ABBOTT REALTIME CT/NG AND ABBOTT MULTI-COLLECT SPECIMEN COLLECTION KIT, MODELS 8L07-91, 9K12-03

K092704 · Abbott Molecular, Inc. · LSK · May 28, 2010 · Microbiology

Device Facts

Record IDK092704
Device NameABBOTT REALTIME CT/NG AND ABBOTT MULTI-COLLECT SPECIMEN COLLECTION KIT, MODELS 8L07-91, 9K12-03
ApplicantAbbott Molecular, Inc.
Product CodeLSK · Microbiology
Decision DateMay 28, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3120
Device ClassClass 1

Intended Use

The Abbott RealTime CT/NG assay is an in vitro polymerase chain reaction (PCR) assay for the direct, qualitative detection of the plasmid DNA of Chlamydia trachomatis and the genomic DNA of Neisseria gonorrhoeae. The assay may be used to test the following specimens from symptomatic individuals: female endocervical swab, clinician-collected vaginal swab, and patient-collected vaginal swab specimens; male urethral swab specimens; and female and male urine specimens. The assay may be used to test the following specimens from asymptomatic individuals: clinician-collected vaginal swab and patient-collected vaginal swab specimens; female and male urine specimens.

Device Story

The Abbott RealTime CT/NG assay is an in vitro PCR-based diagnostic test for Chlamydia trachomatis and Neisseria gonorrhoeae. It utilizes the Abbott m2000 System, which integrates automated sample preparation (m2000sp) with real-time PCR amplification and fluorescence detection (m2000rt). The device processes clinical swabs and urine specimens to detect C. trachomatis plasmid DNA and N. gonorrhoeae genomic DNA. It is intended for use in clinical laboratory settings by trained personnel. The system provides qualitative results, which aid clinicians in diagnosing chlamydial and gonococcal infections. By providing sensitive and specific detection, the assay supports timely treatment, partner notification, and the prevention of complications such as pelvic inflammatory disease, ectopic pregnancy, and infertility.

Clinical Evidence

Multicenter clinical study (n=3,832) across 16 US sites. Compared Abbott RealTime CT/NG against reference NAATs and NG culture. Overall sensitivity/specificity for CT: 95.2%/99.3%; for NG: 97.5%/99.7%. Precision study performed at 3 sites using 15-member panels over 30 runs, demonstrating high reproducibility.

Technological Characteristics

In vitro PCR assay using real-time fluorescence detection. Targets C. trachomatis cryptic plasmid and N. gonorrhoeae multicopy opacity gene. Automated sample preparation and amplification/detection on the Abbott m2000 System. Qualitative output. Reagents include amplification kit and control kit.

Indications for Use

Indicated for the qualitative detection of Chlamydia trachomatis and Neisseria gonorrhoeae DNA in symptomatic and asymptomatic individuals using female endocervical, vaginal (clinician or patient-collected), and male urethral swabs, as well as male and female urine specimens.

