TROVAFLOXACIN 10G, SENSI-DISC

K980683 · Becton Dickinson Microbiology Systems · JTN · May 1, 1998 · Microbiology

Device Facts

Record IDK980683
Device NameTROVAFLOXACIN 10G, SENSI-DISC
ApplicantBecton Dickinson Microbiology Systems
Product CodeJTN · Microbiology
Decision DateMay 1, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1620
Device ClassClass 2

Indications for Use

Antimicrobial Susceptibility Test Discs are used for semi-quantitative in vitro susceptibility testing by standardized agar diffusion test procedures. Trovafloxacin Sensi-Discs® are intended for use in determining the susceptibility to Trovafloxacin of a wide range of bacteria, as described under Indications For Use below. Zone sizes used for interpretation of tests, including control organism limits, were determined by the antimicrobic manufacturer, Roerig, a Division of Pfizer Inc., and received FDA approval under NDA Nos. 20-759 and 20-760.

Device Story

Device consists of high-quality paper discs impregnated with 10 µg of Trovafloxacin; used in clinical microbiology laboratories. Principle of operation: Bauer-Kirby agar diffusion method. Discs are placed on inoculated Mueller Hinton Agar (or specialized media for specific organisms); incubated; zones of inhibition measured. Healthcare providers compare zone sizes against established NCCLS/FDA standards to classify bacteria as susceptible, intermediate, or resistant. Output informs antimicrobial therapy selection; benefits patient by guiding targeted antibiotic treatment.

Clinical Evidence

Bench testing only. Performance is based on established agar diffusion techniques and interpretive criteria defined by the drug manufacturer (Roerig) and NCCLS standards (M2-A6, M100-S8). No clinical trial data provided in the submission.

Technological Characteristics

High-quality paper discs impregnated with 10 µg Trovafloxacin. Form factor: 50-disc cartridges. Principle: Agar diffusion (Bauer-Kirby method). Standards: NCCLS M2-A6 and M100-S8. No software or electronic components.

Indications for Use

Indicated for in vitro agar diffusion susceptibility testing to determine bacterial susceptibility to Trovafloxacin. Target organisms include various aerobic Gram-positive (e.g., S. aureus, S. pneumoniae), aerobic Gram-negative (e.g., E. coli, H. influenzae), anaerobic, and other microorganisms (e.g., Chlamydia, Mycoplasma) as specified in the Roerig package insert.

