EUROIMMUN ANTI-SLA/LP ELISA(LGG)

K112221 · Euroimmun US · NIY · Sep 11, 2012 · Immunology

Device Facts

Record IDK112221
Device NameEUROIMMUN ANTI-SLA/LP ELISA(LGG)
ApplicantEuroimmun US
Product CodeNIY · Immunology
Decision DateSep 11, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5660
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K112221 · Sep 11, 2012EUROIMMUN ANTI-SLA/LP ELISA(LGG)Euroimmun USRetrospective clinical samples from university, hospital, and private laboratoriesRetrospective clinical samples were used to establish clinical sensitivity and specificity for the detection of anti-SLA/LP antibodies in patients with autoimmune hepatitis and various control groups.Retrospective clinical samples; Clinical sensitivity; Clinical specificity; Diagnostic performance

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Retrospective clinical cohort study515 clinically characterized samples including AIH-1, AIH-2, AIH/PBC overlap syndrome, and other control groups (viral hepatitis, toxic liver damages, steatohepatitis, PBC, PSC, and other autoimmune diseases); Sample Size: 515; Number of Sites: Multiple (university, hospital, and private laboratories)Not applicable for this studyClinical sensitivity for AIH type 1 and clinical specificity across various liver and autoimmune conditions

Indications for Use

The EUROIMMUN Anti-SLA/LP ELISA (IgG) test kit is intended for the qualitative determination of IgG class autoantibodies against SLA/LP (soluble liver antigen/liver-pancreas antigen) in human serum and plasma. It is used as an aid in the diagnosis of autoimmune hepatitis, type 1, in conjunction with other laboratory and clinical findings.

Device Story

The EUROIMMUN Anti-SLA/LP ELISA (IgG) is an in vitro diagnostic kit used in clinical laboratories to detect IgG autoantibodies against SLA/LP antigen. The device utilizes a microplate coated with recombinant SLA/LP antigen. Patient serum or plasma samples are added to the wells; if anti-SLA/LP antibodies are present, they bind to the antigen. After washing, HRP-conjugated rabbit anti-human IgG is added, followed by TMB substrate. The resulting color change is measured via a microplate reader at 450 nm (reference 620-650 nm). Results are calculated as a ratio of the sample OD to the cut-off calibrator OD. A ratio ≥ 1.0 is considered positive. The output assists clinicians in the diagnosis of autoimmune hepatitis type 1 when interpreted alongside other clinical and laboratory findings.

Clinical Evidence

Clinical study of 515 samples (65 AIH-1, 68 AIH-2, 15 AIH/PBC overlap, 367 controls). Sensitivity for AIH-1 was 27.7% (95% CI: 17.3-40.2%). Specificity was 100% (95% CI: 99.2-100%) excluding overlap samples, and 99.3% (95% CI: 98.1-100%) including overlap samples. Method comparison with predicate (n=190) showed 97.4% overall agreement.

Technological Characteristics

Qualitative ELISA; 96-well microtiter plate coated with recombinant SLA/LP antigen. Energy source: external microwell plate reader (450nm). Reagents: peroxidase-labelled rabbit anti-human IgG conjugate, TMB chromogen/substrate, 0.5 M sulphuric acid stop solution. Manual or automated processing. No software algorithm; rule-based interpretation (Ratio ≥ 1.0).

Indications for Use

Indicated for qualitative detection of IgG autoantibodies against SLA/LP in human serum and plasma to aid in the diagnosis of autoimmune hepatitis type 1. For prescription use only.

