K151866 · Becton, Dickinson and Company · MDB · Feb 25, 2016 · Microbiology
Device Facts
Record ID
K151866
Device Name
BD BACTEC Peds Plus/F Culture Vials (plastic)
Applicant
Becton, Dickinson and Company
Product Code
MDB · Microbiology
Decision Date
Feb 25, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.2560
Device Class
Class 1
Attributes
Pediatric
Indications for Use
BD BACTEC Peds Plus™/F culture vials (enriched Soybean-Casein Digest broth with CO2) are for aerobic blood cultures. Principal use is with the BACTEC fluorescent series instruments for the qualitative culture and recovery of aerobic microorganisms (mainly bacteria and yeast) from pediatric and non-pediatric blood specimens which are generally less than 5 mL in volume.
Device Story
Plastic culture vial containing enriched soybean-casein digest broth and resins for antimicrobial neutralization; used with BD BACTEC fluorescent series instruments (FX, FX40, 9240, 9050). Inoculated with blood sample; instrument monitors CO2 production via fluorescent sensor in vial base. Instrument detects microbial growth through periodic (10-min) fluorescence measurements; algorithms interpret rate and absolute increase of CO2. Positive vials flagged for subculture/Gram stain; negative vials discarded after 120-hour protocol. Benefits include rapid, automated detection of bloodstream pathogens; plastic material offers lighter weight and durability compared to glass predicate.
Clinical Evidence
No clinical data. Bench-only seeded analytical studies compared plastic vials to glass predicate using 984 paired sets across 15 common bloodstream pathogens. Primary endpoints: Time to Detection (TTD) and Percent Recovery. Results showed equivalent performance: 97.76% recovery (plastic) vs 97.26% (glass); median TTD difference -0.25 hours (95% CI -0.495, -0.08). False negative rates were 1.49% (plastic) vs 1.86% (glass).
Technological Characteristics
Plastic vial containing 40 mL enriched soybean-casein digest broth and SPS anticoagulant. Detection via chemical sensor (silicon rubber base with BCP indicator and red dye) measuring CO2-modulated fluorescence. Continuous monitoring via BACTEC fluorescent series instruments. Resins included for antimicrobial adsorption. Sterilization method not specified.
Indications for Use
Indicated for qualitative aerobic culture and recovery of bacteria and yeast from pediatric and non-pediatric blood specimens (0.5-5.0 mL). Contraindicated for use with very small blood volumes (≤0.5 mL) for fastidious organisms (e.g., Haemophilus) without appropriate supplementation; adequacy of antifungal neutralization for Candida albicans/Fluconazole is unknown.
Regulatory Classification
Identification
A microbial growth monitor is a device intended for medical purposes that measures the concentration of bacteria suspended in a liquid medium by measuring changes in light scattering properties, optical density, electrical impedance, or by making direct bacterial counts. The device aids in the diagnosis of disease caused by pathogenic microorganisms.
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM ASSAY ONLY TEMPLATE
A. 510(k) Number:
K151866
B. Purpose for Submission:
To obtain a substantial equivalence determination for a premarket notification for the BD BACTEC Peds Plus /F Culture Vial (plastic)
C. Measurand:
Aerobic microorganisms from blood (bacteria and yeast)
D. Type of Test:
Liquid culture medium for recovery of microorganisms from blood using fluorescent technology to detect the increased $\mathrm{CO}_{2}$ produced by the growth of microorganisms
E. Applicant:
Becton, Dickinson and Company
F. Proprietary and Established Names:
BD BACTEC Peds Plus/F Culture Vials (plastic)
G. Regulatory Information:
1. Regulation section:
21 CFR 866.2560 Microbial Growth Monitor
2. Classification:
Class I
3. Product code:
MDB System, Blood Culturing
4. Panel:
{2}
83 Microbiology
## H. Intended Use:
### 1. Intended use(s):
BD BACTEC Peds Plus /F culture vials (enriched Soybean-Casein Digest broth with CO₂) are for aerobic blood cultures. Principal use is with the BD BACTEC fluorescent series instruments for the qualitative culture and recovery of aerobic microorganisms (mainly bacteria and yeast) from pediatric and non-pediatric blood specimens which are generally less than 5 mL in volume.
