PRODESSE PROFLU+ ASSAY

K132129 · Gen-Probe Prodesse, Inc. · OCC · Aug 9, 2013 · Microbiology

Device Facts

Record IDK132129
Device NamePRODESSE PROFLU+ ASSAY
ApplicantGen-Probe Prodesse, Inc.
Product CodeOCC · Microbiology
Decision DateAug 9, 2013
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 866.3980
Device ClassClass 2

Intended Use

The Prodesse ProFlu+ Assay is a multiplex real-time PCR in vitro diagnostic test for the qualitative detection and differentiation of Influenza A virus, Influenza B virus, and Respiratory Syncytial Virus (RSV) nucleic acids in nasopharyngeal (NP) swab specimens from individuals with signs and symptoms of respiratory tract infection. The ProFlu+ Assay is intended to aid in the differential diagnosis of Influenza A, Influenza B, and RSV viral infections in humans. Negative results do not preclude influenza or RSV infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions. The ProFlu+ Assay is intended for use by professional laboratory personnel.

Device Story

ProFlu+ Assay is a multiplex RT-PCR in vitro diagnostic test; detects/discriminates Influenza A, Influenza B, and RSV nucleic acids. Input: nasopharyngeal swab specimens in viral transport medium. Process: nucleic acid isolation/purification via MagNA Pure LC or NucliSENS easyMAG; reverse transcription and amplification using Cepheid SmartCycler II. Taqman chemistry utilizes 5'-3' exonuclease activity of Taq polymerase to cleave dual-labeled probes, generating fluorescent signals (FAM, TET, Texas Red, Cy5). Universal Internal Control (UIC) monitors for inhibitors. Output: qualitative detection of viral targets monitored by fluorescent intensity during PCR cycles. Used in clinical laboratories by trained personnel. Results aid differential diagnosis; negative results do not exclude infection; positive results do not rule out co-infections. Benefits: rapid identification of respiratory viral pathogens to guide clinical management.

Clinical Evidence

Retrospective clinical comparison study demonstrated the modified assay with Universal Internal Control meets performance claims of the original device. Analytical studies confirmed limit of detection for H3N2v and H7N9 strains. Stability studies validated performance after 10 freeze-thaw cycles for enzymes.

Technological Characteristics

Multiplex real-time RT-PCR. Analyte: RNA. Instrumentation: bioMérieux NucliSENS easyMAG or Roche MagNA Pure and Cepheid SmartCycler II. Controls: Universal Internal Control (RNA IVT + DNA plasmid) and pooled positive controls. Stability: M-MLV Reverse Transcriptase and RNase Inhibitor II (10 freeze-thaw cycles).

Indications for Use

Indicated for symptomatic patients requiring differential diagnosis of Influenza A, Influenza B, and RSV viral infections via nasopharyngeal swab specimens. Not intended for Influenza C detection.

Regulatory Classification

Identification

A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B; (2) Influenza A subtype H1 and Influenza A subtype H3; (3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B; (4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus; (5) Human Metapneumovirus; (6) Rhinovirus; and (7) Adenovirus.

