BD BACTEC Peds Plus/F Culture Vials Soybean-Casein Digest Broth with Resins in a Plastic Vial
K173873 · Becton, Dickinson and Company · MDB · Mar 16, 2018 · Microbiology
Device Facts
Record ID
K173873
Device Name
BD BACTEC Peds Plus/F Culture Vials Soybean-Casein Digest Broth with Resins in a Plastic Vial
Applicant
Becton, Dickinson and Company
Product Code
MDB · Microbiology
Decision Date
Mar 16, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.2560
Device Class
Class 1
Attributes
Pediatric
Indications for Use
BD BACTEC Peds Plus™F culture vials (enriched Soybean-Casein Digest broth with CO2) are for aerobic blood cultures. Principal use is with the BD BACTEC fluorescent series instruments for the qualitative culture and recovery of aerobic microorganisms (mainly bacteria and yeast) from pediatric and non-pediatric blood specimens which are generally less than 3 mL in volume.
Device Story
Blood culture vial containing enriched soybean-casein digest broth, resins, and CO2; used with BD BACTEC fluorescent series instruments. Inoculated with blood sample; instrument monitors CO2 production via chemical sensor in silicon rubber base. Sensor fluorescence increases as microorganisms metabolize nutrients and release CO2; instrument photo detectors measure fluorescence every 10 minutes. Algorithms in instrument rack microprocessor interpret CO2 production rate and absolute increase to flag positive vials. Positive vials removed for subculture and Gram stain; negative vials discarded after 120-hour protocol. Resins neutralize antimicrobial agents. Used in clinical microbiology labs; operated by laboratory technicians. Provides presumptive detection of viable microorganisms to guide clinical treatment.
Clinical Evidence
No clinical data. Evidence consists of bench-only seeded analytical studies. Evaluated 492 paired sets across four instruments (BACTEC FX, FX40, 9240, 9050) using 41 strains of bacteria and yeast. Primary endpoints: Time to Detection (TTD) and percent recovery. Results showed 100% detection rate for 10-100 CFU/vial. Median TTD difference between plastic and glass was -0.50 hours (95% CI -0.50, -0.33). Antimicrobial neutralization study confirmed resin efficacy for fluconazole/Candida albicans.
Technological Characteristics
Plastic culture vial containing enriched Soybean-Casein Digest broth and resins. Sensing principle: chemical sensor detecting CO2 via fluorescence. Connectivity: compatible with BD BACTEC fluorescent series instruments. Sterilization: not specified. Software: instrument-based monitoring.
Indications for Use
Indicated for aerobic blood culture and recovery of microorganisms (bacteria/yeast) from pediatric and non-pediatric blood specimens, typically <3 mL volume.
Regulatory Classification
Identification
A microbial growth monitor is a device intended for medical purposes that measures the concentration of bacteria suspended in a liquid medium by measuring changes in light scattering properties, optical density, electrical impedance, or by making direct bacterial counts. The device aids in the diagnosis of disease caused by pathogenic microorganisms.
K954924 — BACTEC PEDS PLUS CLUTURE VIALS · Becton Dickinson Diagnostic Instrument Systems · Feb 8, 1996
K151866 — BD BACTEC Peds Plus/F Culture Vials (plastic) · Becton, Dickinson and Company · Feb 25, 2016
K113558 — BD BACTEC Plus Aerobic/F Culture Vials (plastic) · Becton, Dickinson and Company · Mar 16, 2012
K161810 — BD BACTEC Standard/10 Aerobic/F Culture Vials Soybean-Casein Digest Broth in a Plastic Vial · Becton Dickinson · Feb 9, 2017
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
K173873
B. Purpose for Submission:
Update blood volume to 3.0 mL in intended use, remove current product limitation of fluconazole/Candida albicans for BD BACTEC Peds Plus/F Culture Vials (plastic)
C. Measurand:
Aerobic microorganisms from blood (bacteria and yeast)
D. Type of Test:
Liquid culture medium for recovery of microorganisms from blood using fluorescent technology to detect the increased CO₂ produced by the growth of microorganisms
E. Applicant:
Becton, Dickinson and Company
F. Proprietary and Established Names:
BD BACTEC Peds Plus/F Culture Vials Soybean-Casein Digest Broth with Resins in a Plastic Vial
G. Regulatory Information:
1. Regulation section:
21 CFR 866.2560 Microbial Growth Monitor
2. Classification:
Class I
3. Product code:
MDB System, Blood Culturing
{1}
4. Panel:
83 Microbiology
## H. Intended Use:
1. Intended use(s):
BD BACTEC Peds Plus/F culture vials (enriched Soybean-Casein Digest broth with CO₂) are for aerobic blood cultures. Principal use is with the BD BACTEC fluorescent series instruments for the qualitative culture and recovery of aerobic microorganisms (mainly bacteria and yeast) from pediatric and non-pediatric blood specimens which are generally less than 3.0 mL in volume.
