BD Phoenix™ Automated Microbiology System - GN Ciprofloxacin (0.0156–4 µg/mL)
K233986 · Becton, Dickinson and Company · LON · Mar 15, 2024 · Microbiology
Device Facts
Record ID
K233986
Device Name
BD Phoenix™ Automated Microbiology System - GN Ciprofloxacin (0.0156–4 µg/mL)
Applicant
Becton, Dickinson and Company
Product Code
LON · Microbiology
Decision Date
Mar 15, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1645
Device Class
Class 2
Attributes
PCCP
Indications for Use
The BD Phoenix Automated Microbiology System is intended for the in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of Gram Negative aerobic and facultative anaerobic bacteria belonging to the family Enterobacterales and non-Enterobacterales.
Device Story
The BD Phoenix Automated Microbiology System is a broth-based microdilution system for antimicrobial susceptibility testing (AST). It uses a sealed, 136-well polystyrene tray containing dried antimicrobial agents and growth/fluorescent controls. The system utilizes a redox indicator (colorimetric oxidation-reduction) to detect bacterial growth; organisms growing in the presence of an antibiotic reduce the indicator, changing it from blue to pink. The system is used in clinical laboratories by trained personnel. It processes samples prepared via the BD PhoenixSpec Nephelometer or the BD Phoenix AP instrument. The instrument continuously incubates panels at 35°C and takes automated readings every 20 minutes. The software-driven expert system (BDXpert) interprets results based on CLSI/FDA breakpoints to provide MIC values and categorical interpretations (S, I, R). This automated process replaces manual reading, standardizes inoculum preparation, and assists clinicians in selecting appropriate antimicrobial therapy by identifying bacterial resistance profiles.
Clinical Evidence
Performance was evaluated using 129 clinical and challenge Salmonella isolates. Combined testing showed 98.4% Essential Agreement (EA) and 94.6% Category Agreement (CA) using the manual inoculation method. Reproducibility was >95% across three sites. No major or very major errors were observed. Testing included both manual (PhoenixSpec) and automated (BD Phoenix AP) inoculation methods, both meeting acceptance criteria (>90% EA/CA).
Technological Characteristics
Broth-based microdilution system using a 136-well molded polystyrene tray. Employs a redox indicator (oxidation-reduction) for colorimetric growth detection. Operates at 35°C ± 1°C. Connectivity includes barcode scanning and automated instrument reading. Software-driven expert system (BDXpert) interprets results. Inoculum standardized to 0.5 McFarland using nephelometry or automated AP instrument. Cation-adjusted Mueller-Hinton broth with 0.010% Tween 80.
Indications for Use
Indicated for in vitro quantitative determination of antimicrobial susceptibility (MIC) of Gram-negative aerobic/facultative anaerobic bacteria (Enterobacteriaceae, Non-Enterobacteriaceae) and Gram-positive bacteria (Staphylococcus, Streptococcus) from pure culture. This notification specifically covers Ciprofloxacin (0.0156-4 ug/mL).
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
BD Phoenix™ Automated Microbiology System - Moxifloxacin-0.125-8 and Ciprofloxacin-0.25-4 µg/mL (K060217)
Submission Summary (Full Text)
{1}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K233986
B Applicant
Becton, Dickinson and Company
C Proprietary and Established Names
BD Phoenix Automated Microbiology System - GN Ciprofloxacin (0.0156-4 µg/mL)
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LON | Class II | 21 CFR 866.1645 - Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System | MI - Microbiology |
## II Submission/Device Overview:
### A Purpose for Submission:
1. Reporting range extension of ciprofloxacin to the BD Phoenix Gram Negative ID/AST and AST only Phoenix panels from 0.25 – 4 µg/mL to 0.0156 – 4 µg/mL to accommodate the updated FDA-recognized breakpoints for Salmonella species as published in the FDA STIC website. The reporting range extension only applies to Salmonella species.
2. Update the BD Phoenix Gram negative ID/AST and AST only Phoenix panels (0.25 – 4 µg/mL) for the current reporting range to include updated FDA-recognized breakpoints for Enterobacterales and Pseudomonas aeruginosa, as published in the FDA STIC website.
