The AmpliVue® Bordetella Assay is an in vitro diagnostic test for the qualitative detection of Bordetella pertussis nucleic acids isolated from nasopharyngeal swab specimens obtained from patients suspected of having respiratory tract infection attributable to Bordetella pertussis. The AmpliVue® Bordetella Assay utilizes helicase-dependent amplification (HDA) of the insertion sequence IS481 and a self-contained disposable amplification detection device that allows for manual evaluation of assay results. The IS481 sequence can also be found in strains of other organisms (i.e., B. holmesii and B. bronchiseptica). B. holmesii infection may cause clinical illness similar to B. pertussis, and mixed outbreaks involving both B. pertussis and B. holmesii infection have been reported. Additional testing should be performed if necessary to differentiate B. holmesii and B. pertussis. B. bronchiseptica is a rare cause of infection in humans. When clinical factors suggest that B. pertussis may not be the cause of respiratory infection, other clinically appropriate investigation(s) should be carried out in accordance with published guidelines. Negative results for the AmpliVue® Bordetella Assay do not preclude B. pertussis infection and positive results do not rule out co-infection with other respiratory pathogens. Results from the AmpliVue® Bordetella Assay should be used in conjunction with information obtained during the patient's clinical evaluation as an aid in diagnosis of Bordetella pertussis infection and should not be used as the sole basis for treatment or other patient management decisions. The AmpliVue® Bordetella Assay is intended for use in hospital, reference or state laboratory settings. The device is not intended for point-of-care use.
Device Story
The AmpliVue Bordetella Assay is an in vitro diagnostic test for detecting B. pertussis DNA in nasopharyngeal swabs. The device uses helicase-dependent amplification (HDA) to target the IS481 insertion sequence. The workflow involves manual sample processing: heating the sample in process buffer at 95°C, adding it to a reaction tube containing lyophilized HDA reagents (isothermal polymerase, helicase, single-stranded binding protein), and incubating at 64°C for 60 minutes. The reaction tube is then transferred to a self-contained disposable cassette. Inside the cassette, a mechanical activation step releases running buffer, wicking the amplicon onto a lateral flow strip. Results are visualized as colored lines (T2 for B. pertussis, C for process control) on the strip. The device is operated by laboratory personnel in hospital, reference, or state labs. It provides qualitative results to aid clinicians in diagnosing B. pertussis infection, though it does not differentiate from B. holmesii or B. bronchiseptica, which may also contain the IS481 target. Results must be used alongside clinical evaluation.
Clinical Evidence
Clinical study conducted at four US sites with 842 fresh nasopharyngeal swab specimens. Performance compared to a composite reference method (two validated real-time PCR assays followed by bi-directional sequencing). Results for 836 evaluable specimens: Sensitivity 97.0% (64/66; 95% CI: 89.6%-99.2%) and Specificity 98.1% (755/770; 95% CI: 96.8%-98.8%).
Technological Characteristics
The device utilizes isothermal Helicase-Dependent Amplification (HDA) of the IS481 target sequence. It is a manual, non-instrumented, self-contained disposable cassette system. Detection is via lateral flow strip with anti-FAM and anti-DNP antibodies and streptavidin-conjugated color particles. The assay requires a heat block for sample lysis (95°C) and amplification (64°C).
Indications for Use
Indicated for qualitative detection of Bordetella pertussis nucleic acids in nasopharyngeal swab specimens from patients of all ages suspected of respiratory tract infection attributable to B. pertussis. For prescription use in hospital, reference, or state laboratories.
Regulatory Classification
Identification
A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B; (2) Influenza A subtype H1 and Influenza A subtype H3; (3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B; (4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus; (5) Human Metapneumovirus; (6) Rhinovirus; and (7) Adenovirus.
Special Controls
*Classification.* Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.
Predicate Devices
illumigene® Pertussis DNA Amplification Assay (k133673)
K133673 — ILLUMIGENE PERTUSIS DNA AMPLIFICATION ASSAY, ILLUMIGENE PERTUSSIS EXTERNAL CONTROL KIT, ILLUMIPRO-10 AUTOMATED ISOTHERMA · Meridian Bioscience, Inc. · Mar 25, 2014
K152285 — illumigene Pertussis DNA Amplification Assay · Meridian Bioscience, Inc. · Nov 10, 2015
Submission Summary (Full Text)
{0}
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K143206
B. Purpose for Submission:
To obtain a substantial equivalence determination for the Amplivue® Bordetella Assay.
C. Measurand:
Insertion sequence IS481 of *Bordetella pertussis*.
D. Type of Test:
The Amplivue® Bordetella Assay is a helicase-dependent amplification *in vitro* diagnostic test for the qualitative detection *Bordetella pertussis* nucleic acids isolated from nasopharyngeal swab specimens obtained from patients suspected of having respiratory tract infection attributable to *Bordetella pertussis*.