Regulatory Classification

Identification

Chlamydia serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to chlamydia in serum. Additionally, some of these reagents consist of chlamydia antisera conjugated with a fluorescent dye used to identify chlamydia directly from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of disease caused by bacteria belonging to the genus Chlamydia and provides epidemiological information on these diseases. Chlamydia are the causative agents of psittacosis (a form of pneumonia), lymphogranuloma venereum (a venereal disease), and trachoma (a chronic disease of the eye and eyelid).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 1092704 MAY 2 8 2010 # Abbott RealTime CT/NG List No. 8L07-91 510(k) Summary Abbott RealTime CT/NG // multi-Co!lect Specimen Collection Kit May 2010 Admin Sec 3_510(k) Summary_mw019 Admin. Sec. 3 510(k) Summary Page 1 of 44 : {1}------------------------------------------------ # Table of Contents | | | Page | |------|--------------------------------------------------------------------------|------| | 3.0 | Abbott RealTime CT/NG 510(k) Summary | 3 | | 3.1 | Official Correspondent to the File | 3 | | 3.2 | Trade Name: | 3 | | 3.3 | Common Name: | 3 | | 3.4 | Classification Name: | 3 | | 3.5 | Registration Number and Classification Code: | 3 | | 3.6 | Substantially Equivalent Devices | 4 | | 3.7 | Purpose of the Submission | 4 | | 3.8 | Date of Preparation | 4 | | 3.9 | Manufacturer | 4 | | 3.10 | Intended Use | 5 | | 3.11 | Device Description | 5 | | 3.12 | Background on Chlamydia and Gonorrheal Disease | 5 | | 3.13 | Technological Characteristics of the Device as Compared to the Predicate | 7 | | 3.14 | Summary of Nonclinical Studies | 10 | | 3.15 | Summary of Clinical Studies | 19 | | 3.16 | Conclusion Drawn from Clinical Studies | 42 | | 3.17 | References | 43 | : . : . . . : . : · {2}------------------------------------------------ #### 3.0 Abbott RealTime CT/NG 510(k) Summary A summary of the Abbott® RealTime CT/NG assay #### 3.1 Official Correspondent to the File ### Submitted By: Abbott Molecular Inc. 1300 E. Touhy Avenue Des Plaines, IL 60018 phone: (224) 361-7000 eFAX (847) 775-6777 ### Company Contact: Paula Martin Senior Manager Regulatory and Clinical Affairs Phone (224) 361-7333, eFAX (847) 775-6777 e-mail: paula.martin@abbott.com #### 3.2 Trade Name: Abbott RealTime CT/NG (List No. 8L07-91) #### 3.3 Common Name: In vitro polymerase chain reaction (PCR) assay for Chlamydia trachomatis and Neisseria gonorrhoeae. #### 3.4 Classification Name: Nucleic acid test (NAT) #### 3.5 Registration Number and Classification Code: 21 CFR 866.3390 (Class II, Code LSL) describes test reagents used to identifiy Neisseria from clinical specimens. 21 CFR 866.3120 (Class I, Code MKZ) describes test reagents used to identify Chlamydia from clinical specimens. {3}------------------------------------------------ #### 3.6 Substantially Equivalent Devices ### Abbott RealTime CT/NG (List No. 8L07-91) Predicate Devices: GEN-PROBE APTIMA Combo 2 Assay (Assigned 510(k) No. K043224), Becton Dickenson ProbeTec ET Chlamydia trachomatis /Neisseria gonorrhoeae Amplified DNA Assay (Assigned 510(k) No. K012351). The Abbott RealTime CT/NG assay was also compared to cell culture methods intended for the qualitative detection of Neisseria gonorrhoeae. The predicate culture methods are: bioMerieux API® NH Assay, EY Laboratories Gonocheck® II (Assigned 510(k) No. 940162) Assay, Pharmacia Diagnostics Phadebact® Monoclonal GC test. #### 3.7 Purpose of the Submission The purpose of this 510(k) is to gain clearance to market the Abbott RealTime CT/NG (List No. 8L07-91) assay. #### 3.8 Date of Preparation: August, 2009 #### 3.9 Manufacturer Abbott Molecular Inc. is the legal manufacturer of the Abbott RealTime CTNG (List No. 8L07-91) assay and the Abbott multi-Collect Specimen Collection Kit (List No. 9K12-03). Timothy Zurow Director Manufacturing Operations Phone (224) 361-7379. eFAX (847) 775-6777 e-mail: timothy.zurow(@abbott.com Abbott Molecular Inc. 1300 E. Touhy Avenue Des Plaines, IL 60018 Abbott Molecular Inc. Establishment Registration No .: 3005248192 {4}------------------------------------------------ #### Intended Use 3.10 The proposed intended use for the Abbott RealTime CT/NG assay is: The Abbott RealTime CT/NG (List No. 8L07-91) assay is an in vitro polymerase chain reaction (PCR) assay for the direct, qualitative detection of the plasmid DNA of Chlamydia trachomatis and the genomic DNA of Neisseria gonorrhoeae. The assay may be used to test the following specimens from symptomatic individuals: female endocervical swab, clinician-collected vaginal swab, and patient-collected vaginal swab specimens; male urethral swab specimens; and female and male urine specimens. The assay may be used to test the following specimens from asymptomatic individuals: clinician-collected vaginal swab and patient-collected vaginal swab specimens; female and male urine specimens. #### 3.11 Device Description Abbott RealTime CT/NG consists of two reagent kits: - · Abbott RealTime CT/NG Amplification Reagent Kit (List No. 8L07-91) - · Abbott RealTime CT/NG Control Kit (List No. 8L07-80) The Abbott RealTime CT/NG assay uses PCR technology with homogenous real-time fluorescence detection on the m2000 System. The Abbott m2000 System consists of the Abbott m2000sp and Abbott m2000rt instruments. The Abbott m2000 System integrates sample preparation with nucleic acid amplification and detection to generate assay results. The Abbott m2000sp is used for processing samples and the Abbott m2000rt is used for amplification and detection. #### 3.12 Background on Chlamydia and Gonorrheal Disease Chlamydia are non-motile, Gram-negative, obligate intracellular parasites of eukaryotic cells. They form