MICROSCAN SYNERGIES PLUS GRAM NEGATIVE MIC/COMBO PANELS WITH NETILMICIN (0.12-32 UG/ML)
K020182 · Dade Behring, Inc. · LON · Feb 6, 2002 · Microbiology
Device Facts
Record ID
K020182
Device Name
MICROSCAN SYNERGIES PLUS GRAM NEGATIVE MIC/COMBO PANELS WITH NETILMICIN (0.12-32 UG/ML)
Applicant
Dade Behring, Inc.
Product Code
LON · Microbiology
Decision Date
Feb 6, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1645
Device Class
Class 2
Intended Use
MicroScan® Synergies plus™ Gram-Negative MIC/Combo Panels are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic Gram-Negative bacilli (Enterobacteriaceae, glucose non-fermenters, and non-Enterobacteriaceae glucose fermenters. After inoculation, panels are read on the WalkAway® SI System or equivalent (upgraded WalkAway® 40 or WalkAway® 96) according to the Package Insert.
Device Story
MicroScan Synergies plus Gram-Negative MIC/Combo Panels are miniaturized broth dilution susceptibility tests. Panels are inoculated with standardized suspensions of Gram-Negative bacilli and incubated at 35°C in WalkAway SI, 40, or 96 systems. The device determines the Minimum Inhibitory Concentration (MIC) by identifying the lowest antimicrobial concentration that inhibits bacterial growth. Used in clinical microbiology laboratories by technicians to guide antimicrobial therapy. Results assist clinicians in selecting appropriate antibiotic treatments for patients with Gram-Negative infections.
Clinical Evidence
External evaluation conducted using fresh clinical isolates and stock challenge strains. Performance compared against NCCLS frozen reference panels. Results demonstrated 95.1% overall Essential Agreement for Netilmicin. Reproducibility and precision testing performed using the WalkAway system with turbidity inoculum preparation. Quality control testing yielded acceptable results.
Technological Characteristics
Miniaturized broth microdilution panels. Uses automated incubation and reading via WalkAway SI, 40, or 96 systems. Employs turbidity-based inoculum preparation. Designed for aerobic and facultative anaerobic Gram-Negative bacilli. Connectivity via integration with WalkAway instrument platforms.
Indications for Use
Indicated for antimicrobial susceptibility testing of rapidly growing aerobic and facultative anaerobic Gram-Negative bacilli (Enterobacteriaceae, glucose non-fermenters, and non-Enterobacteriaceae glucose fermenters) using Netilmicin (0.12-32 ug/ml). Not for use with Proteus mirabilis.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
MicroScan Dried Gram Negative MIC/Combo Panels
Related Devices
K020319 — MICROSCAN SYNERGIES PLUS GRAM NEGATIVE MIC/COMBO PANELS WITH AMPICILLIN/SULBACTAM (0.5/0.25-64/34 UG/ML) · Dade Behring, Inc. · Apr 17, 2002
K020398 — MICROSCAN SYNERGIES PLUS GRAM NEGATIVE MIC/COMBO PANELS WITH AMPICILLIN (0.25-128 UG/ML) · Dade Behring, Inc. · May 16, 2002
K031602 — MICROSCAN SYNERGIES PLUS GRAM NEGATIVE MIC/COMBO PANELS WITH NITROFURANTOIN (1-256 UG/ML) · Dade Behring, Inc. · Jul 31, 2003
K020173 — MICROSCAN SYNERGIES PLUS GRAM NEGATIVE MIC/COMBO PANELS WITH TETRACYCLINE (0.12-128 UG/ML) · Dade Behring, Inc. · Feb 6, 2002
K023541 — MICROSCAN SYNERGIES PLUS GRAM NEGATIVE MIC/COMBO PANELS WITH MOXIFLOXACIN (0.004-16UG/ML) · Dade Behring, Inc. · Dec 13, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling three stylized waves or lines, stacked on top of each other.