### 2. Indication(s) for use:
Same as Intended Use.
### 3. Special conditions for use statement(s):
**Prescription Use**
- Some organisms may be dependent on having a minimum amount of blood in the medium for optimal growth. Fastidious organisms, such as certain Haemophilus species, require growth factors from the blood specimen, such as NAD, or factor V. Optimal growth of these organisms is dependent on having greater than 0.5 mL blood in the specimen. If the blood specimen volume is very small (0.5 mL or less) an appropriate supplement may be required for recovery of these organisms.
- Studies have demonstrated that the resins present in this medium do not adequately neutralize meropenem preparations. Based on the neutralization study with Candida albicans and Fluconazole, the media showed some degree of neutralization; however the results were inconclusive. The adequacy of antifungal neutralization by resins in the BD BACTEC Peds Plus/F vial (plastic) is unknown.
- The default 5-day (120 hours) protocol was utilized for all analytical testing with the BACTEC Peds Plus/F culture media and protocol lengths of >5 days have not been evaluated.
### 4. Special instrument requirements:
BACTEC fluorescent series instruments BACTEC FX, BACTEC FX40, BACTEC 9240 and BACTEC 9050 and were evaluated using software versions listed below:
| Instrument | Software Version |
| --- | --- |
| BACTEC FX | V4.60 |
| BACTEC 9240 | V4.95 |
| BACTEC 9050 | V2.01 |
| BACTEC FX 40 | V2.00A and V2.49G* |
{3}
* V2.49G was the working build and released as version V2.50
# I. Device Description
The blood sample to be tested is inoculated into one or more vials which are inserted into the BACTEC fluorescent series instrument for incubation and periodic reading. Each vial contains a chemical sensor which can detect increases in $\mathrm{CO}_{2}$ produced by the growth of microorganisms. The sensor is monitored by the instrument every ten minutes for an increase in its fluorescence, which is proportional to the amount of $\mathrm{CO}_{2}$ present. A positive reading indicates the presumptive presence of viable microorganisms in the vial. Detection is limited to microorganisms that will grow in a particular type of culture medium.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
BACTEC Peds Plus/F Culture Vials Soybean-Casein Digest Broth with Resins
2. Predicate $510(\mathrm{k})$ number(s):
K954927
3. Comparison with predicate:
Table 1: Comparison with Predicate Device
| Similarities | | |
| --- | --- | --- |
| Item | Device BD BACTEC Peds Plus/F Culture Vials (plastic) | Predicate BACTEC Peds Plus/F Culture Vials (glass) K954927 |
| Intended Use | This device is for qualitative aerobic culture and recovery of microorganisms from human blood, to be used with the BD BACTEC fluorescent series instruments | Same |
| Sample Type | Human blood | Same |
| Sample Volume | 0.5- 5.0 mL of blood (recommended 1.0- 3.0 mL) | 0.5 - 5.0 mL of blood (optimum 1.0 - 3.0 mL) |
| Growth Medium Volume | 40 mL enriched soybean-casein digest broth | Same |
| Instrument | BD BACTEC fluorescent series | Same |
| Detection Technology | Continuous monitoring; measurement of CO2 increase; resins for absorption of antimicrobials; rocking agitation | Same |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | Device
BD BACTEC Peds Plus/F
Culture Vials (plastic) | Predicate
BACTEC Peds Plus/F
Culture Vials (glass)
K954927 |
| | parameters | |
| Incubation | 35°C (± 1.5°C) up to 120 hours | Same |
| Differences | | |
| --- | --- | --- |
| Vial Material | Plastic | Glass |
| Vial Weight | Lighter than glass | - |
| Vial Height | 5.0 inches | 5.6 inches |
| Sensor Adhesion Promoter | Yes | None |
| Vial Sensor | 2.6 gram per vial, specific for the plastic vial geometry | 1.75 gram per vial |
| Sensor Components | BCP (indicator)- 6.5 mg/vial
Red dye- 4.0 mg/vial | BCP (indicator)- 1.8 mg/vial
Red dye- 1.9 mg/vial |
## K. Standard/Guidance Document Referenced (if applicable):
Not Applicable
## L. Test Principle:
The BD BACTEC Peds Plus /F medium is an enriched soybean- casein digest broth, with each vial containing 40 mL of broth. Sodium polyanetholesulfonate (SPS) is added to the medium as an anticoagulant that inhibits bacteriocidal activities in the blood. The concentration of SPS has been adjusted to accommodate the low blood volumes of 0.5- 5.0 mL per vial.