inclusions in the cytoplasm of the host cell. Chlamydia trachomatis , one of three chlamydial species, is the causative agent of the sexually transmitted disease (STD) chlamydia. Chlamydial infections of the urogenital tract are associated with salpingitis, cervicitis, ectopic pregnancies and tubal factor infertility in women as well as Admin, Sec, 3 510(k) Summary Page 5 of 44 {5}------------------------------------------------ nongonococcal urethritis and epididymitis in men. 1-4 The genital site most commonly affected in women is the cervix, but the infection can be asymptomatic and, if untreated, is likely to ascend to the uterus, fallopian tubes and ovaries causing pelvic inflammatory disease (PID). Neonates born of infected mothers can contract inclusion conjunctivitis, nasopharyngeal infections, and pneumonia due to Chlamydia trachomatis . Infection by Chlamydia trachomatis in men is also often asymptomatic and, if untreated, may lead to epididymitis, a major complication.3 Patients infected with Chlamydia trachomatis may be co-infected with Neisseria gonorrhoeae , the causative agent of gonorrhea. Further, patients with treatment indications for gonorrhea but not chlamydia often harbor Chlamydia trachomatis.7 Chlamydia infections may not respond well to recommended regimens for treating Neisseria gonorrhoeae. Therefore, unless chlamydial infection has been ruled out in patients treated for gonorrhea, dual therapy for gonococcal and chlamydia infections is recommended.5 Cell culture, commonly used to detect Chlamydia trachomatis, has been replaced by more sensitive nucleic acid tests. Since a specific diagnosis of chlamydia may improve treatment compliance and enhance partner notification, the use of these highly sensitive and specific tests is strongly recommended.3 Gonorrhea is one of the most common sexually transmitted diseases in the United States. Over 700,000 new infections of Neisseria gonorrhoeae are estimated to occur each year. In men, gonorrhea infection usually results in acute anterior urethritis accompanied by a purulent exudate.19,11 In women, the infection is most often found in the cervix, but the vagina and uterus also may be infected. Frequently the infection is asymptomatic, especially in women. Without treatment, local complications of gonococcal infection can occur including pelvic inflammatory disease (PID) or acute salpingitis for women and epididymitis for men. 10.11 Rarely, disseminated gonococcal infection, DGI. may occur in untreated patients. 13 Neisseria gonorrhoeae is a Gram-negative, oxidase-positive diplococcus without flagellae.12 Culture is commonly used for the detection of Neisseria gonorrhoeae. Presumptive diagnosis of gonorrhea is based on the morphological examination, Gram Abbott RealTime CT/NG // multi-Collect Specimen Collection Kit May 2010 Admin Sec 3_510(k) Summary_mw019 Admin. Sec. 3 510(k) Summary Page 6 of 44 {6}------------------------------------------------ stain, and oxidase measurement of the culture isolate. Confirmation procedures have been used for definitive identification of Neisseria gonorrhoeae including sugar fermentation, fluorescent antibody staining, nucleic acid hybridization, and agglutination. 4315 Nucleic acid tests are widely available for the sensitive detection of Neisseria gonorrhoeae . ### 3.13 Technological Characteristics of the Device as Compared to the Predicate The primary functional components of the Abbott RealTime CT/NG assay are substantially equivalent to other legally marketed nucleic acid amplification tests (NAAT) intended for the qualitative detection of Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (NG). The Abbott RealTime CT/NG assay has the same general intended uses as the predicate devices. Although there are some technological differences between the Abbott RealTime CT/NG and the predicate devices, these differences do not raise new types of safety or effectiveness questions. These devices are similar in that they are designed to prepare nucleic acids for amplification, amplify specific Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (NG) sequences, detect the amplified products, and report qualitative results. The primary similarities and differences between the Abbott RealTime CT/NG assay and the NAAT predicate devices are shown in Attachment Tables 3.1 and 3.2. {7}------------------------------------------------ | l | | |--------|--| | న | | | | | | le | | | q<br>C | | | r<br>1 | | milarities and Differences Between Abbott RealTime CT/NG and Nucleic Acid Amplification Predicate Devices | Feature | Abbott RealTime CT/NG | Gen-Probe Aptima Combo 2 | Becton Dickenson ProbeTec ET | | |------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------| | Intended Use | The Abbott RealTime CT/NG (List<br>No. 8L07-91) assay is an in vitro<br>polymerase chain reaction (PCR)<br>assay for the direct, qualitative<br>detection of the plasmid DNA of<br><i>Chlamydia trachomatis</i> and the<br>genomic DNA of <i>Neisseria<br/>gonorrhoeae</i> . The assay may be<br>used to test the following<br>specimens from symptomatic<br>individuals: female endocervical<br>swab, clinician-collected vaginal<br>swab, and patient-collected vaginal<br>swab specimens; male urethral<br>swab specimens; and female and<br>male urine specimens. The assay<br>may be used to test the following<br>specimens from asymptomatic<br>individuals: clinician-collected<br>vaginal swab and patient-collected<br>vaginal swab specimens; female<br>and male urine specimens. | The APTIMA Combo 2 Assay is a target<br>amplification nucleic acid probe test that<br>utilizes target capture for the in vitro<br>qualitative detection and differentiation of<br>ribosomal RNA (rRNA) from <i>Chlamydia<br/>trachomatis</i> and/or <i>Neisseria gonorrhoeae</i><br>in clinician collected endocervical, vaginal,<br>and male urethral swab specimens,<br>patientcollected vaginal swab specimens,<br>and female and male urine specimens. The<br>assay may be used to test specimens from<br>symptomatic and asymptomatic individuals<br>to aid in the diagnosis of gonococcal<br>and/or chlamydial urogenital disease using<br>the TIGRIS DTS Automated Analyzer or<br>semiautomated instrumentation as<br>specified.