LIAISON® XL MUREX Anti-HBs, LIAISON® XL MUREX Control Anti-HBs and LIAISON® XL MUREX Anti-HBs Verifiers
P180039 · DiaSorin, Inc. · SEI · Feb 21, 2020 · Microbiology
Device Facts
Record ID
P180039
Device Name
LIAISON® XL MUREX Anti-HBs, LIAISON® XL MUREX Control Anti-HBs and LIAISON® XL MUREX Anti-HBs Verifiers
Applicant
DiaSorin, Inc.
Product Code
SEI · Microbiology
Decision Date
Feb 21, 2020
Decision
APRL
Regulation
21 CFR 866.3173
Device Class
Class 2
Attributes
Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
P180039 · Feb 21, 2020
LIAISON® XL MUREX Anti-HBs, LIAISON® XL MUREX Control Anti-HBs and LIAISON® XL MUREX Anti-HBs Verifiers
DiaSorin, Inc.
Archived clinical samples from patients diagnosed with acute or chronic Hepatitis B
Retrospective clinical samples were used to evaluate the clinical performance (agreement) of the LIAISON XL MUREX Anti-HBs assay in comparison to FDA-approved reference assays.
Retrospective study; Archived samples; Clinical agreement; Hepatitis B
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Clinical Agreement Study (Retrospective Component); Retrospective clinical agreement study
Subjects diagnosed with acute and/or chronic Hepatitis B (Age 17-67); Sample Size: 256 retrospective samples; Number of Sites: 6 countries (Russia, Colombia, Cameroon, Ghana, Nigeria, and the United States)
Six FDA approved reference assays (HBsAg, HBeAg, Anti-HBs, Anti-HBc, Anti-HBc IgM, and Anti-HBe)
Positive and Negative Percent Agreement (PPA/NPA)
Indications for Use
The LIAISON XL MUREX Anti-HBs is an in vitro chemiluminescent immunoassay (CLIA) for the qualitative and quantitative determination of antibody to hepatitis B surface antigen (anti-HBs) in human adult and pediatric (2 – 21 years) serum and plasma (lithium and sodium heparin and K2 EDTA) including separator tubes, on the LIAISON XL Analyzer. Assay results in conjunction with other hepatitis B virus (HBV) serological markers and clinical information may be used as an aid in the diagnosis of HBV infection in patients with symptoms of hepatitis or who may be at risk for HBV infection. The assay results may be used as an aid in the determination of susceptibility to HBV infection in individuals prior to or following HBV vaccination or where vaccination status is unknown. The assay is not approved for use in screening blood, plasma or tissue donors.
Device Story
LIAISON XL MUREX Anti-HBs is an in vitro chemiluminescent immunoassay (CLIA) performed on the automated LIAISON XL Analyzer. It uses human serum or plasma samples. The assay principle is a direct sandwich chemiluminescence immunoassay: anti-HBs in the sample binds to HBsAg-coated magnetic particles and isoluminol-conjugated HBsAg. After washing, starter reagents induce a flash chemiluminescence reaction; a photomultiplier measures the light signal (RLU), which is proportional to the anti-HBs concentration. The analyzer performs all processing and result interpretation. Clinicians use the quantitative/qualitative output alongside other HBV serological markers and clinical history to diagnose HBV infection or assess immunity/susceptibility post-vaccination. The device benefits patients by enabling timely diagnosis, appropriate treatment, and vaccination management, thereby reducing HBV transmission and disease burden.
Clinical Evidence
Clinical agreement study tested 3,082 samples (prospective and retrospective) from 6 countries against 6 FDA-approved reference assays. Results showed 97.1% positive percent agreement (PPA) and 96.5% negative percent agreement (NPA). Subgroup PPA: 97.4% in vaccinated subjects, 99.0% in naturally immune subjects. NPA in non-infected subjects was 97.8%. Study included pediatric (2-21 years) and adult populations. No adverse events reported.
Technological Characteristics
Chemiluminescent immunoassay (CLIA) using magnetic particles coated with recombinant HBsAg (ad and ay subtypes) and isoluminol-conjugated HBsAg. Requires LIAISON XL Analyzer (fully automated). Sample types: serum, lithium/sodium heparin plasma, K2 EDTA plasma. Connectivity: standalone analyzer. Software: embedded firmware for processing and interpretation. Reagents include ProClin 300 preservative. Calibration stable for 6 weeks.
Indications for Use
Indicated for adult and pediatric (2-21 years) patients requiring qualitative or quantitative determination of anti-HBs antibodies. Used as an aid in diagnosing HBV infection in symptomatic patients or those at risk, and for determining susceptibility to HBV infection prior to or following vaccination or when vaccination status is unknown. Not for blood, plasma, or tissue donor screening.
Regulatory Classification
Identification
A hepatitis B virus (HBV) antibody assay is identified as an in vitro diagnostic device intended for prescription use in the detection of antibodies to HBV in human serum, plasma, or other matrices, and as a device that aids in the diagnosis of HBV infection in persons with signs and symptoms of hepatitis and in persons at risk for hepatitis B infection. Results from assays may be qualitative or quantitative, such as quantitative anti-HBs. In addition, results from an anti-HBc IgM (IgM antibodies to core antigen) assay indicating the presence of anti-HBc IgM are indicative of recent HBV infection. Anti-HBs (antibodies to surface antigen) assay results may be used as an aid in the determination of susceptibility to HBV infection in individuals prior to or following HBV vaccination or when vaccination status is unknown. The assay is not intended for screening of blood, plasma, cells, or tissue donors. The assay is intended as an aid in diagnosis in conjunction with clinical findings and other diagnostic procedures.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The labeling required under § 809.10(b) of this chapter must include:
(i) A prominent statement that the assay is not intended for the screening of blood, plasma, cells, or tissue donors.
(ii) A detailed explanation of the principles of operation and procedures for performing the assay.
(iii) A detailed explanation of the interpretation of results.
(iv) Limitations, which must be updated to reflect current clinical practice and disease presentation and management. The limitations must include statements that indicate:
(A) When appropriate, performance characteristics of the assay have not been established in populations of immunocompromised or immunosuppressed patients or other special populations where assay performance may be affected.
(B) Detection of HBV antibodies to a single viral antigen indicates a present or past infection with hepatitis B virus, but does not differentiate between acute, chronic, or resolved infection.
(C) The specimen types for which the device has been cleared, and that use of the assay with specimen types other than those specifically cleared for this device may result in inaccurate assay results.
(D) Diagnosis of hepatitis B infection should not be established on the basis of a single assay result but should be determined by a licensed healthcare professional in conjunction with the clinical presentation, history, and other diagnostic procedures.
(E) A non-reactive assay result may occur early during acute infection, prior to development of a host antibody response to infection, or when analyte levels are below the limit of detection of the assay.
(F) Results obtained with this assay may not be used interchangeably with results obtained with a different manufacturer's assay.
(v) For devices intended for the quantitative detection of HBV antibodies (anti-HBs), in addition to the special controls listed in paragraphs (b)(1) and (2) of this section, labeling required under § 809.10(b) of this chapter must include:
(A) The assay calibrators' traceability to a standardized reference material that FDA has determined is appropriate (
*e.g.,* a recognized consensus standard) and the limit of blank (LoB), limit of detection (LoD), limit of quantitation (LoQ), linearity, and precision to define the analytical measuring interval.(B) Performance results of the analytical sensitivity study testing a standardized reference material that FDA has determined is appropriate (
*e.g.,* a recognized consensus standard).(2) Design verification and validation must include the following:
(i) Detailed device description, including all parts that make up the device, ancillary reagents required but not provided, an explanation of the device methodology, and design of the antigen(s) and capture antibody(ies) sequences, rationale for the selected epitope(s), degree of amino acid sequence conservation of the target, and the design and composition of all primary, secondary and subsequent standards used for calibration.
(ii) Documentation and characterization (
*e.g.,* supplier, determination of identity, and stability) of all critical reagents (including description of the antigen(s) and capture antibody(ies)), and protocols for maintaining product integrity throughout its labeled shelf life.(iii) Risk analysis and management strategies, such as Failure Modes Effects Analysis and/or Hazard Analysis and Critical Control Points summaries and their impact on assay performance.
(iv) Final release criteria to be used for manufactured assay lots with appropriate evidence that lots released at the extremes of the specifications will meet the identified analytical and clinical performance characteristics as well as stability.
(v) Stability studies for reagents must include documentation of an assessment of real-time stability for multiple reagent lots using the indicated specimen types and must use acceptance criteria that ensure that analytical and clinical performance characteristics are met when stability is assigned based on the extremes of the acceptance range.
(vi) All stability protocols, including acceptance criteria.
(vii) When applicable, analytical sensitivity of the assay that is the same or better than that of other cleared or approved assays.
(viii) Analytical performance studies and results for determining the limit of blank (LoB), limit of detection (LoD), cutoff, precision (reproducibility), including lot-to-lot and/or instrument-to-instrument precision, interference, cross reactivity, carryover, hook effect, seroconversion panel testing, matrix equivalency, specimen stability, reagent stability, and cross-genotype antibody detection sensitivity, when appropriate.
(ix) For devices intended for the detection of antibodies for which a standardized reference material (that FDA has determined is appropriate) is available, the analytical sensitivity study and results testing the standardized reference material. Detailed documentation of that study and its results must be provided, including the study protocol, study report, testing results, and all statistical analyses.
(x) For devices with associated software or instrumentation, documentation must include a detailed description of device software, including software applications and hardware-based devices that incorporate software. The detailed description must include documentation of verification, validation, and hazard analysis and risk assessment activities, including an assessment of the impact of threats and vulnerabilities on device functionality and end users/patients as part of cybersecurity review.
(xi) Detailed documentation of clinical performance testing from a clinical study with an appropriate number of HBV reactive and non-reactive samples in applicable risk categories and conducted in the appropriate settings by the intended users. Performance must be analyzed relative to an FDA cleared or approved HBV antibody assay or a comparator that FDA has determined is appropriate. Additional relevant patient groups must be validated as appropriate. The samples must include prospective (sequential) samples for each identified specimen type and, as appropriate, additional characterized clinical samples. Samples must be sourced from geographically diverse areas.