Regulatory Classification

Identification

An antimicrobial susceptibility test disc is a device that consists of antimicrobic-impregnated paper discs used to measure by a disc-agar diffusion technique or a disc-broth elution technique the in vitro susceptibility of most clinically important bacterial pathogens to antimicrobial agents. In the disc-agar diffusion technique, bacterial susceptibility is ascertained by directly measuring the magnitude of a zone of bacterial inhibition around the disc on an agar surface. The disc-broth elution technique is associated with an automated rapid susceptibility test system and employs a fluid medium in which susceptibility is ascertained by photometrically measuring changes in bacterial growth resulting when antimicrobial material is eluted from the disc into the fluid medium. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K980683 #### 1 MAY Date 2/20/98 ## SUMMARY OF SAFETY AND EFFECTIVENESS SUBMITTED BY: Virginia C. Weinknecht Requlatory Affairs Specialist Becton Dickinson Microbiology Systems 7 Loveton Circle Sparks, MD 21152-0999 ### NAME OF DEVICE: | Trade Name: | Trovafloxacin, 10 µg, Sensi-Discs Catalog Numbers 4331715 and 4331716 | |--------------------------|-----------------------------------------------------------------------| | Common Name/Description: | Antimicrobial Susceptibility Test Discs | | Classification Name: | Antimicrobial Susceptibility Test Discs | | INDICATE DEVICE: | Other BBL® Sensi-Discs® such as Ciprofloxacin, 5 mcg, Sensi-Disc® | ### DEVICE DESCRIPTION: PRE ### INTENDED USE: Antimicrobial Susceptibility Test Discs are used for semi-quantitative in vitro susceptibility testing by standardized agar diffusion test procedures. Trovafloxacin Sensi-Discs® are intended for use in determining the susceptibility to Trovafloxacin of a wide range of bacteria, as described under Indications For Use below. Zone sizes used for interpretation of tests, including control organism limits, were determined by the antimicrobic manufacturer, Roerig, a Division of Pfizer Inc., and received FDA approval under NDA Nos. 20-759 and 20-760. ### INDICATIONS FOR USE: Use of BBL® Trovafloxacin Sensi-Discs® for in vitro agar diffusion susceptibility testing is indicated when there is a need to determine the susceptibility of bacteria to Trovafloxacin. Trovafloxacin has been shown to {1}------------------------------------------------ be active against most strains of microorganisms listed below, as described in the Roerig package insert for this antimicrobic. # Active In Vitro and In Clinical Infections Against: Aerobic Gram-Positive Microorganisms Enterococcus faecalis (many strains are only moderately susceptible Staphylococcus aureus (methicillin-susceptible strains) Staphylococcus epidermidis (methicillin-susceptible strains) Streptococcus agalactiae Streptococcus pneumoniae (penicillin-susceptible strains) Streptococcus pyogenes Viridans group streptococci Aerobic Gram-Negative Microorganisms Escherichia coli Gardnerella vaginalis Haemophilus influenzae Haemophilus parainfluenzae Klebsiella pneumoniae Moraxella catarrhalis Neisseria gonorrhoeae Proteus mirabilis Pseudomonas aeruginosa Anaerobic Microorganisms Bacteroides fragilis Peptostreptococcus species Prevotella species Other Microorganisms: Chlamydia pneumoniae Chlamvdia trachomatis Legionella pneumophila Mycoplasma pneumoniae ### Active In Vitro Only Aqainst: Aerobic Gram-Positive Microorganisms Streptococcus pneumoniae (penicillin-resistant strains) Aerobic Gram-Negative Microorganisms Citrobacter freundii Enterobacter aerogenes Morganella morganii Proteus vulgaris Anaerobic Microorganisms Bacteroides distasonis Bacteroides ovatus Clostridium perfringens {2}------------------------------------------------ Active In Vitro Only Against (continued): Other Microorganisms: Mvcoplasma hominis Ureaplasma urealyticum ### PRODUCT DESCRIPTION: Trovafloxacin Susceptibility Test Discs are prepared by impregnating high guality paper with accurately determined amounts of Trovafloxacin supplied by the manufacturer, Roerig. Each Trovafloxacin disc is clearly marked on both sides with the agent and content. Trovafloxacin discs are furnished in cartridges of 50 discs each. Trovafloxacin cartridges are packed as either a single cartridge in a single box, or in a package containing ten cartridges. Agar diffusion methods employing dried filter paper discs impregnated with specific concentrations of antimicrobial agents were developed in the 1940's. In order to eliminate or minimize variability in the testing, Bauer et al. developed a standardized procedure in which Mueller Hinton Agar was selected as the test medium. Various regulatory agencies and standards-writing organizations subsequently published standardized reference procedures based on the Bauer-Kirby method. Among the earliest and most widely accepted of these standardized procedures were those published by the U.S. Food and Drug Administration (FDA) and the World Health