BD Vaginal Panel

K243725 · Becton, Dickinson and Company · PQA · Dec 19, 2024 · Microbiology

Device Facts

Record IDK243725
Device NameBD Vaginal Panel
ApplicantBecton, Dickinson and Company
Product CodePQA · Microbiology
Decision DateDec 19, 2024
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 866.3975
Device ClassClass 2

Intended Use

The BD Vaginal Panel is an automated qualitative in vitro diagnostic test for the direct detection of DNA targets from bacteria associated with bacterial vaginosis (qualitative results reported based on detection and quantitation of targeted organism markers), Candida species associated with vulvovaginal candidiasis, and/or Trichomonas vaginalis. The test utilizes real-time polymerase chain reaction (PCR) for the amplification of specific DNA targets and utilizes fluorogenic target-specific hybridization probes to detect and differentiate DNA from: - Bacterial vaginosis (BV) markers (Individual markers are not reported) o Lactobacillus spp. (L. crispatus and L. jensenii) o Gardnerella vaginalis o Atopobium vaginae o Bacterial Vaginosis Associated Bacteria-2 (BVAB-2) o Megasphaera-1 - Vulvovaginal candidiasis (VVC) markers (Markers are reported in the following groups) o Candida spp. (C. albicans, C. tropicalis, C. parapsilosis, C. dubliniensis) o Candida glabrata o Candida krusei - Trichomonas vaginalis (TV) The assay may be used to detect DNA associated with BV, VVC, and TV, or a subset of these conditions per clinician order, in vaginal swab specimens collected from patients who are symptomatic for vaginitis/vaginosis. The BD Vaginal Panel is intended to aid in the diagnosis of vaginal infections in women with a clinical presentation consistent with bacterial vaginosis, vulvovaginal candidiasis and trichomoniasis. The BD Vaginal Panel is available for use with the BD MAX™ System or the BD COR™ System.

Device Story

BD Vaginal Panel; nucleic acid-based multiplex test system. Modification allows users to select specific assays for reporting based on clinician orders. Operates via clinical laboratory instrumentation; processes vaginal samples to detect specific microbial targets. Output provides diagnostic information to clinicians for patient management. Used in clinical laboratory settings by trained technicians. Modification maintains fundamental scientific technology of previously cleared device.

Clinical Evidence

No new clinical data presented. Clinical utility is unchanged from DEN160001. The manufacturer asserts that enabling result masking for subsets of conditions does not impact analytical or clinical performance, as the detection logic for each condition remains independent and unchanged.

Technological Characteristics

Real-time PCR assay; uses fluorogenic TaqMan hydrolysis probes. Automated extraction via magnetic affinity beads with lyticase. Compatible with BD MAX and BD COR systems. Software-controlled result reporting allows masking of specific targets. No changes to materials or sensing principles compared to predicate.

Indications for Use

Indicated for the detection of nucleic acid sequences from microorganisms associated with vaginitis, bacterial vaginosis, and Trichomonas vaginalis.

