FilmArray RP EZ Control Panel M265

K161573 · Maine Molecular Quality Controls, Inc. · PMN · Sep 2, 2016 · Microbiology

Device Facts

Record IDK161573
Device NameFilmArray RP EZ Control Panel M265
ApplicantMaine Molecular Quality Controls, Inc.
Product CodePMN · Microbiology
Decision DateSep 2, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3920
Device ClassClass 2

Intended Use

FilmArray RP EZ Control Panel M265 is intended for use as an external positive and negative assayed quality control to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Adenovirus, Coronavirus, Human Metapneumovirus, Human Rhinovirus/ Enterovirus, Influenza A, Influenza A subtype H1, Influenza A subtype H3, Influenza A subtype H1-2009, Influenza B, Parainfluenza Virus, Respiratory Syncytial Virus, Bordetella pertussis, Chlamydophila pneumoniae, and Mycoplasma pneumoniae on the FilmArray RP EZ assay performed on the FilmArray systems. FilmArray RP Positive Control is composed of synthetic RNA transcripts specifically designed for and intended to be used solely with the FilmArray RP EZ assay. This product is not intended to replace manufacturer controls provided with the device.

Device Story

FilmArray RP EZ Control Panel M265 provides external positive and negative quality control materials for the FilmArray RP EZ assay. Positive control consists of synthetic RNA transcripts representing all respiratory pathogens detected by the assay; negative control consists of non-specific RNA in buffer. Controls are processed by laboratory personnel identically to patient nasopharyngeal swab samples in Viral Transport Media. The device is used on the FilmArray 2.0 instrument to monitor qualitative amplification, detection, and identification steps. Results are reviewed by clinicians to verify assay performance and ensure validity of patient test results.

Clinical Evidence

No clinical data. Bench testing only. Reproducibility study (3 sites, 182 controls) showed 100% agreement for positive controls and 97.78% for negative controls. Internal precision and lot-to-lot studies confirmed consistent performance across operators, lots, and days, with stable Crossing Point (Cp) values.

Technological Characteristics

Ready-to-use liquid control panel. Positive control: synthetic RNA transcripts in buffers, stabilizers, and preservatives. Negative control: non-specific RNA in buffers, stabilizers, and preservatives. Designed for use with FilmArray 2.0 EZ Configuration system. Monitors extraction, reverse transcription, amplification, detection, and identification steps.

Indications for Use

Indicated for use as external positive and negative quality control material for in vitro nucleic acid testing of respiratory pathogens (Adenovirus, Coronavirus, Human Metapneumovirus, Human Rhinovirus/Enterovirus, Influenza A/B, Parainfluenza, RSV, B. pertussis, C. pneumoniae, M. pneumoniae) using the FilmArray RP EZ assay. Intended for clinical laboratory use.

Regulatory Classification

Identification

An assayed quality control material for clinical microbiology assays is a device indicated for use in a test system to estimate test precision or to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. This type of device consists of single or multiple microbiological analytes intended for use with either qualitative or quantitative assays.

Special Controls

An assayed quality control material for clinical microbiology assays must comply with the following special controls: (1) Premarket notification submissions must include detailed device description documentation and information concerning the composition of the quality control material, including, as appropriate: (i) Analyte concentration; Expected values: (ii) Analyte source: (iii) (iv) Base matrix; (v) Added components; (vi) Safety and handling information; and, (vii) Detailed instructions for use. (2) Premarket notification submissions must include detailed documentation, including line data as well as detailed study protocols and a statistical analysis plan used to establish performance, including: (i) Description of the process for value assignment and validation. (ii) Description of the protocol(s) used to establish stability. (iii) Line data establishing precision/reproducibility. (iv) Where applicable, assessment of matrix effects and any significant differences between the quality control material and typical patient samples in terms of conditions known to cause analytical error or affect assay performance. (v) Where applicable, identify or define traceability or relationship to a domestic or international standard reference material and/or method. (vi) Where applicable, detailed documentation related to studies for surrogate controls. (3) Premarket notification submissions must include an adequate mitigation (e.g., realtime stability program) to the risk of false results due to potential modifications to the assays specified in the device's 21 CFR 809.10 compliant labeling. (4) Your 21 CFR 809.10 compliant labeling must include the following: (i) The intended use in your 21 CFR 809.10(a)(2) and 21 CFR 809.10(b)(2) compliant labeling must include the following: (A) Assayed control material analyte(s); (B) Whether the material is intended for quantitative or qualitative assays: (C) Stating if the material is a surrogate control; (D)The system(s), instrument(s), or test(s) for which the quality control material is intended. (ii) The intended use in your 21 CFR 809.10(a)(2) and 21 CFR 809.10(b)(2) compliant labeling must include the following statement: "This product is not intended to replace manufacturer controls provided with the device." (iii)A limiting statement that reads "Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements."

