K152285 · Meridian Bioscience, Inc. · OZZ · Nov 10, 2015 · Microbiology
Device Facts
Record ID
K152285
Device Name
illumigene Pertussis DNA Amplification Assay
Applicant
Meridian Bioscience, Inc.
Product Code
OZZ · Microbiology
Decision Date
Nov 10, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3980
Device Class
Class 2
Indications for Use
The illumigene Pertussis DNA Amplification Assay, performed on the illumipro-10™, is a qualitative in vitro diagnostic test for the direct detection of Bordetella pertussis in human nasopharyngeal swab samples taken from patients suspected of having respiratory tract infection attributable to Bordetella pertussis. The illumigene Pertussis assay utilizes loop-mediated isothermal DNA amplification (LAMP) technology to detect B. pertussis by targeting the IS481 insertional element of the B. pertussis genome. The IS481 insertional element can also be found in B. holmesii and B. bronchiseptica strains. Respiratory infection with B. pertussis, B. holmesii or B. bronchiseptica may yield positive test results in IS481 assays. B. holmesii infection may cause clinical illness similar to B. pertussis, and mixed outbreaks involving both B. pertussis and B. holmesii infection have been reported. Additional testing should be performed if necessary to differentiate B. holmesii and B. pertussis. B. bronchiseptica is a rare cause of infection in humans. When clinical factors suggest that B. pertussis may not be the cause of respiratory infection, other clinically appropriate investigation(s) should be carried out in accordance with published guidelines. Negative results for the illumigene Pertussis DNA Amplification Assay do not preclude Bordetella pertussis infection and positive results do not rule out co-infection with other respiratory pathogens. Results from the illumigene Pertussis assay should be used in conjunction with information obtained during the patient’s clinical evaluation as an aid in diagnosis of B. pertussis infection and should not be used as the sole basis for treatment or other patient management decisions. illumigene Pertussis is intended for use in hospital, reference or state laboratory settings. The device is not intended for point-of-care use.
Device Story
The illumigene Pertussis DNA Amplification Assay is an in vitro diagnostic test for detecting B. pertussis in nasopharyngeal swabs. It utilizes loop-mediated isothermal DNA amplification (LAMP) to target the IS481 insertional element of the B. pertussis genome. The assay is performed on the illumipro-10 instrument. The device is intended for use in hospital, reference, or state laboratory settings by trained personnel. The system provides qualitative results to aid in the diagnosis of B. pertussis infection. Results are used by clinicians in conjunction with patient clinical evaluation to guide treatment and management decisions. The assay does not differentiate between B. pertussis, B. holmesii, and some B. bronchiseptica strains, as all may yield positive results due to the shared IS481 target.
Clinical Evidence
Clinical evaluation conducted at four sites with 145 symptomatic patients. Compared predicate method vs. pretreatment method. Initial invalid rate without pretreatment was 13.1% (19/145); with pretreatment, invalid rate dropped to 4.1% (6/145). PPA 93.3% (14/15) and NPA 96.3% (105/109) observed. Additional bench validation confirmed 100% agreement for hold-time studies (room temp and refrigerated) for both swab-retained and ESwab samples.
Technological Characteristics
Uses loop-mediated isothermal DNA amplification (LAMP) technology. Targets the IS481 insertional element of the B. pertussis genome. Performed on the illumipro-10 instrument. Qualitative in vitro diagnostic assay.
Indications for Use
Indicated for qualitative detection of Bordetella pertussis in human nasopharyngeal swab samples from patients suspected of respiratory tract infection. For use in hospital, reference, or state laboratory settings.
Regulatory Classification
Identification
A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B; (2) Influenza A subtype H1 and Influenza A subtype H3; (3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B; (4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus; (5) Human Metapneumovirus; (6) Rhinovirus; and (7) Adenovirus.
Special Controls
*Classification.* Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.
