K110968 · Gen-Probe Prodesse, Inc. · OCC · Jun 27, 2011 · Microbiology
Device Facts
Record ID
K110968
Device Name
PROFLU+ ASSAY
Applicant
Gen-Probe Prodesse, Inc.
Product Code
OCC · Microbiology
Decision Date
Jun 27, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3980
Device Class
Class 2
Attributes
Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K110968 · Jun 27, 2011
PROFLU+ ASSAY
Gen-Probe Prodesse, Inc.
Archived prospectively collected clinical samples from 2008-2009 respiratory seasons
The sponsor used a cohort of archived clinical samples to establish performance characteristics of the reformulated ProFlu+ assay by comparing it to the original cleared assay and confirming discordant results with bi-directional sequencing.
Reformulated ProFlu+ Assay Clinical Comparison; Retrospective comparative study using archived clinical samples; Study Period: 2008-2009 respiratory seasons
Symptomatic patients presenting with respiratory symptoms; Sample Size: 233; Number of Sites: 2
Original ProFlu+ Assay (K073029)
Percent positive and negative agreement between reformulated and original assay
Indications for Use
The ProFlu™+ Assay is a multiplex Real-Time PCR (RT-PCR) in vitro diagnostic test for the rapid and qualitative detection and discrimination of Influenza A Virus, Influenza B Virus, and Respiratory Syncytial Virus (RSV) nucleic acids isolated and purified from nasopharyngeal (NP) swab specimens obtained from symptomatic patients. This test is intended for use to aid in the differential diagnosis of Influenza A, Influenza B and RSV viral infections in humans and is not intended to detect Influenza C. Negative results do not preclude influenza or RSV virus infection and should not be used as the sole basis for treatment or other management decisions. Conversely, positive results do not ruleout bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease. The use of additional laboratory testing and clinical presentation must be considered in order to obtain the final diagnosis of respiratory viral infection. Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation (2006 - 2007 respiratory season). Performance characteristics for Influenza A were confirmed when Influenza A/H1, Influenza A/H3, and Influenza A/2009 H1N1 were the predominant Influenza A viruses in circulation (2008 and 2009). When other Influenza A viruses are emerging, performance characteristics may vary. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
Device Story
The ProFlu+ Assay is a multiplex RT-PCR test for qualitative detection of Influenza A, Influenza B, and RSV (A/B) in nasopharyngeal swabs. The process involves: 1) Nucleic acid isolation/purification using Roche MagNA Pure LC or bioMérieux NucliSENS easyMAG systems; 2) Addition of Internal Control (IC) to monitor for inhibitors; 3) RT-PCR amplification and detection on the Cepheid SmartCycler II instrument. The assay uses Taqman chemistry where target-specific primers and dual-labeled probes (reporter/quencher) are used. Taq polymerase 5'-3' exonuclease activity cleaves the probe during amplification, releasing the reporter dye to generate a fluorescent signal. The SmartCycler Dx software monitors fluorescence in real-time to determine results. The device is used in clinical laboratories by trained personnel. Results aid clinicians in differential diagnosis of respiratory viral infections. The assay includes positive RNA transcript controls and an internal control to ensure procedural integrity and monitor for PCR inhibition.
Clinical Evidence
Performance established via comparison of reformulated assay to original ProFlu+ Assay using 233 prospectively collected archived samples (2008-2009). Concordance confirmed by bidirectional sequencing. Influenza A: 100% PPA, 99.4% NPA. Influenza B: 100% PPA, 100% NPA. RSV: 100% PPA, 99.0% NPA. Analytical specificity (cross-reactivity) tested against 58 respiratory pathogens with 100% specificity. Limit of Detection (LoD) for reformulated assay confirmed identical to original for all strains.
Technological Characteristics
Multiplex RT-PCR assay. Targets: Influenza A (Matrix gene), RSV A/B (Polymerase gene), Influenza B (NS1/NS2 genes). Chemistry: Taqman (5'-3' exonuclease activity). Instrumentation: Cepheid SmartCycler II. Detection: FAM, TET, Texas Red, Cy5 fluorophores. Internal control included for inhibition monitoring. Nucleic acid extraction via MagNA Pure LC or NucliSENS easyMAG.
Indications for Use
Indicated for symptomatic patients (adults and children) requiring differential diagnosis of Influenza A, Influenza B, and RSV viral infections via nasopharyngeal swab specimens.
Regulatory Classification
Identification
A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B; (2) Influenza A subtype H1 and Influenza A subtype H3; (3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B; (4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus; (5) Human Metapneumovirus; (6) Rhinovirus; and (7) Adenovirus.
Special Controls
*Classification.* Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## A. 510(k) Number:
K110968
## B. Purpose for Submission:
Modification of the previously cleared ProFlu+ Rt-PCR assay (K092500). Two new Influenza A probes located directly upstream of the current ProFlu+ Influenza A probe and containing a degenerate nucleotide (C or T) at the 3' terminus were added to the supermix to improve the detection of the 2009 H1N1 influenza viruses containing point mutation in the matrix gene at the probe binding.
## C. Measurand:
Respiratory specimen virus nucleic acid (RNA) target sequences. The targeted viruses have been associated with respiratory infections in adults and/or children. Viral types detected: Influenza A, Influenza B, Respiratory Syncytial Virus Type A, Respiratory Syncytial Virus Type B
## D. Type of Test:
Multiplex nucleic acid assay for qualitative determination of Influenza A, Influenza B, Respiratory Syncytial Virus Type A, Respiratory Syncytial Virus Type B) in nasopharyngeal swabs obtained from individuals with signs and symptoms of respiratory tract infections.