Each BD BACTEC Peds Plus/F culture vial contains a chemical sensor in a silicon rubber base that can detect increases in CO₂ produced by the growth of microorganisms. Low volume (0.5- 5 mL) of blood is inoculated into the BD BACTEC Peds Plus/F blood culture vial, which is inserted into the BD BACTEC Fluorescent Series instrument for incubation, agitation and periodic measurement. When microorganisms are present in the blood sample, they metabolize nutrients in the culture medium, releasing CO₂ into the medium. A dye in the sensor reacts with the CO₂, modulating the amount of light that is absorbed by the fluorescent material in the sensor. The instrument's photo detectors monitor the sensor every 10 minutes and measure the level of fluorescence, which is proportional to the amount of CO₂ present in the vial. Positivity of a vial is determined by algorithms resident in the instrument rack's microprocessor. The algorithms use the rate of CO₂ production as well as the absolute increase in CO₂ to interpret the data.
Culture vials flagged as presumptively positive are removed from the instrument for
{5}
subculture and Gram stain in order to identify the microorganisms for further evaluation and proposed patient treatment. Culture vials that are not flagged as positive remain in the instrument until the test protocol has been completed and negative vials are discarded at the end of protocol (120 hours).
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
The BACTEC Peds Plus/F (plastic) vial was evaluated across three lots in the Time To Detection (TTD) and the Percent Recovery studies. Different lots of key raw materials were used to manufacture each lot of culture media.
The results stratified by lot (combining blood volumes, inoculum levels, organisms, and instruments) are shown in Table 2.
Table 2: Positive Percent Recovery and Time To Detection by Lots
| Plastic lot | Total # | Positive Percent Recovery | Median TTD (hrs) |
| --- | --- | --- | --- |
| 1 | 448 | 89.06% | 20.025 |
| 2 | 448 | 91.74% | 20.445 |
| 3 | 448 | 90.4% | 20.325 |
The analysis was acceptable and there was no statistically significant difference across the lots in the precision study.
#### b. Linearity/assay reportable range:
Not applicable
#### c. Traceability, Stability, Expected values (controls, calibrators, or methods):
##### Quality Control
An internal validation study across three lots with inoculum at 10-100 CFU per vial was conducted using the organisms listed below. Haemophilus influenzae, Neisseria meningitidis, Candida albicans, Pseudomonas aeruginosa, Streptococcus pneumoniae, and Alcaligenes faecalis were evaluated by testing 18 vials for each; Streptococcus pyogenes and Escherichia coli were evaluated by testing 21 vials each; Staphylococcus aureus was evaluated by testing six vials. All organisms were detected ≤72 hours, with mean ranges between 7.5 (Streptococcus pyogenes) to 37.7 hours (Neisseria meningitidis).
