K132631 · Phadia US, Inc. · LKP · Aug 29, 2014 · Immunology
Device Facts
Record ID
K132631
Device Name
EIA SMDP IMMUNOASSAY
Applicant
Phadia US, Inc.
Product Code
LKP · Immunology
Decision Date
Aug 29, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5100
Device Class
Class 2
Indications for Use
EliA SmD⁺ is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Sm in human serum and plasma (EDTA, citrate) to aid in the diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA SmD⁺ uses the EliA IgG method on the instrument Phadia 100. EliA SmD⁺ is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Sm in human serum and plasma (EDTA, citrate) to aid in the diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA SmD⁺ uses the EliA IgG method on the instrument Phadia 250.
Device Story
EliA SmD⁺ is a fluoroenzymeimmunoassay for detecting IgG antibodies against SmD3 peptide in human serum/plasma. Input: patient serum/plasma samples. Operation: samples are incubated in wells coated with synthetic SmD3 peptide; bound antibodies are detected via β-galactosidase-labeled anti-human IgG conjugate; fluorescence is measured after adding development solution. Performed on automated Phadia 100 or Phadia 250 instruments. Output: semi-quantitative concentration (U/mL) compared against a stored calibration curve. Used in clinical laboratories by trained personnel. Results aid clinicians in diagnosing SLE by identifying specific autoantibodies. Benefits include automated, standardized detection of Sm antibodies compared to manual ELISA methods.
Clinical Evidence
Clinical performance evaluated using 350 specimens (100 SLE, 250 other autoimmune/infectious conditions). Sensitivity: 23.0% (95% CI: 15.2–32.5%) to 25.0% (95% CI: 16.9–34.7%) depending on equivocal handling. Specificity: 97.6% (95% CI: 94.8–99.1%) to 98.4% (95% CI: 96.0–99.6%). Method comparison with predicate (n=354) showed 97.5% total agreement. Analytical studies included precision (CV <10%), linearity, and interference testing.
Technological Characteristics
In vitro semi-quantitative immunoassay; utilizes EliA IgG method; compatible with Phadia 100 and Phadia 250 automated instruments; detects IgG antibodies to Sm in human serum and plasma.
Indications for Use
Indicated for the in vitro semi-quantitative measurement of IgG antibodies to Sm in human serum and plasma (EDTA, citrate) as an aid in the clinical diagnosis of systemic lupus erythematosus (SLE) in conjunction with other clinical and laboratory findings.
Regulatory Classification
Identification
An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k132631
B. Purpose for Submission:
New device
C. Measurand:
IgG antibodies specific for Sm protein
D. Type of Test:
Fluoroenzymeimmunoassay, Semi-quantitative
E. Applicant:
Phadia US Inc.
F. Proprietary and Established Names:
EliA™ SmD⁺ Immunoassay
G. Regulatory Information:
1. Regulation section:
21 CFR §866.5100 Antinuclear Antibody Immunological Test System
2. Classification:
Class II (Assay)
3. Product code:
LKP, Anti-Sm Antibody, Antigen and Control
4. Panel:
Immunology (82) (Assays)
{1}
H. Intended Use:
1. Intended use(s):
EliA SmD⁺ is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Sm in human serum and plasma (EDTA, citrate) to aid in the diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA SmD⁺ uses the EliA IgG method on the instrument Phadia 100.
EliA SmD⁺ is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Sm in human serum and plasma (EDTA, citrate) to aid in the diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA SmD⁺ uses the EliA IgG method on the instrument Phadia 250.
2. Indication(s) for use:
Same as Intended Use
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Phadia 100 and Phadia 250 instruments
I. Device Description:
EliA™ uses a modular reagent system. The test specific, method specific and general reagents are packaged and purchased as separate units. The reagents on Phadia® 100Phadia® 250 are identical; they are only filled in different containers.
EliASmD⁺ Test-Specific Reagents consist of:
1) EliA™ SmD⁺ wells coated with synthetic SmD3 peptide
2) EliA™ SmD⁺ Positive Control, containing human monoclonal antibodies specific to SmD3 peptide
3) EliA™ IgG/IgM/IgA Negative Control containing normal human serum from healthy donors.
The control materials were cleared k072393 and k131821.
Also required for the test are EliA™ Method-Specific Reagents:
EliA™ IgG Calibrators (human IgG in PBS (0, 4, 10, 20, 100, 600 μg/L), EliA™ IgG Curve Control (human IgG (5μg/L) in PBS), EliA™ Sample Diluent (PBS containing BSA, detergent and 0.095% sodium azide), EliA™ IgG Conjugate (β-galactosidase labeled mouse monoclonal anti-human IgG).