Regulatory Classification

Identification

A device that detects nucleic acid sequences from microorganisms associated with vaginitis and bacterial vaginosis is a qualitative in vitro diagnostic device intended for the detection of microbial nucleic acid sequences in vaginal specimens collected from patients with signs and symptoms of vaginitis or bacterial vaginosis. This device is intended to aid in the diagnosis of vaginitis or bacterial vaginosis when used in conjunction with clinical signs and symptoms and other laboratory findings.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include: (i) Documentation with a detailed device description of device components; ancillary reagents required but not provided; and explanation of the methodology including primer/probe sequence, design, and rationale for sequence selection. (ii) Documentation with information that demonstrates the performance characteristics of the device, including: (A) Limit of Detection; (B) Precision (reproductivity); (C) Analytical specificity; (D) Analytical reactivity (inclusivity); (E) Specimen stability; and (F) Effects of interfering substances. (iii) Detailed documentation from a prospective clinical study. As appropriate to the intended use, the prospective clinical study must be performed on an appropriate study population, including women of various ages and ethnicities. The prospective clinical study must compare the device performance to results obtained from well-accepted comparator methods. (iv) Detailed documentation for device software, including software applications and hardware-based devices that incorporate software. (2) The labeling required under § 809.10(b) of this chapter must include: (i) A detailed explanation of the interpretation of results and acceptance criteria; (ii) For devices with an intended use that includes detection of nucleic acid sequences from bacteria associated with bacterial vaginosis, clinical performance stratified by patient demographics such as race, ethnicity, age, and pregnancy status. (iii) For devices with an intended use that includes detection of nucleic acid sequences from bacteria associated with bacterial vaginosis, a summary of device results in an asymptomatic population with demographic characteristics appropriate to the intended use population. (iv) For devices with an intended use that includes detection of either Candida species or bacteria associated with bacterial vaginosis, a limitation that *Candida* species and bacterial compositions associated with bacterial vaginosis can be present as part of normal vaginal flora and results should be considered in conjunction with available clinical information.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ December 19, 2024 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Becton, Dickinson and Company Joseph Basore Staff Regulatory Affairs Specialist 7 Loveton Circle Sparks, Maryland 21152 ### Re: K243725 Trade/Device Name: BD Vaginal Panel Regulation Number: 21 CFR 866.3975 Regulation Name: Device That Detects Nucleic Acid Sequences From Microorganisms Associated With Vaginitis And Bacterial Vaginosis Regulatory Class: Class II Product Code: PQA, OUY, OOI, NSU Dated: November 22, 2024 Received: December 3, 2024 Dear Joseph Basore: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2}------------------------------------------------ Sincerely, Bryan M. Digitally signed by Bryan M. Bryan M. Grabias -S Grabias -S 12:29:18-05'00' Bryan Grabias, Ph. D. Acting Branch Chief Bacterial Respiratory and Medical Countermeasures Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K243725 Device Name BD Vaginal Panel; BD MAX™ Vaginal Panel #### Indications for Use (Describe) The BD Vaginal Panel is an automated qualitative in vitro diagnostic test for the direct detection of DNA targets from bacteria associated with bacterial vaginosis (qualitative results reported based on detection and quantitation of targeted organism markers), Candida species associated with vulvovaginal candidiasis, and/or Trichomonas vaginalis. The test utilizes real-time polymerase chain reaction (PCR) for the amplification of specific DNA targets and utilizes fluorogenic target-specific hybridization probes to detect and differentiate DNA from: - · Bacterial vaginosis (BV) markers (Individual markers are not reported) - o Lactobacillus spp. (L. crispatus and L. jensenii) - o Gardnerella vaginalis - o Atopobium vaginae - o Bacterial Vaginosis Associated Bacteria-2 (BVAB-2) - o Megasphaera-1 - · Vulvovaginal candidiasis (VVC) markers (Markers are reported in the following groups) o Candida spp. (C. albicans, C. tropicalis, C. parapsilosis, C. dubliniensis) - o Candida glabrata - o Candida krusei - · Trichomonas vaginalis (TV) The assay may be used to detect DNA associated with BV, VVC, and TV, or a subset of these conditions per clinician order, in vaginal swab specimens collected from patients who are symptomatic for vaginitis/vaginosis. The BD Vaginal Panel is intended to aid in the diagnosis of vaginal infections in women with a clinical presentation consistent with bacterial vaginosis, vulvovaginal candidiasis and trichomoniasis. The BD Vaginal Panel is available for use with the BD MAX™ System or the BD COR™ System. