AmpliVue Trichomonas Assay

K143329 · Quidel Corporation · OUY · Mar 17, 2015 · Immunology

Device Facts

Record IDK143329
Device NameAmpliVue Trichomonas Assay
ApplicantQuidel Corporation
Product CodeOUY · Immunology
Decision DateMar 17, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3860
Device ClassClass 2

Intended Use

The AmpliVue® Trichomonas Assay is an in vitro diagnostic test, uses isothermal amplification technology (helicase-dependent amplification, HDA) for the qualitative detection of Trichomonas vaginalis nucleic acids isolated from clinician-collected vaginal swab specimens obtained from symptomatic or asymptomatic females to aid in the diagnosis of trichomoniasis.

Device Story

The AmpliVue Trichomonas Assay is a manual, in vitro diagnostic test for the qualitative detection of T. vaginalis DNA. It utilizes helicase-dependent amplification (HDA), an isothermal amplification technology, to target a conserved multi-copy sequence of the T. vaginalis genome. The process involves eluting a clinician-collected vaginal swab in a lysis tube, followed by heat-based cell lysis. An aliquot of the lysate is transferred to a dilution tube and then to a reaction tube containing lyophilized HDA reagents. The reaction is asymmetric, producing single-stranded amplicons that hybridize with sequence-specific capture probes and a biotinylated detection probe. The resulting hybrid is detected via a lateral flow strip within a self-contained, disposable cassette. The cassette uses a vertical-flow DNA detection strip coated with anti-hapten antibodies to display test (T) and control (C) lines, which are visualized by streptavidin-conjugated color particles. The device is intended for use by healthcare professionals in clinical settings to aid in the diagnosis of trichomoniasis. Results are interpreted visually by the operator. The assay provides a rapid, point-of-care-like diagnostic capability without the need for specialized instrumentation.

Clinical Evidence

Multi-center study (n=992) comparing AmpliVue to composite reference method (Wet Mount and InPouch TV culture). Results: 100% sensitivity (95% CI: 96.9-100%) and 98.2% specificity (95% CI: 97.0-98.9%) across combined symptomatic and asymptomatic populations. 8/16 false positives were positive by an alternative FDA-cleared molecular device.

Technological Characteristics

Isothermal helicase-dependent amplification (HDA) technology. Materials include lysis/dilution buffers, lyophilized HDA reagents, and a lateral flow cassette with a vertical-flow DNA detection strip. Detection uses streptavidin-conjugated color particles. Requires external heat blocks (95°C and 64°C). Standalone, manual, non-instrumented detection. Qualitative result via visual line interpretation.

Indications for Use

Indicated for the qualitative detection of Trichomonas vaginalis nucleic acids in clinician-collected vaginal swab specimens from symptomatic or asymptomatic females to aid in the diagnosis of trichomoniasis.

Regulatory Classification

Identification

A Trichomonas vaginalis nucleic acid assay is a device that consists of primers, probes, enzymes, and controls for the amplification and detection of trichomonas nucleic acids in endocervical swabs, vaginal swabs, and female urine specimens, from women symptomatic for vaginitis, cervicitis, or urethritis and/or to aid in the diagnosis of trichomoniasis in asymptomatic women. The detection of trichomonas nucleic acids, in conjunction with other laboratory tests, aids in the clinical laboratory diagnosis of trichomoniasis caused by Trichomonas vaginalis.

