PREMIER PLATINUM HPSA PLUS, MODELS 601396, 601480

K053335 · Meridian Bioscience, Inc. · LYR · Mar 10, 2006 · Microbiology

Device Facts

Record IDK053335
Device NamePREMIER PLATINUM HPSA PLUS, MODELS 601396, 601480
ApplicantMeridian Bioscience, Inc.
Product CodeLYR · Microbiology
Decision DateMar 10, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3110
Device ClassClass 1

Intended Use

The Premier Platinum HpSA PLUS enzyme immunoassay (EIA) is an in vitro qualitative procedure for the detection of Helicobacter pylori antigens in human stool. Test results are intended to aid in the diagnosis of H. pylori infection and to monitor response during and post-therapy in patients. Accepted medical practice recommends that testing by any current method, to confirm eradication, be done at least four weeks following completion of therapy.

Device Story

Premier Platinum HpSA PLUS is an in vitro diagnostic, microwell-based enzyme-linked immunoassay (EIA) for detecting H. pylori antigen in human stool. The device consists of microwells coated with capture antibodies, enzyme conjugate (detector antibodies), sample diluent, wash buffer, substrate solution, stop solution, and positive control. Used in clinical laboratories by technicians; the assay requires no calibrators. The operator adds stool samples to the microwells; the presence of H. pylori antigen is detected via a colorimetric reaction measured by a dual-wavelength spectrophotometer (450/630 nm). Results are interpreted against a fixed cutoff (OD ≥ 0.100). The output aids clinicians in diagnosing H. pylori infection and monitoring patient response to eradication therapy. By providing a qualitative assessment of bacterial presence, the device assists in determining the success of treatment regimens, potentially reducing unnecessary follow-up or guiding further clinical intervention.

Clinical Evidence

Method comparison study using 291 symptomatic patient samples. Compared subject device to predicate; 100% positive agreement (94/94) and 94.8% negative agreement (183/193). Discrepant analysis against CLO, histology, or UBT confirmed subject device results were accurate in cases where it differed from the predicate. Analytical performance included precision/reproducibility studies (100% reproducibility) and analytical specificity testing against numerous bacterial, yeast, and viral strains, showing no interference.

Technological Characteristics

Qualitative EIA kit; microwells coated with monoclonal anti-H. pylori antibodies. Peroxidase-conjugated monoclonal antibody detection. Energy source: spectrophotometer (optional). Dimensions/form factor: standard microwell plate format. Connectivity: standalone. Sterilization: N/A. Software: None (manual/spectrophotometric interpretation).

Indications for Use

Indicated for patients requiring diagnosis of H. pylori infection or monitoring of response during and post-therapy. Testing for eradication confirmation is indicated at least four weeks post-therapy.

