BD Phoenix Automated Microbiology System - GN Eravacycline (0.125-2 µg/mL)

K251713 · Becton, Dickinson and Company · LON · Aug 8, 2025 · Microbiology

Device Facts

Record IDK251713
Device NameBD Phoenix Automated Microbiology System - GN Eravacycline (0.125-2 µg/mL)
ApplicantBecton, Dickinson and Company
Product CodeLON · Microbiology
Decision DateAug 8, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2
AttributesPCCP

Indications for Use

The BD Phoenix Automated Microbiology System is intended for the in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of Gram Negative aerobic and facultative anaerobic bacteria belonging to the order Enterobacterales and non-Enterobacterales.

Device Story

The BD Phoenix Automated Microbiology System is a broth-based microdilution platform for antimicrobial susceptibility testing (AST). It utilizes sealed, self-inoculating polystyrene panels containing dried antimicrobial agents and a redox indicator. Input consists of pure bacterial culture suspensions standardized to 0.5 McFarland using a nephelometer. The system incubates panels at 35°C; it continuously monitors bacterial growth via colorimetric oxidation-reduction (blue to pink) and turbidity measurements every 20 minutes. The instrument records growth patterns to determine MIC values and categorical interpretations (e.g., susceptible). Results are processed by the BDXpert expert system using CLSI/FDA-derived rules. The system is used in clinical microbiology laboratories by trained technicians. Output provides clinicians with MIC data to guide antibiotic therapy. The device benefits patients by enabling rapid, standardized susceptibility testing to inform appropriate antimicrobial selection.

Clinical Evidence

Performance was evaluated using 785 clinical and 84 challenge isolates. Combined manual inoculation results showed 97.8% Essential Agreement (EA) and 97.1% Category Agreement (CA). Adjusted potential very major error rates were calculated for specific species (E. cloacae 15.8%, E. coli 100%, K. aerogenes 20%, K. pneumoniae 8.1%) due to the susceptible-only category. Trending analysis indicated lower MIC values for C. freundii, C. koseri, and E. coli. Labeling includes limitations for non-susceptible strain detection and confirmatory testing requirements.

Technological Characteristics

Broth-based microdilution system using molded polystyrene panels with 136 micro-wells. Employs colorimetric oxidation-reduction indicator (redox) and turbidity for growth detection. Energy source: electrical (automated instrument). Connectivity: networked system. Software: V2.20.0.0 or higher. Inoculation: manual (PhoenixSpec Nephelometer) or automated (BD Phoenix AP). Sterilization: not specified.

