LIAISON HSV-2 TYPE SPECIFIC IGG ASSAY, CONTROL HSV-2 IGG
K081687 · DiaSorin, Inc. · MYF · Nov 10, 2008 · Microbiology
Device Facts
Record ID
K081687
Device Name
LIAISON HSV-2 TYPE SPECIFIC IGG ASSAY, CONTROL HSV-2 IGG
Applicant
DiaSorin, Inc.
Product Code
MYF · Microbiology
Decision Date
Nov 10, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3305
Device Class
Class 2
Indications for Use
The LIAISON® HSV-2 Type Specific IgG assay is a chemiluminescent immunoassay (CLIA) to be used with the LIAISON® Analyzer for the qualitative determination of type specific IgG antibodies to Herpes Simplex Virus Type 2 (HSV-2) in human serum. The assay is indicated for testing sexually active adults or expectant mothers to aid in the presumptive diagnosis of HSV-2 infection. The LIAISON® HSV-2 Type Specific IgG assay has not been established for use in the pediatrics population, for neonatal screening, or for testing immunocompromised patients. The assay is neither FDA cleared nor approved for testing blood or plasma donors. The LIAISON® Control HSV-2 (negative and positive) are intended for use as assayed quality control samples to monitor the performance of the LIAISON® HSV-2 Type Specific IgG assay.
Device Story
The LIAISON® HSV-2 Type Specific IgG assay is an indirect chemiluminescence immunoassay (CLIA) performed on the LIAISON® Analyzer. It uses human serum samples as input. The device utilizes magnetic particles coated with recombinant HSV-2 gG2 antigen to capture specific IgG antibodies. A mouse monoclonal antibody linked to an isoluminol derivative acts as a conjugate. Following incubation and wash cycles, starter reagents induce a flash chemiluminescence reaction. A photomultiplier measures the light signal (Relative Light Units, RLU), which is proportional to the concentration of HSV-2 IgG. The system provides qualitative results to aid clinicians in the presumptive diagnosis of HSV-2 infection. It is intended for use in clinical laboratory settings by trained personnel. The output assists healthcare providers in determining patient serostatus, which informs clinical management and counseling.
Clinical Evidence
Clinical study of 951 samples (401 sexually active adults, 430 expectant mothers, 120 low prevalence) compared to FDA-cleared immunoblot. Sexually active adults: 98.1% sensitivity, 98.0% specificity. Expectant mothers: 94.8% sensitivity, 97.3% specificity. Low prevalence: 100% sensitivity and specificity. CDC panel agreement: 100%. Analytical studies included precision (overall %CV 8.7-83.2%), cross-reactivity (840 samples), and interference testing.
Technological Characteristics
Indirect chemiluminescence immunoassay (CLIA). Solid phase: magnetic particles coated with recombinant HSV-2 gG2 antigen. Conjugate: mouse monoclonal antibody linked to isoluminol derivative. Detection: photomultiplier measuring Relative Light Units (RLU). System: LIAISON® Analyzer. Qualitative determination.
Indications for Use
Indicated for sexually active adults or expectant mothers to aid in the presumptive diagnosis of HSV-2 infection. Not for pediatric use, neonatal screening, immunocompromised/immunosuppressed patients, or blood/plasma donor testing.
Regulatory Classification
Identification
Herpes simplex virus serological assays are devices that consist of antigens and antisera used in various serological tests to identify antibodies to herpes simplex virus in serum. Additionally, some of the assays consist of herpes simplex virus antisera conjugated with a fluorescent dye (immunofluorescent assays) used to identify herpes simplex virus directly from clinical specimens or tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by herpes simplex viruses and provides epidemiological information on these diseases. Herpes simplex viral infections range from common and mild lesions of the skin and mucous membranes to a severe form of encephalitis (inflammation of the brain). Neonatal herpes virus infections range from a mild infection to a severe generalized disease with a fatal outcome.
Special Controls
*Classification.* Class II (special controls). The device is classified as class II (special controls). The special control for the device is FDA's revised guidance document entitled “Class II Special Controls Guidance Document: Herpes Simplex Virus Types 1 and 2 Serological Assays.” For availability of the guidance revised document, see § 866.1(e).
