K981020 · Zeus Scientific, Inc. · MID · May 18, 1998 · Immunology
Device Facts
Record ID
K981020
Device Name
ZEUS ANTI-CARDIOLIPIN IGS ELISA TEST SYSTEM
Applicant
Zeus Scientific, Inc.
Product Code
MID · Immunology
Decision Date
May 18, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5660
Device Class
Class 2
Intended Use
Autoantibodies directed against phospholipids, and anti-cardiolipin (aCL) in particular, have been associated with recurrent thrombosis, thrombocytopenia, and spontaneous abortions. Detection of these autoantibodies may aid in the diagnosis of such disorders. The Zeus Scientific, Inc. anti-Cardiolipin IgA ELISA Test System is an enzyme-linked immunosorbent assay (ELISA) designed for the semi-quantitative measurement of circulating IgA autoantibodies to cardiolipin. This test is for in vitro diagnostic use.
Device Story
Zeus Scientific, Inc. anti-Cardiolipin IgA ELISA Test System is an in vitro diagnostic device used in clinical laboratories. It utilizes enzyme-linked immunosorbent assay (ELISA) technology to detect IgA autoantibodies to cardiolipin in patient serum samples. The assay provides semi-quantitative measurements, which clinicians use to assist in diagnosing conditions linked to phospholipid autoantibodies, such as recurrent thrombosis, thrombocytopenia, and spontaneous abortions. The test is intended for prescription use by healthcare professionals.
Clinical Evidence
No clinical data provided in the document; substantial equivalence is based on regulatory classification and intended use.
Technological Characteristics
Enzyme-linked immunosorbent assay (ELISA) for semi-quantitative measurement of IgA autoantibodies. In vitro diagnostic device.
Indications for Use
Indicated for the semi-quantitative measurement of circulating IgA autoantibodies to cardiolipin in human serum to aid in the diagnosis of disorders associated with recurrent thrombosis, thrombocytopenia, and spontaneous abortions.
Regulatory Classification
Identification
A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).
Related Devices
K981014 — ZEUS SCIENTIFIC, INC. ANTI-CARDIOLIPIN IGG ELISA TEST SYSTEM · Zeus Scientific, Inc. · Jun 2, 1998
K984173 — THE APTUS (AUTOMATED) APPLICATION OF THE ANTI-CARDIOLIPIN IGA ELISA TEST SYSTEM · Zeus Scientific, Inc. · Dec 22, 1998
K981021 — ZEUS ANTI-CARDIOLIPIN IGM TEST SYSTEM · Zeus Scientific, Inc. · Jun 2, 1998
K971784 — ZEUS SCIENTIFIC, INC., ANTICARDIOLIPIN IGG/IGM ELISA TEST SYSTEM · Zeus Scientific, Inc. · Jul 23, 1997
K984175 — THE APTUS (AUTOMATED) APPLICATION OF THE ANTI-CARDIOLIPIN IGM ELISA TEST SYSTEM · Zeus Scientific, Inc. · Dec 17, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized image of an eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle image.
## 1 8 MAY
Mark J. Kopnitsky Vice President, Research and Development Zeus Scientific, Inc. 200 Evans Way Branchburg, New Jersey 08876
Re: K981020 The Zeus Scientific, Inc. Anti-Cardiolipin IgA Trade Name: ELISA Test System Requlatory Class: II Product Code: MID Dated: March 15, 1998 Received: March 19, 1998
Dear Mr. Kopnitsky:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Requlations.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
{1}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. ್ದರಿ determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
of of of Page
510(k) Number (if known): 198/020
The Zeus Scientific, Inc. anti-Cardiolipin IgA ELISA Test System Device Name:
Indications for Use:
Autoantibodies directed against phospholipids, and anti-cardiolipin (aCL) in particular, have been associated with recurrent thrombosis, thrombocytopenia, and spontaneous abortions. Detection of these autoantibodies may aid in the diagnosis of such disorders. The Zeus Scientific, Inc. anti-Cardiolipin IgA ELISA Test System is an enzyme-linked immunosorbent assay (ELISA) designed for the semi-quantitative measurement of circulating IgA autoantibodies to cardiolipin. This test is for in vitro diagnostic use.
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

| (Division Sign-Off) | |
|-----------------------------------------|---------|
| Division of Clinical Laboratory Devices | |
| 510(k) Number | K981020 |
| Prescription Use | <div style="text-align:center;">✓</div> | OR | Over-The-Counter Use |
|----------------------|-----------------------------------------|----|--------------------------|
| (Per 21 CFR 801,109) | | | (Optional Format 1-2-96) |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.