BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - CEFTRIAXONE (GN) 0.5-64UG/ML
K062944 · Becton, Dickinson & CO · LON · Nov 15, 2006 · Microbiology
Device Facts
Record ID
K062944
Device Name
BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - CEFTRIAXONE (GN) 0.5-64UG/ML
Applicant
Becton, Dickinson & CO
Product Code
LON · Microbiology
Decision Date
Nov 15, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1645
Device Class
Class 2
Intended Use
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus and Streptococcus.
Device Story
BD Phoenix Automated Microbiology System performs rapid identification and antimicrobial susceptibility testing (AST) on bacterial isolates. System components include instrument, software, panels with biochemicals/antimicrobial agents, ID broth, AST broth, and AST indicator solution. Process: pure culture isolates inoculated into sealed, molded polystyrene panels; panels incubated at 35°C in instrument; instrument performs optical readings every 20 minutes measuring redox indicator changes and turbidity. Software interprets data to provide MIC values and categorical interpretations (S, I, R). Used in clinical microbiology laboratories by trained technicians. Output assists clinicians in selecting appropriate antimicrobial therapy for patients with bacterial infections.
Clinical Evidence
Performance evaluated using clinical, stock, and challenge isolates across multiple US sites. Results compared to CLSI reference broth microdilution method. Metrics include Essential Agreement (EA) and Category Agreement (CA). No specific sensitivity/specificity percentages provided in summary, but system demonstrated substantial equivalence to reference methods.
Indicated for the removal of the “Do Not Report” ceftriaxone limitation for Proteus vulgaris/penneri in patients requiring antimicrobial susceptibility testing for Gram-negative aerobic and facultative anaerobic bacteria.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
K173523 — BD Phoenix Automated Microbiology System - GN Meropenem-vaborbactam (0.125/8-32/8 ug/mL) · Becton, Dickinson and Company · Feb 9, 2018
K020322 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM · Becton, Dickinson & CO · Mar 27, 2002
K082852 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-NITROFURANTOIN (GP) 4-128 UG/ML · Becton, Dickinson & CO · Jan 29, 2009
K123404 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM- IMIPENEM (0.0625-32 UG/ML) · Becton, Dickinson & CO · Jun 27, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
| SUBMITTED BY: | Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152 Phone: 410-316-4938 Fax: 410-316-4499 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | NOV 15 2006 |
| CONTACT NAME: | Janine Matlak Regulatory Affairs Specialist |
| DATE PREPARED: | September 27, 2006 |
| DEVICE TRADE NAME: | BD Phoenix™ Automated Microbiology System - Ceftriaxone 0.5-64 µg/mL |
| DEVICE COMMON NAME: | Antimicrobial susceptibility test system-short incubation |
| DEVICE CLASSIFICATION: | Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Device, 21 CFR 866.1645 |
| PREDICATE DEVICES: | VITEK® System (PMA No. N50510) and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ceftriaxone (K060444, April 12, 2006) and Ceftriaxone (K032655, October 6, 2003). |
| INTENDED USE: | The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram- positive bacteria of human origin. |
510(K) SUMMARY
# DEVICE DESCRIPTION:
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:
- BD Phoenix instrument and software. .
- . BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents for AST determinations.
- . BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum.
- BD Phoenix AST Broth used for performing AST tests only. .
- BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth . determination.
{1}------------------------------------------------
The Phoenix AST method is a broth based microdilution test. The Phoenix system utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. Phoenix panels are inoculated with a specified organism density and placed in the instrument.
The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).
# DEVICE COMPARISON:
The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the CLSI (formerly NCCLS) reference broth microdilution method. This premarket notification provides data supporting the removal of the limitation for Proteus vulgaris/penneri and ceftriaxone on the BD Phoenix™ Automated Microbiology System.
# SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:
The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST panels prepared according to CLSI M7). The system has been evaluated as defined in the FDA guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA'', February 5, 2003.
{2}------------------------------------------------
# Clinical Studies
Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Gram-negative Phoenix panel format containing this antimicrobial agent. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method.
The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within + one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant).
Table 1 summarizes the performance for the isolates tested in this study.
