K230651 · Becton, Dickinson and Company · JTN · Jun 7, 2023 · Microbiology
Device Facts
Record ID
K230651
Device Name
BD BBL™ Sensi-Disc™ Lefamulin 20µg (LMU-20)
Applicant
Becton, Dickinson and Company
Product Code
JTN · Microbiology
Decision Date
Jun 7, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1620
Device Class
Class 2
Indications for Use
BD BBL Sensi-Disc Antimicrobial Susceptibility Test (AST) Discs are used in the semi-quantitative agar diffusion test method for in vitro susceptibility testing. BD BBL Sensi-Disc Lefamulin Disc 20 µg (LMU-20) can be used to determine susceptibility to Lefamulin against the following bacteria, as described in the FDA-approved package insert for this antimicrobial agent. Active in vitro and in Clinical Infections Against: Gram-positive Bacteria Streptococcus pneumoniae Staphylococcus aureus (methicillin-susceptible isolates) Gram-negative Bacteria Haemophilus influenzae
Device Story
Device consists of 6mm absorbent paper discs impregnated with 20µg Lefamulin; used in clinical microbiology laboratories for semi-quantitative agar diffusion (Kirby-Bauer) susceptibility testing. Pure bacterial cultures are inoculated onto test media; discs are placed on the surface. Antibiotic diffuses into agar; after incubation, zones of inhibition are measured. Results are compared against established zone diameter breakpoints to determine susceptibility. Output is a zone diameter measurement (mm) used by clinicians to guide antimicrobial therapy decisions. Benefits include standardized, rapid assessment of bacterial susceptibility to Lefamulin, aiding in appropriate antibiotic selection for patients with infections caused by indicated organisms.
Clinical Evidence
Bench testing only. Performance evaluated using 336 clinical and challenge isolates (90 H. influenzae, 90 S. aureus, 90 S. pneumoniae clinical; 66 challenge). Overall Category Agreement (CA) was 98.5%. Reproducibility study (270 data points) showed >95% agreement between disc lots and across readers. Quality control testing met acceptance criteria.
Technological Characteristics
6mm absorbent paper discs impregnated with 20µg Lefamulin. Principle: Kirby-Bauer agar diffusion. Manual measurement of zone diameters. No electronic components, software, or energy source. Sterilization method not specified.
Indications for Use
Indicated for in vitro susceptibility testing of Streptococcus pneumoniae, Staphylococcus aureus (methicillin-susceptible), and Haemophilus influenzae to Lefamulin using the semi-quantitative agar diffusion method. Prescription use only.
Regulatory Classification
Identification
An antimicrobial susceptibility test disc is a device that consists of antimicrobic-impregnated paper discs used to measure by a disc-agar diffusion technique or a disc-broth elution technique the in vitro susceptibility of most clinically important bacterial pathogens to antimicrobial agents. In the disc-agar diffusion technique, bacterial susceptibility is ascertained by directly measuring the magnitude of a zone of bacterial inhibition around the disc on an agar surface. The disc-broth elution technique is associated with an automated rapid susceptibility test system and employs a fluid medium in which susceptibility is ascertained by photometrically measuring changes in bacterial growth resulting when antimicrobial material is eluted from the disc into the fluid medium. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of "ADMINISTRATION".
June 7, 2023
Becton, Dickinson and Company Katherine Cicala Regulatory Affairs Specialist 7 Loveton Circle Sparks, Maryland 21152-0999
Re: K230651
Trade/Device Name: BD BBL Sensi-Disc Lefamulin 20μg (LMU-20) Regulation Number: 21 CFR 866.1620 Regulation Name: Antimicrobial Susceptibility Test Disc Regulatory Class: Class II Product Code: JTN Dated: March 8, 2023 Received: March 9, 2023
Dear Katherine Cicala:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ribhi Shawar -S
Ribhi Shawar, Ph.D. (ABMM) Chief General Bacteriology and Antimicrobial Susceptibility Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230651
Device Name BD BBL Sensi-Disc Lefamulin 20ug (LMU-20)
Indications for Use (Describe)
BD BBL Sensi-Disc Antimicrobial Susceptibility Test (AST) Discs are used in the semi-quantitative agar diffusion test method for in vitro susceptibility testing.
BD BBL Sensi-Disc Lefamulin Disc 20 µg (LMU-20) can be used to determine susceptibility to Lefamulin against the following bacteria, as described in the FDA-approved package insert for this antimicrobial agent.
