PREMIER Platinum HpSA PLUS

K182559 · Meridian Bioscience, Inc. · LYR · Nov 5, 2018 · Microbiology

Device Facts

Record IDK182559
Device NamePREMIER Platinum HpSA PLUS
ApplicantMeridian Bioscience, Inc.
Product CodeLYR · Microbiology
Decision DateNov 5, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3110
Device ClassClass 1

Indications for Use

The PREMIER Platinum HpSA PLUS enzyme immunoassay (EIA) is an in vitro qualitative procedure for the detection of Helicobacter pylori antigens in human stool. Test results are intended to aid in the diagnosis of H. pylori infection and to monitor response during and post-therapy in patients. Accepted medical practice recommends that testing by any current method, to confirm eradication, be done at least four weeks following completion of therapy.

Device Story

PREMIER Platinum HpSA PLUS is a microwell-based enzyme immunoassay (EIA) for qualitative detection of H. pylori antigens in human stool. Device utilizes monoclonal antibodies adsorbed to microwells. Process: diluted patient sample and enzyme conjugate added to microwells; one-hour incubation; wash step; substrate addition; 10-minute incubation; stop solution addition. Color development indicates presence of bound enzyme. Results interpreted visually or spectrophotometrically. Used in clinical laboratory settings. Provides objective, simple interpretation without complex calculations. Aids clinicians in diagnosing H. pylori infection and monitoring treatment effectiveness, relapse, or eradication.

Clinical Evidence

Clinical study compared modified device to predicate using 159 archived, unpreserved stool samples from symptomatic patients. Results showed 100% positive percent agreement (95% CI: 93.7-100.0%) and 100% negative percent agreement (95% CI: 96.4-100.0%). Analytical performance included reproducibility (100% agreement across 300 tests), LoD (4.66 ng/mL), and cross-reactivity/interference testing.

Technological Characteristics

Qualitative enzyme immunoassay (EIA) in microwell format. Utilizes monoclonal antibodies adsorbed to microwells. Detection via colorimetric enzyme-substrate reaction. Interpretation is visual or spectrophotometric. No electronic components or software algorithms.

Indications for Use

Indicated for symptomatic patients requiring diagnosis of H. pylori infection or monitoring of treatment response during and post-therapy. Testing for eradication confirmation is recommended at least four weeks post-therapy.

