EUROIMMUN EUROLINE PROFILE AUTOIMMUNE LIVER DISEASE 8AG (IGG)

K113439 · Euroimmun US · NIY · Apr 12, 2013 · Immunology

Device Facts

Record IDK113439
Device NameEUROIMMUN EUROLINE PROFILE AUTOIMMUNE LIVER DISEASE 8AG (IGG)
ApplicantEuroimmun US
Product CodeNIY · Immunology
Decision DateApr 12, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5660
Device ClassClass 2

Indications for Use

The EUROLINE Profile Autoimmune Liver Disease 8 Ag (IgG) Kit is an immune line-blot strip test intended for the qualitative detection of IgG class antibodies against 8 different antigens: AMA-M2, M2-3E (BPO), Sp100, PML, gp210, LKM-1, LC-1 and SLA/LP in human serum and plasma (EDTA, Li-heparin, Citrate). Detection of these antibodies is used as an aid in the diagnosis of autoimmune liver diseases in conjunction with other laboratory and clinical findings.

Device Story

The EUROLINE Profile Autoimmune Liver Disease 8 Ag (IgG) is an immunoblot test for detecting autoantibodies in human serum or plasma. The device uses antigen-coated membrane strips as the solid phase. Patient samples are incubated on the strips; if specific antibodies are present, they bind to the antigens. An alkaline phosphatase-labeled anti-human IgG conjugate is added, followed by a chromogenic substrate (NBT/BCIP). This produces a visible color reaction at the antigen-specific band site. The test is performed in a clinical or hospital laboratory by trained personnel. Results are evaluated visually by comparing band intensity against a reaction control card. The presence of specific antibodies aids clinicians in diagnosing autoimmune liver diseases, such as primary biliary cirrhosis or autoimmune hepatitis, when combined with other clinical and laboratory findings. The device provides qualitative results for each antibody separately.

Clinical Evidence

Clinical validation involved 734 samples from various patient groups (PBC, AIH, viral hepatitis, PSC, and other controls). Sensitivity for PBC (AMA-M2) was 85.4% (95% CI: 79.8-89.9%). Specificity for PBC across all antigens ranged from 97.2% to 99.8%. For AIH, sensitivity for LKM-1 was 12.3% (overall) and 85.7% (Type 2 AIH). Specificity for AIH across LKM-1, LC-1, and SLA/LP ranged from 98.8% to 99.8%. Method comparison with predicate ELISAs showed high agreement; discrepant samples were primarily from PBC patients. No positive results were found in 150 healthy blood donors.

Technological Characteristics

Qualitative immunoblot (lineblot) assay. Uses purified/recombinant antigens (AMA-M2, M2-3E, Sp100, PML, gp210, LKM-1, LC-1, SLA/LP) coated on membrane strips. Detection via alkaline phosphatase-labeled anti-human IgG conjugate and NBT/BCIP chromogenic substrate. Manual procedure: sample incubation, wash, conjugate incubation, wash, substrate incubation, and visual evaluation. No electronic components or software algorithms; evaluation is visual based on band intensity relative to a control card.

Indications for Use

Indicated for qualitative detection of IgG antibodies against 8 liver-related antigens (AMA-M2, M2-3E, Sp100, PML, gp210, LKM-1, LC-1, SLA/LP) in human serum/plasma to aid diagnosis of autoimmune liver diseases. For prescription use only.

