K183007 · Phadia AB · LKP · Dec 24, 2018 · Immunology
Device Facts
Record ID
K183007
Device Name
EliA SmDP Immunoassay
Applicant
Phadia AB
Product Code
LKP · Immunology
Decision Date
Dec 24, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5100
Device Class
Class 2
Intended Use
EliA SmDP is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Sm in human serum and EDTA-plasma to aid in the clinical diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA SmDP uses the EliA IgG method on the instrument Phadia 2500/5000.
Device Story
EliA SmDP Immunoassay is an in vitro diagnostic test for detecting IgG antibodies to Sm protein in human serum or EDTA-plasma. The device utilizes polystyrene wells coated with synthetic SmD3 peptide. Patient samples are incubated in these wells; if anti-Sm antibodies are present, they bind to the antigen. After washing, enzyme-labeled anti-IgG conjugate is added, followed by a development solution. The resulting fluorescence is measured to quantify antibody levels. The assay is performed on the Phadia 2500/5000 automated instrument system. The system uses disposable pipette tips to prevent carry-over and features an integrated barcode reader for automated reagent tracking. Results are calculated by comparing sample fluorescence to a calibration curve. The device is intended for use in clinical laboratories by trained personnel to assist in the diagnosis of systemic lupus erythematosus (SLE).
Clinical Evidence
No new clinical diagnostic studies were performed. Substantial equivalence was established via analytical performance testing (precision, linearity, LoD/LoQ) and a method comparison study. The instrument comparison study evaluated 105 samples (75 positive, 15 negative, 15 equivocal) across three Phadia 2500/5000 instruments versus the predicate. Passing-Bablok regression showed high correlation (R > 0.99). PPA and NPA were calculated, with TPA ranging from 98.1% to 100.0% (equivocal as positive). Reference range established using 400 healthy subjects.
Technological Characteristics
Solid-phase fluoroimmunoassay. Reagents: SmD3 peptide-coated wells, β-Galactosidase labeled anti-IgG conjugate, 4-Methylumbelliferyl-β-D-galactoside development solution. Instrument: Phadia 2500/5000 (automated, random access). Connectivity: Networked via IDM. Software: Onboard ISW. Sterilization: Not applicable (reagents are ready-to-use). Materials: High-density polyethylene (dilution wells), polyethylene (pipette tips).
Indications for Use
Indicated for the in vitro semi-quantitative measurement of IgG antibodies to Sm in human serum and EDTA-plasma to aid in the clinical diagnosis of systemic lupus erythematosus (SLE).
Regulatory Classification
Identification
An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 24, 2018
Phadia AB Shervl Skinner Associate Director RA/QA Phadia US Inc. 4169 Commercial Avenue Portage, Michigan 49002
Re: K183007
Trade/Device Name: EliA SmDP Immunoassay Regulation Number: 21 CFR 866.5100 Regulation Name: Antinuclear antibody immunological test system Regulatory Class: Class II Product Code: LKP Dated: October 31, 2018 Received: October 31, 2018
Dear Sheryl Skinner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Leonthena R. Carrington -S
Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K183007
Device Name EliA(TM) SmDP Immunoassay
#### Indications for Use (Describe)
EliA SmDP is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Sm in human serum and EDTA-plasma to aid in the clinical diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA SmDP uses the EliA IgG method on the instrument Phadia 2500/5000.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|--------------------------------------------------------------------------------------|-----------------------------------------------|
|--------------------------------------------------------------------------------------|-----------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
This summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92.
## 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
## A. 510(k) Number:
K183007
#### B. Purpose for Submission:
Adding a previously cleared assay on a new instrument platform (Phadia® 2500/5000)
#### C. Measurand:
lgG antibodies specific for Sm protein
#### D. Type of Test:
Semi-quantitative measurement immunoassays
#### E. Applicant:
Phadia AB Rapsgatan 7P P.O. Box 6460 SE-751 37 Uppsala, Sweden Tel: +46-18-16 50 60
510(k) Contact Person: Sheryl Skinner Associate Director, RA/QA ImmunoDiagnostics US Thermo Fisher Scientific Phadia US Inc. 4169 Commercial Avenue Portage, Mi 49002, USA Tel. 269.568.3603 sheryl.skinner@thermofisher.com
Date of Summary Preparation:
October 30, 2018
# F. Proprietary and Established Names:
EliA™ SmDP Immunoassay
{4}------------------------------------------------
#### G. Regulatory Information:
- 1. Regulation section: 21 CFR §866.5100 Antinuclear Antibody Immunological Test System
- 2. Classification: Class II
- 3. Product code: LKP, Anti-Sm Antibody, Antigen and Control
- 4. Panel: Immunology (82)
#### H. Intended use(s):
1. Intended use(s):
EliA SmDP is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Sm in human serum and EDTA-plasma to aid in the clinical diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA SmDP uses the EliA IgG method on the instrument Phadia 2500/5000.