Special Controls

*Classification.* Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;” (2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and (3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K132129 Page 1 of 3 8/6/2013 ### Attachment D 510(k) SUMMARY #### CONTACT Emily Ziegler Scientist I Gen-Probe Prodesse, Inc. 20925 Crossroads Circle Waukesha, W153186 4UG 0.9 2013 #### NAME OF DEVICE Trade Name: Regulation Number: Product Code: Classification Name: Prodesse ProFluTM+ Assay 21 CFR 866.3980 OCC, 001 Nucleic acid amplification assay for detection and differentiation of Influenza A, Influenza B, and RSV #### PREDICATE DEVICE - K110968, ProFlu™+ Assay . #### INTENDED USE The Prodesse ProFlu™+ Assay is a multiplex Real-Time PCR (RT-PCR) in vitro diagnostic test for the rapid and qualitative detection and discrimination of Influenza B Virus, and Respiratory Syncytial Virus (RSV) nucleic acids isolated and purified from nasopharyngeal (NP) swab specimens obtained from symptomatic patients. This test is intended for use to aid in the differential diagnosis of Influenza A, Influenza B and RSV viral infections in humans and is not intended to detect Influenza C. Negative results do not preclude influenza or RSV virus infection and should not be used as the sole basis for treatment or other management decisions. Conversely, positive results do not rule-out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease. The use of additional laboratory testing and clinical presentation must be considered in order to obtain the final diagnosis of respiratory viral infection. Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation (2006 - 2007 respiratory season). Performance characteristics for Influenza A were confirmed when Influenza A/H1, Influenza A/H3, and Influenza A/2009 H1N1 were the predominant Influenza A viruses in circulation (2008 and 2009). When other Influenza A viruses are emerging, performance characteristics may vary. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. ### PRODUCT DESCRIPTION The ProFlu+ Assay enables detection and discrimination of Influenza A Virus, Influenza B Virus, RSV and universal internal control nucleic acid. Nasopharyngeal swab specimens are collected from patients with signs and symptoms of a respiratory infection using a polyester, rayon or nylon tipped swab and placed into viral transport medium. {1}------------------------------------------------ A Universal Internal Control (UIC) is added to each sample prior to nucleic acid isolation to monitor for inhibitors present in the specimens. The isolation and purification of the nucleic acids is performed using either a MagNA Pure LC Instrument (Roche) and the MagNA Purc Total Nucleic Acid Isolation Kit (Roche) or a NucliSENS® easyMAG™ System (bioMérieux) and the Automated Magnetic Extraction Reagents (bioMérieux). The purified nucleic acids are added to Influenza B/RSV Mix along with enzymes included in the ProFlu+ Assay Kit. The Influenza A/Influenza B/RSV Mix contains oligonucleotide primers and target-specific oligonucleotide probes. The primers are complementary to highly conserved regions of genetic sequences for these respiratory viruses. The probes are dual-labeled with a reporter dye attached to the 5'-end and a quencher dye attached to the 3'-end (see table below). | Analyte | Gene Targeted | Probe<br>Fluorophore | AbsorbancePeak | EmissionPeak | Instrument<br>Channel | |----------------------------------|-------------------------------|-------------------------|----------------|--------------|-----------------------| | Influenza A Virus | Matrix | FAM | 495 nm | 520 nm | FAM | | Respiratory<br>Syncytial Virus A | Polymerase | CAL Fluor<br>Orange 560 | 540 nm | 561 nm | TET | | Respiratory<br>Syncytial Virus B | Polymerase | CAL Fluor<br>Orange 560 | 540 nm | 561 nm | TET | | Influenza B Virus | Non-structural<br>NS1 and NS2 | CAL Fluor Red<br>610 | 595 nm | 615 nm | Texas Red | | Universal Internal<br>Control | NA | Quasar 670 | 647 nm | 667 nm | Cy5 | Reverse transcription of the RNA in the sample into complementary DNA (cDNA) and subsequent amplification of DNA is performed in a Cepheid SmartCycler® II instrument. In this process, the probe anneals specifically to the template followed by primer extension and amplification. The ProFity- Assay is based on Tagman chemistry, which utilizes the 5 - 3 exonuclease activity of the Taq polymerase to cleave the probe thus separating the reporter dye from the quencher. This generates an increase in fluorescent signal upon excitation from a light source. With each cycle, additional reporter dye molecules are cleaved from their respective probes, further increasing fluorescent signal. The amount of fluorescence at any given cycle is dependent on the amount of amplification products present at that time. Fluorescent intensity is monitored during each PCR cycle by the SmartCyclerII instrument. # DEVICE COMPARISON The modified ProFlu+ Assay differs from the current kit in the following ways: - Outsourcing of control stock manufacturing leading to a change in control vector; . - Universal Internal Control, consisting of an RNA in vitro transcript and a DNA plasmid, . incorporated into the kit; - Two modified positive controls replacing the four current positive controls. In addition, the 1:10 . dilution step performed by customers has been removed; - . Additional reactivity for Influenza A/Indiana/10/2011 (H3N2v) and Influenza A/Anhui/1/2013 (H7N9). - e Increase the number of allowable freeze-thaw cycles for the M-MLV Reverse Transcriptase and RNase Inhibitor II from five (5) to ten (10). The labeling was updated accordingly to incorporate the modifications listed above. # SUBSTANTIAL EQUIVALENCE - 1. The Intended Use and Warnings or Precautions of the modified device as described in the labeling have not changed. Confidential to Gen-Probe Prodesse, Inc. {2}------------------------------------------------ The modifications detailed in the table below had not had any effect or caused any changes to the 2. fundamental scientific technology of the device. | Modification | Potential Impact of Modification | Verification/Validation Result | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Outsourcing of controls leading to<br>minor changes in sequence<br>Incorporation of a Universal<br>Internal Control, containing both<br>RNA and DNA internal control<br>sequences. | Modification of the internal control<br>may affect the ability of the device to<br>detect the target organisms.