Regulatory Classification

Identification

A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k113439 B. Purpose for Submission: New device C. Measurand: IgG antibodies to AMA-M2, M2-3E (BPO), Sp100, PML, gp210, LKM-1, LC-1, SLA/LP D. Type of Test: Manual and automated read-out, qualitative immunoblot assay E. Applicant: EUROIMMUN US Inc. F. Proprietary and Established Names: EUROLINE Profile Autoimmune Liver Disease 8 Ag (IgG) Kit G. Regulatory Information: 1. Regulation section: 21CFR§866.5090 – Antimitochondrial antibody immunological test system 2 CFR§866.5660 – Multiple autoantibodies immunological test system 2. Classification: Class II 3. Product code: DBM - Antimitochondrial antibody, indirect immunofluorescent, antigen, control NUM - Autoantibodies, nuclear body protein, Sp100 NRI - Autoantibodies, nuclear pore glycoprotein gp210 NIY - Autoantibodies, anti-soluble liver antigen (SLA), autoimmune hepatitis NBS - Autoantibodies, LKM-1(liver/kidney microsome, type 1) 4. Panel: Immunology (82) H. Intended Use: 1. Intended use(s): The EUROLINE Profile Autoimmune Liver Disease 8 Ag (IgG) Kit is an immune line-blot strip test intended for the qualitative detection of IgG class antibodies against 8 different antigens: AMA-M2, M2-3E (BPO), Sp100, PML, gp210, LKM-1, LC-1 and SLA/LP in human serum and plasma (EDTA, Li-heparin, Citrate). {1} Detection of these antibodies is used as an aid in the diagnosis of autoimmune liver diseases in conjunction with other laboratory and clinical findings. 2. Indication(s) for use: Same as intended use 3. Special conditions for use statement(s): For Prescription use only. 4. Special instrument requirements: CanoScan LiDE Series flatbed scanner using ScanGear and EUROLineScan softwares for automated read. # I. Device Description: The EUROLINE Profile Autoimmune Liver Disease 8 Ag (IgG) Kit consists of antigen coated line blot strips, a positive control (100x concentrate), biotin-ExtrAvidin-alkaline phosphatase-labeled goat anti-human IgG conjugate (10X concentrate), sample buffer (ready-to-use), wash buffer (10X concentrate), Nitrobluetetrazoliumchloride/5-Bromo-4-chloro-3-indolylphosphate (NBT/BCIP) substrate solution (ready-to-use), incubation tray and test instruction. # J. Substantial Equivalence Information: 1. Predicate device name(s) and 510(k) number(s): | EUROLINE Profile Autoimmune Liver Disease Autoantibodies | Predicate device | 510(k) number | | --- | --- | --- | | Anti-AMA-M2/M2-3E (BPO) | Quanta Lite M2 EP (MIT3) ELISA | k052262 | | Anti-Sp100 and PML | Quanta Lite Sp100 ELISA | k050662 | | Anti-gp210 | Quanta Lite gp210 ELISA | k040885 | | Anti-LKM-1 and LC-1 | Quanta Lite LKM-1 ELISA | k000535 | | Anti-SLA/LP | Quanta Lite SLA ELISA | k021482 | 2. Comparison with predicate: | Similarities | | | | | --- | --- | --- | --- | | Item | Device | | Predicate | | Intended Use | Qualitative detection of IgG antibodies against 8 different antigens complete to aid in the diagnosis of autoimmune liver diseases. | | Same (when combined) | | Capture antigens | Sp100 | Recombinant Sp100, expressed by cloning the corresponding human cDNA in E.coli | Same | | | gp210 | Recombinant gp210, expressed by cloning the corresponding | Same | | | | human cDNA in E.coli | | {2} | Similarities | | | | | --- | --- | --- | --- | | Item | Device | | Predicate | | | | human cDNA in E.coli | | | | LKM-1 | Recombinant cytochrome P450 IID6, expressed by cloning the corresponding human cDNA in insect cells using a baculovirus vector. | Same | | | SLA/LP | Recombinant SLA/LP, expressed by cloning the corresponding human cDNA in E.coli | Same | | Detection antibody | Goat anti-human IgG | | Same | | Sample dilution | 1:101 | | Same | | Differences | | | | | --- | --- | --- | --- | | Item | Device | | Predicate | | Assay format | Qualitative, positive/negative | | Semi-quantitative | | Solid phase | Membrane test strips | | Polystyrene microwells | | Instrument | Manual visual readout | | Spectrophotometer | | Capture antigens | AMA-M2 | Natively purified from bovine heart | Affinity purified recombinant M2 EP MIT3. | | | M2-3E (BPO) | Recombinant