E. Applicant:
Quidel Corporation
F. Proprietary and Established Names:
Amplivue® Bordetella Assay
G. Regulatory Information:
1. Regulation section:
21 CFR 866.3980 – Respiratory viral panel multiplex nucleic acid assay
2. Classification:
Class II
3. Product code:
OZZ – *Bordetella pertussis* Nucleic Acid Amplification Assay System
4. Panel:
83- Microbiology
{1}
H. Intended Use:
1. Intended use(s):
The AmpliVue® Bordetella Assay is an in vitro diagnostic test for the qualitative detection of *Bordetella pertussis* nucleic acids isolated from nasopharyngeal swab specimens obtained from patients suspected of having respiratory tract infection attributable to *Bordetella pertussis*.
The AmpliVue® Bordetella Assay utilizes helicase-dependent amplification (HDA) of the insertion sequence IS481 and a self-contained disposable amplification detection device that allows for manual evaluation of assay results. The IS481 sequence can also be found in strains of other organisms (i.e., *B. holmesii* and *B. bronchiseptica*). *B. holmesii* infection may cause clinical illness similar to *B. pertussis*, and mixed outbreaks involving both *B. pertussis* and *B. holmesii* infection have been reported. Additional testing should be performed if necessary to differentiate *B. holmesii* and *B. pertussis*. *B. bronchiseptica* is a rare cause of infection in humans. When clinical factors suggest that *B. pertussis* may not be the cause of respiratory infection, other clinically appropriate investigation(s) should be carried out in accordance with published guidelines.
Negative results for the AmpliVue® Bordetella Assay do not preclude *B. pertussis* infection and positive results do not rule out co-infection with other respiratory pathogens. Results from the AmpliVue® Bordetella Assay should be used in conjunction with information obtained during the patient's clinical evaluation as an aid in diagnosis of *Bordetella pertussis* infection and should not be used as the sole basis for treatment or other patient management decisions.
The AmpliVue® Bordetella Assay is intended for use in hospital, reference or state laboratory settings. The device is not intended for point-of-care use.
2. Indication(s) for use:
Same as intended use.
3. Special conditions for use statement(s):
For *in vitro* diagnostic use only
For prescription use only
4. Special instrument requirements:
Heat blocks capable of $95^{\circ}\mathrm{C} \pm 2^{\circ}\mathrm{C}$ and $64^{\circ}\mathrm{C} \pm 2^{\circ}\mathrm{C}$.
{2}
I. Device Description:
The AmpliVue® Bordetella Assay combines simple sample processing, an isothermal amplification technology named helicase-dependent amplification (HDA), and a self-contained disposable amplicon detection device, for the detection of *Bordetella pertussis* from nasopharyngeal swabs.
Patient samples are collected using a nasopharyngeal swab and placed into a liquid medium. Fifty microliters (50 μL) of the sample are then transferred to a process buffer that is provided with the kit and mixed. The Process Buffer tubes are heated at 95 °C for 10 minutes. Fifty microliters (50 μL) of the Process Buffer containing sample is added to a reaction tube containing lyophilized mix of HDA reagents. Included in the reaction mix are the isothermal polymerase, helicase and single stranded binding protein. After completion of the HDA reaction the reaction tube is transferred to the amplicon cartridge containing the running buffer. The amplicon cartridge is closed and inserted into the detection chamber. The detection chamber is activated by depressing the detection chamber handle. Upon activation, the reservoir containing the running buffer and the 0.2 mL tube containing the amplicon is punctured and the solutions are wicked to the lateral flow strip.
Materials Provided:
- 16 Tests per Kit
| Symbol | Component | Quantity | Storage |
| --- | --- | --- | --- |
| 1 | Detection Cassettes | 16/kit | 2° to 30°C |
| 2 | Process Buffer | 16 tubes/kit 1.45mL | 2° to 30°C |
| 3 | Reaction Tubes | 16 tubes/kit | 2° to 8°C |
| 4 | Amplicon Cartridge | 16/kit | 2° to 30°C |
Materials required but not provided:
- External controls for *Bordetella pertussis* (e.g. Quidel Molecular Bordetella Control Set #M117, which contains positive and negative controls, serves as an external processing and extraction control)
- Sterile DNAse-free filter-blocked or positive displacement micropipettor tips
- Micropipettor
- Stopwatch or timer
- Scissors or a blade
- Micro tube tray
- Heat block capable of 95° C ± 2° C temperature
- Heat block with heated lid capable of 64° ± 2° C temperature
- Thermometer
J. Substantial Equivalence Information:
1. Predicate device name(s):
{3}
illumigene® Pertussis DNA Amplification Assay
2. Predicate 510(k) number(s):
k133673
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | AmpliVue® Bordetella Assay (k143206) | illumigene® Pertussis DNA Amplification Assay (k133673) |
| Intended Use | The AmpliVue® Bordetella Assay is an in vitro diagnostic test for the qualitative detection of Bordetella pertussis nucleic acids isolated from nasopharyngeal swab specimens obtained from patients suspected of having respiratory tract infection attributable to Bordetella pertussis.
