VITEK® 2 AST-Gram Negative Cefepime is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 AST-Gram Negative Cefepime is a quantitative test. Cefepime has been shown to be active against most strains of the microorganisms listed below. according to the FDA label for this antimicrobial. Active in vitro and in clinical infections: Enterobacter spp. Escherichia coli Klebsiella pneumoniae Proteus mirabilis Pseudomonas aeruginosa In vitro data available but clinical significance is unknown: Citrobacter koseri (formerly Citrobacter diversus) Citrobacter freundii Pantoea agglomerans (formerly Enterobacter agglomerans) Klebsiella oxytoca Proteus vulgaris Providencia rettgeri Providencia stuartii Serratia marcescens The VITEK @ 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.
Device Story
VITEK® 2 AST-GN Cefepime card; miniaturized version of broth microdilution technique. Input: isolated bacterial colonies diluted in 0.45-0.50% saline. Operation: automated filling/sealing (VITEK 2) or manual (VITEK 2 Compact); card incubation/reading for up to 18 hours. System monitors growth in card wells; calculates Minimum Inhibitory Concentration (MIC) and interpretive category. Output: automated report of MIC and susceptibility category. Used in clinical laboratories; operated by laboratory technicians. Results assist clinicians in selecting appropriate antimicrobial therapy for patients with bacterial infections.
Clinical Evidence
Performance evaluated using 824 clinical isolates and 100 challenge isolates. Combined performance for Enterobacteriaceae and P. aeruginosa showed 93.8% Essential Agreement and 96.1% Category Agreement. Adjusted major discrepancy rate 0.5%; adjusted very major discrepancy rate 1.6%. Reproducibility 99.6-100%. Growth failure rate 0.2%.
Technological Characteristics
Miniaturized broth microdilution card; contains antimicrobial concentrations equivalent to standard mcg/mL concentrations. Automated incubation and optical growth monitoring. Compatible with VITEK 2 and VITEK 2 Compact systems. Software-based automated analysis of growth kinetics to determine MIC.
Indications for Use
Indicated for antimicrobial susceptibility testing of isolated colonies of clinically significant aerobic Gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp., and yeast. Specifically for Cefepime, targets include Enterobacter spp., E. coli, K. pneumoniae, P. mirabilis, P. aeruginosa, and various other Gram-negative species.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM ASSAY ONLY TEMPLATE
A. 510(k) Number:
K161227
B. Purpose for Submission:
To obtain a substantial equivalence determination for Cefepime for testing of gram negative bacilli on the VITEK® 2 and VITEK® 2 Compact Antimicrobial Susceptibility Test (AST) Systems.
C. Measurand:
The VITEK® 2 AST-Gram Negative card contains the following concentrations of cefepime: 0.25, 1, 4, 16 and 32 µg/mL (equivalent standard method concentration by efficacy in µg/mL). The MIC result reporting range for the card is ≤0.12 - ≥32 µg/mL.
D. Type of Test:
Automated quantitative or qualitative antimicrobial susceptibility test for Cefepime
E. Applicant:
bioMérieux, Inc.
F. Proprietary and Established Names:
VITEK® 2 AST- GN Cefepime (≤0.12 - ≥32 µg/mL)
VITEK 2® AST- Gram Negative Cefepime (≤0.12 - ≥32 µg/mL)
G. Regulatory Information:
1. Regulation section:
21 CFR 866.1645 Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System
2. Classification:
Class II
{2}
2
3. Product code:
LON - Fully automated short-term incubation cycle antimicrobial susceptibility system
LTW - Susceptibility Test Cards, Antimicrobial
LTT - Panels, Test, Susceptibility, Antimicrobial
4. Panel:
83 Microbiology
H. Intended Use:
1. Intended use(s):
The VITEK 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.
2. Indication(s) for use:
VITEK 2 Gram Negative Cefepime is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 Gram Negative Cefepime is a quantitative test. Cefepime has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for the antimicrobial.
Active in vitro and in clinical infections:
Enterobacter spp.