<br><br>The assay is also intended for use with<br>testing of gynecological specimens<br>collected in the PreservCyt Solution and<br>processed with the Cytyc ThinPrep 2000<br>System. | The BD ProbeTec ET <i>Chlamydia<br/>trachomatis</i> (CT) and <i>Neisseria<br/>gonorrhoeae</i> (GC) Amplified DNA<br>Assays, when tested with the BD<br>ProbeTec ET System, use Strand<br>Displacement Amplification (SDA)<br>technology for the direct, qualitative<br>detection of <i>Chlamydia trachomatis<br/></i> and <i>Neisseria gonorrhoeae</i> DNA in<br>endocervical swabs, male urethral,<br>swabs, and in female and male urine<br>specimens as evidence of infection<br>with <i>C. trachomatis, N. gonorrhoeae</i> ,<br>or of coinfection with both <i>C.<br/>trachomatis</i> and <i>N. gonorrhoeae</i> .<br>Specimens may be from symptomatic<br>or asymptomatic females and males. A<br>separate Amplification Control is an<br>option for inhibition testing (BD<br>ProbeTec ET CT/GC/AC Reagent<br>Pack). The BD ProbeTec ET CT/GC<br>assays may be performed using either<br>the BD ProbeTec ET System or a<br>combination of the BD ProbeTec ET<br>System and BD Viper instrument. | | | Feature | Current Application<br>Abbott RealTime CT/NG | | Amplified Nucleic Acid Predicate Devices | | | | | | Gen-Probe Aptima Combo 2 | Becton Dickenson ProbeTec ET | | Assay Type | Qualitative | • | Qualitative | Qualitative | | CT Analyte<br>Targets | CT cryptic plasmid DNA | • | CT ribosomal RNA | CT cryptic plasmid DNA | | NG Analyte<br>Targets | NG genomic DNA | • | NG ribosomal RNA | NG genomic DNA | | Input Sample<br>Types | Endocervical swab specimens<br>Self-collected vaginal swab<br>specimens<br>Clinician-collected vaginal<br>swab specimens<br>Male urethral swab specimens<br>Male and female urine<br>specimens | • | Endocervical swab specimens<br>Self-collected vaginal swab<br>specimens<br>Clinician-collected vaginal<br>swab specimens<br>Male urethral swab specimens<br>Male and female urine<br>specimens<br>PreservCyt liquid Pap specimens | Endocervical swab specimens<br>Male urethral swab specimens<br>Male and female urine<br>specimens | | Sample<br>Preparation<br>Procedure | Automated | • | Semi-automated/automated | Manual/ semi-automated | | Amplification<br>Technology | Real-time PCR | | Ribosomal RNA transcriptionmediated<br>amplification (TMA) | Strand displacement DNA<br>amplification (SDA) | | Assay<br>Controls | Negative Control<br>Cutoff Control<br>Internal Control | • | Negative Control<br>Positive Control | Negative Control<br>Positive Control<br>Optional Amplification Control | ### Abbott RealTime CT/NG // multi-Collect Specimen Collection Kit May 2010 Admin Sec 3_510(k) Summary_mw019 Admin. Sec. 3 510(k) Summary {8}------------------------------------------------ Similarities and Differences Between Abbott RealTime CT/NG and Nucleic Acid Amplification Predicate Devices Abbott RealTime CT/NG // multi-Collect Specimen Collection Kit May 2010 Admin Sec 3_510(k) Summary_mw019 Page 9 of 44 ' {9}------------------------------------------------ #### Summary of Nonclinical Studies 3.14 #### Analytical Sensitivity 3.14.1 The Limit of Detection (LOD) claim for the Abbott RealTime CT/NG assay is 320 copies of Chlamydia trachomatis (CT) target DNA and 320 copies of Neisseria gonorrhoeae (NG) target DNA per assay. The assay targets the Chlamydia trachomatis cryptic plasmid (present at approximately 7 to 10 copies per Chlamydia organism) and the multicopy opacity gene of Neisseria gonorrhoeae (repeated up to 11 times per organism). Thus, 320 copies of target DNA is equivalent to approximately 30 to 40 organisms per assay. The LOD of the Abbott RealTime CT/NG assay is defined as CT and NG target DNA concentration detected with a probability of 95% or greater. The CT and NG DNA concentrations detected with 95% probability were determined by testing dilutions of CT and NG target DNA. Probit analysis of the data determined that the concentration of CT DNA detected with 95% probability was 21 copies/assay (95% CI 18 - 28), the concentration of nvCT DNA detected with 95% probability was 29 copies/assay (95% CI 24 - 41), and the concentration of NG DNA detected with 95% probability was 149 copies/assay (95% CI 130 - 176). The claimed assay LOD was confirmed by testing samples that contained 320 copies of CT, nvCT and NG target DNA per assay. The detection rate was 100% (405/405) for CT, 100% (403/403) for nvCT, and 99.5% (403/405) for NG in the assay. An additional study was conducted to challenge the performance of the Abbott RealTime CT/NG assay in samples containing high target numbers of CT, nvCT, or NG in the presence of low target numbers of the opposite analyte. Samples were prepared to contain 320 CT or nvCT target DNA copies and 1 x 10' NG target DNA copies per assay, or 1 x 107 CT or nvCT target DNA copies and 320 NG target DNA copies per assay. The detection rate of 320 copies of CT or nvCT DNA in the presence of high NG target was 100% (405/405). The detection rate of 320 copies of NG DNA in the presence of high CT or nvCT target was 100% (405/405). Abbott Rea!Time CT/NG // multi-Collect Specimen Collection Kit May 2010 Admin Sec 3_510{k) Summary_mw019 Admin. Sec. 3 510(k) Summary Page 10 of 44 {10}------------------------------------------------ The analytical sensitivity of the Abbott RealTime CT/NG assay for detecting Chlumydia