{2}
J. Substantial Equivalence Information:
1. Predicate device name(s):
Varelisa Sm Antibodies
2. Predicate 510(k) number(s):
k042629
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use / Indications for Use | EliA SmD^{P} is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Sm in human serum and plasma (EDTA, citrate) to aid in the diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA SmD^{P} uses the EliA IgG method on the instruments Phadia 100 and Phadia 250. | The Varelisa Sm Antibodies EIA kit is designed for the semiquantitative and qualitative determination of SmD antibodies in serum or plasma to aid in the diagnosis of systemic lupus erythematosus (SLE). |
| Assay Type | ELISA | Same |
| Type of Test | Semi-quantitative | Same |
| Antigen | Synthetic SmD3 peptide | Same |
| Solid Phase | Microwells | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Instrumentation | Phadia 100 and 250 are fully automated immunoassay analyzers | ELISA Reader needed |
| Controls | Positive and Negative Controls sold separately | Positive and Negative Control Sera provided with assay |
| Reaction temperature | 37°C (controlled) | Room temperature, 18-25°C |
| Incubation times | Diluted patient samples: 30 min. Conjugate: 28 min. | Calibrators, Controls, diluted patient samples: 30 |
{3}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | Development Solution: 39 min. | min. Conjugate: 30 min. Substrate: 10 min. |
| Detection antibody (conjugate) | mouse monoclonal anti-human IgG β-Galactosidase | rabbit polyclonal anti-human IgG horse-radish peroxidase |
| Signal | Fluorescence | Optical density |
| Calibration | Option to store total IgG calibration curve for up to 28 days and run curve controls in each assay for calibration | Analyte-specific IgG calibration curve in each test |
| Calibrators | Set of six vials of human IgG at concentrations of 0, 4, 10, 20, 100, and 600 μg/L | Set of six vials of Sm antibody calibrators at concentrations of 0, 3, 7, 16, 40, and 100 U/mL |
| Sample Dilution | 1:50 | 1:101 |
| Reportable Range | 0.8 U/mL – 480 U/mL | 0.5 U/mL - 100 U/mL |
| Cut-off / Results Interpretation | <7 U/mL negative
7-10 U/mL equivocal
>10 U/mL positive | <10 U/mL negative
10-15 U/mL equivocal
>15 U/mL positive |
# K. Standard/Guidance Document Referenced (if applicable):
CLSI EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation
# L. Test Principle:
The EliA $\mathrm{SmD}^{\mathrm{P}}$ Wells are coated with a synthetic SmD3 peptide. If present in the patient's specimen, antibodies to the SmD3 peptide bind to their specific antigen. After washing away non-bound antibodies, enzyme-labeled antibodies against human IgG antibodies (EliA IgG Conjugate) are added to form an antibody-conjugate complex. After incubation, non-bound conjugate is washed away and the bound complex is incubated with a Development Solution. After stopping the reaction, the fluorescence in the reaction mixture is measured. The higher the response value, the higher the amount of antibody bound and detected in the sample tested. To evaluate test results, the response for patient samples is compared directly to the response for calibrators.
The total IgG calibration is based on a set of six WHO-standardized IgG Calibrators derived from human serum. They are used to establish an initial calibration curve, which may be used for up to 28 days on additional assays and can be stored by the instrument. Each additional assay includes calibrator (curve) controls that have to recover in defined ranges to ensure that the stored calibration curve is still valid. Limits for the response of the Curve Controls are defined in the Phadia 100/250 Operator and Panel Software.
{4}
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The precision of the EliA SmD$^{\text{P}}$ assay on the Phadia 100 was assessed in a study with 8 samples. Each sample was tested over 7 runs, 4 replicates/run on 3 Phadia 100 instruments over 7 days for a total of 84 replicate determinations per sample). A new calibration curve was performed on each run. The results are presented in the table below.
| EliA SmD$^{\text{P}}$ on Phadia 100 | | | |
| --- | --- | --- | --- |
| Mean (U/mL) | Intra run (CV%) | Inter run (CV%) | Total |
| 8.5 | 4.8 | 2.5 | 5.4 |
| 17.8 | 6.0 | 3.5 | 6.0 |
| 364.7 | 6.4 | 4.5 | 7.8 |
| 4.0 | 11.2 | 14.3 | 18.2 |
| 11.5 | 3.8 | 4.4 | 5.8 |
| 39.9 | 3.2 | 4.2 | 5.3 |
| 82.1 | 2.7 | 3.9 | 4.7 |
| 293.5 | 5.0 | 6.1 | 7.8 |
The precision of the assay on the Phadia 250 was assessed as described above for the Phadia 100 instrument.