#### Performance of BD Phoenix System for Gram-Negative Organisms by Drug Table 1:
| and and the comments of the first of the first of the first of the first of the first of the first of the first of the first of the first and<br>milmieroni<br>. | No. of Cattle Concession<br>tration<br>oncent. | Company of Concession Carder Carder States Black<br>- | | <br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>- | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|-------------------------------------------------------|---|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| eftriaxone<br>REMELER IS FILLER BE | I to a se an an | | 1 | | t<br>- |
# Conclusions Drawn from Substantial Equivalence Studies
The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", February 5, 2003. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), and Ceftriaxone (K032655, October 6, 2003 and K060444, April 12, 2006).
{3}------------------------------------------------
# SUMMARY INFORMATION FOR CEFTRIAXONE Available Range 0.5-64 µg/mL
## Performance
Accuracy
| Andre ( 8 1 m 3 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1<br>Antimicrobial | promote problem of Address of Address of Active Station and Concession Collection of Concessfully of Concessfully of Concessfully of Concessfully of Concessfully of Concessfu<br>oncentration<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | n<br>- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - | workers and it is a manus<br>2<br>A Mar Lander, 12, 11, 11, 11, 11, 1982, 11 11 | 17 | .<br>and the country of the first of the first of the first of the first of the first of the first of the first of the first of the first for the first and the first and the first<br>. I wanten an a i manufacturer in a season and a con- |
|------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|---------------------------------------------------------------------------------|----|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| THE CONTRACT WIT AND IN CONTRACTOr Concession of the Catalog of<br>t<br>(<br>ettriaxone<br>SHE RE LE RELLE NE BOLDER & | 110 m<br>i | | ﺮ | | C |
# Breakpoints - CLSI, FDA
| Organism | S | I | R |
|-------------------------------------------------------------------|----|-------|-------|
| Enterobacteriaceae | ≤8 | 16,32 | $≥64$ |
| <i>Pseudomonas aeruginosa</i> and other<br>Non-enterobacteriaceae | ≤8 | 16,32 | $≥64$ |
# Recommended Quality Control Organisms
| Quality Control Strain | MIC Range | Source |
|-----------------------------------|-------------------|-----------|
| Escherichia coli ATCC 25922 | 0.03 - 0.12 µg/mL | CLSI, FDA |
| Pseudomonas aeruginosa ATCC 27853 | 8–64 µg/mL | CLSI |
| | 8-32 µg/mL | FDA |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol.
Food and Drug Administration 2098 Gaither Road Rockville MD. 20850
Ms. Janine Matlak Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152
NOV 15 2006
k062944 Re: Trade/Device Name: BD Phoenix 10 Automated Microbiology System Ceftriaxone (0.5-64 ug/mL) - Gram Negative ID/AST or AST · Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: September 27, 2006 Received: September 28, 2006
Dear Ms. Matlak:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalient (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Sally, attorn
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 1 of 1
510(k) Number: K062944
Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent ceftriaxone (0.5-64 ug/mL) - Gram-Negative ID/AST or AST only Phoenix panels.
Indications for Use:
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus and Streptococcus.
This premarket notification is for ceftriaxone for the removal of the limitation for Proteus vulgaris/penneri from the original premarket notification (K032655, October 6, 2003) and subsequent premarket notification (K060444, April 12, 2006).
Ceftriaxone has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
#### Active in Clinical Infections Against:
Acinetobacter calcoaceticus Escherichia coli Proteus mirabilis Enterobacter aerogenes Klebsiella oxytoca Serratia marcescens Enterobacter cloacae Klebsiella pneumoniae Pseudomonas aeruginosa Morganella morganii Proteus vulgaris
### Active In Vitro Against:
Citrobacter koseri (formerly C. diversus) Citrobacter freundii Providencia species (including Providencia rettgeri)
Prescription Use (Per 21 CFR 801.109) Salmonella species (including Salmonella typhi)
Over-the-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Leddi L. Parly
Vision Sign-Off
Shigella species
Office of In Vitro Diagnostic Device Evaluation and Safety
Kof 2944
BD Diagnostic Systems Becton, Dickinson and Company
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.