2. Indication(s) for use:
Same as Intended Use
3. Special conditions for use statement(s):
Prescription Use
Updated limitation specific for antifungal agents/yeast combination under the Antimicrobial Activity section is:
Studies have demonstrated that the resins present in this medium adequately neutralize the antifungal agent fluconazole with Candida albicans. However, other antifungal agent/yeast combinations have not been tested/evaluated.
4. Special instrument requirements:
BACTEC fluorescent series instruments BACTEC FX, BACTEC FX40, BACTEC 9240 and BACTEC 9050 and were evaluated using software versions listed below:
| Instrument | Software Version |
| --- | --- |
| BACTEC FX | V5.20 |
| BACTEC 9240 | V4.95 |
| BACTEC 9050 | V2.01 |
| BACTEC FX 40 | V2.51B |
## I. Device Description:
The blood sample to be tested is inoculated into one or more vials which are inserted into the BACTEC fluorescent series instrument for incubation and periodic reading. Each vial contains a chemical sensor which can detect increases in CO₂ produced by the growth of microorganisms. The sensor is monitored by the instrument every ten minutes for an increase
{2}
in its fluorescence, which is proportional to the amount of $\mathrm{CO}_{2}$ present. A positive reading indicates the presumptive presence of viable microorganisms in the vial. Detection is limited to microorganisms that will grow in a particular type of culture medium.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
BD BACTEC Peds Plus/F Culture Vials (Plastic) Soybean-Casein Digest Broth with Resin in a plastic vial
2. Predicate $510(\mathbf{k})$ number(s):
K151866
3. Comparison with predicate:
Table 1: Comparison with Predicate Device
| Similarities | | |
| --- | --- | --- |
| Item | Device BD BACTEC Peds Plus/F Culture Vials Soybean-Casein Digest Broth with Resins in a Plastic Vial K173873 | Predicate BACTEC Peds Plus/F Culture Vials (Plastic) Soy Soybean- Casein Digest Broth with Resins in a Plastic Vial K151866 |
| Sample Type | Human blood | Same |
| Growth Medium | 40 mL enriched soybean-casein digest broth with CO2and N2 | Same |
| Detection Technology | Continuous monitoring; measurement of CO2increase; resins for absorption of antimicrobials; rocking agitation parameters | Same |
| Incubation | 35°C (± 1.5°C) up to 120 hours | Same |
| Vial Weight | 20.9 gram | Same |
| Vial Height | 5.0 inches | Same |
| Vial Sensor | Specific for the plastic vial geometry | Same |
| Sensor Adhesion Promotor | Required | Same |
{3}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | BD BACTEC Peds Plus /F culture vials (enriched Soybean-Casein Digest broth with CO_{2}) are for aerobic blood cultures. Principal use is with the BD BACTEC fluorescent series instruments for the qualitative culture and recovery of aerobic microorganisms (mainly bacteria and yeast) from pediatric and non-pediatric blood specimens which are generally less than 3 mL in volume. | BD BACTEC Peds Plus /F culture vials (enriched Soybean-Casein Digest broth with CO_{2}) are for aerobic blood cultures. Principal use is with the BD BACTEC fluorescent series instruments for the qualitative culture and recovery of aerobic microorganisms (mainly bacteria and yeast) from pediatric and non-pediatric blood specimens which are generally less than 5 mL in volume. |
| Blood Volume | 3.0 mL | 0.5, 5.0 mL |
| Ingredient (Pyridoxal HCl (Vitamin B_{6}) | 0.001% w/v | 0.0006% w/v |
## K. Standard/Guidance Document Referenced (if applicable):
Not Applicable
## L. Test Principle:
The BD BACTEC Peds Plus /F medium is an enriched soybean- casein digest broth, with each vial containing 40 mL of broth. Sodium polyanetholesulfonate (SPS) is added to the medium as an anticoagulant that inhibits bacteriocidal activities in the blood. The concentration of SPS has been adjusted to accommodate the low blood volumes of 0.5- 5.0 mL per vial.
Each BD BACTEC Peds Plus/F culture vial contains a chemical sensor in a silicon rubber base that can detect increases in CO₂ produced by the growth of microorganisms. Low volume (0.5- 5 mL) of blood is inoculated into the BD BACTEC Peds Plus/F blood culture vial, which is inserted into the BD BACTEC Fluorescent Series instrument for incubation, agitation and periodic measurement. When microorganisms are present in the blood sample, they metabolize nutrients in the culture medium, releasing CO₂ into the medium. A dye in the sensor reacts with the CO₂, modulating the amount of light that is absorbed by the fluorescent material in the sensor. The instrument’s photo detectors monitor the sensor every 10 minutes and measure the level of fluorescence, which is proportional to the amount of CO₂ present in the vial. Positivity of a vial is determined by algorithms resident in the instrument rack’s microprocessor. The algorithms use the rate of CO₂ production as well as the absolute increase in CO₂ to interpret the data.