Indications for Use

Indicated for in vitro quantitative determination of antimicrobial susceptibility by MIC of Gram-negative aerobic and facultative anaerobic bacteria (Enterobacterales and Non-Enterobacterales) and Gram-positive bacteria (Staphylococcus, Enterococcus, Streptococcus) from pure culture. This notification specifically covers Eravacycline (0.125-2 µg/mL) for Enterobacterales.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY ## I Background Information: A 510(k) Number K251713 B Applicant Becton, Dickinson and Company C Proprietary and Established Names BD Phoenix Automated Microbiology System - GN Eravacycline (0.125-2 µg/mL) D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | LON | Class II | 21 CFR 866.1645 - Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System | MI - Microbiology | ## II Submission/Device Overview: A Purpose for Submission: Addition of eravacycline to the BD Phoenix Gram negative ID/AST only Phoenix panels B Measurand: Eravacycline 0.125-2 µg/mL C Type of Test: Antimicrobial susceptibility test (quantitative) colorimetric, oxidation-reduction, growth based ## III Intended Use/Indications for Use: Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} K251713 - Page 2 of 13 A Intended Use(s): The BD Phoenix Automated Microbiology System is intended for the in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of Gram Negative aerobic and facultative anaerobic bacteria belonging to the order Enterobacterales and non-Enterobacterales. B Indication(s) for Use: The BD Phoenix Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacterales and Non-Enterobacterales and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus, and Streptococcus. This premarket notification is for the BD Phoenix Automated Microbiology System with Eravacycline at a concentration of 0.125-2 µg/mL. Testing is indicated for Enterobacterales as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC). The BD Phoenix Automated Microbiology System - GN Eravacycline (0.125-2 µg/mL) has demonstrated acceptable performance with the following organisms: Enterobacterales (Citrobacter amalonaticus, Citrobacter braakii, Citrobacter farmeri, Citrobacter freundii, Citrobacter koseri, Citrobacter youngae, Enterobacter cloacae, Escherichia coli, Klebsiella aerogenes, Klebsiella oxytoca, and Klebsiella pneumoniae) C Special Conditions for Use Statement(s): Rx - For Prescription Use Only The ability of the BD Phoenix AST system to detect non-susceptibility to eravacycline in the following species is unknown because non-susceptible strains were not available or insufficient at the time of comparative testing: C. freundii, C. koseri, C. amalonaticus, C. braakii, C. farmeri, C. youngae and K. oxytoca. For Enterobacter cloacae, two of the five potential very major errors were within essential agreement when compared to the reference method. Due to the susceptible-only category, the potential very major error rate was adjusted to 15.8% (3/19). If deemed necessary for patient care, confirmatory testing with an alternate method could be performed for this organism before reporting Phoenix MIC results of 0.25 µg/mL. For Klebsiella pneumoniae, eight of the eleven potential very major errors were within essential agreement when compared to the reference method. Due to the susceptible-only category, the potential very major error rate was adjusted to 8.1% (3/37). If deemed necessary for patient care, confirmatory testing with an alternate method could be performed for this organism before reporting Phoenix MIC results of 0.25 µg/mL. For Klebsiella aerogenes, neither of the two potential very major errors were within essential agreement when compared to the reference method. Due to the susceptible-only category, the potential very major error rate was adjusted to 20% (2/10). If deemed necessary for patient care, {2} confirmatory testing with an alternate method could be performed for this organism before reporting Phoenix MIC results of 0.25 µg/mL. For Escherichia coli, none of the three potential very major errors were within essential agreement when compared to the reference method. Due to the susceptible-only category, the potential very major error rate was not adjusted and remained at 100% (3/3). These errors occurred with a very low percentage of the total isolates evaluated (3/376, 0.8%). If deemed necessary for patient care, confirmatory testing with an alternate method could be performed for this organism before reporting Phoenix MIC results of ≤0.125 µg/mL or 0.25 µg/mL. ## D Special Instrument Requirements: - BD Phoenix Automated Microbiology System and software (V2.20.0.0 or higher) - PhoenixSpec Nephelometer - BD Phoenix AP Instrument ## IV Device/System Characteristics: ### A Device Description: This submission is for a single drug in the Gram-negative ID/AST or AST only panel. The ID portion of the ID/AST combination panel was not subject for review in this submission. The Phoenix AST method is a broth-based microdilution test. The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. The ID/AST combination panel includes an ID side (51 wells) with dried substrates for bacterial identification and an AST side (85 wells). The AST panel contains a wide range of two-fold doubling dilution concentrations of antimicrobial agents and growth and fluorescent controls at appropriate well locations. The AST panel does not include wells for isolate identification. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. The organism to be tested must be a pure culture and be preliminarily identified as gram-positive or gram-negative. Colonies are then suspended in ID broth and equated to a 0.5 McFarland suspension using a nephelometer device. A further dilution is made into AST broth (a cation-adjusted formulation of Mueller-Hinton broth containing 0.010% Tween 80), to which the redox-buffered oxidation-reduction AST indicator solution is added producing a blue color in the wells. The concentration of organisms in the final AST broth suspension is approximately 5 × 10⁵ CFU/mL. The Phoenix AST Broth is poured into the inoculation port of the AST panel and the inoculum flows into the panel, filling panel wells. Polyethylene caps are applied to seal the inoculation ports. An air admittance port is located in the panel lid to ensure adequate oxygen tension in the panel for the duration of the test. Inoculated panels are barcode scanned and loaded into the BD Phoenix Automated Microbiology System instrument where panels are continuously incubated at 35°C ± 1°C. Continuous measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. The instrument takes readings every 20 minutes. Organisms growing in the presence of a given antimicrobial agent reduce the indicator (changing K251713 - Page 3 of 13 {3} it to a pink color). This signals organism growth and resistance to that antimicrobial agent. Organisms killed or inhibited by the antimicrobial agent do not cause reduction of the indicator and therefore do not produce a color change. The Phoenix instrument reads and records the results of the antimicrobial tests contained in the panel and interprets the reactions (based on the organism identification) to give a minimal inhibitory concentration (MIC) value and category interpretations (susceptible, intermediate, resistant, or not susceptible). AST results are available within 16 hours. This is an auto read result; no manual readings are possible with this system. Additional comments concerning specific organism/antimicrobial combinations are provided from the software-driven expert system (BDXpert), using rules derived from CLSI documentation and/or the FDA-approved drug labeling. ## B Principle of Operation: The BD Phoenix Automated Microbiology System is a broth-based microdilution method that utilizes a redox indicator (colorimetric oxidation-reduction) to enhance detection of organism growth. The MIC is determined by comparing growth in wells containing serial two-fold dilutions of an antibiotic to the growth in "growth control wells" that contain no antibiotic. ## V Substantial Equivalence Information: ## A Predicate Device Name(s): Bd Phoenix Automated Microbiology System - GN Tigecycline (0.25-16 μg/mL) ## B Predicate 510(k) Number(s): K132909 ## C Comparison with Predicate(s): | Device & Predicate Device(s): | Device K251713 | Predicate K132909 | | --- | --- | --- | | Device Trade Name | BD Phoenix Automated Microbiology System – GN Eravacycline (0.125-2 μg/mL) | BD Phoenix Automated Microbiology System – GN Tigecycline (0.25-16 μg/mL) | | General Device Characteristic Similarities | | | | Intended Use | The BD Phoenix Automated Microbiology System is intended for the in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of Gram Negative aerobic and facultative anaerobic bacteria belonging to the order Enterobacterales and non-Enterobacterales. | Same | K251713 - Page 4 of 13 {4} | Device & Predicate Device(s): | Device K251713 | Predicate K132909 | | --- | --- | --- | | Source of Microorganisms for Testing | Bacterial colonies isolated from culture | Same | | Technology | Automated growth-based detection | Same | | Methodology | Determination of MIC using serial two-fold dilution format | Same | | Read Method | Automated | Same | | Inoculation Methods | Manual: BD PhoenixSpec nephelometer Automated: BD Phoenix AP Instrument | Same | | Incubation Time | <16 hours | Same | | General Device Characteristic Differences | | | | Antimicrobial Agent | Eravacycline | Tigecycline | | Reporting Range | 0.125-2 μg/mL | 0.25-16 μg/mL | | Result Reported | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (S, NS) | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (S, I, R) | | Tested Organisms | Citrobacter amalonaticus Citrobacter braakii Citrobacter farmeri Citrobacter freundii Citrobacter koseri Citrobacter youngae Enterobacter cloacae Escherichia coli Klebsiella aerogenes Klebsiella oxytoca Klebsiella pneumoniae | Enterobacter cloacae Escherichia coli Klebsiella pneumoniae Klebsiella oxytoca Citrobacter freundii Citrobacter koseri Enterobacter aerogenes Serratia marcescens | VI Standards/Guidance Documents Referenced: 1. Guidance for Industry and FDA, Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems, August 28, 2009. 