Submission Summary (Full Text)
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K152285
B. Purpose for Submission:
To make three modifications to the upfront sample preparation of the original K133673 illumigene Pertussis DNA Amplification Assay performed on the illumipro-10™. The modifications are:
1. Modify sample preparation to retain the specimen swab in the Sample Buffer Tube at the end of the elution step.
2. Add the Copan Liquid Amies Elution Swab (ESwab™) Collection and Transport System (Copan Diagnostics, K061301) as an acceptable sample type.
3. Add an optional sample pretreatment method to neutralize the interfering activity of biological substances found in the nasopharynx of some patients.
C. Measurand:
DNA target sequence: IS481 Insertion Sequence
D. Type of Test:
The illumigene Pertussis DNA Amplification Assay is a qualitative *in vitro* diagnostic device for the direct detection of *Bordetella pertussis* DNA in human nasopharyngeal swab (NP) samples. The assay uses isothermal loop-mediated DNA amplification (LAMP) technology that targets the IS481 insertional element of the *Bordetella pertussis* genome.
E. Applicant:
Meridian Bioscience Inc.
F. Proprietary and Established Names:
illumigene® Pertussis DNA Amplification Assay
G. Regulatory Information:
1. Regulation section:
866.3980: Respiratory Viral Panel Multiplex Nucleic Acid Assay
{2}
2. Classification:
II
3. Product code:
OZZ, Bordetella pertussis DNA Assay System
4. Panel:
Microbiology (83)
H. Intended Use:
1. Intended use(s):
The illumigene Pertussis DNA Amplification Assay, performed on the illumipro-10™, is a qualitative in vitro diagnostic test for the direct detection of Bordetella pertussis in human nasopharyngeal swab samples taken from patients suspected of having respiratory tract infection attributable to Bordetella pertussis.
The illumigene Pertussis assay utilizes loop-mediated isothermal DNA amplification (LAMP) technology to detect B. pertussis by targeting the IS481 insertional element of the B. pertussis genome. The IS481 insertional element can also be found in B. holmesii and B. bronchiseptica strains. Respiratory infection with B. pertussis, B. holmesii or B. bronchiseptica may yield positive test results in IS481 assays. B. holmesii infection may cause clinical illness similar to B. pertussis, and mixed outbreaks involving both B. pertussis and B. holmesii infection have been reported. Additional testing should be performed if necessary to differentiate B. holmesii and B. pertussis. B. bronchiseptica is a rare cause of infection in humans. When clinical factors suggest that B. pertussis may not be the cause of respiratory infection, other clinically appropriate investigation(s) should be carried out in accordance with published guidelines.
Negative results for the illumigene Pertussis DNA Amplification Assay do not preclude Bordetella pertussis infection and positive results do not rule out co-infection with other respiratory pathogens. Results from the illumigene Pertussis assay should be used in conjunction with information obtained during the patient’s clinical evaluation as an aid in diagnosis of B. pertussis infection and should not be used as the sole basis for treatment or other patient management decisions.
illumigene Pertussis is intended for use in hospital, reference or state laboratory settings. The device is not intended for point-of-care use.
2
{3}
2. Indication(s) for use:
Same as Intended Use
3. Special conditions for use statement(s):
For Prescription Use Only
4. Special instrument requirements:
illumipro-10™ Automated Isothermal Amplification and Detection System
I. Device Description:
The illumigene Molecular Diagnostic Test System is comprised of the illumigene Pertussis DNA Amplification Assay Test Kit, the illumigene Pertussis External Control Kit and the illumipro-10™ Automated Isothermal Amplification and Detection System. The illumigene Pertussis assay utilizes loop-mediated isothermal amplification (LAMP) technology to detect the presence of Bordetella pertussis in human nasopharyngeal swab specimens. The assay targets a 198 base pair sequence of the Bordetella pertussis genome residing in a region of the IS481 insertional element sequence. Each illumigene Pertussis assay is completed using an illumigene Assay Control/Negative Control Reagent containing Control material, an illumigene Sample Buffer tube, an illumigene Pertussis Test Device and Mineral Oil.