(3) For any HBV antibody assay intended for quantitative detection of anti-HBV antibodies, the following special controls, in addition to those special controls listed in paragraphs (b)(1) and (2) of this section, also apply:
(i) Detailed documentation of the metrological calibration traceability hierarchy to a standardized reference material that FDA has determined is appropriate.
(ii) Detailed documentation of the following analytical performance studies conducted, as appropriate to the technology, specimen types tested, and intended use of the device, including upper and lower limits of quantitation (UloQ and LloQ, respectively), linearity using clinical samples, and an accuracy study using the recognized international standard material.
Submission Summary (Full Text)
{0}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
I. GENERAL INFORMATION
Device Generic Name: Antibody to Hepatitis B Surface Antigen (Anti-HBs assay)
Device Trade Name: LIAISON® XL MUREX Anti-HBs
LIAISON® XL MUREX Control Anti-HBs
LIAISON® XL MUREX Anti-HBs Verifiers
Device Procode: LOM
Applicant’s Name and Address: DiaSorin Inc.
1951 Northwestern Avenue
Stillwater, MN 55082
Date(s) of Panel Recommendation: None
Premarket Approval Application (PMA) Number: P180039
Date of FDA Notice of Approval: February 21, 2020
II. INDICATIONS FOR USE
The LIAISON XL MUREX Anti-HBs is an in vitro chemiluminescent immunoassay (CLIA) for the qualitative and quantitative determination of antibody to hepatitis B surface antigen (anti-HBs) in human adult and pediatric (2 – 21 years) serum and plasma (lithium and sodium heparin and K₂ EDTA) including separator tubes, on the LIAISON XL Analyzer. Assay results in conjunction with other hepatitis B virus (HBV) serological markers and clinical information may be used as an aid in the diagnosis of HBV infection in patients with symptoms of hepatitis or who may be at risk for HBV infection. The assay results may be used as an aid in the determination of susceptibility to HBV infection in individuals prior to or following HBV vaccination or where vaccination status is unknown.
The assay is not approved for use in screening blood, plasma or tissue donors.
The LIAISON XL MUREX Control Anti-HBs (negative and positive) is intended for use as assayed quality control samples to monitor the performance of the LIAISON XL MUREX Anti-HBs assay. The performance characteristics of LIAISON XL MUREX Control Anti-HBs have not been established for any other assays or instrument platforms.
The LIAISON XL MUREX Anti-HBs Verifiers (level 1, 2, 3, and level 4) are assayed quality control materials intended for the quantitative verification of calibration and reportable range of the LIAISON XL MUREX Anti-HBs assay. The performance
PMA P180039: FDA Summary of Safety and Effectiveness Data
{1}
characteristics of LIAISON XL MUREX Anti-HBs Verifiers have not been established in connection with any other assay or instrument platforms.
## III. CONTRAINDICATIONS
There are no known contraindications.
## IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the LIAISON XL MUREX Anti-HBs labeling.
## V. DEVICE DESCRIPTION
### Kit Components
**Reagents:** The LIAISON XL MUREX Anti-HBs is an *in vitro* diagnostic device consisting of four (4) reagents provided in individual compartments within a plastic container called the Reagent Integral. The assay configuration allows performance of 100 tests.
The assay is comprised of the following components:
- 1 vial of Magnetic particles coated with HBsAg obtained in mammalian cells by the recombinant DNA technology (balanced ad and ay subtypes), BSA, Phosphate buffer (PBS), < 0.1% sodium azide.
- 1 vial Calibrator 1 containing Human plasma containing low levels of anti-HBs fetal calf serum, EDTA and 0.2% ProClin® 300, preservatives.
- 1 vial Calibrator 2 containing human plasma with high levels of anti-HBs fetal calf serum, EDTA. 0.2% ProClin® 300, preservatives, and an inert blue dye.
- 1 vial Conjugate containing Heat-treated Human HBsAg (balanced ad and ay subtypes) conjugated to an isoluminol derivative, BSA, phosphate buffer (PBS), EDTA, 0.2% ProClin® 300, preservatives, and an inert blue dye.
**Controls:** LIAISON XL MUREX Control Anti-HBs set consists of 2 levels (positive and negative) ready to use controls. Each control solution allows at least 20 tests to be performed. The control set is an additional material required to perform the test.
The controls are comprised of the following components:
- 2 vials of Negative control containing Human serum without anti-HBs antibodies with 0.2% ProClin® 300 and preservatives
- 2 vials of Positive control containing Human plasma with anti-HBs antibodies, 0.2% ProClin® 300 and preservatives
**Verifiers:** LIAISON XL MUREX Anti-HBs Verifier set consists of 4 levels (A-D) ready to use verifiers. Each verifier solution allows at least 20 tests to be performed. The Verifier set is an additional material required to perform the test.
PMA P180039: FDA Summary of Safety and Effectiveness Data
{2}
The Verifiers are comprised of the following components:
- 1 vial per level A to D, containing Human plasma with 4 different levels of anti-HBs, $0.2\mathrm{ProClin}^{\text{®}}$ and preservatives.
In addition, the following Analyzer and accessories are required for performing the LIAISON XL MUREX Anti-HBs, LIAISON XL MUREX Control Anti-HBs and the LIAISON XL MUREX Anti-HBs Verifiers:
- LIAISON XL Analyzer is a fully automated chemiluminescent analyzer, performing the complete sample processing steps of the chemiluminescent assay and interprets the results.
- LIAISON Wash/System Liquid (10x) - phosphate buffer solution, $< 0.1\%$ sodium azide.
- LIAISON XL Starter Kit - catalyst in $4\%$ sodium hydroxide solution and $0.12\%$ hydrogen peroxide solution
# Assay Principle
The qualitative and quantitative determination of anti-HBs is a direct, sandwich chemiluminescence immunoassay (CLIA). Anti-HBs present in samples, calibrators, or controls binds to the solid phase and HBsAg conjugate, thus forming a sandwich. After the incubation, the unbound material is removed with a wash cycle. Subsequently, the starter reagents are added and a flash chemiluminescence reaction is thus induced. The light signal, and hence the amount of isoluminol-HBsAg conjugate, is measured by a photomultiplier as relative light units (RLU) and is indicative of anti-HBs concentration present in the sample.
Table 1: Initial Interpretation of Results for the LIAISON XL MUREX Anti-HBs is as follows:
| Initial Result LIAISON XL Anti-HBs assay | | |
| --- | --- | --- |
| mIU/mL | Results | Retest Procedure |
| < 9.0 | Negative | No retest is required. |
| 9.0 < x < 11.0 | Equivocal | Retest in duplicate with the LIAISON XL Anti-HBs assay. |
| ≥ 11.0 | Positive | No retest is required. |
PMA P180039: FDA Summary of Safety and Effectiveness Data
{3}
Table 2: Final Interpretation of Results for the LIAISON XL MUREX Anti-HBs is as follows:
| Final Result
LIAISON XL Anti-HBs assay | | |
| --- | --- | --- |
| Result After Retest | Final Result | Clinical Interpretation of the Final Results |
| Both retest results are <10 IU/mL | Negative | Anti-HBs concentration detected at < 10 mIU/mL.
Individual is considered to be not immune to infection with HBV. |
| 2 results out of 3
<10 IU/mL | Negative | Anti-HBs concentration detected at < 10 mIU/mL.
Individual is considered to be not immune to infection with HBV. |
| 2 results out of 3
≥10 IU/mL | Positive | Anti-HBs concentration detected at ≥ 10 mIU/mL.
Individual is considered to be immune to infection with HBV. |
| Both retest results are ≥10 IU/mL | Positive | Anti-HBs concentration detected at ≥ 10 mIU/mL.
Individual is considered to be immune to infection with HBV. |
VI. ALTERNATIVE PRACTICES AND PROCEDURES
There are several other alternatives for the detection and quantitation of antibodies to hepatitis B surface antigen (Anti-HBs). There are currently several FDA approved in vitro diagnostic tests commercially available for serological markers of hepatitis B virus (HBV) infection which, when used in conjunction with a patient's medical history, clinical examination and other laboratory findings, may be used as an aid in the diagnosis of HBV infection in patients with symptoms of hepatitis or who may be at risk for HBV infection. The assay may be used as an aid in the determination of susceptibility to hepatitis B virus (HBV) infection in individuals prior to or following HBV vaccination or where vaccination status is unknown. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle.
VII. MARKETING HISTORY
The LIAISON XL MUREX Anti-HBs assay (318220) and LIAISON XL MUREX Control Anti-HBs (318221) are essentially the same as the CE-marked LIAISON XL murex Anti-HBs assay (311220) and LIAISON XL murex Control Anti-HBs (311221) with some minor modifications to raw material manufacturing processes.
The LIAISON XL MUREX Anti-HBs assay (318220), LIAISON XL MUREX Control Anti-HBs (318221) and the LIAISON XL MUREX Anti-HBs Verifiers (318223) have not been marketed in the U.S. or any foreign country.
The CE-marked LIAISON XL murex Anti-HBs assay (311220) and LIAISON XL murex Control Anti-HBs (311221) have been marketed in multiple countries. These devices
PMA P180039: FDA Summary of Safety and Effectiveness Data
{4}
have not been withdrawn from the market in any country for reasons relating to safety and effectiveness.
The following table includes a list all countries where the CE-marked versions have been marketed in the past year.
Table 3: Countries where CE Marked LIAISON XL murex Anti-HBs (311220) and LIAISON XL murex Control Anti-HBs (311221) is sold
| Austria | Australia | Bangladesh |
| --- | --- | --- |
| Belgium | Bulgaria | Brunei |
| Brazil | Bahamas | Switzerland |
| Chile | Colombia | Cyprus |
| Czech Republic | Germany | Denmark |
| Dominican Republic | Egypt | Spain |
| Finland | France | United Kingdom |
| Greece | Croatia | Hungary |
| Ireland | Israel | Iraq |
| Iran | Italy | Kuwait |
| Lebanon | Luxembourg | Morocco |
| Netherlands | Norway | Perù |
| Poland | Portugal | Paraguay |
| Romania | Saudi Arabia | Slovenia |
| Thailand | Tunisia | Turkey |
| South Africa | | |
## VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
Below is a list of the potential adverse effects (e.g., complications) associated with the use of the device. The LIAISON XL MUREX Anti-HBs is intended for *in vitro* diagnostic use, and as a result, there is no direct adverse effect on the patient. Standard good laboratory practices are considered sufficient to minimize risks to the end user.