## E. Applicant:
Gen-Probe Prodesse, Inc.
## F. Proprietary and Established Names
ProFlu+™ assay
Common Name: Respiratory Viral Panel (RVP) multiplex nucleic acid assay
## G. Regulatory Information:
1. Regulation section: 866.3980 - Respiratory viral panel multiplex nucleic acid assay
2. Classification: Class II
3. Product code: OCC, OOI
4. Panel: Microbiology (83)
## H. Intended Use:
1. Intended use:
The ProFlu™+ Assay is a multiplex Real-Time PCR (RT-PCR) in vitro diagnostic test for the rapid and qualitative detection and discrimination of Influenza A Virus, Influenza B Virus, and Respiratory Syncytial Virus (RSV)
{2}
nucleic acids isolated and purified from nasopharyngeal (NP) swab specimens obtained from symptomatic patients. This test is intended for use to aid in the differential diagnosis of Influenza A, Influenza B and RSV viral infections in humans and is not intended to detect Influenza C.
Negative results do not preclude influenza or RSV virus infection and should not be used as the sole basis for treatment or other management decisions.
Conversely, positive results do not rule-out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease. The use of additional laboratory testing and clinical presentation must be considered in order to obtain the final diagnosis of respiratory viral infection.
Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation (2006 – 2007 respiratory season). Performance characteristics for Influenza A were confirmed when Influenza A/H1, Influenza A/H3, and Influenza A/2009 H1N1 were the predominant Influenza A viruses in circulation (2008 and 2009). When other Influenza A viruses are emerging, performance characteristics may vary.
If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
2. Indication(s) for use:
Same as Intended Use
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Roche MagNA Pure LC System with software version 3.0.11 or bioMérieux NucliSENS easyMAG System with Software version 1.0.1 or 2.0. Cepheid SmartCycler II Real Time Instrument with Dx software version 1.7b or 3.0 a/b.
I. Device Description:
The ProFlu+ Assay enables detection and differentiation of Influenza A Virus, Influenza B Virus, Respiratory Syncytial Virus (Types A and B), and Internal Control.
2
{3}
An overview of the procedure is as follows:
1. Collect nasopharyngeal swab specimens from symptomatic patients using a polyester, rayon or nylon tipped swab and place into viral transport medium (refer to Materials Required but not Provided).
2. Add an Internal Control (IC) to every sample to monitor for inhibitors present in the specimens.
3. Perform isolation and purification of nucleic acids using a MagNA Pure LC System (Roche) and the MagNA Pure Total Nucleic Acid Isolation Kit (Roche) or a NucliSENS easyMAG System (bioMérieux) and the Automated Magnetic Extraction Reagents (bioMérieux).
4. Add purified nucleic acids to Influenza A/Influenza B/RSV Mix along with enzymes included in the ProFlu+ Detection Kit. The Influenza A/Influenza B/RSV Mix contains oligonucleotide primers and target-specific oligonucleotide probes. The primers are complementary to highly conserved regions of genetic sequences for these respiratory viruses. The probes are dual-labeled with a reporter dye and a quencher (see table below).
5. Perform reverse transcription of RNA into complementary DNA (cDNA) and subsequent amplification of DNA in a Cepheid SmartCycler II instrument. In this process, the probe anneals specifically to the template followed by primer extension and amplification. The ProFlu+ Assay is based on Taqman reagent chemistry, which utilizes the $5^{\prime} - 3^{\prime}$ exonuclease activity of the Taq polymerase to cleave the probe thus separating the reporter dye from the quencher. This generates an increase in fluorescent signal upon excitation from a light source. With each cycle, additional reporter dye molecules are cleaved from their respective probes, further increasing fluorescent signal. The amount of fluorescence at any given cycle is dependent on the amount of amplification products present at that time. Fluorescent intensity is monitored during each PCR cycle by the real-time instrument.
| Analyte | Gene Targeted | Probe Fluorophore | Absorbance Peak | Emission Peak | Instrument Channel |
| --- | --- | --- | --- | --- | --- |
| Influenza A Virus | Matrix | FAM | 495 nm | 520 nm | FAM |
| Respiratory Syncytial Virus A | Polymerase | CAL Fluor Orange 560 | 540 nm | 561 nm | TET |
| Respiratory Syncytial Virus B | Polymerase | CAL Fluor Orange 560 | 540 nm | 561 nm | TET |
| Influenza B Virus | Non-structural NS1 and NS2 | CAL Fluor Red 610 | 595 nm | 615 nm | Texas Red |
| Internal Control | NA | Quasar 670 | 647 nm | 667 nm | Cy5 |
# Materials Provided
ProFlu+ Assay consists of a Detection Kit and Control Kit.
