Access CEA

K223921 · Beckman Coulter, Inc. · DHX · Sep 22, 2023 · Immunology

Device Facts

Record IDK223921
Device NameAccess CEA
ApplicantBeckman Coulter, Inc.
Product CodeDHX · Immunology
Decision DateSep 22, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.6010
Device ClassClass 2

Indications for Use

The Access CEA assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of Carcinoembryonic Antigen (CEA) levels in human serum, using the Access Immunoassay Systems. CEA measured by the Access Immunoassay Systems is used as an aid in the management of cancer patients in whom changing CEA concentrations have been observed.

Device Story

Access CEA is a two-site immunoenzymatic sandwich assay using mouse monoclonal anti-CEA antibodies. It utilizes paramagnetic particles and chemiluminescent detection to quantify CEA in human serum. The assay is designed for use on Beckman Coulter's Dxl 9000 Access Immunoassay Analyzer. It is intended for clinical laboratory use by trained personnel. The device transforms patient serum samples into quantitative CEA concentration values. Healthcare providers use these results to monitor cancer patients, specifically to track changes in CEA levels over time, which aids in clinical management decisions. The assay provides a standardized, automated method for longitudinal monitoring of disease status.

Clinical Evidence

Bench testing only. Method comparison study (n=153) demonstrated high correlation with the predicate (R=1.00, slope 0.98). Imprecision studies (CLSI EP-05-A3) showed within-laboratory CVs ranging from 2.5% to 5.2% across the measuring range (0.45–865 ng/mL). Analytical performance established: LoB 0.09 ng/mL, LoD 0.1 ng/mL, and LoQ 0.2 ng/mL. Linearity confirmed across the 0.2–1,000 ng/mL range.

Technological Characteristics

Paramagnetic particle, chemiluminescent immunoassay. Two-site immunoenzymatic sandwich format using mouse monoclonal anti-CEA antibodies. Automated operation on Dxl 9000 Access Immunoassay Analyzer. Measuring range 0.2–1,000 ng/mL. Reagents stable at 2–10°C for 28 days. Uses Lumi-Phos PRO substrate.

Indications for Use

Indicated for the quantitative determination of Carcinoembryonic Antigen (CEA) levels in human serum to aid in the management of cancer patients experiencing changing CEA concentrations.

Regulatory Classification

Identification

A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.