5
{6}
Streptococcus pyogenes ATCC 19615
Escherichia coli ATCC 25922
Streptococcus pneumoniae* ATCC 6305
Pseudomonas aeruginosa ATCC 27853
Candida albicans ATCC18804
Neisseria meningitidis ATCC 13090
Alcaligenes faecalis ATCC 8750
Haemophilus influenzae ATCC 19418
Staphylococcus aureus ATCC 25923
*CLSI recommended strain
## d. Detection limit:
The microbial detection limit study was conducted to assess the capability of the culture media to detect low numbers of organisms (expected target level of 0-1 and 1-10 CFU/vial) when present in blood. The study included 15 organisms (12 bacteria and 3 yeast strains) tested at two blood volumes, each at challenging target inoculum levels of 0-1 and 1-10 CFU/vial across three lots with BACTEC FX and BACTEC 9240:
15 org x 2 blood vol x 2 inoculum levels x 3 lots x 2 instrument types = 360
For statistical analysis, 95% two-sided bootstrap confidence intervals for differences were used. The results were stratified by organism and are shown in Table 3.
Table 3: Microbial Detection Limit difference by organism
| Organism Name | Number of Samples | Percent recovery for modified vial (Plastic) | Percent recovery for predicate vial (Glass) | Difference in Percent Recovery between Plastic and Glass | 95% CI lower bound | 95% CI upper bound |
| --- | --- | --- | --- | --- | --- | --- |
| Candida albicans | 24 | 75 | 75 | 0 | -23.1 | 23.1 |
| Candida glabrata | 24 | 83.33 | 79.17 | 4.17 | -20.28 | 28.61 |
| Cryptococcus neoformans | 24 | 75 | 83.33 | -8.33 | -28.06 | 11.39 |
| Enterococcus faecalis | 24 | 66.67 | 66.07 | 0 | -20 | 20 |
| Escherichia coli | 24 | 70.83 | 75 | -4.17 | -28.61 | 20.28 |
| Haemophilus influenzae | 24 | 75 | 79.17 | -4.17 | -28.61 | 20.28 |
| Haemophilus parainfluenzae biotype I | 24 | 41.67 | 45.83 | -4.17 | -22.35 | 14.02 |
| Micrococcus luteus | 24 | 50 | 45.83 | 4.17 | -22.87 | 31.2 |
| Neisseria | 24 | 87.5 | 75 | 12.5 | -8.52 | 33.52 |
{7}
| Organism Name | Number of Samples | Percent recovery for modified vial (Plastic) | Percent recovery for predicate vial (Glass) | Difference in Percent Recovery between Plastic and Glass | 95% CI lower bound | 95% CI upper bound |
| --- | --- | --- | --- | --- | --- | --- |
| gonorrhoeae | | | | | | |
| Neisseria meningitidis | 24 | 79.17 | 79.17 | 0 | -25.83 | 25.53 |
| Pseudomonas aeruginosa | 24 | 100 | 70.83 | 29.17 | 7.62 | 50.71 |
| Staphylococcus aureus | 24 | 83.33 | 58.33 | 25 | 1.19 | 48.81 |
| Staphylococcus epidermidis | 24 | 54.17 | 58.33 | -4.17 | -28.61 | 20.28 |
| Streptococcus pneumoniae | 24 | 45.83 | 20.83 | 25 | -1.46 | 51.46 |
| Streptococcus sanguinis | 24 | 66.67 | 79.17 | -12.5 | -36.48 | 11.48 |
At the low target inoculum level of 1-10 CFU/vial, there were:
- Three Haemophilus parainfluenzae recovery failures with four CFUs on the subculture plate in both modified plastic and predicate glass vials
- One Neisseria meningitidis recovery failure with five CFUs on the subculture plate in the modified plastic vial.
Both organisms are included in the limitation section of the Package Insert.
Recommendation for each is included. For $N$ meningitidis, there were no recovery failures with the higher volume of $5\mathrm{mL}$ of blood. For $H$ parainfluenzae, see "False Negatives" section below for details.
Data was also evaluated at target inoculum level of 0-1 and 1-10 CFU/vial across two blood volumes (0.5 and $5.0\mathrm{mL}$ ) in the Time to Detection (TTD) study as shown in Table 4.