K103603 — ZEUS ELISA HSV GG-2 IGG TEST SYSTEM · Zeus Scientific, Inc. · May 20, 2011
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number: K081687
B. Purpose for Submission: Initial Premarket Notification
C. Measurand: Anti-HSV 2 IgG Antibodies
D. Type of Test: Qualitative, Chemiluminescent Immunoassay (CLIA)
E. Applicant: Diasorin, Inc.
F. Proprietary and Established Names: LIAISON® HSV-2 Type Specific IgG
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| MYF & JJX : Herpes Simplex Virus-2 (IgG Antibody) | Class II | 21CFR 866.3305 | Microbiology (83) |
H. Intended Use:
The LIAISON® HSV-2 Type Specific IgG assay is a chemiluminescent immunoassay (CLIA) technology to be used with the LIAISON® Analyzer for the qualitative determination of type specific IgG antibodies to Herpes simplex virus Type 2 (HSV-2) in human serum. The assay is indicated for testing sexually active adults or expectant mothers to aid in the presumptive diagnosis of HSV-2 infection.
The LIAISON® HSV-2 Type Specific IgG assay has not been established for use in the pediatrics population, for neonatal screening, or for testing immunocompromised or immunosuppressed patients. The assay is neither FDA cleared nor approved for testing blood or plasma donors.
Caution: U.S. Federal Law restricts this device to sale by or on the order of a physician.
3) Special conditions for use statement(s): For Prescription use only
4) Special instrument requirements: LIAISON® Analyzer with Chemiluminescence reader
I. Device Description: The LIAISON® HSV-2 type specific IgG assay is an in vitro diagnostic device consisting of reagents provided within a plastic container/kit called the Reagent Integral. The reagent integral kit contains magnetic particles, calibrators, specimen diluent, and conjugate for 100 tests. Controls are required to run the assay and
{1}
sold separately. The LIAISON® HSV-2 IgG immunoassay is performed on the LIAISON® Analyzer (Model 15970) that was cleared previously (K052499). The magnetic particles are beads coated with HSV-2 IgG2 recombinant antigen. During first incubation, the magnetic particles are incubated with serum to capture IgG antibodies to HSV-2. During the second incubation, HSV-2 IgG and antigen complex bound to the magnetic particles are mixed with the detector antibody conjugate. The detector antibody conjugate is mouse monoclonal to human IgG conjugated to an isoluminol derivative. The unbound material is removed with a wash cycle after each incubation. Subsequently, a flash chemiluminescence reaction is induced by adding the starter reagents. The light signal, and hence the amount of isoluminol-antibody conjugate, is measured by a photomultiplier as relative light units (RLU) that is indicative of IgG anti-HSV-2 present in the samples.
# J. Substantial Equivalence Information:
a) Predicate device name (s):
The HerpeSelect® 1 and 2 immunoblot
b) Predicate Numbers (s):
K000238
Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | LIAISON® HSV-2 Specific IgG Assay | FOCUS Diagnostic HSV 1 and 2 Immunoblot IgG |
| Intended Use | The LIAISON® HSV-2 Type Specific IgG assay is a chemiluminescent immunoassay (CLIA) technology to be used with the LIAISON® Analyzer for the qualitative determination of type specific IgG antibodies to Herpes simplex virus Type 2 (HSV-2) in human serum. The assay is indicated for testing sexually active adults or expectant mothers to aid in the presumptive diagnosis of HSV-2 infectionThe LIAISON® HSV-2 Type | Focus Diagnostics'HerpeSelect® 1 and 2Immunoblot IgG test is intended for qualitatively detecting the presence or absence of human IgG class antibodies to HSV-1 and HSV-2 in human sera. The test is indicated for testing sexually active adults or pregnant women for aiding in the presumptive diagnosis of HSV-1 and HSV-2 infection. The predictive value of a positive or negative result depends on the population's prevalence and the pretest likelihood of HSV-1 |
{2}
| | Specific IgG assay has not been established for use in the pediatrics population, for neonatal screening, or for testing immunocompromised or immunosuppressed patients. The assay is neither FDA cleared nor approved for testing blood or plasma donors. | and HSV-2 infection. The performance of this assay has not been established for use in a pediatric population, for neonatal screening, for testing of immunocompromised patients, for use by a point of care facility or for use with automated equipment. |
| --- | --- | --- |
| Controls | 2 (Negative and Positive) | 2 (Negative and Positive) |
| Sample Matrix | Serum | Serum |
| Reagent Storage | 2-8°C, On-board or in Refrigerator | 2-8°C Refrigerator only |
