fart 807), labeling (21 CF ruality systems (QS) regulation (21 CFR Part 820). {7}------------------------------------------------ #### Page 2 This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to ocgin manoming your antial equivalence of your device to a legally premarket notification. The PDF maing of Gassification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, II you desite specific mornation assectiving of your device, please contact the Office of of questions on the promotion and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Tou may obtain other general information on your responsibilities under the Act from the I bu may obtain other geleral marketing and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Salartys Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ # Indications for Use 510(k) Number (if known): K041049 Device Name: BinaxNow® Influenza A&B Test Indications for Use: The BinaxNOW® Influenza A & B Test is an in vitro immunochromatographic assay for The DillaXNO w - Influenza A co B A and B nucleoprotein antigens in nasopharyngeal (NP) swab and nasal wash/aspirate specimens. It is intended to aid in the rapid (Nr) >Wav and nasar wasn't wasn't level infections. Negative test results should be confirmed by culture. X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Division Sign-Off ාස්ථි Device Evaluation and Safety Page 1 of 1 *10(k) Ko4104 9
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...