2. Indication(s) for use: Same as intended use
3. Special conditions for use statement(s): For prescription use only
Special instrument requirements: ব
Performance studies were obtained from the Phadia® 2500/5000 instrument. This device is not for point-of-care use.
{5}------------------------------------------------
#### Device Description: .
The method-specific reagents are identical with K132631 (EliA SmDº on Phadia 250), but are filled in containers specific for the Phadia 2500/5000 instrument. Each device consists of:
- Test Wells: -EliA SmDP Wells are coated with a synthetic SmD3 peptide – 4 carriers (12 wells each), ready to use;
- EliA Sample Diluent: PBS containing BSA, detergent and 0.095% (w/v) sodium azide - 6 bottles, 48 mL each, ready to use; or 6 bottles, 400 mL each, ready to use;
- -EliA IgG Conjugate 50 or 200: ß-Galactosidase labeled anti-IgG (mouse monoclonal antibodies) in PBS containing BSA and 0.06% (w/v) sodium azide – 6 wedge shaped bottles, 5 mL each, ready to use; or 6 wedge shaped bottles, 19 mL each, ready to use
- EliA IgG Calibrator Strips: Human IgG (0, 4, 10, 20, 100, 600 µg/L) in PBS containing BSA, detergent and 0.095% (w/v) sodium azide - 5 strips, 6 singleuse vials per strip, 0.3 mL each, ready to use;
- -EliA IgG Curve Control Strips: Human IgG (20 µg/L) in PBS containing BSA, detergent and 0.095% (w/v) sodium azide - 5 strips, 6 single-use vials per strip, 0.3 mL each, ready to use;
- EliA IgG Calibrator Well: Coated with mouse monoclonal antibodies 4 carriers (12 wells each), ready to use.
The Phadia EliA Immunodiagnostic System is an automated system for immunodiagnostic testing. The EliA reagents are available as modular packages, each purchased separately. All packages are required to carry out EliA SmDP tests.
#### J. Substantial Equivalence Information:
Predicate device name(s) and 510(k) number(s): 1. EliA SmDP on Phadia 250 instrument, K132631
{6}------------------------------------------------
#### 2. Comparison with predicate device:
#### EliA SmDP Immunoassays on Phadia 250 and Phadia 2500/5000 instruments – Similarities to predicate devices
| Feature | Predicate Device<br>Phadia 250 | New Device<br>Phadia 2500/5000 |
|--------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use<br>EliA SmDP | EliA SmDP is intended for the in<br>vitro semi-quantitative<br>measurement of IgG antibodies<br>directed to Sm in human serum<br>and plasma (EDTA, citrate) as<br>an aid in the clinical diagnosis of<br>systemic lupus erythematosus<br>(SLE) in conjunction with other<br>laboratory and clinical findings.<br>EliA SmDP uses the EliA IgG<br>method on the instrument<br>Phadia 250. | EliA SmDP is intended for the in<br>vitro semi-quantitative<br>measurement of IgG antibodies<br>directed to Sm in human serum<br>and EDTA-plasma to aid in the<br>clinical diagnosis of systemic<br>lupus erythematosus (SLE) in<br>conjunction with other laboratory<br>and clinical findings. EliA SmDP<br>uses the EliA IgG method on the<br>instrument Phadia 2500/5000. |
| Analytical<br>technology:<br>Immuno-<br>fluorescence<br>measurement | Same | Same |
| Assay process | Same | Same |
| Common, dedicated<br>Phadia reagents | Same | Same<br>Introduction of new article<br>numbers for Development<br>Solution, Stop Solution and<br>Washing Solution is only due to<br>larger filling volumes which are<br>required for the bigger<br>instruments Phadia 2500/5000 |
| Result calculation<br>software; Phadia<br>Information Data<br>Manager (IDM) | Same | Same |
| Sample volume | 90 µL (20 µL of non-diluted<br>sample) | 90 µL (20 µL of non-diluted<br>sample) |
| Incubation<br>temperature | 37°C | 37°C |
| Conjugate volume | 90 µL | 90 µL |
| Development<br>Solution Volume | 90 µL | 90 µL |
| Stop Solution<br>Volume | 200 µL | 200 µL |
| Assay set-up | Random access | Random access |