The AmpliVue® Bordetella Assay utilizes helicase-dependent amplification (HDA) of the insertion sequence IS481 and a self-contained disposable amplification detection device that allows for manual evaluation of assay results. The IS481 sequence can also be found in strains of other organisms (i.e., B. holmesii and B. bronchiseptica). B. holmesii infection may cause clinical illness similar to B. pertussis, and mixed outbreaks involving both B. pertussis and B. holmesii infection have been reported. Additional testing should be performed if necessary to differentiate B. holmesii and B. pertussis. B. bronchiseptica is a rare cause of infection in humans. When clinical factors suggest that B. pertussis may | The illumigene® Pertussis DNA Amplification Assay, performed on the illumipro-10™, is a qualitative in vitro diagnostic test for the direct detection of Bordetella pertussis in human nasopharyngeal swab samples taken from patients suspected of having respiratory tract infection attributable to Bordetella pertussis.
The illumigene Pertussis assay utilizes loop-mediated isothermal DNA amplification (LAMP) technology to detect Bordetella pertussis by targeting the IS481 insertional element of the Bordetella pertussis genome. The IS481 insertional element can also be found in Bordetella holmesii and |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | AmpliVue® Bordetella Assay (k143206) | illumigene® Pertussis DNA Amplification Assay (k133673) |
| | not be the cause of respiratory infection, other clinically appropriate investigation(s) should be carried out in accordance with published guidelines.
Negative results for the AmpliVue® Bordetella Assay do not preclude *B. pertussis* infection and positive results do not rule out co-infection with other respiratory pathogens. Results from the AmpliVue® Bordetella Assay should be used in conjunction with information obtained during the patient's clinical evaluation as an aid in diagnosis of *Bordetella pertussis* infection and should not be used as the sole basis for treatment or other patient management decisions. | *Bordetella bronchiseptica* strains. Respiratory infection with *Bordetella pertussis*, *Bordetella holmesii* or *Bordetella bronchiseptica* may yield positive test results in IS481 assays. *B. holmesii* infection may cause clinical illness similar to *B. pertussis*, and mixed outbreaks involving both *B. pertussis* and *B. holmesii* infection have been reported. Additional testing should be performed if necessary to differentiate *B. holmesii* and *B. pertussis*. *B. bronchiseptica* is a rare cause of infection in humans. When clinical factors suggest that *B. pertussis* may not be the cause of respiratory infection, other clinically appropriate investigation(s) should |
{5}
| Similarities | | |
| --- | --- | --- |
| Item | AmpliVue® Bordetella Assay (k143206) | illumigene® Pertussis DNA Amplification Assay (k133673) |
| | | be carried out in accordance with published guidelines.
Negative results for the illumigene Pertussis DNA Amplification Assay do not preclude Bordetella pertussis infection and positive results do not rule out co-infection with other respiratory pathogens.
Results from the illumigene Pertussis assay should be used in conjunction with information obtained during the patient's clinical evaluation as an aid in diagnosis of Bordetella pertussis infection and should not be used as the sole basis for treatment or other patient management decisions.
illumigene Pertussis is intended for use in hospital, reference or |
{6}
| Similarities | | |
| --- | --- | --- |
| Item | AmpliVue® Bordetella Assay (k143206) | illumigene® Pertussis DNA Amplification Assay (k133673) |
| | | state laboratory settings. The device is not intended for point-of-care use. |
| Sample Types | Nasopharyngeal swab specimens obtained from patients suspected of having respiratory tract infection attributable to Bordetella pertussis | Same |
| Sample Heat Lysis | Manual | Same |
| Target Sequence Detected | Bordetella pertussis IS481 insertion element | Same |
| Differences | | |
| --- | --- | --- |
| Item | AmpliVue® Bordetella Assay (k143206) | illumigene® Pertussis DNA Amplification Assay (k133673) |
| DNA Amplification Technology | Helicase-dependent amplification (HDA); self- contained | Loop-Mediated Isothermal Amplification (LAMP); self- contained and automated |
| Self-Contained System Assay after sample preparation | No | Yes |
| Detection Technique | Manual | Automated |
| Instrument | None | illumipro-10™ |
{7}
| Differences | | |
| --- | --- | --- |
| Item | AmpliVue® Bordetella Assay (k143206) | illumigene® Pertussis DNA Amplification Assay (k133673) |
| Testing Time | 85 - 90 minutes | 60 -70 minutes |
## K. Standard/Guidance Document Referenced (if applicable):
Not applicable.