Regulatory Classification

Identification

Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identify Campylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases. Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### 510(k) SUMMARY 510(k) number: K053335 Submitters name and address: Meridian Bioscience, Inc. 3471 River Hills Drive Cincinnati. OH 45244 513-271-3700 Contact: Susan Rolih Date summary prepared: January 30, 2006 Name of the device: Premier Platinum HpSA PLUS Enzyme Immunoassay for the detection of H. pylori antigen in human stool specimens Classification: LYR, CFR section 866.3110 #### Predicate device to which this device is being compared: Premier Platinum HpSA (Meridian Bioscience, Inc., Cincinnati, OH) (K980076, K983255) Device description: Premier Platinum HpSA PLUS is an in vitro diagnostic, microwell-based, enzymelinked immunoassay for the detection of Helicobacter pylori antigen in human stool. The assay is intended for use in clinical laboratories to test for bacterial colonization to aid diagnosis, or monitor a patient's response during therapy to eradicate infection. The assay consists of Microwells coated with specific antibodies (solid phase/capture antibodies), Enzyme Conjugate (detector antibodies), Sample Diluent, Premier 20X Wash Buffer I, Premier Substrate Solution I, Premier Stop Solution I and Positive Control. Sample Diluent also functions as the Negative Control reagent. No calibrators are needed to use this device. Intended use: The Premier Platinum HpSA PLUS enzyme immunoassay (EIA) is an in vitro qualitative procedure for the detection of Helicobacter pylori antigens in human stool. Test results are intended to aid in the diagnosis of H. pylori infection and to monitor response during and post-therapy in patients. Accepted medical practice recommends that testing by any current method, to confirm eradication, be done at least four weeks following completion of therapy. There is no change to the intended use of this device from its predicate. # Comparison charts (Premier Platinum HpSA PLUS vs Predicate Device): | Characteristics | Premier Platinum HpSA<br>PLUS | Premier Platinum HSA<br>(predicate) | |--------------------------------|--------------------------------------|--------------------------------------| | Device Type | | | | In vitro diagnostic device | Yes | Yes | | Control | Includes external control<br>reagent | Includes external control<br>reagent | | Calibrator | No | No | | Intended Use | | | | Detection of H. pylori antigen | Yes | Yes | | Screening test | Yes | Yes | | Diagnostic test | No | No | | Monitoring therapy | Yes | Yes | | Acceptable Sample | | | | Stool | Yes | Yes | {1}------------------------------------------------ | Laboratory Equivalence with (Predicate Device) | Premier Platinum HpSA<br>PLUS | Predicate | |------------------------------------------------|-------------------------------|-------------------------------| | Combined Totals | | | | Agreement, positive tests | 100% | N/A | | Agreeement, negative tests | 94.8% | N/A | | Agreement, overall | 96.5% | N/A | | Performance characteristics | | | | Precision/Reproducibility (intra-assay) | 97% | 100% | | Linearity/reportable range | N/A | N/A | | Analytical limit of detection/sensitivity | ≥ 4.67 ng in stool | ≥ 184 ng in stool | | Assay cutoff | 0.100 at OD450/630 | 0.120 at OD450/630 | | Indeterminant range | None | 0.100 - 0.120 at<br>OD450/630 | # Interpretation of test results The results of bench tests were read using a standard laboratory dual wave length spectrophotometer. Results were interpreted according to the following scale: Spectrophotometric dual wavelength (450/630 nm) Negative < 0.100 Positive ≥ 0.100 Results occurring in the 0.100 to 0.120 OD range were tracked to determine if a significant number of results were obtained such that would justify the inclusion of an equivocal range. The alsence of equivocal results in the studies showed that this criteria was not necessary. # Analytical sensitivity – limit of detection Study design: Serial dilutions of purified H. pylori flagellar antigen and a H. pylori bacterial strain (ATCC 43504) were prepared in stool or Sample Diluent and used to determine the lowest concentration of antigen that would still yield a definitive positive result (Assoco ≥ 0.100 on Premier Platinum HpSA PLUS). Final concentrations were calculated from the data points using linear regression analysis. Conclusions to the study: The analytical limit for H. pylori flagellar antigen is 4.67 ng/mL in stool and 0.69 ng/mL in sample diluent. The limit for H. pylori bacterial strain is 1.0 % organisms/mL in stool and 4.0 1 10111 organisms/mL in Sample Diluent. The