2. CLSI. Performance Standards for Antimicrobial Susceptibility Testing. 35th ed. CLSI supplement M100. Clinical and Laboratory Standards Institute; 2025. 3. CLSI. Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically. 11th ed. CLSI supplement M07. Clinical Laboratory Standards Institute; 2018. K251713 - Page 5 of 13 {5} K251713 - Page 6 of 13 # VII Performance Characteristics (if/when applicable): # A Analytical Performance: # 1. Precision/Reproducibility: Reproducibility of the BD Phoenix Automated Microbiology System - GN Eravacycline $(0.125 - 2\mu \mathrm{g / mL})$ was conducted at three sites using a panel of 12 non-fastidious gram-negative organisms with on-scale results. The isolates were tested at each site in triplicate over three different days using both inoculation methods (manual and BD Phoenix AP) resulting in 324 data points (12 strains x 3 replicates x 3 sites x 3 days = 324) for each inoculation method. The isolates tested in the reproducibility study included Citrobacter freundii (1), Enterobacter cloacae (5), Klebsiella oxytoca (1) and Klebsiella pneumoniae (5). The reproducibility was calculated based on MIC values falling within $\pm 1$ dilution of the mode value for each isolate. As all results were on-scale, there was no calculation of worst case scenario results. The best-case reproducibility was calculated as described in the AST Special Controls Guidance document. The reproducibility results were acceptable as shown in Table 1. Table 1. Reproducibility for Eravacycline (0.125-2 μg/mL) | Inoculation Method | Best Case Reproducibility | | --- | --- | | Manual PhoenixSpec Nephelometer | 100% (324/324) | | Phoenix AP Instrument | 100% (324/324) | # 2. Linearity: Not applicable # 3. Analytical Specificity/Interference: Not applicable # 4. Assay Reportable Range: Not applicable # 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): # Quality Control Testing: The CLSI recommended QC organisms (E. coli ATCC 25922 and P. aeruginosa ATCC 27853) were tested a sufficient number of times (i.e., at least 20/site) at each of three testing sites. The strains were tested using both manual and Phoenix AP inoculation methods and read by the BD Phoenix instrument. Although the majority of QC results were off-scale, the expected range for P. aeruginosa ATCC 27853 was partially on-scale and the expected ranges of the two QC strains included values that span the reporting range. The results are summarized in Table 2. Results were acceptable for greater than $95\%$ of tests performed using both inoculation methods. {6} Table 2. Quality Control Results - Eravacycline | Organism | Concentration (μg/mL) | Reference | BD Phoenix | | | --- | --- | --- | --- | --- | | | | | Manual Inoculation (PhoenixSpec) | Phoenix AP Inoculation | | E. coli ATCC 25922† Expected Range: 0.016-0.125 μg/mL | ≤0.125 | 133 | 89 | 82 | | | 0.25 | | | | | | 0.5 | | | | | | 1 | | | | | | 2 | | | | | | >2 | | | | | P. aeruginosa ATCC 27853†† Expected Range: 2-16 μg/mL | ≤0.125 | 3 | | | | | 0.25 | 1 | | | | | 0.5 | 1 | | | | | 1 | 4 | | | | | 2 | 12 | 33 | 20 | | | >2 | 111 | 55 | 62 | †The lowest dilution of the BD Phoenix Automated Microbiology System – GN Eravacycline MIC range is ≤0.125 μg/mL. Obtaining this value was considered an indicator that the quality control test results were acceptable. ††The highest dilution of the BD Phoenix Automated Microbiology System – GN Eravacycline MIC range is &gt;2 μg/mL. Obtaining this value was considered an indicator that the quality control test results were acceptable. ## Inoculum Density Check: The BD PhoenixSpec Nephelometer was used to prepare the inocula for testing of the clinical, challenge, reproducibility, and QC isolates. The same inoculum suspension was used for both the Phoenix System and the reference method testing. The BD Phoenix AP instrument was used to standardize the inocula for challenge, QC, and reproducibility isolates. Validation data for both the PhoenixSpec and the Phoenix AP instrument was provided and found to be acceptable. ## Growth Failure Rate: The growth failure rate for both inoculation methods was 0%. ## Purity Check: Purity check plates were performed on all isolates from each inoculum preparation. Results were only included for pure cultures. 