Special Controls

*Classification.* Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY ## I Background Information: A 510(k) Number K223921 B Applicant Beckman Coulter, Inc. C Proprietary and Established Names Access CEA D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | DHX | Class II | 21 CFR 866.6010 - Tumor-Associated Antigen Immunological Test System | IM - Immunology | ## II Submission/Device Overview: A Purpose for Submission: Modification of the previously cleared device – the change of substrate and updated analytical measuring interval of the assay on the DxI 9000 Immunoassay Analyzer B Measurand: Carcinoembryonic antigen (CEA) C Type of Test: Quantitative, Chemiluminescent immunoassay Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} K223921 - Page 2 of 9 ## III Intended Use/Indications for Use: ### A Intended Use(s): See Indications for Use below. ### B Indication(s) for Use: The Access CEA assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of Carcinoembryonic Antigen (CEA) levels in human serum, using the Access Immunoassay Systems. CEA measured by the Access Immunoassay Systems is used as an aid in the management of cancer patients in whom changing CEA concentrations have been observed. ### C Special Conditions for Use Statement(s): Rx - For Prescription Use Only The Instruction for use of the device contains the following warnings: The concentration of CEA in a given specimen determined with different manufacturers can vary due to differences in assay methods and reagent specificity. The results reported by the laboratory to the physician must include the identity of the CEA assay used. Values obtained with different assay methods cannot be used interchangeably. If, in the course of monitoring a patient, the assay method used for determining CEA values is changed, additional sequential testing should be carried out to confirm baseline values. ### D Special Instrument Requirements: DxI 9000 Access Immunoassay Analyzer (refer to K221225) ## IV Device/System Characteristics: ### A Device Description: Access CEA Reagent Kit Each Access CEA reagent kit contains two reagent packs (50 tests/pack). Each pack contains the following: - Solid phase: Paramagnetic particles coated with mouse anti-CEA MAb, suspended in TRIS buffered bovine serum albumin (BSA), with < 0.1% sodium azide and 0.1% ProClin 300 - Diluent: Phosphate buffer, protein (bovine, murine) with < 0.1% sodium azide and 0.1% ProClin 300 - Conjugate: Mouse anti-CEA MAb bound to alkaline phosphatase (bovine), diluted in phosphate buffer, protein (bovine), < 0.1% sodium azide and 0.1% ProClin 300 {2} Materials needed but not supplied with reagent kit - Access CEA Calibrators: at zero and approximately 1, 10, 100, 500 and 1,000 ng/mL - Access CEA QC controls: at approximately 3 and 300 ng/mL - Commercial control material - Lumi-Phos PRO substrate - UniCel DxI Wash Buffer II - Access CEA Diluent (optional) The modification of the Access CEA includes: (a) to replace substrate (Lumi-Phos 530) of the Access CEA with a new substrate (Lumi-Phos PRO) (b) to run on DxI 9000 Access Immunoassay Analyzer with the update the analytical measuring interval of 0.2 – 1,000 ng/mL ## B Principle of Operation: The Access CEA assay is a two-site immunoenzymatic “sandwich” assay using two mouse monoclonal anti-CEA antibodies (MAb) which react with different epitopes of CEA. A sample is added to a reaction vessel, along with the first anti-CEA MAb-alkaline phosphatase conjugate and the second anti-CEA MAb bound to paramagnetic particles. After incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is directly proportional to the concentration of analyte in the sample. Analyte concentration is automatically determined from a stored calibration. ## V Substantial Equivalence Information: ### A Predicate Device Name(s): Access CEA ### B Predicate 510(k) Number(s): K981985 K991707 ### C Comparison with Predicate(s): K223921 - Page 3 of 9 {3} K223921 - Page 4 of 9 | Device & Predicate Device(s): | K223921 Candidate Device | K991707 Predicate | K981985 Predicate | | --- | --- | --- | --- | | Device