Failure of the product to perform as intended or human error in the use of the test may lead to a false result. Appropriate Warnings and Precautions for identified risks are contained in the labeling and assay Instructions for Use.
A false positive result in a diagnostic setting includes improper patient management, including treatment for hepatitis B with antiviral medication. Antiviral medical has risks including toxicity and more rarely allergic reactions. Over time, viral resistance in patients who are co-infected but undiagnosed with other viruses using the same antiviral medication, such as HIV, can lead to viral resistance; however, the chance of an undiagnosed co-infection in a patient tested for hepatitis B is exceedingly unlikely.
PMA P180039: FDA Summary of Safety and Effectiveness Data
Page 5
{5}
Furthermore, a false positive result when the device is used as an aid in the determination of susceptibility to HBV infection in individuals prior to or following HBV vaccination or where vaccination status is unknown includes leading a provider to falsely believe that a patient has been vaccinated in the past and/or has current immunity when the patient does not. This could lead to a patient inappropriately missing a vaccination.
Additionally and likely more importantly, because hepatitis B surface antibody is ordered as part of a panel in clinical practice, this risk will likely be mitigated as incongruous test results would lead a clinician to either retest the patient or further investigate the etiology of hepatitis. As the performance of the assay in the clinical trial suggests that false positive results will be uncommon, the true risk of a false negative is likely minimal.
A false negative result when the device is used as an aid in the determination of susceptibility to HBV infection in individuals prior to or following HBV vaccination or where vaccination status is unknown include potentially missing and undertreating a patient who has hepatitis B infection and whose clinical picture warrants antiviral treatment. Again, because hepatitis B surface antibody is ordered as part of a panel in clinical practice, the risk of missing and undertreating a patient who has hepatitis B infection will likely be mitigated as incongruous test results would lead a clinician to either retest the patient or further investigate the etiology of hepatitis. As the performance of the assay in the clinical trial suggests that false negative results will be uncommon, the true risk of a false negative is likely minimal.
For the specific adverse events that occurred in the clinical study, please see Section X below.
## IX. SUMMARY OF NONCLINICAL STUDIES
### A. Laboratory Studies
#### 1. Cut-off Determination
The cut-off was established internally at DiaSorin and verified by testing a total of 113 samples (55 known negative and 58 known positive). A Receiver Operating Characteristics (ROC) analysis was performed on the results of the specimens tested. The assay's cut-off was evaluated with the observed results to demonstrate that its selection represents the best level of specificity, without compromising the sensitivity.
The cut-off value at $10.0\mathrm{mIU/mL}$ is within the optimal range determined by the ROC curve to discriminate between negative and positive results.
The established cut-off Index is $10.0\mathrm{mIU/mL}$, with an equivocal range of $\pm 10\%$ (9.00 - 11.0 mIU/mL).
PMA P180039: FDA Summary of Safety and Effectiveness Data
Page 6
{6}
2. Sensitivity / Seroconversion Panels
The seroconversion sensitivity of the LIAISON XL MUREX Anti-HBs assay has been demonstrated by testing 10 commercial seroconversion panels in comparison to a reference anti-HBs immunoassay in terms of number of days from initial draw to first positive sample, as well as the difference between the last negative results and the first positive results.
The LIAISON XL MUREX Anti-HBs assay yielded a positive result sooner by one blood draw or more than the comparator assay in the 10 panels.
3. Analytical Sensitivity / Dilution Study with Standard
The sensitivity of the LIAISON XL MUREX Anti-HBs assay was evaluated by preparing serial dilutions of the WHO Second International Standard for anti-hepatitis B surface antigen (anti-HBs) immunoglobulin, human NIBSC code 07/164 in serum and plasma negative anti-HBs.
Dilutions were tested in five (5) replicates on one (1) kit lot and on one (1) LIAISON XL Analyzer. The LIAISON XL MUREX Anti-HBs assay demonstrated linearity and accuracy between 3 and 500 mIU/ml.
4. Endogenous Interference
The LIAISON XL MUREX Anti-HBs assay was tested for potential interference of high levels of endogenous substances including Triglycerides (3000 mg/dL), Hemoglobin (1000 mg/dL), Bilirubin (conjugated and unconjugated 20 mg/dL), Albumin (6000 mg/dL) and Cholesterol (350 mg/dL). Ten (10) negative samples or negative pools were spiked with an IgG anti-HBs high positive sample to achieve high negative, low positive and moderate samples. Samples were tested in 26 replicates each on one (1) kit lot. No interference was observed at the levels tested.
5. Analytical Specificity (Cross-Reactivity)
A study was conducted to evaluate the LIAISON XL MUREX Anti-HBs assay for cross-reactivity with specimens from individuals with medical conditions unrelated to HBV infection. A total of 301 samples from 28 unrelated medical conditions were tested in singlicate on one (1) kit lot of LIAISON XL MUREX Anti-HBs and on a reference Anti-HBs assay.
Of the 301 samples, no evidence of cross-reactivity was observed. The results of each potential cross-reactant are shown in in table below.
PMA P180039: FDA Summary of Safety and Effectiveness Data
Page 7
{7}
Table 4: Summary of Cross-Reactivity Testing
| | | N° of observed negative results | |
| --- | --- | --- | --- |
| Potential cross reactant | N° of expected negative samples | LIAISON XL MUREX Anti HBs | Reference Method |
| Anti-nuclear antibodies (ANA) | 10 | 10 | 10 |
| Auto-immune hepatitis | 10 | 10 | 10 |
| C. trachomatis | 11 | 11 | 11 |
| CMV (anti-CMV IgG or IgM positive) | 11 | 11 | 11 |
| EBV (IgM positive) | 11 | 11 | 11 |
| Fatty liver disease | 11 | 11 | 11 |
| HAMA | 11 | 11 | 11 |
| Hemodialysis patient | 11 | 11 | 11 |
| Hepatitis A Virus (IgM positive) | 11 | 11 | 11 |
| Hepatitis C Virus (anti-HCV ) | 11 | 11 | 11 |
| Hepatocellular carcinoma | 11 | 11 | 11 |
| HIV-1 (anti-HIV-1 positive) | 11 | 11 | 11 |
| HIV-2 (anti-HIV-2 positive) | 11 | 11 | 11 |
| HSV (anti-HSV IgG or IgM positive) | 11 | 11 | 11 |
| HTLV-1/2 (anti-HTLV positive) | 11 | 11 | 11 |
| IgG monoclonal gammopathy | 11 | 11 | 11 |
| IgM monoclonal gammopathy | 10 | 10 | 10 |
| Influenza vaccine recipients | 11 | 11 | 11 |
| Multiparous pregnancies | 10 | 10 | 10 |
| Multiple myeloma | 11 | 11 | 11 |
| Multiple transfusion recipients | 10 | 10 | 10 |
| N. gonorrhoea | 11 | 11 | 11 |
| Pregnancy 1sttrimester | 10 | 10 | 10 |
| Pregnancy 2ndtrimester | 10 | 10 | 10 |
| Pregnancy 3rdtrimester | 11 | 11 | 11 |
| Rheumatoid Factor | 11 | 11 | 11 |
| T. pallidum (anti-T. pallidum IgG and/or IgM positive) | 11 | 11 | 11 |
| T.cruzi (anti-T. cruzi positive) | 11 | 11 | 11 |
# 6. Sample Equivalence/Matrix Effect
Twenty-five (25) paired sets of matched serum (with and without gel SST) and plasma (lithium and sodium heparin and $\mathrm{K}_2\mathrm{EDTA}$ ) were tested to determine if these sample types provide equivalent results on the LIAISON XL MUREX Anti-HBs assay. Each sample was divided into three aliquots. Two sets of aliquots were spiked with an anti-HBs high positive sample to achieve two (2) levels of samples: high negative and low positive samples. The third set of aliquots was un-spiked to serve as control samples. The results of the negative and low positive samples did not change the classification of the expected result.
PMA P180039: FDA Summary of Safety and Effectiveness Data
{8}
The results obtained on the serum-plasma paired samples indicate that there is equivalence among serum (with and without Gel SST), $\mathrm{K}_2$ -EDTA, Lithium Heparin and Sodium Heparin plasma, as indicate in the table below:
| Sample Type | Slope | Intercept |
| --- | --- | --- |
| Serum with Gel separator | 0.987 | 0.138 |
| K2-EDTA | 0.971 | -0.036 |
| Lithium Heparin | 1.023 | -0.150 |
| Sodium Heparin | 0.998 | -0.099 |
# 7. Carry-Over Study
The LIAISON XL Analyzer uses disposable tips for sample pipetting. A carry-over study was performed to evaluate the extent of carryover and the associated residual risk for signal carryover in the instrument's measuring cell as the result of a high signal-generating sample.
Two samples were used for this evaluation: one (1) anti-HBs negative human serum sample and one (1) human positive sample/pool with a high level of anti-HBs analyte. The samples were tested in singlicate in five (5) runs in the following sequence: High Pos, Neg, High Pos, Neg, High Pos, Neg, High Pos, Neg, High Pos, Neg.
All acceptance criteria were met, demonstrating that no significant amount of analyte is carried over from one sample reaction into the subsequent sample reactions.
# 8. Stability Studies
# Sample Stability
Studies were performed to determine the storage stability of patient serum and plasma samples at storage temperatures of $2 - 8^{\circ}\mathrm{C}$ , room temperature (RT), $-20^{\circ}\mathrm{C}$ . A multiple freeze/thaw (F/T) study was also performed.
Serum and plasma samples tested contained Anti-HBs analyte levels of negative, high negative and low positive.
$2 - 8^{\circ}\mathrm{C}$ study - samples were tested unstressed $(\mathrm{T} = 0)$ , and again after 1, 2, 3, 4, 5, 7 and 8 days of storage at $2 - 8^{\circ}\mathrm{C}$ for 24 hours per day.
> room temperature study (RT) - samples were tested immediately after preparation and again after 1, 2, 3, 4, and 5 days of storage at RT for 24 hours each day.
- $20^{\circ}\mathrm{C}$ study - samples were tested unstressed $(\mathrm{T} = 0)$ and stored at $-20^{\circ}\mathrm{C}$ or lower for 1, 3, and 4 months.