| Differences | | |
| Item | LIAISON® HSV-2 Specific IgG Assay | FOCUS Diagnostic HSV 1 and 2 Immunoblot IgG |
| Type of Assay | Indirect Chemiluminescence Immunoassay | Immunoblot |
| Sample Handling/processing | Automated | Manual |
| Cutoff | Index Value 1.0 | Dark band on the Cutoff/Positive control strip |
| Calibrators | Two | Cutoff/Positive control |
| Detector | Mouse Monoclonal antibody to human IgG conjugated to an isoluminol derivative | Alkaline phosphate-conjugated goat antihuman IgG |
| Capture Reagent | Magnetic particles coated with HSV-2 IgG2 recombinant antigen | A blend of HSV-1 and HSV-2 native virus antigens, a recombinant gG1 and a recombinant gG2 antigen |
| Equivocal zone | Index Value of >0.90 and <1.1 | Positive anti-human serum, positive HSV common antigen, negative gG-1 and gG-2 |
| Unit of Measure | Index | Positive or Negative reading |
| Measurement System | Photomultiplier (flash chemiluminescence reader) | Visual manual reading |
| Total incubation | 35 minutes | 2 hours 20 minutes |
K. Standard/Guidance Document Referenced (if applicable):
1. Class II Special Controls Guidance Document: Herpes Simplex Virus Types 1 & 2 Serological Assays, April 3, 2007. http://www.fda.gov/cdrh/ode/guidance/1305.pdf
{3}
2. Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11, 2005. http://www.fda.gov/cdrh/ode/guidance/337.pdf
L. Test Principle: The method for qualitative determination of specific IgG to HSV-2 is an indirect chemiluminescence immunoassay (CLIA). HSV-2 gG2 recombinant antigen is used for coating magnetic particles (solid phase) and a mouse monoclonal antibody is linked to an isoluminol derivative (isoluminol-antibody conjugate). During the first incubation, HSV-2 antibodies present in calibrators, samples or controls bind to the solid phase. During the second incubation, the antibody conjugate reacts with HSV-2 IgG already bound to the solid phase. After each incubation, the unbound material is removed with a wash cycle. Subsequently, the starter reagents are added and a flash chemiluminescence reaction is thus induced. The light signal, and hence the amount of isoluminol-antibody conjugate, is measured by a photomultiplier as relative light units (RLU) and is indicative of HSV-2 IgG present in calibrators, samples or controls.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
a. Precision/Reproducibility: A reproducibility/precision study was conducted at two external and one internal. Each site used a different lot of the LIAISON® HSV-2 Type Specific IgG for the study. A coded panel comprised of 8 "engineered" serum samples was prepared by DiaSorin S.p.A. and provided to the sites. The coded panel samples were prepared by spiking positive samples into negative samples to achieve high negative, low positive and high positive results.
The coded panel was tested at all three sites, using four replicates per run in two runs per day with different operators performing each run during five operating days. The mean, standard deviation, and coefficient of variation (%CV) of the results were computed for each of the tested specimens for each of the sites and across sites. The Inter/Intra-operator and Inter/Intra-site reproducibility results are summarized in the following table.
| Sample ID | N | Mean Index | Within run %CV | Between run %CV | Total (by site) %CV | Between site %CV | Overall SD | Overall %CV |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| NC | 120 | 0.04 | 51.4 | 54.7 | 66.4 | 78.3 | 0.03 | 83.2 |
| PC | 120 | 2.25 | 4.1 | 5.7 | 6.7 | 12.0 | 0.29 | 13.1 |
| HSV2A | 120 | 0.91 | 4.9 | 8.9 | 9.7 | 12.2 | 0.14 | 15.0 |
| HSV2B | 120 | 1.19 | 4.5 | 7.4 | 8.2 | 2.8 | 0.10 | 8.7 |
| HSV2C | 120 | 2.06 | 5.1 | 7.9 | 9.2 | 5.0 | 0.22 | 10.8 |
| HSV2D | 120 | 0.77 | 5.0 | 9.5 | 10.4 | 6.8 | 0.10 | 13.6 |
| HSV2E | 120 | 0.88 | 5.1 | 8.8 | 10.1 | 4.0 | 0.09 | 10.7 |
| HSV2F | 120 | 1.38 | 4.2 | 8.5 | 9.3 | 4.3 | 0.14 | 10.5 |
| HSV2G | 120 | 1.37 | 3.8 | 7.5 | 7.8 | 5.8 | 0.13 | 9.8 |
| HSV2H | 120 | 15.7 | 8.1 | 8.0 | 10.7 | 3.7 | 1.75 | 11.2 |
{4}
b. Linearity/assay reportable range: NA
c. Traceability, Stability, Expected values (controls, calibrators, or methods): NA
d. Detection limit: This device is for the HSV-2 antibody detection and was not tested for the limit of detection. There is no standard available for measuring the HSV antibody units in the serum.
e. Analytical specificity:
1. Cross Reactivity: Cross reactivity studies for the HSV-2 Type Specific IgG assay were designed to evaluate potential interference directed against closely related members of the Herpes virus family (EBV, CMV, VZV) and other conditions that may mimic an HSV -2 infection.