Regulatory Classification

Identification

A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k112221 B. Purpose for Submission: New assay C. Measurand: Anti-soluble liver antigen/liver pancreas (SLA/LP) antigen IgG antibodies D. Type of Test: Qualitative E. Applicant: EUROIMMUN US Inc. F. Proprietary and Established Names: EUROIMMUN Anti-SLA/LP (IgG) G. Regulatory Information: 1. Regulation section: 21 CFR §866.5660, Multiple autoantibodies immunological test system 2. Classification: Class II 3. Product code: NIY, autoantibodies, anti-soluble liver antigen (SLA), autoimmune hepatitis 4. Panel: Immunology (82) H. Intended Use: 1. Intended use(s): The EUROIMMUN Anti-SLA/LP ELISA (IgG) test kit is intended for the qualitative determination of IgG class autoantibodies against SLA/LP (soluble liver antigen/liver-pancreas antigen) in human serum and plasma. It is used as an aid in the diagnosis of autoimmune hepatitis, type 1, in conjunction with other laboratory and clinical findings. 2. Indication(s) for use: Same as intended use above. 3. Special conditions for use statement(s): For prescription use only. {1} 4. Special instrument requirements: Microplate reader capable of measuring OD at 450 nm and 620 to 650 nm. I. Device Description: The device is packaged as a kit containing: 12 microplate strips each containing 8 antigen coated wells and frame, one cut-off calibrator (20 RU/mL), human serum-based positive and negative controls, horseradish peroxidase (HRP)-conjugated rabbit anti-human IgG, ready-to-use sample buffer, 10x wash buffer concentrate, 3,3',5,5' tetramethylbenzidine (TMB) substrate and sulfuric acid stop solution. J. Substantial Equivalence Information: 1. Predicate device name(s) and k number(s): INOVA Quanta Lite SLA ELISA, k021482 2. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Detection of anti-SLA/LP IgG autoantibodies to “aid in the diagnosis of autoimmune hepatitis, type 2”. | Same | | Assay | ELISA | Same | | Antigen | Recombinant SLA/LP | Same | | Substrate | TMB | Same | | Controls | One Positive Control One Negative Control | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Assay Type | Qualitative | Semi-quantitative | | Conjugate | Rabbit anti-human IgG labeled with HRP | Goat anti-human IgG labeled with HRP | | Calibrators and Controls | 1 cut-off calibrator 2 controls: 1 positive, 1 negative | 3 controls: 1 high positive, 1 low positive, 1 negative | | Sample Matrix | Serum or plasma (EDTA, heparin, and citrate) | Serum only | | Reported Units | Ratio | U/mL | | Cut-Off Level | Ratio 1.0 | 25 U/mL | K. Standard/Guidance Document Referenced (if applicable): CEN 13640:2002, Stability Testing of In Vitro Diagnostic Reagents. {2} # L. Test Principle: The test wells are coated with recombinant SLA/LP. The cut-off calibrator, controls, and diluted patient samples are added to the wells and autoantibodies recognizing the antigen bind during the first incubation. After washing the wells to remove all unbound proteins, conjugate is added. The conjugate binds to the captured human autoantibodies. Excess unbound conjugate is removed by another wash step. The bound conjugate is visualized with TMB substrate. Microtiter plates are read at 450 nm and a reference wavelength of 620 to $650\mathrm{nm}$ . The controls and patient results are interpreted by comparing them as a ratio of the cut-off. # M. Performance Characteristics (if/when applicable): # 1. Analytical performance: # a. Precision/Reproducibility: The reproducibility of the assay was assessed using patient sera with a range of values. Intra-assay reproducibility was assessed by 20 determinations on one day while inter-assay reproducibility was based on 24 determinations performed over six different days. The same samples were used in both experiments. The