3. Establish a Pre-Determined Change Control Plan (PCCP) to address future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{2}
B Measurand:
Ciprofloxacin (0.0156–4 µg/mL), for Salmonella spp.
Ciprofloxacin (0.25–4 µg/mL), for Enterobacterales and Pseudomonas aeruginosa
C Type of Test:
Antimicrobial Susceptibility Test (Quantitative) colorimetric, oxidation-reduction, growth based.
III Intended Use/Indications for Use:
A Intended Use(s):
The BD Phoenix Automated Microbiology System is intended for the in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of Gram Negative aerobic and facultative anaerobic bacteria belonging to the family Enterobacterales and non-Enterobacterales.
B Indication(s) for Use:
The BD Phoenix Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus, and Streptococcus.
This premarket notification is for the BD Phoenix Automated Microbiology System with Ciprofloxacin at a concentration of 0.0156–4 µg/mL. Ciprofloxacin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
Active In Vitro and in Clinical Infections Against:
- Citrobacter koseri
- Citrobacter freundii
- Enterobacter cloacae
- Escherichia coli
- Klebsiella pneumoniae
- Morganella morganii
- Proteus mirabilis
- Proteus vulgaris
- Providencia rettgeri
- Providencia stuartii
- Pseudomonas aeruginosa
- Salmonella typhi
- Serratia marscescens
- Shigella boydii
- Shigella dysenteriae
- Shigella flexneri
- Shigella sonnei
K233986 - Page 2 of 13
{3}
Active In Vitro but clinical significance is unknown:
Edwardsiella tarda
Klebsiella aerogenes (formerly Enterobacter aerogenes)
Klebsiella oxytoca
Salmonella enteritidis
## C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
Results for the following antimicrobic/organism combination(s) are suppressed from reporting by the BD Phoenix System:
- Ciprofloxacin: Stenotrophomonas maltophilia
## D Special Instrument Requirements:
BD Phoenix Automated Microbiology System and software (V2.20.0.0 or higher)
PhoenixSpec Nephelometer
BD Phoenix AP Instrument
## IV Device/System Characteristics:
## A Device Description:
This submission is for a range extension of a single antimicrobial cleared for use on BD Phoenix ID/AST or AST only panels. The ID portion of the ID/AST combination panel was not subject to review in this submission.
The Phoenix AST method is a broth-based microdilution test. The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. The ID/AST combination panel includes an ID side (51 wells) with dried substrates for bacterial identification and an AST side (85 wells). The AST panel contains a wide range of two-fold doubling dilution concentrations of antimicrobial agents and growth and fluorescent controls at appropriate well locations. The AST panel does not include wells for isolate identification.
The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. The organism to be tested must be a pure culture and be preliminarily identified as gram-positive or gram-negative. Colonies are then suspended in ID broth and equated to a 0.5 McFarland suspension using a nephelometer device. A further dilution is made into AST broth (a cation-adjusted formulation of Mueller-Hinton broth containing 0.010% Tween 80), to which the redox-buffered oxidation-reduction AST indicator solution is added producing a blue color in the wells. The concentration of organisms in the final AST broth suspension is approximately 5 X 10⁵ CFU/mL.
The Phoenix AST Broth is poured into the inoculation port of the AST panel and the inoculum flows into the panel, filling panel wells. Polyethylene caps are applied to seal the inoculation ports. An air admittance port is located in the panel lid to ensure adequate oxygen tension in the panel for the duration of the test. Inoculated panels are barcode scanned and loaded into the BD
K233986 - Page 3 of 13
{4}
Phoenix Automated Microbiology System instrument where panels are continuously incubated at 35 °C ± 1 °C.
Continuous measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. The instrument takes readings every 20 minutes.
Organisms growing in the presence of a given antimicrobial agent reduce the indicator (changing it to a pink color). This signals organism growth and resistance to that antimicrobial agent.