Regulatory Classification

Identification

Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identify Campylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases. Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K182559 B. Purpose for Submission: To obtain a substantial equivalence determination for a modification of the PREMIER Platinum HpSA PLUS C. Measurand: Helicobacter pylori antigen D. Type of Test: Qualitative enzyme immunoassay (EIA) E. Applicant: Merdidian Bioscience, Inc. F. Proprietary and Established Names: PREMIER Platinum HpSA PLUS G. Regulatory Information: 1. Regulation section: 21 CFR 866.3110; Campylobacter fetus serological reagents 2. Classification: Class I 3. Product code: LYR – Campylobacter pylori {1} 4. Panel: 83-Microbiology H. Intended Use: 1. Intended use(s): The PREMIER Platinum HpSA PLUS enzyme immunoassay (EIA) is an in vitro qualitative procedure for the detection of Helicobacter pylori antigens in human stool. Test results are intended to aid in the diagnosis of H. pylori infection and to monitor response during and post-therapy in patients. Accepted medical practice recommends that testing by any current method, to confirm eradication, be done at least four weeks following completion of therapy. 2. Indication(s) for use: Same as the Intended Use. 3. Special conditions for use statement(s): For prescription use only. 4. Special instrument requirements: A spectrophotometric plate reader capable of reading the following wavelengths: 450 nm or 450/630 nm. I. Device Description: The PREMIER Platinum HpSA PLUS test is a microwell-based enzyme immunoassay that detects H. pylori antigens present in human stool. This version of the assay has several modifications compared to the earlier version, including a replacement capture/conjugate monoclonal minor antibody and capture/conjugate monoclonal major antibody propagated by different methods. Additionally, the positive control reagent final acceptance criteria have been modified, concurrent with the minor antibody change. The test utilizes a plurality (mixture) of monoclonal anti-H. pylori capture antibodies adsorbed to microwells. Diluted patient samples and an enzyme conjugate reagent are added to the microwells and incubated for one hour at room temperature. A wash is performed to remove unbound material. Substrate is added and incubated for 10 minutes at room temperature. Color develops in the presence of bound enzyme. Stop solution is added and the results are interpreted visually or spectrophotometrically. No calculations are required and the visual color change makes the interpretation of results objective and simple. The color, if 2 {2} positive, is a yellow color that may be read visually or by using a spectrophotometer at the following wavelengths: 450 nm or 450/630 nm. J. Substantial Equivalence Information: 1. Predicate device name(s): PREMIER Platinum HpSA PLUS 2. Predicate 510(k) number(s): K053335, K980076, K983255 3. Comparison with predicate: 3 {3} | Similarities | | | | --- | --- | --- | | Item | Modified Device (K182559) | Predicate Device (K053335) | | Product Code | LYR | Same | | Intended Use | The Premier Platinum HpSA PLUS enzyme immunoassay (EIA) is an in vitro qualitative procedure for the detection of Helicobacter pylori antigens in human stool. Test results are intended to aid in the diagnosis of H. pylori infection and to monitor response during and post-therapy in patients. Accepted medical practice recommends that testing by any current method, to confirm eradication, be done at least four weeks following completion of therapy. | Same | | Technology | Qualitative Enzyme Immunoassay (EIA), Microwell Format | Same | | Specimen Type | Human Stool (solid, liquid) | Same | | Reagents and Components | • Antibody Coated Microwells • Positive Control • Sample Diluent/Negative Control • Premier 20X Buffer I • Enzyme Conjugate • Premier Substrate I • Premier Stop Solution 1 • Transfer pipettes • Microwell strip sealer • Wooden stick applicators | Same | | Interpretation of Results | Results may be interpreted visually or using a spectrophotometer. | Same | | LoD | 4.66 ng/mL of stool | 4.67 ng/mL of stool | {4} | Differences | | | | --- | --- | --- | | Item | Modified Device (K182559) | Predicate Device (K053335) | | Antibody Coated Microwells | **Major Antibody:** Monoclonal anti-H. pylori capture antibody, derived from tissue culture. **Minor Antibody:** Replacement monoclonal anti-H. pylori capture antibody, derived from tissue culture. | **Major Antibody:** Monoclonal anti-H. pylori capture antibody, derived from ascites. **Minor Antibody:** Monoclonal anti-H. pylori capture antibody, derived from ascites. | | Enzyme Conjugate | **Major Antibody:** Monoclonal anti-H. pylori conjugate antibody, derived from tissue culture. **Minor Antibody:** Replacement monoclonal anti-H. pylori conjugate antibody, sourced from cell culture. | **Major Antibody:** Monoclonal anti-H. pylori conjugate antibody, derived from ascites. **Minor Antibody:** Monoclonal anti-H. pylori conjugate antibody, derived from ascites. | | Positive Control Acceptance Criteria | A450/630 1.8-2.3 | A450/630 1.2-1.8 | K. Standard/Guidance Document Referenced (if applicable): Not applicable L. Test Principle: Enzyme-linked immunosorbent assay M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Studies were performed to demonstrate the inter-laboratory and intra-laboratory reproducibility of the modified device. Testing was performed by two technologists in three separate laboratories over the course of five days. Each technologist was supplied with a panel of blinded specimens prior to testing. The specimen panels were tested according to the package insert method. {5} Contrived specimen panels were prepared by spiking H. pylori purified flaggelar antigen (H. pylori ATCC 43504) into negative pooled stool at antigen concentrations above, near and below the assay limit of detection for H. pylori antigen. Each panel consisted of one negative, three high negatives (0.1x LoD), three low positives (2x LoD), and three moderate positive (5x LoD) specimens, that were randomly sorted within each panel. Negative and positive controls were run on each day of testing. Three kit lots were used during the study. Results were interpreted using either single or dual wavelength spectrophotometric reads. The study results are summarized in Table 1. No differences were observed for single or dual wavelength spectrophotometric reads. The reproducibility performance was acceptable. Table 