expected result was determined from the sample mean. The results that correctly matched the expected result were tallied. The results are summarized in the tables below: Intra-assay reproducibility: | Sample # | Sample Mean Ratio | Ratio Range | Expected Result | # Correct Results | | --- | --- | --- | --- | --- | | 1 | 0.2 | 0.2 – 0.2 | Neg | 20/20 | | 2 | 0.4 | 0.4 – 0.4 | Neg | 20/20 | | 3 | 0.9 | 0.8 – 0.9 | Neg | 20/20 | | 4 | 1.1 | 1.1 – 1.2 | Pos | 20/20 | | 5 | 1.3 | 1.1 – 1.4 | Pos | 20/20 | | 6 | 1.6 | 1.5 – 1.7 | Pos | 20/20 | | 7 | 3.6 | 3.4 – 3.7 | Pos | 20/20 | | 8 | 4.9 | 4.6 – 5.3 | Pos | 20/20 | Inter-assay reproducibility: | Sample # | Sample Mean Ratio | Ratio Range | Expected Result | # Correct Results | | --- | --- | --- | --- | --- | | 1 | 0.2 | 0.2 – 0.3 | Neg | 24/24 | | 2 | 0.4 | 0.3 – 0.4 | Neg | 24/24 | | 3 | 0.8 | 0.8 – 0.9 | Neg | 24/24 | | 4 | 1.2 | 1.0 – 1.3 | Pos | 24/24 | | 5 | 1.2 | 1.0 – 1.3 | Pos | 24/24 | | 6 | 1.7 | 1.6 – 1.8 | Pos | 24/24 | | 7 | 3.9 | 3.4 – 4.2 | Pos | 24/24 | | 8 | 5.4 | 4.7 – 5.8 | Pos | 24/24 | {3} Lot-to-lot reproducibility was assessed using patient sera with a range of values. Each sample/lot combination was tested between six and 38 times. The expected result was determined from the sample mean. The results that correctly matched the expected result were tallied. The results are summarized in the table below: | Sample | Sample Mean Ratio | Ratio Range | Lots Tested | Runs /Lot | Total Replicates | Expected Result | # Correct Results | | --- | --- | --- | --- | --- | --- | --- | --- | | 1 | 5.6 | 5.4 – 6.0 | 3 | 2 | 6 | Pos | 6/6 | | 2 | 7.6 | 7.3 – 8.0 | 3 | 2 | 6 | Pos | 6/6 | | 3 | 9.7 | 9.5 – 9.9 | 3 | 2 | 6 | Pos | 6/6 | | 4 | 0.9 | 0.9 – 0.9 | 3 | 2 | 6 | Neg | 6/6 | | 5 | 3.1 | 2.1 – 4.1 | 38 | 1 | 38 | Pos | 38/38 | | 6 | 3.9 | 3.2 – 4.6 | 38 | 1 | 38 | Pos | 38/38 | | 7 | 0.1 | 0.1 – 0.2 | 16 | 1 | 16 | Neg | 16/16 | | 8 | 0.1 | 0.1 – 0.2 | 10 | 1 | 10 | Neg | 10/10 | | 9 | 0.1 | 0.1 – 0.2 | 10 | 1 | 10 | Neg | 10/10 | b. Linearity/assay reportable range: Not applicable. c. Traceability, Stability, Expected values (controls, calibrators, or methods): There are no international standards for SLA/LP antibodies. Sample results are expressed as ratios of the cut-off calibrator. Calibrators and Controls: The cut-off calibrator and controls are derived from human serum and match pre-specified performance criteria. Each lot of cut-off calibrator and controls are traceable to a master lot. Stability studies: Stability studies are conducted in accordance to CEN 13640:2002. Three production lots of all kit reagents were tested. Real-time stability studies (+2°C to +8°C) support a 12-month unopened stability claim. Other studies support that the re-constituted wash buffer is stable for up to 28 days, and the opened reagents stored at +2°C to +8°C are stable for 12 months. Sample stability: The sponsor recommends following the guidelines in CLSI H18-A3 for sample storage. d. Detection limit: Not applicable. e. Analytical specificity: Endogenous interferents: Five samples containing various levels of anti-SLA/LP levels were spiked {4} with varying concentrations of hemoglobin up to 1,000 mg/dL, triglycerides up to 2,000 mg/dL, or bilirubin up to 40 mg/dL. Comparison of all spiked sample/interferent combinations to the unspiked sample was within ±15% of the original sample concentration except one; a sample with a ratio of 1.0 spiked with 40 mg/dL bilirubin had 119% recovery. Lower concentrations of bilirubin in the same sample showed recovery ±10% of the unspiked sample. ## Cross-reactivity with anti-LKM-1: Twenty-eight (28) sera serologically positive for anti-LKM-1 antibodies were tested with the assay and found negative for anti-SLA/LP antibodies suggesting that antibodies specific for