> Freeze/Thaw (F/T) study - samples were tested unstressed $(\mathrm{T} = 0)$ and after 1, 2, 3, 4, 5, 6, and 7 F/T cycles. Samples were frozen for 12-24 hours at $-20^{\circ}\mathrm{C}$ or lower and thawed at room temperature.
PMA P180039: FDA Summary of Safety and Effectiveness Data
{9}
The results of the studies indicate serum and plasma samples are stable for:
| | Sample Stability Claims | | | |
| --- | --- | --- | --- | --- |
| Sample Matrix | Number of Freeze and Thaw Cycles | Storage at 2-8°C | Storage at -20°C | Storage at Room Temperature |
| Serum | 6 | 7 days | 3 months | 1 day |
| Plasma | 1 | 7 days | 3 months | 1 day |
## Reagent Stability
### Real-Time (Shelf-Life)
Studies were performed to establish the shelf-life for the LIAISON XL MUREX Anti-HBs assay. Three (3) lots of LIAISON XL MUREX Anti-HBs were stored at the recommended storage temperature of 2-8°C throughout the study. Performance was assessed against clinically relevant acceptance criteria using three (3) lots LIAISON XL MUREX Control Anti-HBs (positive and negative) and an internal stability panel consisting of eight (8) samples. Study results demonstrate that reagents are stable and continue to meet acceptance criteria ten (10) months after the date of manufacture for the LIAISON XL MUREX Anti-HBs.
### Reagent On-Board
Stability studies were conducted to determine the length of time the LIAISON XL MUREX Anti-HBs Reagent Integral can be stored on-board the LIAISON XL Analyzer in the refrigerated area once opened.
One (1) lot of the LIAISON XL MUREX Anti-HBs assay was stored on-board the LIAISON XL Analyzer throughout the 13 weeks of the study. The LIAISON XL MUREX Control Anti-HBs (negative and positive) along with the internal stability panel were tested in duplicate at one (1) week intervals up to the 13 weeks.
The LIAISON XL MUREX Anti-HBs assay is stable on-board the LIAISON XL Analyzer for 6 weeks.
### Reagent Open Use
The aim of this study was to assess the open use stability of the LIAISON XL MUREX Anti-HBs kit reagents by simulating normal conditions of use as specified in the instruction for use.
Testing of samples was performed in duplicate, on one (1) lot of LIAISON XL MUREX Anti-HBs and one (1) lot of LIAISON XL MUREX Control Anti-HBs. Results were calculated using the initial (time zero) assay calibration.
PMA P180039: FDA Summary of Safety and Effectiveness Data
Page 10
{10}
The opened Reagent Integral was then removed from the XL Analyzer and stored at 2-8 °C. Kit performance using the opened Reagent Integral was evaluated weekly up to 13 weeks.
The Reagent Integral is stable after opening for 6 weeks when stored at 2-8 °C.
## Temperature Stress/Reagent Transport Study
The transport simulation tests were performed in order to verify that kit reagents maintain their properties during the shipment and delivery conditions to the customer. After being subjected to simulated stress conditions, testing was performed on 1 lot of LIAISON XL MUREX Anti-HBs at midlife and will be tested at expiration.
All testing performed meets acceptance criteria under various simulated transport conditions.
## Calibrator Stability
The LIAISON XL MUREX Anti-HBs calibrators are included on the Reagent Integral. All studies for the Reagent Integral are applicable to the 2 levels of calibrators provided.
## Calibration Interval Stability
The aim of this study was to assess stability of the product calibration by simulating normal condition of use as specified in the instruction for use.
A calibration was performed at time 0 and the Reagent Integral was stored on board the analyzer for the duration of the study. Kit performance was evaluated weekly up to nine (9) weeks by testing the stability panel and LIAISON XL MUREX Control Anti-HBs on one (1) lot of LIAISON XL MUREX Anti-HBs reagents. Results were generated using the initial (time zero) assay calibration and performance was assessed against clinically relevant acceptance criteria.
Results demonstrate that the LIAISON XL MUREX Anti-HBs calibration is stable for six (6) weeks.
## Control stability
### Real-time (Shelf-Life)
Studies were performed to establish the shelf-life for the LIAISON XL MUREX Control Anti-HBs and the LIAISON XL MUREX Anti-HBs Verifiers. Three lots of LIAISON XL MUREX Control Anti-HBs were stored at the recommended storage temperature of 2-8°C throughout the study.
PMA P180039: FDA Summary of Safety and Effectiveness Data
Page 11
{11}
Current results demonstrate that the positive and negative controls are stable and continue to meet acceptance criteria at eighteen (18) months.
## Open use
The aim of this study was to assess stability of the opened Control vials by simulating normal conditions of use, as specified in the instruction for use. Testing was performed in duplicate, on one lot of LIAISON XL MUREX Control Anti-HBs.
LIAISON XL MUREX Control Anti-HBs (negative and positive) were within their established ranges and were assessed against clinically relevant acceptance criteria.
The LIAISON XL MUREX Control Anti-HBs (negative and positive) are stable for 6 weeks after opening when stored at 2-8 °C between uses.
## Temperature Stress/Reagent Transport Study
The transport simulation tests were performed in order to ensure that kit Controls maintain their properties during the shipment and delivery conditions to the customer. After being subjected to simulated stress conditions testing was performed on 1 lot of LIAISON XL MUREX Control Anti-HBs.
All testing performed meets acceptance criteria under various simulated transport conditions.
## 9. Precision
### Internal 20 Days
A precision/reproducibility study was carried out over a period of twenty (20) days on the LIAISON XL MUREX Anti-HBs assay using the LIAISON XL Analyzer. The CLSI document EP05-A3 was consulted in the preparation of the testing protocol. The testing was performed internally at DiaSorin S.p.A.
A coded panel of eight (8) serum-based samples consisting of 1 Negative, 1 High Negative, 1 Low Positive, 1 Moderate Positive and 4 Positive along with 3 lots of the LIAISON XL MUREX Control Anti-HBs (neg and pos) and 3 lots of the LIAISON XL MUREX Anti-HBs Verifiers were tested in 2 replicates per run, 2 runs per day for 20 days using 3 different LIAISON XL MUREX Anti-HBs assay reagents and spanning at least 2 calibration cycles.
The Repeatability of the combined 3 lots of LIAISON XL MUREX Anti-HBs ranged from 1.3% to 3.1%. The with-in Laboratory %CV of the combined 3 lots of LIAISON XL MUREX Anti-HBsc ranged from 4.2% to 15.1%. The results are shown below.
PMA P180039: FDA Summary of Safety and Effectiveness Data
Page 12
{12}
Table 5: Summary of Precision Study
| LIAISON XL MUREX Anti-HBs Assay All 3 Lots Combined | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample ID | N | Mean | Repeatability | | Between-Run | | Between-Day | | Between-Lot | | Within Laboratory |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD |
| Ctrl Neg #RS-764 | 240 | *977 | 32.198 | 3.3% | 7.269 | 0.7% | 24.653 | 2.5% | 20.068 | 2.1% | 45.826 |
| Ctrl Neg #RS-765 | 240 | *940 | 27.365 | 2.9% | 13.132 | 1.4% | 18.266 | 1.9% | 17.476 | 1.9% | 39.501 |
| Ctrl Neg #RS-766 | 240 | *941 | 25.578 | 2.7% | 17.607 | 1.9% | 19.764 | 2.1% | 20.353 | 2.2% | 42.061 |
| Ctrl Pos #RS-767 | 240 | 51.9 | 0.825 | 1.6% | 0.905 | 1.7% | 0.694 | 1.3% | 5.428 | 10.5% | 5.608 |
| Ctrl Pos #RS-768 | 240 | 56.4 | 0.908 | 1.6% | 1.012 | 1.8% | 0.641 | 1.1% | 4.215 | 7.5% | 4.475 |
| Ctrl Pos #RS-769 | 240 | 53.7 | 0.890 | 1.7% | 1.169 | 2.2% | 0.578 | 1.1% | 3.967 | 7.4% | 4.269 |
| Cal Ver #RS-784 A | 240 | 7.56 | 0.236 | 3.1% | 0.154 | 2.0% | 0.181 | 2.4% | 0.95 | 12.6% | 1.007 |
| Cal Ver #RS-784 B | 240 | 20.1 | 0.396 | 2.0% | 0.248 | 1.2% | 0.364 | 1.8% | 2.47 | 12.3% | 2.54 |
| Cal Ver #RS-784 C | 240 | 72.6 | 1.022 | 1.4% | 0.681 | 0.9% | 0.857 | 1.2% | 6.809 | 9.4% | 6.972 |
| Cal Ver #RS-784 D | 240 | 237 | 2.973 | 1.3% | 3.572 | 1.5% | 2.681 | 1.1% | 27.599 | 11.6% | 28.116 |
| Cal Ver #RS-785 A | 240 | 6.94 | 0.214 | 3.1% | 0.103 | 1.5% | 0.146 | 2.1% | 0.658 | 9.5% | 0.715 |
| Cal Ver #RS-785 B | 240 | 18.3 | 0.412 | 2.3% | 0.339 | 1.9% | 0.375 | 2.1% | 1.662 | 9.1% | 1.786 |
| Cal Ver #RS-785 C | 240 | 75.1 | 0.870 | 1.2% | 1.042 | 1.4% | 0.885 | 1.2% | 4.376 | 5.8% | 4.666 |
| Cal Ver #RS-785 D | 240 | 238 | 3.747 | 1.6% | 2.507 | 1.1% | 3.324 | 1.4% | 19.32 | 8.1% | 20.116 |
| Cal Ver #RS-786 A | 240 | 7.15 | 0.223 | 3.1% | 0.080 | 1.1% | 0.203 | 2.8% | 0.7 | 9.8% | 0.767 |
| Cal Ver #RS-786 B | 240 | 19.4 | 0.371 | 1.9% | 0.371 | 1.9% | 0.289 | 1.5% | 1.736 | 8.9% | 1.836 |
| Cal Ver #RS-786 C | 240 | 77.3 | 1.038 | 1.3% | 0.935 | 1.2% | 1.122 | 1.5% | 5.436 | 7.0% | 5.723 |
| Cal Ver #RS-786 D | 240 | 251 | 4.106 | 1.6% | 2.782 | 1.1% | 3.584 | 1.4% | 23.625 | 9.4% | 24.404 |
| AHBS-1-U1 | 240 | *1024 | 27.177 | 2.7% | 18.492 | 1.8% | 18.060 | 1.8% | 22.522 | 2.2% | 43.749 |
| AHBS-1-U2 | 240 | 7.32 | 0.205 | 2.8% | 0.124 | 1.7% | 0.135 | 1.8% | 1.116 | 15.3% | 1.149 |
| AHBS-1-U3 | 240 | 16.9 | 0.386 | 2.3% | 0.618 | 3.7% | 0.000 | 0.0% | 1.841 | 10.9% | 1.98 |
| AHBS-1-U4 | 240 | 26.9 | 0.457 | 1.7% | 0.405 | 1.5% | 0.556 | 2.1% | 3.967 | 14.8% | 4.052 |
| AHBS-1-U5 | 240 | 35.2 | 0.608 | 1.7% | 0.675 | 1.9% | 0.313 | 0.9% | 2.234 | 6.4% | 2.432 |
| AHBS-1-U6 | 240 | 76 | 1.385 | 1.8% | 0.410 | 0.5% | 1.062 | 1.4% | 4.744 | 6.2% | 5.071 |
| AHBS-1-U7 | 240 | 122 | 1.770 | 1.4% | 1.288 | 1.1% | 1.693 | 1.4% | 14.523 | 11.9% | 14.784 |
| AHBS-1-U8 | 240 | 289 | 4.745 | 1.6% | 5.408 | 1.9% | 3.250 | 1.1% | 38.677 | 13.4% | 39.475 |
PMA P180039: FDA Summary of Safety and Effectiveness Data
{13}
# External Precision 5-day Study
A five (5) day precision/reproducibility study was conducted at two (2) external laboratories and at DiaSorin Inc. to verify the precision of the LIAISON® XL MUREX Anti-HBs assay. The CLSI document EP15-A3 was consulted in the preparation of the testing protocol. The coded panel, comprised of eight (8) frozen serum samples was the same panel used in the 20-day precision study.