Detection Kit (100 Reactions)
{4}
4
| Reagents | Description | Quantity/ Tube | Cap Color | Reactions/ Tube |
| --- | --- | --- | --- | --- |
| Influenza A/ Influenza B/ RSV Mix | ➤ Taq DNA polymerase
➤ oligonucleotide primers
➤ oligonucleotide probes
➤ Buffer containing dNTPs (dATP, dCTP, dGTP, dTTP),
➤ MgCl_{2} and stabilizers
➤ Bovine serum albumin | 1030 μL | Brown | 50
(2 tubes provided) |
| M-MLV Reverse Transcriptase | ➤ 10U/μL | 36 μL | Red | 100 |
| RNase Inhibitor II | ➤ 40U/μL | 120 μL | Green | 100 |
| Internal RNA Control III | ➤ Non-infectious in vitro
transcribed RNA | 30 μL | Yellow | 100 |
## Control Kit
| Reagents | Description | Quantity/ Tube | Cap Color | Reactions/ Tube |
| --- | --- | --- | --- | --- |
| Influenza A RNA Control III | ➤ Non-infectious in vitro
transcribed RNA of specific viral sequences | 300 μL | White | 15 |
| Influenza B RNA Control III | ➤ Non-infectious in vitro
transcribed RNA of specific viral sequences | 300 μL | Blue | 15 |
| RSV A RNA Control III | ➤ Non-infectious in vitro
transcribed RNA of specific viral sequences | 300 μL | Purple | 15 |
| RSV B RNA Control III | ➤ Non-infectious in vitro
transcribed RNA of specific viral sequences | 300 μL | Clear | 15 |
## Materials Required but not Provided
### Plasticware and consumables
☐ Polyester, rayon or nylon tipped nasopharyngeal swabs
☐ RNase/DNase-free 1.5 mL polypropylene microcentrifuge tubes
☐ Sterile RNase/DNase-free filter or positive displacement micropipettor tips
☐ MagNA Pure LC System Disposables (Reagent Tubs, Reaction Tips, Tip Trays, Cartridges) or easyMAG System Disposables (Sample Strips and Tips)
☐ Biohit Pipette Tips for use with easyMAG System
☐ Greiner Break Four uncoated plates for use with easyMAG System
☐ Cepheid PCR reaction tubes, 25 μL
☐ Parafilm M or MagNA Pure LC Cartridge Seals
### Reagents
☐ Roche MagNA Pure LC Total Nucleic Acid Isolation Kit) for 192 isolations or bioMérieux NucliSENS easyMAG reagents
{5}
☐ Micro Test M4 Viral Transport Medium Micro Test M5 Viral Transport Medium, Micro Test M6 Viral Transport Medium, Micro Test M4RT Viral Transport Medium, Copan Universal Transport Medium, or BD Universal Viral Transport vial, 3mL
☐ Molecular Grade Water (RNase/DNase Free)
☐ Extraction Control (e.g. previously characterized positive sample)
NOTE: Only qualified lots of the MagNA Pure LC Total Nucleic Acid Isolation Kit can be used with the ProFlu+ Assay. Lots not specifically qualified by Gen-Probe Prodesse, Inc. for use with the ProFlu+ Assay are not verified for use with this assay, and may cause erroneous results.
A list of these qualified extraction reagents is available at www.gen-probe.com. Please notify the reagent manufacturer of issues with this ancillary reagent and Gen-Probe Prodesse, Inc. of the impact on the performance of the ProFlu+ Assay.
## Equipment
☐ – 70°C Freezer
☐ Roche MagNA Pure LC System with software version 3.0.11 or bioMérieux NucliSENS easyMAG System with Software version 1.0.1 or 2.0
☐ Biohit multi-channel pipettor for use with easyMAG System
☐ Cepheid SmartCycler II Real Time Instrument with Dx Software version 1.7b, 3.0a, or 3.0b
☐ Micropipettors (range between 1-10 µL, 10-200 µL and 100-1000 µL)
☐ Mini-centrifuge with adapter for Cepheid Reaction Tubes
☐ Cepheid cooling block
☐ Ice/Ice Bucket or -20°C Cold Block
☐ Biosafety Cabinet
## Interpretation of Specimen Results
The SmartCycler Dx software automatically determines the specimen results. The interpretation of the assay specimen results is as follows:
| Sample ID¹ | Assay Result | IC Result | Warning / Error Code | Influenza A Result | RSV Result | Influenza B Result | Interpretation of Results |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Sample ID | Negative | Pass | | NEG | NEG | NEG | Influenza A, B and RSV nucleic acid not detected |
| Sample ID | Positive | NA* | | POS | NEG | NEG | Influenza A nucleic acid detected |
| Sample ID | Positive | NA* | | NEG | POS | NEG | RSV nucleic acid detected |
| Sample ID | Positive | NA* | | NEG | NEG | POS | Influenza B nucleic acid detected |
| Sample ID | Positive | NA* | | POS | NEG | NEG | Influenza A and RSV nucleic acid detected |
| Sample ID | Positive | NA* | | POS | NEG | NEG | Influenza A and |
{6}
6
| | | | | | | | Influenza B nucleic acid detected |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Sample ID | Positive | NA* | | NEG | POS | POS | RSV and Influenza B nucleic acid detected |
| Sample ID | Positive | NA* | | POS | POS | POS | Influenza A, Influenza B and RSV nucleic acid detected. Triple infections are rare, repeat testing from the purified nucleic acid or collect and test a new sample. |
| Sample ID | Unresolved | Fail | | NEG | NEG | NEG | Unresolved – PCR inhibition or reagent failure. Repeat testing from the purified nucleic acid or collect and test a new sample. |
| Sample ID | ND | ND | 3079² | ND | ND | ND | Not Determined – error code 3079 |
| Sample ID | Invalid | | 4098³ | ND | ND | ND | Not Determined – error code 4098 |
¹ Columns and data not used for interpretation are not included
² Error Code 3079: Warning/Error Code 3079 is periodically observed with Influenza A positives (Influenza A Positive Control, Influenza A positive NP swab samples). Warning/Error Code 3079 occurs when the fluorescence (RFU) signal is too high. In this case, all results for that sample are reported by the Dx software as ND (Not Determined). If a Ct value ≥ 13 is reported in the Influenza A, RSV, and/or Influenza B Ct columns, the sample results can be recorded as POS for the specific analyte(s).