| Reagent packaging<br>size | Various/Common | Various/Common<br>Introduction of new article<br>number for EliA Sample Diluent<br>(83-1071-01) is only due to<br>larger filling volume. |
| Onboard storage of<br>reagents | Yes | Yes |
| Time to 1st result | ~2 h | ~2 h |
| Feature | Predicate Device<br>Phadia 250 | New Device<br>Phadia 2500/5000 |
| Sample matrix;<br>Serum or plasma<br>type as indicated in<br>the DFU dependent<br>on assay | Human serum and plasma<br>(EDTA, citrate) | Human serum or EDTA- plasma,<br>i.e. citrate plasma is omitted as<br>for all EliA tests |
| Daily throughput | ~250 tests | ~2500/5000 tests |
| Sample Dilution | Phadia 250 uses a steel pipette<br>to dilute the samples in Dilution<br>Plates (Art.No. 12-3907-08) | Phadia 2500/5000 uses<br>disposable Pipette Tips in Racks<br>(Art No. 12-3805-04) for<br>pipetting samples in Dilution<br>Well (Art.No. 12-4005-69) |
| Risk for carry-over | The warning "DO NOT REUSE"<br>in the Phadia 250 DFU for EliA<br>Conjugates is due to the fact that<br>a low risk of conjugate<br>contamination by carry-over from<br>samples was identified. In order<br>to reduce the risk, the single use<br>statement for the conjugate was<br>included in the Phadia 250 DFU. | When running EliA tests on the<br>Phadia 2500/5000 instruments,<br>there is no need for this warning<br>statement because these<br>instruments use disposable tips<br>for pipetting samples and a<br>separate pipette for the<br>conjugate, and carry-over from<br>samples to conjugate is<br>impossible. |
| Loading of EliA<br>Carriers | EliA carriers are loaded manually<br>on the Loading Tray from where<br>they can be processed directly or<br>transferred to the cooled storage<br>compartment. | The Phadia 2500/5000<br>instruments do not have such a<br>Loading Tray. The EliA carriers<br>are loaded into racks which are<br>directly transferred to the cooled<br>storage compartment |
| Barcode reader | The Phadia 250 instrument has<br>a built-in barcode reader at the<br>front of the instrument, but the<br>operator needs to scan the<br>barcodes manually by showing<br>the reagents to the barcode<br>reader. Alternatively, the<br>operator can also enter the<br>characters below the barcode<br>manually. | The Phadia 2500/5000<br>instruments dispose of a built-in<br>barcode reader, and the<br>reagents are on a moving belt<br>which conveys them past the<br>barcode reader. The lot-specific<br>information will be read<br>automatically by the instrument<br>during loading. |
| Process time / Time<br>to patient result | Phadia 250 needs 1 minute to<br>process one Well.<br>Phadia 250 provides the results<br>at a one minute interval. | Phadia 2500/5000 instruments<br>process two Wells in parallel in<br>48 seconds.<br>Phadia 2500/5000 provides the<br>results at a 24 seconds interval. |
{7}------------------------------------------------
{8}------------------------------------------------
### EliA SmD² Immunoassays on Phadia 250 and Phadia 2500/5000 instruments – Differences to predicate device
{9}------------------------------------------------
#### K. Standard/Guidance Document Referenced (if applicable):
CLSI EP05-A3; Evaluation of Precision Performance of Quantitative Measurement Methods: September 2014
CLSI EP06-A Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach: April 2003
CLSI EP17-A, Protocols for Determination of Limits of Detection and Limits of Quantification; October 2004.
CLSI EP09-A3, Measurement Procedure Comparison and Bias Estimation Using Patient Samples
#### L. Test Principle:
The EliA wells are molded cups comparable to excised wells from a microtiter plate. They are made of polystyrene and are coated with the respective antigen. The wells are at the same time a holder of the coupled antigen for convenient automation and a reaction chamber with reaction/washing solution handling based on pipetting to add and aspiration to remove liquids.