Special Controls

*Classification.* Class II (special controls). The special controls are set forth in FDA's guideline document entitled: “Class II Special Controls Guideline: Nucleic Acid Amplification Assays for the Detection of*Trichomonas vaginalis;* Guideline for Industry and Food and Drug Administration Staff.” See § 866.1(e) for information on obtaining this document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized graphic of three human profiles facing to the right. The profiles are connected and form a single, flowing shape. The graphic is surrounded by a circular border of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Quidel Corporation Ronald H. Lollar Senior Director, Clinical and Regulatory Affairs 2005 East State Street, Suite 100 Athens OH 45701 March 17, 2015 Re: K143329 Trade/Device Name: Amplivue® Trichomonas Assay Regulation Number: 21 CFR 866.3860 Regulation Name: Trichomonas vaginalis nucleic acid amplification test system Regulatory Class: II Product Code: OUY Dated: February 18, 2015 Received: February 19, 2015 Dear Mr. Lollar: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Sally A. Hojvat -S Sally Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K143329 Device Name AmpliVue® Trichomonas Assay #### Indications for Use (Describe) The AmpliVue® Trichomonas Assay is an in vitro diagnostic test, uses isothermal amplification technology (helicasedependent amplification, HDA) for the qualitative detection of Trichomonas vaginalis nucleic acids isolated from clinician-collected vaginal swab specimens obtained from symptomatic females to aid in the diagnosis of trichomoniasis. Type of Use (Select one or both, as applicable) | <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | |----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------| |----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------| #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### Applicant: Quidel Corporation 12544 High Bluff Drive, Suite 200 San Diego, California 92130 Telephone: 858-552-7910 Fax: 858-646-8045 ### Contact Information: Ronald H. Lollar, Senior Director Clinical and Regulatory Affairs 2005 East State Street, Suite 100 Athens, Ohio 45701 740-589-3300 - Corporate number 740-589-3373 – Desk phone 858-552-6451—Fax Ron.Lollar@quidel.com ### Date of preparation of 510(k) summary: November 17, 2014 ### A. 510(k) Number: K143329 ### B. Purpose for Submission: To obtain substantial equivalence for the AmpliVue Trichomonas Assay ### C. Measurand: Repeated DNA fragment located in T. vaginalis genome ### D. Type of Test: Helicase-dependent amplification (HDA) {4}------------------------------------------------ ### E. Applicant: Quidel Corporation ### F. Proprietary and Established Names: AmpliVue Trichomonas Assay ### G. Regulatory Information: | Product Code | Classification | Regulation Section | Panel | |------------------------------------------------------------------------------------|----------------|--------------------|----------------------| | OUY –<br><i>Trichomonas vaginalis</i> nucleic<br>acid amplification<br>test system | Class II | 21 CFR 866.3860 | Microbiology<br>(83) | ### H. Intended Use: ### 1. Intended Use(s): The AmpliVue® Trichomonas Assay is an in vitro diagnostic test, uses isothermal amplification technology (helicase-dependent amplification, HDA) for the qualitative detection of Trichomonas vaginalis nucleic acids isolated from clinician-collected vaginal swab specimens obtained from symptomatic or asymptomatic females to aid in the diagnosis of trichomoniasis. ### 2. Indication(s) for Use: Same as Intended Use. ### 3. Special conditions for use statement(s): - . For in vitro diagnostic use only - For prescription use only - 4. Special instrument requirements: None {5}------------------------------------------------ #### l. Device Description: The AmpliVue® Trichomonas Assay combines simple processing, an isothermal amplification technology named Helicase-Dependent Amplification (HDA), and a selfcontained disposable amplicon detection device for the detection of T. vaginalis in clinician-collected vaginal swabs from symptomatic and asymptomatic women. The assay targets a conserved multi-copy sequence of the T. vaginalis DNA. The vaginal swab is eluted in a lysis tube, and the cells are lysed by simple heat treatment. After heat treatment, an aliquot of the lysed specimen is transferred into a dilution tube. An aliquot of the diluted sample is added to a reaction tube containing a lyophilized mix of HDA reagents including primers specific for the amplification of a conserved DNA sequence only found in T. vaginalis. The assay also includes an internal control to confirm the integrity of the assay reagents and cassette detection as well as to control for (or determine whether) HDA-inhibitors that may be present within the clinical specimens. The HDA reaction is asymmetric so that an excess of single stranded DNA (amplicon) is formed. The sequence specific capture probes as well as a biotinylated detection probe shared by both target and internal control bind to the corresponding single-stranded amplicons, forming dual labeled probe-amplicon hybrid. After completion of the HDA reaction, the reaction