1. Reproducibility Testing Results Obtained with the Modified Device by Single or Dual Wavelength Spectrophotometric Read | Sampe Type | Clinical Site 1 Percent Agreement | Clinical Site 2 Percent Agreement | Clinical Site 3 Percent Agreement | Total | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | | Percent Agreement | 95% CI | | | | Negative | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 (100%) | 95.9%-100% | | High Negative | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 (100%) | 95.9%-100% | | Low Positive | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 (100%) | 95.9%-100% | | Moderate Positive | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 (100%) | 95.9%-100% | | Negative Control | 10/10 | 100% | 10/10 | 100% | 10/10 | 100% | 30/30 (100%) | 88.6%- 100% | | Positive Control | 10/10 | 100% | 10/10 | 100% | 10/10 | 100% | 30/30 (100%) | 88.6%- 100% | b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): Not applicable d. Detection limit: The lower limit of detection of this assay is $4.66\mathrm{ng / mL}$ , measured by dual wavelength, for unpreserved, pooled negative human stool. e. Analytical specificity: # Cross-reactivity The specificity of the modified PREMIER Platinum HpSA PLUS was tested with the bacterial, viral, and fungal strains listed in Table 2, below. Positive and negative stools (unpreserved pooled negative human stool, previously screened by the {6} modified and predicate device) were spiked with $1.0 \times 10^{7} \mathrm{CFU/mL}$ bacteria and fungi. Viruses were spiked at a final concentration greater than $1 \times 10^{5} \mathrm{TCID}_{50} / \mathrm{mL}$ . None of the microorganisms tested yielded a positive result in the negative stool or interfered with detection of $H.$ pylori antigen in positive stool. Table 2. Microorganisms Tested for Cross-Reactivity | Adenovirus 41 | Klebsiella pneumoniae | | --- | --- | | Aeromonas hydrophila | Lactococcus lactis | | Bacillus subtilis | Listeria monocytogenes | | Borrelia burgdorferi | Peptostreptococcus anaerobius | | Campylobacter coli | Proteus vulgaris | | Campylobacter fetus | Pseudomonas aeruginosa | | Campylobacter jejuni | Pseudomonas fluorescens | | Campylobacter jejuni (2) | Rotavirus | | Campylobacter lari | Salmonella enterica subsp. enterica serovar dublin | | Candida albicans | Salmonella enterica subsp. enterica serovar hilversum | | Citrobacter freundii | Salmonella enterica subsp. enterica serovar heidelberg (Group B) | | Clostridium difficile | Salmonella enterica subsp. enterica serovar minnesota | | Clostridium perfringens | Salmonella enterica subsp. enterica serovar typhimurium | | Enterobacter cloacae | Serratia liquefaciens | | Enterococcus faecalis | Serratia marcescens | | Escherichia coli O157:H7 | Shigella boydii | | Escherichia coli | Shigella dysenteriae | | Escherichia coli 8739 | Shigella flexneri | | Escherichia coli 9637 | Shigella sonnei | | Escherichia fergusonii | Staphylococcus aureus | | Escherichia hermanii | Staphylococcus aureus (Cowan Strain I) | | Escherichia hermanii EMDI-64 | Staphylococcus epidermidis | | Haemophilus influenza | Yersinia enterocolitica | # Interfering substances The following substances, that may be present in human stool, do not interfere with positive or negative test results at the stated concentrations per $500~\mathrm{uL}$ human stool: Barium sulfate $-25\mathrm{mg}$ , Mylanta $-11.5\mathrm{mg}$ , Pepto-Bismol $-0.44\mathrm{mg}$ , Prilosec (omeprazole) $-1\mathrm{mg}$ , Tagamet (cimetidine) $-1\mathrm{mg}$ , TUMS $-10\mathrm{mg}$ , Hemoglobin $-62.5\mathrm{mg}$ , Mucin $-17\mathrm{mg}$ , NSAID Ibuprofen $-0.250\mathrm{mg}$ , Stearic acid $-5.3\mathrm{mg}$ , Paltmitic acid $-2.65\mathrm{mg}$ , White blood cells $-250\mathrm{uL}$ , Whole blood $-250\mathrm{uL}$ . Assay cut-off: Not applicable f. Prozone/Hook Effect: Not applicable {7} 8 2. Comparison studies: a. Method comparison with predicate device: Method comparison testing was done to compare performance of the modified PREMIER Platinum HpSA PLUS, using a replacement capture/conjugate monoclonal minor antibody and capture/conjugate monoclonal major antibody propagated by different methods, to that of the predicate device. Testing was performed on 159 archived, unpreserved clinical stool specimens. The specimens were collected from individuals with signs and symptoms of *H. pylori* infection (from the Intended Use Population). Three lots of the modified device were tested. When compared to results obtained with the predicate, the percent positive and percent negative agreements were 100%. The results are summarized in Table 3 below. Table 3. Performance Comparison between the Modified Device and Predicate Device | | PP HpSA PLUS Predicate | | | | --- | --- | --- | --- | | PP HpSA PLUS Modified | Positive | Negative | Total | | Positive | 57 | 0 | 57 | | Negative | 0 | 102 | 102 | | Total | 57 | 102 | 159 | | | | | 95% CI | | Positive Percent Agreement | 100% (57/57) | | 93.7-100% | | Negative Percent Agreement | 100% (102/102) | | 96.4-100% | The percent positive and negative agreements compared to the predicate device were 100% (57/57, CI = 93.7% - 100%) and 100% (102/102, CI = 96.4% -100%), respectively. These results are acceptable. b. Matrix comparison: Not applicable {8} 9 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: See section M3a. above. c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Studies on the epidemiology of *H. pylori* have shown that this organism is present worldwide. Gastritis caused by *H. pylori* has been shown to correlate with age, ethnic background, family size and socioeconomic class. The prevalence of *H. pylori* infection in a given population can vary from 20% to 90%. In patients diagnosed with duodenal ulcers, however, it has been shown in every age group to be approximately 80%. Currently recommended eradication treatments have an efficacy rate between 75% and 90%. The modified Premier Platinum HpSA PLUS test detects the presence of *H. pylori* antigens in human stool. Expected values for a given population should be determined for each laboratory. The rate of positivity may vary depending on geographic location, method of specimen collection, handling and transportation, test employed and general health environment of the patient population under study. N. Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. O. Conclusion: The submitted information in this premarket notification is complete and supports substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...