ACCESS CEA REAGENTS ON THE ACCESS IMMUNOASSAY ANALYZER MODEL 33200, 33205, 33206 & 33209
Applicant
Beckman Coulter, Inc.
Product Code
DHX · Immunology
Decision Date
Sep 24, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.6010
Device Class
Class 2
Indications for Use
The Access® CEA assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of Carcinoembryonic antigen (CEA) in human serum using the ACCESS Immunoassay System. CEA measured by the Access CEA Immunoassay, is used as an aid in the management of cancer patients.
Device Story
ACCESS CEA Assay is a paramagnetic particle chemiluminescent immunoassay for quantitative measurement of CEA in human serum. Used on the ACCESS Immunoassay Analyzer; system performs automated random access processing. Device utilizes monoclonal antibodies specific to CEA epitopes; binding reaction measured via chemiluminescence. Output is quantitative CEA concentration in serum. Used in clinical laboratory settings by trained personnel. Results assist healthcare providers in monitoring cancer patients; longitudinal data compared against clinical assessments to track disease status. Benefits include automated, high-throughput monitoring of CEA levels to aid clinical decision-making in cancer management.
Clinical Evidence
Bench testing only. Correlation study (n=146) comparing ACCESS CEA to Abbott IMx showed r=0.97, y=1.02x + 1.76. Expected range established in 234 healthy individuals. Longitudinal monitoring of cancer patients showed high concordance with IMx and clinical assessments. Analytical performance: recovery 99.9%, precision <5% (5-500 ng/ml), analytical sensitivity 0.1 ng/ml. No significant interference from CEA-like antigens, common contaminants, or therapeutic agents. No hook effect up to 100,000 ng/ml.
Technological Characteristics
Paramagnetic particle solid phase; chemiluminescent detection; monoclonal antibody-based immunoassay. Designed for use on the ACCESS Immunoassay Analyzer (fully automated, random access). Measures CEA in human serum. No specific material standards cited.
Indications for Use
Indicated for the quantitative determination of carcinoembryonic antigen (CEA) in human serum as an aid in the management of cancer patients.
Regulatory Classification
Identification
A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.
Special Controls
*Classification.* Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.
Predicate Devices
Abbott IMx CEA Microparticle Enzyme Immunoassay (P830066)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date "SEP 24 1998". The month is September, the day is the 24th, and the year is 1998. The text is in a bold, sans-serif font. The image is slightly blurry.
K 981985
## 510(k) Summary of Safety and Effectiveness
#### 1. General Information
Device Generic Name:
Device Trade Name:
Applicant's Name and Address:
Carcinoembryonic Antigen (CEA) Immunological Test System for Management of Cancers
ACCESS® CEA Assay
Beckman Coulter, Inc. 1000 Lake Hazeltine Drive Chaska, MN 55318
Date:
June 4, 1998
#### 2. Predicate Device
Abbott IMx CEA Microparticle Enzyme Immunoassay Abbott Laboratories One Abbott Park Road Abbott Park, IL 60064-3500
PMA Number: P830066
#### 3. Device Description
The ACCESS CEA Immunoassay Reagents and the ACCESS Immunoassay Analyzer comprise the ACCESS Immunoassay System for the quantitative determination of CEA in human serum.
#### 4. Indications for Use
The ACCESS CEA Immunoassay is a paramagnetic particle chemiluminescent immunoassay for the quantitative determination of carcinoembryonic antigen (CEA) in human serum using the ACCESS Immunoassay System. CEA, measured by the ACCESS CEA Immunoassay, is intended for use as an aid in the management of cancer patients.
#### 5. Comparison of Technological Characteristics
Both the ACCESS CEA Immunoassay and the Abbott IMx CEA assay quantitatively measure serum CEA by means of simultaneous immunoassays utilizing the binding of CEA to monoclonal antibodies specific to epitopes on the CEA molecule. Both systems utilize liquid multi-point calibrators.