| EliA SmD$^{\text{P}}$ on Phadia 250 | | | |
| --- | --- | --- | --- |
| Mean (U/mL) | Intra run (CV%) | Inter run (CV%) | Total |
| 7.7 | 6.5 | 1.6 | 6.7 |
| 17.5 | 5.3 | 3.4 | 6.3 |
| 377.7 | 4.0 | 2.6 | 4.7 |
For all samples, the inter-run and intra-run coefficients of variation of EliA SmD$^{\text{P}}$ were within the sponsor's acceptance criteria (<10% and <8% respectively).
To determine the lot-to-lot reproducibility of the assay on the Phadia 250, 3 samples were tested using 3 lots, over 7 runs (1 run/day), 4 replicates/run on 3 Phadia 250 instruments for a total of 252 repetitions. A calibration curve was included in each run. The results are presented in the table below.
{5}
| EliA SmD^{P} Lot-to-Lot Reproducibility | |
| --- | --- |
| Mean (U/mL) | Reproducibility (CV%) |
| 7.7 | 3.8 |
| 17.5 | 9.5 |
| 377.7 | 1.1 |
b. Linearity/assay reportable range:
Six patient serum samples were diluted to produce 7 levels of reactive specimens that were analyzed to assess linearity.
| Linearity of EliA^{TM} SmD^{P} on Phadia 100 | | | | | |
| --- | --- | --- | --- | --- | --- |
| Sample | Dilution range (U/mL) | Slope (95% CI) | Y-Intercept (95% CI) | R^{2} | %CV Range |
| 1 | 9.2 - 78.3 | 0.99 (0.96 to 1.02) | 0.08 (-2.03 to 0.62) | 1.00 | 0.4 - 3.0 |
| 2 | 3.7 - 10.2 | 1.24 (1.20 to 1.28) | -1.42 (-1.70 to 1.14) | 1.00 | 2.21 - 4.5 |
| 3 | 9.7 - 207.5 | 1.067 (1.02 to 1.11) | -0.67 (-4.91 to 3.57) | 1.00 | 1.5 - 5.4 |
| 4 | 5.2 - 252.7 | 1.04 (1.01 to 1.08) | 0.92 (-2.61 to 4.45) | 1.00 | 0.6 - 6.1 |
| 5 | 6.5 - 398.3 | 1.02 (0.99 to 1.05) | 2.41 (-2.17 to 7.00) | 1.000 | 0.6 - 7.4 |
| 6 | 9.2 - 78.3 | 0.99 (0.96 to 1.02) | -0.7051 (-2.03 to 0.62) | 1.000 | 1.4 - 3.0 |
| Linearity of EliA^{TM} SmD^{P} on Phadia 250 | | | | | |
| --- | --- | --- | --- | --- | --- |
| Sample | Dilution range (U/mL) | Slope (95% CI) | Y-Intercept (95% CI) | R^{2} | %CV Range |
| 1 | 5.5 - 74.3 | 1.04 (1.03 to 1.05) | -1.03 (1.36 to 0.70) | 1.00 | 0.3 - 4.5 |
| 2 | 3.8 - 281.3 | 1.09 (1.03 to 1.14) | 1.91 (-3.75 to 7.57) | 1.00 | 0.9 - 5.3 |
| 3 | 6.6 - 210.7 | 1.09 (1.08 to 1.10) | -2.54 (-3.88 to -1.20) | 1.00 | 0.9 - 7.2 |
| 4 | 8.8 - 227.4 | 1.06 (1.04 to 1.09) | 1.192 (-0.95 to 3.34) | 1.00 | 0.5 - 6.5 |
| 5 | 6.1 - 419.8 | 1.12 (1.11 to 1.13) | -0.36 (-2.20 to 1.47) | 1.00 | 1.3 -4.5 |
| 6 | 5.5 - 74.3 | 1.04 (1.03 to 1.05) | -1.03 (-1.36 to -0.70) | 1.00 | 0.3 - 4.5 |
The limits of the measuring range were set to 0.8 - >480 U/mL. The upper limit of the reportable range can vary between 480 and 720 U/mL, due to a lot-specific conversion
{6}
from $\mu \mathrm{g} / \mathrm{L}$ to EliA U/mL. Results above the upper limit are reported as "above".
