Elecsys HSV-2 IgG (08948887160)

K220924 · Roche Diagnostics · MYF · Oct 12, 2022 · Microbiology

Device Facts

Record IDK220924
Device NameElecsys HSV-2 IgG (08948887160)
ApplicantRoche Diagnostics
Product CodeMYF · Microbiology
Decision DateOct 12, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3305
Device ClassClass 2

Intended Use

lmmunoassay for the in vitro qualitative determination of IgG class antibodies to HSV-2 in human serum and lithium-heparin plasma, K2-EDTA plasma, and K3-EDTA plasma. The test is intended for sexually active individuals and expectant mothers as an aid in the presumptive diagnosis of HSV-2 infection. The test results may not determine the state of active lesions or associated disease manifestations, particularly for primary infection. The predictive value of positive or negative results depends on the population's prevalence and the pretest likelihood of HSV-2. The electrochemiluminescence immunoassay "ECLIA" is intended for use on cobas e immunoassay analyzers. This test is not FDA-cleared for screening blood or plasma donors. The performance of this assay has not been established for use in a pediatric population, neonates, or immunocompromised patients or for use at point-of-care facilities.

Device Story

In vitro diagnostic immunoassay for qualitative detection of IgG antibodies to HSV-2 in human serum and plasma; utilizes sandwich test principle with biotinylated recombinant HSV-2 antigens and ruthenium-labeled recombinant HSV-2 antigens. Operated on cobas e immunoassay analyzers in clinical laboratory settings. Device incorporates biotin-binding agent to deplete free biotin and streptavidin interference-reducing agent to improve assay robustness. Analyzers measure electrochemiluminescence (ECLIA) signal; software compares sample signal against calibrated cutoff value to determine qualitative result. Results aid clinicians in presumptive diagnosis of HSV-2 infection; results do not indicate active lesion status or disease manifestation, particularly in primary infection. Benefits include improved biotin tolerance (up to 1200 ng/mL) and reduced streptavidin interference compared to previous version.

Clinical Evidence

No clinical data provided; substantial equivalence supported by non-clinical bench testing demonstrating improved biotin and streptavidin interference tolerance.

Technological Characteristics

Sandwich immunoassay using biotinylated and ruthenium-labeled recombinant HSV-2 antigens. Employs electrochemiluminescence (ECLIA) detection. Reagents include biotin-binding agent and streptavidin interference-reducing agent. Designed for use on cobas e immunoassay analyzers. No changes to core sensing principle compared to predicate.

Indications for Use

Indicated for sexually active individuals and expectant mothers as an aid in the presumptive diagnosis of HSV-2 infection. Not for screening blood/plasma donors. Performance not established for pediatric, neonatal, or immunocompromised populations, or point-of-care use.

Regulatory Classification

Identification

Herpes simplex virus serological assays are devices that consist of antigens and antisera used in various serological tests to identify antibodies to herpes simplex virus in serum. Additionally, some of the assays consist of herpes simplex virus antisera conjugated with a fluorescent dye (immunofluorescent assays) used to identify herpes simplex virus directly from clinical specimens or tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by herpes simplex viruses and provides epidemiological information on these diseases. Herpes simplex viral infections range from common and mild lesions of the skin and mucous membranes to a severe form of encephalitis (inflammation of the brain). Neonatal herpes virus infections range from a mild infection to a severe generalized disease with a fatal outcome.