The precision panel was tested at all three (3) sites on the LIAISON® XL Analyzer using six (6) replicates per run in one (1) run per day for five (5) operating days with multiple technicians performing the testing.
Table 6: Summary of Reproducibility Study
| LIAISON XL MUREX Anti-HBs Assay All 3 Lots Combined | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample ID | N | Mean | Repeatability | | Between-Day/Run | | Within Laboratory Precision | | Between-sites/Lot | | Reproducibility | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Ctrl Neg (All 3 lots) | 90 | 780.611 | 28.577 | 3.7% | 13.741 | 1.8% | 31.710 | 4.1% | 51.133 | 6.6% | 60.167 | 7.7% |
| Ctrl Pos (All 3 lots) | 90 | 53.777 | 1.313 | 2.4% | 0.811 | 1.5% | 1.543 | 2.9% | 1.862 | 3.5% | 2.418 | 4.5% |
| Cal Ver A (All 3 lots) | 90 | 7.278 | 0.217 | 3.0% | 0.082 | 1.1% | 0.232 | 3.2% | 0.227 | 3.1% | 0.325 | 4.5% |
| Cal Ver B (All 3 lots) | 90 | 19.269 | 0.527 | 2.7% | 0.279 | 1.4% | 0.596 | 3.1% | 1.044 | 5.4% | 1.202 | 6.2% |
| Cal Ver C (All 3 lots) | 90 | 74.560 | 1.345 | 1.8% | 1.253 | 1.7% | 1.838 | 2.5% | 1.200 | 1.6% | 2.195 | 2.9% |
| Cal Ver D (All 3 lots) | 90 | 242.422 | 4.881 | 2.0% | 3.828 | 1.6% | 6.203 | 2.6% | 13.707 | 5.7% | 15.045 | 6.2% |
| AHBS-1-U1 | 90 | 847.011 | 22.506 | 2.7% | 14.415 | 1.7% | 26.727 | 3.2% | 54.888 | 6.5% | 61.049 | 7.2% |
| AHBS-1-U2 | 90 | 7.119 | 0.260 | 3.6% | 0.248 | 3.5% | 0.359 | 5.0% | 0.264 | 3.7% | 0.445 | 6.3% |
| AHBS-1-U3 | 90 | 16.532 | 0.531 | 3.2% | 1.198 | 7.2% | 1.311 | 7.9% | 0.892 | 5.4% | 1.585 | 9.6% |
| AHBS-1-U4 | 90 | 25.670 | 1.265 | 4.9% | 2.254 | 8.8% | 2.585 | 10.1% | 0.981 | 3.8% | 2.765 | 10.8% |
| AHBS-1-U5 | 90 | 33.063 | 1.140 | 3.4% | 2.018 | 6.1% | 2.317 | 7.0% | 1.000 | 3.0% | 2.524 | 7.6% |
| AHBS-1-U6 | 90 | 72.782 | 2.392 | 3.3% | 4.628 | 6.4% | 5.210 | 7.2% | 2.950 | 4.1% | 5.987 | 8.2% |
| AHBS-1-U7 | 90 | 113.361 | 4.672 | 4.1% | 8.672 | 7.7% | 9.850 | 8.7% | 8.769 | 7.7% | 13.188 | 11.6% |
| AHBS-1-U8 | 90 | 282.244 | 8.281 | 2.9% | 10.400 | 3.7% | 13.294 | 4.7% | 20.494 | 7.3% | 24.429 | 8.7% |
*Precision calculations are based on signal (RLU)
# 10. Pediatric and Adult Sample Equivalency
Pediatric samples were tested to determine if these types of samples provide equivalent results to adult human serum.
PMA P180039: FDA Summary of Safety and Effectiveness Data
{14}
A total of thirty (30) negative pediatric patient samples were used for this study. The pediatric samples encompassed the age range of two (2) months to twenty-one (21) years. Ten (10) pediatric samples were spiked with an anti-HBs high positive sample to obtain high negative samples. Ten (10) pediatric samples were spiked with an IgG anti-HBs high positive sample to obtain low positive samples. Ten (10) pediatric samples were spiked with an anti-HBs high positive sample to obtain moderate positive samples. Adult negative pool samples were used as controls and were spiked with an anti-HBs high positive sample to achieve the same three (3) levels of samples: high negative, low positive and moderate positive samples.
The samples were tested in duplicate, with the LIAISON XL MUREX Anti-HBs assay. Percent (%) recovery of the analyte from the pediatric and adult blood was calculated for each sample.
All acceptance criteria were met demonstrating acceptable performance of pediatric samples. It can be concluded that pediatric samples react in the same way as the adult samples and are acceptable for use in the LIAISON XL MUREX Anti-HBs assay.
## B. Animal Studies
Not Applicable.
## C. Additional Studies
Not Applicable.
## X. SUMMARY OF PRIMARY CLINICAL STUDY(IES)
The applicant performed a clinical study to establish a reasonable assurance of safety and effectiveness for the detection and quantitation of antibodies to hepatitis-B surface antigen with the LIAISON XL MUREX Anti-HBs using samples that would routinely be tested for hepatitis in the US. Data from this clinical study were the basis for the PMA approval decision. A summary of the clinical study is presented below.
## A. Study Design
A multi-site clinical agreement study was conducted to evaluate the clinical performance of the LIAISON XL MUREX Anti-HBs assay on samples that would routinely be tested for hepatitis and samples that were selected from individuals that were diagnosed with acute or chronic Hepatitis B infection.
The clinical agreement study involved the testing of 3,082 samples on six (6) FDA approved reference assays, each detecting a unique serological marker (HBsAg, HBeAg, Anti-HBs, Anti-HBc, Anti-HBc IgM, and Anti-HBe) in order to determine the HBV classification for each of the samples tested.
PMA P180039: FDA Summary of Safety and Effectiveness Data
Page 15
{15}
The samples were collected from 6 different countries: Russia, Colombia, Cameroon, Ghana, Nigeria, and the United States. The U.S. samples were from multiple states including Ohio, Pennsylvania, Indiana, Florida, California, Texas, New Jersey, Tennessee, Massachusetts, and Puerto Rico.
The prospective (unselected) subjects were defined as follows:
- Pediatric and adult male (38.2%), female (61.7%) and unknown gender (0.1%) subjects at risk for hepatitis due to medical conditions (dialysis, transplantation), occupation, lifestyle, behavior or a known exposure event.
- Subjects showing signs or symptoms and individuals living in an area with a higher probability of HBV infection.
- The demographic breakdown of the prospective population was as follows: American Indian/Alaskan Native (0.1%), Asian (0.8%), Black/African American (31.2%), Caucasian (62.5%), Other (5.2%), and Unknown (0.2%) with an age range of 2 - 98 years of age.
The retrospective (selected/archived) samples were from male (69.5%), female (21.9%), and unknown gender (8.6%) subjects diagnosed with acute and/or chronic Hepatitis having an age range of 17 - 67 years of age from the following ethnicities: Asian (1.6%), Black/African American (23.8%), Caucasian (73%), other (1.2%) and 0.4% Unknown.
The distribution of LIAISON XL MUREX Anti-HBs reactive and non-reactive results by age and gender of the overall prospective population are presented below.