Culture vials flagged as presumptively positive are removed from the instrument for
{4}
subculture and Gram stain in order to identify the microorganisms for further evaluation and proposed patient treatment. Culture vials that are not flagged as positive remain in the instrument until the test protocol has been completed and negative vials are discarded at the end of protocol (120 hours).
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
The BD BACTEC Peds Plus/F vial (plastic) was evaluated across three lots of culture media vials in the Time To Detection (TTD) and the Percent Recovery studies using exactly 3.0 mL of blood per vial.
The TTD results with 95% confidence intervals (CI) stratified by vial lot (combining total results for organisms and instruments) are shown in Table 2.1.
| Lot # | Total # | Median Time to Detection (TTD) in Hours (95% CI) | | | |
| --- | --- | --- | --- | --- | --- |
| | | 0-1 CFU | 1-10 CFU | 10-100 CFU | Overall |
| 1 | 204 | 28.09(17.24, 40.72) | 19.88(15.85,29.33) | 17.05(15.41,17.61) | 17.23(16.49,18.62) |
| 2 | 208 | 19.62(12.31,26.08) | 21.56(16.14,32.25) | 17.09(15.82,17.93) | 17.45(16.57,18.55) |
| 3 | 203 | 40.57(14.24,88.11) | 21.07(16.51,28.26) | 17.05(15.78,17.74) | 17.57(16.56,18.68) |
The percent recovery results with 95% CI stratified by vial lot (combining total results for inoculum levels, organisms, and instruments) are shown in Table 2.2.
| Lot # | Total # | Positive | Percent Recovery (95% CI) |
| --- | --- | --- | --- |
| 1 | 224 | 204 | 91.07% (88.84, 95.98) |
| 2 | 224 | 208 | 92.86% (87.95, 95.09) |
| 3 | 224 | 203 | 90.62% (88.83, 95.54) |
The analysis was acceptable and there was no significant difference across the three lots evaluated with 3.0 mL blood in this study.
#### b. Linearity/assay reportable range:
Not applicable
#### c. Traceability, Stability, Expected values (controls, calibrators, or methods):
##### Quality Control
An internal validation study across three lots with inoculum at 10-100 CFU and 3.0
{5}
mL of blood per vial was conducted using nine organisms: Streptococcus pyogenes, Escherichia coli, Pseudomonas aeruginosa, and Streptococcus pneumoniae were evaluated by testing nine vials each; Haemophilus influenzae, Neisseria meningitidis, Candida albicans, and Alcaligenes faecalis were evaluated by testing six vials each; Staphylococcus aureus was evaluated by testing three vials. All organisms were detected $\leq 72$ hours, with mean ranges between 10 (Escherichia coli) to 36.5 (Neisseria meningitidis) hours.
| Streptococcus pyogenes | ATCC 19615 |
| --- | --- |
| Escherichia coli | ATCC 25922 |
| Streptococcus pneumoniae* | ATCC 6305 |
| Pseudomonas aeruginosa | ATCC 27853 |
| Candida albicans | ATCC18804 |
| Neisseria meningitidis | ATCC 13090 |
| Alcaligenes faecalis | ATCC 8750 |
| Haemophilus influenzae | ATCC 19418 |
| Staphylococcus aureus | ATCC 25923 |
*CLSI recommended QC strain for blood culture media
# d. Detection limit:
# Microbial Detection Limit (MDL)
The microbial detection limit study was conducted to assess the capability of the culture media to detect low level of organisms (expected target level of 0-1 and 1-10 CFU/vial) when present in blood. The study included 15 organisms (12 bacteria and 3 yeast strains) tested with $3.0~\mathrm{mL}$ of blood, at challenging target inoculum levels of 0-1 and 1-10 CFU/vial across three lots with BACTEC FX and BACTEC 9240 instruments:
15 strains x 1 blood volume (3.0 mL) x 2 inoculum levels x 3 lots x 2 instruments = 180 paired sets
For statistical analysis, $95\%$ two-sided bootstrap confidence intervals for differences were used. The results were stratified by organism and are shown in Table 3.1 below.