Special Controls

*Classification.* Class II (special controls). The device is classified as class II (special controls). The special control for the device is FDA's revised guidance document entitled “Class II Special Controls Guidance Document: Herpes Simplex Virus Types 1 and 2 Serological Assays.” For availability of the guidance revised document, see § 866.1(e).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below that. October 12, 2022 Roche Diagnostics Cynthia Aker Regulatory Affairs Principal 9115 Hague Road PO Box 50416 Indianapolis, Indiana 46250 Re: K220924 Trade/Device Name: Elecsys HSV-2 IgG (08948887160) Regulation Number: 21 CFR 866.3305 Regulation Name: Herpes simplex virus serological assays Regulatory Class: Class II Product Code: MYF Dated: March 29, 2022 Received: March 31, 2022 Dear Cynthia Aker: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Maria Garcia, Ph.D. Assistant Director Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use Submission Number (if known) Device Name Elecsys HSV-2 IgG (08948887160) Indications for Use (Describe) lmmunoassay for the in vitro qualitative determination of IgG class antibodies to HSV-2 in human serum and lithium-heparin plasma, K2-EDTA plasma, and K3-EDTA plasma. The test is intended for sexually active individuals and expectant mothers as an aid in the presumptive diagnosis of HSV-2 infection. The test results may not determine the state of active lesions or associated disease manifestations, particularly for primary infection. The predictive value of positive or negative results depends on the population's prevalence and the pretest likelihood of HSV-2. The electrochemiluminescence immunoassay "ECLIA" is intended for use on cobas e immunoassay analyzers. This test is not FDA-cleared for screening blood or plasma donors. The performance of this assay has not been established for use in a pediatric population, neonates, or immunocompromised patients or for use at point-of-care facilities. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary Prepared on: 2022-09-26 | Contact Details | 21 CFR 807.92(a)(1) | | |------------------------------------|------------------------------------------------------------------|--------------| | Applicant Name | Roche Diagnostics | | | Applicant Address | 9115 Hague Road PO Box 50416 Indianapolis IN 46250 United States | | | Applicant Contact Telephone | 1-463-336-2138 | | | Applicant Contact | Mrs. Cynthia Aker | | | Applicant Contact Email | cynthia.aker@roche.com | | | Device Name | 21 CFR 807.92(a)(2) | | | Device Trade Name | Elecsys HSV-2 IgG (08948887160) | | | Common Name | Herpes simplex virus serological assays | | | Classification Name | Enzyme Linked Immunosorbent Assay, Herpes Simplex Virus, Hsv-2 | | | Regulation Number | 866.3305 | | | Product Code | MYF | | | Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) | | | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | K121895 | Elecsys HSV-2 IgG | MYF | | Device Description Summary | 21 CFR 807.92(a)(4) | | The Elecsys HSV-2 IgG immunoassay makes use of a sandwich test principle using biotinylated recombinant HSV-2-specific antigens and HSV-2-specific recombinant antigens labeled with a ruthenium complex. The Elecsys HSV-2 IgG immunoassay is intended for the The and HSI qualitative determination of lgG class antibodies to HSV-2 in human serum and in the presumptive diagnosis of HSV-2 infection. It is intended for use on the cobas e immunoassay analyzers. Results are determined automatically by the software by comparing the electrochemiluminescence signal obtained from the reaction product of the sample with the signal of the cutoff value previously obtained by calibration. Intended Use/Indications for Use 21 CFR 807.92(a)(5) {4}------------------------------------------------ lmmunoassay for the in vitro qualitative determination of IgG class antibodies to HSV-2 in human serum and lithium-heparin plasma, K2-EDTA plasma, and K3-EDTA plasma. The test is individuals and expectant mothers as an aid in the presumptive diagnosis of HSV-2 infection. The test results may not determine the state of active lesions or associated disease manifestations, particularly for primary infection. The predictive or negative results depends on the population's prevalence and the pretest likelihood of HSV-2. The electrochemiluminescence immunoassay "ECLIA" is intended for use on cobas e immunoassay analyzers. This test is not FDA-cleared for screening blood or plasma donors. The performance of this assay has not been established for use in a pediation, neonates, or immunocompromised patients or for use at point-of-care facilities. ## Indications for Use Comparison Elecsys HSV-2 lgG (updated assay Mat. No. 0894887160) is substantially equivalent to Elecsys HSV-2 lgG, deared under K121895. The intended use of Elecsys HSV-2 lgG was updated to remove analyzers that are no longer supported for use with Roche assays. The intended use was also updated to add the following sentence: "The test results may not determine the state of active lesions or associated disease manifestations, particularly for primary infection." The indications for use of the Elecsys HSV-2 IgG did not change from the predicate device. # Technological Comparison Roche Diagnostics has updated the current Elecsys HSV-2 lgG assay in order to improve the biotin tolerance from < 70 ng/mL to ≤ 1200 ng/mL and to reduce streptavidin interference. A technical solution was implemented by adding an of the reagents, which allows depletion of biotin samples by binding free biotin, and by adding a streptavidin interference reducing agent to enhance the streptavidin tolerance. No other technological characteristics were changed. The information submitted in this Premarket Notification supports a substantial equivalent decision. Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) 21 CFR 807.92(a)(5) # 21 CFR 807.92(a)(6)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...