Organisms killed or inhibited by the antimicrobial agent do not cause reduction of the indicator and therefore do not produce a color change. The Phoenix instrument reads and records the results of the antimicrobial tests contained in the panel and interprets the reactions (based on the organism identification) to give a minimal inhibitory concentration (MIC) value and category interpretations (susceptible, intermediate, resistant, or not susceptible). AST results are available within 16 hours. This is an auto read result; no manual readings are possible with this system.
Additional comments concerning specific organism/antimicrobial combinations are provided from the software-driven expert system (BDXpert), using rules derived from CLSI documentation and/or the FDA-approved drug labeling.
## B Principle of Operation:
The BD Phoenix Automated Microbiology System is a broth-based microdilution method that utilizes a redox indicator (colorimetric oxidation-reduction) to enhance detection of organism growth. The MIC is determined by comparing growth in wells containing serial two-fold dilutions of an antibiotic to the growth in "growth control wells" that contain no antibiotic.
## V Substantial Equivalence Information:
## A Predicate Device Name(s):
BD Phoenix Automated Microbiology System - Moxifloxacin-0.125-8 and Ciprofloxacin-0.25 - 4 µg/ml
## B Predicate 510(k) Number(s):
K060217
## C Comparison with Predicate(s):
| Device & Predicate Device(s): | Device K233986 | Predicate K060217 |
| --- | --- | --- |
| Device Trade Name | BD Phoenix Automated Microbiology System – GN Ciprofloxacin (0.0156-4 µg/mL) | BD Phoenix Automated Microbiology System – GN Ciprofloxacin (0.25 – 4 µg/mL) |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The BD Phoenix Automated Microbiology System is intended for the in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by | Same |
K233986 - Page 4 of 13
{5}
K233986 - Page 5 of 13
| Device & Predicate Device(s): | Device K233986 | Predicate K060217 |
| --- | --- | --- |
| | minimal inhibitory concentration (MIC) of Gram Negative aerobic and facultative anaerobic bacteria belonging to the family Enterobacterales and non-Enterobacterales | |
| Antimicrobial Agent | Ciprofloxacin | Same |
| Source of Microorganisms for Testing | Bacterial colonies isolated from culture. | Same |
| Technology | Automated growth-based detection | Same |
| Methodology | Determination of MIC using serial two-fold dilution format | Same |
| Read Method | Automated | Same |
| Inoculation Methods | Manual: BD PhoenixSpec nephelometer
Automated: BD Phoenix AP Instrument | Same |
| Result Reported | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (S, I, R) | Same |
| Incubation Time | < 16 hours | Same |
| General Device Characteristic Differences | | |
| Indicated Organisms | Active In Vitro and in Clinical Infections Against:
Citrobacter koseri
Citrobacter freundii
Enterobacter cloacae
Escherichia coli
Klebsiella pneumoniae
Morganella morganii
Proteus mirabilis
Proteus vulgaris
Providencia rettgeri
Providencia stuartii
Pseudomonas aeruginosa
Salmonella typhi
Serratia marscescens
Shigella boydii
Shigella dysenteriae
Shigella flexneri
Shigella sonnei
Active In Vitro but clinical significance is unknown:
Edwardsiella tarda
Klebsiella aerogenes (formerly Enterobacter aerogenes) | Gram negative aerobic and facultative anaerobic bacteria belonging to Enterobacterales and non – Enterobacterales |
{6}
| Device & Predicate Device(s): | Device K233986 | Predicate K060217 |
| --- | --- | --- |
| | Klebsiella oxytoca
Salmonella enteritidis | |
| Breakpoints | Enterobacterales:
(S/I/R) ≤0.25 / 0.5 / ≥1 | Enterobacterales:
(S/I/R) ≤1 / 2 / ≥4 |
| | P. aeruginosa:
(S/I/R) ≤0.5 / 1 / ≥2 | P. aeruginosa:
(S/I/R) ≤1 / 2 / ≥4 |
| | Salmonella spp.: (S/I/R) ≤0.0625 / 0.125 – 0.5 / ≥1 | Salmonella spp.: (S/I/R) ≤1 / 2 / ≥4 |
| Reporting Range | Enterobacterales:
0.25 – 4 μg/mL | 0.25 – 4 μg/mL |
| | P. aeruginosa:
0.25 – 4 μg/mL | |
| | Salmonella spp.: 0.0156-4 μg/mL | |
VI Standards/Guidance Documents Referenced:
Standards/Guidance Documents Referenced
1. Guidance for Industry and FDA, Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems, August 28, 2009.
2. CLSI. Performance Standards for Antimicrobial Susceptibility Testing. 33rd ed. CLSI supplement M100. Clinical and Laboratory Standards Institute; 2023.