Table 3.1: Percent Recovery Difference for 0-1 and 1-10 CFU Combined (by Organism) Using ${3.0}\mathrm{\;{mL}}$ of Blood
| Organism | Paired sets (#) | Percent recovery for Peds Plus/F (Plastic) | Percent recovery for Peds Plus/F (Glass) | Difference in percent recovery between Plastic and Glass (95% CI) |
| --- | --- | --- | --- | --- |
| Candida albicans | 12 | 83.33% | 75.00% | 8.33% (-16.67%, 33.33%) |
| Candida glabrata | 12 | 58.33% | 75.00% | -16.67% (-41.67%, 0%) |
| Crytococcus neoformans | 12 | 83.33% | 66.67% | 16.67% (-16.67%, 50%) |
{6}
| Organism | Paired sets (#) | Percent recovery for Peds Plus/F (Plastic) | Percent recovery for Peds Plus/F (Glass) | Difference in percent recovery between Plastic and Glass (95% CI) |
| --- | --- | --- | --- | --- |
| Enterococcus faecalis | 12 | 83.33% | 91.67% | -8.33% (-25.00%, 0%) |
| Escherichia coli | 12 | 66.67% | 83.33% | -16.67% (-58.33%, 25.00%) |
| Haemophilus influenzae | 12 | 75.00% | 66.67% | 8.33% (-16.67%, 33.33%) |
| H. parainfluenzae | 12 | 83.33% | 83.33% | 0.00% (-25.00%, 25.00%) |
| Micrococcus luteus | 12 | 50.00% | 58.33% | -8.33% (-25.00%, 0%) |
| Neisseria gonorrhoeae | 12 | 75.00% | 83.33% | -8.33% (-33.33%, 16.67%) |
| Neisseria meningitidis | 12 | 58.33% | 66.67% | -8.33% (-33.33%, 16.67%) |
| Pseudomonas aeruginosa | 12 | 75.00% | 91.67% | -16.67% (-41.67%, 0%) |
| Staphylococcus aureus | 12 | 75.00% | 83.33% | -8.33% (-41.67%, 25%) |
| Staphylococcus epidermidis | 12 | 66.67% | 41.67% | 25.00% (8.13%, 50%) |
| Streptococcus pneumoniae | 12 | 41.67% | 33.33% | 8.33% (-67.67%, 41.67%) |
| Streptococcus sanguinis | 12 | 50.00% | 58.33% | -8.33% (-25.00%, 0%) |
| Total | 180 | 68.33% | 70.56% | -2.22% (-9.44%, 4.44%) |
There were 180 paired sets tested; 104 sets were positive in both the predicate glass and the modified plastic vials, 23 were positive in the glass vial only, and 19 were positive in the plastic vial only. Of the 104 paired sets that were positive, 18 were at the target inoculum level of 0-1 CFU/vial and 86 at the target inoculum level of 1-10 CFU/vial. These paired sets were further evaluated by Time To Detection (TTD) in hours with $95\%$ CI using bootstrap percentile method. The performance is noted in the table below:
Table 3.2: Median TTD Difference for 0-1 and 1- 10 CFU/Vial in Hours
| Target Inoculum Level | Paired Sets (#) | Median TTD (Plastic) | Median TTD (Glass) | Median of TTD Differences (Plastic-Glass) |
| --- | --- | --- | --- | --- |
| 0-1 CFU | 18 | 22.02 (15.09, 38.13) | 23.02 (15.44, 37.56%) | -0.34 (-1.17, 0.25) |
| 1-10 CFU | 86 | 20.84 (18.5, 23.24) | 21.73 (17.77, 24.84) | -0.5 (-0.67, -0.17) |
The MDL study demonstrated that the Peds Plus/F vial (plastic) performed equivalently when compared to the comparator glass device at the challenging low target inoculum levels of 0-1 and 1-10 CFU/vial with $3.0~\mathrm{mL}$ of blood.
e. Analytical specificity:
Not applicable
f. Assay cut-off:
Not applicable
2. Comparison studies:
{7}
Performance of the BD BACTEC Peds Plus/F (plastic) blood culture vials was evaluated in seeded internal analytical studies to demonstrate equivalent performance to the comparator device, the BD BACTEC Peds Plus/F (glass) blood culture vials. The comparisons were made using the following parameters: Time to detection, percent recovery, false negative rate, and antimicrobial neutralization capability. For statistical analysis, $95\%$ two-sided bootstrap confidence intervals for differences were used.
a. Method comparison with predicate device:
# Instrument Time to Detection (TTD) study
The TTD (in hours) was recorded as part of the combined Percent Recovery and the Microbial Detection Limit (MDL) studies. MDL studies were conducted using the BD BACTEC FX and 9240 instruments at challenging inoculum levels 0-1 and 1-10 CFU per vial. The Percent Recovery study represented by the standard inoculum of 10 to 100 CFU per vial across four instruments (9050, 9240, FX, and FX40). Studies were conducted with $3.0~\mathrm{mL}$ of blood over three media lots.
TTD result stratified by target inoculum levels is shown in Table 4.1 below.
Table 4.1: Median TTD Difference By Inoculum Levels in Hours
| Target Inoculum Level | Median TTD (Plastic) | Median TTD (Glass) | Median of TTD Differences (Plastic-Glass) |
| --- | --- | --- | --- |
| 0-1 CFU | 22.02 (15.09, 38.13) | 23.02 (15.25, 38.05) | -0.34 (-1.08, 0.25) |
| 1-10 CFU | 20.84 (18.02, 23.24) | 21.73 (17.93, 24.91) | -0.5 (-0.67, -0.17) |
| 10-100 CFU | 17.07 (16.56, 17.56) | 17.52 (17.00, 18.01) | -0.50 (-0.50, -0.33) |
TTD results in hours with $95\%$ CI for the standard inoculum level of 10- 100 CFU/vial were evaluated by testing 41 isolates (32 bacteria and 3 yeasts noted in Table 4.2 below), across three lots tested with four BACTEC instruments. The results stratified by organisms and lots are shown in Table 4.2 and 4.3 below.