limits are lower than those reported for the Predicate Device, ### Linearitv Linearity does not apply to the endpoint produced by this device. #### Interfering substance testing # Drugs, Nonmicrobial Substances Study design: Selected drugs and other nonmicrobial substances that might be present in stool specimens were added to five known H. pylori-positive and five known negative samples. The final concentrations of the additives per 500 ul. of human stool are as follows: TUMS - 10 mg, Pepto Bismoluna of the Tagamet - 1 mg, Prilosec OTC - 1 mg, barium suifate - 10 mg, whole blood - 100 mucin - 6.7 mg, human hemoglobin (to make dark stool) – 15 mg, steric + palmitic acids (to make fatty stool) – 7.9 mg. The spiried samples were tested in triplicate and in parallel with an unspiked control. Acceptance criteria required the values within replicates be similar to each other and to the value obtained with the unspiked specmers. None of the potentially interfering substances had a significant effect on negative test results. Notes correlated closely with unspiked samples. Conclusions to the study. Trues or nonmicrobial substances that might be present as co-contaminants of stool samples do not adversely affect results obtained with Premier Platinum HpSA PLUS. These data correlate with data published for the predicate device. # Microbial/Viral organisms (potentially cross-reactive or interfering species) Study design: The bacteria, yeast and viruses selected were that might be expected to be present in human stool samples either as part of normal flora or from a disease state. The final concentration of bracteria {2}------------------------------------------------ or yeast in each sample was ≥ #4 McFarland Standard (1.2 X 10 organisms/mL). The final concentration of viruses in each sample was not determined. Unspiked samples were tested in parallel to provide a reference against which the reactions with spiked samples could be compared. Samples were tested in triplicate. Acceptance criteria required that the values within replicates be similar to each other and to the value obtained with the unspiked specimen. See data in Tables below. None of the potential co-contaminants adverselv affected the final positive or negative test results. Conclusions to the study: Microbial and viral organisms that might be present as co-contaminants of stool samples do not adversely affect results obtained with Premier Platinum HpSA PLUS. These data correlate with data published for the predicate device. | Sample | 42 | | | | 32 | | | | |----------|-------|-------|-------|-------|-------|-------|-------|-------| | | Run 1 | Run 2 | Run 3 | Avg | Run 1 | Run 2 | Run 3 | Avg | | No Spike | 0.465 | 0.409 | 0.415 | 0.430 | 1.472 | 1.316 | 1.099 | 1.296 | | No Spike | 0.536 | 0.553 | 0.494 | 0.528 | 1.861 | 1.979 | 1.900 | 1.913 | | No Spike | 0.417 | 0.468 | 0.403 | 0.429 | 1.464 | 1.550 | 1.706 | 1.573 | | No Spike | 0.445 | 0.460 | 0.478 | 0.472 | 1.308 | 1.345 | 1.562 | 1.405 | | No Spike | NT | NT | NT | NT | 1.412 | 1.468 | 1.330 | 1.403 | #### Effects of various microbial organisms on positive test results. | | | NT = not tested | |--|--|-----------------| |--|--|-----------------| | | 42 | | | | 32 | | | | |-------------------------------|-------|-------|-------|-------|-------|-------|-------|-------| | Sample | Run 1 | Run 2 | Run 3 | Avg | Run 1 | Run 2 | Run 3 | Avg | | Adenovirus | 0.460 | 0.438 | 0.412 | 0.437 | 1.517 | 1.511 | 1.340 | 1.456 | | Aeromonas hydrophila | 0.508 | 0.557 | 0.469 | 0.511 | 1.837 | 1.766 | 1.860 | 1.821 | | Borrelia burgdorferi | 0.480 | 0.306 | 0.293 | 0.360 | 1.239 | 1.307 | 1.230 | 1.259 | | Campylobacter lari | 0.421 | 0.522 | 0.410 | 0.451 | 1.721 | 1.733 | 1.848 | 1.767 | | Campylobacter fetus | NT | NT | NT | NT | 1.505 | 1.633 | 1.623 | 1.587 | | Campylobacter jejuni | NT | NT | NT | NT | 1.544 | 1.378 | 1.489 | 1.470 | | Campylobacter jejuni 2 | NT | NT | NT | NT | 1.520 | 1.526 | 1.396 | 1.481 | | Campylobacter jejuni solution | NT | NT | NT | NT | 1.402 | 1.661 | 1.612 | 1.558 | | Campylobacter lari | NT | NT | NT | NT | 1.594 | 1.512 | 1.565 | 1.557 | | Candida albicans | 0.452 | 0.436 | 0.423 | 0.437 | 1.682 | 1.642 | 1.839 | 1.721 | | Citrobacter freundii | 0.563 | 0.557 | 0.527 | 0.549 | 2.264 | 2.264 | 2.181 | 2.236 | | Clostridium difficile | 0.543 | 0.600 | 0.516 | 0.553 | 1.878 | 1.996 | 2.047 | 1.974 | | Clostridium perfringens | 0.514 | 0.496 | 0.516 | 0.509 | 1.799 | 1.833 | 1.922 | 1.851 | | Enterobacter cloacae | 0.428 | 0.572 | 0.553 | 0.518 | 2.373 | 2.374 | 2.469 | 2.405 | | Enterococcus faecalis | 0.506 | 0.517 | 0.555 | 0.526 | 2.033 | 1.963 | 2.134 | 2.043 | | Escherichia coli 0157:H7 | 