6. Detection Limit: Not applicable 7. Assay Cut-Off: Not applicable K251713 - Page 7 of 13 {7} B Comparison Studies: 1. Method Comparison with Predicate Device: Results obtained with the BD Phoenix Automated Microbiology System - GN Eravacycline (0.125-2 µg/mL) panel were compared to results obtained with the CLSI frozen broth microdilution reference panel. Reference panels were prepared according to CLSI M07 guidelines. The range of dilutions evaluated with the reference panels was the same as that used for the BD Eravacycline panel. The BD Phoenix Spec Nephelometer, the primary inoculation method, was used to obtain a 0.50 – 0.60 McFarland for all challenge, clinical, QC, and reproducibility isolates. The BD Phoenix AP instrument, the secondary inoculation method, was used to test challenge, QC, and reproducibility isolates. It is designed to standardize the ID broth inoculum equivalent to the BD Phoenix Spec Nephelometer, add the preset amount of AST indicator broth to the AST broth tube, and transfer the required aliquot of ID broth inoculum to AST broth tubes. Clinical: Clinical testing was conducted at two external sites and one internal site using 626 (79.7%) fresh and 159 (20.3%) stock isolates for a total of 785 clinical isolates. These consisted of Citrobacter freundii (23 isolates), Citrobacter koseri (26 isolates), Citrobacter amalonaticus (3 isolates), Citrobacter braakii (4 isolates), Citrobacter farmeri (1 isolate), Citrobacter youngae (1 isolate), Enterobacter cloacae (60 isolates), Escherichia coli (359 isolates), Klebsiella aerogenes (57 isolates), Klebsiella oxytoca (51 isolates), and Klebsiella pneumoniae (200 isolates). Challenge: A total of 84 challenge isolates were evaluated at three U.S. sites using Citrobacter freundii (3 isolates), Citrobacter koseri (2 isolates), Enterobacter cloacae (21 isolates), Escherichia coli (17 isolates), Klebsiella aerogenes (4 isolates), K. oxytoca (3 isolates) and Klebsiella pneumoniae (34 isolates). Four (4) of the challenge isolates (two (2) E. cloacae and two (2) K. pneumoniae isolates) were only tested with the manual inoculation method. A total of 71 non-susceptible strains of Enterobacterales were evaluated. No non-susceptible strains of the following species were evaluated: C. freundii, C. amalonaticus, C. braakii, C. farmeri, and C. youngae. Only one isolate each of C. koseri and K. oxytoca was non-susceptible. As a result, the following limitation was added to the device labeling: The ability of the BD Phoenix AST system to detect non-susceptibility to eravacycline in the following species is unknown because non-susceptible strains were not available or insufficient at the time of comparative testing: C. freundii, C. koseri, C. amalonaticus, C. braakii, C. farmeri, C. youngae and K. oxytoca. In addition, since no interpretive category is defined other than "susceptible" for eravacycline, the following general statement in the device labeling applies: For some organism/antimicrobial combinations, the absence or rare occurrence of resistant strains precludes defining any result categories other than susceptible. For strains yielding results suggestive of a nonsusceptible category, organism K251713 - Page 8 of 13 {8} identification and antimicrobial susceptibility test results should be confirmed. Subsequently, the isolates should be saved and submitted to a reference laboratory that will confirm the result using the CLSI reference dilution method. Results for clinical and challenge isolates were evaluated separately and combined. The performance of testing eravacycline using the manual inoculation method is illustrated in Table 3 below. Table 3. Eravacycline (0.125-2 μg/mL) Results, Manual Inoculation | | Tot | EA No. | EA % | Eval Tot | Eval EA No. | Eval EA % | CA Tot | CA % | No. NS | No. S | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Enterobacterales [≤ 0.5 (S)] | | | | | | | | | | | | | | | Clinical | 785 | 769 | 98.0 | 348 | 332 | 95.4 | 768 | 97.8 | 41 | 744 | NA | 2 | 15 | | Challenge | 84 | 81 | 96.4 | 55 | 52 | 94.6 | 76 | 90.5 | 30 | 54 | NA | 2 | 6 | | Combined | 869 | 850 | 97.8 | 403 | 384 | 95.3 | 844 | 97.1 | 71 | 798 | NA | 4 | 21 | EA - Essential Agreement CA - Category Agreement Eval - Evaluable isolates NS - Non-Susceptible isolates S - Susceptible isolates NA - Not Applicable min - Minor errors maj - Major errors vmj - Very Major errors Essential Agreement (EA) occurs when there is agreement between the MIC result of the reference method and that of the BD Phoenix within plus or minus one serial two-fold dilution of the antibiotic. Evaluable results are those that are on scale for both the BD Phoenix and the reference method or those in which an off-scale result is at least two doubling dilutions from the on-scale result. Category Agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation of the BD Phoenix. For Enterobacterales using manual inoculation method, the combined clinical and challenge results (869 isolates) were acceptable at 97.8% and 97.1% for EA and CA respectively. There were four (4) potential major errors and 21 potential very major errors (Table 3). Since no interpretive category is defined other than "susceptible" for eravacycline for all organisms evaluated, isolates for which the MIC values are above the susceptible breakpoint are reported as non-susceptible. When categorical errors occur, these are considered potential errors. Additionally, due to the lack of a category other than "susceptible" for eravacycline when testing Enterobacterales, further analysis of the errors is performed, and adjustments are made by considering the MIC values where the errors occurred. All four (4) of the potential major errors had MIC values that were