fusion protein, produced in E.coli | | | | PML | Recombinant PML, expressed by cloning the corresponding human cDNA in E.coli | Not included | | | LC-1 | Recombinant LC-1 expressed by cloning the corresponding human cDNA in insect cells using a baculovirus vector. | Not included | | Sample type | Serum or plasma (EDTA, Li-heparin, Citrate) | | Serum | | Controls | One positive control, 100X concentrate | | 3 controls (high positive, low positive, negative) | | Conjugate | biotin-ErxtAvidin-alkaline phosphatase, 10X concentrate | | horseradish peroxidase | | Substrate | NBT/BCIP | | 3,5',5',5' Tetramethylbenzidine | | Interpretation of results | positive/negative compared to reaction control card | | Units | {3} K. Standard/Guidance Document Referenced (if applicable): DIN EN 13640:2002: Stability testing of in vitro diagnostic reagents L. Test Principle: The principle of the EUROLINE Profile Autoimmune Liver Disease 8 Ag (IgG) Kit is that of an enzyme linked immunosorbent assay (ELISA), using a membrane as the solid phase instead of microtiter wells. Different purified antigens have been coated and applied in easy to read lines (bands) onto a membrane strip. Autoantibodies in patient samples bind to the bands and are detected via a secondary antibody linked to an enzyme. The strips are evaluated visually by comparison of the band intensity with the reaction control card or scanned and then evaluated with EUROLineScan. The control band on the strips contains (non-specific) anti-human IgG, which reacts with the sample IgG to show a color reaction if the incubation was performed correctly and so represents a function test on each single strip. The positive control contains a mixture of the targeted antibodies which bind to the antigen coated on the blot strips. A strip incubated with the positive control shows a positive result. If either the control band or the strip incubated with the positive control is negative, test results are invalid and should be repeated. The qualitative results are reported for each individual antibody separately as positive or negative. The intensity of the reaction is not reported but only as a means to distinguish between a positive and negative reaction and not as an indication of disease status. The interpretation of the test results does not include a combined score or diagnosis. M. Performance Characteristics (if/when applicable): 1. Analytical performance: Test results were evaluated by visual or automated reading of the test strips. For visual evaluation, the reaction intensity of each antigen band on the test strip was compared to the two color bars on the reaction control card that correspond to the intensities just below (negative) and just above (positive) the cut-off. For automated evaluation, the test strips are scanned using a flatbed scanner and evaluated with EUROLineScan with a cut-off of 11 grey scale units. a. Precision/Reproducibility: i. Visual evaluation. Assay reproducibility was determined by testing 5 to 9 samples that cover the complete range of results (negative, positive and near to cut-off) for each antigen (see table below). The intra-assay reproducibility is based on 20 replicates tested in one day and the inter-assay reproducibility on 20 different runs, each run performed by the same reader on a different day. The lot-to-lot reproducibility was tested in 3 different runs using 3 different lots. The reproducibility data showed no positive sample was 4 {4} found negative and vice versa. The inter-reader reproducibility of 3 samples that cover the complete range of results (negative, positive and near to cut-off) for each antigen (see table below) was evaluated by 3 different technicians and under 3 different light conditions (sunlight, neon light and electric bulb light). No deviation was observed between the individual readings and from the light conditions. Intra-assay reproducibility | Antigen band | Sample characterization | Sample Number | Visual evaluation result | | | --- | --- | --- | --- | --- | | | | | % pos | % neg | | AMA-M2 | pos | 2 | 100% | 0% | | | neg (near to cut-off) | 1 | 0% | 100% | | | neg | 2 | 0% | 100% | | M2-3E | pos | 2 | 100% | 0% | | | neg (near to cut-off) | 1 | 0% | 100% | | | neg | 2 | 0% | 100% | | Sp100 | pos | 1 | 100% | 0% | | | neg (near to cut-off) | 1 | 0% | 100% | | | neg | 3 | 0% | 100% | | PML | pos | 1 | 100% | 0% | | | pos (near to cut-off) | 1 | 100% | 0% | | | neg | 3 | 0% | 100% | | gp210 | pos | 1 | 100% | 0% | | | pos (near to cut-off) | 1 | 100% | 0% | | | neg | 3 | 0% | 100% | | LKM-1 | pos | 2 | 100% | 0% | | | neg (near to cut-off) | 1 | 0% | 100 | | | neg | 2 | 0% | 100% | | LC-1 | pos | 1 | 100% | 0% | | | neg (near to cut-off) | 1 | 0% | 100% | | | neg | 3 | 0% | 100% | | SLA/LP | pos | 1 | 100% | 0% | | | neg (near to cut-off) | 1 | 0% | 100% | | | neg | 3 | 0% | 100% | Inter-assay reproducibility | Antigen band | Sample characterization | Sample Number | Visual evaluation result | | | --- | --- | --- | --- | --- | | | | | % pos | % neg | | AMA-M2 | pos | 2 | 100% | 0% | | | pos (near to cut-off) | 3 | 100% | 0% | | | neg (near to cut-off) | 1 | 40% | 60% | | | neg (near to cut-off) | 1 | 55% | 45% | | | neg | 2 | 0% | 100% | | M2-3E | pos | 2 | 100% | 0% | | | pos (near to cut-off) | 1 | 100% | 0% | | | neg (near to cut-off) | 3 | 95% | 5% | | | neg (near to cut-off) | 1 | 90% | 10% | | | neg | 2 | 0% | 100% | {5} 6 | Antigen band | Sample characterization | Sample Number | Visual evaluation result | | | --- | --- | --- | --- | --- | | | | | % pos | % neg | | Sp100 | pos | 2 | 100% | 0% | | | pos (near to cut-off) | 5 | 100% | 0% | | | neg | 2 | 0% | 100% | | PML | pos | 1 | 100% | 0% | | | pos (near to cut-off) | 1 | 100% | 0% | | | neg (near to cut-off) | 2 | 80% | 20% | | | neg (near to cut-off) | 2 | 75% | 25% | | | neg | 2 | 0% | 100% | | gp210 | pos | 2 | 100% | 0% | | | pos (near to cut-off) | 1 | 100% | 0% | | | neg (near to cut-off) | 1 | 45% | 55% | | | neg (near to cut-off) | 1 | 40% | 60% | | | neg | 2 | 0% | 100% | | LKM-1 | pos | 1 | 100% | 0% | | | pos (near to cut-off) | 3 | 100% | 0% | | | neg (near to cut-off) | 2 | 80% | 20% | | | neg | 2 | 0% | 100% | | LC-1 | pos | 1 | 100% | 0% | | | pos (near to cut-off) | 1 | 100% | 0% | | | neg (near to cut-off) | 1 | 90% | 10% | | | neg (near to cut-off) | 1 | 80% | 20% | | | neg (near to cut-off) | 1 | 75% | 25% | | | neg | 2 | 0% | 100% | | SLA/LP | pos | 1 | 100% | 0% | | | pos (near to cut-off) | 1 | 100% | 0% | | | neg (near to cut-off) | 2 | 55% | 45% | | | neg (near to cut-off) | 1 | 45% | 55% | | | neg (near to cut-off) | 1 | 40% | 60% | | | neg | 2 | 0% | 100% | Inter-reader reproducibility: | Antigen band | Sample characterization | Sample Number | Visual evaluation result | | | --- | --- | --- | --- | --- | | | | | % pos | % neg | | AMA-M2 | pos | 2 | 100% | 0% | | | neg (near to cut-off) | 1 | 0% | 100% | | | neg | 2 | 0% | 100% | | M2-3E | pos | 3 | 100% | 0% | | | neg | 2 | 0% | 100% | | Sp100 | pos | 2 | 100% | 0% | | | neg (near to cut-off) | 1 | 0% | 100% | | | neg | 2 | 0% | 100% | | PML | pos | 2 | 100% | 0% | | | neg | 3 | 0% | 100% | | gp210 | pos | 2 | 100% | 0% | | | neg | 3 | 0% | 100% | | LKM-1 | pos | 2 | 100% | 0% | | | neg (near to cut-off) | 1 | 0% | 100% | | | neg | 2 | 0% | 100% | {6} 7 | Antigen band | Sample characterization | Sample Number | Visual evaluation result | | | --- | --- | --- | --- | --- | | | | | % pos | % neg | | LC-1 | pos | 2 | 100% | 0% | | | neg (near to cut-off) | 1 | 0% | 100% | | | neg | 2 | 0% | 100% | | SLA/LP | pos | 1 | 100% | 0% | | | neg (near to cut-off) | 1 | 0% | 100% | | | neg | 3 | 0% | 100% | ii. Automated evaluation. The reproducibility of the automatic evaluation was investigated by repeated scanning and analysis of test strips incubated with 16 different serum samples displaying signals for different antibodies near the cut off. The processed test strips were scanned using three different flatbed scanners, each scanner was used 10 times scanning the same strips and the results were evaluated by EUROLineScan. The scanner intensity values of the 30 analyses were used to calculate coefficients of variation (CV's). The results are shown in the tables below. Samples near the cut-off (n = 3 x 