ZEUS ANTI-CARDIOLIPIN IGM TEST SYSTEM

K981021 · Zeus Scientific, Inc. · MID · Jun 2, 1998 · Immunology

Device Facts

Record IDK981021
Device NameZEUS ANTI-CARDIOLIPIN IGM TEST SYSTEM
ApplicantZeus Scientific, Inc.
Product CodeMID · Immunology
Decision DateJun 2, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5660
Device ClassClass 2

Intended Use

Autoantibodies directed against phospholipids, and anti-cardiolipin (aCL) in particular, have been associated with recurrent thrombosis, thrombocytopenia, and spontaneous abortions. Detection of these autoantibodies may aid in the diagnosis of such disorders. The Zeus Scientific, Inc. anti-Cardiolipin IgM ELISA Test System is an enzyme-linked immunosorbent assay (ELISA) designed for the semi-quantitative measurement of circulating IgM autoantibodies to cardiolipin. This test is for in vitro diagnostic use.

Device Story

The Zeus Scientific, Inc. anti-Cardiolipin IgM ELISA Test System is an in vitro diagnostic assay used in clinical laboratories. It detects circulating IgM autoantibodies to cardiolipin in human serum samples. The device utilizes enzyme-linked immunosorbent assay (ELISA) technology to measure these autoantibodies semi-quantitatively. Healthcare professionals, such as laboratory technicians, perform the test to assist in the clinical diagnosis of conditions linked to phospholipid autoantibodies, including recurrent thrombosis, thrombocytopenia, and spontaneous abortions. The output provides a semi-quantitative measurement that clinicians use to support diagnostic decision-making for patients presenting with these clinical symptoms.

Clinical Evidence

No clinical data provided in the document; substantial equivalence is based on the device's intended use and technological characteristics as an ELISA test system.

Technological Characteristics

Enzyme-linked immunosorbent assay (ELISA) for semi-quantitative measurement of IgM autoantibodies. In vitro diagnostic test system.

Indications for Use

Indicated for the semi-quantitative measurement of circulating IgM autoantibodies to cardiolipin in human serum to aid in the diagnosis of disorders associated with recurrent thrombosis, thrombocytopenia, and spontaneous abortions.

Regulatory Classification

Identification

A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Image /page/0/Picture/2 description: The image shows a logo or emblem that includes the words "DEPARTMENT OF HEALTH" arranged vertically along the left side. To the right of the text is a stylized graphic consisting of three curved lines or shapes, stacked vertically. The overall design appears to be a combination of text and abstract elements, possibly representing a health-related organization or department. 2 1998 JUN Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Mr. Mark J. Kopnitsky Vice President, Research and Development Zeus Scientific, Inc. 200 Evans Way Branchburq, New Jersey 08876 Re : K981021/S1 Anti-Cardiolipin IgM ELISA Test System Trade Name: Requlatory Class: II Product Code: MID Dated: May 20, 1998 Received: May 21, 1998 Dear Mr. Kopnitsky: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) , it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations. {1}------------------------------------------------ Paqe 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655. This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" Sincerely yours, Steven Butman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Page 510(k) Number (if known): 198107 | The Zeus Scientific, Inc. anti-Cardiolipin IgM ELISA Test System Device Name: Indications for Use: Autoantibodies directed against phospholipids, and anti-cardiolipin (aCL) in particular, have been associated with recurrent thrombosis, thrombocytopenia, and spontaneous abortions. Detection of these autoantibodies may aid in the diagnosis of such disorders. The Zeus Scientific, Inc. anti-Cardiolipin IgM ELISA Test System is an enzyme-linked immunosorbent assay (ELISA) designed for the semi-quantitative measurement of circulating IgM autoantibodies to cardiolipin. This test is for in vitro diagnostic use. (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) eborah Moore, Acting Branch Chief (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number K981021 Prescription Use (Per 21 CFR 801,109) OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...