Table 4.2: Median TTD Difference for 10- 100 CFU/Vial in Hours (by Organism)
| Organism | Paired Sets (#) | Median TTD (Plastic) | Median TTD (Glass) | Median of TTD Differences (Plastic-Glass) |
| --- | --- | --- | --- | --- |
| Abiotrophia defectiva | 12 | 16.84 (16.54, 17.53) | 17.67 (17.00, 18.15) | -1.08 (-1.50, 0.16) |
| Acinetobacter lwoffii | 12 | 21.02 (20.3, 21.82) | 22.06 (22.02, 22.57) | -1 (-1.75, -0.5) |
| Aerococcus viridans | 12 | 13.36 (13.00,13.82) | 13.89 (13.43, 14.57) | -0.50 (-1, -0.08) |
| Aggregatibacter actinomycetemcomitans | 12 | 25.07 (24.05, 26.52) | 25.82 (25.31, 27.02) | -1 (-1.5, -0.5) |
| Alcaligenes faecalis | 12 | 20.91 (20.49, 21.96) | 22.49 (21.99, 23.05) | -1.5 (-1.59, -1) |
| Bacillus subtilis | 12 | 21.54 (21.08, 22.57) | 22.55 (22.01, 23.16) | -0.75 (-2, 0) |
| Candida albicans | 12 | 28.56 (27.26, 29.25) | 27.59 (26.78, 28.54) | 0.50 (-0.50, 1.42) |
| Candida glabrata | 12 | 44.64 (39.53, 46.04) | 40.41 (36.86, 44.77) | 2.25 (-1.00, 4.58) |
| Cardiobacterium hominis | 12 | 39.94 (39.11, 41.4) | 41.45 (40.58, 41.98) | -1.17 (-2.34, -1.00) |
| Corynebacterium jeikeium | 12 | 29.02 (28.55, 29.57) | 29.30 (28.51, 29.63) | -0.08 (-1.08, 0.17) |
{8}
| Organism | Paired Sets (#) | Median TTD (Plastic) | Median TTD (Glass) | Median of TTD Differences (Plastic-Glass) |
| --- | --- | --- | --- | --- |
| Cryptococcus neoformans | 12 | 61.81 (59.73, 63.80) | 65.86 (63.92, 69.43) | -4.67 (-7.08, -2.5) |
| Eikenella corrodens | 12 | 28.57 (27.54, 28.67) | 29.38 (28.60, 29.91) | -1.08 (-1.66, -0.42) |
| Enterobacter cloacae | 12 | 9.58 (9.29, 9.80) | 9.75 (9.49, 9.93) | -0.08 (-0.33, 0) |
| Enterococcus faecalis | 12 | 9.66 (9.50, 9.74) | 9.87 (9.57, 10.03) | -0.25 (-0.42, 0) |
| Escherichia coli | 12 | 10.07 (9.79, 10.29) | 10.24 (9.87, 10.46) | -0.08 (-0.25, 0.08) |
| Granulicatella adiacens | 12 | 15.04 (14.52, 15.29) | 15.1 (14.93, 15.5) | -0.25 (-0.66, 0.08) |
| Haemophilus influenzae | 12 | 17.09 (16.95, 17.39) | 17.62 (17.32, 18.12) | -0.5 (-1.00, 0) |
| H. influenzae, type b | 12 | 17.31 (17.12, 17.56) | 17.55 (17.06, 18.06) | -0.09 (-0.50, 0) |
| H. influenzae, type a | 12 | 16.02 (15.64, 17.13) | 16.34 (16.00, 17.38) | -0.25 (-0.92, 0.25) |
| H. parainfluenzae | 12 | 33.04 (31.41, 34.84) | 30.73 (28.99, 33.55) | 1.5 (-0.09, 4) |
| Kingella kingae | 12 | 17.6 (17.09, 18.12) | 17.82 (17.19, 18.27) | 0 (-0.92, 0.75) |
| Klebsiella pneumoniae | 12 | 9.12 (8.91, 9.24) | 9.17 (9.07, 9.41) | -0.25 (-0.33, 0.08) |
| Leuconostoc citreum | 12 | 17.46 (17.05, 18.37) | 18.06 (17.57, 18.71) | -0.5 (-1, 0) |
| Micrococcus luteus | 12 | 38.74 (38.34, 39.95) | 39.89 (39.3, 41.37) | -1.08 (-2.92, 0.5) |
| Neisseria gonorrhoeae | 12 | 21.12 (20.09, 21.38) | 22.12 (21.25, 22.55) | -1.08 (-1.50, 0.42) |
| Neisseria meningitidis | 12 | 19.44 (18.88, 19.96) | 21.51 (20.80, 22.09) | -2.27 (-2.92, -0.75) |
| Pediococcus acidilactici | 12 | 19.52 (19.45, 20.11) | 19.66 (19.45, 20.11) | 0 (-0.66, 0.66) |
| Proteus mirabilis | 12 | 13.37 (13.11, 13.54) | 13.91 (13.68, 14.17) | -0.58 (-0.92, -0.16) |
| Providencia stuartii | 12 | 11.54 (11.42, 11.61) | 11.87 (11.64, 12.02) | -0.33 (-0.58, -0.08) |