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ The BD MAX™ Vaginal Panel performed on the BD MAX™ System is an automated qualitative in vitro diagnostic test for the direct detection of DNA targets from bacterial vaginosis (qualitative results reported based on detection and of targeted organism markers), Candida species associated with vulvovaginal candidiasis, and/or Trichomonas vaginalis. The test utilizes real-time polymerase chain reaction (PCR) for the amplification of specific DNA targets and utilizes fluorogenic target-specific hybridization probes to detect and differentiate DNA from: - · Bacterial vaginosis markers (Individual markers not reported) - o Lactobacillus spp. (L. crispatus and L. jensenii) - o Gardnerella vaginalis - o Atopobium vaginae - o Bacterial Vaginosis Associated Bacteria-2 (BVAB-2) - o Megasphaera-1 - · Vulvovaginal candidiasis (VVC) markers (markers are reported in the following groups) - o Candida spp. (C. albicans, C. tropicalis, C. parapsilosis, C.dubliniensis) - o Candida glabrata - o Candida krusei {4}------------------------------------------------ #### • Trichomonas vaginalis (TV) The assay may be used to detect BV, VVC, and/or TV, or a subset of these conditions per clinician order, in vaginal swab specimens collected from patients who are symptomatic for vaginitis/vaginosis. The BD MAX™ Vaginal is intended to aid in the diagnosis of vaginal infections in women with a clinical presentation consistent with bacterial vaginosis, vulvovaginal candidiasis and trichomoniasis. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ #### 510(k) Summary BD Vaginal Panel #### Summary Preparation Date: 11/22/2024 #### Submitted by: BD Diagnostic Solutions Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152 #### Contact: Joseph Basore, Ph.D., RAC Staff Regulatory Affairs Specialist Tel: 616-301-4068 Email: Joseph.Basore@bd.com #### Proprietary Names: For the instrument: BD COR™ System, BD MAXTM System For the assay: BD Vaginal Panel, BD MAXTM Vaginal Panel #### Common Names: For the instrument: High-throughput molecular system Bacterial Vaginosis Assay For the assay: Vaginitis Assay TV Assay Candida Assay #### Regulatory Information Regulation section: 21 CFR 866.3975 – Device that detects nucleic acid sequences from microorganisms associated with vaginitis and bacterial vaginosis Classification: Class II (Special Controls) Panel: Microbiology (83) {6}------------------------------------------------ ### Product Code(s): - POA Vaginitis and Bacterial Vaginosis Nucleic Acid Detection System - OUY Trichomonas vaginalis Nucleic Acid Amplification Test System - Real Time Nucleic Acid Amplification System 00I - Instrumentation for Clinical Multiplex Test Systems NSU ### Predicate Device BD Vaginal Panel (DEN160001, K191957, K201017, K223653) ### Device Establishment Registration Number: 1119779 ### Intended Use ### BD Vaginal Panel (For use with BD COR™ System) The BD Vaginal Panel is an automated qualitative in vitro diagnostic test for the direct detection of DNA targets from bacteria associated with bacterial vaginosis (qualitative results reported based on detection and quantitation of targeted organism markers), Candida species associated with vulvovaginal candidiasis, and/or Trichomonas vaginalis. The test utilizes real-time polymerase chain reaction (PCR) for the amplification of specific DNA targets and utilizes fluorogenic target-specific hybridization probes to detect and differentiate DNA from: - Bacterial vaginosis (BV) markers (Individual markers are not reported) ● - Lactobacillus spp. (L. crispatus and L. jensenii) - Gardnerella vaginalis - Atopobium vaginae ● - Bacterial Vaginosis Associated Bacteria-2 (BVAB-2) ● - Megasphaera-1 ● - Vulvovaginal candidiasis (VVC) markers (Markers are reported in the following groups) ● - Candida spp. (C. albicans, C. tropicalis, C. parapsilosis, C. dubliniensis) ● - Candida glabrata ● - Candida krusei . - Trichomonas vaginalis (TV) . The assay may be used to detect DNA associated with BV, VVC, and TV, or a subset of these conditions per clinician order, in vaginal swab specimens collected from patients who are symptomatic for vaginitis/vaginosis. The BD Vaginal Panel is intended to aid in the diagnosis of vaginal infections in women with a clinical