MICROSCAN SYNERGIES PLUS GRAM NEGATIVE MIC/COMBO PANELS WITH NITROFURANTOIN (1-256 UG/ML)
Applicant
Dade Behring, Inc.
Product Code
LON · Microbiology
Decision Date
Jul 31, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1645
Device Class
Class 2
Intended Use
The MicroScan Synergies plus™ Gram-Negative MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-negative bacilli. After inoculation, panels are incubated for 4.5 - 18 hours at 35°C +/- 1°C, in a WalkAway SI or equivalent, and read by the MicroScan® Instrumentation. Additionally, the panels may be incubated in a non-CO2 incubator and the AST portions can be read visually, according to the Package Insert.
Device Story
MicroScan Synergies plus Gram-Negative MIC/Combo Panels are miniaturized broth dilution susceptibility tests. Panels contain dehydrated antimicrobial agents in Mueller-Hinton broth. Clinical laboratory technicians inoculate panels with standardized bacterial suspensions; panels are rehydrated with water. Incubation occurs in WalkAway SI system or equivalent (4.5-18 hours at 35°C). Device determines Minimum Inhibitory Concentration (MIC) by identifying the lowest antimicrobial concentration inhibiting bacterial growth. Results are read via MicroScan instrumentation or visual inspection. Data assists clinicians in selecting appropriate antibiotic therapy for gram-negative infections, potentially improving patient outcomes by ensuring effective treatment.
Clinical Evidence
Bench testing only. Study compared MicroScan results to NCCLS M7 reference broth dilution method using 531 clinical isolates and 60 challenge organisms. Primary endpoints were Essential Agreement (EA) and Category Agreement (CA). Combined <16-hour results showed 99% EA and 87.1% CA. Reproducibility study across three sites showed >95% reproducibility. No major or very major errors observed in overnight (≥16h) readings. Minor errors noted near interpretive cutoffs.
Technological Characteristics
Miniaturized broth microdilution panels; dehydrated antimicrobial agents in Mueller-Hinton broth; rehydrated with water. Incubation at 35°C +/- 1°C. Compatible with WalkAway SI automated instrumentation or visual reading. Analyzes gram-negative bacilli susceptibility to Nitrofurantoin (1-256 µg/ml).
Indications for Use
Indicated for determining quantitative/qualitative antimicrobial susceptibility of rapidly growing aerobic and facultative anaerobic gram-negative bacilli (Enterobacteriaceae, glucose non-fermenters, non-Enterobacteriaceae glucose fermenters) isolated from solid media. Contraindicated for Morganella morganii, Proteus spp., Providencia spp., or Serratia spp. due to natural resistance to Nitrofurantoin.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
K020182 — MICROSCAN SYNERGIES PLUS GRAM NEGATIVE MIC/COMBO PANELS WITH NETILMICIN (0.12-32 UG/ML) · Dade Behring, Inc. · Feb 6, 2002
K162740 — MicroScan Dried Gram-Negative MIC/Combo Panels with Imipenem (0.25 - 8 ug/mL) · Beckman Coulter, Inc. · Mar 21, 2017
K020398 — MICROSCAN SYNERGIES PLUS GRAM NEGATIVE MIC/COMBO PANELS WITH AMPICILLIN (0.25-128 UG/ML) · Dade Behring, Inc. · May 16, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
AUG 3 0 2004
Mr. Robert Eusebio Manager Regulatory Affairs Dade MicroScan, Inc. 1584 Enterprise Boulevard West Sacramento, CA 95691
k031602 Re: Trade/Device Name: MicroScan® Synergies Plus Gram Negative MIC/Combo Panels with Nitrofurantoin (1 - 256 µg/ml) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System Regulatory Class: II Product Code: LON, JWY Dated: July 30, 2004 Received: August 3, 2004
### Dear Mr. Eusebio:
This letter corrects our substantially equivalent letter of July 31, 2003, regarding the trade name which was changed to MicroScan® Synergies Plus to better reflect the intended use of the device.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent [(for the indications for use stated in the enclosure)] to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-3084. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at their toll free number (800) 638-2041 or at (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely vours.