tube is transferred to a cassette for rapid detection with the test result displayed as test and/or control lines in the window of the cassette. The dual-labeled probe-amplicon hybrid is then detected by the lateral flow strip within the cassette. The bottom line captures the test amplicon and the top line captures the control amplicon. The biotin label binds the streptavidin-conjugated color particles for visualization and the test result is shown as colored lines visible to the naked eye. The cassette is comprised of two individual components: an amplicon cartridge that holds the running buffer and a single 0.2 mL thin wall reaction tube containing the amplified product; and the detection chamber which houses the amplicon cartridge and a verticalflow DNA detection strip embedded into the cassette. The DNA detection strip is coated with different anti-hapten antibodies that serve as the T. vaginalis test (T) line and the control (C) line in the assay. A razor blade and a plastic pin located at the bottom of the detection chamber opens the HDA reaction tube and the running buffer bulb when the handle of the cassette is closed. The mixture flows through a fiberglass paper connected to the DNA detection strip that contains a fiberglass pad pre-loaded with streptavidinconjugated color particles for color visualization. Detection of T. vaginalis DNA is reported whenever the T2 (Test line 2) is visible through the detection window of the cassette. The presence of the C line is not required for positive results. No detection of T. vaginalis DNA is reported when only the C line is displayed. The assay is regarded as invalid when neither line is displayed. {6}------------------------------------------------ ### Materials Provided: AmpliVue® Trichomonas Assay Kit: M211 16 Tests per kit | Component | Quantity | Storage | |----------------------------------------------|---------------|------------| | Detection Cassettes | 16/kit | 2° to 30°C | | Lysis Buffer Tubes (1.0 mL solution/tube) | 16 tubes/kit | 2° to 8°C | | Dilution Buffer Tubes (1.5 mL solution/tube) | 16 tubes/kit, | 2° to 8°C | | Reaction Tubes | 16 tubes/kit | 2° to 8°C | | Amplicon Cartridge | 16/kit | 2° to 30°C | ### Materials required but not provided: - . External controls for Trichomonas vaginalis (e.g. Quidel Molecular Trichomonas Assay Control Kit: M119, which contains positive and negative controls. This positive control contains intact non-viable, trophozoites and has been titered to be near the limit of detection for the assay. This negative control is the same matrix as the positive control, but is trophozoite-free. These controls serve as an external processing and extraction control - Sterile DNAse-free filter-blocked or positive displacement micropipettor tips ● - Micropipettor - Stopwatch or timer - Scissors or a blade - Micro tube tray - Heat block capable of 95° C ± 2° C temperature ● - Heat block with heated lid capable of 64° ± 1° C temperature ● - Thermometer ● #### J. Substantial Equivalence Information: - 1. Predicate device name(s): APTIMA Trichomonas vaginalis Assay (PANTHER® System) - 2. Predicate 510(k) number(s): K122062 {7}------------------------------------------------ ## 3. Comparison with predicate: | Similarities | | | |------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Item | AmpliVue® Trichomonas Assay | APTIMA Trichomonas<br>vaginalis Assay (PANTHER®<br>System) (k122062) | | Intended Use | The AmpliVue® Trichomonas Assay<br>is an in vitro diagnostic test, uses<br>isothermal amplification<br>technology (helicase-dependent<br>amplification, HDA) for the<br>qualitative detection of<br><i>Trichomonas vaginalis</i> nucleic acids<br>isolated from clinician-collected<br>vaginal swab specimens obtained<br>from symptomatic or<br>asymptomatic females to aid in the<br>diagnosis of trichomoniasis. | The APTIMA Trichomonas<br>vaginalis Assay is an in vitro<br>qualitative nucleic acid<br>amplification test (NAAT) for<br>the detection of ribosomal<br>RNA (rRNA) from<br><i>Trichomonas vaginalis</i> to aid<br>in the diagnosis of<br>trichomoniasis using the<br>PANTHER System.<br>The assay may be used to<br>test the following specimens<br>from symptomatic or<br>asymptomatic women:<br>clinician-collected<br>endocervical swabs,<br>clinician-collected vaginal<br>swabs, and specimens<br>collected in PreservCyt<br>Solution. | | Differences | | | | Item | AmpliVue® Trichomonas Assay | APTIMA Trichomonas<br>vaginalis Assay (k122062) | | Sample Types | Clinician-collected Vaginal<br>Swabs | Clinician-collected Vaginal<br>Swabs, Endocervical Swabs,<br>ThinPrep in PreservCyt<br>solution | | Target Sequence<br>Detected | Repeated DNA fragment located<br>in <i>T. vaginalis</i> genome | <i>T. vaginalis</i> ribosomal RNA<br>(rRNA) | | Amplification<br>Technology | Helicase-dependent<br>amplification (HDA) | Transcription Mediated<br>Amplification (TMA)<br>Hybridization Protection<br>Assay (HPA) | | Self-Contained<br>System Assay after<br>sample preparation | No | Yes | | Detection Technique | Manual | Automated | | Instrument | None | <i>Panther</i> | | Performance<br>Characteristics | Clinician-collected Vaginal<br>Specimens: Asymptomatic<br>Sensitivity 100% (95% CI 94.1 – 