Escherichia coli
Klebsiella pneumoniae
Proteus mirabilis
Pseudomonas aeruginosa
In vitro data available but clinical significance is unknown:
Citrobacter koseri (formerly Citrobacter diversus)
Citrobacter freundii
Pantoea agglomerans (formerly Enterobacter agglomerans)
Klebsiella oxytoca
Proteus vulgaris
Providencia rettgeri
Providencia stuartii
Serratia marcescens
{3}
3
3. Special conditions for use statement(s):
Prescription use only
The following limitations are included in the device labeling:
Although within Essential Agreement, the lack of an intermediate category has shown major and very major discrepancies when compared to the reference method. Testing should be repeated using an alternative testing/reference method prior to reporting results for the following antibiotic/organism:
- Cefepime: Pseudomonas aeruginosa when the VITEK 2 MIC is 8μg/mL or 16μg/mL
Perform an alternative method of testing prior to reporting of results for the following antibiotic/organism combination(s):
- Cefepime: Hafnia alvei, Morganella spp.
Users outside the U.S. should perform an alternative method of testing prior to reporting of results for the following antibiotic/organism combination(s):
- Cefepime: Bordetella bronchiseptica
4. Special instrument requirements:
VITEK 2 Systems 7.01 (PC version) software for VITEK 2 and VITEK 2 Compact Systems
I. Device Description:
The VITEK 2 AST card is a miniaturized, abbreviated and automated version of the doubling dilution technique for determining the minimum inhibitory concentration (MIC). Each VITEK 2 test card contains 64 microwells. A control well containing only culture medium is included on all cards, with the remaining wells containing premeasured amounts of a specific antimicrobial agent in a culture medium base. A suspension of organism from a pure culture is prepared in a tube containing 0.45-0.5% sterile saline and standardized to a McFarland 0.5 using the DensiCHEK Plus. The VITEK 2 System automatically fills, seals and places the card into the incubator/reader; manual methods can also be used for the inoculation of test cards for use in the VITEK 2 System. The VITEK 2 Compact has a manual filling and sealing operation. The VITEK 2 Systems monitor the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card.
VITEK 2 AST-GN Cefepime has the following concentrations in the card: 0.25, 1, 4, 16 and 32 μg/mL (equivalent standard method concentration by efficacy in μg/mL). The MIC result reporting range for the card is ≤0.12 - ≥32 μg/mL.
{4}
# J. Substantial Equivalence Information:
1. Predicate device name(s):
VITEK 2 AST- GN Doxycycline
2. Predicate $510(\mathrm{k})$ number(s):
K121546
3. Comparison with predicate:
Table 1: Comparison with the Predicate Device
| Similarities | | |
| --- | --- | --- |
| Item | Device VITEK 2 AST- GN Cefepime | Predicate VITEK 2 AST- GN Doxycycline (K121546) |
| Intended Use | The VITEK 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast. | Same |
| Test Method | Automated quantitative antimicrobial susceptibility test for use with the VITEK 2 and VITEK 2 Compact Systems to determine the in vitro susceptibility of Gram negative bacilli | Same |
| Inoculum | Saline suspension of organisms | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Antimicrobial | Cefepime | Doxycycline |
| Antimicrobial Concentration | 0.25, 1, 4, 16, 32 | 1, 4, 16 |
| Reading Algorithm | Growth pattern analysis-Unique to Cefepime | Discriminate analysis-Unique to Doxycycline |
# K. Standard/Guidance Document Referenced (if applicable):
- CLSI M100-S25: Performance Standards for Antimicrobial Susceptibility Testing
- CLSI M07-A10: Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria
{5}
That Grow Aerobically
- Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems
L. Test Principle:
The VITEK 2 and VITEK 2 Compact Systems utilize automated growth-based detection using attenuation of light measured by an optical scanner. The optics used in the systems use visible light to directly measure organism growth. Transmittance optics are based on an initial light reading of a well before significant growth has begun. Periodic light transmittance samplings of the same well measure organism growth by how much light is prevented from going through the well. The VITEK 2 System monitors the growth of each well in the card over a defined period of time. An interpretive call is made between 4 and 16 hours for a “rapid” read but may be extended to 18 hours in some instances. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic on the card.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
A reproducibility study was conducted at three sites using ten isolates of gram negative bacilli that were consistent with the intended use. Isolates were tested in triplicate over three days for a total of 270 data points. The isolates tested in the reproducibility study included Klebsilla pneumoniae (three isolates), Pseudomonas aeruginosa (two isolates), Enterobacter cloacae (two isolates), Serratia marcescens (two isolates), and E. coli (one isolate). Inocula were prepared using both manual and automatic dilution methods for testing in the VITEK 2. Inocula were prepared by the manual testing method only in the VITEK 2 Compact. The mode MIC value was determined and the reproducibility was calculated based on MIC values falling within ± 1 dilution of the mode MIC value.