## L. Test Principle:
The AmpliVue® Bordetella Assay detects *Bordetella pertussis* DNA isolated from nasopharyngeal swab specimens obtained from symptomatic patients suspected of having respiratory tract infection attributable to *Bordetella pertussis*. The assay consists of three major steps: 1) specimen preparation, 2) isothermal Helicase-Dependent Amplification (HDA) of a target sequence of B. pertussis, and 3) detection of the amplified DNA by target-specific hybridization probes via a colorimetric reaction on a lateral-flow strip which is embedded in a self-contained disposable cassette to prevent amplicon contamination.
Patient samples are collected using a nasopharyngeal swab and placed into a liquid medium. Fifty microliters (50 μL) of the sample are then transferred to a process buffer that is provided with the kit and mixed. The Process Buffer tubes are heated at 95 °C for 10 minutes. Fifty microliters (50 μL) of the Process Buffer containing sample is added to a reaction tube containing lyophilized mix of HDA reagents.
A HDA reaction is carried out in the Reaction Tube which contains lyophilized HDA reagents, dNTPs, primers and probes. Incubation at 64°C for 60 minutes results in isothermal amplification of the target sequence by *B. pertussis* specific primers. The amplified DNA is detected by a set of specific detection probes included in the Reaction Tube: *B. pertussis* target hybridizes to two specific probes labeled with Biotin (BioTEG) and 6-carboxyfluorescein (FAM). A competitive process control (PRC) is included in the Lysis Tube to monitor specimen processing and inhibitory substances in clinical samples, reagent failure or device failure. The PRC target is amplified and hybridizes to the PRC specific probes labeled with Biotin (BioTEG) and 2,4-dinitrophenyl (DNP).
Following completion of the HDA reaction, the Reaction Tube is transferred to a proprietary Cassette for detection. The Cassette is comprised of two components: 1) an Amplicon Cartridge that holds the running buffer and the 0.2 mL Reaction Tube and 2) the Detection Chamber which houses the Amplicon Cartridge and an embedded vertical-flow DNA detection strip. The DNA detection strip is coated with anti-FAM and anti-DNP antibodies. Once the Cassette is closed, a razor blade and plastic pin located at the bottom of the Cassette opens the Reaction Tube and running buffer bulb, resulting in the release of their
{8}
contents. The contents flows through a fiberglass paper connected to the DNA detection strip that is attached to a fiberglass pad pre-loaded with streptavidin-conjugated color particles. The B. pertussis amplicon with biotin- and FAM-labeled probes is captured by the anti-FAM antibodies at the test (T2) line, and the Process Control amplicon with biotin- and DNP-labeled probes is captured by the anti-DNP antibodies at the control (C) line. The streptavidin-conjugated color particles bind to the biotin in the probe-amplicon hybrid and the test results are displayed in the Cassette window as colored T2 and/or C lines that are visible to the naked eye.
Detection of B. pertussis is reported when the T2 line is visible through the detection window of the Cassette. No detection of B. pertussis is reported when only the C line is displayed. The assay is regarded as invalid when none of the lines are displayed.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Precision studies for the AmpliVue® Bordetella Assay were conducted by two (2) operators three times (3x) per day for twelve (12) days with a panel of three (3) simulated samples that include moderate positive (5x LoD) and low positive (2x LoD), and Bordetella pertussis (BP) negative. The study results are acceptable. The results are shown in the Table I below.
| TABLE I: PRECISION | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Category | Operator #1 | | Operator #2 | | Combined | |
| | #expected results/# tested | % Agreement | #expected results/# tested | % Agreement | #expected results/# tested | % Agreement |
| BP Low Positive (4,716 cfu/mL) | 36/36 | 100% | 36/36 | 100% | 72/72 | 100% |
| BP Moderate Positive (11,790 cfu/mL) | 36/36 | 100% | 36/36 | 100% | 72/72 | 100% |
| BP Negative | 36/36 | 100% | 36/36 | 100% | 72/72 | 100% |
The reproducibility of the AmpliVue® Bordetella Assay was evaluated at three (3) laboratory sites (two external, one in-house). Reproducibility was assessed using a panel of three (3) simulated samples that include moderate positive and low positive (5x and 2x LoD), and Bordetella pertussis (BP) negative. The panels and controls were processed and tested on the AmpliVue® Bordetella Assay at each site by 2 operators for 5 non-consecutive days (2 operators x 3 replicates x 5 days x 3 sites = 90 results per concentration). The LoD values were based on the values obtained in the LoD study. The results are shown in the Table II below.
{9}
| TABLE II: REPRODUCIBILITY | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Category | SITE | | | | | | Overall Percent Agreement | 95% Confidence Interval |
| | Site #1 | | Site #2 | | Site #3 | | | |
| | #expected results/# tested | % Agreement | #expected results/# tested | % Agreement | #expected results/# tested | % Agreement | | |
| BP Low Positive (4,716 cfu/mL) | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 | 95.9% to 100% |
| BP Moderate Positive (11,790 cfu/mL) | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 | 95.9% to 100% |
| BP Negative | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 | 95.9% to 100% |
| BP Positive Control | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 | 95.9% to 100% |
| BP Negative Control | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 | 95.9% to 100% |
The results suggest that there are no significant differences between different users and different sites on different days. Reproducibility studies are acceptable.
b. Linearity/assay reportable range:
Not applicable – This assay is qualitative.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
Not applicable. This assay is qualitative.
