VITEK® 2 Gram Negative Trimethoprim/sulfamethoxazole is designed for antimicrobial susceptibility testing of gram-negative organisms and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 Gram Negative Trimethoprim/sulfamethoxazole is a quantitative test. Trimethoprim/sulfamethoxazole has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections Escherichia coli (including susceptible enterotoxigenic strains implicated in traveler's diarrhea) Klebsiella species Enterobacter species Morganella morganii Proteus vulgaris Proteus mirabilis Shigella flexneri Shigella sonnei The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 and VITEK® 2 Compact Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, and S. pneumoniae.
Device Story
VITEK®2 AST card system performs automated antimicrobial susceptibility testing. Input: standardized organism suspension from pure culture. Process: card inoculated via vacuum filling; incubated at 35.5°C; optically monitored every 15 minutes. Algorithm: qualitative growth-based detection using predetermined growth thresholds. Output: susceptibility report. Used in clinical laboratories; operated by laboratory personnel. Benefits: automated, standardized determination of bacterial susceptibility to trimethoprim/sulfamethoxazole, facilitating clinical antibiotic selection.
Clinical Evidence
Bench testing only. External evaluation compared VITEK 2 AST-GN SXT performance against CLSI broth microdilution reference method using fresh/stock clinical isolates and stock challenge strains. Demonstrated 100% overall Category Agreement. Reproducibility and Quality Control results were acceptable.
Technological Characteristics
Miniaturized card format; automated incubation/reading; optical growth monitoring; quantitative microdilution methodology; software-based analysis of growth kinetics; compatible with VITEK 2 and VITEK 2 Compact platforms.
Indications for Use
Indicated for susceptibility testing of Escherichia coli, Klebsiella species, Enterobacter species, Morganella morganii, Proteus mirabilis, Proteus vulgaris, Shigella flexneri, and Shigella sonnei to trimethoprim/sulfamethoxazole in clinical laboratories.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
VITEK® 2 Gram Negative Ertapenem (k041982)
Related Devices
K091899 — VITEK 2 GRAM NEGATIVE MEROPENEM · bioMerieux, Inc. · Aug 14, 2009
K172944 — VITEK 2 AST-GN Ceftazidime/Avibactam <=0.12 >=16 µg/mL · bioMerieux, Inc. · Dec 19, 2017
K161217 — VITEK 2 Gram Negative Ceftriaxone (<=0.25->=64 ug/mL) · bioMerieux, Inc. · Jan 12, 2017
K050976 — VITEK 2 GRAM POSITIVE TRIMETHOPRIM/SULFAMETHOXAZOLE · bioMerieux, Inc. · May 23, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
K073406
Image /page/0/Picture/1 description: The image shows the logo for bioMérieux. The logo features a stylized image of a cell or sphere with radiating lines, bisected by a curved line. Below the image is the company name, "BIOMÉRIEUX," in a sans-serif font.
### 510(k) SUMMARY
JAN 2 3 2008
## VITEK® 2 Gram Negative Trimethoprim/sulfamethoxazole
#### 510(k) Submission Information:
| | Submitter's Name: | bioMérieux, Inc. |
|----|----------------------|----------------------------------------------------------------------------------------------------|
| | Address: | 595 Anglum Road<br>Hazelwood, MO 63042 |
| | Contact Person: | Nancy Weaver<br>Associate Director, Regulatory Affairs |
| | Phone Number: | 314-731-8695 |
| | Fax Number: | 314-731-8689 |
| | Date of Preparation: | November 29, 2007 |
| B. | Device Name: | |
| | Formal/Trade Name: | VITEK® 2 Gram Negative<br>Trimethoprim/sulfamethoxazole<br>( $≤$ 1/19 - $≥$ 16/304 µg/ml) |
| | Classification Name: | Fully Automated Short-Term Incubation Cycle<br>Antimicrobial Susceptibility Device, 21 CFR 866.164 |
| | Common Name: | VITEK 2 AST-GN SXT |
| C. | Predicate Device: | |
| | | VITEK 2 Gram Negative Ertapenem (K041982). |
#### D. 510(k) Summary:
B
C
VITEK® 2 Gram Negative Trimethoprim/sulfamethoxazole is designed for antimicrobial susceptibility testing of Escherichia coli, Klebsiella sp., Enterobacter sp., Morganella morganii, Proteus mirabilis, Proteus vulgaris, Shigella flexneri and Shigella sonnei. VITEK 2 Gram Negative Trimethoprim/Sulfamethoxazole is a quantitative test. It is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. The antimicrobial presented in VITEK 2 AST Cards is in concentrations equivalent by efficacy to standard method concentrations in mog/ml. The VITEK 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology.