Using VITEK 2 and automatic dilution, all results were on scale and the reproducibility was 100%.
Using VITEK 2 and manual dilution, there was one off-scale result. Reproducibility was at 99.6% for best an worst case scenarios.
Using VITEK 2 Compact and manual dilution, all results were on-scale and the reproducibility was 100%.
The reproducibility results were acceptable.
5
{6}
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Inoculum Density Check: The inoculum density was monitored using the DensiCHEK Plus™ instrument. The DensiCHEK Plus™ was standardized daily with all results recorded and within expected range.
Purity Check: A purity check of all organisms was performed at the time of VITEK2 card inoculation. Only results obtained with pure cultures were evaluated.
Growth Failure Rate: There were two isolates that failed to grow in the clinical study. Complete test results were available for 822 isolates in a total of 824 clinical isolates. The growth failure rate was 0.2% and was acceptable.
Quality Control Testing: The FDA recommended QC organisms (E. coli ATCC 25922 and Pseudomonas aeruginosa ATCC 27853) were tested using both the VITEK 2 card and the reference method at each site. Both the automatic dilution and manual dilution methods were used for the VITEK 2 and the manual dilution method was used for the VITEK 2 Compact.
The expected range for E. coli ATCC 25922 with cefepime is 0.015 – 0.12 µg/mL. Even though the cefepime concentrations included in the VITEK 2 AST-Gram Negative card are 0.25, 1, 4, 16 and 32 µg/mL, the reporting range is ≤0.12 – ≥32 µg/mL. Therefore, all results for the QC strain E. coli ATCC 25922 were off scale for the VITEK 2 and VITEK 2 Compact Systems as both VITEK systems report the lowest end of the scale as ≤0.12 µg/mL (Table 2). However, Pseudomonas aeruginosa ATCC 27853 was tested where the expected range is 0.5–4 µg/mL and all results were on-scale. All results were within the expected range.
Table 2: Quality Control Results for VITEK 2 (Automatic and Manual Dilution Methods) and for VITEK 2 Compact (Manual Dilution Method)
| | | VITEK 2 Automatic-Dilution | | VITEK 2 Manual Dilution | | VITEK 2 Compact Manual Dilution | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Organism | Conc. (μg/mL) | Ref. | Test | Ref. | Test | Ref. | Test |
| E. coli
ATCC 25922
Expected Range:
0.015- 0.12μg/mL | ≤0.008 | | | | | | |
| | 0.016 | 1 | | 1 | | 1 | |
| | 0.03 | 166 | | 120 | | 117 | |
| | 0.06 | 66 | | 44 | | 43 | |
| | (≤)0.125* | | 233 | | 165 | | 161 |
| | 0.25 | | | | | | |
| | 0.5 | | | | | | |
| | 1 | | | | | | |
{7}
| | | VITEK 2 Automatic-Dilution | | VITEK 2 Manual Dilution | | VITEK 2 Compact Manual Dilution | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Organism | Conc. (μg/mL) | Ref. | Test | Ref. | Test | Ref. | Test |
| | 2 | | | | | | |
| | 4 | | | | | | |
| | 8 | | | | | | |
| | 16 | | | | | | |
| | 32 | | | | | | |
| | ≥64 | | | | | | |
| | | | | | | | |
| P. aeruginosaATCC 27853Expected Range:0.5-4μg/mL | 0.125 | | | | | | |
| | 0.25 | | | | | | |
| | 0.5 | 15 | 1 | 6 | 1 | 6 | 1 |
| | 1 | 178 | | 130 | 1 | 129 | |
| | 2 | 40 | 231 | 28 | 163 | 26 | 161 |
| | 4 | 1 | 2 | 1 | | 1 | |
| | 8 | | | | | | |
| | 16 | | | | | | |
| | 32 | | | | | | |
| | ≥64 | | | | | | |
* The lowest end point of the VITEK 2 Cefepime MIC range is ≤0.12 μg/mL. Obtaining this value was considered as an indicator that the quality control test results were acceptable.
