Positive Blood Culture Processor For Inoculum Preparation Used For Antimicrobial Susceptibility Testing

Page Type
Product Code
Definition
Processing device that includes associated reagents, when applicable, that is intended to recover microbes from positive blood cultures for inoculum preparation. The prepared inoculum is intended for subsequent use in antimicrobial susceptibility testing.
Physical State
In vitro diagnostic device
Technical Method
Processor to prepare an inoculum for subsequent use in antimicrobial susceptibility testing.
Target Area
For antimicrobial susceptibility testing using an inoculum of microbial cells recovered from positive blood cultures.
Regulation Medical Specialty
Microbiology
Review Panel
Microbiology
Submission Type
510(K)
Device Classification
Class 2
Regulation Number
866.1650
GMP Exempt?
No
Summary Malfunction Reporting
Ineligible
Implanted Device
No
Life-Sustain/Support Device
No
Third Party Review
Not Third Party Eligible

CFR § 866.1650 A cellular analysis system for multiplexed antimicrobial susceptibility testing

§ 866.1650 A cellular analysis system for multiplexed antimicrobial susceptibility testing.

(a) Identification. A cellular analysis system for multiplexed antimicrobial susceptibility testing is a multiplex qualitative and/or quantitative in vitro diagnostic device intended for the identification and determination of the antimicrobial susceptibility results of organisms detected in samples from patients with suspected microbial infections. This device is intended to aid in the determination of antimicrobial susceptibility or resistance when used in conjunction with other laboratory findings.

(b) Classification. Class II (special controls). The special controls for this device are:

(1) Design verification and validation must include:

(i) Detailed device description documentation, including the device components, ancillary reagents required but not provided, a detailed explanation of the methodology, including primer/probe sequence, design, rationale for sequence selection, and details of the antimicrobial agents, as applicable.

(ii) Detailed documentation from the following analytical and clinical performance studies: limit of detection, inclusivity, precision, reproducibility, interference, cross-reactivity, carryover, and cross-contamination, quality control and additional studies, as applicable to specimen type and assay intended use.

(iii) Detailed documentation from an appropriate clinical study. The study, performed on a study population consistent with the intended use population, must compare the device performance to results obtained from well-accepted reference methods.

(iv) Detailed documentation for device software, including software applications and hardware-based devices that incorporate software.

(2) The labeling required under § 809.10(b) of this chapter must include:

(i) Limitations and protocols regarding the need for correlation of results by standard laboratory procedures, as applicable.

(ii) A detailed explanation of the interpretation of results and acceptance criteria.

(iii) A detailed explanation of the principles of operation and procedures for assay performance and troubleshooting.

[90 FR 24963, June 13, 2025]

Positive Blood Culture Processor For Inoculum Preparation Used For Antimicrobial Susceptibility Testing

Page Type
Product Code
Definition
Processing device that includes associated reagents, when applicable, that is intended to recover microbes from positive blood cultures for inoculum preparation. The prepared inoculum is intended for subsequent use in antimicrobial susceptibility testing.
Physical State
In vitro diagnostic device
Technical Method
Processor to prepare an inoculum for subsequent use in antimicrobial susceptibility testing.
Target Area
For antimicrobial susceptibility testing using an inoculum of microbial cells recovered from positive blood cultures.
Regulation Medical Specialty
Microbiology
Review Panel
Microbiology
Submission Type
510(K)
Device Classification
Class 2
Regulation Number
866.1650
GMP Exempt?
No
Summary Malfunction Reporting
Ineligible
Implanted Device
No
Life-Sustain/Support Device
No
Third Party Review
Not Third Party Eligible

CFR § 866.1650 A cellular analysis system for multiplexed antimicrobial susceptibility testing

§ 866.1650 A cellular analysis system for multiplexed antimicrobial susceptibility testing.

(a) Identification. A cellular analysis system for multiplexed antimicrobial susceptibility testing is a multiplex qualitative and/or quantitative in vitro diagnostic device intended for the identification and determination of the antimicrobial susceptibility results of organisms detected in samples from patients with suspected microbial infections. This device is intended to aid in the determination of antimicrobial susceptibility or resistance when used in conjunction with other laboratory findings.

(b) Classification. Class II (special controls). The special controls for this device are:

(1) Design verification and validation must include:

(i) Detailed device description documentation, including the device components, ancillary reagents required but not provided, a detailed explanation of the methodology, including primer/probe sequence, design, rationale for sequence selection, and details of the antimicrobial agents, as applicable.

(ii) Detailed documentation from the following analytical and clinical performance studies: limit of detection, inclusivity, precision, reproducibility, interference, cross-reactivity, carryover, and cross-contamination, quality control and additional studies, as applicable to specimen type and assay intended use.

(iii) Detailed documentation from an appropriate clinical study. The study, performed on a study population consistent with the intended use population, must compare the device performance to results obtained from well-accepted reference methods.

(iv) Detailed documentation for device software, including software applications and hardware-based devices that incorporate software.

(2) The labeling required under § 809.10(b) of this chapter must include:

(i) Limitations and protocols regarding the need for correlation of results by standard laboratory procedures, as applicable.

(ii) A detailed explanation of the interpretation of results and acceptance criteria.

(iii) A detailed explanation of the principles of operation and procedures for assay performance and troubleshooting.

[90 FR 24963, June 13, 2025]