Table 7: Demographic Summary of the Prospective Population
| Age Range | Gender | LIAISON XL MUREX | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | + | | - | | Total |
| | | n | % | n | % | |
| 0-9 | F | 5 | 100.0% | 0 | 0.0% | 5 |
| | M | 3 | 30.0% | 7 | 70.0% | 10 |
| 10-19 | F | 21 | 52.5% | 19 | 47.5% | 40 |
| | M | 8 | 53.3% | 7 | 46.7% | 15 |
| 20-29 | F | 251 | 64.5% | 138 | 35.5% | 389 |
| | M | 130 | 55.8% | 103 | 44.2% | 233 |
| 30-39 | F | 253 | 53.4% | 221 | 46.6% | 474 |
| | M | 111 | 49.6% | 113 | 50.4% | 224 |
| 40-49 | F | 121 | 41.7% | 169 | 58.3% | 290 |
| | M | 68 | 33.8% | 133 | 66.2% | 201 |
| 50-59 | F | 106 | 45.7% | 126 | 54.3% | 232 |
| | M | 70 | 37.8% | 115 | 62.2% | 185 |
| 60-69 | F | 62 | 39.2% | 96 | 60.8% | 158 |
| | M | 50 | 46.3% | 58 | 53.7% | 108 |
PMA P180039: FDA Summary of Safety and Effectiveness Data
Page 16
{16}
PMA P180039: FDA Summary of Safety and Effectiveness Data
Page 17
| Age Range | Gender | LIAISON XL MUREX | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | + | | - | | Total |
| | | n | % | n | % | |
| 70-79 | F | 15 | 32.6% | 31 | 67.4% | 46 |
| | M | 16 | 41.0% | 23 | 59.0% | 39 |
| 80-89 | F | 8 | 57.1% | 6 | 42.9% | 14 |
| | M | 2 | 28.6% | 5 | 71.4% | 7 |
| 90-98 | F | 4 | 100.0% | 0 | 0% | 4 |
| | M | 0 | 0% | 0 | 0% | 0 |
| Unk | F | 1 | 100.0% | 0 | 0% | 1 |
| | M | 1 | 100.0% | 0 | 0% | 1 |
| Total | | 1306 | 48.8% | 1370 | 51.2% | 2676 |
## Hepatitis B Status Classification
Hepatitis B Status Classification was based on testing all the samples with FDA approved HBV assays for HBsAg, HBeAg, Anti-HBc, Anti-HBc IgM, Anti-HBe and Anti-HBs. HBV classification for the prospective and retrospective specimens is presented below.
Table 8: Hepatitis B Status Classification
| HBV Classification | HBsAg | HBeAg | Total Anti-HBc | Anti-HBc IgM | Anti-HBe | Anti-HBs | Prospective (n) | Retrospective (n) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Acute | R | NR | NR | NR | NR | NR | 12 | 97 |
| Acute | R | R | NR | NR | NR | NR |
| Acute | R | R | R | R | NR | NR |
| Acute | R | R | R | R | R | NR |
| Acute | R | R | R | R | EQV | NR |
| Acute | R | NR | R | EQV | R | NR |
| Acute | R | NR | R | R | EQV | NR |
| Acute | R | R | R | R | NR | NR |
| Acute | R | R | R | R | R | R |
| Acute | R | R | R | R | R | R |
| Acute | R | R | R | R | R | R |
| Rate Acute | R | NR | R | R | R | NR | 2 | 32 |
| Late Acute | R | NR | R | R | R | R |
| Late Acute | R | NR | R | R | R | R | 2 | 32 |
| Chronic | R | NR | NR | NR | R | NR | 76 | 68 |
{17}
| HBV Classification | HBsAg | HBeAg | Total Anti-HBc | Anti-HBc IgM | Anti-HBe | Anti-HBs | Prospective (n) | Retrospective (n) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Chronic | R | NR | R | NR | NR | R | | |
| Chronic | R | R | R | NR | NR | R | | |
| Chronic | R | R | R | NR | NR | NR | | |
| Chronic | R | EQV | R | NR | NR | NR | | |
| Chronic | R | NR | R | NR | R | NR | | |
| Chronic | R | NR | R | NR | NR | NR | | |
| Chronic | R | NR | R | NR | R | R | | |
| Chronic | R | EQV | R | NR | NR | NR | | |
| Early Recovery | NR | NR | R | R | R | NR | 48 | 9 |
| Early Recovery | NR | NR | R | EQV | R | R | | |
| Early Recovery | NR | NR | R | R | NR | NR | | |
| Early Recovery | NR | NR | R | NR | R | NR | | |
| Early Recovery | NR | NR | R | NR | NR | NR | | |
| Early Recovery | NR | NR | R | R | NR | R | | |
| Early Recovery | NR | NR | R | R | R | R | | |
| Recovery | NR | NR | R | NR | R | R | 131 | 36 |
| Recovery | NR | NR | NR | NR | R | R | | |
| Recovery | NR | NR | R | NR | EQV | R | | |
| Immune Due to Natural Infection | NR | NR | R | NR | NR | R | 104 | 3 |
| Immune Due to Natural Infection | NR | NR | R | NR | NR | EQV | | |
| HBV Vaccine Response | NR | NR | NR | NR | NR | R | 1144 | 8 |
| HBV Vaccine Response | NR | NR | NR | NR | NR | EQV | | |
| Not Previously Infected | NR | NR | NR | NR | NR | NR | 1302 | 1 |
PMA P180039: FDA Summary of Safety and Effectiveness Data
{18}
| HBV Classification | HBsAg | HBeAg | Total Anti-HBc | Anti-HBc IgM | Anti-HBe | Anti-HBs | Prospective (n) | Retrospective (n) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Not Interpretable | NR | NR | NR | NR | R | NR | 7 | 2 |
| Not Interpretable | NR | NR | NR | R | NR | NR | | |
| Not Interpretable | NR | R | NR | NR | NR | NR | | |
| Not Interpretable | NR | R | NR | NR | NR | R | | |
| Not Interpretable | NR | R | R | R | NR | EQV | | |
| Not Interpretable | NR | R | R | R | NR | R | | |
| Not Interpretable | R | NR | NR | NR | NR | R | | |
| Total | | | | | | | 2826 | 256 |
## Clinical Agreement Study Analysis
Comparison results of the LIAISON XL MUREX Anti-HBs to the reference anti-HBs assay are presented with Negative and Positive percent (%) agreement and 95% confidence intervals for combined prospective and retrospective specimens for each of the HBV Classification categories. In addition, Pediatric Cumulative (prospective and retrospective) Clinical Agreement results are presented below.
Table 9: Cumulative Clinical Agreement (Combined Prospective & Retrospective)
| HBV Classification | Reference Anti-HBs assay | | | | | | Total |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | Reactive | | Eqv | | Non reactive | | |
| | LIAISON XL MUREX Anti-HBs | | LIAISON XL MUREX Anti-HBs | | LIAISON XL MUREX Anti-HBs | | |
| | Positive | Negative | Positive | Negative | Positive | Negative | |
| Acute | 2 | 2 | 0 | 1 | 0 | 104 | 109 |
| Late Acute | 1 | 1 | 0 | 0 | 0 | 32 | 34 |
| Chronic | 3 | 1 | 0 | 0 | 1 | 139 | 144 |
| Early Recovery | 8 | 0 | 0 | 0 | 9 | 40 | 57 |
| Recovery | 163 | 4 | 0 | 0 | 0 | 0 | 167 |
| Immune Due to Natural Infection | 103 | 0 | 3 | 1 | 0 | 0 | 107 |
PMA P180039: FDA Summary of Safety and Effectiveness Data
{19}
| HBV Classification | Reference Anti-HBs assay | | | | | | Total |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | Reactive | | Eqv | | Non reactive | | |
| | LIAISON XL MUREX Anti-HBs | | LIAISON XL MUREX Anti-HBs | | LIAISON XL MUREX Anti-HBs | | |
| | Positive | Negative | Positive | Negative | Positive | Negative | |
| HBV Vaccine Response | 1106 | 14 | 16 | 16 | 0 | 0 | 1152 |
| Not Previously Infected | 0 | 0 | 0 | 0 | 29 | 1274 | 1303 |
| Not Interpretable | 3 | 0 | 0 | 1 | 0 | 5 | 9 |
| Total | 1389 | 22 | 19 | 19 | 39 | 1594 | 3082 |
Table 10: Cumulative Clinical Agreement (Combined Prospective & Retrospective)
| HBV Classification | Positive Percent Agreement (PPA) | Negative Percent Agreement (NPA) |
| --- | --- | --- |
| Acute | 2/5 (40.0%)
95% CI: 11.8% to 76.9% | 104/104 (100.0%)
95% CI: 96.4% to 100.0% |
| Late Acute | 1/2 (50.0%)
95% CI: 9.5% to 90.5% | 32/32 (100.0%)
95% CI: 89.3% to 100.0% |
| Chronic | 3/4 (75.0%)
95% CI: 30.1% to 95.4% | 139/140 (99.3%)
95% CI: 96.1% to 99.9% |
| Early Recovery | 8/8 (100.0%)
95% CI: 67.6% to 100.0% | 40/49 (81.6%)
95% CI: 68.6% to 90.0% |
| Recovery | 163/167 (97.6%)
95% CI: 94.0% to 99.1% | N/A |
| Immune Due to Natural Infection | 103/104 (99.0%)
95% CI: 94.8% to 99.8% | 0/3 (0.0%)
95% CI: 0.0% to 56.2% |
| HBV Vaccine Response | 1106/1136 (97.4%)
95% CI: 96.3% to 98.1% | 0/16 (0.0%)
95% CI: 0.0% to 19.4% |
| Not Previously Infected | N/A | 1274/1303 (97.8%)
95% CI: 96.8% to 98.4% |
| Not Interpretable | 3/4 (75.0%)
95% CI: 30.1% to 95.4% | 5/5 (100.0%)
95% CI: 56.6% to 100.0% |
| Total | 1389/1430 (97.1%)
95% CI: 96.1% to 97.9% | 1594/1652 (96.5%)
95% CI: 95.5% to 97.3% |
## Clinical Endpoints
With regard to safety, as an *in vitro* diagnostic test, the LIAISON XL MUREX Anti-HBs test involves taking a sample of plasma or serum from a patient. The test, therefore, presents
PMA P180039: FDA Summary of Safety and Effectiveness Data
{20}
no more safety hazard to an individual being tested than other tests where blood samples are drawn. Safety issues regarding false positive and negative test results are discussed in section VIII.
With regard to effectiveness, the clinical performance of the LIAISON XL MUREX Anti-HBs was evaluated versus a FDA approved anti-HBs test for patients at risk for infection with hepatitis B and for patients with signs and symptoms of hepatitis.
With regard to success/failure criteria, the assay performed well with a positive percent agreement (PPA) of 97.1% a negative percent agreement (NPA) of 96.5% among subjects in various stages of HBV infection, a PPA of 97.4% in HBV-vaccinated subjects, a PPA of 99.0% in subjects immune due to natural infection, and an NPA of 97.8% among subjects not previously infected with HBV.
## B. Accountability of PMA Cohort
The clinical agreement study involved the testing of 3,082 samples on six (6) FDA approved reference assays, each detecting a unique serological marker (HBsAg, HBeAg, Anti-HBs, Anti-HBc, Anti-HBc IgM, and Anti-HBe in order determine the HBV classification for each of the samples tested.