High dose hook effect: A hook effect was not observed when analyzing a high positive sample that had a concentration up to 8 times above the upper limit of the measuring range.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
i) Traceability:
Calibrators: There is no international reference standard for IgG antibodies that specifically recognize Sm or the SmD3 peptide. The instrument measures specific IgG concentrations in $\mu \mathrm{g} / \mathrm{L}$ . By using a conversion factor given by the lot-specific code of the EliA SmD Well, the results are automatically converted to EliA Units/mL.
EliA IgG Calibrators and Curve Controls are traceable (via an unbroken chain of calibrations) to the International Reference Preparation (IRP) 67/86 of Human Serum Immunoglobulins A, G and M from WHO. New batches of IgG Calibrators are compared to a secondary standard (standardized with the IRP) or the IRP directly and adjusted accordingly to meet the correct concentration.
The calibrators are required to perform an initial calibration curve, which can be stored in the Phadia instrument and may be used for up to 28 days on this and additional IgG assays. Each additional assay outside of a calibration run includes curve controls that have to fall within defined ranges to verify that the stored calibration curve is still valid.
Controls: EliA ANA Positive Control is prepared from selected pooled human sera and contains IgG antibodies to dsDNA, RNP, Sm, Ro, La, Scl-70, CENP and Jo-1. This control was cleared under k072393. The EliA™ IgG/IgM/IgA Negative Control was prepared from normal human serum. This control was cleared under k131821. The target ranges for the EliA™ Controls for the two platforms are summarized below:
| Instrument | EliATM Positive SmDP Control |
| --- | --- |
| Phadia 100 | 31.7 – 73. U/mL |
| Phadia 250 | 31.1 – 72.5 U/mL |
| Instrument | EliATM IgG/IgM/IgA Negative Control |
| Phadia 100 | ≤ 4 U/mL |
| Phadia 250 | ≤ 4 U/mL |
ii) Kit Stability:
{7}
Closed and open stability - An accelerated stability study initially determined the shelf-life of the EliA SmD$^{\text{P}}$ Well was 24 months. A real-time stability study supported a 24 months stability claim. All studies were performed on three batches of EliA SmD$^{\text{P}}$ Well. Other required components (previously reviewed) of the assay method have a shelf life of 18 to 24 months. The sponsor notes that it is important to store the wells in dry conditions at 2-8°C.
On-board stability - The on-board stability of the EliA SmD$^{\text{P}}$ Wells packed in carriers were tested for 2, 4 and 6 weeks at 10°C and 80% humidity in duplicates in one run using 3 positive and 2 negative samples only on the Phadia 250 instrument since for Phadia 100 instrument the reagents are stored outside the instrument and are only loaded as needed for an assay. The on-board stability for the EliA™ SmD$^{\text{P}}$ Wells was determined to be 28 days at 2-8°C.
iii) Sample Storage
The sponsor recommends following the guidelines in CLSI H18-A3 for sample storage. Separated serum/plasma should remain at room temperature for no longer than eight hours. If assays will not be completed within eight hours, serum/plasma should be refrigerated (2 to 8°C). If assays are not completed within 48 hours, or the separated serum/plasma will be stored beyond 48 hours, serum/plasma should be frozen at or below -20°C. Freezing and thawing should be avoided.
d. Detection limit:
The Limit of Blank (LOB), Limit of Detection (LOD) and Limit of Quantitation were determined based on CLSI EP17-A, “Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline, EP17-A”. On each system (Phadia 100 and Phadia 250), six samples with undetectable levels of analyte were spiked with EliA Sample Diluent and were measured in 72 replicates. The sample with the lowest value was chosen to be the blank sample while the other 5 samples were deemed ‘low positives’ for this experiment. For Phadia 100 and Phadia 250, each sample was run in a total of six runs, three runs in two instruments.
