VITEK 2 GRAM NEGATIVE LEVOFLOXACIN

K072038 · bioMerieux, Inc. · LON · Sep 12, 2007 · Microbiology

Device Facts

Record IDK072038
Device NameVITEK 2 GRAM NEGATIVE LEVOFLOXACIN
ApplicantbioMerieux, Inc.
Product CodeLON · Microbiology
Decision DateSep 12, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2
AttributesAI/ML

Intended Use

VITEK® 2 Gram Negative Levofloxacin is designed for antimicrobial susceptibility testing of Enterobacter cloacae, Escherichia coli, Klebsiella pneumoniae, Proteus mirabilis, Pseudomonas aeruginosa, Serratia marcescens, Acinetobacter baumannii, Acinetobacter Iwoffi, Citrobacter koseri, Citrobacter freundii, Enterobacter aerogenes, Enterobacter sakazakii, Klebsiella oxytoca, Morganella morganii, Pantoea agglomerans, Proteus vulgaris, Providencia rettgeri, Providencia stuartii, Pseudomonas fluorescens. VITEK 2 Gram Negative Levofloxacin is a quantitative test. It is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 and VITEK® 2 Compact Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gramnegative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, and S. pneumoniae.

Device Story

VITEK 2 Gram Negative Levofloxacin is an in vitro diagnostic test for determining antimicrobial susceptibility of gram-negative bacilli. Input: pure culture organism suspension in saline, standardized to McFarland 0.5 via DensiChek turbidity meter. Process: automated vacuum-filling of 64-microwell test cards containing antibiotic and culture medium; incubation at 35.5°C; periodic optical scanning (transmittance) every 15 minutes to measure growth inhibition. Output: quantitative Minimum Inhibitory Concentration (MIC) and category interpretation (S-I-R). Used in clinical laboratories; operated by laboratory personnel. Output assists clinicians in selecting appropriate antibiotic therapy for patients with bacterial infections.

Clinical Evidence

Performance evaluated via clinical and challenge studies (n=690 total isolates). Compared against CLSI broth microdilution reference method. Combined results: 98.6% Essential Agreement (EA) and 95.8% Category Agreement (CA). Observed 1 very major discrepancy and 1 major discrepancy. Reproducibility >95%. QC testing performed with E. coli ATCC 25922 and P. aeruginosa ATCC 27853.

Technological Characteristics

Miniaturized broth microdilution card; automated incubation and optical growth monitoring. Connectivity via VITEK 2 or VITEK 2 Compact instrument platforms. Quantitative MIC determination.