0.569 | 0.534 | 0.527 | 0.543 | 2.177 | 2.228 | 2.243 | 2.216 | | Escherichia coli 8739 | 0.500 | 0.263 | 0.499 | 0.421 | 1.882 | 1.890 | 1.952 | 1.908 | | Escherichia coli 9637 | 0.493 | 0.536 | 0.498 | 0.509 | 1.850 | 1.782 | 1.864 | 1.832 | | Escherichia fergusonii | 0.533 | 0.528 | 0.517 | 0.526 | 2.237 | 2.167 | 2.124 | 2.176 | | Escherichia hermannii | 0.437 | 0.440 | 0.378 | 0.418 | 1.469 | 1.643 | 1.714 | 1.609 | | Escherichia hermannii EMDI-64 | 0.488 | 0.436 | 0.443 | 0.456 | 1.838 | 1.530 | 1.748 | 1.705 | | Helicobacter pylori | OUT* | OUT* | OUT* | OUT* | OUT | OUT | OUT | OUT | | Klebsiella pneumoniae | 0.533 | 0.534 | 0.519 | 0.529 | 2.051 | 2.269 | 2.153 | 2.158 | | Lactobacillus lactis | 0.474 | 0.456 | 0.420 | 0.450 | 1.486 | 1.642 | 1.684 | 1.604 | | Listeria monocytogenes | 0.441 | 0.427 | 0.391 | 0.420 | 1.672 | 1.623 | 1.582 | 1.626 | | Peptostreptococcus anaerobius | 0.422 | 0.341 | 0.381 | 0.381 | 1.559 | 1.481 | 1.516 | 1.519 | | Proteus vulgaris | 0.509 | 0.487 | 0.476 | 0.491 | 1.880 | 1.950 | 1.962 | 1.931 | {3}------------------------------------------------ | Sample | 42 | | | | 32 | | | | |-------------------------------------------------------|-------|-------|-------|-------|-------|-------|-------|-------| | | Run 1 | Run 2 | Run 3 | Avg | Run 1 | Run 2 | Run 3 | Avg | | Pseudomonas aeruginosa | 0.217 | 0.560 | 0.536 | 0.438 | 2.208 | 2.272 | 2.158 | 2.213 | | Pseudomonas fluorescens | 0.448 | 0.427 | 0.460 | 0.445 | 1.620 | 1.785 | 1.749 | 1.718 | | Rotavirus | 0.384 | 0.407 | 0.445 | 0.412 | 1.356 | 1.361 | 1.387 | 1.368 | | Salmonella enterica serovar Hilversum | 0.441 | 0.433 | 0.429 | 0.434 | 1.290 | 1.661 | 1.617 | 1.523 | | Salmonella enterica subsp. Enterica serovar Hilversum | 0.519 | 0.537 | 0.534 | 0.530 | 2.232 | 2.281 | 2.336 | 2.283 | | Sample | 42 | | | | 32 | | | | |-------------------------------------------------------|-------|-------|-------|-------|-------|-------|-------|-------| | | Run 1 | Run 2 | Run 3 | Avg | Run 1 | Run 2 | Run 3 | Avg | | Salmonella enterica subsp. Enterica serovar Minnesota | 0.565 | 0.505 | 0.600 | 0.557 | 2.512 | 2.529 | 2.296 | 2.446 | | Salmonella Group B | 0.413 | 0.431 | 0.388 | 0.411 | 1.632 | 1.597 | 1.784 | 1.671 | | Salmonella typhimurium | 0.549 | 0.588 | 0.569 | 0.569 | 2.336 | 2.180 | 2.215 | 2.244 | | Serratia liquefaciens | 0.445 | 0.429 | 0.457 | 0.444 | 1.508 | 1.610 | 1.653 | 1.590 | | Serratia liquefaciens | 0.482 | 0.541 | 0.503 | 0.509 | 2.180 | 1.990 | 2.125 | 2.098 | | Serratia marcescens | 0.393 | 0.470 | 0.329 | 0.397 | 1.019 | 1.411 | 1.617 | 1.349 | | Shigella boydii | 0.540 | 0.552 | 0.520 | 0.537 | 2.171 | 2.091 | 2.235 | 2.166 | | Shigella flexneri | 0.365 | 0.549 | 0.562 | 0.492 | 2.321 | 2.317 | 2.337 | 2.325 | | Shigella dysenteriae | 0.536 | 0.534 | 0.535 | 0.535 | 2.294 | 2.241 | 2.275 | 2.270 | | Shigella sonnei | 0.420 | 0.465 | 0.325 | 0.403 | 1.598 | 1.606 | 1.564 | 1.589 | | Staphylococcus aureus | 0.618 | 0.526 | 0.527 | 0.557 | 2.243 | 2.255 | 2.189 | 2.229 | | Staphylococcus aureus (Cowans 1) | 0.505 | 0.461 | 0.477 | 0.481 | 1.919 | 1.805 | 1.808 | 1.844 | | Staphylococcus epidermidis | 0.587 | 0.518 | 0.570 | 0.558 | 2.294 | 2.200 | 2.121 | 2.205 | | Streptococcus faecalis | 0.501 | 0.513 | 0.469 | 0.494 | 1.963 | 1.965 | 1.954 | 1.961 | | Yersinia enterocolitica | 0.500 | 0.543 | 0.504 | 0.516 | 1.985 | 1.881 | 1.490 | 1.785 | | Yersinia enterocolitica | 0.503 | 0.490 | 0.491 | 0.495 | 1.974 | 1.762 | 1.992 | 1.909 | *Note: "> 3.000" signifies the signal exceeded the high limit of the plate reader used in the study was Auses 2.999. | Sample | 2-128 | | | | |-------------------------------|-------|-------|-------|-------| | | Run 1 | Run 2 | Run 3 | Avg | | Unspiked Sample | 0.386 | 0.382 | 0.323 | 0.364 | | Campylobacter lari | 0.399 | 0.398 | 0.390 | 0.396 | | Campylobacter jejuni | 0.468 | 0.442 | 0.456 | 0.455 | | Campylobacter jejuni solution | 0.432 | 0.427 | 0.409 | 0.423 | | Campylobacter jejuni 2 | 0.414 | 0.410 | 0.393 | 0.406 | | Campylobacter fetus | 0.398 | 0.427 | 0.383 | 0.403 | {4}------------------------------------------------ | Sample | 2-42 | | | | |----------|-------|-------|-------|-------| | | Run 1 | Run 2 | Run 3 | Avg | | No Spike | 0.011 | 0.009 | 0.014 | 0.011 | | No Spike | 0.006 | 0.005 | 0.008 | 0.006 | | No Spike | 0.006 | 0.007 | 0.014 | 0.009 | | No Spike | 0.005 | 0.007 | 0.007 | 0.006 | | No Spike | 0.006 | 0.008 | 0.006 | 0.007 | | | 2-42 | | | | |-------------------------------------------------------|---------|---------|---------|---------| | Sample | Run 1 | Run 2 | Run 3 | Avg | | Adenovirus | 0.007 | 0.007 | 0.006 | 0.007 | | Aeromonas hydrophila | 0.004 | 0.006 | 0.006 | 0.005 | | Borrelia bergdorferi | 0.012 | 0.006 | 0.007 | 0.008 | | Campylobacter lari | 0.008 | 0.007 | 0.004 | 0.006 | | Campylobacter fetus | 0.007 | 0.008 | 0.004 | 0.006 | | Campylobacter jejuni | 0.016 | 0.006 | 0.008 | 0.010 | | Campylobacter jejuni 2 | 0.000 | 0.006 | 0.007 | 0.004 | | Campylobacter jejuni solution | 0.004 | 0.009 | 0.007 | 0.007 | | Campylobacter lari | 0.005 | 0.007 | 0.006 | 0.006 | | Candida albicans | 0.006 | 0.008 | 0.008 | 0.007 | | Citrobacter freundii | 0.005 | 0.016 | 0.006 | 0.009 | | Clostridium difficile | 0.010 | 0.011 | 0.009 | 0.010 | | Clostridium perfringens | 0.010 | 0.008 | 0.008 | 0.009 | | Enterobacter cloacae | 0.003 | 0.003 | 0.004 | 0.003 | | Enterococcus faecalis | 0.002 | 0.002 | 0.007 | 0.004 | | Escherichia coli 0157:H7 | 0.005 | 0.034 | 0.064 | 0.034 | | Escherichia coli 8739 | 0.005 | 0.005 | 0.007 | 0.006 | | Escherichia coli 9637 | 0.003 | 0.007 | 0.006 | 0.005 | | Escherichia fergusonii | 0.004 | 0.002 | 0.007 | 0.004 | | Escherichia hermannii | 0.008 | 0.007 | 0.008 | 0.008 | | Escherichia hermannii EMDI-64 | 0.006 | 0.008 | 0.007 | 0.007 | | Helicobacter pylori | > 3.000 | > 3.000 | > 3.000 | > 3.000 | | Klebsiella pneumoniae | 0.004 | 0.004 | 0.007 | 0.005 | | Lactobacillus lactis | 0.007 | 0.007 | 0.017 | 0.010 | | Listeria monocytogenes | 0.008 | 0.005 | 0.009 | 0.007 | | Peptostreptococcus anaerobius | 0.008 | 0.008 | 0.009 | 0.008 | | Proteus vulgaris | 0.004 | 0.000 | 0.007 | 0.004 | | Pseudomonas aeruginosa | 0.011 | 0.026 | 0.021 | 0.019 | | Pseudomonas fluorescens | 0.006 | 0.007 | 0.007 | 0.007 | | Rotavirus | 0.006 | 0.005 | 0.005 | 0.005 | | Salmonella enterica serovar Hilversum | 0.008 | 0.007 | 0.009 | 0.008 | | Salmonella enterica subsp. Enterica serovar Hilversum | 0.002 | 0.010 | 0.045 | 0.019 | | Salmonella enterica subsp. Enterica serovar Minnesota | 0.005 | 0.008 | 0.005 | 0.006 | | Salmonella Group B | 0.007 | 0.006 | 0.007 | 0.007 | {5}------------------------------------------------ | Sample | Run 1 | Run 2 | Run 3 | Avg | |----------------------------------|-------|-------|-------|-------| | Salmonella typhimurium | 0.003 | 0.007 | 0.005 | 0.005 | | Serratia liquefaciens | 0.007 | 0.007 | 0.007 | 0.007 | | Serratia liquefaciens | 0.000 | 0.003 | 0.004 | 0.002 | | Serratia marcescens | 0.007 | 0.008 | 0.008 | 0.008 | | Shigella boydii | 0.006 | 0.000 | 0.003 | 0.003 | | Shigella flexneri | 0.004 | 0.027 | 0.004 | 0.012 | | Shigella dysenteriae | 0.006 | 0.005 | 0.050 | 0.020 | | Shigella sonnei | 0.007 | 0.002 | 0.009 | 0.006 | | Staphylococcus aureus | 0.008 | 0.004 | 0.008 | 0.007 | | Staphylococcus aureus (Cowans 1) | 0.007 | 0.009 | 0.007 | 0.008 | | Staphylococcus epidermidis | 0.008 | 0.003 | 0.007 | 0.006 | | Streptococcus faecalis | 0.006 | 0.005 | 0.007 | 0.006 | | Yersinia enterocolitica | 0.005 | 0.004 | 0.008 | 0.006 | | Yersinia enterocolitica | 0.004 | 0.003 | 0.006 | 0.004 | ### Performance Evaluation Data Summarized Comparison of Premier Platinum HpSA PLUS to Premier Platinum HpSA: Tests with 291 samples from symptomatic patients collected either prior to or following treatment were used to demonstrate that Premier Platinum HpSA PLUS performed similarly to Premier Platinum HpSA. Thirty three of these samples were originally evaluated in an earlier trial to demonstrate the effectiveness of Premier Platinum HpSA. Test performance including 95% confidence intervals is detailed in the following table. | | | PP HpSA | | | |--------------|-----------------|-----------------|---------------|--| | PP HpSA PLUS | Positive | Negative | Indeterminate | | | Positive | 94 | 10 | 3 | | | Negative | 0 | 183 | 1 | | | Correlation | 277/287 (96.5%) | 93.7% - 98.3% | | | | | 95% CI | | | | | Agreement | Positive test | 94/94 = 100% | | | | | Negative test | 183/193 = 94.8% | | | Eight of the ten samples that were positive by Premier Platinum HpSA PLUS, but negative by Premier Platinum HpSA, were positive by CLO, histology or UBT testing. The three samples that were positive by Premier Platinum HpSA PLUS but indeterminate by Premier Platinum HpSA were positive by CLO, histology or UBT testing. The one sample that was negative by Premier Platinum HpSA PLUS but indeterminate by Premier Platinum HpSA was negative by CLO, histology or UBT testing. #### Analysis of samples producing discordant results Samples producing discordant results between Premier Platinum HpSA PLUS and the predicate were evaluated against test data from other conventional tests such as CLO, Histology, or UBT to determine the trueness of the results. The results of that evaluation are provided shown in the Table below. {6}------------------------------------------------ | Sample<br>Number | PP HpSA PLUS Results | CLO/Histology/UBT<br>Results | PP HpSA Results<br>(Predicate) | Interpretation using<br>CLO/Hist/UBT | |------------------|----------------------|------------------------------|--------------------------------|--------------------------------------| | UC82 | Positive | Negative | Negative | FP--PPHpSAPLUS | | 2 | Positive | Positive | Indeterminate | TP--PPHpSAPLUS | | 3-44 | Negative | Negative | Indeterminate | TN--PPHpSAPLUS | | U082 | Positive | Positive | Negative | TP--PPHpSAPLUS | | U004 | Positive | Positive | Negative | TP--PPHpSAPLUS | | U120 | Positive | Positive | Negative | TP--PPHpSAPLUS | | U159 | Positive | Positive | Negative | TP--PPHpSAPLUS | | U056 | Positive | Positive | Negative | TP--PPHpSAPLUS | | U137 | Positive | Positive | Indeterminate | TP--PPHpSAPLUS | | U161 | Positive | Positive | Negative | TP--PPHpSAPLUS | | P026 | Positive | Positive | Negative | TP--PPHpSAPLUS | | P040 | Positive | Negative | Negative | FP--PPHpSAPLUS | | P172 | Positive | Positive | Indeterminate | TP--PPHpSAPLUS | | P173 | Positive | Positive | Negative | TP--PPHpSAPLUS | Legend: FP = false positive, TP = true positive, TN = true negative #### Performance Comparison Table | Performance Characteristics (rounded) in Direct Comparison to | Premier Platinum HpSA | Premier Platinum HpSA | |----------------------------------------------------------------|-----------------------|-----------------------| | Clinical Status or Condition | PLUS | (Predicate) | | Estimated Clinical Sensitivity | N/A | 96.1% | | Estimated Clinical Specificity | N/A | 95.7% | | Predictive Value of a Positive Test | N/A | 96.1% | | Predictive Value of a Negative Test | N/A | 95.7% | | Laboratory Equivalence with (Predicate Device) Combined Totals | | | | Agreement, positive tests | 100% | N/A | | Agreement, negative tests | 94.8% | N/A | | Correlation | 96.5% | 95.9% | | Performance characteristics | | | | Precision/Reproducibility | 100% | 100% | | Linearity/reportable range | N/A | N/A | | Limit of detection | ≥ 4.67 ng in stool | ≥ 184 ng in stool | | Assay cutoff | 0.100 at OD450/630 | 0.120 at OD450/630 | #### Therapeutic Monitoring Study design: A panel of frozen, archival specimens from four patients who were monitored during eradication therapy and tested using the predicate Premier Platinum HpSA (K980076 and K983255) were assessed using the Premier Platinum HpSA PLUS assay. One of the panels represented a low positive state with the predicate at the beginning of eradication therapy (See Figure 1.) The remaining three panels represented strongly positive states. (See Figures 2-4.) That data obtained with Premier Platinum HpSA PLUS was compared to that originally obtained with the predicate. In the case of strongly positive samples, the eradication curves for the two tests are substantively the same. The eradication curve for Premier Platinum HpSA PLUS differs from that of the predicate with low positive samples at the beginning of therapy since it produces stronger test results. However, by week four following treatment, the curves are identical. Conclusions to the study: Premier Platinum HpSA performs similarly to the predicate when used to monitor the effectiveness of eradication therapy. {7}------------------------------------------------ Image /page/7/Figure/2 description: Figure 1 shows a graph of the absorbance of two different substances, Predicate 450/630 and PLUS 450/630, for Patient 41. The x-axis represents the specimen number, ranging from 1 to 7, while the y-axis represents the absorbance. The absorbance of Predicate 450/630 is relatively low, ranging from 0.02 to 0.27, while the absorbance of PLUS 450/630 is higher, ranging from 0.00 to 3.00. Image /page/7/Figure/3 description: The image contains the text "Figure 2.". The text is in a simple sans-serif font. The figure number is likely a reference to a larger document or publication. Image /page/7/Figure/4 description: This image shows a graph titled "Patient 51" with "Absorbance" on the y-axis and "Specimen" on the x-axis. There are two lines on the graph, one labeled "Predicate 450/630" and the other labeled "PLUS 450/630". The graph shows the absorbance values for each specimen, with the values for "Predicate 450/630" being 3.000, 1.945, 0.034, 0.026, and 0.025, and the values for "PLUS 450/630" being 3.000, 2.868, 0.041, 0.007, and 0.006. Figure 3. Image /page/7/Figure/6 description: This image is a graph for patient 71 showing absorbance versus specimen number. There are two lines on the graph, one for predicate 450/630 and one for PLUS 450/630. Both lines start at 3.000 for specimens 1 and 2, but then drop to around 0.030 for