100)<br>Specificity 98.3% (95% CI 96.9 – 99.1)<br>Symptomatic<br>Sensitivity 100% (95% CI 93.9 – 100)<br>Specificity 97.9% (95% CI 95.5 – 99.0) | Clinician-collected Vaginal<br>Specimens:<br>Asymptomatic<br>Sensitivity 100% (95% CI<br>75.8 – 100)<br>Specificity 97.3% (95% CI<br>94.6 – 98.7)<br>Symptomatic<br>Sensitivity 100% (95% CI<br>93.7 – 100)<br>Specificity 98.8% (95% CI<br>97.0 – 99.5) | {8}------------------------------------------------ ## AmpliVue Trichomonas Assay 11/17/2014 Page 6 of 18 # 510(k) Summary {9}------------------------------------------------ ### K. Test Principle: The AmpliVue® Trichomonas Assay combines simple processing, an isothermal amplification technology named Helicase-Dependent Amplification (HDA), and a selfcontained disposable amplicon detection device for the detection of T. vaginalis in clinician-collected vaginal swabs from symptomatic and asymptomatic women. The assay targets a conserved multi-copy sequence of the T. vaginalis DNA. The vaginal swab is eluted in a lysis tube, and the cells are lysed by simple heat treatment. After heat treatment, an aliquot of the lysed specimen is transferred into a dilution tube. An aliquot of the diluted sample is added to a reaction tube containing a lyophilized mix of HDA reagents including primers specific for the amplification of a conserved DNA sequence only found in T. vaginalis. The assay also includes an internal control to confirm the integrity of the assay reagents and cassette detection as well as to control for (or determine whether) HDA-inhibitors that may be present within the clinical specimens. The HDA reaction is asymmetric so that an excess of single stranded DNA (amplicon) is formed. The sequence specific capture probes as well as a biotinylated detection probe shared by both target and internal control bind to the corresponding single-stranded amplicons, forming dual labeled probe-amplicon hybrid. After completion of the HDA reaction, the reaction tube is transferred to a cassette for rapid detection with the test result displayed as test and/or control lines in the window of the cassette. The dual-labeled probe-amplicon hybrid is then detected by the lateral flow strip within the cassette. The bottom line captures the test amplicon and the top line captures the control amplicon. The biotin label binds the streptavidin-conjugated color particles for visualization and the test result is shown as colored lines visible to the naked eye. The cassette is comprised of two individual components: an amplicon cartridge that holds the running buffer and a single 0.2 mL thin wall reaction tube containing the amplified product; and the detection chamber which houses the amplicon cartridge and a verticalflow DNA detection strip embedded into the cassette. The DNA detection strip is coated with different anti-hapten antibodies that serve as the T. vaginalis test (T) line and the control (C) line in the assay. A razor blade and a plastic pin located at the bottom of the detection chamber opens the HDA reaction tube and the running buffer bulb when the handle of the cassette is closed. The mixture flows through a fiberglass paper connected to the DNA detection strip that contains a fiberglass pad pre-loaded with streptavidinconjugated color particles for color visualization. Detection of T. vaginalis DNA is reported whenever the T2 (Test line 2) is visible through the detection window of the cassette. The presence of the C line is not required for positive results. No detection of T. vaginalis DNA is reported when only the C line is displayed. The assay is regarded as invalid when neither line is displayed. {10}------------------------------------------------ ### L. Performance Characteristics: - 1. Analytical performance: - a. Precision/Reproducibility: ### Reproducibility In order to confirm the reproducibility of the AmpliVue Trichomonas Assay a blinded and randomized study four-member panel containing Trichomonas vaginalis positive samples (3× LoD, 1/9× LoD) and a negative sample were tested at three (3) test sites (one in-house laboratory and two (2) clinical sites). Each site tested a reproducibility panel and Assay Controls for five (5) days in triplicate. Testing was done by two operators at each site. Each operator ran the panel once a day using one lot of AmpliVue Trichomonas Assay. The AmpliVue Trichomonas Assay generated reproducible results in this study. | | SITE | | | | | | | | | | | | |------------------------------------------------------|-----------------------------------|----------------|----------------------|-----------------------------------|----------------|-----------------------------------|----------------|--------|-------------------|-----------|----------|-------------------| | Concentration | Site #1 | | | Site #2 | | Site #3 | | | Overall Percent | | 95% CI | | | | #expected<br>results /#<br>tested | %<br>Agreement | 95% CI | #expected<br>results /#<br>tested | %<br>Agreement | #expected<br>results/ #<br>tested | %<br>Agreement | 95% CI | Agreement | | | | | High Negative*<br>(34<br>trophozoites<br>/mL) | 9/30 | 30% | 16.7%<br>to<br>47.9% | 8/30 | 27% | 14.2% to<br>47.9% | 19/30 | 63% | 45.5% to<br>78.1% | 36/<br>90 | 40% | 30.5% to<br>50.3% | | Low