presentation consistent with bacterial vaginosis, vulvovaginal candidiasis and trichomoniasis. {7}------------------------------------------------ The BD Vaginal Panel is available for use with the BD MAX™ System or the BD COR™ System. ### BD MAX™ Vaginal Panel The BD MAX™ Vaginal Panel performed on the BD MAX™ System is an automated qualitative in vitro diagnostic test for the direct detection of DNA targets from bacteria associated with bacterial vaginosis (qualitative results reported based on detection and of targeted organism markers), Candida species associated with vulvovaginal candidiasis, and/or Trichomonas vaginalis. The test utilizes real-time polymerase chain reaction (PCR) for the amplification of specific DNA targets and utilizes fluorogenic target-specific hybridization probes to detect and differentiate DNA from: - Bacterial vaginosis (BV) markers (Individual markers are not reported) ● - Lactobacillus spp. (L. crispatus and L. jensenii) - Gardnerella vaginalis ● - Atopobium vaginae ● - Bacterial Vaginosis Associated Bacteria-2 (BVAB-2) - Megasphaera-1 ● - Vulvovaginal candidiasis (VVC) markers (Markers are reported in the following groups) ● - Candida spp. (C. albicans, C. tropicalis, C. parapsilosis, C.dubliniensis) - Candida glabrata - Candida krusei ● - Trichomonas vaginalis ● The assay may be used to detect DNA associated with BV, VVC, and TV, or a subset of these conditions per clinician order, in vaginal swab specimens collected from patients who are symptomatic for vaginitis/vaginosis. The BD MAX™ Vaginal Panel is intended to aid in the diagnosis of vaginal infections in women with a clinical presentation consistent with bacterial vaginosis, vulvovaginal candidiasis and trichomoniasis. ### Special Conditions for Use Statement: For Prescription Use Only Special Instrument Requirements: BD COR™ System, BD MAX™ System ### Device Description {8}------------------------------------------------ The BD Vaginal Panel is an automated, qualitative in vitro diagnostic test for bacterial vaginosis markers. Candida species associated with vulyovaginal candidiasis, and Trichomonas vaginalis. From a single specimen, a vaginal swab collected from a patient who is symptomatic for vaginitis or vaginosis, the test provides qualitative (positive) results for the presence of the following conditions and/or organisms: - Bacterial vaginosis (based on detection of Lactobacillus spp. (L. crispatus and L. . jensenii), Gardnerella vaginalis, Atopobium vaginae, Bacterial Vaginosis Associated Bacteria-2, and Megasphaera-1) - Candida spp. (C. albicans, c. tropicalis, C. parapsilosis, and C. dubliniensis) . - Candida glabrata ● - Candida krusei ● - Trichomonas vaginalis ● The BD Vaginal Panel is available for use with the BD MAX™ System or the BD COR™ System under the trade names BD MAX™ Vaginal Panel and BD Vaginal Panel, respectively. The assay previously provided results for all five targets for every specimen run, regardless of whether the ordering clinician desires evaluation of all reported conditions. After clearance of this 510(k) submission, BD will release updates to the BD MAX™ and BD COR ™ software to allow users the versatility to mask results, per specimen, based on the order received from the clinician. This change allows a laboratory to use the BD Vaginal Panel to test a specimen for an individual target group (BV, VVC, or TV) or any two of the three targeted conditions, in addition to the current configuration that reports all three conditions simultaneously. ### Test Principle The BD Vaginal Panel is designed for use with the BD Molecular Swab Collection Kit. Samples are transported to the testing laboratory in Sample Buffer Tubes. The Sample Buffer Tubes are vortexed to release cells from the swab into the buffer, after which they are loaded on the BD MAX™ System or the BD COR™ System, along with instrument-specific consumables. No further operator intervention is necessary, and the following automated procedures occur. The bacterial cells are lysed, and DNA is extracted, captured, and concentrated on magnetic beads. After an elution step, the DNA solution is added to reagents containing specific primers and probes used to amplify and detect the genetic targets if present. After reconstitution, the instrument dispenses a fixed volume of PCR-ready solution containing extracted nucleic acids into the PCR Cartridge. Microvalves in the cartridge are sealed by the system prior to initiating PCR to contain the amplification mixture and thus prevent evaporation and contamination. Upon amplification, signal detection and interpretation are performed automatically by the instrument using real time PCR. The Extraction reagent includes a Sample Processing Control, which monitors the integrity of the reagents