*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed device description documentation and information concerning the composition of the quality control material, including, as appropriate: (i) Analyte concentration; (ii) Expected values; (iii) Analyte source; (iv) Base matrix; (v) Added components; (vi) Safety and handling information; and (vii) Detailed instructions for use. (2) Premarket notification submissions must include detailed documentation, including line data as well as detailed study protocols and a statistical analysis plan used to establish performance, including: (i) Description of the process for value assignment and validation. (ii) Description of the protocol(s) used to establish stability. (iii) Line data establishing precision/reproducibility. (iv) Where applicable, assessment of matrix effects and any significant differences between the quality control material and typical patient samples in terms of conditions known to cause analytical error or affect assay performance. (v) Where applicable, identify or define traceability or relationship to a domestic or international standard reference material and/or method. (vi) Where applicable, detailed documentation related to studies for surrogate controls. (3) Premarket notification submissions must include an adequate mitigation (e.g., real-time stability program) to the risk of false results due to potential modifications to the assays specified in the device's 21 CFR 809.10 compliant labeling. (4) Your 21 CFR 809.10 compliant labeling must include the following: (i) The intended use of your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following: (A) Assayed control material analyte(s); (B) Whether the material is intended for quantitative or qualitative assays; (C) Stating if the material is a surrogate control; and (D) The system(s), instrument(s), or test(s) for which the quality control material is intended. (ii) The intended use in your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following statement: “This product is not intended to replace manufacturer controls provided with the device.” (iii) A limiting statement that reads “Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white. Public Health Service September 2, 2016 Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Maine Molecular Quality Controls, Inc. Joan Gordon President 23 Mill Brook Road Saco, ME 04072 Re: K161573 Trade/Device Name: FilmArray RP EZ Control Panel M265 Regulation Number: 21 CFR 866.3920 Regulation Name: Assayed quality control material for clinical microbiology assays Regulatory Class: Class II (Special Controls) Product Code: PMN Dated: June 6, 2016 Received: June 7, 2016 Dear Ms. Gordon: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the {1}------------------------------------------------ Page 2 - Ms. Gordon electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Steven R. Gitterman -S for Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K161573 Device Name FilmArray RP EZ Control Panel M265 #### Indications for Use (Describe) FilmArray RP EZ Control Panel M265 is intended for use as an external positive and negative assayed quality control to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Adenovirus, Coronavirus, Human Metapneumovirus, Human Rhinovirus, Influenza A, Influenza A subtype H1, Influenza A subtype H3, Influenza A subtype H1-2009, Influenza Virus, Respiratory Syncytial Virus, Bordetella pertussis, Chlamydophila pneumoniae, and Mycoplasma pneumoniae on the FilmArray RP EZ assay performed on the FilmArray systems. FilmArray RP Positive Control is composed of synthetic RNA transcripts specifically designed for and intended to be used solely with the FilmArray RP EZ assay. This product is not intended to replace manufacturer controls provided with the device. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for MMQCI, which stands for Maine Molecular Quality Controls, Inc. The logo features the letters "MMQCI" in a combination of maroon and yellow. The "MM" and "CI" are in maroon, while the "Q" is in yellow. Below the letters, the full name of the company, "MAINE MOLECULAR QUALITY CONTROLS, INC.", is written in a smaller font size. A yellow brushstroke-like design is on the left side of the logo. 23 Mill Brook Road Mill Brook Business Park Saco, Maine 04072 www.mmaci.com Phone: 207-885-1072 Fax: 207-885-1079 510 (k) Summary #### 510(k) Number: K161573 Purpose for submission: New product ### Applicant Information: | Applicant: | Maine Molecular Quality Controls, Inc. | |------------|-----------------------------------------| | Address: | 23 Mill Brook Road<br>Saco, Maine 04072 | | Contact Person: | Joan Gordon, President MMQCI | | |-----------------|------------------------------|--| | Phone: | 207-885-1072 extension 201 | | | Fax: | 207-885-1079 | | | Email Address: | jgordon@mmqci.com | | Preparation Date: May 31, 2016 #### Device | Device Trade Name: | FilmArray RP EZ Control Panel M265, model M265 | |---------------------|-------------------------------------------------------------------| | Device Common Name: | Quality Control Material for Microbiology Assays | | Device Type: | Assayed quality control material for clinical microbiology assays | | Class: | Class II (Special controls) | | Regulation: | 21 CFR 866.3920 | | Panel: | Microbiology - 83 | | Product code: | PMN | #### Predicate Device DEN150058; Amplichek II, Bio-Rad Laboratories #### Device Description FilmArray RP EZ Control Panel M265, P/N M265, is a quality control panel consisting of 2 controls, FilmArray RP EZ Positive Control, P/N M266, and FilmArray RP EZ Negative Control. P/N M267. The Positive Contains non-infectious surrogate control material: a solution of synthetic RNA transcripts in buffers, stabilizers and preservatives. The RNA carries segments of all respiratory pathogens detected by the FilmArray RP EZ assay (Table 1. below) on the FilmArray 2.0 EZ Configuration system. The RNA in the Negative Control is non-specific RNA in buffers, stabilizers and preservatives. Each liquid control of FilmArray RP EZ Control Panel M265 is processed separately according to FilmArray RP EZ assay manufacturer's Instructions for Use for patient samples (nasopharyngeal swabs {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Maine Molecular Quality Controls, Inc. (MMQCI). The logo features the acronym "MMQCI" in bold, dark red letters. Below the acronym, the full name of the company, "Maine Molecular Quality Controls, Inc.," is written in a smaller, lighter font. A stylized, abstract graphic in shades of yellow and gold is positioned to the left of the text, adding a visual element to the logo. (NPS) obtained from individuals suspected of respiratory tract infection and placed in Viral Transport Media (VTM)). | Respiratory Pathogens | | |-------------------------------|-----------------------------| | Adenovirus | Influenza A H1-2009 | | Coronavirus | Influenza B | | Human Metapneumovirus | Parainfluenza Virus | | Human Rhinovirus/ Enterovirus | Respiratory Syncytial Virus | | Influenza A | Bordetella pertussis | | Influenza A subtype H1 | Chlamydophila pneumoniae | | Influenza A subtype H3 | Mycoplasma pneumoniae | Table 1. Respiratory pathogens detected by FilmArray RP EZ assay ### Device Intended Use FilmArray RP EZ Control Panel M265 is intended for use as an external positive and negative assayed quality control to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Adenovirus, Coronavirus, Human Metapneumovirus, Human Rhinovirus/ Enterovirus, Influenza A, Influenza A subtype H1, Influenza A subtype H3, Influenza A subtype H1-2009, Influenza B, Parainfluenza Virus, Respiratory Syncytial Virus, Bordetella pertussis, Chlamydophila pneumoniae, and Mycoplasma pneumoniae on the FilmArray RP EZ assay performed on the FilmArray systems. FilmArray RP Positive Control is composed of synthetic RNA transcripts specifically designed for and intended to be used solely with the FilmArray RP EZ assay. This product is not intended to replace manufacturer controls provided with the device. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for MMQCI, which stands for Maine Molecular Quality Controls, Inc. The logo features the letters "MMQCI" in a serif font, with "MM" in a larger size and a darker color than "QCI". Below the letters, the full name of the company is written in a smaller font. To the left of the letters, there is an abstract design in yellow, resembling a brush stroke or a stylized wave. | Characteristic | Candidate Device:<br>FilmArray RP EZ<br>Control Panel M265<br>(K161573) | Predicate Device<br>Amplichek II<br>(DEN150058) | |---------------------------------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------| | Intended Use | External assayed