Positive<br>(307<br>trophozoites<br>/mL) | 30/30 | 100% | 88.6%<br>to<br>100% | 30/30 | 100% | 88.6% to<br>100% | 30/30 | 100% | 88.6% to<br>100% | 90/<br>90 | 100<br>% | 95.9% to<br>100% | | Moderate<br>Positive<br>(921<br>trophozoites<br>/mL) | 30/30 | 100% | 88.6%<br>to<br>100% | 30/30 | 100% | 88.6% to<br>100% | 30/30 | 100% | 88.6% to<br>100% | 90/<br>90 | 100<br>% | 95.9% to<br>100% | | Negative | 30/30 | 100% | 88.6%<br>to | 30/30 | 100% | 88.6% to<br>100% | 30/30 | 100% | 88.6% to<br>100% | 90/<br>90 | 100<br>% | 95.9% to<br>100% | {11}------------------------------------------------ ### AmpliVue Trichomonas Assay 11/17/2014 Page 9 of 18 ### 510(k) Summary | | SITE | | | | | | | | | | | | |---------------------|-----------------------------------|----------------|---------------------|-----------------------------------|----------------|------------------|-----------------------------------|----------------|------------------|-----------------|----------|------------------| | Concentration | Site #1 | | | Site #2 | | | Site #3 | | | Overall Percent | | 95% CI | | | #expected<br>results /#<br>tested | %<br>Agreement | 95% CI | #expected<br>results /#<br>tested | %<br>Agreement | 95% CI | #expected<br>results/ #<br>tested | %<br>Agreement | 95% CI | Agreement | | | | | | | 100% | | | | | | | | | | | Positive<br>Control | 30/30 | 100% | 88.6%<br>to<br>100% | 30/30 | 100% | 88.6% to<br>100% | 30/30 | 100% | 88.6% to<br>100% | 90/<br>90 | 100<br>% | 95.9% to<br>100% | | Negative<br>Control | 30/30 | 100% | 88.6%<br>to<br>100% | 30/30 | 100% | 88.6% to<br>100% | 30/30 | 100% | 88.6% to<br>100% | 90/<br>90 | 100<br>% | 95.9% to<br>100% | * The High Negative sample has a detection target range of 20 to 80%. Data generated by all sites fall into this range. > The results suggest that there are no significant differences between different users and different sites on different days. Reproducibility studies are acceptable. - b. Linearity/assay reportable range: Not applicable – This assay is qualitative. - c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability: Not applicable. This assay is qualitative. Specimen Stability: Not applicable Controls: Controls (Quidel Molecular Trichomonas Control Set #M119, which contains positive and negative controls, serves as an external processing and extraction control) were run on the AmpliVue Trichomonas Assay each day of testing. These controls are described as follows: {12}------------------------------------------------ - a. The internal control is used to detect HDA inhibitory specimens and to confirm the integrity of assay reagents and cassette detection. The internal control is included in the reaction tube. - b. External assay positive control serves as the assay positive control. The positive control listed in "Materials Required But Not Provided" contains intact non-viable, trophozoites and has been titered to be near the limit of detection for the assay. Transfer 50 µL of positive control into a labeled dilution buffer tube and proceed with processing as described above in Step 1 of Amplification. The external assay positive control is intended to monitor substantial reagent and cassette failure - ﻥ External assay negative control serves as the assay negative control. The negative control listed in "Materials Required But Not Provided" is the same matrix as the positive control, but is trophozoite-free. Transfer 50 µL of negative control into a labeled dilution buffer tube and proceed with processing as described above in Step 1 of Amplification. The external assay negative control is intended to detect reagent or environment contamination or carry-over by either T. vaginalis DNA or amplicons. - d. Detection limit: The analytical sensitivity (limit of detection or LoD) of the AmpliVue * Trichomonas Assay was determined using quantified (trophozoite/mL) stocks of two (2) T. vaqinalis stocks, one metronidazole-susceptible G3 and one metronidazole-resistant CDC888 serially diluted in negative matrix. The LoD is defined as the lowest concentration at which 95% of all replicates tested positive. The strains were freshly grown and quantified using a hemocytometer. The cells were serially diluted in Liguid Stuart medium to five (5) concentrations in the preliminary LoD determination study. Each test dilution was run as 10 replicates in the presence of negative matrix on three reagent lots. LoD was confirmed by testing each reference strain with 20 replicates on three reagent lots in the negative vaginal swab matrix. The assay LoD for Trichomonas vaginalis strain G3 is 307 trophozoites/mL and for strain CDC888 the LoD is 921 trophozoites/mL. {13}------------------------------------------------ - e. Analytical specificity: ### Cross Reactivity: A study was performed to evaluate the cross-reactivity of the AmpliVue Trichomonas Assay with forty-five (45) microorganisms (36 bacteria, 4 yeasts, 4 viruses, 1 parasite) potentially found in specimens collected to test for Trichomonas vaginalis infection. Cross-reactive microorganisms were tested at clinically relevant levels of viruses (≥10 pfu/mL or genomic copies/mL) and bacteria, yeast and parasite (≥10 °cfu/mL or genomic copies/mL) in the device. All