3. CLSI. Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically. 11th ed. CLSI supplement M07. Clinical Laboratory Standards Institute; 2018.
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Reproducibility of the BD Phoenix Automated Microbiology System - GN Ciprofloxacin (0.0156-4 μg/mL) was conducted at three clinical sites using a panel of 12 Salmonella isolates with on-scale MICs. The isolates were tested at each site in triplicate over three different days using both inoculation methods (manual and BD Phoenix AP) resulting in 324 data points (12 strains x 3 replicates x 3 sites x 3 days = 324) for each inoculation method. The isolates tested in the reproducibility study included Salmonella enterica ssp. enterica serovar Paratyphi A (1), and Salmonella species (11). The reproducibility was calculated
K233986 - Page 6 of 13
{7}
based on MIC values falling within ±1 dilution of the mode value for each isolate. As all results were on-scale, there was no calculation of worst-case scenario results. Quality control performed during performance of the reproducibility study was determined to be acceptable. The results of the study demonstrate that for ciprofloxacin with Salmonella spp., there was an overall reproducibility across test sites of greater than 95% (± 1 dilution) agreement when compared to the test mode; results were determined to be acceptable. (Table 1)
Table 1. Reproducibility for Ciprofloxacin (0.0156-4 μg/mL)
| Inoculation Method | Best Case Reproducibility |
| --- | --- |
| Manual | 100% |
| Phoenix AP | 99.7% |
As there is no change in the reporting range when testing Enterobacterales and P. aeruginosa with the BD Phoenix Automated Microbiology System – GN Ciprofloxacin (0.25-4 μg/mL), reproducibility data remains the same as was previously evaluated during review of K060217 and was determined to be acceptable (refer to the K060217 Decision Summary).
2. Linearity:
Not applicable
3. Analytical Specificity/Interference:
Not applicable
4. Assay Reportable Range:
This submission is for the extension of the ciprofloxacin concentrations tested and reporting range for Salmonella spp. The reporting range for Enterobacterales and P. aeruginosa remain at 0.25 – 4 μg/mL.
Table 2. Reporting Ranges for Ciprofloxacin
| Organism | Testing/Reporting Range |
| --- | --- |
| Salmonella spp. | 0.0156-4 μg/mL |
| Enterobacterales | 0.25-4 μg/mL |
| P. aeruginosa | 0.25-4 μg/mL |
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
The CLSI recommended QC organisms (E. coli ATCC 25922 and P. aeruginosa ATCC 27853) were tested a sufficient number of times (i.e., at least 20/site) at each of three testing sites. QC strains were tested using both manual and Phoenix AP inoculation methods and read by the BD Phoenix instrument. Results were acceptable for greater than 95% of tests performed using both inoculation methods.
K233986 - Page 7 of 13
{8}
Table 3. Quality Control for Ciprofloxacin
| Organism | Concentration (μg/mL) | Reference | BD Phoenix Results | |
| --- | --- | --- | --- | --- |
| | | | Manual Inoculation | Phoenix AP Inoculation |
| E. coliATCC 25922 | ≤0.015 | 128 | | |
| | 0.03 | 4 | | |
| | 0.06 | 1 | | |
| | 0.12 | | | |
| | 0.25* | | 89 | 82 |
| | 0.5 | | | |
| | 1 | | | |
| | 2 | | | |
| | 4 | | | |
| P. aeruginosaATCC 27853 | ≤0.015 | | | |
| | 0.03 | | | |
| | 0.06 | | | |
| | 0.12 | | | |
| | 0.25* | 80 | | |
| | 0.5 | 53 | 87 | 80 |
| | 1 | | 1 | 2 |
| | 2 | | | |
| | 4 | | | |
Grey shaded rows indicate the BD Phoenix Automated Microbiology System - GN Ciprofloxacin (0.25-4 $\mu \mathrm{g} / \mathrm{mL}$ ) expected range.