specimens 3-9. {8}------------------------------------------------ Image /page/8/Figure/2 description: The image shows a graph and table for Patient 87. The graph plots absorbance on the y-axis versus specimen number on the x-axis for two different predicates, 450/630 and PLUS 450/630. Both predicates start with an absorbance of 3.000 for specimens 1 and 2, but then drop to 0.030 and 0.009 respectively for specimen 3. The absorbance values for specimens 4 through 7 are all below 0.033. Figure 4. ### Reproducibility Assay precision, intra-assay variability and inter-assay variability were assessed with a reference panel prepared from high positive samples (n = 2), low low negative samples (n = 2), and low positive and high neqative specimens (n = 1 each). The latter were diluted to near the assay limit of sensitivity. Nine replicates each of the low positive and high negative samples were included in the panel to bring the total cohort to 22 reference specimens. Each reference specimen was coded to prevent its identification during testing. Each was evaluated twice per day for three consecutive days by three different laboratories. In accordance with the IFU, values of < 0.100 are interpreted as negative when results are read at A450/630. High negative samples (OD values just below 0.100) produced weakly positive results (OD values just above 0.100) in 42 out of 162 times tests. It is expected that high negative samples tested at the cut-off will produce weakly positive results 50% of the time. (See EP12-A, User protocol for evaluation of qualitative performance; approved guideline; NCCLS/CLSI, Vol. 22, no.14, 2002.) Low positive and low negative samples produced the correct results 100% of the time. Reproducibility was 100% with no intra-assay variability for samples prepared above or below the limit of analytical sensitivity. {9}------------------------------------------------ 51し、, notification Meridian Bioscience, Inc. _US Premier Platinum HpS, | esuits with reproducibility test par | | | | | | | | | | | | | | | | | | | | |--------------------------------------|--------|---------------|----------------|-------------|-----------------------|-------------|--------|---------------|-------------------|-------------|-------|-------|--------------|-------|-------------|-------------|-------|--------|-------------| | sults at 450/630 nr | | | | echnologist | | | | | | echnologist | | | | | | echnologist | | | | | Sample II | Sample | l, keçi L keç | | | r und 7 Áseo | | Day 3 | l unu l L fea | guna s Ligar Link | Day 2 | Day 2 | Day 3 | gung Esta g | t ung | grums Liked | Day 2 | Day 2 | l, ung | Day 3 Run 2 | | | Qual | | g unst L. feet | und i z keq | | Day 3 Run 1 | Run 2 | | | run 1 | run 2 | Run | | | | run 1 | run 2 | | | | | Resul | | | | | | | | | | | | | | | | | | | | HP # | 1.510 | 2.162 | 1.997 | 1.830 | 1.924 | 1.852 | 1 754 | 2.092 | .986 | 1.941 | 1.842 | .880 | 1.745 | 2.380 | 2.300 | 1.962 | 2.36 | 2.233 | 2.364 | | 2 HP #2 | 1.06 | 1.445 | 1.436 | 1.283 | | 1.217 | 1.244 | .537 | .406 | 1.302 | 1.307 | 293 | 1.154 | 1.858 | 1.743 | 1.400 | 1.698 | 1.689 | 1.614 | | Cut off LP # | 0.133 | 0.217 | 0.20€ | 0.176 | ତି ପ୍ରତିଶତ ସହ ପ୍ରତିଶତ | 0.188 | 0.162 | 0.274 | 0.234 | 0.192 | 0.203 | 0.186 | 0.175 | 0.292 | 0.290 | 0.294 | 0.289 | 0.290 | 0.301 | | Cut off LP # | 0.133 | 0.206 | 0.197 | 0.171 | 0.214 | 0.166 | 0.165 | 0.271 | 0.237 | 0.213 | 0.189 | 0.173 | 0.185 | 0.310 | 0.294 | 0.268 | 0.273 | 0.299 | 0.299 | | Cut off LP # | 0.133 | 0.179 | 0.195 | 0.195 | 0.207 | 0.176 | 0.179 | 0.197 | 0.234 | 0.199 | 0.185 | 0.158 | 0.178 | 0.318 | 0.311 | 0.305 | 0.298 | 0.299 | 0.290 | | Cut off LP # | 0.133 | 0.185 | 0.215 | 0.187 | 0.210 | 0.182 | 0.149 | 0.265 | 0.219 | 0.219 | 0.175 | 0.163 | 0.167 | 0.295 | 0.312 | 0.279 | 0.308 | 0.295 | 0.336 | | Cut off LP # | 0.133 | 0.180 | 0.211 | 0.179 | 0.210 | 0.176 | 0.172 | 0.243 | 0.185 | 0.214 | 0.164 | 0.149 | 0.167 | 0.286 | 0.275 | 0.288 | 0.318 | 0.296 | 0.316 | | Cut off LP # | 0.133 | 0.20 | 0.219 | 0.22 | 0.220 | 0.188 | 0.178 | 0.239 | 0.215 | 0.214 | 0.146 | 0.137 | 0.156 | 0.300 | 0.292 | 0.287 | 0.315 | 0.305 | 0.318 | | Cut off LP# | 0.133 | 0.207 | 0.187 | 0.206 | 0.203 | 0.189 | 0.131 | 0.217 | 0.220 | 0.203 | 0.140 | 0.133 | 0.158 | 0.293 | 0.29 | 0.30€ | 0.287 | 0.308 | 0.300 | | 0 Cut off LP # | 0.133 | 0.186 | 0.187 | 0.202 | 0.223 | 0.179 | 0.163 | 0.233 | 0.201 | 0.211 | 0.118 | 0.097 | 0.140 | 0.309 | 0.307 | 0.267 | 0.298 | 0.290 | 0.299 | | 1 Cut off LP # | 0.133 | 0.193 | 0.192 | 0.175 | 0.173 | 0.159 | 0.156 | 0.238 | 0.200 | 0.204 | 0.14( | 0.124 | 0.145 | 0.331 | 0.317 | 0.30 | 0.298 | 0.315 | 0.314 | | 2 Cut off HN # | 0.075 | 0.094 | 0.104 | 