³ An Invalid assay run will display Error Code 4098
* Detection of the Internal Control in the Cy5 detection channel is not required for positive result. High viral load can lead to reduced or absent Internal Control signal.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
Luminex Molecular Diagnostics, xTAG™ Respiratory Viral Panel (RVP)
Gen-Probe Prodesse, ProFlu+ Assay
2. Predicate 510(k) number(s):
K063765, K073029
3. Comparison with predicate:
| Features | New ProFlu+ Assay | Current ProFlu+ Assay | Luminex RVP |
| --- | --- | --- | --- |
| 510(k) | TBD | K073029 | K063765 |
| Regulation | 866.3980 | 866.3980 | 866.3980 |
| Product Code | OCC | OCC | OCC, OEM, OEP |
| Device Class | Class II | Class II | Class II |
| Intended Use | For the in vitro qualitative detection and differentiation of influenza A, influenza | For the in vitro qualitative detection and differentiation of influenza A, influenza | Direct and differential qualitative detection of influenza types A |
{7}
| Features | New ProFlu+ Assay | Current ProFlu+ Assay | Luminex RVP |
| --- | --- | --- | --- |
| | B, and RSV viral nucleic acids. | B, and RSV viral nucleic acids. | and B, RSV types A and B, Parainfluenza types 1, 2 and 3, Adenovirus and Rhinovirus viral nucleic acids. |
| Technology/Detection | Real Time RT-PCR Detection | Real Time RT-PCR Detection | RT-PCR Detection: Amplified products are coupled to microspheres and detected using spectrofluorometric analysis. |
| Specimen Types | NP swabs | NP swabs | NP swabs |
| Nucleic Acid Isolation | Roche MagNA Pure LC System and bioMerieux NucliSENS easyMAG | Roche MagNA Pure LC System and bioMerieux NucliSENS easyMAG | NucliSENS® miniMAG extraction Kit (bioMerieux) QIAamp® MiniElute® Virus Spin Kit (Qiagen) |
| Instrument /Assay Platform | Cepheid SmartCycler II System | Cepheid SmartCycler II System | Luminex 100 or 200 |
| Assay Controls | Influenza A, Influenza B, RSV A, RSV B positive RNA transcript controls and an Internal RNA control provided | Influenza A, Influenza B, RSV A, RSV B positive RNA transcript controls and an Internal RNA control provided | Bacteriophage lambda positive control and E. coli MS2 phage Internal Control –ancillary reagents not provided |
# K. Standard/Guidance Document referenced (if applicable):
- User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline (CLSI EP12-A 2002)
- Protocols for Determination of Limits of Detection and Limits of Quantitation (CLSI EP-17A 2004)
- Molecular Diagnostic Methods for Infectious Diseases; Approved Guideline (CLSI MM3-A2 2006)
- Nucleic Acid Sequencing Methods in Diagnostic Laboratory Medicine; Approved Guideline (CLSI MM9-A 2004)
- Collection, Transport, Preparation, and Storage of Specimens for Molecular Methods (CLSI MM13A 2006)
Medical devices - Quality management systems - Requirements for regulatory purpose (ISO 13485:2003)
Medical devices - Application of risk management to medical devices (ISO 14971:2007)
- Stability Testing of In Vitro Diagnostic Reagents (CEN 13640:2002)
{8}
- Format for Traditional and Abbreviated 510(k)s - Guidance for Industry and FDA Staff
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm084365.html
- Nucleic Acid Based In Vitro Diagnostic Devices for Detection of Microbial Pathogens - Draft Guidance for Industry and FDA Staff
http://www.fda.gov/ohrms/dockets/98fr/05d-0434-gd10001.pdf
- Guidance for Off-the-Shelf Software Use in Medical Devices; Final
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm073778.html
- Statistical Guidance on Reporting Results from Studies Evaluating Diagnostic Tests; Draft Guidance for Industry and FDA Reviewers
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm071148.html
- Guidance on Informed Consent for In Vitro Diagnostic Device Studies Using Leftover Human Specimens that are Not Individually Identifiable - Guidance for Sponsors, Institutional Review Boards, Clinical Investigators and FDA Staff
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm078384.html
- Draft Guidance for Industry and FDA Staff: Establishing the Performance Characteristics of In Vitro Diagnostic Devices for the Detection or Detection and Differentiation of Influenza Viruses
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm079171.html
- In Vitro Diagnostic Devices to Detect Influenza A Viruses: Labeling and Regulatory Path - Guidance for Industry and FDA Staff
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm078538.html
- Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm180307.html
L. Test Principle:
ProFlu+ assay enables detection and differentiation of Influenza A Virus, Influenza B Virus, Respiratory Syncytial Virus (Types A and B), and Internal Control.
Nucleic acids are isolated and purified from nasopharyngeal swab specimens using a MagNA Pure LC System (Roche) and the MagNA Pure Total Nucleic Acid Isolation Kit (Roche) or a NucliSENS easyMAG System (bioMérieux) and the Automated Magnetic Extraction Reagents (bioMérieux). Purified nucleic acids are added to Influenza A/Influenza B/RSV Mix along with enzymes included in the ProFlu+ Detection Kit. The Influenza A/Influenza B/RSV Mix contains oligonucleotide primers and target-specific oligonucleotide probes. The primers are complementary to highly conserved regions of genetic sequences for these respiratory viruses. The
{9}
probes are dual-labeled with a reporter dye and a quencher attached to the 3' end or internally.
First RNA is reverse transcribed into complementary DNA (cDNA) which is then subsequently amplified in a Cepheid SmartCycler II instrument. In this process, the probe anneals specifically to the template followed by primer extension and amplification. Taq polymerase cleaves the probe utilizing its $5^{\prime} - 3^{\prime}$ exonuclease activity thus separating the reporter dye from the quencher. This generates an increase in fluorescent signal upon excitation from a light source. With each cycle, additional reporter dye molecules are cleaved from their respective probes, further increasing fluorescent signal. The amount of fluorescence at any given cycle is dependent on the amount of amplification products present at that time. Fluorescent intensity is monitored during each PCR cycle by the real-time instrument.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
Please refer to previously FDA-cleared 510(k) Premarket Notification, K092500 for additional information.