trachomatis serovars A through L was determined by testing dilutions of each serovar. Serovars A through K and L1 through L3 were detected at less than 1 Inclusion Forming Units (IFU) per assay. Additionally, nvCT was diluted and was also detected at less than 1 IFU per assay. The analytical sensitivity of the Abbott RealTime CT/NG assay for detecting 28 different isolates of Neisseria gonorrhoeae was determined by testing dilutions of each isolate. All isolates were detected at less than 1 Colony Forming Unit (CFU)/assay. #### 3.14.2 Evaluation of Potential Cross-Reactants A total of 111 strains of bacteria, viruses, parasites, yeast, and fungi were tested for potential cross reactivity in the Abbott RealTime CT/NG assay (Table 3.6). These included organisms that are phylogenetically related to CT and NG, and those that can be found in the urogenital tract. Purified DNA or RNA was diluted to a final concentration of 1 x 10' copies/assay. HBV DNA and HCV RNA were added directly into the PCR reaction at approximately 4 x 105 and 6 x 10° copies per reaction, respectively. All results were negative for both CT and NG. Additionally, a total of 32 culture isolates were tested for potential cross reactivity in the Abbott RealTime assay. These included 27 organisms listed in Table 3.6, and Neisseria cinerea, Neisseria lactamica, Neisseria sicca, Ca Ski cells containing HPV 16, and Hela cells containing HPV 18. Ca Ski cells containing HPV 16 and Hela cells containing HPV 18 were tested at 106 cells per assay, C. pneumoniae and C. psittaci were tested at 106 EB per assay, HSV-1 and HSV-2 were tested at 106 genomes per assay, and the rest of the organisms were tested at 106 Colony Forming Units (CFU) per assay. All results were negative for both CT and NG. Abbott RealTime CT/NG // multi-Collect Specimen Collection Kit May 2010 Admin Sec 3_510(k) Summary_mw019 {11}------------------------------------------------ | Microorganism/Virus | | | |-------------------------------------|-----------------------------------------|-------------------------------| | Achromobacter xerosis | Haemophilus ducreyi* | Proteus vulgaris | | Acinetobacter calcoaceticus | Haemophilus influenzae | Providencia stuartii | | Acinetobacter Iwoffii | Helicobacter pylori | Pseudomonas aeruginosa* | | Actinomyces israelii | Hepatitis B virus (HBV) | Pseudomonas putida | | Aerococcus viridans | Hepatitis C virus (HCV) | Rahnella aquatilis | | Aeromonas hydrophila | Herpes Simplex Virus, type I* | Rhizobium radiobacter | | Alcaligenes faecalis | Herpes Simplex Virus, type II* | Rhodospirillum rubrum | | Arcanobacterium pyogenes | Human immunodeficiency virus (HIV-1) | Ruminococcus productus | | Bacillus subtilis | Human Papilloma Virus 16 | Salmonella typhimurium | | Bacteroides fragilis | Human Papilloma Virus 18 | Salmonella enterica | | Bacteroides ureolyticus | Kingella denitrificans | Serratia marcescens* | | Bifidobacterium adolescentis | Kingella kingae | Staphylococcus aureus* | | Bifidobacterium breve | Klebsiella oxytoca | Staphylococcus epidermidis* | | Brevibacterium linens | Klebsiella pneumoniae | Staphylococcus saprophyticus* | | Campylobacter jejuni | Lactobacillus acidophilus* | Streptococcus agalactiae* | | Candida albicans* | Lactobacillus brevis* | Streptococcus bovis | | Candida glabrata | Lactobacillus delbrueckii subsp. lactis | Streptococcus mitis. | | Candida parapsilosis | Lactobacillus jensenii | Streptococcus mutans | | Candida tropicalis | Legionella pneumophila | Streptococcus pneumoniae | | Chlamydia pneumoniae* | Listeria monocytogenes | Streptococcus pyogenes | | Chlamydia psittaci* | Micrococcus luteus* | Streptococcus salivarius | | Chromobacterium violaceum | Mobiluncus mulieris | Streptococcus sanguinis | | Chryseobacterium<br>meningosepticum | Moraxella (Branhamella) catarrhalis | Streptomyces griseinus | | Citrobacter freundii | Moraxella lacunata | Trichomonas vaginalis | | Clostridium perfringens | Moraxella osloensis | Ureaplasma urealyticum | | Corynebacterium genitalium* | Morganella morganii | Veillonella parvula | | Corynebacterium xerosis | Mycobacterium gordonae | Vibrio parahaemolyticus | | Cryptococcus neoformans | Mycobacterium smegmatis* | Weissella paramesenteroides | | Cytomegalovirus | Mycoplasma genitalium | Yersinia enterocolitica | | | Microorganism/Virus | | | Deinococcus radiodurans | Mycoplasma hominis | | | Derxia gummosa | Neisseria flava* | | | Eikenella corrodens | Neisseria meningitidis-A* | | | Enterobacter cloacae* | Neisseria meningitidis-B* | | | Enterobacter aerogenes | Neisseria meningitidis-C* | | | Enterococcus avium | Neisseria meningitidis-D* | | | Enterococcus faecalis* | Neisseria perflava* | | | Enterococcus faecium | Pantoea agglomerans | | | Escherichia coli* | Peptostreptococcus anaerobius | | | Fusobacterium nucleatum | Plesiomonas shigelloides | | | Gardnerella vaginalis | Propionibacterium acnes | | | Gemella haemolysans | Proteus mirabilis* | | Table 3.6 Potentially Cross-Reactive Microorganisms/Viruses * Tested with purified DNA or RNA and with culture isolates. {12}------------------------------------------------ # Table 3.6 (Continued) # Potentially Cross-Reactive Microorganisms/Viruses * Tested with purified DNA or RNA and with culture isolates. . {13}------------------------------------------------ #### Evaluation of Potentially Interfering Substances 3.14.3 The potential for interference in the Abbott RealTime CT/NG assay was assessed with substances that may be found in swab and/or urine specimens. Substances were spiked into a swab and/or urine matrix containing 320 copies of CT and NG target DNA per assay, and into a swab and/or urine matrix without CT or NG DNA. No interference in the performance of the Abbott RealTime CT/NG assay was observed in the presence