organisms were diluted in Liquid Stuart medium and tested in negative matrix in triplicate in the AmpliVue Trichomonas assay. The organisms included in the cross-reactivity study and their tested concentrations are shown in the table below. | Microorganism | Concentration<br>Tested | |------------------------------|----------------------------| | Acinetobacter Iwoffi | 4.55×106 CFU/mL | | Actinomyces israelii | 6.63×106 CFU/mL | | Atopobium vaginae | 3.60×106 CFU/mL | | Bacteroides fragilis | 4.2×106 CFU/mL | | Bifidobacterium adolescentis | 1.00×106 CFU/mL | | Campylobacter jejuni | 1.72×106 CFU/mL | | Candida albicans | 2.00×106 CFU/mL | | Candida glabrata | 7.87×106 CFU/mL | | Candida parapsilosis | 2.87×106 CFU/mL | | Candida tropicalis | 2.15×106 CFU/mL | | Chlamydia trachomatis | 7.83×106 CFU/mL | | Clostridium difficile | 6.77×106 CFU/mL | | Clostridium perfringens | 1.06×106 CFU/mL | | Corynebacterium genitalium | 3.61×106 CFU/mL | | Cryptococcus neoformans | 1.92×106 CFU/mL | | Enterobacter aerogenes | 1.18×106 CFU/mL | | Enterococcus faecalis | 2.20×106 CFU/mL | | Escherichia coli | 1.13×106 CFU/mL | | Fusobacterium nucleatum | 8.05×106 CFU/mL | | Gardnerella vaginalis | 1.20×106 CFU/mL | | Haemophilus ducreyi | 2.97×106 genomic copies/mL | | HIV-1 Subtype B RNA | 1.14×106 genomic copies/mL | | Herpes simplex virus I | 7.96×106 TCID50/mL | | Herpes simplex virus II | 2.27×105 TCID50/mL | {14}------------------------------------------------ AmpliVue Trichomonas Assay 11/17/2014 Page 12 of 18 #### 510(k) Summary | Microorganism | Concentration<br>Tested | |------------------------------------|-----------------------------------------| | HPV 16 (SiHa) | 4.3×106 genomic copies/mL | | Klebsiella oxytoca | 1.63×106 CFU/mL | | Lactobacillus acidophilus | 2.00×106 CFU/mL | | Lactobacillus jensenii | 4.06×106 CFU/mL | | Lactobacillus vaginalis | 1.11×106 CFU/mL | | Listeria monocytogenes | 6.13×106 CFU/mL | | Mobiluncus curtisii | 3.2×106 CFU/mL | | Mycoplasma hominis | 1.30×106 CFU/mL | | Neisseria gonorrhoeae | 3.20×106 CFU/mL | | Pentatrichomonas hominis | 4.5×106 CFU/mL | | Peptostreptococcus anaerobius | 8.1×106 genomic copies/mL | | Prevotella bivia | 3.01×106 CFU/mL | | Propionibacterium acnes | 6.63×106 CFU/mL | | Proteus mirabilis | 1.19×106 CFU/mL | | Pseudomonas aeruginosa | 1.32×106 CFU/mL | | Staphylococcus aureus MRSA | 7.52×106 CFU/mL | | Staphylococcus epidermidis MRSE | 1.75×106 CFU/mL | | Streptococcus pyogenes | 6.38×106 CFU/mL | | Streptococcus agalactiae | 2.20×106 CFU/mL | | Trichomonas tenax | 6.3×106 CFU/mL | | Ureaplasma urealyticum | 1.23×106 genomic copies/ml | | Microorganism | Concentration<br>Tested | | Acinetobacter Iwoffi | 4.55×106 CFU/mL | | Actinomyces israelii | 6.63×106 CFU/mL | | Atopobium vaginae | 3.60×106 CFU/mL | | Bacteroides fragilis | 4.2×106 CFU/mL | | Bifidobacterium adolescentis | 1.00×106 CFU/mL | | Campylobacter jejuni | 1.72×106 CFU/mL | | Candida albicans | 2.00×106 CFU/mL | | Candida glabrata | 7.87×106 CFU/mL | | Candida parapsilosis | 2.87×106 CFU/mL | | Candida tropicalis | 2.15×106 CFU/mL | | Chlamydia trachomatis | 7.83×106 CFU/mL | | Clostridium difficile | 6.77×106 CFU/mL | | Clostridium perfringens | 1.06×106 CFU/mL | | Corynebacterium genitalium | 3.61×106 CFU/mL | | Cryptococcus neoformans | 1.92×106 CFU/mL | | Enterobacter aerogenes | 1.18×106 CFU/mL | | Enterococcus faecalis | 2.20×106 CFU/mL | | Escherichia coli | 1.13×106 CFU/mL | | Fusobacterium nucleatum | 8.05×106 CFU/mL | | Gardnerella vaginalis | 1.20×106 CFU/mL | | Haemophilus ducreyi | 2.97×106 genomic<br>copies/mL | | HIV-1 Subtype B RNA | 1.14×106 genomic<br>copies/mL | | Herpes simplex virus I | 7.96×106 TCID50/mL | | Herpes simplex virus II | 2.27×105 TCID50/mL | | HPV 16 (SiHa) | 4.3×106 genomic<br>copies/mL | | Klebsiella oxytoca | 1.63×106 CFU/mL | | Lactobacillus acidophilus | 2.00×106 CFU/mL | | Lactobacillus jensenii | 4.06×106 CFU/mL | | Lactobacillus vaginalis | 1.11×106 CFU/mL | | Listeria monocytogenes | 6.13×106 CFU/mL | | Mobiluncus curtisii | 3.2×106 CFU/mL | | Mycoplasma hominis | 1.30×106 CFU/mL | | Neisseria gonorrhoeae | 3.20×106 CFU/mL | | Pentatrichomonas hominis | 4.5×106 CFU/mL | | Peptostreptococcus anaerobius | 8.1×106 genomic<br>copies/mL | | Microorganism | Concentration<br>Tested | | Propionibacterium acnes | $6.63\times10^{6}$ CFU/mL | | Proteus mirabilis | $1.19\times10^{6}$ CFU/mL | | Pseudomonas aeruginosa | $1.32\times10^{6}$ CFU/mL | | Staphylococcus aureus MRSA | $7.52\times10^{6}$ CFU/mL | | Staphylococcus epidermidis<br>MRSE | $1.75\times10^{6}$ CFU/mL | | Streptococcus pyogenes | $6.38\times10^{6}$ CFU/mL | | Streptococcus agalactiae | $2.20\times10^{6}$ CFU/mL | | Trichomonas tenax | $6.3\times10^{6}$ CFU/mL | | Ureaplasma urealyticum | $1.23\times10^{6}$ genomic<br>copies/ml | No cross-reactivity was seen with the AmpliVue Trichomonas Assay with any of forty-five (45) microorganisms (36 bacteria, 4 yeasts, 4 viruses, 1 parasite) tested. ### Interference: Microorganism A study was conducted to determine if the AmpliVue® Trichomonas assay is inhibited in the presence of forty-five (45) microorganisms (36 bacteria, 4 yeasts, 4 viruses, 1 parasite) potentially found in specimens collected to