# b. Linearity/assay reportable range:
Not applicable.
# c. Traceability, Stability, Expected values (controls, calibrators, or methods):
There are no changes to the controls for the New ProFlu+ Assay. Please refer to previously FDA-cleared 510(k) Premarket Notification, K092500.
# d. Detection limits:
The analytical sensitivity (limit of detection or LoD) of the ProFlu+ Assay was determined using 6 Influenza A strains, 1 strain each of Influenza B, RSV A and RSV B. Each viral strain was extracted using the Roche MagNA Pure LC instrument and tested in replicates of 60 per concentration. Analytical sensitivity (LoD) defined as the lowest concentration at which $\geq 95\%$ of all replicates tested positive are summarized in the table below. The LoDs for the reformulated ProFlu+ Assay were identical to the original ProFlu+ Assay for all strains tested.
| Viral Strain | LoD Concentration |
| --- | --- |
| Influenza A/ Virginia/1/06 (H1N1) | 1x10^0 TCID50/mL |
| Influenza A/New Caledonia/12/99 (H1N1) | 1x10^3 TCID50/mL |
| Influenza A/Port Chalmers/1/73 (H3N2) | 1x10^2 TCID50/mL |
| Influenza A/California/07/04 (H3N2) | 1x10^0 TCID50/mL |
| Influenza A/California/04/09 (2009 H1N1) | 1x10^1 TCID50/mL |
{10}
e. Analytical specificity:
# Cross Reactivity
The analytical specificity of the ProFlu+ Assay was evaluated by testing a panel of 58 cultures consisting of 31 viral, 26 bacterial, and 1 yeast strain representing common respiratory pathogens or flora commonly present in nasopharynx. Bacteria and yeast were tested at concentrations of $10^{5}$ to $10^{8}$ CFU/mL. Viruses were tested at concentrations of $10^{3}$ to $10^{6}$ TCID $_{50}$ /mL, except where noted. Samples were extracted using the Roche MagNA Pure LC instrument and tested in triplicate. Analytical specificity of the ProFlu+ Assay was $100\%$ .
| Strains | Concentration | Influenza A | RSV | Influenza B |
| --- | --- | --- | --- | --- |
| H1N1 IA New Caledonia | 103TCID50/ml | + | - | - |
| H3N2 IA Port Chalmers | 103TCID50/ml | + | - | - |
| IB Wisconsin | 103TCID50/ml | - | - | + |
| RSV A Long | 102TCID50/ml | - | + | - |
| RSV B Wash | 103TCID50/ml | - | + | - |
| Adenovirus 1/Adenoid 71 | 106TCID50/mL | - | - | - |
| Adenovirus 7 | 106TCID50/mL | - | - | - |
| Coronavirus 229E | 106TCID50/mL | - | - | - |
| Coxsackie B4 | 104TCID50/mL | - | - | - |
| Coxsackie B5/10/2006 | 105TCID50/mL | - | - | - |
| Cytomegalovirus | 104TCID50/mL | - | - | - |
| Echovirus 2 | 106TCID50/mL | - | - | - |
| Echovirus 3 | 105TCID50/mL | - | - | - |
| Echovirus 6 | 105TCID50/mL | - | - | - |
| Echovirus 11 | 106TCID50/mL | - | - | - |
| Enterovirus 68 | 103TCID50/mL | - | - | - |
| Enterovirus 70 | 103TCID50/mL | - | - | - |
| Epstein Barr Virus | 108copies/mL | - | - | - |
| HSV Type 1 MacIntyre Strain | 105TCID50/mL | - | - | - |
| HSV Type 2 G strain | 105TCID50/mL | - | - | - |
| Human Metapneumovirus A2 | 104TCID50/mL | - | - | - |
| Human Rhinovirus 1a | 103TCID50/mL | - | - | - |
| Human Rhinovirus | 103TCID50/mL | - | - | - |
| Measles/7/2000 | 104TCID50/mL | - | - | - |
| Mumps Virus | 103TCID50/mL | - | - | - |
| Parainfluenza Type 1 | 103TCID50/mL | - | - | - |
| Parainfluenza Type 2 | 105TCID50/mL | - | - | - |
| Parainfluenza Type 3 | 106TCID50/mL | - | - | - |
| Parainfluenza Type 4 | 104TCID50/mL | - | - | - |
| Poliovirus 1 | 106TCID50/mL | - | - | - |
| Varicella Zoster Virus | 104TCID50/mL | - | - | - |
| Vibrio 1 | 106TCID50/mL | - | - | - |
| Vibrio 2 | 106TCID50/mL | - | - | - |
| Vibrio 3 | 106TCID50/mL | - | - | - |
| Vibrio 4 | 106TCID50/mL | - | - | - |
| Vibrio 5 | 106TCID50/mL | - | - | - |
{11}
| Strains | Concentration | Influenza A | RSV | Influenza B |
| --- | --- | --- | --- | --- |
| Bordetella pertussis | 10^6 cfu/mL | - | - | - |
| Bordetella bronchoiseptica | 10^7 cfu/mL | - | - | - |
| Chlamydia pneumonia | 10^6 TCID50/mL | - | - | - |
| Chlamydia trachomatis | 10^6 TCID50/mL | - | - | - |
| Legionella micdadei | 10^7 cfu/mL | - | - | - |
| Legionella pneumophila | 10^7 cfu/mL | - | - | - |
| Mycobacterium intracellulare | 10^6 cfu/mL | - | - | - |
| Mycobacterium tuberculosis | 10^8 cfu/mL | | | |
| Mycoplasma pneumonia | 10^6 cfu/mL | - | - | - |
| Haemophilus influenza | 10^8 cfu/mL | - | - | - |
| Pseudomonas aeruginosa | 10^7 cfu/mL | - | - | - |
| Proteus vulgaris | 10^7 cfu/mL | - | - | - |