as well as the process steps involved in DNA extraction, amplification, and detection, and checks for the presence of potential assay inhibitors. The amplified DNA targets are detected using hydrolysis (TaqMan®) probes, labeled at one end with a fluorescent reporter dye (fluorophore), and at the other end, with a quencher moiety. {9}------------------------------------------------ Probes labeled with different fluorophores are used to detect the target analytes in different optical channels of the instrument. When the probes are in their native state, the fluorescence of the fluorophore is quenched due to its proximity to the quencher. However, in the presence of target DNA, the probes hybridize to their complementary sequences and are hydrolyzed by the 5'-3' exonuclease activity of the DNA polymerase as it synthesizes the nascent strand along the DNA template. As a result, the fluorophores are separated from the quencher molecules and fluorescence is emitted. The amount of fluorescence detected in the optical channels used for the BD Vaginal Panel is directly proportional to the quantity of the corresponding probe that is hydrolyzed. The instrument monitors these signals at each cycle of the PCR and interprets the data at the end of the reaction to provide qualitative test results for each analyte (i.e., positive or negative). The BD Vaginal Panel is divided into two Master Mixes, as described below. - The Vaginitis Master Mix contains reagents for detection of the following targets: . - Candida spp (C. albicans, C. tropicalis, C. parapsilosis and C. dubliniensis) ● - Candida glabrata - Candida krusei - Trichomonas vaginalis - The Vaginosis Master Mix contains reagents for detection of the following . bacterial vaginosis markers: - Gardnerella vaginalis ● - Lactobacillus crispatus and Lactobacillus jensenii ● - Atopobium vaginae ● - Bacterial Vaginosis Associated Bacteria-2 (BVAB-2) ● - Megasphaera-1 ● ### Substantial Equivalence1 Table 1 provides the similarities and differences between the submitted device and the legally marketed predicate device. <sup>1</sup> The term "substantial equivalence" as used in this 510(k) notification is limited to the definition of substantial equivalence as found in the Federal Food, Drug and Cosmetic Act, as amended and as applied under 21 CFR 807, Subpart E under which a device can be marketed without pre-market approval or reclassification of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of substantial equivalence herein shall be construed as an admission against interest under the US Patent Laws or their application by the courts. {10}------------------------------------------------ BD Diagnostic Solutions | Items | Predicate - BD Vaginal Panel<br>(K223653) | Subject -BD MAX Vaginal Panel | Subject - BD Vaginal Panel (For use<br>with BD CORT™ System) | | |-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------| | Regulation | 866.3975 | Same | Same | | | Product Code | PQA, OUY, OOI, NSU | Same | Same | | | Device Class | II | Same | Same | | | Intended Use | The BD Vaginal Panel is an automated<br>qualitative <i>in vitro</i> diagnostic test for the<br>direct detection of DNA targets from<br>bacteria associated with bacterial<br>vaginosis (qualitative results reported<br>based on detection and quantitation of<br>targeted organism markers), <i>Candida</i><br>species associated with vulvovaginal<br>candidiasis, and <i>Trichomonas vaginalis</i><br>from vaginal swabs in patients who are<br>symptomatic for vaginitis/vaginosis. The<br>test utilizes real-time polymerase chain<br>reaction (PCR) for the amplification of<br>specific DNA targets and utilizes<br>fluorogenic target-specific hybridization<br>probes to detect and differentiate DNA<br>from: | The BD MAX™ Vaginal Panel<br>performed on the BD MAX™ System is<br>an automated qualitative <i>in vitro</i><br>diagnostic test for the direct detection of<br>DNA targets from bacteria associated<br>with bacterial vaginosis (qualitative<br>results reported based on detection and of<br>targeted organism markers), <i>Candida</i><br>species associated with vulvovaginal<br>candidiasis, and/or <i>Trichomonas</i><br><i>vaginalis</i> . The test utilizes real-time<br>polymerase chain reaction (PCR) for the<br>amplification of specific DNA targets and<br>utilizes fluorogenic target-specific<br>hybridization probes to detect and<br>differentiate DNA from: | The BD Vaginal Panel is an automated<br>qualitative <i>in vitro</i> diagnostic test for the<br>direct detection of DNA targets from<br>bacteria associated with bacterial<br>vaginosis (qualitative results reported<br>based on detection and quantitation of<br>targeted organism markers), <i>Candida</i><br>species associated with vulvovaginal<br>candidiasis, and/or <i>Trichomonas</i><br><i>vaginalis</i> . The test utilizes real-time<br>polymerase chain reaction (PCR) for the<br>amplification of specific DNA targets and<br>utilizes fluorogenic target-specific<br>hybridization probes to detect and<br>differentiate DNA from: | | | | • Bacterial vaginosis markers<br>(Individual markers not reported)<br>○ Lactobacillus spp. ( <i>L.