| Organism/Condition | N | Comparator HSV-2 Assay | LIAISON® HSV-2 Positive | LIAISON® HSV-2 Negative | LIAISON® HSV-2 Equivocal |
| --- | --- | --- | --- | --- | --- |
| Epstein Barr Virus (IgG) | 114 | Negative | 0 | 114 | 0 |
| Cytomegalovirus (IgG) | 35 | Negative | 0 | 35 | 0 |
| Cytomegalovirus (IgM) | 6 | Negative | 0 | 6 | 0 |
| Toxoplasmosis (IgG) | 27 | Negative | 0 | 27 | 0 |
| Rubella (IgG) | 87 | Negative | 0 | 87 | 0 |
| Varicella-Zoster (IgG) | 114 | Negative | 0 | 114 | 0 |
| HSV-1 IgG | 383 | Negative | 4 | 377 | 2 |
| HAMA Samples | 6 | Negative | 0 | 6 | 0 |
| Chlamydia trachomatis | 8 | Negative | 0 | 8 | 0 |
| Human Papilloma Virus | 6 | Negative | 1 | 5 | 0 |
| Treponema pallidum | 13 | Negative | 2 | 10 | 1 |
| Gardnerella vaginalis | 4 | Negative | 0 | 4 | 0 |
| Candida albicans | 7 | Negative | 0 | 7 | 0 |
| Neisseria Gonorrhea | 3 | Negative | 1 | 2 | 0 |
| Rheumatoid Factor (RF) | 15 | Negative | 0 | 15 | 0 |
| Anti-Nuclear Antibody (ANA) | 12 | Negative | 0 | 12 | 0 |
| Total | 840 | | 8 | 829 | 3 |
2. Interference: The device performance was evaluated with the presence of interferents. The assay performance was not affected by hemolysis (at $138.5\mathrm{mg / dL}$ Hemoglobin), lipemia (at $3000\mathrm{mg / dL}$ triglycerides), icterus (at $20\mathrm{mg / dL}$ of bilirubin), cholesterol (at 500 $\mathrm{mg / dL}$ ) and serum Albumin at $(10,000\mathrm{mg / dL})$
{5}
f. Assay cut-off: NA
2. Comparison studies:
a. Method comparison with reference method:
The HerpeSelect® 1 and 2 immunoblot
b. Matrix comparison: NA
3. Clinical studies:
a. Clinical Sensitivity: NA
b. Clinical specificity: NA
Performance Characteristics
Summary of Clinical Studies
DiaSorin tested 951 samples collected in the Northeastern United States and compared the LIAISON® HSV-2 Type Specific IgG assay to an FDA cleared Immunoblot. The samples were classified as “At risk” samples (n=401) from sexually active adults (at risk for a Sexually Transmitted Disease (STD)), Expectant Mothers (n=430) and “Low Prevalence” samples (n=120) from patients seen at the clinic for anything other than an STD. The studies were conducted at two (2) independent external laboratories.
The sample populations were divided between the sites. Site 1 tested a total of 460 samples (201 At risk, 199 Expectant Mothers and 60 Low prevalence. Site 2 tested a total of 491 samples (200 at risk, 231 Expectant Mothers and 60 Low prevalence). Site 3 tested a total of 951 samples (401 At risk, 430 Expectant Mothers and 120 Low prevalence). Equivocal samples were repeat tested as per the Instructions for use. Any repeat equivocal samples on the predicate device were sent to a Reference Laboratory in the Pacific Northwest for Western Blot testing.
{6}
7
# Sexually Active Adults (401)
Four hundred one (401) samples were obtained from Sexually Active Adults who were seen at STD clinics in the Northeastern U.S. were tested with the LIAISON® HSV-2 Type Specific IgG assay and the HSV-2 predicate method Immunoblot.
| LIAISON®
HSV-2 Type
Specific IgG | Predicate Device | | | Total |
| --- | --- | --- | --- | --- |
| | Positive | Equivocal | Negative | |
| Positive | 104 | 1 | 5 | 110 |
| Equivocal | 0 | 0 | 0 | 0 |
| Negative | 1 | 1 | 289 | 291 |
| Total | 105 | 2* | 294 | 401 |
| | Percent | 95% Confidence Intervals |
| --- | --- | --- |
| Sensitivity | 98.1% (104/106) | 95.6 - 99.9% |
| Specificity | 98.0% (289/295) | 96.0 – 99.1% |
# Expectant Mother Population (430)
Four hundred thirty (430) samples collected from expectant mothers in the Northeastern U.S. were tested with the LIAISON® HSV-2 Type Specific IgG assay and the HSV-2 predicate method Immunoblot.
