K053126 · Binax, Inc. · PSZ · Nov 30, 2005 · Microbiology
Device Facts
Record ID
K053126
Device Name
BINAXNOW INFLUENZA A & B TEST
Applicant
Binax, Inc.
Product Code
PSZ · Microbiology
Decision Date
Nov 30, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3328
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K053126 · Nov 30, 2005
BINAXNOW INFLUENZA A & B TEST
Binax, Inc.
Retrospective frozen clinical samples from physician offices, clinics, and hospitals
Retrospective clinical samples were used to evaluate the performance (sensitivity and specificity) of the BinaxNOW Influenza A & B Test compared to the predicate Binax NOW Flu A and Flu B tests.
Retrospective clinical samples; Performance validation; Clinical site collection
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective study using frozen clinical samples
Symptomatic patients (64% pediatric, 36% adult); Sample Size: 306 (Flu A) and 303 (Flu B) samples; Number of Sites: Multiple sites (US and Sweden)
Binax NOW® Flu A Test and Binax NOW® Flu B Test
Sensitivity and specificity for detection of influenza A and B
Indications for Use
The BinaxNOW® Influenza A & B Test is an in vitro immunochromatographic assay for the qualitative detection of influenza A and B nucleoprotein antigens in nasopharyngeal swab and nasal wash/aspirate specimens. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. Negative test results should be confirmed by culture.
Device Story
The BinaxNOW® Influenza A & B Test is a lateral flow immunochromatographic assay used in clinical settings to detect influenza A and B nucleoprotein antigens. The device consists of a cardboard, book-shaped hinged test unit containing a test strip with immobilized monoclonal antibodies. Nasopharyngeal swab or nasal wash/aspirate specimens are eluted into a solution and applied to the test strip. If antigens are present, they bind to the antibodies, creating visible pink-to-purple sample lines within 15 minutes. A blue control line confirms assay validity. Healthcare providers interpret the results visually to aid in the differential diagnosis of influenza infections. The device provides rapid results, allowing for timely clinical decision-making, though negative results must be confirmed by culture. It is intended for use by healthcare professionals.
Clinical Evidence
Clinical performance was evaluated in prospective and retrospective studies. A prospective study (n=114) compared the device to cell culture/DFA, showing 75% sensitivity/100% specificity for Flu A and 50% sensitivity/100% specificity for Flu B. A retrospective study (n=306/303) compared the device to individual Flu A/B tests, showing 100% sensitivity/96% specificity for Flu A and 93% sensitivity/97% specificity for Flu B. Reproducibility studies across 3 sites showed 97% agreement. Analytical studies confirmed detection of various influenza strains and H5N1, with no cross-reactivity against 36 common respiratory microorganisms.
Technological Characteristics
Lateral flow immunochromatographic assay. Employs monoclonal antibodies immobilized on a membrane support within a cardboard, book-shaped hinged device. Qualitative visual readout. No electronic components or software. Manual operation.
Indications for Use
Indicated for qualitative detection of influenza A and B nucleoprotein antigens in nasopharyngeal (NP) swab and nasal wash/aspirate specimens to aid in rapid differential diagnosis of influenza A and B viral infections in symptomatic patients. Negative results require culture confirmation.
Regulatory Classification
Identification
An influenza virus antigen detection test system is a device intended for the qualitative detection of influenza viral antigens directly from clinical specimens in patients with signs and symptoms of respiratory infection. The test aids in the diagnosis of influenza infection and provides epidemiological information on influenza. Due to the propensity of the virus to mutate, new strains emerge over time which may potentially affect the performance of these devices. Because influenza is highly contagious and may lead to an acute respiratory tract infection causing severe illness and even death, the accuracy of these devices has serious public health implications.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The device's sensitivity and specificity performance characteristics or positive percent agreement and negative percent agreement, for each specimen type claimed in the intended use of the device, must meet one of the following two minimum clinical performance criteria:
(i) For devices evaluated as compared to an FDA-cleared nucleic acid based-test or other currently appropriate and FDA accepted comparator method other than correctly performed viral culture method:
(A) The positive percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The negative percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(ii) For devices evaluated as compared to correctly performed viral culture method as the comparator method:
(A) The sensitivity estimate for the device when testing for influenza A must be at the point estimate of at least 90 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 80 percent. The sensitivity estimate for the device when testing for influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The specificity estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(2) When performing testing to demonstrate the device meets the requirements in paragraph (b)(1) of this section, a currently appropriate and FDA accepted comparator method must be used to establish assay performance in clinical studies.