Active in vitro and in Clinical Infections Against: Gram-positive Bacteria Streptococcus pneumoniae Staphylococcus aureus (methicillin-susceptible isolates)
Gram-negative Bacteria Haemophilus influenzae
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
Summary Preparation Date: June 06, 2023
#### I Background Information
#### A 510(k) Number
K230651
#### B Applicant
BD Diagnostic Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, Maryland 21152 Establishment Registration Number: 1119779
#### C Proprietary and Established Name
BD BBL Sensi-Disc Lefamulin 20 ug (LMU-20)
#### D Regulatory Information
| Product Code | Classification | Regulation Section | Panel |
|--------------|----------------|----------------------------------------------------------------|-------------------|
| JTN | Class II | 21 CFR 866.1620 -<br>Antimicrobial Susceptibility<br>Test Disc | MI - Microbiology |
#### II Submission/Device Overview:
#### A Purpose for Submission:
To obtain substantial equivalence determination for Lefamulin Antimicrobial Susceptibility Test Disc.
#### B Measurand:
Lefamulin Disc 20 µg (LMU-20)
#### C Type of Test:
Antimicrobial Susceptibility Test Disc
#### III Intended Use/Indications for Use:
#### A Intended Use(s):
See Indications for Use below.
{5}------------------------------------------------
#### B Indication(s) for Use:
BD BBL Sensi-Disc Antimicrobial Susceptibility Test (AST) Discs are used in the semiquantitative agar diffusion test method for in vitro susceptibility testing.
BD BBL Sensi-Disc Lefamulin Disc 20 µg (LMU-20) can be used to determine susceptibility to Lefamulin against the following bacteria, as described in the FDA-approved package insert for this antimicrobial agent.
## Active in vitro and in Clinical Infections Against:
Gram-positive Bacteria Streptococcus pneumoniae Staphylococcus aureus (methicillin-susceptible isolates)
Gram-negative Bacteria Haemophilus influenzae
#### C Special Conditions for Use Statement(s):
- . Rx - For Prescription Use Only
- Per the FDA-Recognized Susceptibility Test Interpretive Criteria website, the safety and . efficacy of antimicrobial drugs, for which antimicrobial susceptibility is tested by this AST device, may or may not have been established in adequate and well-controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labeling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved.
- Limitations
- The current absence of resistant isolates precludes defining any results other than ● "Susceptible." Isolates yielding results other than "Susceptible" should be submitted to a reference laboratory for further testing.
#### D Special Instrument Requirements:
Not applicable.
#### IV Device/System Characteristics:
#### A Device Description:
The BD BBL Sensi-Disc Lefamulin 20 µg (LMU-20) device is a 6 mm disc prepared by impregnating high quality absorbent paper with accurately determined amounts of Lefamulin. Discs are clearly marked on both sides with the code LMU-20. The code designates the agent Lefamulin (LMU) and the drug content (20 ug).
BD BBL Sensi-Disc Antimicrobial Susceptibility Test Discs are furnished in cartridges containing 50 discs each. The last disc in each cartridge is marked "X" and contains the drug as coded. BD BBL Sensi-Disc Antimicrobial Susceptibility Test Discs can be dispensed using a BD BBL Sensi-Disc Dispenser.
{6}------------------------------------------------
#### B Principle of Operation:
A suitable therapeutic agent can be determined using filter paper discs impregnated with specified concentrations of antimicrobial agents placed on the surface of a suitable test medium. The test is performed by inoculating pure cultures of clinical isolates onto the test medium and placing the AST disc on the surface of the medium. The antibiotic within the disc diffuses into the agar. After incubation, the zones of inhibition around the discs are measured and compared against recognized zone diameter ranges for the specific antimicrobial agent/organism combinations being tested.