| LIAISON®
HSV-2 Type
Specific IgG | Predicate Device | | | Total |
| --- | --- | --- | --- | --- |
| | Positive | Equivocal | Negative | |
| Positive | 91 | 0 | 8 | 99 |
| Equivocal | 1 | 0 | 1 | 2 |
| Negative | 4 | 0 | 325 | 329 |
| Total | 96 | 0 | 334 | 430 |
| | Percent | 95% Confidence Intervals |
| --- | --- | --- |
| Sensitivity | 94.8% (91/96) | 89.4 – 97.9% |
| Specificity | 97.3% (325/334) | 95.3 – 98.6% |
{7}
8
# Low Prevalence Population (120)
One hundred twenty (120) “Low Prevalence” samples were obtained from patients who were seen at clinics (not for STD) in the Northeastern U.S. were tested with the LIAISON® HSV-2 Type Specific IgG assay and the HSV-2 predicate method Immunoblot.
| LIAISON® HSV-2 Type Specific IgG | Predicate Device | | | Total |
| --- | --- | --- | --- | --- |
| | Positive | Equivocal | Negative | |
| Positive | 20 | 0 | 0 | 20 |
| Equivocal | 0 | 0 | 0 | 0 |
| Negative | 0 | 0 | 100 | 100 |
| Total | 20 | 0 | 100 | 120 |
| | Percent | 95% Confidence Intervals |
| --- | --- | --- |
| Sensitivity | 100.0% (20/20) | 86.1 – 100.0% |
| Specificity | 100.0% (100/100) | 97.0 – 100.0% |
# Agreement with a CDC Panel
A serum panel was obtained from the Centers for Disease Control and Prevention and tested by the LIAISON® HSV-2 Type Specific IgG assay. The panel consisted of 52% positive and 48% negative samples. The LIAISON® HSV-2 Type Specific IgG assay demonstrated 100% total agreement with the CDC positive results (52/52) and negative results (48/48).
4. Clinical cut-off: NA
5. Expected values/Reference range:
The prevalence may vary depending upon geographical location, age, gender, type of test employed, specimen collection and handling procedures as well as clinical history of the patient. The observed and the hypothetical predictive values for the sexually active adults and expectant mothers populations are shown below. The positive predictive value (PPV) will decrease proportionally to the prevalence of HSV-2 infection as reflected in the table. The calculations are based on LIAISON® HSV-2 sensitivity and specificity of 98.1% and 98.0%, respectively, in a sexually active adult population and 94.8% and 97.3%, respectively, in an expectant mothers population and both 100% in a low prevalence population.
{8}
Observed Prevalence with Sexually Active Adults, Expectant Mothers and Low Prevalence populations
| Population | Sero-Status | Observed Prevalence | |
| --- | --- | --- | --- |
| | | LIAISON | Predicate |
| Sexually Active Adults* | HSV-2 Negative | 72.7% (290/399) | 73.7%(294/399) |
| | HSV-2 Positive | 27.3% (109/399) | 26.3%(105/399) |
| Expectant Mothers** | HSV-2 Negative | 76.9% (329/428) | 77.8%(333/428) |
| | HSV-2 Positive | 23.1% (99/428) | 22.2% (95/428) |
| Low Prevalence | HSV-2 Negative | 83.3% (100/120) | 83.3% (100/120) |
| | HSV-2 Positive | 16.7% (20/120) | 16.7% (20/120) |
*Excludes 2 Indeterminate results for Predicate device
**Excludes 2 Equivocal results for LIAISON device
HSV-2 Prevalence vs. Hypothetical Predictive Values
| Prevalence | Sexually Active Adults | | Expectant Mothers | | Low Prevalence Adults | |
| --- | --- | --- | --- | --- | --- | --- |
| | PPV | NPV | PPV | NPV | PPV | NPV |
| 80% | 99.5% | 92.8% | 99.3% | 82.4% | 100% | 100% |
| 70% | 99.1% | 95.7% | 98.8% | 88.9% | 100% | 100% |
| 60% | 98.7% | 97.2% | 98.1% | 92.6% | 100% | 100% |
| 50% | 98.0% | 98.1% | 97.2% | 94.9% | 100% | 100% |
| 40% | 97.0% | 98.7% | 95.9% | 96.6% | 100% | 100% |
| 30% | 95.5% | 99.2% | 93.8% | 97.8% | 100% | 100% |
| 20% | 92.5% | 99.5% | 89.8% | 98.7% | 100% | 100% |
| 10% | 84.5% | 99.8% | 79.6% | 99.4% | 100% | 100% |
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.