(3) Annual analytical reactivity testing of the device must be performed with contemporary influenza strains. This annual analytical reactivity testing must meet the following criteria:
(i) The appropriate strains to be tested will be identified by FDA in consultation with the Centers for Disease Control and Prevention (CDC) and sourced from CDC or an FDA-designated source. If the annual strains are not available from CDC, FDA will identify an alternative source for obtaining the requisite strains.
(ii) The testing must be conducted according to a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(iii) By July 31 of each calendar year, the results of the last 3 years of annual analytical reactivity testing must be included as part of the device's labeling. If a device has not been on the market long enough for 3 years of annual analytical reactivity testing to have been conducted since the device received marketing authorization from FDA, then the results of every annual analytical reactivity testing since the device received marketing authorization from FDA must be included. The results must be presented as part of the device's labeling in a tabular format, which includes the detailed information for each virus tested as described in the certificate of authentication, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where the analytical reactivity testing data can be found; or
(B) In the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.
(4) If one of the actions listed at section 564(b)(1)(A)-(D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services (HHS) determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate. The procedure and location of testing may depend on the nature of the emerging virus.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.
{1}------------------------------------------------
NOV 3 0 2005
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| The assigned 510(k) number is: | K025312 |
|--------------------------------|---------|
|--------------------------------|---------|
The purpose of this 510(k) submission is to expand the Analytical Reactivity claims of the currently 510(k) cleared BinaxNOW® Influenza A & B Test (510(k) #K041049).
# SUBMITTER
Binax, Inc. 10 Southgate Road Scarborough, Maine 04074 (207) 730-5700 (Office) (207) 730-5710 (FAX)
## CONTACT PERSON
Angela Drysdale angela.drysdale@binax.com {email)
## DATE PREPARED
November 04, 2005
## TRADE NAME
BinaxNOW® Influenza A & B Test
#### COMMON NAME
NOW® Flu A/B Test, NOW® Influenza A/B, NOW® Influenza A & B, Binax NOW® Influenza A & B, Binax NOW® Influenza A/B
#### CLASSIFICATION NAME
Antigen, CF (including CF Controls), Influenza Virus A, B, C (per 21 CFR 866.3330)
# PREDICATE DEVICES
Binax NOW® Influenza A & B Test; K041049 Binax NOW® Flu A Test; K021649 Binax NOW® Flu B Test: K021646
# DEVICE DESCRIPTION:
The BinaxNOW® Influenza A & B Test is an immunochromatographic membrane assay that uses highly sensitive monoclonal antibodies to detect influenza type A & B nucleoprotein antigens in nasopharyngeal specimens. These antibodies and a control antibody are immobilized onto a membrane support as three distinct lines and combined with other reagents/pads to construct a test strip. This test strip is mounted inside a cardboard, bookshaped hinged test device.
Swab specimens require a sample preparation step, in which the sample is eluted off the swab into elution solution or transport media. Nasal wash/aspirate samples require no preparation. Sample is added to the top of the test strip and the test device is closed. Test results are interpreted at 15 minutes based on the presence of pink-to-purple colored Sample Lines. The blue Control Line turns pink in a valid assay.
{2}------------------------------------------------
# INTENDED USE
The BinaxNOW® Influenza A & B Test is an in vitro immunochromatographic assay for the qualitative detection of influenza A and B nucleoprotein antigens in nasopharyngeal swab and nasal wash/aspirate specimens. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. Negative test results should be confirmed by culture.
## TECHNOLOGICAL CHARACTERISTICS
The Expanded Claim BinaxNOW® Influenza A & B Test is exactly the same as the currently 510/k) cleared BinaxNOW® Influenza A & B Test. Both tests use identical lateral flow immunochromatographic technology. Both tests are rapid immunoassays that employ specific antibodies immobilized onto a solid phase to capture and visualize influenza nucleoprotein antigens.