Inoculum Density Check. The inoculum density was monitored using the DensiCHEK Plus instrument. The DensiCHEK Plus was standardized weekly with all results recorded and in expected range.
d. Detection limit:
Not Applicable
e. Analytical specificity:
Not Applicable
f. Assay cut-off:
Not Applicable
2. Comparison studies:
a. Method comparison with predicate device:
Results obtained with the bioMérieux VITEK 2 AST - Gram Negative card with cefepime were compared to results obtained with the CLSI broth microdilution reference panel. The VITEK 2 AST-Gram Negative card with cefepime contains the
{8}
following concentrations of cefepime: 0.25, 1, 4, 16, and $32\mu \mathrm{g / mL}$ (equivalent standard method concentration by efficacy in $\mu \mathrm{g / mL}$ ) and the reporting range is $\leq 0.12 - \geq 32\mu \mathrm{g / mL}$ (i.e. $\leq 0.125$ , 0.25, 0.5, 1, 2, 4, 8, 16, and $\geq 32$ ). The reference panel contained two-fold serial dilutions with a range of 0.008 to $128\mu \mathrm{g / mL}$ . The testing conditions for the reference method were:
Medium: Mueller Hinton broth (Cation-Adjusted) with the appropriate dilutions of antimicrobial solution added
Inoculum: Direct colony suspension
- Incubation: $35^{\circ}\mathrm{C}$ ; 16-20 hours
Test inocula were standardized using the DensiCHEK Plus instrument. VITEK 2 AST - Gram Negative cards were inoculated using automatic dilution (for reading on the VITEK 2 instrument) or using a manual dilution method (for reading on the VITEK 2 instrument or on the VITEK 2 COMPACT instrument). Reference panels were inoculated as outlined in the CLSI document M07-A10.
A total of 824 clinical isolates were evaluated at three external sites with VITEK 2 AST - Gram Negative cards inoculated by automatic dilution and interpreted using the VITEK 2 instrument. There were two isolates that failed to grow (99.8% growth rate). The majority of isolates were fresh (641 isolates, 78.0%); 181 isolates (22.0%) were stock isolates. There were 531 Enterobacteriaceae and 291 Pseudomonas aeruginosa isolates tested in the clinical studies.
