The BD Phoenix ™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.
Device Story
BD Phoenix™ Automated Microbiology System performs automated antimicrobial susceptibility testing (AST). System uses molded polystyrene trays with 136 micro-wells containing dried reagents and AST indicator. Pure culture isolates are suspended in broth (0.5 McFarland), diluted in AST broth, and inoculated into panels. Instrument continuously incubates panels at 35°C; monitors colorimetric oxidation-reduction (blue to pink/colorless) via optical scanning every 20 minutes. Software interprets growth patterns to determine MIC values and categorical interpretations (SIR) using CLSI-derived rules. System provides automated results in 4-16 hours; no manual readings. Healthcare providers use MIC results to guide antimicrobial therapy for bacterial infections.
Clinical Evidence
Performance evaluated via multi-site clinical study (4 sites) comparing Phoenix™ results to CLSI broth dilution reference method. Combined data for Cefepime (n=1789) showed 95.2% EA and 92.9% CA. Combined data for Ceftriaxone (n=1872) showed 95.8% EA and 90.9% CA. Reproducibility >95%. Quality control testing met acceptance criteria.
Technological Characteristics
Broth-based microdilution AST system. Uses redox indicator (colorimetric oxidation-reduction) for growth detection. Instrumentation includes barcode scanner, incubator (35°C), and optical reader. Trays are molded polystyrene with 136 micro-wells. Connectivity: Automated system with software-driven expert rules. Sterilization: Not specified.
Indications for Use
Indicated for in vitro quantitative determination of antimicrobial susceptibility by MIC of Gram-negative aerobic/facultative anaerobic bacteria (Enterobacteriaceae, Non-Enterobacteriaceae) and Gram-positive bacteria (Staphylococcus, Enterococcus, Streptococcus) from pure culture.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
VITEK® System (PMA No. N50510)
BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321)
BD Phoenix™ Automated Microbiology System Cefepime-0.5-64 µg/mL (K032675)
BD Phoenix™ Automated Microbiology System Ceftriaxone-0.5-64 µg/mL (K032655)
Related Devices
K020322 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM · Becton, Dickinson & CO · Mar 27, 2002
K082852 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-NITROFURANTOIN (GP) 4-128 UG/ML · Becton, Dickinson & CO · Jan 29, 2009
K062944 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - CEFTRIAXONE (GN) 0.5-64UG/ML · Becton, Dickinson & CO · Nov 15, 2006
K033458 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - CONFIRMATORY ESBL TEST · Becton, Dickinson & CO · May 28, 2004
K123404 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM- IMIPENEM (0.0625-32 UG/ML) · Becton, Dickinson & CO · Jun 27, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(K) SUMMARY
| SUBMITTED BY: | Becton, Dickinson and Company<br>7 Loveton Circle<br>Sparks, MD 21152<br>Phone: (410) 316 - 4278<br>Fax: 410-316-4499 |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT NAME: | Monica E Giguere<br>Regulatory Affairs Specialist |
| DATE PREPARED: | March 30, 2006 |
| DEVICE TRADE NAME: | BD Phoenix <sup>™</sup> Automated Microbiology System --<br>Cefepime-0.5-64 µg/mL and Ceftriaxone-0.5-64 µg/mL, |
| DEVICE COMMON NAME: | Antimicrobial susceptibility test system-short incubation |
| DEVICE CLASSIFICATION: | Fully Automated Short-Term Incubation Cycle Antimicrobial<br>Susceptibility Device, 21 CFR 866.1645 |
| PREDICATE DEVICES: | VITEK <sup>®</sup> System (PMA No. N50510) and BD Phoenix <sup>™</sup><br>Automated Microbiology System with Gatifloxacin (K020321,<br>May 23, 2002), Cefepime-0.5-64 µg/mL (K032675, March 1,<br>2004) and Ceftriaxone -0.5-64 µg/mL (K032655,<br>October 6, 2003) |
| INTENDED USE: | The BD Phoenix <sup>™</sup> Automated Microbiology System is<br>intended for the rapid identification and <i>in vitro</i> antimicrobial<br>susceptibility testing of isolates from pure culture of most<br>aerobic and facultative anaerobic Gram-negative and Gram-<br>positive bacteria of human origin. |
#### DEVICE DESCRIPTION:
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:
- . BD Phoenix instrument and software.
- BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . for AST determinations.
- BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. .
- . BD Phoenix AST Broth used for performing AST tests only.
{1}------------------------------------------------
- BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth . determination.
The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. Phoenix panels are inoculated with a specified organism density and placed into the instrument.
The instrument houses the panels where they are continuously incubated at a nominal temperature of 35℃. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).
# DEVICE COMPARISON:
The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the CLSI reference broth microdilution method. This premarket notification provides data for additional organism groups with Cefepime - 0.5-64 ug/mL and Ceftriaxone - 0.5-64 µg/mL and the BD Phoenix™ Automated Microbiology System.
# SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:
The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems: Guidance for Industry and FDA," February 5, 2003.
# Site Reproducibility
Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel chosen isolates. Each site tested the isolates in triplicate on three different days using one lot of Phoenix panels containing the antimicrobial agents and associated reagents.
The results of the study demonstrate for each antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the isolates tested.
{2}------------------------------------------------
## Clinical Studies
Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Phoenix panel formats containing antimicrobial agents. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method.
The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within ± one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant).
Table 1 summarizes the performance for the isolates tested in this study.
#### Table 1: Performance of BD Phoenix System by Drug
| Antimicrobial | Concentration | EA (n) | EA (%) | CA (n) | CA (%) |
|---------------|---------------|--------|--------|--------|--------|
| Cefepime | 0.5-64 μg/mL | 1789 | 95.2 | 1789 | 92.9 |
| Ceftriaxone | 0.5-64 μg/mL | 1872 | 95.8 | 1872 | 90.9 |
#### Conclusions Drawn from Substantial Equivalence Studies
The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems: Guidance for Industry and FDA." February 5, 2003. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Cefepime-0.5-64 µg/mL (K032675, March 1, 2004) and Ceftriaxone -0.5-64 µg/mL (K032655, October 6, 2003).
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" is arranged in a circular fashion around the emblem.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
APR 1 2 2006
Ms. Monica E. Giguere Regulatory Affairs Specialist BD Diagnostic Systems 7 Loveton Circle Sparks. MD 21152
Re: k060444/ S1
Trade/Device Name: BD Phoenix™ Automated Microbiology System Cefepime (0.5-64 ug/mL) and Ceftriaxone (0.5-64 ug/mL) on Gram-negative ID/AST or AST Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: February 20, 2006 Received: February 21, 2006
Dear Ms. Giguere:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
#### Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Sales, a Forg
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 2
510(k) Number: K060444(S)
Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent Cefepime (0.5-64 µg/mL) and Ceftriaxone (0.5-64 µg/mL) on Gram-negative ID/AST or AST only Phoenix panels with additional organism groups
Indications for Use:
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus, and Streptococcus.
This premarket notification is for additional organism groups and BD Phoenix™ Automated Microbiology System with cefepime (0.5-64 µg/mL) and ceftriaxone (0.5-64 µg/mL) on Gramnegative ID/AST or AST only Phoenix panels.
Cefepime has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
#### Active In Vitro and in Clinical Infections Against:
Pseudomonas aeruginosa
Enterobacter species Escherichia coli Klebsiella pneumoniae
#### Active In Vitro Against:
Acinetobacter lwoffi Citrobacter koseri Citrobacter freundii Hafnia alvei Klebsiella oxytoca Morganella morganii Pantoea agglomerans Providencia rettgeri Providencia stuartii Serratia marcescens
Division Sign-Off
*Signature illegible for Sally Hojjat*
Office of In Vitre Diagnostic Device Evaluation and Safety
-10(k)_
{6}------------------------------------------------
Ceftriaxone has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
# Active In Vitro and in Clinical Infections Against:
Acinetobacter calcoaceticus Enterobacter aerogenes Enterobacter cloacae Escherichia coli
Klebsiella oxytoca Klebsiella pneumoniae Morganella morganii Proteus mirabilis
Serratia marcescens Pseudomonas aeruginosa
### Active In Vitro Against:
Citrobacter koseri (formerly C. diversus) Citrobacter freundii Providencia species (including Providencia rettgeri) Salmonella species Shigella species
Prescription Use (Per 21 CFR 801.109) Over-the-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
\$\int*{a}^{b}a*{1}h_{2}r\$
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K060444/SCC1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.