## PERFORMANCE SUMMARY
## CLINICAL STUDIES
## BinaxNOW® Influenza A & B Test Performance vs. Cell Culture / DFA - Prospective Study
The performance of the BinaxNQW® Influenza A & B Test was compared to cell culture and/or DFA, and to the Binax NOW® Flu A Test and the Binax NOW® Flu B Test, in a prospective study conducted in 2004 outside the US. Nasopharyngeal (NP) swab and nasal wash / aspirate specimens, collected at multiple sites from children (less than 18 years of age) and adults (18 years or older) presenting with influenza-like symptoms, were evaluated in the Binax test at a central testing lab.
Forty-seven percent (47%) of the population tested was male, 53% female, 40% pediatric (< 18 years), and 60% adult (> 18 years). No differences in test performance were observed based on patient age or gender. There were no invalid tests reported.
One hundred and thirteen (113) NP swab specimens and 1 wash/aspirate specimen were tested. One hundred and eight (108) of the 114 samples tested were influenza negative by culture/DFA, and 6 samples were influenza positive. When compared to culture/DFA, the BinaxNOW® Test was 75% (3/4) sensitive and 100% (110/110) specific for detection of influenza A and 50% (1/2) sensitive and 100% (112/112) specific for detection of influenza B. There was 100% agreement between the BinaxNOW® Influenza A & B Test and the individual Flu A and Flu B Tests.
BinaxNOW® A & B Test specificity by sample type versus cell culture / DFA, including 95% confidence intervals, is listed below.
BinaxNOW® Influenza A & B Test Specificity vs. Cell Culture/DFA
| FLU A SPECIFICITY | | | | |
|-------------------|-----|-----|--------|---------|
| Sample | -/+ | +/- | % Spec | 95% CI |
| NP Swab | 109 | 0 | 100% | 97-100% |
| Wash/Aspirate | 1 | 0 | 100% | 16-99% |
| Overall | 110 | 0 | 100% | 97-100% |
| FLU B SPECIFICITY | | | |
|-------------------|-----|--------|---------|
| -/- | +/- | % Spec | 95% CI |
| 111 | 0 | 100% | 97-100% |
| 1 | 0 | 100% | 16-99% |
| 112 | 0 | 100% | 97-100% |
{3}------------------------------------------------
BinaxNOW® Influenza A & B Test Performance vs. Binax NOW® Flu A and Flu B Tests:
Performance of the BinaxNOW® Influenza A & B Test was compared to the current NOW® Flu A Test on 306 retrospective frozen clinical samples and to the NOW® Flu B Test on 303 retrospective frozen clinical samples. All clinical samples were collected from symptomatic patients at multiple physician offices, clinics and hospitals located in the Southern, Northeastern and Midwestern regions of the United States and from one hospital in Sweden. Fifty-three percent (53%) of the population tested was male, 47% female, 64% pediatric (< 18 years) and 36% adult (≥ 18 years). Nasal wash/aspirate specimens comprised approximately 57% of the samples tested, while NP swabs represented 42%. No differences in test performance were observed based on patient age and gender or based on sample type tested.
The BinaxNOW® Influenza A & B Test was 100% sensitive and 96% specific for detection of influenza A vs. the NOW® Flu A Test and 93% sensitive and 97% specific for detection of influenza B vs. the NOW® Flu B Test. Test performance by virus type (A vs. B), by sample type (swab vs. wash/aspirate), and overall, including 95% confidence intervals, is detailed in the following tables.