Trade Name | Access CEA | Access CEA | | | **General Device Characteristic Similarities** | | | | | Intended Use/Indications For Use | The Access CEA assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of Carcinoembryonic Antigen (CEA) levels in human serum, using the Access Immunoassay Systems. CEA measured by the Access Immunoassay Systems is used as an aid in the management of cancer patients *in whom changing CEA concentrations have been observed*. | Same | The Access CEA Immunoassay is a paramagnetic particle chemiluminescent immunoassay for the quantitative determination of carcinoembryonic antigen (CEA) in human serum using the ACCESS Immunoassay System. CEA measured by the ACCESS CEA Immunoassay, is used as an aid in the management of cancer patients. | | Measurand | Carcinoembryonic antigen | Same | Same | | Sample Type | Human serum | Same | Same | | Assay Technology | Chemiluminescent | Same | Same | | Measurement | Quantitative | Same | Same | | Antibody | Monoclonal (mouse) anti-human CEA antibodies | Same | Same | | Operating Principle | One-step sandwich | Same | Same | | Calibrators | Multi-point calibrators containing purified human CEA | Same | Same | | QC Controls | Bi-level controls containing human CEA | Same | Same | | Reagent Stability | Stable at 2 to 10°C for 28 days after initial use | Same | Same | | Method | Automated | Same | Same | | **General Device Characteristic Differences** | | | | | Instrument | DxI 9000 Access Immunoassay Analyzer | Access 2 Immunoassay System; UniCel DxI 800 Immunoassay System; UniCel DxI 600 Immunoassay System | | | Substrate | Lumi-Phos PRO substrate | Access Substrate | | | Measuring Range | 0.2 – 1,000 ng/mL | 0.1 – 1,000 ng/mL | | {4} VI Standards/Guidance Documents Referenced: The following Clinical and Laboratory Standards Institute (CLSI) guidelines were used: - CLSI EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline—Third Edition - CLSI EP06 2nd Edition, Evaluation of Linearity of Quantitative, Measurement Procedures - CLSI EP09c 3rd Edition, Measurement Procedure Comparison and Bias Estimation Using Patient Samples - CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline– Second Edition VII Performance Characteristics (if/when applicable): A Analytical Performance: All results presented below met the manufacturer’s pre-determined acceptance criteria. 1. Precision/Reproducibility: Precision testing was performed in accordance with CLSI guideline EP05-A3. a) Within-Laboratory Precision: The studies were performed at a single site using one lot of the modified Access CEA reagent on one DxI 9000 Access Immunoassay Analyzer. Seven levels of human serum samples were run in a minimum of three replicates per run, two runs daily over the course of 20 days, resulting a minimum of 120 datapoints for each sample. The data were analyzed for repeatability (within-run), between-run, between-day, and within-laboratory precision. The mean (ng/mL), standard deviation (SD) (ng/mL) and percent coefficient of variation (%CV) are summarized in table below. | Sample | Mean (ng/mL) | N | Within-Run | | Between-Run | | Between-Day | | Within-Laboratory | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | 1 | 0.5 | 126 | 0.0 | 3.5 | 0.0 | 1.5 | 0.0 | 2.0 | 0.0 | 4.3 | | 2 | 5.1 | 120 | 0.1 | 2.1 | 0.0 | 0.8 | 0.1 | 1.2 | 0.1 | 2.5 | | 3 | 11.0 | 126 | 0.2 | 1.9 | 0.3 | 2.5 | 0.2 | 1.5 | 0.4 | 3.5 | | 4 | 89.0 | 120 | 1.9 | 2.2 | 0.8 | 0.9 | 1.1 | 1.3 | 2.4 | 2.7 | | 5 | 153 | 120 | 2.8 | 1.8 | 1.9 | 1.3 | 2.3 | 1.5 | 4.1 | 2.7 | | 6 | 525 | 120 | 18.5 | 3.5 | 6.6 | 1.2 | 16.0 | 3.1 | 25.3 | 4.8 | | 7 | 865 | 120 | 22.4 | 2.6 | 5.3 | 0.6 | 38.6 | 4.5 | 45.0 | 5.2 | K223921 - Page 5 of 9 {5} b) Lot-to-Lot Precision: The study was performed at a single site using three modified Access CEA with three lots of Lumi-Phos PRO Substrate, three Access CEA Calibrator lots on two instruments. Six levels of human serum samples were run in five replicates per run, one run per day for two days, resulting a total of 180 datapoints for each sample. The data were analyzed for the repeatability (within-run), between-day, between-substrate