#### V Substantial Equivalence Information:
#### A Predicate Device Name:
HardyDisk AST Lefamulin 20μg (LMU20)
#### B Predicate 510(k) Number:
K192326
#### C Comparison with Predicate:
| Device &<br>Predicate Device: | Device:<br>K230651 | Predicate:<br>K192326 |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade | BD BBL Sensi-Disc Lefamulin | HardyDisk AST Lefamulin 20 |
| Name | 20 µg (LMU-20) | ug (LMU20) |
| | General Device Characteristic Similarities | |
| Regulation | 866.1620 | Same |
| Product Code | JTN | Same |
| Intended Use | See Indications for Use below. | HardyDisk AST Disks are used<br>for semi-quantitative in vitro<br>susceptibility testing by the agar<br>diffusion test procedure (Kirby-<br>Bauer) of rapidly growing and<br>certain fastidious bacterial<br>pathogens. Standardized methods<br>for agar diffusion testing have<br>been described for<br>Enterobacteriaceae,<br>Staphylococcus spp.,<br>Pseudomonas spp.,<br>Acinetobacter spp., Listeria<br>monocytogenes, Enterococcus<br>spp., and by modified<br>procedures, Haemophilus spp.,<br>Neisseria gonorrhoeae, N.<br>meningitidis and Streptococcus<br>spp., including Streptococcus<br>pneumoniae. |
| Device & | Device: | Predicate: |
| Predicate Device: | K230651 | K192326 |
| Indications for Use | BD BBL Sensi-Disc<br>Antimicrobial Susceptibility<br>Test (AST) Discs are used in<br>the semi-quantitative agar<br>diffusion test method for in<br>vitro susceptibility testing.<br>BD BBL Sensi-Disc Lefamulin<br>Disc 20 µg (LMU-20) can be<br>used to determine susceptibility<br>to Lefamulin against the<br>following bacteria, as described<br>in the FDA-approved package<br>insert for this antimicrobial<br>agent.<br>Active in vitro and in Clinical<br>Infections Against:<br>Gram-positive Bacteria<br>Streptococcus pneumoniae<br>Staphylococcus aureus<br>(methicillin-susceptible<br>isolates)<br>Gram-negative Bacteria<br>Haemophilus influenzae | HardyDisk AST Disks are used<br>for semi-quantitative in vitro<br>susceptibility testing by the agar<br>diffusion test procedure (Kirby-<br>Bauer) of rapidly growing and<br>certain fastidious bacterial<br>pathogens. Standardized methods<br>for agar diffusion testing have<br>been described for<br>Enterobacteriaceae,<br>Staphylococcus spp.,<br>Pseudomonas spp.,<br>Acinetobacter spp., Listeria<br>monocytogenes, Enterococcus<br>spp., and by modified<br>procedures, Haemophilus spp.,<br>Neisseria gonorrhoeae, N.<br>meningitidis and Streptococcus<br>spp., including Streptococcus<br>pneumoniae.<br>Use of HardyDisk AST<br>Lefamulin 20µg (LMU20) for in<br>vitro agar diffusion susceptibility<br>testing is indicated when there is<br>the need to determine the<br>susceptibility of bacteria to<br>Lefamulin.<br>HardyDisk AST Lefamulin at<br>concentration 20µg can be used<br>to determine the zone diameter<br>(mm) of Lefamulin against the<br>following bacteria for which<br>Lefamulin has been shown to be<br>active both clinically and in<br>vitro:<br>Streptococcus pneumoniae<br>Staphylococcus aureus<br>(methicillin-susceptible isolates)<br>Haemophilus influenzae |
| Antimicrobial<br>Agent | Lefamulin | Same |
| Antimicrobial | | |
| Agent | 20 µg | Same |
| Concentration | | |
| Device &<br>Predicate Device: | Device:<br>K230651 | Predicate:<br>K192326 |
| Interpretation | The user will interpret the zone<br>diameter according to the<br>established interpretive criteria<br>for the drug. | Same |
| Methodology | Kirby-Bauer Disk Diffusion<br>Susceptibility Test Protocol<br>requires the user to determine<br>categorical interpretations<br>(S/I/R) using the measured zone<br>diameters. | Same |
| Result<br>Interpretation<br>Method | Measurement of zone size. | Same |
| General Device Characteristic Differences | | |
| Manufacturing<br>Specifications | BD's specifications | Hardy Diagnostics'<br>specifications |
{7}------------------------------------------------
{8}------------------------------------------------
#### VI Standards/Guidance Documents Referenced:
- CLSI. Performance Standards for Antimicrobial Susceptibility Testing. 33rd ed. CLSI ● supplement M100. Clinical and Laboratory Institute; 2023.
- CLSI. Performance Standards for Antimicrobial Disk Susceptibility Tests. 13th ed. CLSI ● standard M02. Clinical and Laboratory Standards Institute; 2018.