10 = 30) | Sample | EUROLineScan Intensity values | | | | --- | --- | --- | --- | | | Mean value | SD | % CV | | AMA-M2 | 13 | 0.7 | 5.5 | | M2-3E | 14 | 0.4 | 2.7 | | Sp100 | 12 | 0.4 | 3.1 | | PML | 13 | 0.4 | 3.4 | | gp210 | 12 | 0.0 | 0.0 | | LKM-1 | 14 | 0.5 | 3.9 | | LC-1 | 12 | 0.2 | 1.5 | | SLA/LP | 14 | 0.7 | 4.9 | Positive samples (n = 3 x 10 = 30) | Sample | EUROLineScan Intensity values | | | | --- | --- | --- | --- | | | Mean value | SD | % CV | | AMA-M2 | 69 | 0.7 | 1.0 | | M2-3E | 37 | 0.4 | 1.0 | | Sp100 | 47 | 0.6 | 1.3 | | PML | 66 | 0.4 | 0.7 | | gp210 | 51 | 0.7 | 1.3 | | LKM-1 | 73 | 0.6 | 0.8 | | LC-1 | not available | | | | SLA/LP | 39 | 0.4 | 0.9 | iii. Comparability of visual and automatic reading The use of the EUROLineScan software compared to visual reading was investigated using a panel of serum sample preparations covering the whole range of antigens. The samples were tested in 20 different runs, each run on a different day using the same kit lot. Each run was performed by the same technician and evaluation was performed both {7} visually using the reaction control card and automatically using the EUROLineScan software with a cut-off of 11 grey scale units. The results are shown in the table below. Only a few results from sample preparations exactly at the cut-off gave discrepant results between visual and automatic reading. | Antigen | Sample Characterization | Visual evaluation result | | EUROLineScan result | | Grey scale units result | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | % pos | % neg | % pos | % neg | Mean | Range | | AMA-M2 | pos (near to cut-off) | 100% | 0% | 100% | 0% | 15 | 13 - 18 | | | pos | 100% | 0% | 100% | 0% | 68 | 60 - 76 | | | Neg | 0% | 100% | 0% | 100% | 1 | 0 - 1 | | | neg | 0% | 100% | 0% | 100% | 1 | 0 - 2 | | | pos | 100% | 0% | 100% | 0% | 54 | 48 - 58 | | | pos (near to cut-off) | 100% | 0% | 100% | 0% | 13 | 11 - 15 | | | pos (near to cut-off) | 100% | 0% | 100% | 0% | 14 | 11 - 16 | | | neg (near to cut-off) | 40% | 60% | 40% | 60% | 10 | 6 - 12 | | | neg (near to cut-off) | 55% | 45% | 55% | 45% | 11 | 8 - 12 | | M2-3E | pos (near to cut-off) | 100% | 0% | 100% | 0% | 18 | 14 - 23 | | | pos | 100% | 0% | 100% | 0% | 47 | 38 - 57 | | | neg | 0% | 100% | 0% | 100% | 2 | 2 - 3 | | | neg | 0% | 100% | 0% | 100% | 3 | 2 - 4 | | | pos | 100% | 0% | 100% | 0% | 40 | 32 - 45 | | | pos (near to cut-off) | 95% | 5% | 95% | 5% | 13 | 10 - 15 | | | neg (near to cut-off) | 95% | 5% | 95% | 5% | 12 | 10 - 15 | | | neg (near to cut-off) | 90% | 10% | 90% | 10% | 12 | 10 - 15 | | | pos (near to cut-off) | 95% | 5% | 95% | 5% | 13 | 10 - 15 | | Sp100 | pos | 100% | 0% | 100% | 0% | 45 | 38 - 49 | | | pos (near to cut-off) | 100% | 0% | 100% | 0% | 14 | 11 - 17 | | | neg | 0% | 100% | 0% | 100% | 0 | 0 - 1 | | | neg | 0% | 100% | 0% | 100% | 0 | 0 - 1 | | | pos | 100% | 0% | 100% | 0% | 86 | 75 - 96 | | | pos (near to cut-off) | 100% | 0% | 100% | 0% | 14 | 11 - 17 | | | pos (near to cut-off) | 100% | 0% | 100% | 0% | 14 | 11 - 17 | | | pos (near to cut-off) | 100% | 0% | 100% | 0% | 14 | 11 - 18 | | | pos (near to cut-off) | 100% | 0% | 100% | 0% | 14 | 11 - 19 | | PML | pos (near to cut-off) | 100% | 0% | 100% | 0% | 16 | 12 - 20 | | | pos | 100% | 0% | 100% | 0% | 68 | 62 - 78 | | | neg | 0% | 100% | 0% | 100% | 1 | 1 - 2 | | | neg | 0% | 100% | 0% | 100% | 1 | 0 - 2 | | | neg (near to cut-off) | 75% | 25% | 75% | 25% | 12 | 8 - 14 | | | neg (near to cut-off) | 75% | 25% | 75% | 25% | 12 | 10 - 14 | | | neg (near to cut-off) | 80% | 20% | 80% | 20% | 12 | 7 - 14 | | | neg (near to cut-off) | 80% | 20% | 80% | 20% | 11 | 10 - 13 | | gp210 | pos (near to cut-off) | 100% | 0% | 100% | 0% | 13 | 11 - 15 | | | pos | 100% | 0% | 100% | 0% | 53 | 46 - 58 | | | neg | 0% | 100% | 0% | 100% | 1 | 1 - 2 | | | neg | 0% | 100% | 0% | 100% | 0 | 0 - 1 | | | pos | 100% | 0% | 100% | 0% | 135 | 131 - 140 | | | neg (near to cut-off) | 45% | 55% | 45% | 55% | 11 | 8 - 12 | | | neg (near to cut-off) | 40% | 60% | 40% | 60% | 10 | 9 - 12 | | | neg (near to cut-off) | 45% | 55% | 45% | 55% | 11 | 9 - 13 | | | neg (near to