The values determined are presented here:
| EliA™ SmD$^{\text{P}}$ | LoB | LoD | LoQ |
| --- | --- | --- | --- |
| Phadia 100 | 0.39 U/mL | 0.69 U/mL | 0.69 U/mL |
| Phadia 250 | 0.45 U/mL | 0.73 U/mL | 0.73 U/mL |
e. Analytical specificity:
Comparison to Reference Sera: Twelve reference sera were obtained from the Centers for Disease Control and Prevention (CDC), Atlanta, GA. These sera have known reactivity for a number of autoantibodies that include DNA, SS-B/La, U1-RNP, SS-A/Ro, PCNA, Pol III, PM/Scl, CENP, Scl-70, Jo-1, PM/Scl, rRNP/Rib P. In addition,
{8}
two of the reference sera were known to be positive for Sm. Among the 12 sera, only the two sera that were known to be positive for Sm were found to be reactive in the EliA SmDP assay.
Endogenous Interference: Interferences were assessed by testing three positive serum samples: two samples around the cut-off (11 and 12 U/mL), and a high positive (>110 U/mL). Each sample was spiked with the interfering substances or substance-specific blanks and analyzed using one lot of EliA SmD⁶ Well and one lot of system reagents in two runs, each in three replicates (n=6). The data demonstrated that SmD⁶ was not adversely affected by high levels of the following substances tested up to the concentrations listed in the table below:
| Potential Interfering Compound | Concentration |
| --- | --- |
| Bilirubin F | 19.2 mg/dL |
| Bilirubin C | 20.1 mg/dL |
| Hemoglobin | 496 mg/dL |
| Lipemic factor (ClinOleic) | 1 g/dL (1%) |
| Rheumatoid factor | 500 IU/mL |
f. Assay cut-off:
Based on the results of the expected values/reference range study described below in Section M.5, the 99th percentile lies below the upper limit of the equivocal range for the EliA SmD⁶ assay. The cut-off between equivocal and positive was set to 10 U/mL. The assay cutoffs were set as follows:
| Decision Point | Interpretation |
| --- | --- |
| <7 U/mL | Negative |
| 7-10 U/mL | Equivocal |
| >10 U/mL | Positive |
Where specimens yield equivocal results, the sponsor recommends retesting the patient again after 6-8 weeks.
2. Comparison studies:
a. Method comparison with predicate device:
One hundred twenty samples from individuals diagnosed with SLE and 250 specimens from individuals with other autoimmune and infectious conditions were tested (see 3.a. below for a detailed list). In addition, 20 specimens with known anti-Sm reactivity, but without clinical diagnosis, were added to the analysis.
{9}
Of the 370 specimens tested, 16 were excluded as being outside of the measuring range for one or both methods. For analysis of agreement with the predicate device, 354 sera were measured on the Phadia 250 instrument. All samples were within the reportable range for both assays.
| | Varelisa Sm (Sm Units) | | | | |
| --- | --- | --- | --- | --- | --- |
| | | Positive: >15 | Indeterminate: 10-15 | Negative: <10 | Total |
| EliATM SmD^{P} (U/mL) | Positive: >10 | 39 | 4 | 2 | 45 |
| | Equivocal: 7-10 | 2 | 0 | 3 | 5 |
| | Negative: <7 | 1 | 10 | 293 | 304 |
| | Total | 42 | 14 | 298 | 354 |
Agreements were calculated by grouping each assays' equivocal results with its test negative results, and then agreement was calculated again by grouping each assays' equivocal results with the test positive results:
| Equivocal results considered as negative | Varelisa Sm (Sm Units) | | | |
| --- | --- | --- | --- | --- |
| | | Positive: >15 | Negative: <15 | Total |
| EliA SmD^{P} (U/mL) | Positive: >10 | 39 | 6 | 45 |
| | Negative: <10 | 3 | 306 | 309 |
| | Total | 42 | 312 | 354 |
Positive percent agreement: 92.9% (39/42) 95% CI: 80.5 - 98.5%
Negative percent agreement: 98.1% (306/312) 95% CI: 95.9 - 99.3%
Total percent agreement: 97.5% (345/354) 95% CI: 95.2 - 98.8%
| Equivocal results considered as positive | Varelisa Sm (Sm Units) | | | |
| --- | --- | --- | --- | --- |
| | | Positive: >10 | Negative: <10 | Total |
| EliA SmD^{P} (U/mL) | Positive: >7 | 45 | 5 | 50 |
| | Negative: <7 | 11 | 293 | 304 |