Table 4: Time To Detection at target 0-1, 1-10 CFU across 0.5 and ${5.0}\mathrm{\;{mL}}$ blood
| Target Inoculum level | Blood Volume | Median TTD (Plastic) | Median TTD (Glass) | Median TTD difference (Plastic - Glass) |
| --- | --- | --- | --- | --- |
| 0-1 CFU | 0.5mL | 26.4 | 27.60 | -0.165 |
| | 5mL | 32.65 | 33.69 | -0.995 |
| 1-10 CFU | 0.5mL | 28.52 | 30.67 | -0.5 |
| | 5mL | 24.81 | 25.03 | -0.33 |
In summary, both data sets demonstrated that the modified plastic device performed equivalently when compared to the predicate glass device.
# e. Analytical specificity:
Not applicable
{8}
f. Assay cut-off:
Not applicable
## 2. Comparison studies:
Performance of the BD BACTEC Peds Plus/F (plastic) blood culture vials was evaluated in seeded internal analytical studies to demonstrate comparable performance to the predicate device, the BD BACTEC Peds Plus/F (glass) blood culture vials. Comparison results were acceptable. The comparisons were made using the following parameters: Time to detection, percent recovery, false negative rate, false positive rate and antimicrobial neutralization capability. For statistical analysis, 95% two-sided bootstrap confidence intervals for differences were used.
a. Method comparison with predicate device:
## Instrument Time to Detection (TTD) study
The TTD analysis was evaluated in the combined Percent recovery and the Microbial Detection Limit studies using the BD BACTEC FX and 9240 instruments at three inoculum levels, across two blood volumes (0.5 mL and 5.0 mL) over three media lots. The Percent Recovery study represented the standard inoculum of 10 to 100 CFU per vial and the Microbial Detection Limit study represented the challenge inoculum levels of 0 to 1 and 1 to 10 CFU per vial. The performance is demonstrated in Table 5.
Table 5: Summary of TTD study
| | | Median of TTD for Plastic (95% CI) | Median of TTD for Glass (95% CI) | Median of TTD difference for Plastic - Glass (95% CI) |
| --- | --- | --- | --- | --- |
| Inoculum Levels | 0 to 1 | 30.908 (23.645, 46.965) | 31.28 (23.81, 42.025) | -0.75 (-1.75, 0.085) |
| | 1 to 10 | 26.733 (23.205, 30.255) | 27.84 (23.655, 31.785) | -0.415 (-0.75, 0) |
| | 10 to 100 | 24.135 (21.31, 27.47) | 24.425 (21.435, 27.09) | -0.08 (-0.33, 0.25) |
| Blood Volume | 0.5 mL | 27.34 (23.65, 31.075) | 27.862 (24.34, 31.395) | -0.25 (-0.585, 0.25) |
| | 5 mL | 25.165 (22.255, 27.85) | 25.105 (22.18, 28.31) | -0.33 (-0.5, -0.08) |
| Media Lots | Lot 1 | 26.265 (22.31, 31.01) | 27.415 (23.045, 31.59) | -0.415 (-0.92, -0.075) |
| | Lot 2 | 25.935 (22.29, 30.1) | 25.52 (21.925, 29.605) | -0.08 (-0.415, 0.335) |
| | Lot 3 | 25.6 (22.375, 29.65) | 26.315 (22.78, 30.02) | -0.335 (-0.745, 0.08) |
For all inoculum levels data combined, it was observed that the estimated median for the modified plastic vial and the predicate glass vial was 17.5 hours and 18.11 hours respectively; the median TTD difference was -0.25 hours, with 95% CI (-0.495, -0.08),
{9}
favoring the modified plastic vial. The TTD study demonstrated that the modified plastic device performed equivalently when compared to the predicate glass device.