*The lower end range extension only applies to the Salmonella species. Therefore, the BD Phoenix instrument will not report an MIC below $0.25\mu \mathrm{g / mL}$ for E. coli or P. aeruginosa. Obtaining a value of $\leq 0.25\mu \mathrm{g / mL}$ for E. coli ATCC 25922 was considered an indicator that the quality control test results were acceptable. The expected range for E. coli remains $\leq 0.25\mu \mathrm{g / mL}$ and P. aeruginosa remains $\leq 0.25 - 1\mu \mathrm{g / mL}$ (refer to the K060217 Decision Summary).
Inoculum Density Check: The BD PhoenixSpec Nephelometer (primary inoculation method) was used to prepare the inocula for testing of the clinical, challenge, reproducibility, and QC isolates. The same inoculum suspension was used for both the Phoenix System and the reference method testing. The ability of the BD Phoenix AP instrument to prepare an appropriate inoculum was evaluated with challenge, QC, and reproducibility isolates. Validation data for both the PhoenixSpec and the Phoenix AP instrument was provided and found to be acceptable.
Growth Failure Rate: The growth failure rate for both inoculum preparation methods was $0\%$ .
Purity Check: Purity check plates were prepared for all isolates from each inoculation preparation method. Results were only included for pure cultures.
6. Detection Limit:
Not applicable
K233986 - Page 8 of 13
{9}
7. Assay Cut-Off:
Not applicable
B Comparison Studies:
1. Method Comparison with Predicate Device:
**Ciprofloxacin (0.0156 - 4 μg/mL)**
Results obtained with the BD Phoenix Automated Microbiology System – GN Ciprofloxacin (0.0156 - 4 μg/mL) panel were compared to results obtained with the CLSI frozen broth microdilution reference panel. Reference panels were prepared according to CLSI M07 guidelines. The range of dilutions evaluated with the reference panels was the same as that used for the BD Ciprofloxacin panel (0.0156 - 4 μg/mL).
**Medium**: BD Phoenix AST Broth – a cation-adjusted formulation of Mueller Hinton broth used in conjunction with the AST Indicator Solution, a redox indicator.
**Inoculum**: The BD Phoenix Spec Nephelometer, the primary inoculation method, was used to obtain a suspension approximating a 0.50 - 0.60 McFarland standard for all challenge and clinical isolates.
**Incubation**: 35°C ± 1°C
Clinical testing was conducted at three U.S. sites using 47 *Salmonella* isolates (7 contemporary [14.9%] and 40 stock [85.1%]) isolates. The low number of contemporary isolates tested was considered acceptable due to the prevalence of *Salmonella* spp. Strains tested included: *Salmonella enterica* ssp. *arizonae* (2 isolates), *Salmonella enterica* spp. *enterica serovar* Cholerasuis (3 isolates), *Salmonella enterica* spp. *enterica serovar* Typhi (4 isolates) and *Salmonella* spp. (38 isolates).
A total of 82 challenge isolates were evaluated at 3 sites. Strains tested included *Salmonella enterica* spp. *enterica serovar* Paratyphi A (9 isolates), *Salmonella enterica* spp. *enterica serovar* Typhi (3 isolates) and *Salmonella* spp. (70 isolates).
Results obtained for tests inoculated manually were acceptable.