The samples were collected from 6 different countries: Russia, Colombia, Cameroon, Ghana, Nigeria, and the United States. The U.S. samples were from multiple states including Ohio, Pennsylvania, Indiana, Florida, California, Texas, New Jersey, Tennessee, Massachusetts, and Puerto Rico.
## C. Study Population Demographics and Baseline Parameters
The demographics of the study population are typical for an anti-HBs detection study performed in the US.
The prospective (unselected) subjects were defined as follows:
- Pediatric and adult male (38.2%), female (61.6%) and unknown gender (0.2%) subjects at risk for hepatitis due to medical conditions (dialysis, transplantation), occupation, lifestyle, behavior or a known exposure event.
- Subjects showing signs or symptoms and individuals living in an area with a higher probability of HBV infection.
- The demographic breakdown of the prospective population was as follows: American Indian/Alaskan Native (0.1%), Asian (0.8%), Black/African American (31.2%), Caucasian (62.5%), Other (5.2%), and Unknown (0.2%) with an age range of 2 - 98 years of age.
The retrospective (selected/archived) samples were from male (69.5%), female (21.9%), and unknown gender (8.6%) subjects diagnosed with Acute and/or Chronic Hepatitis having an age range of 17 - 67 years of age from the following ethnicities: Asian (1.6%),
PMA P180039: FDA Summary of Safety and Effectiveness Data
Page 21
{21}
Black/African American (23.8%), Caucasian (73%), other (1.2%) and 0.4% Unknown. The table below shows the demographic distribution of the cohort.
Table11: Prospective and Retrospective Demographic Summary
| | Adult | | | | Pediatric (2-21) | | | | Unknown Age | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Prospective | | Retrospective | | Prospective | | Retrospective | | Prospective | | Retrospective | |
| Race | n | % | n | % | n | % | n | % | n | % | n | % |
| American Indian/Alaskan Native | 2 | 0.1% | 0 | 0.0% | 0 | 0.0% | 0 | 0.0% | 0 | 0.0% | 0 | 0.0% |
| Asian | 21 | 0.8% | 4 | 1.9% | 3 | 1.9% | 0 | 0.0% | 0 | 0.0% | 0 | 0.0% |
| Black/African American | 832 | 31.2% | 57 | 27.7% | 64 | 39.8% | 4 | 13.8% | 0 | 0.0% | 0 | 0.0% |
| Native Hawaiian or Other Pacific Islander | 0 | 0.0% | 0 | 0.0% | 0 | 0.0% | 0 | 0.0% | 0 | 0.0% | 0 | 0.0% |
| White | 1664 | 62.5% | 141 | 68.4% | 89 | 55.3% | 25 | 86.2% | 2 | 100.0% | 21 | 100.0% |
| Unknown | 6 | 0.2% | 1 | 0.5% | 0 | 0.0% | 0 | 0.0% | 0 | 0.0% | 0 | 0.0% |
| Other | 138 | 5.2% | 3 | 1.5% | 5 | 3.1% | 0 | 0.0% | 0 | 0.0% | 0 | 0.0% |
| Total | 2663 | 100.0% | 206 | 100.0% | 161 | 100.0% | 29 | 100.0% | 2 | 100.0% | 21 | 100.0% |
## D. Safety and Effectiveness Results
### 1. Safety Results
With regard to safety, as an in vitro diagnostic test, the LIAISON XL MUREX Anti-HBs test involves taking a sample of plasma or serum from a patient. The test, therefore, presents no more safety hazard to an individual being tested than other tests where blood samples are drawn.
There were no adverse effects that occurred in the PMA clinical study.
### 2. Effectiveness Results
The analysis of effectiveness was based on the 3,082 evaluable patients. Key effectiveness outcomes are presented in the tables below.
PMA P180039: FDA Summary of Safety and Effectiveness Data
{22}
Table 12: Clinical Comparison for Prospectively Collected Samples
| HBV Classification | Roche Anti-HBs | | | | | | Total |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | + | | Eqv | | - | | |
| | LIAISON XL MUREX Anti-HBs | | LIAISON XL MUREX Anti-HBs | | LIAISON XL MUREX Anti-HBs | | |
| | + | - | + | - | + | - | |
| Acute | 0 | 0 | 0 | 0 | 0 | 12 | 12 |
| Late Acute | 0 | 0 | 0 | 0 | 0 | 2 | 2 |
| Chronic | 3 | 1 | 0 | 0 | 1 | 71 | 76 |
| Early Recovery | 5 | 0 | 0 | 0 | 8 | 35 | 48 |
| Recovery | 127 | 4 | 0 | 0 | 0 | 0 | 131 |
| Immune Due to Natural Infection | 100 | 0 | 3 | 1 | 0 | 0 | 104 |
| HBV Vaccine Response | 1099 | 13 | 16 | 16 | 0 | 0 | 1144 |
| Not Previously Infected | 0 | 0 | 0 | 0 | 29 | 1273 | 1302 |
| Not Interpretable | 1 | 0 | 0 | 1 | 0 | 5 | 7 |
| Total | 1335 | 18 | 19 | 18 | 38 | 1398 | 2826 |
Table 13: Percent Agreement for Prospectively Collected Samples
| HBV Classification | Positive Percent Agreement (PPA) | Negative Percent Agreement (NPA) |
| --- | --- | --- |
| Acute | N/A | 12/12 (100.0%)
95% CI: 75.7% to 100.0% |
| Late Acute | N/A | 2/2 (100.0%)
95% CI: 34.2% to 100.0% |
| Chronic | 3/4 (75.0%)
95% CI: 30.1% to 95.4% | 71/72 (98.6%)
95% CI: 92.5% to 99.8% |
| Early Recovery | 5/5 (100.0%)
95% CI: 56.6% to 100.0% | 35/43 (81.4%)
95% CI: 67.4% to 90.3% |
| Recovery | 127/131 (96.9%)
95% CI: 92.4% to 98.8% | N/A |
| Immune Due to Natural Infection | 100/101 (99.0%)
95% CI: 94.6% to 99.8% | 0/3 (0.0%)
95% CI: 0.0% to 56.2% |
| HBV Vaccine Response | 1099/1128 (97.4%)
95% CI: 96.3% to 98.1% | 0/16 (0.0%)
95% CI: 0.0% to 19.4% |
| Not Previously Infected | N/A | 1273/1302 (97.8%)
95% CI: 96.8% to 98.4% |
| Not Interpretable | 1/2 (50.0%)
95% CI: 9.5% to 90.5% | 5/5 (100.0%)
95% CI: 56.6% to 100.0% |
| Total | 1335/1371 (97.4%)
95% CI: 96.4% to 98.1% | 1398/1455 (96.1%)
95% CI: 95.0% to 97.0% |
PMA P180039: FDA Summary of Safety and Effectiveness Data
{23}
Table 14: Clinical Comparison for Retrospectively Collected Samples
| HBV Classification | Roche Anti-HBs | | | | | | Total |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | + | | Eqv | | - | | |
| | LIAISON XL MUREX Anti-HBs | | LIAISON XL MUREX Anti-HBs | | LIAISON XL MUREX Anti-HBs | | |
| | + | - | + | - | + | - | |
| Acute | 2 | 2 | 0 | 1 | 0 | 92 | 97 |
| Late Acute | 1 | 1 | 0 | 0 | 0 | 30 | 32 |
| Chronic | 0 | 0 | 0 | 0 | 0 | 68 | 68 |
| Early Recovery | 3 | 0 | 0 | 0 | 1 | 5 | 9 |
| Recovery | 36 | 0 | 0 | 0 | 0 | 0 | 36 |
| Immune Due to Natural Infection | 3 | 0 | 0 | 0 | 0 | 0 | 3 |
| HBV Vaccine Response | 7 | 1 | 0 | 0 | 0 | 0 | 8 |
| Not Previously Infected | 0 | 0 | 0 | 0 | 0 | 1 | 1 |
| Not Interpretable | 2 | 0 | 0 | 0 | 0 | 0 | 2 |
| Total | 54 | 4 | 0 | 1 | 1 | 196 | 256 |
Table 15: Percent Agreement for Retrospectively Collected Samples
| HBV Classification | Positive Percent Agreement (PPA) | Negative Percent Agreement (NPA) |
| --- | --- | --- |
| Acute | 2/5 (40.0%)
95% CI: 11.8% to 76.9% | 92/92 (100.0%)
95% CI: 96% to 100.0% |
| Late Acute | 1/2 (50.0%)
95% CI: 9.5% to 90.5% | 30/30 (100.0%)
95% CI: 88.6% to 100.0% |
| Chronic | N/A | 68/68 (100.0%)
95% CI: 94.7% to 100.0% |
| Early Recovery | 3/3 (100.0%)
95% CI: 43.8% to 100.0% | 5/6 (83.3%)
95% CI: 43.6% to 97.0% |
| Recovery | 36/36 (100.0%)
95% CI: 90.4% to 100.0% | N/A |
| Immune Due to Natural Infection | 3/3 (100.0%)
95% CI: 43.8% to 100.0% | N/A |
| HBV Vaccine Response | 7/8 (87.5%)
95% CI: 52.9% to 97.8% | N/A |
| Not Previously Infected | N/A | 1/1 (100.0%)
95% CI: 20.7% to 100.0% |
| Not Interpretable | 2/2 (100.0%)
95% CI: 34.2% to 100.0% | N/A |
| Total | 54/59 (91.5%)
95% CI: 81.6% to 96.8% | 196/197 (99.5%)
95% CI: 97.8% to 99.9% |
PMA P180039: FDA Summary of Safety and Effectiveness Data
Page 24
{24}
Table 16: Pediatric- Percent Agreement for Prospectively Collected Samples
| HBV Classification | Prospective Age in Years | Positive Percent Agreement (PPA) | Negative Percent Agreement (NPA) |
| --- | --- | --- | --- |
| Acute | 2-12 | NA | NA |
| | 13-18 | NA | NA |
| | 19-21 | NA | NA |
| Late Acute | 2-12 | NA | NA |
| | 13-18 | NA | NA |
| | 19-21 | NA | NA |
| Chronic | 2-12 | NA | NA |
| | 13-18 | NA | NA |
| | 19-21 | NA | 4/5 (80.0%)
37.6%-96.4% |
| Early Recovery | 2-12 | NA | NA |
| | 13-18 | NA | NA |
| | 19-21 | NA | NA |
| Recovery | 2-12 | NA | NA |
| | 13-18 | NA | NA |
| | 19-21 | 4/5 (80.0%)
37.6%-96.4% | NA |
| Immune Due to Natural Infection | 2-12 | NA | NA |
| | 13-18 | 2/2 (100.0%)
34.2%-100.0% | NA |
| | 19-21 | 1/1 (100.0%)
20.7%-100.0% | NA |
| HBV Vaccine Response | 2-12 | 16/16 (100.0%)
80.6%-100.0% | 0/1 (0.0%)
0.0%-79.3% |
| | 13-18 | 14/16 (87.5%)
64.0%-96.5% | NA |
| | 19-21 | 26/27 (96.3%)
81.7%-99.3% | 0/2 (0.0%)
0.0%-65.8% |
| Not Previously Infected | 2-12 | NA | 12/12 (100.0%)
75.7%-100.0% |
| | 13-18 | NA | 11/13 (84.6%)
57.8%-95.7% |
| | 19-21 | NA | 55/59 (93.2%)
83.8%-97.3% |
| Not Interpretable | 2-12 | NA | 1/1 (100.0%)
20.7%-100.0% |
| | 13-18 | NA | NA |
| | 19-21 | NA | 1/1 (100.0%)
20.7%-100.0% |
PMA P180039: FDA Summary of Safety and Effectiveness Data
{25}
Table 17: Pediatric- Percent Agreement for Retrospectively Collected Samples
| HBV Classification | Retrospective Age in Years | Positive Percent Agreement (PPA) | Negative Percent Agreement (NPA) |
| --- | --- | --- | --- |
| Acute | 2-12 | NA | NA |
| | 13-18 | NA | 6/6 (100.0%)
61.0%-100.0% |
| | 19-21 | NA | 14/14 (100.0%)
78.5%-100.0% |
| Late Acute | 2-12 | NA | NA |
| | 13-18 | NA | 3/3 (100.0%)
43.8%-100.0% |
| | 19-21 | NA | 4/4 (100.0%)
51.0%-100.0% |
| Chronic | 2-12 | NA | NA |
| | 13-18 | NA | NA |
| | 19-21 | NA | 2/2 (100.0%)
34.2%-100.0% |
| Early Recovery | 2-12 | NA | NA |
| | 13-18 | NA | NA |
| | 19-21 | NA | 1/1 (100.0%)
20.7%-100.0% |
| Recovery | 2-12 | NA | NA |
| | 13-18 | NA | NA |
| | 19-21 | NA | NA |
| Immune Due to Natural Infection | 2-12 | NA | NA |
| | 13-18 | NA | NA |
| | 19-21 | NA | NA |