The EliA wells are coated with a synthetic SmD3 peptide. If present in the patient's specimen, antibodies to SmD3 peptide bind to the specific antigen. After washing away non-bound antibodies, enzyme-labeled antibodies against human IgG antibodies (EliA IgG Conjugate) are added to form an antibody-conjugate complex. After incubation, non-bound conjugate is washed away and the bound complex is incubated with a Development Solution. After stopping the reaction, the fluorescence in the reaction mixture is measured. The higher the response value, the higher the amount of antibody bound and detected in the sample tested. To evaluate test results, the response for patient samples is compared directly to the response for calibrators.
#### M. Performance Characteristics (if/when applicable):
1. Analytical performance:
- a. Precision/Reproducibility:
To determine the precision of the assay, the variability was assessed in a study with a total of 21 runs (3 instruments x 7 runs).
The study was performed with 1 run/day over a period of 7 days. Each sample was tested in four replicates/run giving in total 84 replicates per sample. The data was calculated against the calibration curve from Day 1.
We included only one lot of EliA SmDP Well on the Phadia 2500/5000 instrument, as data for inter-lot-variation has already been shown in K132631. The results are summarized in the table below:
{10}------------------------------------------------
| Mean<br>(EliA<br>U/mL) | n | Within-Run | | Between-<br>Run | | Between-<br>Instrument | | Total<br>Imprecision | |
|------------------------|----|------------|-----|-----------------|------|------------------------|------|----------------------|------|
| | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 2.2 | 84 | 0.2 | 8.7 | 0.2 | 11.0 | 0.4 | 20.5 | 0.5 | 24.8 |
| 7.1 | 84 | 0.3 | 3.6 | 0.2 | 2.8 | 0.3 | 4.7 | 0.5 | 6.5 |
| 10.0 | 84 | 0.4 | 3.7 | 0.2 | 1.9 | 0.8 | 7.7 | 0.9 | 8.8 |
| 146.6 | 84 | 5.1 | 3.5 | 3.7 | 2.5 | 7.2 | 4.9 | 9.5 | 6.5 |
| 408.1 | 84 | 15.5 | 3.8 | 9.8 | 2.4 | 31.1 | 7.6 | 36.1 | 8.8 |
#### EliA SmDP on Phadia 2500/5000
#### b. Linearity/assay reportable range:
Four patient serum samples were diluted in EliA Sample Diluent and tested with one batch of EliA SmDP Immunoassay and one set of system reagents on Phadia 2500/5000. The ratios of observed/expected values were calculated. The results are summarized below:
| EliA SmDP on Phadia 2500/5000 | |
|-------------------------------|--|
| PAUL 20 | |
| Dilution<br>range<br>(EliA U/mL) | Slope | 95% Cl | Intercept | 95% Cl | R2 |
|----------------------------------|-------|-------------|-----------|----------------|------|
| 6.5 – 285.5 | 1.04 | 0.95 – 1.12 | 4.21 | 4.21 – -5.70 | 0.99 |
| 8.7 – 392.8 | 0.98 | 0.94 – 1.02 | -0.56 | -0.56 – -7.29 | 1.00 |
| 27.1 – 537.4 | 1.01 | 0.97 – 1.05 | -10.58 | -21.14 – -0.03 | 1.00 |
| 0.9 - 11.4 | 1.01 | 0.99 – 1.02 | -0.09 | -0.19 – 0.01 | 1.00 |
The linear range and the measuring range are set to 1.6 EliA U/mL (LoQ) to 480 EliA U/mL (upper limit of measuring range),
The reportable range (Limit of Detection, upper limit of measuring range) for EliA SmDP is from 0.8 to 480 EliA U/mL. Concentration values between LoD and LoQ may show a higher uncertainty.
- Traceability, Stability, Expected values (controls, calibrators, or methods): C. The EliA IgG method was previously reviewed in K061165.
- d. Detection limit:
The limit of blank (LoB) and limit of detection (LoD) studies were performed on the Phadia 2500/5000 instrument. One blank sample and five low level samples were measured in twelve replicates in each of six runs spread over six different days.
The LoD for EliA SmDP is 0.8 EliA U/mL, determined consistent with the quidelines in CLSI document EP17-A and with proportions of false positives (a) less than 5% and false negatives (ß) less than 5%: based on 432 determinations with 72 blank and 360 low level replicates; and LoB of 0.3 EliA U/mL.
{11}------------------------------------------------
The LoQ for EliA SmDP is 1.6 EliA U/mL, determined consistent with the guidelines in CLSI document EP17-A, based on 360 determinations; and a target uncertainty goal of 20%.