quality<br>control to monitor in vitro<br>lab nucleic acid test | Same | | Physical format | Ready-to-Use Liquid | Same | | Directions for Use | Process like patient sample | Same | | Composition | Synthetic RNA transcripts | Intact microorganisms | | Assay Steps Monitored | Extraction, reverse<br>transcription,<br>amplification, detection,<br>identification | Same except does not<br>monitor extraction<br>(new device not<br>encapsulated) | | Number of targets<br>monitored in one assay | Multiple | Same | # Substantial Equivalence #### Summary Performance Data #### All Test Results Three lots of FilmArray RP EZ Control Panel M265, FilmArray RP EZ Positive Control and FilmArray RP EZ Negative Control, were manufactured and tested using the FilmArray RP EZ assay on the FilmArray instrument 2.0 EZ Configuration system to confirm performance. The lots were manufactured incorporating variables including multiple key component lots, different operator and different days over four months. An internal study was performed by testing the 3 lots at MMQCI (Saco, Maine) over multiple days with 4 reagent lots by one operator. An external study was performed to assess the ability of the intended user to test the FilmArray RP EZ Control Panel M265 in a CLIA-waived setting. The external performance study tested the 3 lots of FilmArray RP EZ Control Panel M265 at 2 CLIA-waived sites and 1 external site under simulated CLIA-waived conditions over a period of 10 days using 3 FilmArray reagent lots, incorporating multiple operators. All FilmArray RP EZ Positive Control respiratory pathogen analytes were correctly detected in the internal and external study (Correct Positive Control Result). All but 2 FilmArray RP EZ Negative Controls gave Correct Negative Control Results of no respiratory pathogens detected. There were 3 Invalid results caused by internal control pouch failures and 2 false positive results. It is not possible to determine the cause of the 2 false positive results. However, since the results were from 2 different control lots and 2 different pouch lots at the same testing site, environmental contamination is possible, especially since testing occurred during flu season. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for MMQCI, which stands for Maine Molecular Quality Controls, Inc. The logo features the letters "MMQCI" in a combination of maroon and yellow colors. The word "MAINE MOLECULAR QUALITY CONTROLS, INC." is written in a smaller font size below the letters. A yellow, curved design element is placed behind the letters, adding a visual flair to the logo. | Table 1. Summary of All Test Results | | | | | | | | | |--------------------------------------|----------------|---------|------------------------------------------|--------------------------------------------|--------------------------------------------|------------------------------------------|--------------------------------------------|--------------------------------------------| | Number<br>of sites | Total<br>Tests | Invalid | Correct<br>Positive<br>Control<br>Result | Incorrect<br>Positive<br>Control<br>Result | Percent<br>Correct*<br>Positive<br>Control | Correct<br>Negative<br>Control<br>Result | Incorrect<br>Negative<br>control<br>Result | Percent<br>Correct*<br>Negative<br>Control | | 4 | 271 | 3 | 129 | 0 | 100% | 137 | 2 | 98.6% | *The 3 Invalid samples were re-tested according to BioFire instructions and were not included in the Percent Correct analysis. # External Site Testing Three lots of FilmArray RP EZ Control Panel M265, FilmArray RP EZ Positive Control and FilmArray RP EZ Negative Control, were manufactured by MMQCI. They were tested by the FilmArray RP EZ assay on the FilmArray instrument 2.0 EZ Configuration system at two CLIAwaived clinical settings and one simulated CLIA-waived test setting where users were screened to have training and educational backgrounds consistent with those in the CLIA-waived testing environment. Testing consisted of three positive and three negative controls run per day, and spanned a period of 10 days (total of 60 control runs per site). Multiple operators participated in testing at each of the three locations. Three lots each of external control material and three lots of RP EZ pouches were tested across all sites. A total of 182 external controls were tested (91 positive and 91 negative). Two tests