Organization (WHO). The procedure was adopted as a consensus standard by the National Committee for Clinical Laboratory Standards (NCCLS) and is periodically updated. The latest NCCLS documents are M2-A6 (1/97) and M100-S7 (1/97). Discs containing a wide variety of antimicrobial agents are applied to the surface of Mueller Hinton Agar plates {or Haemophilus Test Medium Agar for Haemophilus influenzae or Mueller Hinton Aqar with 5% Sheep Blood for Streptococcus pneumoniae] inoculated with pure cultures of clinical isolates. Following incubation, the plates are examined and the zones of inhibition surrounding the discs are measured and compared with established zone size ranges for individual antimicrobial agents in order to determine the agent(s) most suitable for use in antimicrobial therapy. The determination as to whether the organism in question is susceptible (S), intermediate (I), or resistant (R) to an antimicrobial agent is made by comparing zone sizes to those found in the respective organism tables of National Committee for Clinical Laboratory Standards (NCCLS) Document M2-A6 ("Performance Standards for Antimicrobial Disk Susceptibility tests - Sixth Edition, Approved Standard", 1/97) and of NCCLS Document M100-S8 {3}------------------------------------------------ ("Performance Standards for Antimicrobial Susceptibility Testing", Eighth Informational Supplement, 1/98). PERFORMANCE DATA: See attached Roerig product insert section on Susceptibility Tests - Diffusion Techniques for TROVAN™ Tablets (trovafloxacin mesylate tablets) and TROVAN™ I.V. (alatrofloxacin mesylate injection) for intravenous infusion. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes and a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the caduceus. The seal is black and white. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 MAY 1 1998 1998 Virginia C. Weinknecht Regulatory Affairs Specialist Becton Dickinson Microbiology Systems 7 Loveton Circle Sparks, Maryland 21152-0999 Re: K980683 Trade Name: Trovafloxacin 10μg, Sensi-Disc® Regulatory Class: II Product Code: JTN Dated: February 20, 1998 Received: February 23, 1998 Dear Ms. Weinknecht: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Ouality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (OS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations. {5}------------------------------------------------ ### Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ Page 1 of 2 510(k) Number (if known): Device Name: Trovafloxacin, 10 ug, Sensi-Disc® Indications For Use: Use of BBL Trovafloxacin Sensi-Discs® for in vitro agar diffusion susceptibility testing is indicated when there is a need to determine the susceptibility of bacteria to Trovafloxacin has been shown to be active against most strains of microorganisms listed below, as described in the Roerig package insert for this antimicrobic. ### Active In Vitro and In Clinical Infections Against: ### Aerobic Gram-Positive Microorganisms Enterococcus faecalis (many only mod. suscept.) Streptococcus agalactiae Staphylococcus aureus (meth. suscept.) Streptococcus pneumoniae (pen. suscept.) Staphylococcus epidermidis (meth. suscept.) Streptococcus pyogenes Viridans group streptococci #### Aerobic Gram-Negative Microorganisms | Escherichia coli | Moraxella catarrhalis | |---------------------------------|------------------------| | Gardnerella vaginalis | Neisseria gonorrhoeae | | Haemophilus influenzae | Proteus mirabilis | | Haemophilus parainfluenzae | Pseudomonas aeruginosa | | Klebsiella pneumoniae | | | <b>Anaerobic Microorganisms</b> | | | Bacteroides fragilis | Prevotella species | | Peptostreptococcus species | | | <b>Other Microorganisms</b> | | | Chlamydia pneumoniae | Legionella pneumophila | | Chlamydia trachomotis | Mycoplasma pneumoniae | ### Active In Vitro Only Against: | <b>Aerobic Gram-Positive Microorganisms</b> | | |---------------------------------------------|----------------------------------------------------------------| | | <i>Streptococcus pneumoniae</i> (penicillin-resistant strains) | | <b>Aerobic Gram-Negative Microorganism</b> | | | <i>Citrobacter freundii</i> | <i>Morganella morganii</i> | | <i>Enterobacter aerogenes</i> | <i>Proteus vulgaris</i> | | <b>Anaerobic Microorganisms</b> | | | <i>Bacteroides distasonis</i> | <i>Clostridium perfringens</i> | | <i>Bacteroides ovatus</i> | | | <b>Other Microorganisms</b> | | | <i>Mycoplasma hominis</i> | <i>Ureaplasma urealyticum</i> | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | (Division Sign-Off) | | |-----------------------------------------|---------| | Division of Clinical Laboratory Devices | | | 510(k) Number | K980683 | OR | Prescription Use Per 21 CFR 801.109 | Over-The-Counter Use Optional Format 1-2-96 | |-------------------------------------|---------------------------------------------| |-------------------------------------|---------------------------------------------|
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...