A total of 100 challenge isolates (80 Enterobacteriaceae and 20 Pseudomonas aeruginosa) were evaluated at one external site. The challenge set was tested with both automatic dilution and manual dilution methods on the VITEK 2 System and with the manual dilution method on the VITEK2 Compact System. The performance is provided in Table 3-A below:
Table 3-A: Performance† of Clincal and Challenge Isolates, VITEK 2 Auto-Dilution
| | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | #R | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Enterobacteriaceae ≤2 (S), 4-8 (I), ≥16 (R) | | | | | | | | | | | | |
| Clinical | 531 | 513 | 96.6 | 41 | 31 | 75.6 | 519 | 97.7 | 81 | 10 | 1 | 1 |
| Challenge | 80 | 77 | 96.3 | 12 | 9 | 75.0 | 80 | 100 | 0 | 0 | 0 | 0 |
| Combined | 611 | 590 | 96.6 | 53 | 40 | 75.5 | 599 | 98.0 | 81 | 10 | 1 | 1 |
| Pseudomonas aeruginosa ≤8 (S), --, ≥16 (R) | | | | | | | | | | | | |
| Clinical | 291 | 256 | 88.0 | 273 | 239 | 87.5 | 268 | 92.1 | 41 | 0 | 20 | 3 |
| Challenge | 20 | 19 | 95.0 | 20 | 19 | 95.0 | 19 | 95.0 | 2 | 0 | 0 | 1 |
| Combined | 311 | 275 | 88.4 | 293 | 258 | 88.1 | 287 | 92.3 | 43 | 0 | 20 | 4 |
| Enterobacteriaceae + P. aeruginosa | 922 | 865 | 93.8 | 346 | 298 | 86.1 | 886 | 96.1 | 124 | 10 | 21 (2.7%) | 5 (4.0%) |
$\dagger$ EA - Essential Agreement (+/- 1 dilutions)
CA - Category Agreement
EVAL - Evaluable isolates
R or NS - Resistant or non-susceptible isolates
Essential Agreement (EA) occurs when there is agreement between the result of the reference method and min - minor discrepancies
maj - major discrepancies
vmj - very major discrepancies
{9}
that of VITEK 2 test card within plus or minus one serial two-fold dilution of the antibiotic. Evaluable results are those that are on scale for both the VITEK 2 test card and the reference method. Category Agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation of the VITEK 2 test card.
## Enterobacteriaceae, ≤2 (S), 4-8 (I), ≥16 (R)
The performance of Enterobacteriaceae was acceptable at 96.6% EA, 98.0% CA, major discrepancy rate of 0.2% (1/523), and very major discrepancy rate of 1.2% (1/81).
## Pseudomonas aeruginosa, ≤8 (S), --, ≥16 (R)
### i. Category Agreement
Major and very major discrepancies were observed when P. aeruginosa results were analyzed separately. Of the 20 major discrepancies, 17 were within essential agreement [i.e. VITEK 2: 16 (R), Reference: 8 (S)]. Of the four very major discrepancies, three were within essential agreement [i.e. VITEK 2: 8 (S), Reference: 16 (R)]. These 17 major and 3 very major discrepancies would have been considered minor discrepancies if there was an "intermediate" interpretation category for cefepime/P. aeruginosa.
Because of the lack of "intermediate" interpretation category, an adjustment was made in the overall performance evaluation of the discrepancy (error) rates by conducting an analysis to determine whether the MIC values for these isolates were within essential agreement or not.
As a result of this analysis, the adjusted major discrepancy rate for P. aeruginosa became 1.1% (3/268). The combined major discrepancy rate for Enterobacteriaceae and P. aeruginosa became 0.5% (4/791). Similarly, the adjusted very major discrepancy rate for P. aeruginosa became 2.3% (1/43). The combined very major discrepancy rate for Enterobacteriaceae and P. aeruginosa became 1.6% (2/124). These performance data are acceptable based on the criteria outlined in the AST Special Controls guidance document.
The adjusted performance is noted in Tables 3-B and 3-C below:
| | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | #R | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Clinical | 291 | 256 | 88.0 | 273 | 239 | 87.5 | 268 | 92.1 | 41 | 0 | 3
1.2%
(3/250) | 1
2.4%
(1/41) |
| Challenge | 20 | 19 | 95.0 | 20 | 19 | 95.0 | 19 | 20 | 2 | 0 | 0 | 0 |
| Combined | 311 | 275 | 88.4 | 293 | 258 | 88.1 | 287 | 92.3 | 43 | 0 | 3
1.1%
(3/268) | 1
2.3%
(1/43) |
{10}
Table 3-C: Combined Performance (Original Enterobacteriaceae plus Adjusted P. aeruginosa)
| | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | #R | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Enterobacteriaceae + P. aeruginosa | 922 | 865 | 93.8 | 346 | 298 | 86.1 | 886 | 96.1 | 124 | 10 | 4 0.5% (4/791) | 2 1.6%† (2/124) |
† The acceptable number of discrepancies is 3 when 124 of resistant organisms are tested (Table 8, AST Special Controls Guidance)
The original $P$ aeruginosa performance with 20 major and 4 very discrepancies was mitigated by adding a limitation to address the need for repeat testing $P$ aeruginosa by alternative methods when the VITEK 2 Systems MIC is $8\mu \mathrm{g} / \mathrm{mL}$ or $16\mu \mathrm{g} / \mathrm{mL}$ . The limitation is noted below:
Although within Essential Agreement, the lack of an intermediate category has shown major and very major discrepancies when compared to the reference method. Testing should be repeated using an alternative testing/reference method prior to reporting results for the following antibiotic/organism:
- Cefepime: Pseudomonas aeruginosa when the VITEK 2 MIC is $8\mu \mathrm{g} / \mathrm{mL}$ or $16\mu \mathrm{g} / \mathrm{mL}$
For implementation of the repeat testing limitation, bioMerieux indicated that they will issue a customer communication requesting VITEK 2 uers to create a BioART rule stating the limitation in the VITEK 2 System for Pseudomonas aeruginosa when testing cefepime. Implementation of this BioART rule will assure that alternative testing is performed in accordance with the above limitation. The limitation will be included in the next software update release.