were excluded from final data analysis due to Invalid results caused by internal pouch control failures. Data for the remaining 180 control tests are shown below. | Table 2. Summary of External Test Results for 3 Control Lots | | | | | | | | | |--------------------------------------------------------------|----------------|---------|------------------------------------------|--------------------------------------------|--------------------------------------------|------------------------------------------|--------------------------------------------|--------------------------------------------| | External<br>Site | Total<br>Tests | Invalid | Correct<br>Positive<br>Control<br>Result | Incorrect<br>Positive<br>Control<br>Result | Percent<br>Correct*<br>Positive<br>Control | Correct<br>Negative<br>Control<br>Result | Incorrect<br>Negative<br>control<br>Result | Percent<br>Correct*<br>Negative<br>Control | | 1 | 61 | 1 | 30 | 0 | 100% | 30 | 0 | 100% | | 2 | 61 | 1 | 30 | 0 | 100% | 30 | 0 | 100% | | 3 | 60 | 0 | 30 | 0 | 100% | 28 | 2 | 93.3% | | All Sites | 182 | 2 | 90 | 0 | 100% | 88 | 2 | 97.8% | *The 2 Invalid samples were re-tested according to BioFire instructions and were not included in the Percent Correct analysis. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for MMQCI, which stands for Maine Molecular Quality Controls, Inc. The logo features the letters "MMQCI" in a serif font, with the "MM" in a larger size than the "QCI". The words "MAINE MOLECULAR QUALITY CONTROLS, INC." are written in a smaller font below the letters. A yellow brush stroke design is on the left side of the logo. # Reproducibilitv NOTE: The FilmArray 2.0 is a random access instrument. Therefore, 'runs' refer to replicates tested on the same day. Within-run Testing: Within-run reproducibility was demonstrated by 1 operator testing 2 lots of FilmArray RP EZ Control Panel M265 with 1 lot of FilmArray RP EZ pouches on the FilmArray 2.0, each within one day. All results were correct and reproducible. | Table 3. Summary of Within-run Reproducibility Results | | | | | | |--------------------------------------------------------|--------------|--------------|-----------------|-----------|-----------------| | Control | Control Lot# | No. of Tests | Date of Testing | Pouch Lot | Correct Results | | FilmArray RP EZ Positive Control | A06JAN16A | 6 | 1-7-16 | 239115 | 6/6 | | FilmArray RP EZ Negative Control | M41DEC15A | 5 | 12-31-15 | 239115 | 5/5 | Lot-to-Lot Testing: Lot-to-lot reproducibility was demonstrated by testing 3 lots of FilmArray RP EZ Positive Control using the same pouch lot. All results were correct and reproducible. | Control Lot # | Number of Tests | Pouch Lot | Correct Results | |---------------|-----------------|-----------|-----------------| | M40DEC15A | 4 | 238915 | 4/4 | | A28DEC15A | 3 | 238915 | 3/3 | | A06JAN16A | 5 (1 Invalid) | 238915 | 4/4 | Precision Testing: Precision was demonstrated by testing 3 lots of FilmArray RP EZ Positive Control and 3 lots of FilmArray RP EZ Negative Control at MMQCI over multiple days with 4 reagent lots by one operator using one FilmArray 2.0 instrument. All FilmArray RP EZ Positive Control respiratory pathogen analytes were correctly detected. One test was excluded from final data analysis due to Invalid results caused by an internal pouch control failure. | Table 5. Summary of Precision Testing for 3 Control Lots at MMQCI | | | | | | | |-------------------------------------------------------------------|---------------|-----------------|---------|--------------------|----------------------|---------------------| | Control | Control Lot # | No. of<br>Tests | Invalid | Correct<br>Results | Incorrect<br>Results | Percent<br>Correct* | | FilmArray RP EZ Positive Control | M40DEC15A | 12 | 0 | 12 | 0 | 100% | | FilmArray RP EZ Positive Control | A28DEC15A | 13 | 0 | 13 | 0 | 100% | | FilmArray RP EZ Positive Control | A06JAN16A | 15 | 1 | 14 | 0 | 100% | | FilmArray RP EZ Negative Control | A29OCT15B | 16 | 0 | 16 | 0 | 100% | | FilmArray RP EZ Negative Control | M41DEC15A | 16 | 0 | 16 | 0 | 100% | | FilmArray RP EZ Negative Control | A04JAN16A | 17 | 0 | 17 | 0 | 100% | | | TOTAL | 89 | 0 | 88 | 0 | 100% | *The Invalid sample was re-tested according to BioFire instructions and was not included in the Percent Correct analysis.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...