AUG 3 0 2004
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. Robert Eusebio Manager Regulatory Affairs Dade MicroScan, Inc. 1584 Enterprise Boulevard West Sacramento, CA 95691
Re: k020182 Trade/Device Name: MicroScan® Synergies plus™ Gram-Negative MIC/Combo Panels with Netilmicin (0.12-32 ug/ml) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: July 30, 2004 Received: August 3, 2004
Dear Mr. Eusebio:
This letter corrects our substantially equivalent letter of February 6, 2002, regarding the trade name which was changed to MicroScan® Synergies Plus to better reflect the intended use of the device.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent [(for the indications for use stated in the enclosure)] to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-3084. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at their toll free number (800) 638-2041 or at (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
fu. Paolo
Aldington
Sally A. Hojvat, M.Sc., Ph.D Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known): K020182
Device Name: MicroScan® Synergies plus™ Gram-Negative MIC/Combo Panels with Netilmicin
## (0.12 - 32 ug/ml)
media of rapidly growing aerobic and facultative anaerobic Gram-Negative bac!II (Enterobacteriaceae, glucose non-fermenters, and non-Enterobacteriaceae glucose fermenters. After inoculation, panels are read on the WalkAway® SI System or equivalent (upgraded WalkAway® 40 or WalkAway® 96) according to the Package Insert.
This particular submission is for the antimicrobial Netilmicin on the Synergies plus™ Gram-Negative MIC/Combo Panels.
The Gram-Negative organisms which may be used for Netilmicin susceptibility testing in this panel are:
Acintobacter spp Citrobacter koseri (diversus) Citrobacter freundii Escherichia coli Enterobacter spp Klebsiella pneumoniae Proteus vulgaris Pseudomonas aeruginosa Serratia spp Shigella spp Salmonella spp
The MicroScan® Synergies plus™ Gram-Negative with Netilmicin is not intended for use with:
Proteus mirabilis
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off
on Sign-Off
Office of In Vitro Diagnostic Device
Evaluation and Seitro Diagnost
valuation and Safety
510(k) 020182
Page 1 of _ l
{3}------------------------------------------------
K020182
# 510(k) Summary
## 510(k) Submission Information:
| Device Manufacturer: | Dade Behring Inc. |
|----------------------|------------------------------------------------------------|
| Contact name: | Maureen Mende, Group Manager Regulatory Affairs |
| Fax: | 916-374-3144 |
| Date prepared: | January 17, 2002 |
| Product Name: | Microdilution Minimum Inhibitory Concentration (MIC) Panel |
| Trade Name: | MicroScan® rapID/S plus™ Gram-Negative MIC/Combo panel |
| Intended Use: | To determine antimicrobial agent susceptibility |
| 510(k) Notification: | Antimicrobials: Netilmicin |
| Predicate device: | MicroScan Dried Gram Negative MIC/Combo Panels |
#### 510(k) Summary:
MicroScan® rapID/S plus™ Gram-Negative MIC/Combo Panels are designed for use in determing quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-negative bacilli. The MicroScan® rapID/S plus ® Cramgrowing aeroon and racullanter and on the WalkAway Sy System or equivalent (upgraded WalkAway® 40 or WalkAway® 96 instruments).
The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test that have I le allumic oblar susceptions broth to concentrations bridging the range of clinical interest and are occar unders in Macher Innoin arout form. rapID/S plus™ panels are inoculated with a standardized suspension of the organism and incubated at 35°C in the WalkAway® SY System or equivalent Standardized Suspension of the organism and ancentration (MIC) for the test organism is determined by the lowest antimicrobial concentration showing inhibition of growth.
The proposed MicroScan® rapID/S plus™ Gram-Negative MIC/Combo Panel demonstrated substantially The proposed thinklobano rapon's prosvith an NCCLS frozen Reference Panel, as defined in the FDA equivalent performance when criteria for Assessment of Antimicrobial Susceptibility Devices", dated March 8, 2000. The Premarket Notification (510[k]) presents data in support of the MicroScan® rapID/S plus™ Gram-Negative MIC/Combo Panel with Netilmicin.
The external evaluation was conducted with fresh and stock Efficacy isolates and stock Challenge strains. The external evaluations were designed to confirm the acceptability of the proposed rapID/S plus ™ Gram-The extential evaluations were designed to ver with an NCCLS frozen Reference panel. Challenge strains were compared to Expected Results determined prior to the evaluation. The rapID/S plus "Foram-Negative Were compared acceptable performance with an overall Essential Agreement of 95.1% for Netilmicin when compared with the frozen Reference panel.
Instrument reproducibility testing demonstrated acceptable reproducibility and precision with Turbidity inculum preparation method and the WalkAway® System or equivalent (upgraded WalkAway® 40 or WalkAway® 96 instruments).
Quality Control testing demonstrated acceptable results for Netilmicin.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.