Nasopharyngeal swab specimens are eluted with illumigene Sample Buffer by inserting and retaining the swab in the Sample Buffer Tube followed by vortexing. In the case of the ESwab System the collection tube is simply vortexed prior to the next steps. An optional pretreatment/treatment step can also be performed to eliminate inhibitory activity of biological substances found in a patient's sample by addition of the Pretreatment Reagent. Eluted sample, from the Sample Buffer Tube or the ESwab System collection tube is added to an Assay/Control/Negative Control Tube, vortexed and heated at 95°C for 10 minutes. The heat-treated Specimen/Control sample is added to the illumigene Test Device. Mineral oil is added to the illumigene Test Device to prevent evaporation. DNA amplification occurs in the illumigene Test Device. ESwab™ comprises of 1 mL of Liquid Amies and a FLOQSwab™. The volume of ESwab medium utilized for testing is 25 μL as it delivers the same number of organisms as does a rayon swab in transport medium (worst-case swab type).
The illumipro-10™ heats each illumigene Pertussis Test Device containing prepared Sample and Control material, facilitating amplification of target DNA. When B. pertussis is present in the specimen, target DNA is amplified and magnesium pyrophosphate is generated, forming a precipitate in the reaction mixture.
The illumipro-10™ monitors the absorbance characteristics of the reaction solutions at the assay Run Start (Signalinitial, Si) and at the assay Run End (Signalfinal, Sf). The illumipro-10™ calculates the ratio of Sf to Si and compares the ratio to an established cut-off value. The illumipro-10™ software calculates this ratio for both the TEST chamber and the CONTROL
3
{4}
chamber. $\mathrm{S_f}:\mathrm{S_i}$ information obtained from each illumipro- $10^{\mathrm{TM}}$ instrument is not available to the end-user. The illumipro- $10^{\mathrm{TM}}$ instrument reports and stores final results for the end-user as Positive, Negative, or Invalid.
# J. Substantial Equivalence Information:
1. Predicate device name(s): illumigene® Pertussis DNA Amplification Assay
2. Predicate 510(k) number(s): K133673
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device illumigene® Pertussis (K152285) | Predicate illumigene® Pertussis DNA Amplification Assay (K133673) |
| Intended Use | Same as predicate | Qualitative detection of a Bordetella pertussis DNA sequence from nasopharyngeal specimens |
| Detection | Same as predicate | Self-contained and Automated |
| Analyte | Same as predicate | DNA |
| Controls | Same as predicate | Internal Control Provided |
| Organisms Detected | Same as predicate | Bordetella pertussis |
| External Controls | Same as predicate | External positive control included in illumigene Pertussis External Control Kit |
| Test Format | Same as predicate | DNA Amplification; Loop-Mediated Isothermal Amplification (LAMP) |
| Assay Target | Same as predicate | Bordetella pertussis IS481 Insertional Element |
| Instrumentation | Same as predicate | illumipro-10TM |
| Sample Processing | Same as predicate | Manual extraction (for below for differences) |
| Reading Method | Same as predicate | Visible Light Transmission |
| Differences | | |
| --- | --- | --- |
| Item | Device illumigene® Pertussis (K152285) | Predicate illumigene® Pertussis DNA Amplification Assay (K133673) |
| Sample Processing | Sample preparation retains the specimen swab in the Sample Buffer Tube at the | Specimen swab removed from the Sample Buffer Tube at the end of the elution step. |
{5}
| Differences | | |
| --- | --- | --- |
| Item | Device
illumigene® Pertussis
(K152285) | Predicate
illumigene® Pertussis DNA
Amplification Assay
(K133673) |
| | end of the elution step. | |
| Sample Processing | Copan Liquid Amies Elution Swab (ESwab™) Collection and Transport System (Copan Diagnostics, K061301) added as an acceptable sample type. | Not applicable |
| Sample Processing | Optional sample pretreatment method included to neutralize the interfering activity of biological substances found in the nasopharynx of some patients. Pretreatment Reagent can be added to untested and previously tested patient specimens. | Not applicable |
K. Standard/Guidance Document Referenced (if applicable):
None
L. Test Principle:
The illumigene Pertussis assay is based on loop mediated isothermal amplification technology (LAMP). Loop mediated amplification of DNA is accomplished by the use of specially designed primers that provide specific and continuous isothermal amplification. Magnesium pyrophosphate is produced as a by-product of LAMP amplification. The magnesium pyrophosphate forms a white precipitate in the reaction solution, giving the reaction solution a turbid appearance. Change in sample absorbance created by precipitation of magnesium pyrophosphate indicates the presence of target DNA and is considered a positive reaction. The absence of target DNA results in no detectable change in sample absorbance and is considered a negative reaction.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility: See predicate K133673 for original studies.