Table 4. Ciprofloxacin (0.0156 - 4 μg/mL) Results, Manual Inoculation
| | Tot | EA N | EA % | Eval Tot | Eval EA N | Eval EA % | CA Tot | CA % | No. R | No S | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Salmonella enterica ssp. arizonae | | | | | | | | | | | | | |
| Clinical | 2 | 1 | 50 | 2 | 1 | 50 | 2 | 100 | 1 | 0 | 0 | 0 | 0 |
| Total | 2 | 1 | 50 | 2 | 1 | 50 | 2 | 100 | 1 | 0 | 0 | 0 | 0 |
| Salmonella enterica ssp. enterica serovar Choleraesuis | | | | | | | | | | | | | |
| Clinical | 3 | 3 | 100 | 1 | 1 | 100 | 3 | 100 | 0 | 2 | 0 | 0 | 0 |
| Total | 3 | 3 | 100 | 1 | 1 | 100 | 3 | 100 | 0 | 2 | 0 | 0 | 0 |
| Salmonella enterica ssp. enterica serovar Paratyphi A | | | | | | | | | | | | | |
| Challenge | 9 | 9 | 100 | 5 | 5 | 100 | 8 | 88.9 | 0 | 8 | 1 | 0 | 0 |
| Total | 9 | 9 | 100 | 5 | 5 | 100 | 8 | 88.9 | 0 | 8 | 1 | 0 | 0 |
| Salmonella enterica ssp. enterica serovar Typhi | | | | | | | | | | | | | |
K233986 - Page 9 of 13
{10}
| | Tot | EA N | EA % | Eval Tot | Eval EA N | Eval EA % | CA Tot | CA % | No. R | No S | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Clinical | 4 | 4 | 100 | 1 | 1 | 100 | 4 | 100 | 3 | 0 | 0 | 0 | 0 |
| Challenge | 3 | 3 | 100 | 0 | 0 | NA | 3 | 100 | 0 | 3 | 0 | 0 | 0 |
| Total | 7 | 7 | 100 | 1 | 1 | 100 | 7 | 100 | 3 | 3 | 0 | 0 | 0 |
| Salmonella spp. not differentiated | | | | | | | | | | | | | |
| Clinical | 38 | 37 | 97.4 | 30 | 29 | 96.7 | 38 | 100 | 8 | 13 | 0 | 0 | 0 |
| Challenge | 70 | 70 | 100 | 50 | 50 | 100 | 64 | 91.4 | 2 | 43 | 6 | 0 | 0 |
| Total | 108 | 107 | 99.1 | 80 | 79 | 98.8 | 102 | 94.4 | 10 | 56 | 6 | 0 | 0 |
| All Salmonella spp. and serovars Combined | | | | | | | | | | | | | |
| Clinical | 47 | 45 | 95.8 | 34 | 32 | 94.1 | 47 | 100 | 12 | 15 | 0 | 0 | 0 |
| Challenge | 82 | 82 | 100 | 55 | 55 | 100 | 75 | 91.5 | 3 | 54 | 7 | 0 | 0 |
| Total | 129 | 127 | 98.4 | 89 | 87 | 97.8 | 122 | 94.6 | 15 | 69 | 7 | 0 | 0 |
EA - Essential Agreement
CA - Category Agreement
Eval - Evaluable isolates
R - Resistant isolates
S - Susceptible isolates
min - minor errors
maj - major errors
vmj - very major errors
Essential Agreement (EA) occurs when there is agreement between the MIC result of the reference method and that of the BD Phoenix within plus or minus one serial two-fold dilution of the antibiotic. Evaluable results are those that are on scale for both the BD Phoenix and the reference method or those in which an off-scale result is at least two doubling dilutions from the on-scale result. Category Agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation of the BD Phoenix.
The BD Phoenix AP instrument, the secondary inoculation method, was evaluated using challenge, QC, and reproducibility isolates. The BD Phoenix AP instrument is designed to standardize the ID broth inoculum equivalent to the BD Phoenix Spec Nephelometer. Following suspension standardization, the instrument adds the preset amount of AST indicator broth to the AST broth tube and transfers the required aliquot of ID broth inoculum to AST broth tubes. Results obtained for tests inoculated using the BD Phoenix AP instrument were acceptable.