The results are summarized in the table below:
| EliA SmDP (EliA U/mL) | LoB | LoD | LoQ |
|-----------------------|-----|-----|-----|
| Phadia 2500/5000 | 0.3 | 0.8 | 1.6 |
- e. Analytical specificity:
Interference: Previously reviewed in K132631
Carry-over: Phadia 2500/5000 instruments use disposable tips for pipetting samples and a separate pipette for the conjugate, therefore carry-over from samples to conjugate is impossible.
- Assay cut-off: f.
The ranges (negative, equivocal, positive) recommended for the evaluation of the test results were derived from the clinical studies (s. K132631).
#### EliA SmD® Well
| < 7 EliA U/mL | Negative |
|------------------|-----------|
| 7 – 10 EliA U/mL | Equivocal |
| > 10 EliA U/mL | Positive |
- 2. Comparison studies:
- Method comparison with predicate device (Instrument comparison): a. See 2c Instrument Comparison below
- b. Matrix comparison: Previously reviewed under K132631.
- Instrument comparison C.
In the Method Comparison studies for EliA SmDP, more than 100 samples (≥10% of the samples within ±25% of the medical decision point) were run in sinqle replicates on one Phadia 250 and one Phadia 2500/5000 instrument. The acceptance criteria for the method comparison (the slope for the regression lines should be 0.9 – 1.1 for single replicate to single replicate and intercept close to 0) were met for EliA SmDP.
{12}------------------------------------------------
#### A.6-10
#### EliA SmDº:
| Instrument | Intercept | 95% Cl | Slope | 95% Cl |
|---------------|-----------|----------------|-------|--------------|
| PH2500/5000 A | 0.07 | -0.38 to 0.53 | 1.00 | 0.98 to 1.04 |
| PH2500/5000 B | 0.06 | -0.39 to 0.24 | 1.03 | 1.01 to 1.05 |
| PH2500/5000 C | -0.34 | -0.70 to -0.08 | 1.06 | 1.03 to 1.09 |
#### equivocal results considered positive
| criteria | PH2500/5000 A | PH2500/5000 B | PH2500/5000 C |
|----------|----------------|----------------|----------------|
| PPA | <b>98.9%</b> | <b>98.9%</b> | <b>100.0%</b> |
| 95% CI | 94.0% - 100.0% | 94.0% - 100.0% | 96.0% - 100.0% |
| NPA | <b>93.3%</b> | <b>100.0%</b> | <b>100.0%</b> |
| 95% CI | 68.1% - 99.8% | 73.5% - 100.0% | 78.2% - 100.0% |
| TPA | <b>98.1%</b> | <b>99.0%</b> | <b>100.0%</b> |
| 95% CI | 93.3% - 99.8% | 94.7% - 100.0% | 96.5% - 100.0% |
equivocal results considered negative
| criteria | PH2500/5000 A | PH2500/5000 B | PH2500/5000 C |
|----------|----------------|----------------|----------------|
| PPA | <b>98.7%</b> | <b>100.0%</b> | <b>100.0%</b> |
| 95% CI | 92.8% - 100.0% | 95.2% - 100.0% | 95.2% - 100.0% |
| NPA | <b>86.7%</b> | <b>88.9%</b> | <b>93.3%</b> |
| 95% CI | 69.3% - 96.2% | 70.8% - 97.6% | 77.9% - 99.2% |
| TPA | <b>95.2%</b> | <b>97.1%</b> | <b>98.1%</b> |
| 95% CI | 89.2% - 98.4% | 91.6% - 99.4% | 93.3% - 99.8% |
- 3. Clinical studies:
- a. Clinical sensitivity: Not applicable.
- b. Clinical specificity: Not applicable.
- c. Other clinical supportive data (when a. and b. are not applicable): Clinical performance values were reviewed in K132631.
- Clinical cut-off: 4. Same as assay cut-off.
{13}------------------------------------------------
#### 5. Expected values/Reference range:
The frequency distribution for Sm antibodies was investigated in a group of apparently healthy subjects equally distributed by age and gender, using sera from a Caucasian population obtained from a blood bank. The results are given in the table below:
| Test | n = | Median<br>(EliA U/mL) | 95th<br>percentile | 99th<br>percentile |
|----------------------------------|-----|-----------------------|--------------------|--------------------|
| EliA SmDP on Phadia<br>2500/5000 | 400 | 1.1 | 2.2 | 4.2 |
#### N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
#### O. Conclusion:
All available data support that both instrument platforms, Phadia 250 and Phadia 2500/5000 perform substantially equivalent when using the EliA SmDP immunoassays.
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.