| | FLU A SENSITIVITY | | | | FLU A SPECIFICITY | | | |
|---------------|-------------------|-----|--------|---------|-------------------|-----|--------|---------|
| Sample | +/+ | -/+ | % Sens | 95% CI | +/- | +/+ | % Spec | 95% CI |
| NP Swab | 30 | 0 | 100% | 89-100% | 96 | 1 | 99% | 94-100% |
| Wash/Aspirate | 47 | 0 | 100% | 93-100% | 123 | 9 | 93% | 88-96% |
| Overall | 77 | 0 | 100% | 95-100% | 219 | 10 | 96% | 88-96% |
| Sample | FLU B SENSITIVITY | | % Sens | 95% CI | FLU B SPECIFICITY | | % Spec | 95% CI |
|---------------|-------------------|-----|--------|--------|-------------------|-----|--------|---------|
| | +/+ | -/+ | | | +/- | -/+ | | |
| NP Swab | 2 | 0 | 100% | 29-99% | 126 | 1 | 99% | 96-100% |
| Wash/Aspirate | 12 | 1 | 92% | 66-98% | 152 | 9 | 94% | 90-97% |
| Overall | 14 | 1 | 93% | 70-98% | 278 | 10 | 97% | 94-98% |
# Binax NOW® Flu A and Flu B Test Performance vs. Cell Culture
Performance of the Binax NOW® Flu A and Flu B Tests was compared to cell culture on 373 prospective clinical samples collected as part of a multi-center study conducted during the 2002 Flu season at physician offices and clinics located in the Western and mid-Atlantic United States. Fifty-four percent (54%) of the population tested was male, 46% female, 90% pediatric (< 18 years) and 10% adult (≥ 18 years). Nasal wash/aspirates comprised 51% of the samples tested, while NP swabs represented 49%. No differences in performance were observed based on patient age and gender or based on sample type tested.
The Binax NOW® Flu A Test was 80% sensitive and 93% specific while the Binax NOW® Flu B Test was 65% sensitive and 97% specific when compared to cell culture. The performance of the two tests by sample type (swab vs. wash/aspirate) and overall, including 95% confidence intervals, is detailed in the following tables.
{4}------------------------------------------------
| Binax NOW® Flu A Test vs. Cell Culture | | | | | | | |
|----------------------------------------|-------------------|-----|--------|--------|----------------|-----|--------|
| | FLU A SENSITIVITY | | | | FLU A SPECIFIC | | |
| Sample | +/+ | +/- | % Sens | 95% CI | -/- | +/- | % Spec |
| NP Swab | 29 | 8 | 78% | 62-88% | 133 | 12 | 92% |
| Wash/Aspirate | 40 | 9 | 82% | 69-90% | 133 | 9 | 94% |
| Overall | 69 | 17 | 80% | 71-87% | 266 | 21 | 93% |
#### Binax NOW Flu B Test vs. Cell Culture
| Sample | FLU B SENSITIVITY | | | |
|---------------|-------------------|-----|--------|--------|
| | +/+ | +/- | % Sens | 95% CI |
| NP Swab | 21 | 15 | 58% | 42-73% |
| Wash/Aspirate | 29 | 12 | 71% | 56-83% |
| Overall | 50 | 27 | 65% | 54-75% |
| FLU B SPECIFICITY | | | |
|-------------------|-----|--------|--------|
| +/- | +/- | % Spec | 95% CI |
| 142 | 4 | 97% | 93-99% |
| 146 | 4 | 97% | 93-99% |
| 288 | 8 | 97% | 95-99% |
CITY
95% Cl
88-85% 89-97% 89-95%
## ANALYTICAL STUDIES
#### Analytical Sensitivity
The BinaxNOW® test limit of detection (LOD), defined as the concentration of influenza virus that produces positive BinaxNOW test results approximately 95% of the time, was identified by evaluating different concentrations of inactivated Flu A/Beijing and inactivated Flu B/Harbin in the BinaxNOW test.
Twelve (12) different operators each interpreted 2 devices run at each concentration for a total of 24 determinations per level. The following results identify a concentration of 1.03 x 10° ng/ml as the LOD for Flu A/Beijing and 6.05 x 101 ng/ml for Flu B/Harbin.
| Influenza A/Beijing | | | |
|-------------------------------|---------------|---------------|--|
| Concentration<br>(ng/ml) | #<br>Detected | %<br>Detected | |
| $1.03 \times 10^2$ (LOD) | 23/24 | 96 | |
| $5.60 \times 10^1$ (Cutoff) | * | 50 | |
| $3.27 \times 10^1$ (High Neg) | 4/24 | 17 | |
| True Negative | 0/24 | 0 | |
| Influenza B/Harbin | | | |
| Concentration<br>(ng/ml) | #<br>Detected | %<br>Detected | |
| $6.05 x 10^1$ (LOD) | 23/24 | 96 | |
| $2.42 x 10^1$ (Cutoff) | 11/24 | 46 | |
| $1.51 x 10^1$ (High Neg) | 6/24 | 25 | |
| True Negative | 0/24 | 0 | |
*Linear regression was used to calculate a line equation, which was then used to project the cutoff concentration of Flu A/Beijing.