Table 5. Ciprofloxacin (0.0156 - 4 μg/mL) Results, Phoenix AP Inoculation Method
| | Tot | EA N | EA % | Eval Tot | Eval EA N | Eval EA % | CA Tot | CA % | No. R | No. S | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Salmonella enterica ssp. enterica serovar Paratyphi A | | | | | | | | | | | | | |
| Challenge | 9 | 9 | 100 | 5 | 5 | 100 | 8 | 88.9 | 0 | 8 | 1 | 0 | 0 |
| Total | 9 | 9 | 100 | 5 | 5 | 100 | 8 | 88.9 | 0 | 8 | 1 | 0 | 0 |
| Salmonella enterica ssp. enterica serovar Typhi | | | | | | | | | | | | | |
| Challenge | 3 | 3 | 100 | 0 | 0 | NA | 3 | 100 | 0 | 3 | 0 | 0 | 0 |
| Total | 3 | 3 | 100 | 0 | 0 | NA | 3 | 100 | 0 | 3 | 0 | 0 | 0 |
| Salmonella spp. | | | | | | | | | | | | | |
| Challenge | 70 | 70 | 100 | 35 | 35 | 100 | 67 | 95.7 | 1 | 44 | 3 | 0 | 0 |
| Total | 70 | 70 | 100 | 35 | 35 | 100 | 67 | 95.7 | 1 | 44 | 3 | 0 | 0 |
| All Salmonella spp. and serovars Combined | | | | | | | | | | | | | |
| Challenge | 82 | 82 | 100 | 40 | 40 | 100 | 78 | 95.1 | 1 | 55 | 4 | 0 | 0 |
| Total | 82 | 82 | 100 | 40 | 40 | 100 | 78 | 95.1 | 1 | 55 | 4 | 0 | 0 |
# Ciprofloxacin (0.25 - 4 μg/mL)
The BD Phoenix Automated Microbiology System - GN Ciprofloxacin (0.25-4 μg/mL) was originally cleared in K060217 and contained broad indications for testing gram-negative aerobic and facultative anaerobic bacteria belonging to the family of Enterobacteriaceae and non-Enterobacteriaceae species. Since FDA no longer recognizes breakpoints for
K233986 - Page 10 of 13
{11}
Acinetobacter spp. and other Non-Enterobacterales, the indications for use has been revised and limited to *P. aeruginosa* and specific Enterobacterales species, as noted on the FDA drug label for ciprofloxacin.
Since no changes were made to the design or dilution range for testing species other than for testing *Salmonella*, the performance of the BD Phoenix Automated Microbiology System - GN Ciprofloxacin (0.25-4 µg/mL) with Enterobacterales and *P. aeruginosa* using revised interpretive criteria currently recognized by the FDA, the sponsor provided their reanalysis of the original data to confirm that performance continues to be met in accordance with the AST Special Controls Guidance.
The combined clinical and challenge data for 2836 isolates (tested with the manual inoculation method) were re-analyzed with the updated breakpoints. This re-analysis included 2,169 Enterobacterales (2,085 clinical isolates and 84 challenge isolates) and 532 *P. aeruginosa* (524 clinical isolates and 8 challenge isolates). In addition, results from 152 non-indicated species (147 clinical isolates and 5 challenge isolates, representing <10% of the total number of isolates tested) were also re-analyzed. The performance data included in the device labeling included combined performance for Enterobacterales, *P. aeruginosa* and non-indicated species. To provide clarity on the performance provided in the device labeling, the following was included in a footnote to the performance table:
Enterobacterales (2169), Pseudomonas aeruginosa (532) and non-indicated species (152) were included in the above combined performance.
When applying the new breakpoints, the CA for Enterobacterales was >90% and considered acceptable. Reanalyzed results for *P. aeruginosa* breakpoints, showed a CA of 88.7% due mainly to an increased percentage of minor errors. The EA of evaluable results was 85.3%. To address the unacceptable CA, the following limitation was added to the device labeling:
Due to the occurrence of a high number of minor errors, isolates of *P. aeruginosa* that provide MICs of 1 µg/mL should be retested with an alternate method.
## Trending
A trending analysis was conducted using the combined data (clinical and challenge) obtained from the manual inoculation method. This trending calculation takes into account MIC values that are determined to be one or more doubling dilutions lower or higher than the reference method irrespective of whether the device MIC values are on-scale or not. Results that are not clearly at least one dilution lower, at least one dilution higher or in exact agreement with the CLSI reference method are not considered in the trending analysis.
