IMMULITE 2000 CARCINOEMBRYONIC ANTIGEN (CEA) KIT

K980855 · Dpc Cirrus · DHX · Apr 21, 1998 · Immunology

Device Facts

Record IDK980855
Device NameIMMULITE 2000 CARCINOEMBRYONIC ANTIGEN (CEA) KIT
ApplicantDpc Cirrus
Product CodeDHX · Immunology
Decision DateApr 21, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.6010
Device ClassClass 2

Indications for Use

The DPC IMMULITE 2000 CEA is a solid-phase, two site chemiluminescent enzyme immunometric assay for use with the IMMULITE 2000 Automated Analyzer and designed for the quantitative measurement of carcinoembryonic antigen (CEA) in serum. It is intended for in vitro diagnostic use as an aid in the management of cancer patients and in the assessment of prognosis.

Device Story

IMMULITE 2000 CEA is a solid-phase, two-site chemiluminescent enzyme immunometric assay. Device processes serum samples on the IMMULITE 2000 Automated Analyzer to quantify CEA levels. Used in clinical laboratory settings by trained personnel. Output provides quantitative CEA concentration, assisting clinicians in cancer patient management and prognostic assessment.

Clinical Evidence

No clinical data provided in the document; bench testing only.

Technological Characteristics

Solid-phase, two-site chemiluminescent enzyme immunometric assay. Designed for use with the IMMULITE 2000 Automated Analyzer. In vitro diagnostic device.

Indications for Use

Indicated for the quantitative measurement of carcinoembryonic antigen (CEA) in serum as an aid in the management of cancer patients and in the assessment of prognosis.

Regulatory Classification

Identification

A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.

Special Controls

*Classification.* Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wing segments, representing health, human services, and the USA. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle. ## APR 2 1 1998 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Mr. Robert M Di Tullio Director, Quality and Requlatory Affairs Diagnostic Products Corporation (DPC) B13-B18, 111 Canfield Avenue Randolph, New Jersey 07869 Re : K980855 IMMULITE 2000 Carcinoembryonic Antigen (CEA) Kit Trade Name: Requlatory Class: II Product Code: DHX Dated: March 4, 1998 March 5, 1998 Received: Dear Mr. Tullio: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations. {1}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. TO determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" Sincerely yours, Steven Putman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ 510(k) Number (if known): Device Name: IMMULITE 2000 CEA Indications for Use: The DPC IMMULITE 2000 CEA is a solid-phase, two site chemiluminescent enzyme immunometric assay for use with the IMMULITE 2000 Automated Analyzer and designed for the quantitative measurement of carcinoembryonic antigen (CEA) in serum. It is intended for in vitro diagnostic use as an aid in the management of cancer patients and in the assessment of prognosis. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Peter E. Macias (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number K980855 Prescription Use OR Over-The-Counter Use ----------
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...