- . CLSI. Methods for Antimicrobial Susceptibility Testing of Anaerobic Bacteria. 9th ed. CLSI standard M11. Wayne, PA: Clinical and Laboratory Standards Institute; 2018.
#### VII Performance Characteristics (if/when applicable):
#### Analytical Performance: A
## 1. Precision/Reproducibility:
Reproducibility was conducted at one external site using 15 organisms, tested in triplicate with two disc lots on three separate days using one lot of appropriate media (Mueller Hinton agar (MHA) for Staphylococcus aureus, Haemophilus Test Medium for Haemophilus influenzae, and MHA with 5% Sheep Blood for Streptococcus pneumoniae). Each test was visually read by three independent readers with results masked, resulting in 270 data points for evaluation (15 organisms x 2 disc lots x 1 media lot x 3 days x 3 independent readers = 270 data points).
The reproducibility study included 5 Haemophilus influenzae, 5 Staphylococcus aureus, and 5 Streptococcus pneumoniae. Reproducibility was calculated as the percent of results which were within ±3 mm difference in zone diameter comparing test results with the modal or median, if no mode is present, zone diameter value. Summary results between disk lots and across readers are shown in Table 1 below.
{9}------------------------------------------------
| Between Disc Lots | | | Across Readers | | | |
|-------------------|-----------|-----------|----------------|-----------|-----------|-------------|
| Lot #1 | Lot #2 | All Lots | Reader #1 | Reader #2 | Reader #3 | All Readers |
| 100.00% | 100.00% | 100.00% | 96.67% | 100.00% | 100.00% | 98.89% |
| (135/135) | (135/135) | (270/270) | (87/90) | (90/90) | (90/90) | (267/270) |
### Table 1: Reproducibility Summary
The reproducibility performance between disc lots and across readers is >95% and meets the acceptance criteria.
- 2. Linearity:
Not applicable.
- 3. Analytical Specificity/Interference:
Not applicable.
- 4. Assay Reportable Range:
Not applicable.
- 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
## Quality Control (QC) Testing:
The CLSI-recommended quality control (QC) isolates, Haemophilus influenzae (ATCC 49247), Staphylococcus aureus (ATCC 25923), and Streptococcus pneumoniae (ATCC 49619), were tested a sufficient number of times (i.e., a minimum of 20 replicates per lot per reader). One Comparator disk lot and two BD disc lots were used. Each test was visually read by three independent readers, resulting in 129 data points (69 and 60 data points for Lot A and B, respectively) and 69 Comparator disk data points. The performance is shown in Table 2 below.
| QC Organism | Zone Diameter<br>(mm) | BD Lot 11<br>(N) | BD Lot 21<br>(N) | Comparator<br>disk2<br>(N) |
|-------------------------------------------------------------------------|-----------------------|------------------|------------------|----------------------------|
| Haemophilus influenzae<br>ATCC 49247<br><br>Expected Range:<br>22-28 mm | 19 | | | |
| | 20 | | | |
| | 21 | | | |
| | 22 | | | 6 |
| | 23 | | | 14 |
| | 24 | 1 | | 21 |
| | 25 | 9 | 4 | 24 |
| | 26 | 33 | 30 | 4 |
| | 27 | 25 | 25 | |
| | 28 | 1 | 1 | |
| | 29 | | | |
| | 30 | | | |
| | 31 | | | |
{10}------------------------------------------------
| QC Organism | Zone Diameter<br>(mm) | BD Lot 1¹<br>(N) | BD Lot 2¹<br>(N) | Comparator<br>disk²<br>(N) |
|----------------------------|-----------------------|------------------|------------------|----------------------------|
| | 23 | | | |
| | 24 | | | |
| | 25 | | | 1 |
| | 26 | | | 3 |
| | 27 | 11 | 6 | 15 |
| Staphylococcus aureus ATCC | 28 | 17 | 16 | 31 |
| 25923 | 29 | 30 | 31 | 21 |
| Expected Range: | 30 | 13 | 7 | |
| 26-32 mm | 31 | | | |
| | 32 | | | |
| | 33 | | | |
| | 34 | | | |
| | 35 | | | |
| | 16 | | | |
| | 17 | | | |
| | 18 | | | |
| | 19 | | | |
| | 20 | | | |
| Streptococcus pneumoniae | 21 | | | 1 |
| ATCC 49619 | 22 | | | 7 |
| | 23 | 3 | | 26 |
| Expected Range: | 24 | 21 | 16 | 24 |
| 19-27 mm | 25 | 33 | 28 | 8 |
| | 26 | 10 | 16 | 3 |
| | 27 | 2 | | |
| | 28 | | | |
| | 29 | | | |
| | 30 | | | |
ATCC = American Type Culture Collection
'Two BD disc lots were tested (Lot A and Lot B).