Organism groups or species for which the difference between the percentage of isolates with higher vs. lower readings was >30% and for which the confidence interval was determined to be statistically significant were considered to show evidence of trending. Trending that showed higher or lower MIC values compared to the reference is addressed in the labeling.
Results for *Salmonella* spp. showed high trending for *Salmonella enterica* spp. arizonae, *Salmonella enterica* spp. enterica serovar Cholerasuis and *Salmonella enterica* spp. enterica
K233986 - Page 11 of 13
{12}
serovar Paratyphi A but were not statistically significant. Results for Salmonella spp. overall showed no trending.
Table 7. Trending of Ciprofloxacin (0.0156 - 4 μg/mL) Results
| Organism | Total Evaluable for Trending | ≥ 1 Dilution lower No. (%) | Exact No. (%) | ≥ 1 Dilution Higher No. (%) | Percent Difference (CI) | Trending Noted |
| --- | --- | --- | --- | --- | --- | --- |
| Salmonella enterica ssp. arizonae | 2 | 1 (50) | 1 | 0 (0) | -50% (-91% to 27%) | Yes* |
| Salmonella enterica ssp. enterica serovar Choleraesuis | 2 | 0 (0) | 1 | (50) | 50% (-27% to 91%) | Yes* |
| Salmonella enterica ssp. enterica serovar Paratyphi A | 9 | 1 (11.1) | 4 | 4 (44.4) | 33% (-8 to 64%) | Yes* |
| Salmonella enterica ssp. enterica serovar Typhi | 1 | 0 (0) | 1 | 0 (0) | 0% (-79% to 79%) | No |
| Salmonella species | 91 | 10 (11.0) | 65 | 16 (17.6) | 7% (-4% to 17%) | No |
| All Salmonella | 105 | 12 (11.4) | 72 | 21 (20) | 9% (-1% to 18%) | No |
*Not statistically significant.
As required under 511A(2)(2)(B) of the Federal Food, Drug and Cosmetic Act, the following statement is added to the Precautions section of the device labeling:
Per the FDA-Recognized Susceptibility Test Interpretive Criteria website, the safety and efficacy of antimicrobial drugs for which antimicrobial susceptibility is tested by this AST device, may or may not have been established in adequate and well-controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labelling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved.
2. Matrix Comparison:
Not applicable
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable
2. Clinical Specificity:
Not applicable
K233986 - Page 12 of 13
{13}
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable
## D Clinical Cut-Off:
Not applicable
## E Expected Values/Reference Range:
Table 8. FDA-Identified Interpretive Criteria for Ciprofloxacinᵃ
| Organism | Interpretive Criteria for Ciprofloxacin (μg/mL) | | |
| --- | --- | --- | --- |
| | Susceptible | Intermediate | Resistant |
| Salmonella spp. | ≤0.0625 | 0.125 – 0.5 | ≥ 1 |
| Enterobacterales | ≤ 0.25 | 0.5 | ≥ 1 |
| P. aeruginosa | ≤ 0.5 | 1 | ≥ 2 |
ᵃAccording to CLSI M100-Ed33 and FDA STIC Website
https://www.fda.gov/Drugs/DevelopmentApprovalProcess/DevelopmentResources/ucm410971.htm
## VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission included a predetermined change control plan (PCCP) that was reviewed and accepted by FDA, as described in the Antimicrobial Susceptibility Test (AST) System Devices – Updating Breakpoints in Device Labeling guidance. This PCCP addresses future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage
(https://www.fda.gov/Drugs/DevelopmentApprovalProcess/DevelopmentResources/ucm410971.htm). The PCCP outlined the specific procedures and acceptance criteria that Becton, Dickinson and Company intends to use to evaluate the BD Phoenix Automated Microbiology System - GN Ciprofloxacin when revised breakpoints for ciprofloxacin are published on the FDA STIC webpage. The PCCP included with the submission indicated that if specific criteria are met, Becton, Dickinson and Company will update the BD Phoenix Automated Microbiology System - GN Ciprofloxacin device label to include (1) the new breakpoints, (2) an updated performance section after re-evaluation of data in this premarket notification with the new breakpoints, and (3) any new limitations as determined by their evaluation.
K233986 - Page 13 of 13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.