of the substances listed in Table 3.7. Abbott RealTime CT/NG // multi-Collect Specimen Collection Kit May 2010 Admin Sec 3_510(k) Summary_mw019 Admin. Sec. 3 510(k) Summary Page 14 of 44 {14}------------------------------------------------ | able<br>1 | 1 | | | |-----------|---|--|--| |-----------|---|--|--| | Substance | Matrix | Highest Concentration Tested | |----------------------------------------------------|----------------|------------------------------| | Zovirax® Cream 5% | Swab | 0.25% | | CLOTRIMAZOLE Vaginal Cream (2%) | Swab | 0.25% | | Delfen® | Swab | 0.25% | | KY® Jelly | Swab | 0.25% | | Lubrin® | Swab | 0.25% | | Metrogel-Vaginal® | Swab | 0.25% | | Miconazole® 3 Suppository | Swab | 0.25% | | Monostat-1TM Dose Treatment (tioconazole ointment) | Swab | 0.25% | | Norforms® Deodorant Suppositories. | Swab | 0.25% | | Terazol-3® Vaginal Cream | Swab | 0.25% | | Vagi gard® Povidone-Iodine Medicated Douche | Swab | 0.25% | | Vagi gard® Moisturizing Gel | Swab | 0.25% | | Vagisil® Anti-itch Creme | Swab | 0.25% | | Vagisil® Intimate Lubricant | Swab | 0.25% | | Yeast gard® | Swab | 0.25% | | Bilirubin | Urine | 10 mg/mL | | Glucose | Urine | 10 mg/mL | | pH 4 (acidic) Urine | Urine | N/A | | pH 9 (alkaline) Urine | Urine | N/A | | Protein: BGG | Urine | 5% | | Blood | Swab and Urine | 5% | | Leukocytes | Swab and Urine | 1 x 106 cell/mL | Substances That Do Not Interfere with the Abbott RealTime CT/NG Assay Interference in the performance of the Abbott RealTime CT/NG assay may be observed with the following substances: - Talcum powder at concentrations greater than 0.1% in urine specimens. . - Phenazopyridine hydrochloride (the active ingredient in URISTAT) at concentrations . greater than 3 mg/mL in urine specimens. - Mucus at concentrations greater than 0.1% for urine specimens and 1% for swab . specimens. {15}------------------------------------------------ #### 3.14.4 Carryover Potential carryover was determined by performing a study in which high copy CT positive samples were interspersed with negative samples arranged in a checkerboard pattern. The positive samples were CT DNA at a concentration of 101 copies/ml. The carryover rate is defined as the number of CT negative samples that are reported as positive or equivocal over the total number of CT-negative samples tested. Each run included 47 negative samples and 46 positive samples. A total of 14 runs were evaluated using two lots of the RealTime CTNG amplification reagents on four m2000sp and m2000rt instrument pairs. A total of 656 valid negative samples were evaluated for potential carryover effect. A total of 5 false positive and 1 equivocal results were observed. The carryover rate was 0.91%. #### Precision Study 3.14.5 A precision study was performed at three sites, two external and one internal. Each site was provided a fifteen-member panel. Nine panel members targeted different combinations of CT and NG concentrations and six panel members targeted different combinations of nvCT and NG concentrations. The source material for CT was Vero/LGV-II, strain 434. The source material for nvCT was strain 68226. The source material for NG was ATCC isolate 27628 and 31426. Five replicates of each panel member were tested in each run. Thirty runs (10 per site) were performed for a total of 150 replicates of each panel member. The study included three amplification reagent lots. Each site tested two amplification reagent lots. A variance components analysis for a nested model was performed on delta cycle (DC) values, and the results are summarized in Tables 3.8 and 3.9, respectively, Admin. Sec. 3 510(k) Summary Page 16 of 44 {16}------------------------------------------------ | Panel<br>Membera | No.<br>Testedb | No.<br>Positive | Mean<br>Delta<br>Cycle | Within-<br>Run<br>Component<br>SDc | Between-<br>Run<br>Component<br>SDc | Between-<br>Lot<br>Component<br>SDc | Between-<br>Site<br>Component<br>SDc | Total<br>SDc,d | |------------------|----------------|-----------------|------------------------|------------------------------------|-------------------------------------|-------------------------------------|--------------------------------------|----------------| | 1 | 150 | 150 | 15.29 | 0.265 | 0.204 | 0.110 | 0.135 | 0.377 | | 2 | 150 | 150 | 15.67 | 0.411 | 0.245 | 0.000 | 0.179 | 0.511 | | 3 | 150 | 150 | 3.75 | 0.466 | 0.234 | 0.255 | 0.000 | 0.581 | | 4 | 150 | 150 | 9.45 | 0.503 | 0.103 | 0.022 | 0.000 | 0.514 | | 5 | 150 | 0 | ... | ... | ... | ... | ... | ... | | 6 | 149 | 149 | 17.35 | 0.229 | 0.193 | 0.153 | 0.159 | 0.371 | | 7 | 150 | 0 | ... | ... | ... | ... | ... | ... | | 8 | 147 | 0 | ... | ... | ... | ... | ... | ... | | 9 | 150 | 125 | 1.59 | 0.674 | 0.248 | 0.312 | 0.000 | 0.783 | | 10 | 149 | 149 | 15.69 | 0.334 | 0.250 | 0.205 | 0.286 | 0.545 | | 11 | 150 | 150 | 15.59 | 0.428 | 0.180 | 0.216 | 0.289 | 0.588 | | 12 | 150 | 140 | 3.79 | 0.461 | 0.458 | 0.329 | 0.000 | 0.728 | | 13 | 150 | 150 | 9.02 | 0.269 | 0.274 | 0.165 | 0.261 | 0.493 | | 14 | 150 | 150 | 15.63 | 0.284 | 0.265 | 0.109 | 0.413 | 0.578 | | 15 | 147 | 50 | 1.81 | 0.575 | 0.376 | 0.518 | 0.000 | 0.860 | Table 3.8 Precision Study: CT Results ª Chlamydia trachomatis (CT) concentrations were targeted approximately to 4500 IFU/assay in members 1, 2, and 6 and to 45 IFU/assay in member 4. Member 3 was targeted approximately to 0.75 IFU/assay and member 9 to 0.2 IFU/assay both below the claimed assay LOD. New variant strain (nvCT) concentrations were targeted approximately to 50 IFU/assay in members 10, 11, and 14 and 1 IFU/assay in member 13. Members 12 and 15 were targeted to less than 0.1 IFU/assay, below the claimed assay LOD. Members 5, 7, and 8 did not contain any CT or nvCT organisms. b Invalid replicates were excluded from the analysis. & The SD is based on positive replicates only. For member 9, analysis of all replicates with a cycle number (n=133), including those beyond the assay cutoff, resulted in a total SD of 0.960. For member 15, analysis of all replicates with a cycle number (n=52), including those beyond the assay cutoff, resulted in a total SD of 1.037. 