Two BD disc lots were tested (Lot A and Lot B).
2One Comparator disk lot was tested.
The BD disc QC performance is > 95% and is acceptable.
## Inoculum Density Check:
Colony counts were conducted for all QC and reproducibility isolates, as well as 14.0% of clinical isolates. All were within the expected range.
- 6. Detection Limit:
Not applicable.
- 7. Assay Cut-Off:
{11}------------------------------------------------
Not applicable.
#### B Comparison Studies:
#### 1. Method Comparison with Predicate Device:
The BD BBL Sensi-Disc Lefamulin 20 ug (LMU-20) was compared with an FDA cleared disk of the same antimicrobial, mass/concentration, and content. The study was conducted at one testing site. Three independent operators participated in reading of test results with isolates evenly distributed to mimic testing at multiple sites. Testing was performed with one lot of discs from each test manufacturer (BD and comparator FDA-cleared disk), utilizing appropriate media following the method outlined in CLSI M02 and M100. An equivalent procedure to CLSI M11 was utilized for conducting colony counts.
## Clinical:
Clinical testing was performed at one U.S. site with both the BD BBL Sensi-Disc Lefamulin disc and the comparator FDA cleared disk using a total of 270 clinical isolates including 90 Haemophilus influenzae, 90 Staphylococcus aureus (methicillin-susceptible isolates), and 90 Streptococcus pneumoniae.
## Challenge:
Challenge testing was performed at one U.S site. A total of 66 challenge isolates were tested which included 22 Haemophilus influenzae, 22 Staphylococcus aureus (methicillinsusceptible isolates), and 22 Streptococcus pneumoniae.
Performance results for the total 336 clinical and challenge isolates are shown in Table 3. Performance of 39 MRSA isolates is not included in this table due to the lack of FDArecognized breakpoints for Lefamulin when testing MRSA.
| | Total | CA# | CA% | S (#) | NS (#) | VMJ* | MAJ | |
|------------------------|----------------------------------------------------------|-----|-------|-------|--------|------|-----|--|
| Haemophilus influenzae | | | | | | | | |
| Clinical | 90 | 89 | 98.9 | 90 | 0 | 0 | 1 | |
| Challenge | 22 | 19 | 86.4 | 13 | 9 | 3 | 0 | |
| Combined | 112 | 108 | 96.4 | 103 | 9 | 3 | 1 | |
| | Staphylococcus aureus (methicillin-susceptible isolates) | | | | | | | |
| Clinical | 90 | 90 | 100.0 | 90 | 0 | 0 | 0 | |
| Challenge | 22 | 22 | 100.0 | 10 | 12 | 0 | 0 | |
| Combined | 112 | 112 | 100.0 | 100 | 12 | 0 | 0 | |
| | Streptococcus pneumoniae | | | | | | | |
| Clinical | 90 | 89 | 98.9 | 89 | 1 | 1 | 0 | |
| Challenge | 22 | 22 | 100.0 | 19 | 3 | 0 | 0 | |
| Combined | 112 | 111 | 99.1 | 108 | 4 | 1 | 0 | |
| Overall | | | | | | | | |
| Clinical | 270 | 268 | 99.3 | 269 | 1 | 1 | 1 | |
| Challenge | 66 | 63 | 95.5 | 42 | 24 | 3 | 0 | |
| Combined | 336 | 331 | 98.5 | 311 | 25 | 4 | 1 | |
Table 3: Overall Performance of the BD BBL Sensi-Disc Lefamulin Disc vs. Comparator disk
* Due to the lack of an interpretive category other than susceptible and considering that a 3 mm disk zone diameter is equivalent to one MIC doubling dilution, because all potential VMJ errors had disk zone diameter measurement differences
{12}------------------------------------------------
| | Total | CA# | CA% | S (#) | NS (#) | VMJ* | MAJ |
|------------------------------------------------------------------------------------------------------------------|-------|-----|-----|-------|--------|------|-----|
| ≤3 mm compared to the FDA-cleared comparator disk and were ≤ an MIC doubling dilution-equivalent compared to the | | | | | | | |
historical MIC result, the adjusted potential VMJ error rate is 0%.