<br>Additionally, it may change the<br>clinical performance of the ProFlu+<br>Assay. | The UIC did not affect the ability of the<br>ProFlu+ Assay to detect target<br>organisms at the limit of detection as<br>evinced by the results of Analytical<br>Sensitivity, IC Interference, Extractor<br>Equivalency, and Sample Stability<br>studies. Additionally, the results of a<br>retrospective clinical comparison study<br>demonstrated the modified ProFlu+<br>Assay with UIC continues to meet the<br>performance claims for the current<br>ProFlu+ Assay. | | Outsourcing of controls leading to<br>minor changes in sequence<br>Two modified positive controls<br>(Inf. A/Inf. B/RSV A Control and<br>RSV B Control) replacing current<br>four positive controls.<br>Modified positive controls<br>provided "at use" concentration,<br>no dilution is necessary. | Changes in the sequences, format<br>(pooled vs. individual), or<br>concentration may affect the<br>performance of the modified positive<br>controls in terms of stability or<br>ability to detect global assay failures. | Stability studies demonstrated current<br>stability claims are met. Clinical<br>validation of the modified positive<br>controls demonstrated their continued<br>ability to monitor for global assay<br>failures. | | Influenza A H3N2v and H7N9*<br>Reactivity Claims | NA | Results of the Reactivity Study<br>demonstrated the ability of the ProFlu+<br>Assay to detect A/Indiana/10/2011<br>(H3N2v) and A/Anhui/1/2013 (H7N9)<br>nucleic acids at concentrations near the<br>limit of detection of the assay. | | Increase the number of freeze-<br>thaw cycles for the M-MLV<br>Reverse Transcriptase and RNase<br>Inhibitor II from five (5) to ten<br>(10) | Increasing the number of freeze-thaw<br>cycles for the MMLV Reverse<br>Transcriptase and RNase Inhibitor II<br>may affect the assay performance. | Stability studies demonstrated that<br>ProFlu+ Assay performance was not<br>affected when the MMLV Reverse<br>Transcriptase and the RNase Inhibitor<br>II underwent 10 freeze-thaw cycles. | * Although this test has been shown to detect A/Anhui/1/2013 H7N9 RNA and influenza A/ Indiana/10/2011 (H3N2v) virus cultured from positive human respiratory specimens, the performance characteristics of this device with clinical specimens that are positive for H7N9 or H3N2v influenza viruses have not been established. The Prodesse ProFlu™+ Assay can distinguish between influenza A and B viruses, but it cannot differentiate influenza A subtypes. - Verification and validation studies performed demonstrated that all clinical and analytical 3. performance/functionality remains unchanged from the previous device. - The appropriate Design Control activities were performed; 4. - A Risk Analysis was performed and did not raise any new concerns of safety and efficacy a. associated with the modifications. - A declaration of conformity with design controls has been submitted. b. The modified ProFlu+ Assay is substantially equivalent to the current legally marketed device, ProFlu+ Assay. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is positioned within a circle, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged around the circle's perimeter. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Emily Ziegler Scientist I Gen-Probe Prodesse, Inc. 20925 Crossroads Circle Waukesha, WI 53186 August 9, 2013 Re: K132129 Trade/Device Name: ProFlu™+ Assay Regulation Number: 21 CFR 866.3980 Regulation Name: Respiratory Virus Panel Multiplex Nucleic *Acid Assay Regulatory Class: Class II Product Code: OCC, OOI Dated: July 9, 2013 Received: July 10, 2013 Dear Ms. Ziegler: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2 - Emily Ziegler forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Uwe Scherf - S for Sally Hojvat, M.Sc., Ph.D. Director, Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indication for Use 510(k) Number (if known): K132129 · Device Name: ProFluTM+ Assay Indications for Use: The ProFlu™+ Assay is a multiplex Real-Time PCR (RT-PCR) in vitro diagnostic test for the rapid and qualitative detection and discrimination of Influenza A Virus. Influenza B Virus, and Respiratory Syncytial Virus (RSV) nucleic acids isolated and purified from nasopharyngeal (NP) swab specimens obtained from symptomatic patients. This test is intended for use to aid in the differential diagnosis of Influenza A. Influenza B and RSV viral infections in humans and is not intended to detect Influenza C. Negative results do not preclude influenza or RSV virus infection and should not be used as the sole basis for treatment or other management decisions. Conversely, positive results do not rule-out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease. The use of additional laboratory testing and clinical presentation must be considered in order to obtain the final diagnosis of respiratory viral infection. Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation (2006 - 2007 respiratory season). Performance characteristics for Influenza A were confirmed when Influenza A/H1, Influenza A/H3, and Influenza A/2009 H1N1 were the predominant Influenza A viruses in circulation (2008 and 2009). When other Influenza A viruses are emerging, performance characteristics may vary. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. | Prescription Use<br>X<br>(21 CFR Part 801 Subpart D) | And/Or | Over the Counter Use<br>(21 CFR Part 801 Subpart C) | |------------------------------------------------------|--------|-----------------------------------------------------| |------------------------------------------------------|--------|-----------------------------------------------------| (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health (CDRH) Tamara V. Feldblyum -S 2013.08.08 12:31:20 -04'00'
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...