{6}
b. Linearity/Assay Reportable Range: See predicate K133673 for original studies.
c. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): See predicate K133673 for original studies.
d. Detection Limit: See predicate K133673 for original studies. Limit of Detection (LoD) studies were not repeated for the device modifications because the addition of the Pretreatment Reagent is limited to the minor dilution effects of adding the Pretreatment Reagent to 500 µL of nasopharyngeal sample or to the ESwab medium. The calculated difference for the sample input dilution factor between the predicate device and the modified device is 7%. Two validation studies were performed to support that the LoD was not affected by the modifications. Samples for validation studies used to support the proposed modifications to the predicate device (K133673) were spiked close to the LoD of the assay (see Comparison Studies below).
e. Analytical Reactivity: See predicate K133673 for original studies.
f. Analytical Specificity: See predicate K133673 for original studies.
g. Microbial Interference: See predicate K133673 for original studies.
h. Interfering Substances: See predicate K133673 for original studies.
i. Fresh/Frozen Studies: See predicate K133673 for original studies.
j. Swab Type and Transport Media Equivalence Study: See predicate K133673 for original studies.
k. Assay cut-off: See predicate K133673 for original studies.
l. Carryover Study: See predicate K133673 for original studies.
2. Comparison studies:
a. Method comparison with predicate device:
A validation study was performed to evaluate the effect of retaining the swab head and the addition of the optional Pretreatment Reagent on eluted specimen hold times. Step 1 of the Specimen Preparation section of the predicate package insert allows users to hold eluted specimens "at room temperature (21-30°C) for up to 48 hours or refrigerated (2-8°C) for up to 7 days prior to testing" after which the swab head was removed from the Sample Buffer Tube. The retained swab-Pretreatment Reagent combination represents the worst case condition; this condition was assessed in the following hold-time study.
Negative matrix was prepared from donor flocked nasopharyngeal (NP) swabs
6
{7}
(Copan FLOQswab, worst-case swab type). Swabs were screened by illumigene Pertussis to confirm their negative status. Contrived negative samples were prepared with the donor flocked nylon NP swabs which were placed in Sample Buffer tubes containing 500 μL of buffer and 50 μL of Pretreatment Reagent. Contrived positive samples differed from negative contrived samples in that Sample Buffer tubes containing the donor flocked nylon NP swabs were inoculated with quantified B. pertussis strain BAA-589 to a final concentration 2 x LoD; which when diluted with Pretreatment Reagent represented concentrations near the LoD.
Once prepared, the contrived positive and negative samples were stored at 2-8°C and just above the upper limit for room temperature (i.e., 21-30°C) for defined time periods to assess performance. Tests were performed according to the method described in the subject package insert. The illumipro-10™ instruments utilized in this study used the software version, 2.00:602, the same version used for the predicate. The maximum time points selected for the study were in excess of the limits for hold times defined in the package insert.