anti-LKM-1 do not bind to the SLA/LP antigen coating the assay wells. Sample ratios ranged from 0.1 to 0.7. ## f. Assay cut-off: The assay cut-off is 1.0. Results ≥ 1.0 are positive. ## 2. Comparison studies: ### a. Method comparison with predicate device: One hundred and sixty-seven (167) clinically characterized samples were tested with the predicate device and the EUROIMMUN Anti-SLA/LP ELISA (IgG). These samples were from patients with autoimmune hepatitis (AIH), primary biliary cirrhosis (PBC), AIH/PBC overlap syndrome, and viral hepatitis. Twenty-three (23) samples intended to challenge the assay cut-off were created by mixing positive and negative samples. | | Predicate | | | | | | --- | --- | --- | --- | --- | --- | | | | Positive | Borderline | Negative | Total | | EUROIMMUNE Anti-SLA/LP ELISA | Positive | 30 | 1 | 5 | 36 | | | Negative | 0 | 0 | 154 | 58 | | | Total | 30 | 1 | 159 | 190 | If borderline sample is considered positive: Positive agreement 31/31 = 100% 95% CI: 89.0% - 100% Negative agreement 154/159 = 96.9% 95% CI: 92.8% - 98.6% If borderline sample is considered negative: Positive agreement 30/30 = 100% 95% CI: 88.6% - 100% Negative agreement 154/160 = 96.2% 95% CI: 92.1% - 98.3% ### b. Matrix comparison: The purpose of this study was to demonstrate that the new devices give the same results for lithium heparin plasma, citrate plasma and EDTA plasma as for serum samples collected from the same patient. Normal sample pairs were spiked with anti-SLA/LP positive sera to create 16 samples that spanned the assay range. Serum sample ratios ranged from 0.3 - 8.9. As compared to serum samples, bias of the heparin plasma, citrate plasma and EDTA plasma {5} samples ranged from 89% to 113%, 90% to 120%, and 88% to 115%, respectively. ## 3. Clinical studies: ### a. Clinical Sensitivity and Specificity: Five hundred and fifteen (515) clinically characterized samples from several university, hospital, and private laboratories were analyzed with the EUROIMMUN Anti-SLA/LP ELISA (IgG) test to assess the clinical sensitivity and specificity. The patient samples consisted of 65 AIH-1, 68 AIH-2, 111 PBC, 15 AIH/PBC overlap syndrome, 47 hepatitis B, 71 hepatitis C, 19 primary sclerosing cholangitis (PSC), 24 steatohepatitis, 14 toxic liver damage, and 81 samples from other autoimmune diseases. | | Autoimmune Hepatitis Type 1 | | | | | --- | --- | --- | --- | --- | | | | Positive | Negative | Total | | Anti-SLA/LP | Positive | 18 | 3* | 21 | | | Negative | 47 | 447 | 494 | | | Total | 65 | 450 | 515 | * Samples were from AIH/PBC patients Sensitivity: 27.7% (18/65) 95% CI: 18.3% – 39.6% Specificity: 99.3% (447/450) 95% CI: 98.1% – 100.0% The results for non-AIH-1 subjects are shown below: | Results by Diagnosis | Anti-SLA/LP (IgG) | | | --- | --- | --- | | | n | Positive (%) | | Autoimmune hepatitis type 2 (AIH-2) | 68 | 0 (0.0%) | | AIH/PBC overlap syndrome | 15 | 3 (20.0%) | | Viral hepatitis | 118 | 0 (0.0%) | | Toxic liver damages | 14 | 0 (0.0%) | | Steatohepatitis | 24 | 0 (0.0%) | | Primary biliary liver cirrhosis (PBC) | 111 | 0 (0.0%) | | Primary sclerosing cholangitis (PSC) | 19 | 0 (0.0%) | | Other autoimmune diseases* | 81 | 0 (0.0%) | * Other autoimmune diseases include: MCTD, celiac disease, Type 1 diabetes, Hashimoto thyroiditis, Graves' disease, and ulcerative colitis ### b. Other clinical supportive data: Not applicable. ## 4. Clinical cut-off: Not applicable. {6} 5. Expected values/Reference range: The levels of anti-SLA/LP antibodies in 150 apparently healthy blood donor samples (Caucasian, mixed age and sex) were analyzed with the EUROIMMUN Anti-SLA/LP ELISA (IgG). No samples were positive, and ratios ranged from 0 to 0.8. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 7
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...