lot, between-calibrator lot and total precision. The results are summarized in table below. | Sample | Mean (ng/mL) | N | Within-Run | | Between-Day | | Between-Substrate Lot | | Between-Calibrator Lot | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | 1 | 1.0 | 177 | 0.0 | 0.0 | 0.0 | 3.3 | 0.0 | 2.1 | 0.0 | 0.0 | 0.0 | 4.1 | | 2 | 6.0 | 180 | 0.1 | 1.9 | 0.1 | 1.8 | 0.0 | 0.3 | 0.2 | 2.6 | 0.2 | 3.9 | | 3 | 77.0 | 180 | 1.4 | 1.8 | 1.0 | 1.3 | 1.4 | 1.8 | 1.1 | 1.5 | 2.5 | 3.3 | | 4 | 322 | 180 | 6.1 | 1.9 | 4.8 | 1.5 | 2.3 | 0.7 | 3.7 | 1.1 | 11.0 | 3.4 | | 5 | 609 | 180 | 12.7 | 2.1 | 13.2 | 2.2 | 11.0 | 1.8 | 4.6 | 0.7 | 22.0 | 3.6 | | 6 | 841 | 180 | 15.6 | 1.9 | 17.7 | 2.1 | 10.5 | 1.2 | 7.0 | 0.8 | 26.7 | 3.2 | * 3 data point were excluded due to 'instrument error' c) Instrument-to-Instrument Reproducibility: The study was performed using three DxI 9000 Access Immunoassay Analyzers with a single lot of the modified Access CEA reagent. Five levels of serum samples were run in five replicates per run, one run per day over the course of five days, resulting a total of 75 datapoints for each sample. The results are summarized in table below: | Sample | Mean (ng/mL) | N | Within-Run (Repeatability) | | Between-Day | | Between-Instrument | | Reproducibility | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | 1 | 5.4 | 75 | 0.1 | 1.9 | 0.1 | 1.9 | 0.0 | 0.7 | 0.2 | 2.8 | | 2 | 95.0 | 75 | 1.8 | 1.9 | 0.8 | 0.8 | 1.5 | 1.5 | 2.4 | 2.5 | | 3 | 215 | 75 | 5.9 | 2.7 | 2.5 | 1.1 | 2.9 | 1.3 | 7.0 | 3.3 | | 4 | 475 | 75 | 10.8 | 2.3 | 6.6 | 1.4 | 5.1 | 1.1 | 13.6 | 2.9 | | 5 | 828 | 75 | 22.3 | 2.7 | 47.4 | 5.7 | 0.0 | 0.0 | 52.4 | 6.3 | 2. Linearity: Linearity of the modified Access CEA was performed in accordance with CLSI guideline EP06 2nd Edition. Two series of samples were prepared by mixing two High samples with Low sample span the 'analytical measuring interval' (AMI) of the modified Access CEA and to provide eight dilution levels. Low sample was obtained from a single human serum sample that was specifically stripped of CEA antigen to achieve a concentration below the low end of the AMI of the assay. Two High samples were pooled human serum with CEA concentrations of $38.63\mathrm{ng / mL}$ and $1042.74\mathrm{ng / mL}$ . The low sample in the two series was run in replicates of eight, and all other dilution levels were run in replicates of four, and the 'measured value' (mean CEA ng/mL) was compared to its predicted value for deviation percentage using a weighted linear regression analysis. This deviation was then compared to the allowable deviation from linearity (ADL). The results are summarized in table below. K223921 - Page 6 of 9 {6} | Sample | Range (ng/mL) | Slope (95% CI) | Intercept (95% CI) | R² | % Deviation* (Deviation**) | | --- | --- | --- | --- | --- | --- | | 1 | 0.1 – 38.6 | 0.952 (0.936 – 0.967) | 0.003 (-0.009 – 0.015) | 0.999 | -7% – 5% (0.001) | | 2 | 0.1 – 1042.7 | 0.966 (0.950 – 0.983) | 0.003 (-0.006 – 0.012) | 0.999 | -9% – 3% (0.00003) | | Combined 1 and 2 | 0.1 – 1042.7 | 0.959 (0.949 – 0.969) | 0.003 (-0.003 – 0.010) | 1.000 | -8% – 4% (0.000 – 0.001) | * % Deviation from Linearity for concentrations > 2.1 ng/mL ** Deviation (ng/mL) from Linearity for concentrations ≤ 2.1 ng/mL The data support the linearity interval from 0.1 to 1042.7 ng/mL with the deviations from linearity within ±10%. The study results support the linearity of the claimed analytical measuring interval (AMI): 0.2 – 1,000 ng/mL. 3. Assay Reportable Range: The reportable range is the same as the claimed measuring range: 0.2 to 1,000 ng/mL. 4. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): a) Traceability: The traceability was established in K991707. b) Stability: The Access CEA Reagent kit and the Lumi-Phos PRO substrate are packaged separately. The proposed change is to replace substrate (Lumi-Phos 530) of the Access CEA with a new substrate (Lumi-Phos PRO). The shelf-life claims and on-board stability claims of Lumi-Phos PRO substrate were established in K221225. Access CEA Reagent kit stability was established in K991707. The real-time stability study was conducted to verify the stability claims for the modified Access CEA. The results support the shelf-life of 12 months for the reagent pack stored at 2-10°C. 5. Detection Capability: CLSI guideline EP17-A2 were followed to determine the Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantitation (LoQ) for the modified Access CEA on DxI 9000 Access Immunoassay Analyzer. LoB: Four blank samples (individual native serum samples from which CEA antigen was specifically depleted) were tested in five replicates per run, one run per day over three days using two modified Access CEA reagent lots (N = 60 replicates for each lot) on two DxI 9000 Access Immunoassay Analyzers. The LoB for each reagent lot was calculated using the non-parametric approach and the maximum observed LoB is taken as the reported value for K223921 - Page 7 of 9 {7} the measurement procedure. LoB determined using the 95% non-parametric percentile of the replicates for each of the two reagent lots was 0.09 ng/mL and 0.03 ng/mL. The claimed LoB for the modified Access CEA on DxI 9000 Access Immunoassay Analyzer is 0.09 ng/mL. **LoD:** Five to seven serum samples containing low levels of CEA analyte were run in nine replicates per run, one run per day over five days using three modified Access CEA reagent lots (N = minimum 225 replicates for each lot) on three DxI 9000 Access Immunoassay Analyzers. The LoD was determined by fitting the precision profile model between within-laboratory SD and concentration. LoD calculated from LoB + "SD multiplied by the 95th percentile of the standard normal distribution" resulted in 0.1 ng/mL for each reagent lot. The claimed LoD for the modified Access CEA on DxI 9000 Access Immunoassay Analyzer is 0.1 ng/mL. **LoQ:** Nine to eleven serum samples containing low levels of CEA analyte were run in nine replicates of per run, one run per day over five days using three modified Access CEA reagent lots (N=minimum 405 replicates for each lot) on three DxI 9000 Access Immunoassay Analyzers. A variance components model was used to estimate the within-run and within-laboratory (total) %CV for each sample on each instrument and reagent lot combination. The LoQ was set to be the CEA concentration which met the within-laboratory imprecision of 20% CV. The claimed LoQ for the modified Access CEA on DxI 9000 Access Immunoassay Analyzer is 0.2 ng/mL. 6. **Assay Cut-Off:** Not applicable **B Comparison Studies:** 1. **Method Comparison with Predicate Device:** CLSI guideline EP09c 3rd Edition were followed to compare the modified Access CEA on DxI 9000 Access Immunoassay Analyzer to the predicate device, the Access CEA on the Access 2 Immunoassay System. Patient samples falling within the analytical measuring interval of the Access CEA assay were evaluated. A total of 153 deidentified serum samples were tested with three Access CEA reagent lots and three calibrator lots, using one Lumi-Phos PRO substrate lots on three DxI 9000 Access Immunoassay Analyzers (candidate) and using one Access Substrate lots on three Access 2 instruments (predicate). Two commercial quality controls were run in duplicate each day to verify the systems were in control. The comparison between paired measurements was analyzed using Passing-Bablok method by fitting the observed modified Access CEA Assay on DxI 9000 Access Immunoassay Analyzer (dependent variable, y) into a linear regression model, with the observed values from the predicate (x, predicate). The results are summarized in the following table: | N | Range (ng/mL) | Slope (95% CI) | Intercept (95% CI) | R | | --- | --- | --- | --- | --- | | 153 | 0.46 – 1071 | 0.98 (0.97 – 0.99) | 0.06 (0.00 – 0.17) | 1.00 | K223921 - Page 8 of 9 {8} 2. Matrix Comparison: Not applicable. The assay is for serum samples only. C Clinical Studies: Refer to K981985 D Clinical Cut-Off: No cutoff for CEA monitoring has been recommended. E Expected Values/Reference Range: Refer to K981985 VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. K223921 - Page 9 of 9
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...