in essential agreement with the reference MIC value; therefore, the adjusted potential major error rate is 0% (0/798) and is acceptable. Ten (10) of the 21 potential very major errors had MIC values that were in essential agreement with the reference MIC values; therefore, the adjusted potential very major error rate is 15.5% (11/71). When evaluating by individual species, three (3) adjusted potential very major errors were due to E. cloacae isolates (15.8% (3/19)); three (3) adjusted potential very major errors were due to E. coli isolates (100% (3/3)); two (2) adjusted potential very major errors were due to K. aerogenes (20% (2/10)); and three (3) adjusted potential very major errors were due to K. pneumoniae (8.1% (3/37)). Based on the total number of isolates evaluated in the clinical study, the high EA and CA performance for each species, and additional labeling mitigations, the performance was considered acceptable. K251713 - Page 9 of 13 {9} To address the high adjusted potential very major error rate with $E.$ cloacae, $E.$ coli, $K.$ aerogenes and $K.$ pneumoniae, the following limitations were added to the device labeling: ## Enterobacter cloacae For Enterobacter cloacae, two of the five potential very major errors were within essential agreement when compared to the reference method. Due to the susceptible-only category, the potential very major error rate was adjusted to $15.8\%$ (3/19). If deemed necessary for patient care, confirmatory testing with an alternate method could be performed for this organism before reporting Phoenix MIC results of 0.25 $\mu \mathrm{g} / \mathrm{mL}$. ## Escherichia coli For Escherichia coli, none of the three potential very major errors were within essential agreement when compared to the reference method. Due to the susceptible-only category, the potential very major error rate was not adjusted and remained at $100\%$ (3/3). These errors occurred with a very low percentage of the total isolates evaluated (3/376, $0.8\%$). If deemed necessary for patient care, confirmatory testing with an alternate method could be performed for this organism before reporting Phoenix MIC results of $\leq 0.125~\mu \mathrm{g / mL}$ or $0.25~\mu \mathrm{g / mL}$. ## Klebsiella pneumoniae For Klebsiella pneumoniae, eight of the eleven potential very major errors were within essential agreement when compared to the reference method. Due to the susceptible-only category, the potential very major error rate was adjusted to $8.1\%$ (3/37). If deemed necessary for patient care, confirmatory testing with an alternate method could be performed for this organism before reporting Phoenix MIC results of $0.25\mu \mathrm{g} / \mathrm{mL}$. ## Klebsiella aerogenes For Klebsiella aerogenes, neither of the two potential very major errors were within essential agreement when compared to the reference method. Due to the susceptible-only category, the potential very major error rate was adjusted to $20\%$ (2/10). If deemed necessary for patient care, confirmatory testing with an alternate method could be performed for this organism before reporting Phoenix MIC results of 0.25 $\mu \mathrm{g} / \mathrm{mL}$. The performance of testing eravacycline using the BD Phoenix AP inoculation method is illustrated in Table 4 below. Table 4. Eravacycline (0.125-2 μg/mL) Results, Phoenix AP Inoculation | | Tot | EA No. | EA % | Eval Tot | Eval EA No. | Eval EA % | CA Tot | CA % | No. NS | No. S | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Enterobacterales [≤ 0.5 (S)] | | | | | | | | | | | | | | | Challenge | 80 | 78 | 97.5 | 48 | 46 | 95.8 | 74 | 92.5 | 22 | 58 | NA | 5 | 1 | EA - Essential Agreement NS - Non-Susceptible isolates min - Minor errors K251713 - Page 10 of 13 {10} CA - Category Agreement Eval – Evaluable isolates S – Susceptible isolates NA – Not Applicable maj – Major errors vmj – Very Major errors Essential Agreement (EA) occurs when there is agreement between the MIC result of the reference method and that of the BD Phoenix within plus or minus one serial two-fold dilution of the antibiotic. Evaluable results are those that are on scale for both the BD Phoenix and the reference method or those in which an off-scale result is at least two doubling dilutions from the on-scale result. Category Agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation of the BD Phoenix. For Enterobacterales using the BD Phoenix AP inoculation method, the challenge results (80) were acceptable at 97.5% and 92.5% for EA and CA respectively. There were five potential major errors and one potential very major error (Table 4). Since no category is defined other than “susceptible” for eravacycline for all organisms evaluated, isolates for which the MIC values are above the susceptible breakpoint are reported as non-susceptible. When categorical errors occur, these are considered potential errors. Additionally, due to the lack of a category other than “susceptible” for eravacycline when testing Enterobacterales, further analysis of the errors is performed, and adjustments are made by considering