4 The total variability contains within-run, between-lot, and between-lot, and between-site variability. {17}------------------------------------------------ | Table | 1<br>0<br>J.J | | |-------|---------------|--| | | | | | Panel<br>Membera | No.<br>Testedb | No.<br>Positive | Mean<br>Delta<br>Cycle | Within-Run<br>Component<br>SDc | Between-<br>Run<br>Component<br>SDc | Between-<br>Lot<br>Component<br>SDc | Between-<br>Site<br>Component<br>SDc | Total<br>SDc,d | |------------------|----------------|-----------------|------------------------|--------------------------------|-------------------------------------|-------------------------------------|--------------------------------------|----------------| | 1 | 150 | 150 | 13.32 | 0.295 | 0.157 | 0.048 | 0.000 | 0.337 | | 2 | 150 | 150 | 7.64 | 0.419 | 0.182 | 0.000 | 0.123 | 0.473 | | 3 | 150 | 150 | 8.03 | 0.288 | 0.146 | 0.000 | 0.000 | 0.323 | | 4 | 149 | 0 | ... | ... | ... | ... | ... | ... | | 5 | 150 | 150 | 7.59 | 0.245 | 0.184 | 0.028 | 0.000 | 0.308 | | 6 | 149 | 0 | ... | ... | ... | ... | ... | ... | | 7 | 150 | 150 | 13.45 | 0.512 | 0.105 | 0.133 | 0.000 | 0.539 | | 8 | 147 | 0 | ... | ... | ... | ... | ... | ... | | 9 | 150 | 69 | 0.51 | 0.326 | 0.000 | 0.000 | 0.029 | 0.327 | | 10 | 149 | 149 | 13.29 | 0.207 | 0.147 | 0.051 | 0.213 | 0.335 | | 11 | 150 | 150 | 7.27 | 0.271 | 0.159 | 0.046 | 0.110 | 0.336 | | 12 | 150 | 150 | 7.24 | 0.220 | 0.180 | 0.000 | 0.088 | 0.297 | | 13 | 150 | 0 | ... | ... | ... | ... | ... | ... | | 14 | 150 | 0 | ... | ... | ... | ... | ... | ... | | 15 | 147 | 47 | 0.50 | 0.348 | 0.102 | 0.000 | 0.103 | 0.377 | Precision Study: NG Results ª Neisseria gonorrhoeae (NG) concentrations were targeted approximately to 2000 CFU/assay in members 1, 7, and 10; to 20 to 50 CFU/assay in members 2, 3, 11, and 12. Members 9 and 15 were targeted to 0.1 CFU/assay, below the claimed assay LOD. Members 4, 6, 8, 13, and 14 did not contain any NG organisms. b Invalid replicates were excluded from the analysis. 6 For member 9, analysis of all replicates with a cycle number (n=147), including those beyond the assay cutoff, resulted in a total SD of 1.156. For member 15, analysis of all replicates with a cycle number (n=138), including those beyond the assay cutoff, resulted in a total SD of 1.201. d The total variability contains within-run, between-lot, and between-site variability. {18}------------------------------------------------ #### 3.15 Summary of Clinical Studies Performance characteristics of the Abbott RealTime CT/NG assay were established in a multi-center clinical study conducted in the United States. Specimens were collected from subjects at 16 geographically diverse sites that included physician private practices, public and private STD clinics, and a hospital emergency room. A total of 3,832 male and female, asymptomatic and symptomatic subjects were enrolled. Study subjects were classified as symptomatic if the subject reported STD-related symptoms. Specimens collected from each female subject included urine, endocervical swabs, self-collected vaginal swab, and clinician-collected vaginal swabs. Specimens collected from each male subject included urine and urethral swabs. Specimen testing methods included the Abbott RealTime CT/NG assay, two commercially available nucleic acid amplification tests (NAAT) for CT and NG, and culture for NG. The NAATs and the NG culture were used as reference assays in the clinical study. '. For females, self-collected vaginal swab and urine specimens were collected first, followed by endocervical swab for culture. Remaining swab specimen collection was randomized to minimize bias. For males, urethral swab for culture was collected first. Remaining swab specimen collection was randomized to minimize bias. Urine specimen was collected after the swab specimens. For each subject, a patient infected status was determined based on the combined results from the reference assays. A female subject was categorized as infected for CT or NG if a minimum of two positive results (at least one from each reference NAAT) was reported. For CT, female subjects with positive results on both reference urine specimens and negative results on all three reference swab specimens (clinician-collected vaginal swab from NAAT 1 and endocervical swab specimens from both reference assays) were categorized as infected for urine and not infected for swab specimens. A male subject was categorized as infected for CT or NG if a minimum of two positive results was reported. If the reference NG culture assay result was positive, the subject was categorized as infected regardless of NAAT results. Admin. Sec. 3 510(k) Summary Page 19 of 44 {19}------------------------------------------------ A female subject was categorized as not infected with CT or NG if at least one of the reference NAATs reported negative results for all sample types and if the NG culture assay result was negative. A male subject was categorized as not infected with CT or NG if a total of at least two negative results were reported by the reference NAATs and if the NG culture assay result was negative. If patient infected status could not be determined due to missing and/or indeterminate results from the reference assays, the subject was excluded from the analysis. Patient infected status could not be determined for 4 subjects for CT and 7 subjects for NG. Tables 3.10 through 3.28 summarize the clinical trial data. Abbott RealTime