| Pseudomonas aeruginosa | 12 | 14.54 (14.26, 14.85) | 15.05 (14.91, 15.43) | -0.75 (-1, -0.25) |
| Rothia mucilaginosa | 12 | 16.49 (15.82, 16.95) | 17.17 (16.74, 17.45) | -0.58 (-1.42, 0.08) |
| Staphylococcus aureus | 12 | 11.26 (10.91, 11.61) | 11.38 (11.16, 11.61) | -0.16 (-0.34, 0) |
| Staph epidermidis | 12 | 17.63 (17.29, 17.88) | 18.04 (17.62, 18.60) | -0.5 (-1, 0.16) |
| Stenotrophomonas maltophilia | 12 | 35.06 (32.94, 37.48) | 34.52 (33.98, 36.27) | 0.5 (-2.83, 1.84) |
| Streptococcus agalactiae (Group B Strep) | 12 | 8.95 (8.79, 9.18) | 9.04 (8.75, 9.35) | -0.08 (-0.17, 0.17) |
| Streptococcus pneumoniae | 12 | 12.53 (12.14, 12.86) | 12.96 (12.60, 13.05) | -0.25 (-0.50, 0) |
| Streptococcus pneumoniae | 12 | 11.38 (11.28, 11.54) | 11.59 (11.41, 11.77) | -0.25 (-0.42, -0.17) |
| Streptococcus pneumoniae | 12 | 11.78 (11.57, 11.96) | 11.82 (11.70, 12.03) | 0.00 (-0.42, 0.17) |
| Streptococcus pneumoniae | 12 | 13.45 (13.12, 13.84) | 14.29 (13.71, 14.62) | -0.75 (-1.25, 0) |
| Streptococcus pyogenes (Group A Strep.) | 12 | 12.08 (11.77, 12.66) | 12.72 (12.34, 12.93) | -0.66 (-0.92, -0.08) |
| Streptococcus sanguinis | 12 | 14.00 (13.58, 14.11) | 14.58 (14.33, 14.33) | -0.58 (-1, -0.33) |
| Total | 492 | 17.07 (16.53, 17.56) | 17.52 (17.00, 18.00) | -0.50 (-0.50, -0.33) |
Table 4.3: Median TTD Difference for 10- 100 CFU/Vial in Hours (by Lots)
| Lot # | Paired Sets (#) | Median TTD (Plastic) | Median TTD (Glass) | Median of TTD Differences (Plastic-Glass) |
| --- | --- | --- | --- | --- |
| 1 | 164 | 17.05 (15.45, 17.61) | 17.61 (16.00, 18.45) | -0.5 (-0.67, -0.5) |
| 2 | 164 | 17.09 (15.67, 18.00) | 17.55 (16.11, 18.11) | -0.5 (-0.67, -0.42) |
| 3 | 164 | 17.05 (15.54, 17.63) | 17.19 (15.97, 18.27) | -0.16 (-0.33, 0) |
When stratified by organisms, it was noted that there were only four organisms that favored the comparator glass vial: Candida glabrata, Haemophilus parainfluenzae, Candida albicans, and Stenotrophomonas maltophilia.
{9}
The TTD study demonstrated that the modified plastic device performed equivalently compared to the comparator glass device when stratified by target inoculum level, organisms or lot numbers. It was observed that the median TTD for the modified plastic vial and the predicate glass was 17.07 and 17.52 hours respectively. The median difference was -0.50 hours (95% CI -0.50, -0.33), favoring the plastic vial by approximately 30 minutes as noted in Tables 4.1 and 4.2 above.
## Percent Recovery (Detection) Study
The percent recovery (detection) was evaluated in a study of 492 paired sets (41 strains x 3 lots x 4 instruments= 492) at the standard inoculum level of 10 to 100 CFU/vial on four instruments across three lots with 3.0 mL of blood using a diverse set of 41 strains that are commonly isolated from blood. The strains tested is listed in Table 4.2 above.
At standard inoculum level of 10-100 CFU/mL, the detection rate was 100% for both predicate glass and modified plastic devices. Difference in percent recovery was 0%. 95% CI is calculated using a score method for correlated proportion, shown in Table 5.1 for overall performance and in Table 5.2 stratified by lots.