Indications for Use

Indicated for antimicrobial susceptibility testing of specific aerobic gram-negative bacilli (e.g., E. coli, P. aeruginosa, Klebsiella spp.) from isolated colonies. Used as a laboratory aid to determine in vitro susceptibility to antimicrobial agents.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K072038 Image /page/0/Picture/1 description: The image shows the logo for bioMérieux. The logo consists of the company name in a stylized font, with a graphic above it. The graphic is a black circle with lines emanating from it, and a curved line extending above and below the circle. ## 510(k) SUMMARY : 1 : 1 : 10 : ## VITEK® 2 Gram Negative Levofloxacin #### 510(k) Submission Information: | Submitter's Name: | bioMérieux, Inc. | |--------------------------------------------|-----------------------------------------------------------------------------------------------------| | Address: | 595 Anglum Road<br>Hazelwood, MO 63042 | | Contact Person: | Jolyn Tenilado<br>Senior Regulatory Affairs Specialist | | Phone Number: | 314-731-8386 | | Fax Number: | 314-731-8689 | | Date of Preparation: | July 20, 2007 | | B. Device Name: | | | Formal/Trade Name: | VITEK® 2 Gram Negative Levofloxacin (≤ 0.12 - ≥ 8<br>µg/ml) | | Classification Name: | Fully Automated Short-Term Incubation Cycle<br>Antimicrobial Susceptibility Device, 21 CFR 866.1645 | | Common Name: | VITEK 2 AST-GN Levofloxacin | | C. Predicate Device:<br>D. 510(k) Summary: | VITEK 2 Gram Negative Ertapenem (K041982). | VITEK® 2 Gram Negative Levofloxacin is designed for antimicrobial susceptibility testing of Enterpbacter cloacae, Escherichia coli, Klebsiella pneumoniae, Proteus mirabilis, Pseudomonas aeruginosa. Serratia marcescens. Acinetobacter baumannii, Acinetobacter twoffi, Citrobacter koseri, Čitrobacter freundii, Enterobacter aerogenes, Enterobacter sakazakii, Klebsiella oxytoca, Morqanella morqanii, Pantoea agglomerans, Proteus vulgaris, Providencia rettgeri, Providencia stuartii. Pseudomonas fluorescens. It is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. The antimicrobial presented in VITEK 2 AST Cards is in concentrations equivalent by efficacy to standard method concentrations in mog/m!. The VITEK 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology. The bacterial isolate to be tested is diluted to a standardized concentration in 0.45% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK 2 Compact has a manual filling and sealing operation. The VITEK 2 monitors the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MC value along with the interpretive category result for each antibiotic contained on the card. ### bioMérieux, Inc. {1}------------------------------------------------ VITEK 2 Gram Negative Levofloxacin demonstrated substantially equivalent performance when compared with the CLSI broth microdilution reference method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA, Issued Feb. 5, 2003 .* The Premarket Notification (510[k]) presents data in support of VITEK 2 Gram Negative Levofloxacin. An external evaluation was conducted with fresh and stock clinical isolates and stock challenge strains. The external evaluations were designed to confirm the acceptability of VITEK 2 Gram Negative Levofloxacin by comparing its performance with the CLSI broth microdilution reference method. The data is representative of performance on both the VITEK 2 and VITEK 2 Compact instrument platforms, as evidenced in the AST equivalency study presented in the VITEK 2 Compact 510(k), K050002. VITEK 2 Gram Negative Levofloxacin demonstrated acceptable performance of 95,8% overall Category Agreement. Reproducibility and Quality Control demonstrated acceptable results. *Note: This device clinical trial was initiated prior to the March 5, 2007 issuance of the revised guidance, which is why the older guidance document is cited. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal for the Department of Health & Human Services USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the top half of the circle. In the center of the seal is a stylized image of an eagle with its wings spread. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 SEP 1 2 2007 Ms. Joyln Tenllado Senior Regulatory Affairs Specialist bioMérieux, Inc. 595 Anglum Road Hazelwood, MO 63042 k072038 Re: Trade/Device Name: VITEK® 2 Gram Negative Levofloxacin (≤0.12 - ≥8μg/ml) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: July 20, 2007 Received: July 25, 2007 Dear Ms. Tenllando: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been or clossified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that clannt require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publiah further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act 's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as et freth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Sally attaym Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use K072038 510(k) Number (if known):_ Device Name: VITEK® 2 Gram Negative Levofloxacin (≤ 0.12 – ≥ 8 µg/ml) Indications For Use: VITEK® 2 Gram Negative Levofloxacin is designed for antimicrobial susceptibility testing of Enterobacter cloacae, Escherichia coli, Klebsiella pneumoniae, Proteus mirabilis, Pseudomonas aeruginosa, Serratia marcescens, Acinetobacter baumannii, Acinetobacter Iwoffi, Citrobacter koseri, Citrobacter freundii, Enterobacter aerogenes, Enterobacter sakazakii, Klebsiella oxytoca, Morganella morganii, Pantoea agglomerans, Proteus vulgaris, Providencia rettgeri, Providencia stuartii, Pseudomonas fluorescens. VITEK 2 Gram Negative Levofloxacin is a quantitative test. It is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 and VITEK® 2 Compact Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gramnegative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, and S. pneumoniae. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Rourrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) vision Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety Page 1 of 1 510(k)_14072038 p. 10
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...