CT/NG test results were compared to the patient infected status for calculation of assay sensitivity and specificity. A total of 6,555 CT and 6,569 NG results were used in the analysis. The results were analyzed by gender, sample type, and the presence of symptoms. The overall sensitivity and specificity for CT was 95.2% and 99.3%, respectively. The overall sensitivity and specificity for NG was 97.5% and 99.7%, respectively. Sensitivity and specificity for CT for female subjects and male subjects are presented in Tables 3.10 and 3.11, respectively. Sensitivity and specificity for NG for female subjects and male subjects are presented in Tables 3.12 and 3.13, respectively. A comparison of patient infected status, individual test results from the reference assays and Abbott RealTime CT/NG assay was performed. CT results for infected and noninfected female subjects are presented in Tables 3.14 and 3.15, and for infected and noninfected male subjects in Tables 3.16 and 3.17. NG results for infected and non-infected female subjects are presented in Tables 3.18 and 3.19, and for infected and non-infected male subjects in Tables 3.20 and 3.21. The prevalence of CT and NG in this study was dependent on several factors including age, gender, clinic type, and the method of testing. The prevalence per collection site determined by the Abbott RealTime CT/NG assay for endocervical swab specimens is presented in Table 3.22, for clinician-collected and self-collected vaginal swab specimens Admin. Sec. 3 510(k) Summary Page 20 of 44 {20}------------------------------------------------ is presented in Table 3.23; for female urine specimens in Table 3.24; and for male urethral swab and male urine specimens in Tables 3.25 and 3.26, respectively. The Positive and Negative Predictive Values (PPV and NPV) were calculated using hypothetical prevalence rates and the Abbott RealTime CT/NG assay sensitivity and specificity determined from the clinical study. The overall sensitivity and specificity for CT was 95.2% and 99.3%, respectively. The overall sensitivity and specificity for NG was 97.5% and 99.7%, respectively. Estimates of the PPV and NPV for the Abbott RealTime CT/NG assay are presented in Table 3.27 for CT and Table 3.28 for NG. . {21}------------------------------------------------ hlamydia trachomatis Clinical Sensitivity and Specificit | Specimen | Symptoms | n | True Pos | False Pos | True Neg | False Neg | Sensitivity (95% C.I.) | Specificity (95% C.I.) | |---------------------|--------------|-----|----------|-----------|----------|-----------|------------------------|------------------------| | Endocervical Swab | Symptomatic | 616 | 60 | 1 | 551 | 4 | 93.8 (84.8, 98.3) | 99.8 (99.0, 100.0) | | Clinician-Collected | Symptomatic | 615 | 63 | 0 | 551 | 1 | 98.4 (91.6, 100.0) | 100.0 (99.3, 100.0) | | Vaginal Swab | Asymptomatic | 594 | 35 | 4 | 554 | 1 | 97.2 (85.5, 99.9) | 99.3 (98.2, 99.8) | | Self-Collected | Symptomatic | 587 | 62 | 6 | 518 | 1 | 98.4 (91.5, 100.0) | 98.9 (97.5, 99.6) | | Vaginal Swab | Asymptomatic | 586 | 36 | 5 | 544 | 1 | 97.3 (85.8, 99.9) | 99.1 (97.9, 99.7) | | Urine | Symptomatic | 737 | 73 | 2 | 655 | 7 | 91.3 (82.8, 96.4) | 99.7 (98.9, 100.0) | | | Asymptomatic | 686 | 43 | 2 | 638 | 3 | 93.5 (82.1, 98.6) | 99.7 (98.9, 100.0) | Abbott May 2010 Admin Sec 3_510(k) Summary_mw019 Admin. Sec. 3 510(k) Summary Page 22 of 44 {22}------------------------------------------------ hlamydia trachomatis Clinical Sensitivity and Specificit | Abbott RealTime CT/NG // multi-Collect Specimen Collection Kit | |----------------------------------------------------------------| | May 2010 | | Admin Sec 3_510(k) Summary_mw019 | Admin. Sec. 3 510(k) Summary Page 23 of 44 | Specimen | Symptoms | n | True<br>Pos | False<br>Pos | True<br>Neg | False<br>Neg | Sensitivity (95% C.I.) | Specificity (95% C.I.) | |------------------|--------------|-----|-------------|--------------|-------------|--------------|------------------------|------------------------| | Urethral<br>Swab | Symptomatic | 669 | 128 | 9 | 523 | 9 | 93.4<br>(87.9, 97.0) | 98.3<br>(96.8, 99.2) | | Urine | Symptomatic | 822 | 171 | 6 | 637 | 8 | 95.5<br>(91.4, 98.1) | 99.1<br>(98.0, 99.7) | | | Asymptomatic | 643 | 84 | 4 | 552 | 3 | 96.6<br>(90.3, 99.3) | 99.3<br>(98.2, 99.8) | {23}------------------------------------------------ leisseria gonorrhoeae Clinical Sensitivity and Specifici Female Specimens Abbott RealTime CT/NG // multi-Collect Specimen Collection Kit May 2010 ・ Admin Sec 3_510(k) Sunmary_mw019 · · Admin. Sec. 3 510(k) Summary Page 24 of 44 . {24}------------------------------------------------ sitivity Clinical Sen able 3.13 | Neisseria gonorrhoeae Clinical Sensitivity and Specificity | |------------------------------------------------------------| | Male Specimens | | Abbott RealTime CT/NG // multi-Collect Specimen Collection Kit | |----------------------------------------------------------------| | May 2010 | | Admin Ser 3 510(k) Summary mw019 | Urethra Swab (98.0, 99.7) 99.2 (96.3, 99.7) 98.7 ಕ 587 ഗ 228 823 Symptomatic Urine (99.4, 100.0 100.0 (71.5, 100.0) 100.0 0 632 0 l | 643 symptomati (97.6, 99.7 99.0 (97.1, 100.0) 99.5 I 482 S 188 676 Symptomatic Specificity (95% C.) Sensitivity (95% C.I. iləsi Bənə əsasən Esma Şəsi Əni J so astronomic so fi u Symptoms Specimen Admin. Sec. 3 510(k) Summary Page 25 of 44 Admin. Sec. 3 0(k) Summary Page 25 of 44 {25}------------------------------------------------ | | NAAT 1 | | | NAAT 2 | | RealTime CT/NG | | | | | No. of Subjects | | | |--|--------|-----|----|--------|----|----------------|-----|-----|----|-------------------------------|------------------------------|-------|--| | | E | CCV | FU | E | FU | E | CCV | SCV | FU | Symp…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...