Stability:
A study was performed to determine the stability of samples in nasal matrix that are used in the AmpliVue® Bordetella Assay. A contrived positive sample was prepared by spiking a pooled negative clinical sample with a bacterial stock of *Bordetella pertussis* that was prepared and frozen during the LoD study at a concentration of 2x LoD. The contrived positive sample was tested with AmpliVue Bordetella Validation Lot reagents using the AmpliVue Bordetella assay workflow, to establish a 0 hour (h) time point. The contrived positive sample was aliquoted into four tubes and stored at 2°C to 8°C. An aliquot of the specimen was tested at 24 h, 48 h, 72 h, and 96 h time points. A total of 15 positive samples were assayed for each time point along with a positive and negative control.
Based on this study, clinical samples are expected to be stable between 2°C to 8°C for
{10}
up to 96 hours.
## Controls:
Controls (Quidel Molecular Bordetella Control Set #M117, which contains positive and negative controls, serves as an external processing and extraction control) were run on the AmpliVue® Bordetella Assay each day of testing. All Bordetella pertussis positive controls were detected accurately (100%, 111/111). All Bordetella pertussis negative controls were detected accurately (100%, 111/111) in the clinical study. These controls are described as follows:
a. The process control is used to monitor sample processing, to detect HDA inhibitory specimens and to confirm the integrity of assay reagents and cassette detection. The process control is included in the Lysis Buffer tube.
b. The external positive control may be treated as a patient specimen. The control should be sampled and tested as if it were a swab specimen and processed as described in the Assay Procedure. The external positive control is intended to monitor substantial reagent and cassette failure.
c. The external negative control may be treated as a patient specimen. The control should be sampled and tested as if it were a swab specimen and processed as described in the Assay Procedure. The external negative control is used to detect reagent or environmental contamination (or carry-over) by B. pertussis DNA or amplicon.
## d. Detection limit:
The analytical sensitivity (limit of detection or LoD) of the AmpliVue® Bordetella Assay was determined using quantified (CFU/mL) cultures of two (2) Bordetella pertussis bacterial stocks, BP A639 and E431 serially diluted in negative nasal matrix. The LoD is defined as the lowest concentration at which 95% of all replicates tested positive.
The two bacterial strains of Bordetella pertussis (BP) were freshly grown. The organisms were cultured onto Bordet-Gengou (BG) Blood Agar plates from commercially obtained glycerol stock. The resulting colonies were harvested and collected in Phosphate Buffered Saline (PBS), and the density of the bacterial cell suspension was determined using the McFarland assay. After a cell suspension of 0.25 McFarland units was established, the bacteria were serially diluted in PBS. BP CFU/mL concentrations were confirmed by plating onto fresh BG Blood Agar and colony counts.
After the LoD was determined, 20 replicates of the LoD dilutions of BP A639 and E431 were tested using two additional validation lots of buffers and amplification reagents. If 19 or 20 of 20 replicates of the dilution showed positive results for BP, the three lots were deemed equivalent. For the three validation lots of reagents to be
11
{11}
deemed equivalent, the LoD across the three reagent lots must not exceed three halving concentrations (i.e. must not exceed 8 fold).
The assay LoD for Bordetella pertussis is 3.93 CFU/assay or 2,358 CFU/mL (sample input), which was demonstrated on three Validation Lots. The LoD study results are shown in Table III below.
| Table III: LoD for Bordetella pertussis | | |
| --- | --- | --- |
| Strain | Strain ID | CFU/mL |
| Bordetella pertussis strain 1 | A639 | 2,358 |
| Bordetella pertussis strain 2 | E431 | 760.8 |
These study results are acceptable.
e. Analytical Reactivity (Inclusivity):
The reactivity of the AmpliVue® Bordetella Assay was evaluated against an additional six (6) strains of Bordetella pertussis. The testing was performed at concentrations near the level of detection (LoD).