| Proteus mirabilis | 10^7 cfu/mL | - | - | - |
| Neisseria gonorrhoeae | 10^7 cfu/mL | - | - | - |
| Neisseria meningitides | 10^7 cfu/mL | - | - | - |
| Neisseria mucosa | 10^7 cfu/mL | - | - | - |
| Klebsiella pneumonia | 10^7 cfu/mL | - | - | - |
| Escherichia coli | 10^7 cfu/mL | - | - | - |
| Moraxella catarrhalis | 10^6 cfu/mL | - | - | - |
| Corynebacterium diphtheriae | 10^7 cfu/mL | - | - | - |
| Lactobacillus plantarum | 10^7 cfu/mL | - | - | - |
| Streptococcus pneumoniae | 10^8 cfu/mL | - | - | - |
| Streptococcus pyogenes | 10^7 cfu/mL | - | - | - |
| Streptococcus salivarius | 10^6 cfu/mL | - | - | - |
| Staphylococcus epidermidis | 10^7 cfu/mL | - | - | - |
| Staphylococcus aureus | 10^7 cfu/mL | - | - | - |
| Candida albicans | 10^7 cfu/mL | - | - | - |
# Reactivity
The reactivity of the ProFlu+ Assay was evaluated against multiple strains of Influenza A, Influenza B, and Respiratory Syncytial Viruses. The panel consisted of 13 Influenza A subtype H1N1, 13 Influenza A subtype H3N2, 9 swine-origin Influenza A, 2 Influenza A subtype H5N1, 10 Influenza B, and 2 Respiratory Syncytial Virus strains. Each viral strain was extracted using the Roche MagNA Pure LC and tested in triplicate in each assay.
| Viral Strain | Concentration | Influenza A | RSV | Influenza B |
| --- | --- | --- | --- | --- |
| A/Taiwan/42/06 (H1N1) | 1x10^1 TCID50/mL | + | - | - |
| A/Henan/8/05 (H1N1) | 1x10^1 TCID50/mL | + | - | - |
| A/Fujian/156/00 (H1N1) | 1x10^1 TCID50/mL | + | - | - |
| Brazil/1137/99 (H1N1) | 1x10^3 TCID50/mL | + | - | - |
{12}
| Viral Strain | Concentration | Influenza A | RSV | Influenza B |
| --- | --- | --- | --- | --- |
| A/Kentucky/2/06 (H1N1) | 1x102TCID50/mL | + | - | - |
| A/Hawaii/15/01 (H1N1) | 1x103TCID50/mL | + | - | - |
| A/Brisbane/59/2007 (H1N1) | 1x102TCID50/mL | + | - | - |
| A/Solomon Islands/03/06 (H1N1) | 2x101TCID50/mL | + | - | - |
| A/Jiangxi/160/05 (H1N1) | 2x101TCID50/mL | + | - | - |
| A/WS/33 (H1N1) | 5x101.75TCID50/mL | + | - | - |
| A1/Mal/302/54 (H1N1) | 5x105.25TCID50/mL | + | - | - |
| A/PR/8/34 (H1N1) | 1x106TCID50/mL | + | - | - |
| VR 546 A1/Denver/1/57 (H1N1) | 5x105.25TCID50/mL | + | - | - |
| A/Hiroshima/52/05 (H3N2) | 2x100TCID50/mL | + | - | - |
| A/Victoria/512/05 (H3N2) | 2x100TCID50/mL | + | - | - |
| VR 822 A/Victoria/3/75 (H3N2) | 2x103TCID50/mL | + | - | - |
| A/Brazil/02/99 (H3N2) | 2x102TCID50/mL | + | - | - |
| A/New York/55/2004 (H3N2) | 2x100TCID50/mL | + | - | - |
| A/Hong Kong/2831/05 (H3N2) | 2x101TCID50/mL | + | - | - |
| A/Bahamas/2686/99 (H3N2) | 2x102TCID50/mL | + | - | - |
| A/Fuijan/411/02 (H3N2) | 2x102TCID50/mL | + | - | - |
| A/Kentucky/03/06 (H3N2) | 2x102TCID50/mL | + | - | - |
| A/Costa Rica/07/99 (H3N2) | 2x102TCID50/mL | + | - | - |
| A/Hong Kong/218/06 (H3N2) | 2x102TCID50/mL | + | - | - |
| VR 544 A/Hong Kong/8/68 (H3N2) | 2x102CEID50/mL | + | - | - |
| VR 547 A/Aichi/2/68 (H3N2) | 2x102CEID50/mL | + | - | - |
| 2009 H1N1 Clinical Isolate #1 | 2x102TCID50/mL | + | - | - |
| 2009 H1N1 Clinical Isolate #2 | 2x103TCID50/mL | + | - | - |
| 2009 H1N1 Clinical Isolate #3 | 2x102TCID50/mL | + | - | - |
| 2009 H1N1 Clinical Isolate #4 | 2x101TCID50/mL | + | - | - |
| 2009 H1N1 Clinical Isolate #5 | 2x100TCID50/mL | + | - | - |
| A/New Jersey/8/76 (Swine-origin) | 2x104CEID50/mL | + | - | - |
| A/South Dakota/03/2008 (Swine-origin) | 2x103TCID50/mL | + | - | - |
| A/Wisconsin/10/1998 (swine-origin) | 2x103TCID50/mL | + | - | - |
| A/Iowa/2006 (swine-origin) | 2x103TCID50/mL | + | - | - |
| H5N1/VN/1203 2.7ng/μL RNA | 2.7ng/μL | + | - | - |
| H5N1/HK/486 1.4ng/μL RNA | 1.4ng/μL | + | - | - |
| B/Hawaii/11/05 | 1x102TCID50/mL | - | - | + |
| B/Michigan/2/06 | 1x102TCID50/mL | - | - | + |
| B/Hawaii/33/2004 | 1x102TCID50/mL | - | - | + |
| B/Ohio/1/2005 | 1x102TCID50/mL | - | - | + |
| B/Florida/2/06 | 1x102TCID50/mL | - | - | + |
| B/St. Petersburg/04/06 | 1x102TCID50/mL | - | - | + |
| B/Michigan/4/06 | 1x102TCID50/mL | - | - | + |
| B/Malaysia 2506/2004 | 1x102TCID50/mL | - | - | + |
| B/Florida/7/2004 | 1x102TCID50/mL | - | - | + |
| B/Lee/40 | 1x102TCID50/mL | - | - | + |
| RSV A strain A2 | 1x102TCID50/mL | - | + | - |
| RSV B strain 9320 | 1x102TCID50/mL | - | + | - |
$\diamond$ Strains not re-cultured and titered. The original culture/titer from ATCC was used in this study.