Table 5.1: Overall Percent Recovery Difference for 10- 100 CFU/Vial
| Paired Sets (#) | Percent Recovery (Plastic) | Percent Recovery (Glass) | Difference between Percent Recovery of Plastic and Glass |
| --- | --- | --- | --- |
| 492 | 100% | 100% | 0% (-0.77%, 0.77%) |
Table 5.2: Percent Recovery Difference for 10- 100 CFU/Vial by Lots
| Lot# | Paired Sets (#) | Percent Recovery (Plastic) | Percent Recovery (Glass) | Difference between Percent Recovery of Plastic and Glass |
| --- | --- | --- | --- | --- |
| 1 | 164 | 100% | 100% | 0% (-2.29%, 2.29%) |
| 2 | 164 | 100% | 100% | 0% (-2.29%, 2.29%) |
| 3 | 164 | 100% | 100% | 0% (-2.29%, 2.29%) |
The study demonstrated that the modified plastic device performed equivalently when compared to the comparator glass device. The detection rate was 100% for organisms listed in Table 4.2 at standard inoculum level of 10-100 CFU/vial with 3.0 mL of blood for both the comparator glass and the plastic vials.
## Haemophilus spp. Detection Re-calculation
There were four Haemophilus spp. tested in the percent recovery (detection) study with 3.0 mL of blood at the inoculum level of 10-100 CFU/vial. The detection rate was 100% (Table 5.1 above) due to the increased blood volume of 3.0 mL in this study.
The Haemophilus spp. detection rate was evaluated originally (K151866) for three
{10}
target inoculum levels (0-1, 1-10, and 10-100) at 0.5 mL blood with the detection rate range of 8% - 40%. When stratified, the dataset of standard inoculum level 10- 100 CFU/vial with 0.5 and 5.0 mL blood was re-calculated and presented separately in the current submission.
There were a total of 96 paired sets (4 strains x 2 blood vol. x 3 lots x 4 instruments =96) in this re-calculation. The detection rate was 69% (33/48) and 100% (48/48) for 0.5 mL blood and 5.0 mL blood respectively. The detection rate is updated and presented separately according to blood volume at inoculum level of 10- 100 CFU/vial:
The Haemophilus species detection rates were 69% with 0.5 mL blood, and 100% with 5.0 mL
## False Negative Rates (Instrument-negative, subculture-positive)
A false negative was defined as a vial that was instrument-negative at the end of protocol (120 hours) yet contains viable organisms by subcultured onto appropriate culture media. The data for this study was generated from the Instrument Time to Detection, Percent Recovery, and Microbial Detection Limit studies.
A total of 110 instrument negative vials were sub-cultured at the end of protocol; 53 glass and 57 plastic vials. All subcultures were negative. There were no false negatives observed in either the glass or plastic device.
## False Positive Rate ((Instrument-positive, subculture-negative)
A false positive was defined as instrument positive but subculture negative. False positivity of the BACTEC Peds Plus/F plastic vial was assessed with 0.5, 3, and 5 mL fresh human blood in the previous submission (K151866). No instrument false positive signals were detected in the evaluation.
The plastic device performed equivalently when compared to the glass device in the false positive study.
## BD BACTEC Instrument Compatibility
The BACTEC instrument compatibility study was evaluated from the MDL and Percent Recovery studies (180+492= 672 paired sets) across four fluorescent series instruments: BD BACTEC FX, FX40, 9240, and 9050. The study included target inoculum levels of 0-1, 1-10, and 10-100 CFU/mL with 3.0 mL of blood. All four instruments were tested for standard inoculum level of 10-100 CFU/vial; FX and 9240 were tested for inoculum levels of 0-1, 1-10 CFU/mL. Of the 672 paired sets, 596 sets recovered organisms in both the comparator glass and plastic vials. The BACTEC instrument compatibility study is shown in Tables 6.1 and 6.2.