CA - Category Agreement S - Susceptible isolates
NS - Nonsusceptible isolates
MAJ - major errors VMJ - very major errors
Category Agreement (CA) is when the BD result interpretation agrees exactly with the FDA cleared comparator result interpretation. The overall performance of the BD BBL Sensi-Disc Lefamulin disc as compared to the Comparator disk for Haemophilus influenzae group (Table 3) was acceptable with 96.4% (108/112) CA. There were 3 very major errors and 1 major error.
The overall performance of the BD BBL Sensi-Disc Lefamulin disc as compared to the Comparator disk for Staphylococcus aureus (methicillin-susceptible isolates) group (Table 3) was acceptable with 100.0% (112/112) CA. There were no very major or major errors.
The overall performance of the BD BBL Sensi-Disc Lefamulin disc as compared to the Comparator disk for Streptococcus pneumoniae group (Table 3) was acceptable with 99.1% (111/112) CA. There was 1 very major error and no major errors.
As required under 511A(b)(2)(C)(ii)(1) of the Federal Food, Drug and Cosmetic Act, the following statement is included in the 'Warnings and Precautions' section in the device labeling to address testing of non-indicated species:
"Per the FDA-Recognized Susceptibility Test Interpretive Criteria website, the safety and efficacy of antimicrobial drugs, for which antimicrobial susceptibility is tested by this AST device, may or may not have been established in adequate and well-controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labeling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved."
## Resistant Isolates:
A total of 336 clinical and challenge isolates were tested when the BD BBL Sensi-Disc Lefamulin was compared to the FDA cleared comparator disk. However, an insufficient number of resistant isolates were available for testing. The following limitation was added in the device labeling:
"The current absence of resistant isolates precludes defining any results other than "Susceptible." Isolates yielding results other than "Susceptible" should be submitted to a reference laboratory for further testing."
- 2. Matrix Comparison
Not applicable.
#### C Clinical Studies:
- 1. Clinical Sensitivity:
{13}------------------------------------------------
Not applicable
- 2. Clinical Specificity:
Not applicable
- 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable
#### D Clinical Cut-Off:
Not applicable
#### E Expected Values/Reference Range:
The FDA-identified interpretive criteria for Lefamulin are listed in Table 4.
| | | Table 4: Interpretive Categories and Breakpoints for Lefamulin |
|--|--|----------------------------------------------------------------|
| | | |
| | Minimum Inhibitory<br>Concentration (µg/mL) | | | Disk Diffusion<br>(zone diameter in mm) | | |
|---------------------------------------------------------------------|---------------------------------------------|---|---|------------------------------------------|---|---|
| Pathogen | S | I | R | S | I | R |
| <i>Staphylococcus aureus</i> (methicillin-<br>susceptible isolates) | $ \u22640.25 $ | - | - | $ \u226523 $ | - | - |
| <i>Streptococcus pneumoniae</i> | $ \u22640.5 $ | - | - | $ \u226519 $ | - | - |
| <i>Haemophilus influenzae</i> | $ \u22642 $ | - | - | $ \u226518 $ | - | - |
1 According to the FDA STIC Website, which recognizes CLSI M100-Ed33.
Note: The current absence of resistant isolates precludes defining any results other than "Susceptible". Isolates yielding MIC results other than "Susceptible" should be submitted to a reference laboratory for further testing.
S = Susceptible
I = Intermediate
R = Resistant
#### VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
#### IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission included a breakpoint change protocol that was reviewed and accepted by FDA. This protocol addresses future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage (https://www.fda.gov/drugs/developmentresources/antibacterial-susceptibility-test-interpretive-criteria). The protocol outlined the specific procedures and acceptance criteria that Becton, Dickinson, and Company (BD Diagnostic Systems) intends to use to evaluate the BD BBL Sensi-Disc Lefamulin 20µg (LMU-20) when revised breakpoints for Lefamulin are published on the FDA STIC webpage. The breakpoint change protocol included with the submission indicated that if specific criteria are met, BD Diagnostic Systems will update the Lefamulin 20ug device label to include (1) the new breakpoints, (2) an updated performance section after re-evaluation of data in this premarket notification with the new breakpoints, and (3) any new limitations as determined by their evaluation.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.