cut-off) | 40% | 60% | 40% | 60% | 11 | 9 - 13 | {8} | Antigen | Sample Characterization | Visual evaluation result | | EUROLineScan result | | Grey scale units result | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | % pos | % neg | % pos | % neg | Mean | Range | | LKM-1 | pos (near to cut-off) | 100% | 0% | 100% | 0% | 13 | 12 - 16 | | | pos | 100% | 0% | 100% | 0% | 71 | 63 - 79 | | | neg | 0% | 100% | 0% | 100% | 1 | 0 - 2 | | | neg | 0% | 100% | 0% | 100% | 1 | 0 - 1 | | | pos (near to cut-off) | 100% | 0% | 95% | 5% | 14 | 10 - 17 | | | pos (near to cut-off) | 100% | 0% | 100% | 0% | 13 | 12 - 16 | | | neg (near to cut-off) | 80% | 20% | 80% | 20% | 11 | 10 - 14 | | | neg (near to cut-off) | 80% | 20% | 80% | 20% | 12 | 10 - 14 | | LC-1 | pos | 100% | 0% | 100% | 0% | 52 | 36 - 59 | | | pos | 100% | 0% | 100% | 0% | 98 | 89 - 104 | | | pos (near to cut-off) | 100% | 0% | 100% | 0% | 12 | 11 - 16 | | | neg | 0% | 100% | 0% | 100% | 1 | 0 - 2 | | | neg | 0% | 100% | 0% | 100% | 0 | 0 - 2 | | | neg (near to cut-off) | 55% | 45% | 40% | 60% | 11 | 8 - 17 | | | neg (near to cut-off) | 90% | 10% | 90% | 10% | 12 | 10 - 16 | | | neg (near to cut-off) | 75% | 25% | 75% | 25% | 12 | 10 - 14 | | | neg (near to cut-off) | 80% | 20% | 80% | 20% | 12 | 10 - 14 | | SLA/LP | pos (near to cut-off) | 100% | 0% | 100% | 0% | 16 | 13 - 18 | | | pos | 100% | 0% | 100% | 0% | 41 | 36 - 45 | | | neg | 0% | 100% | 0% | 100% | 1 | 0 - 1 | | | neg | 0% | 100% | 0% | 100% | 0 | 0 - 1 | | | neg (near to cut-off) | 55% | 45% | 55% | 45% | 11 | 8 - 12 | | | neg (near to cut-off) | 40% | 60% | 40% | 60% | 11 | 8 - 14 | | | neg (near to cut-off) | 45% | 55% | 45% | 55% | 11 | 10 - 14 | | | neg (near to cut-off) | 55% | 45% | 55% | 45% | 11 | 8 -14 | b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability: There are no reference standards for these analytes. Kit Stability: i. Real-time and Accelerated stability The stability studies were performed in accordance with EN 13640 (Stability Testing of In Vitro Diagnostic Reagents) to demonstrate unopened kit shelf-life stability (from the date of manufacture when stored at recommended temperature $2 - 8^{\circ}\mathrm{C}$ ) and opened kit shelf-life stability. The acceptance criteria are that results do not differ more than one result level [pos, pos (near the cutoff), neg (near the cutoff) and neg] from the reference run. For real-time stability study, three lots of all kit components stored at recommended storage temperature $2 - 8^{\circ}\mathrm{C}$ were evaluated at different occasions with three samples per antigen band. Original sealed kits and opened kits were demonstrated to be stable for up to 18 and 12 months, respectively, when stored at $2 - 8^{\circ}\mathrm{C}$ . {9} For accelerated study, three lots of all kits components were stored for 7 days at 37°C and evaluated with 3 samples per antigen band. The same lots stored at 4°C were used as reference. Reagents were shown to be stable for 7 days at 37°C. ii. Transport stability To simulate transportation of kits from Europe to US, one lot of an example EuroLine test kit with equal composition and technology as EUROLINE Profile Autoimmune Liver Disease 8 Ag (IgG) Kit was stored for 7 days at 4°C and -20°C and evaluated with 4 different samples. Transportation has no significant influence on the results. d. Detection limit: Not applicable e. Analytical specificity: Cross-reactivity: Cross-reactivity was investigated using a panel of samples serologically positive for antibodies to granulocyte cytoplasm (n = 10), thyroid gland antigens (n=10), islet cell antigens (n=10), cardiolipin/anti-phospholipid syndrome (n=5) and the CDC ANA reference panel (n=12). Clinical samples from the following groups were also included: autoimmune hepatitis type 1 (n=84), primary sclerosing cholangitis (n=19), systemic lupus erythematosus (n=10), rheumatoid arthritis (n=50), celiac disease (n=7), non-alcoholic steatohepatitis (n=30) and viral hepatitis (HBV, HCV; n=39). Out of 286 total samples, 9 samples tested positive. Interferences: Interference testing was performed for each antigen using at