Luddu tu. Poole
for
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known): K031602
MicroScan® Synergies plus™ Gram-Negative MIC/Combo Panels with Nitrofurantoin Device Name: (1 - 256 ug/ml)
#### Indications For Use:
The MicroScan Synergies plus™ Gram-Negative MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-negative bacilli. After inoculation, panels are incubated for 4.5 - 18 hours at 35°C +/- 1°C, in a WalkAway SI or equivalent, and read by the MicroScan® Instrumentation. Additionally, the panels may be incubated in a non-CO2 incubator and the AST portions can be read visually, according to the Package Insert.
This particular submission is for the addition of the antimicrobial Nitrofurantoin, at concentrations of 1 to 256 ug/ml, to the test panel.
The gram-negative organisms which may be used for Nitrofurantoin susceptibility testing in this panel are:
> Escherichia coli Citrobacter amalonaticus Citrobacter freundii Citrobacter koseri (diversus) Klebsiella oxytoca Klebsiella ozaenae Klebsiella spp. * Enterobacter spp. *
* Some strains of Klebsiella species and Enterobacter species are resistant to Nitrofurantoin.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Lieddute. Pook
**Division Sign-Off**
Office of In Vitro Diagnostic Device Evaluation and Safety
Page 1 of _ 1
510(k)_031602
{3}------------------------------------------------
K03/602
## JUL 3 1 2003
### 510(k) Summary
#### 510(k) Submission Information:
| Device Manufacturer: | Dade Behring Inc. |
|----------------------|-------------------------------------------------------------|
| Contact name: | Robert Eusebio, Regulatory Affairs Manager |
| Fax: | 916-374-3144 |
| Date prepared: | May 15, 2003 |
| Product Name: | Microdilution Minimum Inhibitory Concentration (MIC) Panels |
| Trade Name: | MicroScan® rapID/S plus™ Gram-Negative MIC/Combo Panels |
| Intended Use: | To determine antimicrobial agent susceptibility |
| 510(k) Notification: | New antimicrobial - Nitrofurantoin |
| Predicate device: | MicroScan® rapID/S plus™ Gram Negative MIC/Combo Panels |
### 510(k) Summary:
MicroScan® rapID/S plus™ Gram-Negative MIC/Combo Panels are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-negative bacilli.
The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test that have been diluted in Mueller-Hinton Broth and dehydrated. Various antimicrobial agents are diluted in broth to concentrations bridging the range of clinical interest. Panels are rehydrated with water, after inoculation with a standardized suspension of the organism. After incubation in the WalkAway® SJ System, or equivalent, for 4.5 - 18 hours, the minimum inhibitory concentration (MIC) for the test organism is read by determining the lowest antimicrobial concentration showing inhibition of growth.
The proposed MicroScan® rapID/S plus" Gram-Negative MIC/Combo Panel demonstrated substantially equivalent performance when compared with an NCCLS frozen Reference Panel, as defined in the FDA document "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", dated February 5, 2003. The Premarket Notification (510[k]) presents data in support of the MicroScan® rapID/S plus™ Gram-Negative MIC/Combo Panel with Nitrofuranton.
The external evaluation was conducted with fresh and stock Efficacy isolates and stock Challenge strains. The external evaluations were designed to confirm the acceptability of the proposed rapID/S plus" (fram-Negative Panel by comparing its performance with an NCCLS frozen Reference panel. Challenge strains were compared to Expected Results determined prior to the evaluation. The rapID/S plus™ Gram-Negative Panel demonstrated acceptable performance with an overall Essential Agreement of >99% for Nitrofurantoin when compared with the frozen NCCLS Reference panel.
Instrument reproducibility testing demonstrated acceptable reproducibility and precision with Nitrofurantoin.
Quality Control testing demonstrated acceptable results for Nitrofurantoin.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.