The ACCESS CEA Immunoassay Reagents are designed for use on the ACCESS Immunoassay Analyzer, a fully automated random access system, with up to 24 different reagents on board. The Abbott IMx Immunoanalyzer is an automated batch mode system in which a single analyte is assaved at one time. The ACCESS Immunoassay Analyzer uses magnetic particle solid phase enzyme immunoassays with chemiluminescent measurement, while the Abbott IMx CEA Immunoassay uses a glass-fiber matrix as the solid phase with a fluorescent detection method.
#### 6. Summary of Studies
Correlation: A comparison of CEA values from 146 samples, ranging from 0 to 1000 ng/ml, run with both the ACCESS CEA Immunoassay and the Abbott IMx CEA Immunoassay demonstrated very good agreement with the following statistical data: r = 0.97; y = 1.02x + 1.76.
Expected Range: In a population of 234 apparently healthy males and females, 98.7% had CEA values of 5.0 ng/ml or less, with the remaining 1.3% having values in the 5.1 to 10.0 nq/ml range.
{1}------------------------------------------------
Monitoring Data: Longitudinal serum samples from previously diagnosed cancer patients were compared using the ACCESS CEA and IMx CEA assays. The monitoring patient data demonstrate that the Access CEA results are highly concordant to both the IMx CEA results and the concurrent clinical assessments.
Recovery: Linearity studies performed by diluting human serum samples with ACCESS CEA Sample Diluent provided an average recovery of 99.9%, with sample mean recoveries ranging from 91.1 to 109.1%. Recovery of exogenous CEA spiked into serum samples resulted in an overall mean recovery of 103.0%, with mean recoveries ranging from 99.0 to 106.2%.
Precision: Within-run, between-run and total imprecision of the Access CEA immunoassay were less than 5% when tested at concentrations between 5.0 and 500 ng/ml.
Specificity: There was no significant interference from CEA-like antigens (NCA, NCA-2. NCA-50 and NFA-1), potential sample contaminants (albumin, bilirubin, HAMA, hemoglobin, lipemia and rheumatoid factor), and therapeutic agents (bleomycin-, cisplatin, cyclophosphamide, doxorubicin, fluorouracil, leucovorin, methotrexate, mitomycin, tamoxifen, vinblastine, and vincristine).
Hook Effect: There was no hook effect observed at CEA concentrations up to 100,000 ng/ml.
Analytical Sensitivity: The lowest level of detection of CEA distinguishable from zero (ACCESS CEA Calibrator S0) with 95% confidence is 0.1 ng/ml.
#### 7. Conclusion
The ACCESS CEA Immunoassay Reagents, when used in conjunction with the ACCESS Immunoassay Analyzer, are substantially equivalent to the Abbott IMx CEA Immunoassay. The ACCESS CEA Immunoassay is appropriate for monitoring patients with cancer.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular fashion around the eagle.
# SEP 2 1998
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Ellen Voss Regulatory Affairs Beckman Coulter 1000 Lake Hazeltine Drive Chaska, Minnesota 55318-1084
Re: K981985/S1 Trade Name: ACCESS® CEA Assay Regulatory Class: II Product Code: DHX Dated: July 31, 1998 Received: July 31, 1998
Dear Ms. Voss : .......
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
the country of the same
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of themact = " for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
### Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a leqally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet "ram address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Gutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page _________________________________________________________________________________________________________________________________________________________________________ 1 of of 1
510(k) Number (if known):
Device Name: ACCESS® CEA
### Indications For Use:
The Access® CEA assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of Carcinoembryonic antigen (CEA) in human serum using the ACCESS Immunoassay System. CEA measured by the Access CEA Immunoassay, is used as an aid in the management of cancer patients.
(PLEASE DO NOT WRITE BELOW THIS LINE--CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Peter E. Malm
(Division Sign Division of Clinical Laboratory De 510(k) Number
Prescription Use (Per 21 CFR 801.109)
OR
Over-The Counter Use _________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.