least 5 samples that cover the complete range of results (negative, positive and near to cut-off) for each antigen. Each sample was spiked with three different levels of endogenous interfering substances, namely hemoglobin, triglycerides and bilirubin. Evaluation of test results was performed visually and no significant interference was observed for concentrations up to 500 mg/dl for hemoglobin, 2000 mg/dl for triglycerides and 40 mg/dl for bilirubin. f. Assay cut-off: The cut-off is predefined by the visual evaluation. A sample is positive if the respective band is clearly visible. The cutoff for each antigen on the EUROLINE was set to the lowest limit of a clearly visible band. To achieve the uniform cut-off for each antigen, during the production several sample strips for each antigen are manufactured with the antigen used in different dilutions. These sample strips are then processed with positive and negative reference sera and the optimal dilution for each antigen is identified. Afterwards the strips are manufactured using each antigen in its optimal dilution. To confirm the assay cut-off, 261 clinically characterized samples from PBC and AIH patients, as well as 171 control samples were tested. The results 10 {10} showed that negative and positive results can clearly be discriminated by the assay. ## 2. Comparison studies: ### a. Method comparison with predicate device: A total of 295 clinically characterized samples (99 from patients with autoimmune hepatitis, 89 from patients with primary biliary cirrhosis, 15 from patients with AIH/PBC overlap syndrome, 42 from patients with viral hepatitis, and 50 from patients with rheumatoid arthritis (RA)) were evaluated and results were compared to the results for the predicate devices. In addition, 6 to 29 artificial samples with antibody concentrations close to the cut-off were created for each antigen by mixing positive samples with negative sample as diluent of the same matrix. The results and the positive and negative agreement between the device and the predicates are shown below. Borderline results (from the predicate) were not included in the agreement calculations. | | Predicate ELISA | | | Positive agreement* Negative agreement* % (95% C.I.)* | | | | --- | --- | --- | --- | --- | --- | --- | | | | | positive | | borderline | negative | | EUROLINE Profile Autoimmune Liver Disease 8 Ag (IgG) Kit | n = 281 | | M2 EP (MIT3) | | | Pos: 95.1 % (88.9 – 98.4 %) Neg: 98.9 % (95.9 – 99.9 %) | | | AMA-M2 and/or M2-3E | positive | 97 | 0 | 2 | | | | | negative | 5 | 4 | 173 | | | | n = 293 | | Sp100 | | | Pos: 94.4 % (81.3 – 99.3 %) Neg: 99.2 % (97.2 – 99.9 %) | | | Sp100 | positive | 34 | 1 | 2 | | | | | negative | 2 | 2 | 252 | | | | n = 296 | | gp210 | | | Pos: 100 % (89.4 – 100 %) Neg: 88.5 % (84.0 – 92.1 %) | | | gp210 | positive | 33 | 2 | 30*** | | | | | negative | 0 | 0 | 231 | | | | n = 325 | | LKM-1 | | | Pos: 92.6 % (82.1 – 97.9 %) Neg: 98.9 % (96.8 – 99.8 %) | | | LKM-1 | positive | 50 | 0 | 3 | | | | | negative | 4 | 1 | 267 | | | | n = 298 | | SLA | | | Pos: 100 % (88.4 – 100 %) Neg: 99.6 % (97.9 – 100.0 %) | | | SLA/LP | positive | 30 | 0 | 1 | | | negative | | 0 | 1 | 266 | | | * Calculations do not include borderline samples; ** 20 PBC, 1 artificial; *** 16 PBC, 7 AIH, 6 AIH/PBC, 1 RA. ### b. Matrix comparison: Comparisons between serum and each of EDTA, heparin, and citrated plasma were performed using 8-11 sample pairs selected to cover the complete range of results (negative, positive and close to cut-off) for each antigen. The results found no positive sample was found negative and vice versa. ## 1. Clinical studies: ### a. Clinical Sensitivity: A total of 734 clinically characterized samples obtained from different study sites were analyzed by the EUROLINE Profile Autoimmune Liver Disease 8 Ag (IgG) Kit. The sites providing the indicated panel of samples are as follow: {11} The results of the clinical studies are shown in the following tables: Sensitivity for Primary biliary liver cirrhosis (PBC) | Panel | n (men, women) | Mean age (age range) | EUROLINE Profile Autoimmune Liver Disease 8 Ag (IgG) Kit