# Recovery (Detection) Study
The percent recovery (detection) was evaluated in a study of 984 paired sets at the standard inoculum level of 10- 100 CFU/vial on four instruments across three lots using a diverse set of microorganisms frequently isolated in blood. Of the 984 paired sets, 953 sets recovered organisms in both the plastic vial and the glass vial and 18 were negative by both vials. Nine paired sets were recovered in the plastic vials but not in the glass vials and four were recovered in the glass vials but not in the plastic vials. The performance is demonstrated in Table 6A and 6B.
Table 6A: Recovery Study Results
| | Predicate (Glass) | | | |
| --- | --- | --- | --- | --- |
| | | Detected | Not Detected | Total |
| Modified (Plastic) | Detected | 953 | 9 | 962 |
| | Not Detected | 4 | 18 | 22 |
| | Total | 957 | 27 | 984 |
In summary, the percent recovery was:
- $97.76\%$ , (962)/(984) for the modified plastic vial
- $97.26\%$ , (957)/(984) for the predicate glass vial
The percent recovery study demonstrated that the modified plastic vial performed equivalently when compared to the predicate glass vial.
The performance was evaluated for each organism and this information is shown in Table 6B.
Table 6B: Percent Recovery (10- 100 CFU/vial) Summary by Organisms
| Organism Name | Number of Samples | Percent recovery for Modified vial (Plastic) | Percent recovery for Predicate vial (Glass) | Difference in recovery between Plastic and Glass | 95% CI Lower bound | 95% CI Upper bound |
| --- | --- | --- | --- | --- | --- | --- |
| Abiotrophia defectiva | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Acinetobacter lwoffii | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Aerococcus viridans | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Aggregatibacter actinomycetemcomitans | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Alcaligenes faecalis | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Bacillus subtilis | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Candida albicans | 24 | 75 | 70.83 | 4.17 | -17.38 | 25.71 |
| Candida glabrata | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Cardiobacterium hominis | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Corynebacterium jeikeium | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
{10}
| Organism Name | Number of Samples | Percent recovery for Modified vial (Plastic) | Percent recovery for Predicate vial (Glass) | Difference in recovery between Plastic and Glass | 95% CI Lower bound | 95% CI Upper bound |
| --- | --- | --- | --- | --- | --- | --- |
| Cryptococcus neoformans | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Eikenella corrodens | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Enterobacter cloacae | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Enterococcus faecalis | 24 | 95.83 | 100 | -4.17 | -18.21 | 9.88 |
| Escherichia coli | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Granulicatella adiacens | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Haemophilus influenzae | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Haemophilus influenzae biotype | 48 | 79.17 | 79.17 | 0 | -8.17 | 8.17 |
| Haemophilus parainfluenzae biotype I | 24 | 79.17 | 62.5 | 16.67 | -2.19 | 35.53 |
| Kingella kingae | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Klebsiella pneumoniae | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Leuconostoc citreum | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Micrococcus luteus | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Neisseria gonorrhoeae | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Neisseria meningitidis | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Pediococcus acidilactici | 24 | 100 | 95.83 | 4.17 | -9.88 | 18.21 |
| Proteus mirabilis | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Providencia stuartii | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Pseudomonas aeruginosa | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Rothia mucilaginosa | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Staphylococcus aureus | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Staphylococcus epidermidis | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Stenotrophomonas maltophilia | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Streptococcus agalactiae (Group B Strep) | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Streptococcus pneumoniae | 96 | 100 | 100 | 0 | -2.89 | 2.89 |
| Streptococcus pyogenes (Group A Strep.) | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
| Streptococcus sanguinis | 24 | 100 | 100 | 0 | -11.55 | 11.55 |
# False Positive Rates (Instrument-positive, subculture-negative)
False positivity was assessed with vials inoculated with fresh human blood of 0.5, 3, and $5\mathrm{mL}$ , but no organisms were added to the vials. There were a total of 288 pair sets across three lots using BACTEC FX and BACTEC 9240 to be completed at the protocol of 120 hours incubation.
