VITEK® 2 Streptococcus Tetracycline (<=0.25 - >=16 µg/mL)

K223481 · bioMerieux, Inc. · LON · Feb 3, 2023 · Microbiology

Device Facts

Record IDK223481
Device NameVITEK® 2 Streptococcus Tetracycline (<=0.25 - >=16 µg/mL)
ApplicantbioMerieux, Inc.
Product CodeLON · Microbiology
Decision DateFeb 3, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2
AttributesAI/ML

Intended Use

VITEK® 2 Streptococcus Tetracycline is designed for antimicrobial susceptibility testing of Streptococcus species and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Streptococcus Tetracycline is a quantitative test. Tetracycline has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections: Streptococcus pneumoniae Streptococcus pyogenes* *The VITEK® 2 Streptococcus Susceptibility Card also reports the susceptibility of the following additional organisms as listed on the FDA Susceptibility Test Interpretative Criteria website (STIC): Streptococcus spp. B-Hemolytic Group (other than S. pyogenes). The VITEK® 2 Streptococcus Susceptibility Card is intended for use with the VITEK® 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Streptococcus pneumoniae, beta-hemolytic Streptococcus, and Viridans Streptococcus to antimicrobial agents when used as instructed.

Device Story

VITEK 2 AST-Streptococcus Tetracycline card; miniaturized, automated doubling dilution technique for MIC determination. Input: bacterial isolate diluted in saline. Operation: card contains 64 wells with premeasured antimicrobial concentrations; system fills, seals, and incubates card; optical scanner measures light attenuation every 15 minutes to monitor growth. Output: MIC value and interpretive category (S, I, R). Used in clinical laboratories; operated by technicians. Results assist clinicians in selecting appropriate antimicrobial therapy for Streptococcus infections. This submission updates device labeling to reflect current FDA-recognized breakpoints.

Clinical Evidence

Bench testing only. Re-analysis of 1127 clinical/challenge isolates (auto-dilution) and 163 challenge isolates (manual dilution) from original submission (K111893), plus 63 additional challenge isolates. Primary endpoints: Essential Agreement (EA) and Category Agreement (CA) against CLSI broth microdilution reference. Results: Combined EA 98.3% (S. pneumoniae), 99.7% (S. pyogenes), 97.7% (S. spp. β-Hemolytic). CA 99.7%, 100%, and 96.0% respectively. Minor discrepancies noted; limitation added for S. dysgalactiae equisimilis.

Technological Characteristics

Miniaturized 64-well AST card; automated growth-based detection via light transmittance optics. Energy source: visible light scanner. Connectivity: VITEK 2/VITEK 2 Compact systems. Software: Discriminant Analysis algorithm. Sterilization: Not specified.