the MIC values where the errors occurred. All five (5) of the potential major errors had MIC values that were in essential agreement with the reference MIC values; therefore, the adjusted potential major error rate is 0% (0/58) which is acceptable. The one potential very major error had an MIC value that was in essential agreement with the reference MIC value; therefore, the adjusted potential very major error rate is 0% (0/22) and is acceptable. ## MIC Trending A trending analysis was conducted using the combined data (clinical and challenge) obtained from the manual inoculation method. This trending calculation takes into account MIC values that are determined to be one or more doubling dilutions lower or higher than the reference method irrespective of whether the device MIC values are on-scale or not. Organism groups or species for which the difference between the percentage of isolates with higher vs. lower readings was &gt;30% and for which the confidence interval was determined to be statistically significant were considered to show evidence of trending. Trending that showed higher or lower MIC values compared to the reference is addressed in the labeling. Evaluation of results for species within Enterobacterales with eravacycline using the manual inoculation method are summarized in Table 5. A trend toward lower MIC values was observed for Citrobacter freundii, Citrobacter koseri, and Escherichia coli when compared to the CLSI broth microdilution reference method. To address the MIC trending, the following footnote was added to the device labeling: BD Phoenix Eravacycline MIC values tended to be in exact agreement or at least one doubling dilution lower when testing Citrobacter freundii, Citrobacter koseri, and Escherichia coli compared to the reference broth microdilution method. K251713 - Page 11 of 13 {11} Table 5. Trending of Eravacycline (0.125-2 μg/mL) Results with Manual Inoculation | Organism | Total Evaluable for Trending | ≥ 1 Dilution Lower No. (%) | Exact No. | ≥ 1 Dilution Higher No. (%) | Percent Difference (CI) | Trending Noted | | --- | --- | --- | --- | --- | --- | --- | | Citrobacter freundii | 22 | 11 (50.0) | 10 | 1 (4.6) | -45% (-65%, -20) | Yes, Low | | Citrobacter koseri | 14 | 10 (71.4) | 4 | 0 (0.0) | -71% (-88%, -38%) | Yes, Low | | Enterobacter cloacae | 78 | 24 (30.8) | 49 | 5 (6.4) | -24% (-36%, -12%) | No | | Escherichia coli | 94 | 75 (79.8) | 14 | 5 (5.3) | -74% (-82%, -63%) | Yes, Low | | Klebsiella aerogenes | 58 | 14 (24.1) | 40 | 4 (6.9) | -17% (-30%, -4%) | No | | Klebsiella oxytoca | 34 | 9 (26.5) | 12 | 13 (38.2) | 12% (10%, 32%) | No | | Klebsiella pneumoniae | 227 | 40 (17.6) | 162 | 25 (11.0) | -7% (-13%, 0%) | No | As required under 511A(2)(2)(B) of the Federal Food, Drug and Cosmetic Act, the following statement is added to the Precautions section of the device labeling: Per the FDA-Recognized Susceptibility Test Interpretive Criteria website, the safety and efficacy of antimicrobial drugs for which antimicrobial susceptibility is tested by this AST device, may or may not have been established in adequate and well-controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labelling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved. 2. Matrix Comparison: Not applicable C Clinical Studies: 1. Clinical Sensitivity: Not applicable 2. Clinical Specificity: Not applicable 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): Not applicable D Clinical Cut-Off: K251713 - Page 12 of 13 {12} Not applicable ## E Expected Values/Reference Range: Table 6. FDA-Recognized Interpretive Criteria for Eravacycline | Organisms | Minimum Inhibitory Concentrations (μg/mL)^{a} | | | | --- | --- | --- | --- | | | Susceptible | Intermediate | Resistant | | Enterobacterales^{b} | ≤0.5 | - | - | aAccording to the FDA STIC Webpage bClinical efficacy was shown for Citrobacter freundii, Enterobacter cloacae, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae. ## VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. ## IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission included a predetermined change control plan (PCCP) that was previously reviewed and accepted by FDA in submission K233986 cleared on March 15, 2024. This PCCP addresses future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage (https://www.fda.gov/Drugs/DevelopmentApprovalProcess/DevelopmentResources/ucm410971.htm). The PCCP outlined the specific procedures and acceptance criteria that BD intends to use to evaluate the BD Phoenix Automated Microbiology System – GN Eravacycline when revised breakpoints for eravacycline are published on the FDA STIC webpage. The PCCP included with the submission indicated that if specific criteria are met, BD will update the eravacycline device label to include (1) the new breakpoints, (2) an updated performance section after re-evaluation of data in this premarket notification with the new breakpoints, and (3) any new limitations as determined by their evaluation. K251713 - Page 13 of 13
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...