48 vials x 3 lots x 2 instruments= 288 vials per device
No instrument false positive signals were detected. The modified device performed equivalently when compared to the predicate device in the false positive study.
{11}
# False Negative Rates (Instrument-negative, subculture positive)
All inoculated paired sets (1344; 984+ 360) that were instrument negative at the end of protocol (120 hours) were subcultured onto appropriate culture media plates. This combined data set was evaluated for the false negative rates. A false negative is a vial that was instrument-negative at the end of protocol yet contains viable organisms upon subculturing onto appropriate culture media. There were a total of 83 paired sets where both the modified and the predicate devices were negative at 120 hours; there were 46 sets where the predicate device only detected (i.e. 46 plastic vials subcultured) and 66 vials where the new device detected (i.e. 63 glass vials subcultured, 3 vials discarded inadvertently).
$(83\times 2) + 46 + 63 = 275$
Of the 275 vials subcultured, 129 were plastic vials and 146 were glass vials. There were 20 false negatives (1.49%, 20/1344) from the plastic vials and 25 false negatives (1.86%, 25/1344) from the glass vials. There was no significant difference in false negative rates between the predicate and the modified devices. The false negative rates with 95% CI were demonstrated in Table 7.
Table 7: False Negative Rate (BACTEC Instrument Negative, Terminal Subculture Positive)
| Total paired sets | Percent False Neg Rate for Modified (Plastic) | Percent False Neg Rate for Predicate (Glass) | Difference in False Rates between Plastic and Glass | 95% CI (lower bound) | 95% CI (Upper bound) |
| --- | --- | --- | --- | --- | --- |
| 1344 | 1.49 (20/1344) | 1.86 (25/1344) | -0.37 | -1.01 | 0.26 |
The majority of the false negatives were Haemophilus species with $0.5\mathrm{mL}$ of bagged blood added. The Haemophilus species were detected at target inoculum of 1-10, 10-100 CFU with increased volume of $5\mathrm{mL}$ blood. Further, the original set of Haemophilus species were re-tested with $0.5\mathrm{mL}$ of fresh blood and all detected in both vials.
The Haemophilus species false negatives are noted in the limitations section of the Package Insert (PI). The recommended limitation states: "If the blood specimen volume is very small (0.5 mL or less), an appropriate supplement may be required for recovery of these organisms."
# BD BACTEC Instrument Compatibility Study
The BACTEC instrument compatibility study was evaluated from the Percent Recovery study (i.e. 984 paired sets) dataset across four fluorescent- series instruments: BD BACTEC FX, FX40, 9240, and 9050. The study included 0.5 and $5.0\mathrm{mL}$ of blood at inoculum level of 10-100CFU per vial. There were 953 sets recovered organisms in both the plastic vial and the glass vial and the Time To Detection (TTD) comparison between the modified plastic and the predicate glass vials was evaluated. The performance data is shown in in Table 8.
{12}
Table 8: TTD Difference (Modified Plastic TTD- Predicate Glass TTD)
| Instrument | Blood (mL) | Detected in modified and Predicate | Median Difference (hour) |
| --- | --- | --- | --- |
| BACTEC FX | 0.5, 5 | 238 | -0.250 |
| BACTEC FX40 | 0.5, 5 | 240 | -0.333 |
| BACTEC 9240 | 0.5, 5 | 235 | -0.250 |
| BACTEC 9050 | 0.5, 5 | 240 | 0.000 |
| Combined | | 953 | -0.249 |
The study demonstrated that the four instruments performed equivalently and they are compatible.
# Antimicrobial Neutralization Capability
The study was to demonstrate the nonionic and cationic resins in the culture media to enhance the recovery of organisms by adsorption of commonly used antibiotics in the blood samples. The non-resin vials, BACTEC Standard /10 Aerobic/F were inoculated at the same time to demonstrate the antimicrobial activity (no growth) in the vial. Eleven antimicrobials representative of their classes were evaluated at the MIC level of selected organisms. The antimicrobials evaluated were: Meropenem, Gentamicin, Ciprofloxacin, Clindamycin, Vancomycin, Cefazolin, Cefoxitin, Cefotaxime, Cefepime, Piperacillin/Tazobactam and Fluconazole. The performance data is shown in Table 9.