Indications for Use

Indicated for antimicrobial susceptibility testing of Streptococcus pneumoniae, Streptococcus pyogenes, and other beta-hemolytic Streptococcus species and Viridans Streptococcus in clinical laboratories using VITEK 2 or VITEK 2 Compact systems.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a bold, sans-serif font and the full name in a smaller, sans-serif font below it. The overall design is clean and professional, reflecting the agency's role in regulating food and drugs in the United States. February 3, 2023 bioMérieux, Inc Jared Bronson Regulatory Affairs Specialist 595 Anglum Rd. Hazelwood, Missouri 63042 Re: K223481 Trade/Device Name: VITEK 2 Streptococcus Tetracycline (<0.25 - ≥16 ug/mL) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System Regulatory Class: Class II Product Code: LON Dated: November 17, 2022 Received: November 18, 2022 Dear Jared Bronson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR {1}------------------------------------------------ 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Ribhi Shawar -S Ribhi Shawar, Ph.D. (ABMM) Branch Chief General Bacteriology and Antimicrobial Susceptibility Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K223481 Device Name VITEK® 2 Streptococcus Tetracycline (≤0.25 ->16 ug/mL) #### Indications for Use (Describe) VITEK® 2 Streptococcus Tetracycline is designed for antimicrobial susceptibility testing of Streptococcus species and is intended for use with the VITEK® 2 and VITEK® 2 COMPACT Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK®2 Streptococcus Tetracycline is a quantitative test. Tetracycline has been shown to be active against the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections: Streptococcus pneumoniae Streptococcus pyogenes* *The VITEK® 2 Streptococcus Susceptibility Card also reports the susceptibility of the following additional organisms as listed on the FDA Susceptibility Test Interpretative Criteria website (STIC): Streptococcus spp. B-Hemolytic Group (other than S. pyogenes). The VITEK® 2 Streptococcus Susceptibility Card is intended for use with the VITEK® 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Streptococcus pneumoniae, beta-hemolytic Streptococcus, and Viridans Streptococcus to antimicrobial agents when used as instructed. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <div style="display:inline-block;"><input checked="" type="checkbox"/></div> Prescription Use (Part 21 CFR 801 Subpart D) | |---------------------------------------------------------------------------------------------------------------------------| | <div style="display:inline-block;"><input type="checkbox"/></div> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for bioMérieux. The logo is a circle with the top half in dark blue and the bottom half in a gradient of yellow to green. The word "BIOMÉRIEUX" is written in white letters in the center of the blue portion of the circle. ### 510(k) SUMMARY ## VITEK® 2 Streptococcus Tetracycline (≤ 0.25 -> 16 µg/mL) #### 510(k) Submission Information: | Submitter's Name: | bioMérieux, Inc. | |----------------------|----------------------------------------------------------------------------------------------------------------------------| | Address: | 595 Anglum Road<br>Hazelwood, MO 63042 | | Contact Person: | Jared Bronson<br>Regulatory Affairs Specialist | | Phone Number: | 314-731-8799 | | Fax Number: | 314-731-8689 | | Date of Preparation: | February 02, 2023 | | B. Device Name: | | | Formal/Trade Name: | VITEK® 2 Streptococcus Tetracycline (≤ 0.25 - ≥ 16<br>µg/mL) | | Classification Name: | 21 CFR 866.1645<br>Fully Automated Short-Term Incubation Cycle<br>Antimicrobial Susceptibility System<br>Product Code: LON | | Common Name: | VITEK® 2 Streptococcus Tetracycline | | C. Predicate Device: | VITEK® 2 Streptococcus Tetracycline (≤ 0.25 - ≥ 16<br>µg/mL) (K111893) | ### D. Device Description: The principle of the VITEK® 2 AST cards is based on the microdilution minimum inhibitory concentration (MIC) technique reported by MacLowry and Marsh(1) and Gerlach(2). The VITEK® 2 AST card is essentially a miniaturized, abbreviated and automated version of the doubling dilution technique(3). {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for bioMérieux, a French multinational biotechnology company. The logo consists of a circle divided horizontally into two sections. The top section is a dark blue color and contains the company name "BIOMÉRIEUX" in white, sans-serif capital letters. The bottom section is a gradient of yellow to green. Each VITEK® 2 AST card contains 64 wells. A control well which only contains microbiological culture media is resident on all cards. The remaining wells contain premeasured portions of a specific antibiotic combined with culture media. The bacterial or yeast isolate to be tested is diluted to a standardized concentration with 0.45 -0.5% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling, sealing and loading operation. The VITEK® 2 Systems monitor the growth of each well in the card over a defined period of time. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card. VITEK® 2 Streptococcus Tetracycline has the following concentrations in the card: 0.125, 0.5, 1, and 4 ug/mL (equivalent standard method concentration by efficacy in ug/mL). # E. Substantial Equivalence Information: The similarities and differences of the VITEK® 2 Streptococcus Tetracycline when compared to the predicate device, VITEK® 2 Streptococcus Tetracycline (K111893), are described in the following table. The differences between both devices include the addition of Streptococcus pneumoniae and Beta hemolytic streptococcus spp. breakpoints and removal of Streptococcus spp. breakpoint, as well as narrowing the indications for use to Streptococcus pneumoniae and Streptococcus pyogenes. | Device and Predicate<br>Device(s): | Device: | Predicate Device: K111893 | |------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Trade Name | VITEK® 2 Streptococcus Tetracycline<br>(≤ 0.25 – ≥ 16 µg/mL) | VITEK® 2 Streptococcus Tetracycline | | Intended Use | VITEK® 2 Streptococcus Tetracycline is<br>designed for antimicrobial<br>susceptibility testing of <i>Streptococcus</i><br>species and is intended for use with the<br>VITEK® 2 and VITEK® 2 Compact<br>Systems as a laboratory aid in the<br>determination of <i>in vitro</i> susceptibility<br>to antimicrobial agents. VITEK® 2<br>Streptococcus Tetracycline is a<br>quantitative test. Tetracycline has been<br>shown to be active against most strains<br>of the microorganisms listed below,<br>according to the FDA label for this<br>antimicrobial. | VITEK® 2 Streptococcus Tetracycline is<br>designed for antimicrobial susceptibility<br>testing of <i>Streptococcus</i> species.<br>VITEK® 2 Streptococcus Tetracycline is<br>a quantitative test intended for use with<br>the VITEK® 2 and the VITEK® 2<br>Compact Systems as a laboratory aid in<br>the determination of <i>in vitro</i><br>susceptibility to antimicrobial agents.<br>Tetracycline has an antimicrobial activity<br>against the microorganisms listed below,<br>according to the FDA label for this<br>antimicrobial. | | | The VITEK® 2 Streptococcus<br>Susceptibility Card is intended for use<br>with the VITEK® 2 Systems in clinical<br>laboratories as an <i>in vitro</i> test to<br>determine the susceptibility of<br><i>Streptococcus pneumoniae</i><br><i>Streptococcus pneumoniae</i> , beta-<br>hemolytic <i>Streptococcus</i> , and<br><i>Viridans Streptococcus</i> to<br>antimicrobial agents when used as<br>instructed. | The VITEK® 2 Antimicrobial<br>Susceptibility Test (AST) is intended to<br>be used with the VITEK® 2 System for<br>the automated quantitative or qualitative<br>susceptibility testing of isolated colonies<br>for the most clinically significant aerobic<br>gram-negative bacilli, <i>Staphylococcus</i><br>spp., <i>Enterococcus</i> spp., <i>Streptococcus</i><br>spp. and clinically significant yeast. | | Test Methodology | Automated quantitative antimicrobial<br>susceptibility test for use with the<br>VITEK® 2 and VITEK® 2 Compact<br>Systems to determine the <i>in vitro</i><br>susceptibility of microorganisms | Same | | Antimicrobial Agent | Tetracycline | Same | | Inoculum | Saline suspension of organism | Same | | Test Card | <i>Streptococcus</i> (AST-ST) Susceptibility<br>Card | Same | | Analysis Algorithms | Discriminant Analysis | Same | | Instrument | VITEK® 2 and VITEK® 2 Compact<br>Systems | Same | | Concentrations | 0.125, 0.5, 1, 4 | Same | | | General Device Characteristic Differences | | | Breakpoints | <i>Streptococcus pneumoniae</i><br>$S ≤ 1$ , I 2, $R ≥ 4$<br><br>Beta hemolytic <i>Streptococcus</i> species<br>(including <i>S. pyogenes</i> )<br>$S ≤ 2$ , I 4, $R ≥ 8$ | <i>Streptococcus</i> spp.<br>$S ≤ 4$ , I 8, $R ≥ 16$ | | Indications for Use | <i>Active in vitro</i> and <i>in clinical</i><br><i>infections</i> :<br><i>Streptococcus pneumoniae</i><br><i>Streptococcus pyogenes</i> *<br><br>*The VITEK® 2 <i>Streptococcus</i><br>Susceptibility Card also reports the<br>susceptibility of the following<br>additional organisms as listed on the<br>FDA Susceptibility Test<br>Interpretative Criteria website | <i>Active in vitro</i> and <i>in clinical infections</i> :<br><i>Streptococcus pneumoniae</i><br><i>Streptococcus pyogenes</i><br><i>Viridans</i> group <i>streptococci</i> | ### Table 1: Substantial Equivalence {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for bioMérieux. The logo is a circle that is split into two sections. The top section is blue and