The bacterial isolate to be tested is diluted to a standardized concentration in 0.45% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK 2 Compact has a manual filling and sealing operation. The VITEK 2 monitors the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.
#### bioMérieux, Inc.
595 Anglum Road, Hazelwood, Missouri 63042-2320, USA 800/638-4835 Fax: 314/731-Phone: 314/731-8500 8700
{1}------------------------------------------------
VITEK 2 Gram Negative Trimethoprim/sulfamethoxazole demonstrated substantially equivalent performance when compared with the CLSI broth microdilution reference method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA, Issued Feb. 5, 2003. *
The Premarket Notification (510[k]) presents data in support of VITEK 2 Gram Negative Trimethoprim/sulfamethoxazole. An external evaluation was conducted with fresh and stock clinical isolates and stock challenge strains. The external evaluations were designed to confirm the acceptability of VITEK 2 Gram Negative Trimethoprim/sulfamethoxazole by comparing its performance with the CLSI broth microdilution reference method. The data is representative of performance on both the VITEK 2 and VITEK 2 Compact instrument platforms, as evidenced in the AST equivalency study presented in the VITEK 2 Compact 510(k), K050002. VITEK 2 Gram Negative Trimethoprim/sulfamethoxazole demonstrated acceptable performance of 100% overall Category Agreement. Reproducibility and Quality Control demonstrated acceptable results.
*Note: This device clinical trial was initiated prior to the March 5, 2007 issuance of the revised guidance, which is why the earlier guidance document is cited.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES U.S.A." arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a representation of human figures. The symbol is composed of three curved lines that converge at the bottom, creating a sense of movement or flow.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JAN 23 2008
Ms. Nancy Weaver Associate Director, Regulatory Affairs bioMérieux, Inc. 595 Anglum Road Hazelwood, MO 63042
k073406 Re: Trade/Device Name: VITEK 2® Gram Negative Trimethoprim/sulfamethoxazole (≤ 1/19 – 16/304 µg/ml) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Device Regulatory Class: Class II Product Code: LON Dated: November 29, 2007 Received: December 4, 2007
Dear Ms. Weaver:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sally attaym
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indication for Use
510(k) Number (if known): k07 3406
Device Name: VITEK® 2 Gram Negative Trimethoprim/sulfamethoxazole (< 1/19 - > 16/304 ug/ml)
Indication For Use:
VITEK® 2 Gram Negative Trimethoprim/sulfamethoxazole is designed for antimicrobial susceptibility testing of gram-negative organisms and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 Gram Negative Trimethoprim/sulfamethoxazole is a quantitative test. Trimethoprim/sulfamethoxazole has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.
Active in vitro and in clinical infections
Escherichia coli (including susceptible enterotoxigenic strains implicated in traveler's diarrhea)
Klebsiella species Enterobacter species Morganella morganii Proteus vulgaris
Proteus mirabilis Shigella flexneri Shigella sonnei
The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 and VITEK® 2 Compact Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, and S. pneumoniae.
Prescription Use _ X (21 CFR Part 801 Subpart D) Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
And/Or
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)
.
Judith McRobb
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K073406
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.