Each strain was tested as three replicates in the AmpliVue® Bordetella Assay. The study was performed in multiple experiments of no more than 14 assays per experiment. For each experiment, three replicates of up to four strains were performed, along with a positive and a negative run control. All six (6) strains were detected by the AmpliVue® Bordetella Assay in this study at the 1x LoD concentration (3.93 CFU/assay or 2,358 CFU/mL). The inclusivity study results and the final organism concentrations tested are shown in Table IV below.
| Table IV: Bordetella pertussis Inclusivity | | |
| --- | --- | --- |
| Bacterial Strain | Concentration CFU/mL | Strain Detected |
| 9797 | 2,358 | Yes |
| 9340 | 2,358 | Yes |
| BAA-1335 | 2,358 | Yes |
| BAA-589 | 2,358 | Yes |
| 51445 | 2,358 | Yes |
| 10380 | 2,358 | Yes |
An in silico BLAST analysis of primers used in the AmpliVue® Bordetella Assay against twenty-three (23) Bordetella strains in the NCBI database was done to further demonstrate inclusivity. The Bordetella primers matched 1068 Bordetella sequences, including eight (8) B. pertussis complete genome/WGS sequences and 25 gene/CDS
{12}
sequences. The primers and probes all have 3-4 mismatches in less than 10% of all sequences, indicating that the region is well conserved. The primers, however, did not match one WGS strain, B. pertussis B1917. In addition, the primers will also detect B. bronchiseptica and B. holmesii strains. The primers did not show evidence of cross-reactivity with other organisms.
f. Cross Reactivity:
A study was performed to evaluate the cross-reactivity of the AmpliVue® Bordetella Assay with seventy-nine (79) other microorganisms potentially found in specimens collected to test for Bordetella pertussis (BP) infection. Cross-reactive microorganism was tested at clinically relevant levels of viruses (10⁵ PFU/mL) and bacteria (10⁶ CFU/mL) in the device. The organisms and their concentrations included in the cross-reactivity study are shown in Table V below.
| Table V: Strains Included in Cross-Reactivity | | |
| --- | --- | --- |
| Strain | Strain ID | CFU/mL |
| Acinetobacter baumanii | 2.90 x10⁶ | CFU/mL |
| Arcanobacterium haemolyticum | 1.15 x10⁶ | CFU/mL |
| Bacteroides fragilis | 1.19 x10⁶ | CFU/mL |
| Bordetella avium | 3.85 x10⁶ | CFU/mL |
| Bordetella bronchiseptica | 9.45 x10⁶ | CFU/mL |
| Bordetella bronchiseptica | 1.17 x10⁶ | CFU/mL |
| Bordetella bronchiseptica (ATCC 4617) | 7.74 x10⁶ | CFU/mL |
| Bordetella bronchiseptica | 1.97 x10⁶ | CFU/mL |
| Bordetella hinzii | 1.40 x10⁶ | CFU/mL |
| Bordetella holmesii (ZeptoMetrix | 3.83 x10⁶ | CFU/mL |
| Bordetella holmesii (ATCC 51541) | 4.10 x10⁶ | CFU/mL |
| Bordetella holmesii (ATCC 700053) | 4.70 x10⁶ | CFU/mL |
| Bordetella holmesii (ATCC 700052) | 4.00 x10⁶ | CFU/mL |
| Bordetella parapertussis A747 | 1.00 x10⁶ | CFU/mL |
| Bordetella petrii | 6.26 x10⁶ | CFU/mL |
| Bordetella trematum | 9.24 x10⁶ | CFU/mL |
| Burkholderia cenocepacia | 2.35 x10⁶ | CFU/mL |
| Burkholderia cepacia | 2.52 x10⁶ | CFU/mL |
| Burkholderia multivorans | 1.95 x10⁶ | CFU/mL |
| Burkholderia thailandensis | 3.95 x10⁶ | CFU/mL |
| Chlamydia trachomatis | 7.83 x10⁶ | CFU/mL |
| Chlamydophila pneumoniae | 2.10 x10⁶ | CFU/mL |
| Corynebacterium diptheriae | 4.00 x10⁶ | CFU/mL |
{13}
| Enterobacter aerogenes | 1.31 x10^{6} | CFU/mL |
| --- | --- | --- |
| Enterococcus faecalis | 3.45 x10^{6} | CFU/mL |
| Escherichia coli | 8.42 x10^{6} | CFU/mL |
| Fusobacterium necrophorum | 3.10 x10^{6} | CFU/mL |
| Haemophilus influenzae | 2.13 x10^{6} | CFU/mL |
| Klebsiella pneumoniae | 1.61 x10^{6} | CFU/mL |