Daily testing of external control reagents yielded the expected results. Eluted specimens containing swab heads and Pretreatment Reagent produced the expected results at (1) time 0 (baseline), (2) at hold times of 25+ and 49+ hours at room temperature, and (3) at hold times of 4 days and 8 days under refrigeration. There were no significant differences in Si:Si ratios of the different time points for the same sample resulting in 100% agreement between positive and negatives to expected positive or negative result. These data are comparable to those previously obtained in hold studies with eluted samples (K133673). The eluted sample hold times (i.e., 48 hours when held at 21-30°C and 7 days if held at 2-8°C) can be applied to eluted specimens retaining swab specimens with or without Pretreatment Reagent. No invalids were observed during this study.
Another validation study was performed to evaluate the impact of holding ESwab medium samples containing Pretreatment Reagent (worst case condition) under three testing conditions 1) time 0 (baseline), (2) at hold times of 25+ and 49+ hours at room temperature, and (3) at hold times of 4 days and 8 days under refrigeration. The maximum time points selected for the study were in excess of the limits for hold times defined in the package insert. Nasal wash from negative donors was used as negative matrix. Nasal wash samples were screened by illumigene Pertussis to confirm their negative status. Contrived negative samples were then prepared by adding the dry FLOQSwab component of the ESwab system to the tube component of the system that contained 1000 μL of ESwab medium (liquid Amies), followed by the addition of 50 μL of the negative matrix. Fifty μL of Pretreatment Reagent was added to each contrived specimen. Contrived positive samples differed from negative contrived samples in that the negative matrix added to each ESwab system was inoculated with quantified B. pertussis strain BAA-589 to a final concentration 2 x LoD; which when diluted with Pretreatment Reagent represented concentrations near the LoD. Prepared samples were stored at 2-8°C or just above the upper limit for room temperature (i.e., 21-30°C) storage for defined time periods.
7
{8}
Daily testing of external control reagents yielded the expected results. ESwab medium specimens containing Pretreatment Reagent produced the expected results at (1) time 0 (baseline), (2) at hold times of 25+ and 49+ hours at room temperature, and (3) at hold times of 4 days and 8 days under refrigeration. There were no significant differences in S_f to S_i ratios of the different time points for the same sample resulting in 100% agreement between positive and negatives to expected result. Performance of this sample was not affected by the presence of Pretreatment Reagent. No invalids were observed in the study.
b. Matrix comparison:
1) Matrix Equivalency Study: See predicate K133673 for original studies.
2) Sample Buffer (Elution buffer) Equivalence Study: See predicate K133673 for original studies.
3. Clinical studies:
a. Clinical Sensitivity: See predicate K133673 for original studies.
b. Clinical specificity: See predicate K133673 for original studies.
c. Other clinical supportive data (when a. and b. are not applicable):
Preliminary design work identified that some healthy, asymptomatic persons carried similar biological inhibitors in their nasopharynx as some symptomatic patients. Specimens from asymptomatic patients were used to establish the optimum pretreatment procedure. Final validation of the method and Pretreatment Reagent was performed using specimens both from specifically selected asymptomatic (Study 1) and symptomatic (Study 2) populations to enrich for specimens containing inhibitory activity.
Study 1: Rayon and flocked nylon swabs, previously cleared for use with the predicate (K133673) device, were used to collect specimens from healthy, asymptomatic subjects. All swabs were tested immediately using the predicate testing method with retesting of all invalid results. The same set of specimens were then treated with the Pretreatment Reagent and tested after pretreatment with any invalid result retested. All predicate, invalid and pretreatment testing was completed within a 24 hour period. The results of this study are shown in the Table below.
| | Rayon Swabs | | Flocked Nylon Swabs | |
| --- | --- | --- | --- | --- |
| | Predicate Method | Pre-treatment Method | Predicate Method | Pre-treatment Method |
| Negative | 54 | 54 | 44 | 54 |
| INVALID | 0 | 0 | 10 | 0 |
{9}
Invalids were not observed with rayon swab samples. In contrast, invalid results were observed in 10 of 54 flocked nylon swab samples which were resolved by the use of the Pretreatment Reagent. Testing using the Pretreatment Reagent method did not produce any false-positive or invalid test results during this study.