# f. Interference Studies:
{13}
Competitive Interference of the ProFlu+ Assay was evaluated using simulated samples containing pairs of target viruses (Influenza A and Influenza B, Influenza A and RSV, Influenza B and RSV) at two different concentrations. One of the concentrations was near the Limit of Detection (LoD) while the other concentration was 1000x the LoD. Samples were extracted using the Roche MagNA Pure LC instrument and tested in triplicate. The presence of two viruses at varying concentrations in a single sample had no effect on the analytical sensitivity (limit of detection or LoD) of the ProFlu+ Assay.
g. Extraction equivalency:
Extraction equivalency of the bioMérieux NucliSENS easyMAG and Roche MagNA Pure LC instruments was evaluated by performing a limit of detection study and a reproducibility study using both instrument systems. For the limit of detection study, Influenza A, Influenza B, RSV A or RSV B virus was diluted into NP swab matrix to the study's limit of detection. Each viral strain dilution was extracted in replicates of 10 on each automated extractor and tested using the ProFlu+ Assay. For the reproducibility study, a panel of 6 simulated samples that included medium and low (near the assay limit of detection) Influenza A and RSV A positive and negative samples was tested. The panel was run on each automated extractor 2 times per day for a total of 5 days.
The bioMérieux NucliSens easyMAG instrument performed equivalently to the Roche MagNA Pure LC instrument with respect to limit of detection and reproducibility (percent agreement). The limits of detection were the same for both instruments and the overall reproducibility percent agreement was 100%.
h. Carry-over/Cross-Contamination
In an internal study there was no evidence of carry-over/cross contamination with the ProFlu+ Assay using either the Roche MagNA Pure LC or the bioMérieux NucliSens easyMAG automated nucleic acid extraction instruments.
2. Comparison studies:
Clinical Comparison Results
The ProFlu+ Assay's supermix was reformulated and performance characteristics were established by comparing the reformulated assay to the original ProFlu+ Assay. All samples positive for Influenza A, Influenza B or RSV using either the current ProFlu+ Assay and/or the reformulated "New" ProFlu+ Assay were confirmed using bidirectional sequencing. The sequencing assays targeted either a different gene than the ProFlu+ Assay or targeted a different region of the same gene as the ProFlu+ Assay. Prospectively collected archived samples from respiratory season years 2008 and 2009 that were collected at two clinical study sites (Columbus, OH and Albuquerque, NM) were used for this study.
13
{14}
"True" influenza A, influenza B or RSV positives were considered as any sample that tested positive for the respective analyte by the original ProFlu+ Assay.
"True" influenza A, influenza B or RSV negatives were considered as any sample that tested negative by the original ProFlu+ Assay.
Influenza A Comparison Results
| | Current ProFlu+ Assay | | | | |
| --- | --- | --- | --- | --- | --- |
| | | Positive | Negative | Total | Comments |
| New Pro Flu+ | Positive | 60 | 1* | 61 | Percent Positive Agreement 100% (93.98%-100%) 95% CI |
| | Negative | 0 | 172 | 172 | Percent Negative Agreement 99.4% (96.80%-99.90%) 95% CI |
| | Total | 60 | 173 | 233 | |
* Sample was positive for Influenza A using bi-directional sequencing.
Influenza B Comparison Results
| | Current ProFlu+ Assay | | | | |
| --- | --- | --- | --- | --- | --- |
| | | Positive | Negative | Total | Comments |
| New Pro Flu+ Assay | Positive | 14 | 0 | 14 | Percent Positive Agreement 100% (78.47% - 100%) 95% CI |
| | Negative | 0 | 219 | 219 | Percent Negative Agreement 100% (98.28% - 100%) 95% CI |
| | Total | 14 | 219 | 233 | |
RSV Comparison Results
| | Current ProFlu+ Assay | | | | |
| --- | --- | --- | --- | --- | --- |
| | | Positive | Negative | Total | Comments |
| New Pro Flu+ | Positive | 35 | 2* | 37 | Percent Positive Agreement 100% (90.11% - 100%) 95% CI |
| | Negative | 0 | 196 | 196 | Percent Negative Agreement 99.0% (96.39%-99.72%) 95% CI |
| | Total | 35 | 198 | 233 | |
* Two samples were positive for RSV using bi-directional sequencing.