11
{11}
Table 6.1: Summary of Compatibility Study
| Instrument | Paired Sets (#) | Median of Time to Detection (TTD) in Hours (95% CI) | | |
| --- | --- | --- | --- | --- |
| | | Plastic | Glass | Median of TTD differences |
| FX 40 | 123 | 17.15 (15.57, 18.1) | 17.62 (15.94, 18.6) | -0.50 (-0.50, -0.17) |
| FX | 174 | 17.59 (16.59, 19.48) | 17.78 (17.14, 19.63) | -0.50 (-0.75, -0.33) |
| 9050 | 123 | 17.00 (16.00, 18.83) | 17.50 (15.67, 19) | -0.33 (-0.50, -0.17) |
| 9240 | 176 | 17.45 (16.52, 19.32) | 17.63 (16.53, 20.23) | -0.33 (-0.50, -0.17) |
Table 6.2: Median of TTD for All Inoculum Levels
| Target Inoculum Level | Instrument | Paired Sets (#) | Median Time to Detection (TTD) in Hours (95% CI) | | |
| --- | --- | --- | --- | --- | --- |
| | | | Plastic | Glass | Median of TTD differences |
| 0 to 1 | FX | 9 | 17.09 (11.65, 40.72) | 18.58 (12.15, 37.55) | -0.50 (1.33, 2.17) |
| | 9240 | 9 | 25.91 (17.24, 51.71) | 25.91 (18.39, 56.87) | -0.17 (-5.00, 3.5) |
| 1 to 10 | FX | 42 | 21.56 (17.11, 28.50) | 22.59 (17.18, 28.58) | -0.50 (-0.83, 0.09) |
| | 9240 | 44 | 19.74 (16.41, 29.29) | 20.74 (17.45, 29.51) | -0.34 (-0.67, -0.17) |
| 10 to 100 | FX40 | 123 | 17.15 (15.57, 18.1) | 17.62 (17.07, 18.60) | -0.50 (-0.50, -0.17) |
| | FX | 123 | 17.14 (15.32, 17.64) | 17.57 (15.77, 18.62) | -0.50 (-0.83, -0.33) |
| | 9050 | 123 | 17.00 (15.33, 18.83) | 17.50 (16.00, 19.00) | -0.50 (-0.50, -0.33) |
| | 9240 | 123 | 16.70 (14.87, 17.58) | 16.53 (15.07, 18.01) | -0.33 (-0.50, -0.17) |
The study demonstrated that the four instruments performed equivalently when the plastic vials were compared to the comparator glass vials.
## Antimicrobial Neutralization Capability
The Antimicrobial Neutralization Capability is to support the removal of the current limitation listed in the BACTEC Peds Plus/F Culture Vials (plastic) package insert, "Based on the neutralization study with Candida albicans and fluconazole, the media showed some degree of neutralization; however the results were inconclusive. The adequacy of antifungal neutralization by resins in the BD BACTEC Peds Plus/F vial (plastic) in unknown."
In the antimicrobial neutralization study, fluconazole was evaluated at 1.0 and 2.0 $\mu \mathrm{g} / \mathrm{mL}$ with two susceptible strains of Candida albicans (expected fluconazole MIC: $0.5\mu \mathrm{g} / \mathrm{mL}$ ) at target inoculum of 10- 100 CFU/vial with $3.0~\mathrm{mL}$ of blood in the Peds Plus/F Culture Vials Soybean-Casein Digest Broth with Resins. Each paired set was tested in triplicates for each strain at each fluconazole concentration:
2 strains x 2 test concentration x 3 (triplicates) = 12 paired-sets
One set of non-resin vials (BACTEC Standard/10 Aerobic/F) with and without fluconazole were tested as controls for each strain at each fluconazole concentration. The non-resin vial with fluconazole served as negative (no growth) and the non-resin vial without fluconazole served as positive (growth) controls. The results are shown in Table 7.
{12}
13
Table 7: Fluconazole/Candida albicans Combination Neutralization
| Organism | Fluconazole (μg/mL) | | Resin Peds Plus/F (# Detected /# tested) | | Non-resin Aerobic (# Detected /# tested) | |
| --- | --- | --- | --- | --- | --- | --- |
| | Expected MIC | Test Conc. | Plastic | Glass | With Antifungal | w/o Antifungal (Growth Control) |
| C. albicans #1 | 0.5 | 1.0 | 3/3 | 3/3 | 0/1 | 1/1 |
| | | 2.0 | 3/3 | 3/3 | 0/1 | 1/1 |
| C. albicans #2 | 0.5 | 1.0 | 3/3 | 3/3 | 0/1 | 1/1 |
| | | 2.0 | 3/3 | 3/3 | 0/1 | 1/1 |
The study demonstrated that resins in the Peds Plus/F media adequately neutralize fluconazole when testing Candida albicans at 10- 100 CFU/vial with 3.0 mL of blood. It supports the removal of the limitation. The following information specific for the antifungal agent/yeast combination is included in the Antimicrobial Activity section under limitation of the product labeling:
"Studies have demonstrated that the resins present in this medium adequately neutralize the antifungal agent fluconazole with Candida albicans. However, other antifungal agents/yeast combination have not been tested/evaluated."
b. Matrix comparison:
In seeded analytical studies, the performance of BD BACTEC Peds Plus/F culture medium in plastic vial was compared to that in glass vial, with 3.0 mL of blood, 32 common blood bloodstream bacteria and three yeast species across four fluorescent series instruments (BACTEC FX, FX40, BACTEC 9240, and BACTEC 9050).
3. Clinical studies:
Not applicable; seeded analytical studies to compare the plastic blood culture vials to the glass blood culture vials (comparator).
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
{13}
Not applicable
5. Expected values/Reference range:
Seeded analytical studies demonstrated equivalent performance of the BD BACTEC Peds Plus/F (plastic) blood culture medium when compared to the BD BACTEC Peds Plus/F (glass) blood culture medium.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
14
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.