0.074 | 0.12 | 0.076 | 0.078 | 0.131 | 0.097 | 0.098 | 0.059 | 0.060 | 0.066 | 0.096 | 0.094 | 0.09€ | 0.094 | 0.095 | 0.090 | | 3 Cut off HN # | 0.075 | 0.10 | 0.103 | 0.069 | 0.105 | 0.074 | 0.074 | 0.152 | 0.112 | 0.083 | 0.090 | 0.097 | 0.105 | 0.08 | 0.093 | 0.095 | 0.093 | 0.085 | 0.094 | | 4 Cut off HN # | 0.075 | 0.073 | 0.104 | 0.088 | 0.092 | 0.082 | 0.07 * | 0.161 | 0.118 | 0.090 | 0.105 | 0.088 | 0.10 | 0.074 | 0.086 | 0.084 | 0.08 | 0.087 | 0.09 | | Cut off HN # | 0.075 | 0.058 | 0.109 | 0.096 | 0.079 | 0.080 | 0.090 | 0.142 | 0.115 | 0.094 | 0.105 | 0.083 | 0.099 | 0.096 | 0.08 | 0.096 | 0.087 | 0.089 | 0.084 | | Cut off HN # | 0.075 | 0.080 | 0.105 | 0.09 | 0.089 | 0.085 | 0.071 | 0.147 | 0.112 | 0.102 | 0.079 | 0.073 | 0.103 | 0.092 | 0.093 | 0.093 | 0.085 | 0.090 | 0.081 | | 7 Cut off HN # | 0.075 | 0.083 | 0.116 | 0.096 | 0.101 | 0.098 | 0.087 | 0.137 | 0.111 | 0.107 | 0.085 | 0.073 | 0.09 | 0.090 | 0.098 | 0.093 | 0.09 | 0.093 | 0.082 | | Cut off HN # | 0.075 | 0.088 | 0.113 | 0.089 | 0.113 | 0.088 | 0.097 | 0.128 | 0.101 | 0.113 | 0.064 | 0.068 | 0.083 | 0.082 | 0.08 | 0.094 | 0.080 | 0.089 | 0.08 | | 9 Cut off HN # | 0.075 | 0.098 | 0.118 | 0.095 | 0.089 | 0.088 | 0.098 | 0.140 | 0.105 | 0.118 | 0.060 | 0.058 | 0.083 | 0.085 | 0.082 | 0.096 | 0.085 | 0.093 | 0.078 | | 0 Cut off HN # | 0.075 | 0.090 | 0.103 | 0.083 | 0.112 | 0.092 | 0.091 | 0.126 | 0.104 | 0.109 | 0.046 | 0.053 | 0.069 | 0.085 | 0.086 | 0.09 | 0.093 | 0.092 | 0.088 | | 21 LN # | 0.005 | 0.005 | 0.00€ | 0.006 | 0.005 | 0.005 | 0.007 | 0.001 | 0.002 | 0.004 | 0.00 | 0.00 | 0.002 | 0.003 | 0.004 | 0.003 | 0.004 | 0.004 | 0.003 | | 2 LN #2 | 0.005 | 0.000 | 0.005 | 0.006 | 0.006 | 0.006 | 0.009 | 0.001 | 0.000 | 0.004 | 0.00 | 0.000 | 0.002 | 0.005 | 0.005 | 0.004 | 0.005 | 0.004 | 0.003 | | rage high negative va | | 0.085 | 0.109 | 0.14 | 0.108 | 0.087 | 0.088 | 0.086 | 0.100 | 0.102 | 0.077 | 0.093 | 0.088 | 0.085 | 0.084 | 0.073 | 0.089 | 0.090 | 0.085 | | rage low positive valu | | 0.195 | 0.20 | 0.242 | 0.216 | 0.304 | 0.299 | 0.190 | 0.210 | 0.208 | 0.162 | 0.288 | 0.298 | 0.178 | 0.162 | 0.14 | 0.163 | 0.300 | 0.308 | | cent Correlatic | | 00% | 100% | 100% | 95% | 100% | 100% | 20% | 00% | 100% | 95% | 95% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | | rrelation of cut off Specimen | | | | | | | | | | | 97% | | | | | | | | | . . -+ -ﮯ۔ ducibility ith الله R Legend: LP = low positive, HP = High positive, LN = Low negative, HN = High negative, {10}------------------------------------------------ # CONCLUSIONS Premier Platinum HpSA PLUS: - 1. Can be used reliably for the rapid detection of H. pylori in human stool specimens - 2. Performs similarly to the existing FDA approved Premier Platinum HpSA (K980076 and K983255). {11}------------------------------------------------ Image /page/11/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three overlapping, curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 MAR 1 0 2006 Ms. Susan Rolih Vice President, Regulatory Affairs and Quality Assurance Meridian Bioscience, Inc. 3471 River Hills Drive Cincinnati, OH 45244 k053335 Re: > Trade/Device Name: Premier Platinum HpSA PLUS Regulation Number: 21 CFR 866.3110 Regulation Name: Campylobacter Fetus Serological Reagents Regulatory Class: Class I Product Code: LYR Dated: January 31, 2006 Received: February 1, 2006 Dear Ms. Rolih: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {12}------------------------------------------------ #### Page 2 -- This letter will allow you to begin marketing your device as described in your Section 510(k) I mis lotel will and my your he FDA finding of substantial equivalence of your device to a legally premated predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours, Sales, a Hogg Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {13}------------------------------------------------ Premier Platinum HpSA PLUS 510(k) Notification Meridian Bioscience, Inc. # INDICATIONS FOR USE STATEMENT Premier Platinum HpSA PLUS **510(K) Number:** K053335 The Premier Platinum HpSA PLUS enzyme immunoassay (EIA) is an in vitro qualitative procedure for The Trention of Helicobacter pylor antigens in human stool. Test results are intended to aid in the the decodion of H. pylor infection and to monitor response during and post-therapy in patients. Accepted ulaginosis or i. pyron invocion that testing by any current method, to confirm eradication, be done at least four weeks following completion of therapy. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) freddie lu. loob Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety C1934) K05 3335 Section 4, Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...