test for Trichomonas vaginalis infection. Potentially interfering microorganisms were tested at clinically relevant levels of viruses (≥10 pfu/mL or genomic copies/mL) and bacteria, yeast and parasite (≥10°cfu/mL or genomic copies/mL) in the device. All organisms were diluted in Liquid Stuart medium and tested in negative matrix in triplicate in the AmpliVue Trichomonas assay. The organisms included in the interference study and their tested concentrations are shown in the table below. {15}------------------------------------------------ AmpliVue Trichomonas Assay 11/17/2014 Page 13 of 18 # 510(k) Summary {16}------------------------------------------------ No interference was seen with the AmpliVue "Trichomonas Assay with any of forty-five (45) microorganisms (36 bacteria, 4 yeasts, 4 viruses, 1 parasite) tested. ### Interference: Substances A study was conducted to determine if the AmpliVue "Trichomonas assay is inhibited in the presence of a panel of thirteen (13) substances potentially present in specimens collected to test for Trichomonas vaginalis infection. Each of the potential interfering substances was tested in three replicates in the presence and absence of near LoD (2×) levels of two strains of Trichomonas vaginalis in the AmpliVue "Trichomonas Assay. Substances were introduced into the assay at concentrations which were medically relevant. | Substances | Final Conc. | |-------------------------------------------------------------------------|-------------| | K-Y Personal Lubricant Jelly | 1% | | Ortho Options Gynol II Extra<br>Strength Vaginal Contraceptive<br>Jelly | 1% | | Summer's Eve Ultra Extra Strength<br>Feminine Deodorant Spray | 1% | | Vagisil Creme Maximum Strength | 1% | | Estradiol | 1% | | Mucin from Porcine Stomach | 1% | | Glacial acetic acid | 1% | | CVS Vinegar & Water Extra<br>Cleansing Disposable Douche | 1% | | Seminal fluid | 1% | {17}------------------------------------------------ | Substances | Final Conc. | |--------------------------------|---------------------------------------------------------------------------------------------------------| | Whole blood with EDTA | 10% | | Summer's Eve Douche, Medicated | 1% | | Acyclovir (Acycloguanosine) | 5% (w/v), active<br>concentration in Zovirax<br>cream<br>1% of active ingredient of<br>Zovirax cream | | Metronidazole | 0.75% (w/v), active<br>concentration in<br>Vandazole gel<br>1% of active ingredient of<br>Vandazole gel | There was no evidence of interference caused by the substances tested. ### Analytical Reactivity (Inclusivity): A study was performed to verify the in silico inclusivity results with functional testing of the AmpliVue Trichomonas Assay using twenty (20) additional strains of Trichomonas vaginalis tested in triplicate at concentrations near LoD. | Bacterial Strain | Strain Detected<br>(Yes/No) | |------------------|-----------------------------| | CDC899 | Yes | | CDC938 | Yes | | CDC963 | Yes | | CDC1031 | Yes | | CDC1256 | Yes | | PMGH25 | Yes | | BUSH20 | Yes | | CDC911 | Yes | | MOR31 | Yes | | CDC1080 | Yes | | B7708/1839 | Yes | | F1623 | Yes | | CDC1095 | Yes | | SD1 | Yes | | SA-384 | Yes | | CDC948 | Yes | | SD10 | Yes | | SA-A53 | Yes | {18}------------------------------------------------ | Bacterial Strain | Strain Detected<br>(Yes/No) | |------------------|-----------------------------| | CDC1230 | Yes | | SA-A19 | Yes | #### f. Assay cut-off: Not applicable. ### 2. Clinical studies: ### a. Clinical Sensitivity: A multi-center study was performed to evaluate the AmpliVue Trichomonas Assay using nine hundred ninety-two (992) clinician-collected vaginal swab specimens obtained from symptomatic (n=342) or asymptomatic (n=650) patients. The clinician categorized the patients as symptomatic or asymptomatic at the of specimen collection. The study was performed April to November 2014 at four locations in the United States and one location in Canada. Specimens were obtained from each subject after informed consent was obtained. For each subject, three (3) vaginal specimens were collected using polyester or rayon Swabs w/ Liquid Stuart's, and one (1) vaginal specimen collected with a collection swab from a FDA-cleared molecular device. The four (4) clinician collected vaginal swabs were used for reference and AmpliVue testing. The first two polyester/rayon swabs were randomized, one swab was tested for the Wet Mount (reference method) and the other swab was used for the InPouch TV Culture (reference method). The third swab was used for testing the AmpliVue Trichomonas Assay. The FDA-cleared molecular device collection swab was used for discordant testing. All sensitivity and specificity calculations were based on a composite reference method of Wet Mount and InPouch TV culture. As specimen was considered positive if either test was positive. One (1) specimen was removed from the study due to a delay in the culture inoculation. Eight (8) specimens yielded invalid results upon initial testing with the AmpliVue "Trichomonas Assay (0.8%). These specimens were re-tested according to the instruction provided in the Package Insert. Six (6) of the specimens yielded valid results when re-tested (5 negative and 1 positive result). Two (2) specimens