To demonstrate comparable analytical sensitivity of the BinaxNOW® Influenza A & B Test and the individual NOW® Flu A and Flu B Tests, the Flu A and B cutoff levels identified above were evaluated in the NOW® Flu A and Flu B Tests.
The A/Beijing cutoff sample detected 50% of the time in the BinaxNOW® Influenza A & B Test was also detected 50% (12/24) of the time in the NOW® Flu A Test when tested by six (6) operators interpreting a total of 24 devices. Likewise, the B/Harbin cutoff sample detected 46% of the time in the BinaxNOW® Influenza A & B Test was detected 10% (4/40) of the time in the NOW® Flu B Test when tested by ten (10) operators interpreting 40 devices.
These data demonstrate that the analytical sensitivity of the BinaxNOW® Influenza A & B Test is equivalent or better than that of the individual NOW® Flu A and B Tests.
{5}------------------------------------------------
## Avian Influenza (H5N1) Analytical Reactivity Testing
The two (2) live avian influenza (H5N1) strains isolated from humans listed below tested positive in the BinaxNOW Influenza A & B Test at concentrations ranging from 10-10 TCIDstml.
| Avian Influenza Strain | Concentration |
|------------------------|---------------|
| A/Hong Kong/156/97 | 1.3 x 102 |
| A/Vietnam/1194/04 | 1.0 x 104 |
#### Reactivity Testing
The seven (7) live influenza A strains and five (5) live influenza B strains listed below tested positive in the BinaxNOW® Influenza A & B Test at concentrations ranging from 10'-10' CEIDss/ml. Atthough the specific influenza strains causing infection in humans can vary year to year, all contain the conserved nucleoproteins targeted by the BinaxNOW® test.
| Influenza Strain | ATCC |
|--------------------------|--------|
| Flu A/NWS/33 | VR-825 |
| Flu A/NWS/33 | VR-215 |
| Flu A/Hong Kong/8/68 | VR-544 |
| Flu A/Aichi/2/68 | VR-547 |
| Flu A/New Jersey/8/76 | VR-897 |
| Flu A/Mal/302/54 | VR-98 |
| Flu A/Port Chalmers/1/73 | VR-810 |
| Flu B/Lee/40 | VR-101 |
| Flu B/Brigit | VR-786 |
| Flu B/Russia/69 | VR-790 |
| Flu B/Hong Kong/5/72 | VR-791 |
| Flu B/R75 | VR-789 |
#### Analytic Specificity (Cross-Reactivity)
To determine the analytical specificity of the BinaxNOW® Influenza A & B Test, 36 commensal and pathogenic microorganisms (27 bacteria, 8 viruses and 1 yeast) that may be present in the nasal cavity or nasopharynx were tested. All of the following microorganisms were negative when tested at concentrations ranging from 104 to 10 TCIDs /ml (viruses), 10' to 108 organisms/ml (bacteria) and 10° organisms/ml (yeast).
| Bacteria | Viruses | Yeast |
|------------------------------------|-----------------------------------|------------------|
| Acinetobacter | Adenovirus | Candida albicans |
| Bordetella pertussis | Coronavirus | |
| Enterococcus faecalis | Coxsackie B4 | |
| Escherichia coli | Cytomegalovirus (CMV) | |
| Gardnerella vaginalis | Parainfluenza 1 | |
| Haemophilus influenzae | Parainfluenza 2 | |
| Klebsiella pneumoniae | Parainfluenza 3 | |
| Lactobacillus casei | Respiratory Syncytial Virus (RSV) | |
| Legionella pneumophila | | |
| Listeria monocytogenes | | |
| Moraxella catarrhalis | | |
| Neisseria gonorrhoeae | | |
| Neisseria meningitidis | | |
| Neisseria sicca | | |
| Neisseria subflava | | |
| Proteus vulgaris | | |
| Pseudomonas aeruginosa | | |
| Serratia marcescens | | |
| Staphylococcus aureus | | |
| Staphylococcus aureus | | |
| (Cowan protein A producing strain) | | |
| Staphylococcus epidermidis | | |
| Streptococcus, Group A | | |
| Streptococcus, Group B | | |
{6}------------------------------------------------
Streptococcus, Group C Streptococcus, Group F Streptococcus mutans Streptococcus pneumoniae
#### Interfering Substances
The following substances, naturally present in respiratory specimens or that may be artificially introduced into the nasal cavity or nasopharynx, were evaluated in the BinaxNOW® Influenza A & B Test at the concentrations listed and were found not to affect test performance. Whole blood (1%) did not interfere with the interpretation of negative BinaxNOW® test results, but did interfere with the interpretation of Flu A LOD positive samples. Therefore, visibly bloody samples may not be appropriate for use in this test.