contains the word "BIOMÉRIEUX" in white sans-serif font. The bottom section is a gradient of yellow and green. # VITEK® 2 Streptococcus Tetracycline {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for bioMérieux, a French multinational biotechnology company. The logo consists of a blue circle on top and a yellow-green gradient circle on the bottom. The company name, "BIOMÉRIEUX", is written in white, sans-serif font in the center of the blue circle. | (STIC): Streptococcus spp. ß- | | |--------------------------------|--| | Hemolytic Group (other than S. | | | pyogenes). | | ### F. Intended Use: VITEK® 2 Streptococcus Tetracycline is designed for antimicrobial susceptibility testing of Streptococcus species and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Streptococcus Tetracycline is a quantitative test. Tetracycline has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections: Streptococcus pneumoniae Streptococcus pyogenes* *The VITEK® 2 Streptococcus Susceptibility Card also reports the susceptibility of the following additional organisms as listed on the FDA Susceptibility Test Interpretative Criteria website (STIC): Streptococcus spp. B-Hemolytic Group (other than S. pyogenes). The VITEK® 2 Streptococcus Susceptibility Card is intended for use with the VITEK® 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Streptococcus pneumoniae, beta-hemolytic Streptococcus, and Viridans Streptococcus to antimicrobial agents when used as instructed. # G. Performance Overview and Conclusion: VITEK® 2 Streptococcus Tetracycline demonstrated substantially equivalent performance when compared with the Broth Microdilution reference method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA (Issued August 28, 2009). The Premarket Notification (510[k]) presents data in support of VITEK® 2 Streptococcus Tetracycline. An external evaluation was conducted with fresh and stock clinical isolates, as well as a set of challenge strains. The external evaluations were designed to confirm the acceptability of VITEK® 2 Streptococcus Tetracycline by comparing its performance with the CLSI broth {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for bioMérieux. The logo is a circle with a blue top half and a green bottom half. The word "BIOMÉRIEUX" is written in white letters across the blue portion of the circle. microdilution reference method incubated at 16-24 hrs. The data is representative of performance on both the VITEK® 2 and VITEK® 2 Compact instrument platforms. VITEK® 2 Streptococcus Tetracycline (≤0.25 - ≥16 µg/mL) demonstrated acceptable performance of 98.3% overall Essential Agreement and 97.9% overall Category Agreement with the reference method as presented in Table 2 below: | Antimi-<br>crobial | Comment | Essential Agreement Category | | | | Category Agreement | | | | |--------------------|-------------------------------------------------------------------|------------------------------|-----|-----|-----|---------------------|-----------------|-----------------|------------------| | | | % Error | | | | % Error | | | | | | | %EA | VME | ME | mE | %CA | VME | ME | mE | | Tetracycline | Streptococcus pneumoniae. | (284/289)<br>98.3% | N/A | N/A | N/A | (288/289)<br>99.7% | (0/83)<br>0.0% | (0/206)<br>0.0% | (1/289)<br>0.3% | | | Streptococcus pyogenes. | (307/308)<br>99.7% | N/A | N/A | N/A | (308/308)<br>100.0% | (0/28)<br>0.0% | (0/280)<br>0.0% | (0/308)<br>0.0% | | | Beta-hemolytic Streptococcus (Other than Streptococcus pyogenes). | (518/530)<br>97.7% | N/A | N/A | N/A | (509/530)<br>96.0% | (2/348)<br>0.6% | (1/166)<br>0.6% | (18/530)<br>3.4% | # Table 2: VITEK® 2 Streptococcus Tetracycline Performance Reproducibility and Quality Control demonstrated acceptable results. H. The performance data presented in the submission support a substantial equivalence decision. VITEK® 2 AST-Streptococcus Tetracycline (≤0.25 ->16 µg/mL) is substantially equivalent to VITEK® 2 AST Streptococcus Tetracycline (K111893). # G. References: - 1. MacLowry, J.D. and Marsh, H.H., Semi-automatic Microtechnique for Serial Dilution Antibiotic Sensitivity Testing in the Clinical laboratory, Journal of Laboratory Clinical Medicine, 72:685-687, 1968. {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for bioMérieux. The logo is a blue circle on top and a yellow circle on the bottom. The word "BIOMÉRIEUX" is written in white letters in the center of the blue circle. - Gerlach, E.H., Microdilution 1: A Comparative Study, p. 63-76. Current Techniques for 2. Antibiotic Susceptibility Testing. A. Balows (ed.), Charles C. Thomas, Springfield, IL, 1974. - Barry, A.L., The Antimicrobic Susceptibility Test, Principles and Practices, Lea and Febiger, 3. Philadelphia, PA, 1976.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...