| Lactobacillus acidophilus | 2.00 x10^{6} | CFU/mL |
| Lactobacillus plantarum | 7.97 x10^{6} | CFU/mL |
| Legionella pneumophilia | 1.76 x10^{6} | CFU/mL |
| Moraxella catarrhalis | 9.90 x10^{6} | CFU/mL |
| Morganella morganii | 1.57 x10^{6} | CFU/mL |
| Mycobacterium avium | 1.84 x10^{6} | CFU/mL |
| Mycobacterium tuberculosis | 1.80 x10^{6} | CFU/mL |
| Mycoplasma pneumoniae | 3.16 x10^{6} | CFU/mL |
| Neisseria gonorrhoeae | 2.45 x10^{6} | CFU/mL |
| Neisseria meningitidis | 7.07 x10^{6} | CFU/mL |
| Neisseria mucosa | 1.66 x10^{6} | CFU/mL |
| Parvimonas micra | 1.55 x10^{6} | CFU/mL |
| Proteus mirabilis | 1.06 x10^{6} | CFU/mL |
| Proteus vulgaris | 3.40 x10^{6} | CFU/mL |
| Pseudomonas aeruginosa | 2.60 x10^{6} | CFU/mL |
| Staphylococcus aureus (MRSA) | 7.10 x10^{6} | CFU/mL |
| Staphylococcus epidermidis | 2.14 x10^{6} | CFU/mL |
| Stenotrophomonas maltophilia | 1.90 x10^{6} | CFU/mL |
| Streptococcus pneumoniae | 1.00 x10^{6} | CFU/mL |
| Streptococcus pyogenes | 1.29 x10^{6} | CFU/mL |
| Streptococcus salivarius | 1.70 x10^{6} | CFU/mL |
| Candida albicans | 3.00 x10^{6} | CFU/mL |
| Adenovirus 31 | 3.55 x10^{5} | TCID_{50}/mL |
| Adenovirus 31 | 2.74 x10^{7} | DNA copies/mL |
| Coronavirus 229E | 1.51 x10^{6} | TCID_{50}/mL |
| Coronavirus NL63 | 1.41 x10^{5} | TCID_{50}/mL |
| Coronavirus OC43 | 8.51 x10^{6} | TCID_{50}/mL |
| Coxsackievirus B4 | 1.08 x10^{5} | TCID_{50}/mL |
| Coxsackievirus B5/10/2006 | 1.02 x10^{5} | TCID_{50}/mL |
| Echovirus 6 | 1.02 x10^{6} | TCID_{50}/mL |
| Echovirus 7 | 1.05 x10^{5} | TCID_{50}/mL |
14
{14}
| Echovirus 9 | 1.41 x105 | TCID50/mL |
| --- | --- | --- |
| Echovirus 11 | 1.51 x106 | TCID50/mL |
| Enterovirus 70 | 1.78 x106 | TCID50/mL |
| Enterovirus 71 | 4.17 x105 | TCID50/mL |
| Epstein-Barr Virus | 1.34 x106 | Virus particles/mL |
| HSV Type 1 (Mclnytre) | 6.65 x106 | TCID50/mL |
| HSV Type 2 (G) | 2.27 x106 | TCID50/mL |
| Influenza A/Mexico/4108/2009 | 2.88 x106 | Virus particles/mL |
| Influenza B/Florida/04/2006 | 2.82 x106 | Virus particles/mL |
| Measles virus | 1.95 x106 | TCID50/mL |
| Metapneumovirus A1 | 3.80 x106 | TCID50/mL |
| Mumps virus | 5.89 x106 | TCID50/mL |
| Parainfluenza Type 1 (#2) | 3.97 x106 | TCID50/mL |
| Parainfluenza Type 2 (Greer) | 3.15 x106 | TCID50/mL |
| Parainfluenza Type 3 (C234) | 2.56 x106 | TCID50/mL |
| Parainfluenza Type 4 (VR-1377) | 1.37 x106 | TCID50/mL |
| Respiratory Syncytial Virus | 1.15 x106 | TCID50/mL |
| Rhinovirus 1A | 1.26 x106 | TCID50/mL |
| Varicella Zoster Virus | 1.70 x106 | DNA copies/mL |
The Cross Reactivity study tested a panel of 79 microorganisms. This study determined that 1 of 4 Bordetella bronchiseptica strains (strain 4617) and 4 of 4 Bordetella holmesii strains tested were cross-reactive with the AmpliVue® BordetellaAssay. These results can be expected as $5\%$ of all Bordetella bronchiseptica strains and all Bordetella holmesii strains are known to carry the IS481 target sequence.
These cross-reactive results are noted in the intended use and limitation sections.
# g. Interference:
A panel of seventeen (17) chemical and biological substances potentially present in Bordetella pertussis infection specimens were evaluated for interference with the AmpliVue® Bordetella Assay. Each substance was tested in three replicates at concentrations which were medically significant in the presence and absence of near LOD (2x) levels of $B$ . pertussis in the AmpliVue® Bordetella Assay.
None of the substances tested were found to interfere with the AmpliVue® Bordetella Assay. The interference substances and their concentrations included in the interference study are shown in Table VI below.