Substance 1 OTC mouthwash 3 OTC nasal sprays 3 OTC throat drops 2 OTC throat sprays 4-acetamidophenol Acetylsalicylic acid Albuterol Chlorpheniramine Dextromethorphan Diphenhydramine Guaiacol glycerol ether Oxymetazoline Phenylephrine Phenylpropanolamine Rebetol Relenza Rimantadine Synagis Tamiflu
Concentration 20% ારત જિલ્લાના એક એવા ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં મુખ્યત્વે ખેત-ઉત્પાદન છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપા નદર્ભ 20% 10 ma/ml 15 mg/ml 20 mg/ml 5 mg/ml 10 mg/ml 5 mg/ml 20 mg/ml 0.05% 50 mg/ml 20 mg/ml 500 ng/ml 20 mg/ml 500 ng/ml 0.1 mg/ml 50 mg/ml
#### Transport Media
The following transport media were tested in the BinaxNOW® Influenza A & B Test as negative samples (no virus present) and after inoculation with the LOD levels of Influenza A & B. Media did not impact BinaxNOW test performance, with the media alone testing negative in the NOW test and media inoculated with LOD Influenza A & B testing positive on the appropriate test line in BinaxNOW" test.
Amies Media Hank's Balanced Salt Solution M4 Media M4-RT Media M5 Media Stuart's Media Saline
#### Reproducibility
A blind study of the BinaxNOW® Influenza A & B Test was conducted at 3 separate sites using panels of blind coded specimens containing negative, low positive, and moderate positive samples. Participants tested each sample multiple times on 3 different days. There was 97% (242/250) agreement with expected test results, with no significant differences within run (replicates tested by one operator), between run (3 different days) or between sites (3 sites),
{7}------------------------------------------------
Signed
Rita Calnan
Date _ 04-Nov-2005
Rita Calnan Director of Quality & Regulatory Affairs
1) Dowdle, W.R., Kendal, A.P., and Noble, G.R. 1980. Influenza Virus, p 836-884. Manual of Clinical Microbiology, 3rd
edition, In Lennette, et. Al (ed.). American So
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a symbol. The symbol consists of four stylized, curved lines that resemble a human figure or a flame. The text is in all capital letters.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
NOV 3 0 2005
Ms. Angela Drysdale Clinical Affairs Specialist Binax, Inc. 10 Southgate Road Portland, Maine 04103
Re: k053126
Trade/Device Name: BinaxNOW® Influenza A & B Test Regulation Number: 21 CFR 866.3330 Regulation Name: Influenza Virus Serological Reagents Regulatory Class: Class II Product Code: GNX Dated: November 4, 2005 Received: November 7, 2005
Dear Ms. Drysdale:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{9}------------------------------------------------
#### Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally premated predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Sally a Hogg
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
# STATEMENT OF INDICATIONS FOR USE
510(k) Number (if known):
K053126
Device Name: BinaxNOW® Influenza A & B Test
Indications For Use:
The BinaxNOW® Influenza A & B Test is an in vitro immunochromatographic assay for the qualitative detection of influenza A and B nucleoprotein antigens in nasopharyngeal (NP) swab and nasal wash/aspirate specimens. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. Negative test results should be confirmed by culture.
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Prescription Use
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Per 21 CFR 801.109) (Optional Format 1-2-96)
Sal, arl
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510/k)_ K 0 5 312 6 -----
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.