3. Clinical studies:
Clinical performance characteristics of the ProFlu+ Assay were established during a prospective study at 3 U.S. clinical laboratories and a retrospective study at 1 U.S. site during the 2006-2007 respiratory virus season (February - April). Please refer to previously FDA-cleared 510(k) Premarket Notification, K092500 for additional information.
4. Clinical cut-off: N/A
5. Expected values/Reference range:
{15}
The prevalence of Influenza and RSV varies each year with epidemics occurring during the fall and winter months in the US. Variables that affect the rate of positivity observed in respiratory testing include: the efficiency and timing of specimen collection, handling and transport of the specimen, the time of year, age of the patient, and local disease prevalence. During the 2006-2007 U.S. respiratory season, the combined prevalence of Influenza A and Influenza B was 13.2%⁸ and in 2005-2006 the combined prevalence was 12.1%⁹. The prevalence of RSV during the 2005-2006 season was 16.2%¹⁰. In the 2007 ProFlu+ multicenter clinical study (samples collected between February and April), the prevalence as observed with culture of Influenza A was 15.8%, Influenza B was 5.4% and RSV was 4.2%. As influenza and RSV seasons overlap, dual positive infections can occur. During this study, culture and the ProFlu+ Assay each detected one Influenza A and RSV dual-positive (although not the same sample) and the ProFlu+ Assay detected one Influenza A and Influenza B dual-positive out of the 891 total samples included in the study. Because the incidence of a triple infection of Influenza A, Influenza B, and RSV is low, it is recommended that the samples undergo repeat testing if nucleic acids from all three analytes are detected in a single sample.
The performance of the modified ProFlu+ assay has been demonstrated using a subset of archived prospectively collected clinical samples from 2008 - 2009 influenza seasons. They were selected to include 20 Influenza A 2009 H1N1, 20 Influenza A H1, 20 Influenza A H3, 10 Influenza B, 10 RSV positive and 121 negative samples.
## N. Instrument Name:
Roche MagNA Pure LC System with software version 3.0.11 or bioMérieux NucliSENS easyMAG System with Software version 1.0.1 or 2.0. Cepheid SmartCycler II Real Time Instrument with Dx software version 1.7b or 3.0 a/b.
## O. System Descriptions:
### 1. Modes of Operation:
The Roche MagNA Pure LC is an automated nucleic acid isolation and purification system based upon binding of nucleic acids to glass particles and has the capability to process a total of 32 reactions within one run. Nucleic acid is purified in multiple plastic reaction tips and cartridges by several steps that include cell lysis and binding of nucleic acid to magnetic glass particles, wash steps, and a heated elution to unbind the nucleic acid from the glass particles.
The bioMérieux NucliSens easyMAG is an automated nucleic acid isolation and purification system that is based upon the same silica extraction technology as the MagNA Pure. The easyMAG is capable of processing a total of 24 reactions with variable sample types, sample volumes, and elution volumes within a single run.
15
{16}
Nucleic acid is purified within a single cartridge by several steps that include lysis and binding of nucleic acid to high affinity magnetic silica beads, a series of wash steps and heated elution of purified nucleic acid from the silica beads.
The Cepheid Smart Cycler II Real Time instrument with Dx software version 1.7b, 3.0a/b is used to perform polymerase chain reaction (PCR) amplification and detection of nucleic acid. Other nucleic acid amplification tests that use the Smart Cycler II instrument have previously received 510(k) clearance: these tests include Prodesse's ProFAST+ Assay (K101855), ProParaflu+ Assay (K091053), ProGastro Cd Assay (K090239), ProhMPV+ Assay (K082688) and others. The Cepheid SmartCycler instrument is an integrated nucleic acid amplification and detection instrument system based on Cepheid's proprietary microprocessor-controlled I-CORE module. For purified DNA samples, the SmartCycler instrument enables PCR for the amplification of DNA and hybridization of fluorogenic target-specific probes for the detection of the amplified cDNA.
2. Software:
FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:
Yes ☐ X ☐ or No ☐
3. Specimen Identification:
User manually enters Patient ID/Sample ID.
4. Specimen Sampling and Handling:
Not applicable
5. Calibration:
Not applicable
6. Quality Control:
Positive Control (PC): The ProFlu+ Assay Control kit contains four RNA controls that consist of non-infectious in vitro transcribed RNA of specific Influenza A, Influenza B, RSV A or RSV B viral sequences. The PC does not go through nucleic acid isolation and purification, but is included with each RT-PCR run. The PC in conjunction with the IC is used to test for procedural errors in order to verify reagent and system performance.
Internal Control (IC): The IC is a non-infectious RNA transcript that is distinguished from the viral targets of the assay by means of a unique IC primer and probe set present in the ProFlu+ Assay. The IC is incorporated into every
16
{17}
sample and is carried through all steps of the procedure from nucleic acid isolation and purification through amplification to monitor for inhibitors present in the specimen or reaction tube. The IC also serves as a general process control ensuring that each step of the procedure was performed correctly, assay and instrument parameters were set correctly, and that general reagents were working.
Negative Control (NC): A Negative Control is not provided with the kit, but is required and described in the ProFlu+ Assay Instructions for Use. Viral transport medium spiked with the IC is to be used as the Negative Control and is processed starting from nucleic acid isolation. The Negative Control serves to monitor for contamination.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In the "Performance Characteristics" Section above:
Not applicable
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
17
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.