{15}
| Table VI: Interference Substances | |
| --- | --- |
| Common Name | Test Concentration |
| Cepacol Sore Throat Lozenges | 5% w/v |
| Halls Cherry Menthol-Lyptus Cough Drops | 15% w/v |
| Children's Dimetapp | 15% v/v |
| Chloraseptic Sore Throat Lozenges | 10% w/v |
| Ricola Original Swiss Sugar-Free Herb Cough Suppressant Throat | 15% w/v |
| Sucrets Complete Lozenges - Vapor Cherry | 5% w/v |
| Mucin (Bovine Submaxillary Gland, Type I-S) | 5 mg/ml |
| Blood (human), EDTA anticoagulated | 5% v/v |
| Neo-Synephrine | 15% v/v |
| Afrin Nasal Spray Original | 15% v/v |
| Zicam Non-Drowsy Allergy Relief Nasal Gel | 5% v/v |
| Rite Aid Brand Saline Nasal Spray | 15% v/v |
| Zanamivir (Relenza) | 5 mg/ml |
| Tobramycin | 4 μg/ml |
| Mupirocin | 10 mg/ml |
| Oseltamivir Phosphate (Tamiflu) | 10 mg/ml |
| Ricola Original Swiss Sugar Free Herb Cough Suppressant Throat | 15% w/v |
h. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable.
b. Matrix comparison:
The compatibility of eight (8) different types of media with the AmpliVue Bordetella assay was determined testing four (4) strains each of pre-titered BP bacterial strains at 1x LoD concentrations. A diluted bacterial strain and one of the eight types of media (Tris EDTA, Molecular Grade Water, E-swab, M4, M4-RT, M5, 0.85% Saline) were added to three Process Buffer tubes and were tested in the AmpliVue Bordetella assay according to the established workflow using Validation Lot reagents
The study was performed in multiple assay runs, with up to four strains tested per assay run. Each experiment included a positive and a negative run control. Any media that resulted in BP- positive detection for 3 of 3 replicates was considered to be compatible with the AmpliVue Bordetella assay.
All four Bordetella pertussis strains were determined to be compatible with the 8 different media types when tested at the assay LoD. Based on these results, Tris
{16}
EDTA, Molecular Grade Water, E-swab, M4, M4-RT, M5, 0.85% Saline will be considered to be compatible with the AmpliVue Bordetella assay.
# 3. Clinical studies:
# a. Clinical Sensitivity:
Performance characteristics of the AmpliVue® Bordetella Assay was established in 2014 (April to August 2014) at four locations in the United States. Eight hundred forty two (842) fresh nasopharyngeal swab specimens were obtained from female and male patients suspected of having respiratory tract infection attributable to Bordetella pertussis. Specimens were collected and transported to each laboratory for testing with the AmpliVue® Bordetella Assay.
Clinical performance was based on comparison of the AmpliVue Bordetella Assay results to those obtained by Composite Reference Method that included two manufacturer validated, IS481-targeted real-time PCR assays (PCR1 and PCR2) followed by bi-directional sequencing from PCR positive specimens. The PCR1 and PCR2 assay protocols included 37 amplification cycles. Bi-directional sequencing was performed for all specimens producing amplicon prior to the end of 37-cycle amplification. Specimens were considered positive when bi-directional sequence sequencing results from either comparator PCR assay confirmed the presence of Bordetella pertussis amplicon. Specimens were considered negative when neither comparator PCR assay produced Bordetella pertussis amplicon at the end of the 37-cycles.
Eight hundred forty two (842) fresh nasopharyngeal swab specimens were tested as described above. Six (6) specimens $(0.7\%)$ were invalid (in both the initial and repeat test neither the T2 or control lines were detected) and have been removed from additional analysis. The table below details the comparison data of the AmpliVue Bordetella Assay and the Composite Reference Method for the remaining eight hundred thirty six (836) specimens.
| Table VII: Clinical Performance Data for the Amplivue® Bordetella Assay vs. Composite Reference Method | | | |
| --- | --- | --- | --- |
| All Sites | | | |
| Amplivue® Bordetella Assay | Composite Reference Culture | | |
| | Positive | Negative | Total |
| Positive | 64 | 15 | 79 |
| Negative | 2 | 755 | 757 |
| Total | 66 | 770 | 836 |
| Sensitivity: 97.0% (64/66) 95% CI (89.6%-99.2%)Specificity: 98.1% (755/770) 95% CI (96.8%-98.8%) | | | |
{17}
b. Clinical specificity:
See table above.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not Applicable.
5. Expected values/Reference range:
The prevalence of Bordetella pertussis detected with the AmpliVue® Bordetella Assay has been calculated for the combined sites based on the age of the patient. Six (6) specimens (0.7%) were invalid (in both the initial and repeat test neither the T2 or control lines were detected) and have been removed from the Expected Values table. All clinical specimens collected during this study were collected between April, 2014 and August 2014. Table VIII below presents the data for the remaining eight hundred forty two (842) specimens.
| Table VIII: Combined Study – Expected Values (N=836) | | | |
| --- | --- | --- | --- |
| Bordetella pertussis | | | |
| Age | Total # | Total Positive | Prevalence |
| ≤ 2 years | 137 | 8 | 5.8% |
| 3 to 12 years | 274* | 27 | 9.9% |
| 13 to 21 years | 145 | 30 | 20.7% |
| ≥ 22 years | 280** | 14 | 5.0% |
| * Four (4) specimens were invalid ** Two (2) specimens were invalid | | | |
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10, 21 CFR 801.109, and special controls.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.