# ii. Essential Agreement
The essential agreement for $P$ aeruginosa in the comparative studies was lower than the acceptance criterion of $90\%$ when analyzed separately. It was further evaluated by trending as demonstrated in Table 4 below:
Table 4: Trending of On-Scale Clinical and Challenge Isolate Results- P. aeruginosa
| | # of Isolates | Difference in MIC as Compared to the CLSI Reference Method | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | ≥3 dil. lower | 2 dil. lower | 1 dil. Lower | Exact | 1 dil. higher | 2 dil. higher | ≥3 dil higher |
| Clinical | 283 | 3 | 4 | 19 | 118 | 113 | 24 | 2 |
| Challenge | 20 | 0 | 0 | 2 | 7 | 10 | 1 | 0 |
| Combined | 303 | 3 (1.0%) | 4 (1.3%) | 21 (6.9%) | 125 (41.3%) | 123 (40.6%) | 25 (8.3%) | 2 (0.7%) |
The following footnote pertinent to cefepime is added to the VITEK 2 Performance Characteristics for Gram-Negative Antimicrobial Susceptibility Testing Table (Table 127, Product Information Manual):
{11}
The EA for VITEK2 Cefepime with Pseudomonas aeruginosa was 88.4%. The trending of on-scale results demonstrated: 41.3% equivalent, 40.6% one doubling dilution higher, and 6.9% one doubling dilution lower when compared to the CLSI reference broth microdilution.
## Challenge
The challenge set of 80 Enterobacteriaceae and 20 Pseudomonas aeruginosa isolates were evaluated with both automatic dilution and manual dilution methods on the VITEK 2 System and with the manual dilution method on the VITEK2 Compact System. The performance is shown in Table 5.
Table 5: Performance of VITEK 2 and VITEK 2 Compact
| | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | #R | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | VITEK 2 | | | | | | | | | | | |
| Auto-dilution | 100 | 96 | 96.0 | 32 | 28 | 87.5 | 99 | 99.0 | 2 | 0 | 0 | 1 |
| Manual Dilution | 100 | 95 | 95.0 | 32 | 27 | 84.4 | 97 | 97.0 | 2 | 2 | 0 | 1 |
| | VITEK 2 Compact | | | | | | | | | | | |
| Manual Dilution | 100 | 96 | 96.0 | 32 | 28 | 87.5 | 97 | 97.0 | 2 | 2 | 0 | 1 |
The performance of the VITEK 2 Compact (manual dilution) was considered acceptable based on the reproducibility (100%), QC (Table 2), and challenge (Table 5) studies.
b. Matrix comparison:
Not Applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable
4. Clinical cut-off:
{12}
Not Applicable
5. Expected values/Reference range:
Table 6: Breakpoints for Cefepime (FDA Approved labeling)
| Organism | Cefepime FDA MIC Breakpoints (μg/mL) | | |
| --- | --- | --- | --- |
| | S | I | R |
| Enterobacteriaceae | ≤2 | 4-8 | ≥16 |
| Pseudomonas aeruginosa | ≤8 | -- | ≥16 |
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.