Positive (% positive) (95% C.I.) | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | | AMA-M2 | M2-3E | Sp100 | PML | gp210 | | PBC | 205 (20,185) | 58 y (22-90 y) | 175 (85.4 %) (79.8 - 89.9 %) | 160 (78.0 %) (71.8 - 83.5 %) | 54 (26.3 %) (20.5 - 32.9 %) | 58 (28.3 %) (22.2 - 35.0 %) | 68 (33.2 %) (26.8 - 40.1 %) | Sensitivity for Autoimmune hepatitis (AIH) | Panel | n (men, women) | Mean age (age range) | EUROLINE Profile Autoimmune Liver Disease 8 Ag (IgG) Kit Positive (% positive) (95% C.I.) | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | | LKM-1 | LC-1 | SLA/LP | | | | AIH | 163 (20, 59, 84 un- known) | 38 y (1-86 y, 84 un- known) | 20 (12.3 %) (7.7 - 18.3 %) | 15 (9.2 %) (5.2 - 14.7 %) | 15 (9.2 %) (5.2 - 14.7 %) | | | | Type 1 | 142 (16, 42, 84 un- known) | 48 y (20-86 y 84 un- known) | 2 (1.4 %) (0.2 - 5.0 %) | 1 (0.7 %) (0.0 - 3.9 %) | 15 (10.6 %) (6.0 - 16.8 %) | | | | Type 2 | 21 (4, 17) | 12 y (1-45 y) | 18 (85.7 %) (63.7 - 97.0 %) | 14 (66.7 %) (43.0 - 85.4 %) | 0 (0.0 %) (0.0 - 16.1 %) | | | b. Clinical specificity: Specificity for PBC | Panel | n (men, women) | Mean age (age range) | EUROLINE Profile Autoimmune Liver Disease 8 Ag (IgG) Kit Negative (% negative) (95% C.I.) | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | | AMA-M2 | M2-3E | Sp100 | PML | gp210 | | AIH | 163 (20, 59, | 38 y (1-86 y, | 162 (99.4 %) | 162 (99.4 %) | 162 (99.4 %) | 163 (100 %) | 154 (94.5 %) | | | 84 un- known) | 84 un- known) | | | | | | | | | | | | | | | {12} 13 | Panel | n (men, women) | Mean age (age range) | EUROLINE Profile Autoimmune Liver Disease 8 Ag (IgG) Kit Negative (% negative) (95% C.I.) | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | | AMA-M2 | M2-3E | Sp100 | PML | gp210 | | | 84 unknown) | 84 unknown) | | | | | | | Viral hepatitis | 39 (16, 23) | 48 y (25-84 y) | 39 (100 %) | 39 (100 %) | 39 (100 %) | 39 (100 %) | 39 (100 %) | | PBC | 19 (12, 7) | 48 y (21-73 y) | 19 (100 %) | 19 (100 %) | 19 (100 %) | 19 (100 %) | 19 (100 %) | | Further controls* | 308 (140, 167,1 unknown) | 18 y (0-83 y) | 304 (98.7 %) | 307 (99.7 %) | 308 (100 %) | 304 (98.7 %) | 302 (98.1 %) | | Total | 529 | | 524 (99.1 %) (97.8 - 99.7 %) | 527 (99.6 %) (98.6 - 100 %) | 528 (99.8 %) (99.0 - 100 %) | 525 (99.2 %) (98.1 - 99.8 %) | 514 (97.2 %) (95.4 - 98.4 %) | Specificity for AIH | Panel | n (men, women) | Mean age (age range) | EUROLINE Profile Autoimmune Liver Disease 8 Ag (IgG) Kit Negative (% negative) (95% C.I.) | | | | --- | --- | --- | --- | --- | --- | | | | | LKM-1 | LC-1 | SLA/LP | | PBC | 205 (20, 185) | 58 y (22-90 y) | 204 (99.5 %) | 205 (100 %) | 199 (97.1 %) | | Viral hepatitis | 39 (16, 23) | 48 y (25-84 y) | 39 (100 %) | 39 (100 %) | 39 (100.0 %) | | PSC | 19 (12, 7) | 48 y (21-73 y) | 19 (100 %) | 19 (100 %) | 19 (100.0 %) | | Further controls* | 308 (140,167, 1 unkn.) | 18 y (0-83 y) | 308 (100 %) | 306 (99.4%) | 307 (99.7 %) | | Total | 571 | | 570 (99.8 %) (99.0 - 100 %) | 569 (99.6 %) (98.7 - 100 %) | 564 (98.8 %) (97.5 - 99.5 %) | *from the following groups: systemic lupus erythematosus (n = 10), Sjögren’s syndrome (n = 5), systemic sclerosis (n = 5), myositis (n = 4), rheumatoid arthritis (n = 50), diabetes (n = 9), celiac disease (n = 7), A1 antitrypsin deficiency (n = 30), Alagille syndrome (n = 29), biliary atresia (n = 35), giant cell hepatitis (n = 16), non-alcoholic steatohepatitis (n = 30), hemochromatosis (n = 17), progressive familial intrahepatic cholestatis (n = 31), Wilson’s disease (n = 30) c. Other clinical supportive data (when a. and b. are not applicable): Not applicable. 4. Clinical cut-off: See Assay cut-off. 5. Expected values/Reference range: The levels of analytes were analyzed in a panel of 150 healthy blood donors of mixed age and sex. None of these samples were found positive. It is {13} recommended that each laboratory determine its own normal range based on the population and equipment used. **N. Proposed Labeling:** The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. **O. Conclusion:** The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 14
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...