Study 2: A multi-site clinical evaluation was conducted from April to June 2015 at four selected test sites with prospectively collected specimens from symptomatic patients. Of the 164 specimens collected for this study, 145 met study criteria. The swabs used in this study were previously cleared for use with the predicate (K133673) device. The majority of specimens submitted to the test sites were collected with rayon tipped swabs; one sample was collected a flocked nylon swab and three samples with a polyester swab. The majority of the specimens were tested on the same day of collection. Collected specimens were tested immediately after collection, first using the predicate testing swab method, with retesting of all invalid results. Following testing with the predicate method the same specimens were then treated with the Pretreatment Reagent and tested after pretreatment with any invalid result retested. All predicate, invalid and pretreatment testing was completed within a 24 hour period. Of the nineteen specimens producing invalid test results by the predicate method, 15 were repeatedly invalid on retesting; three were negative and one positive on retesting. Following addition of Pretreatment Reagent these nineteen specimens, that initially gave invalid results, gave 17 negative and 2 positive results. Six specimens gave a result with the predicate method, but produced an invalid result upon treatment with the Pre-treatment Reagent. The data is presented in the Table below.
| | Test Sites Percent Agreement (Samples in Agreement/Total Tested) | | | | |
| --- | --- | --- | --- | --- | --- |
| | Site 1 | Site 2* | Site 3 | Site 4 | Total |
| Percent Positive Agreement (PPA) | 87.5% (6/7) (CI: 48.7-97.4%) | 0% (0/0) | 100% (2/2) (CI: 34.2-100%) | 100% (6/6) (CI: 61.8-100%) | 93.3% (14/15) (CI: 70.2-98.8%) |
| Percent Negative Agreement (NPA) | 94.3% (50/53) (CI: 84.6-98.1%) | 93.3% (14/15) (CI: 70.2-98.8%) | 100% (15/15) (CI: 79.6-100%) | 100% (26/26) (CI: 87.1-100%) | 96.3% (105/109) (CI: 90.9-98.6%) |
| Initial Invalid Rate without Pretreatment Reagent | 1.6% (1/63) | 0% (0/16) | 37.0% (10/27) | 20.5% (8/39) | 13.1% (19/145) |
| Initial invalid rate with Pretreatment Reagent | 4.8% (3/63) | 6.25% (1/16) | 3.7% (1/27) | 2.6% (1/39) | 4.1% (6/145) |
* Site 2 data represents first pass data only as this site did not retest INVALID untreated specimens, as required by the package insert, and before treating INVALID samples with Pretreatment Reagent. All other sites repeated INVALID test samples before pretreating specimens.
4. Clinical cut-off:
Not applicable
{10}
5. Expected values/Reference range:
See predicate K133673
N. Instrument Name:
*illumipro-10™* Automated Isothermal Amplification and Detection System
O. System Descriptions:
1. Modes of Operation:
See predicate K133673 for original description.
2. Software:
See predicate K133673 for original description.
3. Specimen Identification:
See predicate K133673 for original description.
4. Specimen Sampling and Handling:
Nasopharyngeal swab specimens are eluted with *illumigene* Sample Buffer by inserting and retaining the swab in the Sample Buffer Tube followed by vortexing. In the case of the ESwab System, the collection tube is simply vortexed prior to the next steps. An optional pretreatment/treatment step can also be performed to eliminate inhibitory activity of biological substances found in a patient’s specimen by addition of the Pretreatment Reagent. Specimens are processed manually prior to inoculation of test cartridges which are then placed into the *illumipro-10™* instrument for automated amplification and detection.
5. Calibration:
Calibration of the *illumipro-10™* is not required.
6. Quality Control:
See predicate K133673 for original description.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The “Performance Characteristics” Section above:
None
{11}
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.