yielded {19}------------------------------------------------ a second invalid result (0.2%). The table below shows the sensitivity, specificity, PPV, and NPV of the AmpliVue Trichomonas Assay and the prevalence of T. vaginalis (by asymptomatic, symptomatic clinician designations and combined). | Performance Characteristics of the AmpliVue® Trichomonas Assay by Symptom Status compared to the Composite | | | | | | | | | | | | |------------------------------------------------------------------------------------------------------------|----------------|-----|-----|-----|-----|----|-------|--------------------------|--------------------------|------------------------|----------------------| | Reference Method | | | | | | | | | | | | | Site<br>Number | Symptom Status | N | TP | FP | TN | FN | Prev% | Sensitivity%<br>(95% CI) | Specificity%<br>(95% CI) | PPV %<br>(95% CI) | NPV %<br>(95% CI) | | Combined | Asymptomatic | 647 | 61 | 10 | 576 | 0 | 9.4 | 100<br>(94.1 to 100) | 98.3<br>(96.9 to 99.1) | 85.9<br>(76.0 to 92.2) | 100<br>(99.3 to 100) | | | Symptomatic | 342 | 59 | 6 | 277 | 0 | 17.3 | 100<br>(93.9 to 100) | 97.9<br>(95.5 to 99.0) | 90.8<br>(81.3 to 95.7) | 100<br>(98.6 to 100) | | | All | 989 | 120 | 16* | 853 | 0 | 12.1 | 100<br>(96.9 to 100) | 98.2<br>(97.0 to 98.9) | 88.2<br>(81.7 to 92.6) | 100<br>(99.6 to 100) | | Site 1 | Asymptomatic | 133 | 26 | 3 | 104 | 0 | 19.5 | 100<br>(87.1 to 100) | 97.2<br>(92.1 to 99.0) | 89.7<br>(73.6 to 96.4) | 100<br>(96.4 to 100) | | | Symptomatic | 163 | 27 | 2 | 134 | 0 | 16.6 | 100<br>(87.5 to 100) | 98.5<br>(94.8 to 99.6) | 93.1<br>(78.0 to 98.1) | 100<br>(97.2 to 100) | | | All | 296 | 53 | 5 | 238 | 0 | 17.9 | 100<br>(93.2 to 100) | 97.9<br>(95.3 to 99.1) | 91.4<br>(81.4 to 96.3) | 100<br>(98.4 to 100) | | Site 2 | Asymptomatic | 46 | 5 | 1 | 40 | 0 | 10.9 | 100<br>(56.6 to 100) | 97.6<br>(87.4 to 99.6) | 83.3<br>(43.6 to 97.0) | 100<br>(91.2 to 100) | | | Symptomatic | 69 | 17 | 1 | 51 | 0 | 24.6 | 100<br>(81.6 to 100) | 98.1<br>(89.9 to 99.7) | 94.4<br>(74.2 to 99.0) | 100<br>(93.0 to 100) | | | All | 115 | 22 | 2 | 91 | 0 | 19.1 | 100<br>(85.1 to 100) | 97.8<br>(92.5 to 99.4) | 91.7<br>(74.2 to 97.7) | 100<br>(95.9 to 100) | | Site 3 | Asymptomatic | 206 | 20 | 3 | 183 | 0 | 9.7 | 100<br>(83.9 to 100) | 98.4<br>(95.4 to 99.4) | 87.0<br>(67.9 to 95.5) | 100<br>(97.9 to 100) | | | Symptomatic | 41 | 7 | 2 | 32 | 0 | 17.1 | 100<br>(64.6 to 100) | 94.1<br>(80.9 to 98.4) | 77.8<br>(45.3 to 93.7) | 100<br>(89.3 to 100) | | | All | 247 | 27 | 5 | 215 | 0 | 10.9 | 100<br>(87.5 to 100) | 97.7<br>(94.8 to 99.0) | 84.4<br>(68.2 to 93.1) | 100<br>(98.2 to 100) | | Site 4 | Asymptomatic | 260 | 10 | 3 | 247 | 0 | 3.8 | 100<br>(72.2 to 100) | 98.8<br>(96.5 to 99.6) | 76.9<br>(49.7 to 91.8) | 100<br>(98.5 to 100) | | | Symptomatic | 35 | 3 | 1 | 31 | 0 | 8.6 | 100<br>(43.8 to 100) | 96.9<br>(84.3 to 99.4) | 75.0<br>(30.1 to 95.4) | 100<br>(89.0 to 100) | | | All | 295 | 13 | 4 | 278 | 0 | 4.4 | 100<br>(77.2 to 100) | 98.6<br>(96.4 to 99.4) | 76.5<br>(52.7 to 90.4) | 100<br>(98.6 to 100) | | Site 5 | Asymptomatic | 2 | 0 | 0 | 2 | 0 | 0 | N/A | 100<br>(34.2 to 100) | N/A | 100<br>(34.2 to 100) | | | Symptomatic | 34 | 5 | 0 | 29 | 0 | 14.7 | 100<br>(56.6 to 100) | 100<br>(88.3 to 100) | 100<br>(56.6 to 100) | 100<br>(88.3 to 100) | | | All | 37 | 5 | 0 | 31 | 0 | 13.5 | 100<br>(56.6 to 100) | 100<br>(89.0 to 100) | 100<br>(56.6 to 100) | 100<br>(89.0 to 100) | * Eight (8) of sixteen (16) Composite negative/AmpliVue positive specimens were positive by a FDA-cleared Trichomonas vaginalis molecular device. ### b. Clinical specificity: {20}------------------------------------------------ See Section 3a. - 4. Clinical cut-off: Not applicable - 5. Expected values: The prevalence of T. vaginalis (by asymptomatic clinician designations and combined) detected by the AmpliVue Trichomonas Assay in the multi-center study was calculated and is provided in the table below. | Symptom<br>Status | Combined | Site 1 | Site 2 | Site 3 | Site 4 | Site 5 | |-------------------|----------|--------|--------|--------|--------|--------| | Asymptomatic | 11.0% | 21.8% | 12.8% | 11.2% | 5.0% | 0.0% | | Symptomatic | 19.0% | 17.8% | 26.1% | 22.0% | 11.4% | 14.7% | | Combined | 13.7% | 19.6% | 20.7% | 13.0% | 5.8% | 13.5% | Positive and Negative Predictive Values The estimated positive predictive value (PPV) and negative predictive value (NPV) of the AmpliVue Trichomonas Assay across different hypothetical prevalence rates are shown in the table below. These calculations are based on the overall estimated sensitivity and specificity for clinician-collected vaginal swab specimens in the AmpliVue Trichomonas Assay clinical study. | Hypothetical PPV and NPV of the AmpliVue Trichomonas Assay | | | |------------------------------------------------------------|---------|---------| | Prevalence % | PPV (%) | NPV (%) | | 1 | 33.3 | 100 | | 2 | 50.0 | 100 | | 5 | 72.5 | 100 | | 10 | 85.5 | 100 | | 15 | 89.8 | 100 | | 20 | 93.0 | 100 | | 25 | 94.4 | 100 | ### Conclusion: The analytical and clinical study results for the AmpliVue Trichomonas Assay support the determination of substantial equivalence to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...