</i><br><i>crispatus</i> and <i>L. jensenii</i> )<br>○ <i>Gardnerella vaginalis</i><br>○ <i>Atopobium vaginae</i><br>○ Bacterial Vaginosis<br>Associated Bacteria-2<br>○<br>(BVAB-2)<br>○ <i>Megasphaera-1</i> | • Bacterial vaginosis (BV) markers<br>(Individual markers are not reported)<br>○ Lactobacillus spp. ( <i>L.</i><br><i>crispatus</i> and <i>L. jensenii</i> )<br>○ <i>Gardnerella vaginalis</i><br>○ <i>Atopobium vaginae</i><br>○ Bacterial Vaginosis<br>Associated Bacteria-2<br>○<br>(BVAB-2)<br>○ <i>Megasphaera-1</i> | • Bacterial vaginosis (BV) markers<br>(Individual markers are not reported)<br>○ Lactobacillus spp. ( <i>L.</i><br><i>crispatus</i> and <i>L. jensenii</i> )<br>○ <i>Gardnerella vaginalis</i><br>○ <i>Atopobium vaginae</i><br>○ Bacterial Vaginosis<br>Associated Bacteria-2<br>○<br>(BVAB-2)<br>○ <i>Megasphaera-1</i> | | | Items | Predicate - BD Vaginal Panel<br>(K223653) | Subject -BD MAX Vaginal Panel | Subject - BD Vaginal Panel (For use<br>with BD CORTM System) | | | | Candida spp. (C. albicans, C. tropicalis, C. parapsilosis, C. dubliniensis) Candida glabrata Candida krusei Trichomonas vaginalis The BD Vaginal Panel is intended to aid in the diagnosis of vaginal infections in women with a clinical presentation consistent with bacterial vaginosis, vulvovaginal candidiasis and trichomoniasis.<br><br>The BD Vaginal Panel is available for use with the BD MAXTM System or the BD CORTM System. | Vulvovaginal candidiasis (VVC) markers (markers are reported in the following groups) Candida spp. (C. albicans, C. tropicalis, C. parapsilosis, C.dubliniensis) Candida glabrata Candida krusei Trichomonas vaginalis (TV) The assay may be used to detect BV, VVC, and/or TV, or a subset of these conditions per clinician order, in vaginal swab specimens collected from patients who are symptomatic for vaginitis/vaginosis. The BD MAXTM Vaginal Panel is intended to aid in the diagnosis of vaginal infections in women with a clinical presentation consistent with bacterial vaginosis, vulvovaginal candidiasis and trichomoniasis. | Vulvovaginal candidiasis (VVC) markers (Markers are reported in the following groups) Candida spp. (С. albicans, C. tropicalis, C. parapsilosis, C. dubliniensis) Candida glabrata Candida krusei Trichomonas vaginalis (TV) The assay may be used to detect DNA associated with BV, VVC, and TV, or a subset of these conditions per clinician order, in vaginal swab specimens collected from patients who are symptomatic for vaginitis/vaginosis. The BD Vaginal Panel is intended to aid in the diagnosis of vaginal infections in women with a clinical presentation consistent with bacterial vaginosis, vulvovaginal candidiasis and trichomoniasis.<br><br>The BD Vaginal Panel is available for use with the BD MAXTM System or the BD CORTM System. | | | Indications for Use | Symptomatic patients | Same | Same | | | Specimen Type | Clinician and patient-collected female vaginal swab | Same | Same | | | Collection/ Transport<br>Device | BD Molecular Swab Collection Kit | Same | Same | | | Technology | PCR | Same | Same | | | Items | Predicate - BD Vaginal Panel (K223653) | Subject -BD MAX Vaginal Panel | Subject - BD Vaginal Panel (For use with BD CORT™ System) | | | Organisms Detected | <i>Lactobacillus</i> spp. ( <i>L. crispatus</i> and <i>L.jensenii</i> ) <i>Gardnerella vaginalis Atopobium vaginae</i> Bacterial Vaginosis Associated Bacteria-2 (BVAB-2) <i>Megasphaera-1 Candida</i> spp. ( <i>C. albicans, C. tropicalis, C. parapsilosis, C. dubliniensis</i> ) <i>Candida glabrata Candida krusei Trichomonas vaginalis</i> | Same | Same | | | Sample Prep/ Results | BD MAX™ System or BD CORT™ System | BD MAX™ System | BD CORT™ System | | | Assay Controls | Sample Processing Control | Same | Same | | | Target Detection | Channel – Dyes | Targets | | | | | ROX | Vaginitis MM<br>Candida group1 | Vaginosis MM<br><i>Lactobacillus crispatus/jensenii</i> | | | | FAM | <i>Trichomonas vaginalis</i> | <i>Atopobium vaginae</i> | | | | VIC – Cal Fluor Orange 560 | <i>Candida glabrata</i> | <i>Megasphaera a-1/ BVAB-2</i> | | | Items | Predicate - BD Vaginal Panel<br>(K223653) | Subject -BD MAX Vaginal Panel | Subject - BD Vaginal Panel (For use<br>with BD COR™ System) | | | | Cy5 –<br>Quasar<br>670 | | | | | | Candida<br>krusei | | | | | | Gardnerella<br>vaginalis | | | | | | Cy5.5 –<br>Quasar<br>705 | | | | | | SPC | |…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...