Table 9: Antimicrobial-Organism Recovery Comparison at MIC level
| | | | | | Number Detected | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Antimicrobial | Organism | Strain | MIC | Test conc. (μg/mL) | Glass (resin) | Plastic (resin) | Non-resin medium |
| Meropenem | Enterococcus faecalis | 29212 | 4 | 5 | 3 | 3 | 0 |
| | Staphylococcus aureus | 25923 | 0.065 | 0.65 | 0 | 1 | 0 |
| Gentamycin | Escherichia coli | 25922 | 2 | 4 | 3 | 3 | 0 |
| | Staphylococcus aureus | 25923 | 0.25 | 1 | 3 | 3 | 0 |
| Ciprofloxacin | Staphylococcus aureus | 25923 | 0.5 | 1 | 3 | 3 | 0 |
| Clindamycin | Escherichia coli | 25922 | 4 | 4 | 3 | 3 | 3 |
| | Staphylococcus aureus | 25923 | 0.125 | 1 | 3 | 3 | 0 |
| Vancomycin | Staphylococcus aureus | 25923 | 2 | 5 | 3 | 3 | 0 |
| | Enterococcus faecalis | 29212 | 4 | 5 | 3 | 3 | 0 |
| Cefazolin | Escherichia coli | 25922 | 0.5 | 4 | 3 | 3 | 0 |
| | Staphylococcus aureus | 25923 | 2 | 4 | 3 | 3 | 0 |
{13}
| | | | | | Number Detected | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Antimicrobial | Organism | Strain | MIC | Test conc. (μg/mL) | Glass (resin) | Plastic (resin) | Non-resin medium |
| Cefoxitin | Escherichia coli | 25922 | 4 | 5 | 3 | 3 | 0 |
| | Staphylococcus aureus | 25923 | 2 | 4 | 3 | 3 | 0 |
| Cefotaxime | Staphylococcus aureus | 25923 | 1 | 2 | 3 | 3 | 0 |
| Cefepime | Staphylococcus aureus | 25923 | 1 | 1.5 | 3 | 3 | 0 |
| Piperacillin/Tazobactam | Enterobacter aerogenes | 13048 | 4 | 4 | 3 | 3 | 0 |
| Fluconazole | Candida albicans | 18804 | 1 | 4 | 3 | 3 | 3 |
The study indicated some degree of neutralization with Candida albicans and fluconazole. However, the results were inconclusive for the antifungal neutralization in the BACTEC Peds Plus/F culture media. The statement below is included in the limitation section of the PI:
"Based on the neutralization study with Candida albicans and Fluconazole, the media showed some degree of neutralization; however the results were inconclusive. The adequacy of antifungal neutralization by resins in the BD BACTEC Peds Plus/F vial (plastic) is unknown."
E. coli is intrinsically resistant to clindamycin and the result was as expected. The modified device performed equivalently when compared to the predicate device during antimicrobial neutralization evaluation.
# b. Matrix comparison:
In seeded analytical studies, the performance of BD BACTEC Peds Plus/F culture medium in plastic vial was compared to that in glass vial, with two human blood volumes, 15 common blood bloodstream pathogens (bacteria and yeast) across four fluorescent series instruments: BACTEC FX, FX40, BACTEC 9240, and BACTEC 9050.
# 3. Clinical studies:
Not applicable; seeded analytical studies to compare the new plastic blood culture vials to the glass blood culture vials (predicate).
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
{14}
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Seeded analytical studies demonstrated equivalent performance of the BD BACTEC Peds Plus/F (plastic) blood culture medium when compared to the BD BACTEC Peds Plus/F (glass) blood culture medium.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.