| HBV Vaccine Response | 2-12 | NA | NA |
| | 13-18 | NA | NA |
| | 19-21 | 1/1 (100.0%)
20.7%-100.0% | NA |
| Not Previously Infected | 2-12 | NA | NA |
| | 13-18 | NA | NA |
| | 19-21 | NA | NA |
| Not Interpretable | 2-12 | NA | NA |
| | 13-18 | NA | NA |
| | 19-21 | NA | NA |
PMA P180039: FDA Summary of Safety and Effectiveness Data
Page 26
{26}
PMA P180039: FDA Summary of Safety and Effectiveness Data
Page 27
3. **Subgroup Analyses**
The study design enabled an assessment of assay performance by subgroup as depicted in tables above which show subjects stratified by stage of HBV infection and HBV vaccine response.
4. **Pediatric Extrapolation**
In this premarket application, existing clinical data from adult subjects was not leveraged to support approval of a pediatric patient population.
Samples from pediatric patients were tested in the clinical study in order to support a pediatric claim.
E. **Financial Disclosure**
The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study included 3 investigators. None of the clinical investigators had disclosable financial interests/arrangements as defined in sections 54.2(a), (b), (c), and (f). The information provided does not raise any questions about the reliability of the data.
XI. **PANEL MEETING RECOMMENDATION AND FDA’S POST-PANEL ACTION**
In accordance with the provisions of section 515(c)(3) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Microbiology Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.
XII. **CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES**
A. **Effectiveness Conclusions**
The effectiveness of the LIAISON XL MUREX Anti-HBs test for the qualitative and quantitative determination of antibodies to hepatitis-B surface antigen in human serum and plasma (lithium and sodium heparin and K2 EDTA) samples including separator tubes, on the LIAISON XL Analyzer has been demonstrated in the following patient populations: adults and pediatric patients (2 – 21 years). The results of this test may be used as an aid in the diagnosis of hepatitis B virus (HBV) infection in patients with symptoms of hepatitis or who may be at risk for hepatitis B infection. The results of this test may be used as an aid in the determination of susceptibility to HBV infection in individuals prior to or following HBV vaccination or where vaccination status is unknown. The positive agreement of the assay is 97.1% with a two-sided 95%
{27}
confidence interval (CI) of 96.1% - 97.1% and the negative percent agreement is 96.5% with a two-sided 95% CI of 95.5 - 97.3%.
## B. Safety Conclusions
The risks of the device are based on nonclinical laboratory studies as well as data collected in a clinical study conducted to support PMA approval as described above. Based on the results of these studies the LIAISON XL MUREX Anti-HBs assay when used according to the manufacturer's instructions can aid the physician in the diagnosis of HBV infection and in the determination of susceptibility to HBV infection. The positive agreement of the assay is 97.1% with a two-sided 95% confidence interval (CI) of 96.1% - 97.1% and the negative percent agreement is 96.5% with a two-sided 95% CI of 95.5 - 97.3%.
## C. Benefit-Risk Determination
The probable benefits of the device are also based on data collected in a clinical study conducted to support PMA approval as described above. The benefits of the assay are the determination of susceptibility to hepatitis B virus infection before or following vaccination, or as part of a hepatitis B panel, the appropriate diagnosis and treatment of hepatitis B infection. Treatment or vaccination for appropriate patients can mitigate the sequelae of hepatitis B infection and may result in improved morbidity and mortality in these patients. Known sequelae of hepatitis B infection include continued symptoms, increases in all-cause mortality, liver disease-related complications and death, hepato-cellular carcinoma rates, and need for liver transplantation.
Additionally, diagnosis and appropriate treatment or appropriate vaccination for hepatitis B infection can potentially decrease transmission and disease burden in the general population and particularly in populations at high risk for hepatitis B infection. While the performance of the LIAISON XL MUREX Anti-HBs assay in the clinical study suggests that patients will benefit from the assay, low prevalence of certain HBV classifications (namely acute infection, chronic infection, and early recovery) is a source of potential uncertainty when analyzing the prospective and retrospective samples. The wide confidence intervals for those subgroups is expected due to the biology of hepatitis B infection and is acceptable.
The probable risks of the device are also based on data collected in a clinical study conducted to support PMA approval as described above.
Risks of false positive results when the device is used as an aid in the diagnosis of hepatitis B virus (HBV)
infection in patients with symptoms of hepatitis or who may be at risk for HBV infection include improper patient management, including treatment for hepatitis B with antiviral medication. Antiviral medication has risks including toxicity and more rarely allergic reactions. Over time, viral resistance in patients who are coinfected
PMA P180039: FDA Summary of Safety and Effectiveness Data
Page 28
{28}
but undiagnosed with other viruses that are treated with the same antiviral medication, such as HIV, can lead to viral resistance, however the chance of an undiagnosed coinfection in a patient treated for hepatitis B is unlikely.
Risks of false positive results when the device is used as an aid in the determination of susceptibility to HBV infection in individuals prior to or following HBV vaccination or where vaccination status is unknown includes leading a provider to falsely believe that a patient has been vaccinated in the past and/or has current immunity when the patient, in fact, does not. This could lead to a patient inappropriately missing a vaccination.
Risks of false negative results when the device is used as an aid in the diagnosis of HBV infection in patients with symptoms of hepatitis or who may be at risk for HBV infection include potentially missing and undertreating a patient who has hepatitis B infection. Missing and under-treating a patient with hepatitis B infection whose clinical picture warrants antiviral treatment could result in the known sequelae of HBV infection and may result in high morbidity and mortality in these patients. Additionally, missing a diagnosis of hepatitis B infection will not allow for clinicians to potentially decrease transmission and disease burden in the general population, particularly in populations at high risk for hepatitis B infection.
Risks of false negative results when the device is used as an aid in the determination of susceptibility to HBV infection in individuals prior to or following HBV vaccination or where vaccination status is unknown include unnecessary repeated vaccination for hepatitis B. Although vaccination has risks such as local reactions, administrations of extra doses of single-antigen hepatitis B vaccine is not inherently harmful. In fact, it is common practice to administer higher dose vaccinations in an accelerated schedule for patients undergoing organ transplantation, for example.
1. Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
In conclusion, given the available information above, the data support the claimed intended use, the probable benefits outweigh the probable risks.
D. Overall Conclusions
The data in this application support the reasonable assurance of safety and effectiveness of this device when used in accordance with the indications for use. The probable clinical benefits outweigh the potential risks for the proposed assay considering the performance of the device in the clinical trial and the low risk and associated risk mitigations in clinical practice. The proposed assay labelling will facilitate accurate assay implementation and interpretation of results. The clinical performance observed in the prospective and retrospective clinical trial suggests that errors will be uncommon and that the assay may provide substantial benefits to
PMA P180039: FDA Summary of Safety and Effectiveness Data
Page 29
{29}
patients as an accurate and sensitive aid in the diagnosis of HBV infection when used in conjunction with other laboratory results and clinical information or for the determination of susceptibility to HBV infection in individuals prior to or following HBV vaccination or where vaccination status is unknown.
## XIII. CDRH DECISION
CDRH issued an approval order on February 21, 2020.
The applicant’s manufacturing facilities have been inspected and found to be in compliance with the device Quality System (QS) regulation (21 CFR 820).
## XIV. APPROVAL SPECIFICATIONS
Directions for use: See device labeling.
Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the device labeling.
Post-approval Requirements and Restrictions: See approval order.
PMA P180039: FDA Summary of Safety and Effectiveness Data
Page 30
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.