| | Total | 56 | 298 | 354 |
Positive percent agreement: 80.4% (45/56) 95% CI: 67.6 - 89.8%
Negative percent agreement: 98.3% (293/298) 95% CI: 96.1 - 99.5%
Total percent agreement: 95.5% (338/354) 95% CI: 92.8 - 97.4%
b. Matrix comparison:
Twenty-five negative subjects and twenty-five positive subjects provided sera and plasma specimens (EDTA and citrate) and these were tested for $\mathrm{SmD}^{\mathrm{P}}$ on the Phadia 250. The sponsor's acceptance criteria were that each plasma sample should yield results within $20\%$ of the serum specimen and that all negative specimens should remain negative. A Passing & Bablok analysis was performed and the results are depicted in the following table:
{10}
| | Range tested (U/mL) | Slope (95% CI) | Intercept (95% CI) | R² |
| --- | --- | --- | --- | --- |
| Serum vs. Citrate plasma | 0.5 – 433.9 | 0.99 (0.98 to 1.01) | -0.09 (-0.82 to 0.15) | 1.00 |
| Serum vs. EDTA plasma | 0.5 – 413.1 | 1.00 (0.97 to 1.01) | 0.19 (-0.39 to 0.70) | 1.00 |
c. Instrument comparison:
In order to determine if the performance of EliA $\mathrm{SmD}^{\mathrm{P}}$ is equivalent on the instruments Phadia 100 and Phadia 250, 36 samples distributed over the measuring range were tested: 25 positive, 7 equivocal and 4 negative samples. All samples were run on 3 different Phadia 100 and 3 different Phadia 250 instruments. Each sample was tested once on each instrument. The results from Phadia 100 Instrument 1 were compared to the results from Phadia 250 Instrument 1, and so on for the other two instrument pairs. The results of a Weighted Deming regression analyses performed on each pair are shown below.
| EliATM SmDP Weighted Deming Regression: Phadia 100 vs Phadia 250 | | |
| --- | --- | --- |
| Instrument Pair | Slope (95% CI) | Y-Intercept (95% CI) |
| 1 | 0.95 (0.92 to 0.97) | -0.78 (-1.13 to -0.44) |
| 2 | 0.97 (0.92 to 1.01) | -0.64 (-1.33 to 0.05) |
| 3 | 0.98 (0.94 to 1.03) | -0.47 (-1.04 to 0.10) |
3. Clinical studies:
a. Clinical Sensitivity and Specificity:
One hundred samples from individuals diagnosed with SLE and 250 specimens from individuals with other autoimmune and infectious conditions were tested on the Phadia 250 instrument.
| Equivocal results considered as positive | SLE Diagnosis | | | |
| --- | --- | --- | --- | --- |
| | | Positive: | Negative: | Total |
| EliA SmDP(U/mL) | Positive: > 10 | 75 | 4 | 79 |
| | Equivocal: 7-10 | 2 | 2 | 4 |
| | Negative: < 7 | 23 | 244 | 267 |
| | Total | 100 | 250 | 350 |
Clinical sensitivity and specificity were calculated by grouping the assay's equivocal results with its test negative results, and then sensitivity and specificity was calculated again by grouping the assay's equivocal results with the test positive results:
{11}
12
| Equivocal results considered as negative | SLE Diagnosis | | | |
| --- | --- | --- | --- | --- |
| | | Positive: | Negative: | Total |
| EliA SmD^{P} (U/mL) | Positive: >10 | 23 | 4 | 27 |
| | Negative: < 10 | 77 | 246 | 323 |
| | Total | 100 | 250 | 350 |
Sensitivity: 23.0% (23/100) 95% CI: 15.2 – 32.5%
Specificity: 98.4% (246/250) 95% CI: 96.0 - 99.6%
| Equivocal results considered as positive | SLE Diagnosis | | | |
| --- | --- | --- | --- | --- |
| | | Positive: | Negative: | Total |
| EliA SmD^{P} (U/mL) | Positive: >7 | 25 | 6 | 31 |
| | Negative: <7 | 75 | 244 | 319 |
| | Total | 100 | 250 | 350 |
Sensitivity: 25.0% (25/100) 95% CI: 16.9 - 34.7%
Specificity: 97.6% (244/250) 95% CI: 94.8 – 99.1%
Performance on Non-SLE Specimens:
One hundred fifty specimens from individuals having other conditions were tested for SmD<sup>P</sup> reactivity, including dermatomyositis, polymyositis, scleroderma, mixed connective tissue disease, rheumatoid arthritis, Hashimoto thyroiditis, Graves' Disease and Sjögren's Syndrome. In this study, no more than 5% of the specimens that were tested were found to be positive for SmD<sup>P</sup> at a level > 10 U/mL.
| | Performance on Non-SLE Specimens | | |
| --- | --- | --- | --- |
| Non-SLE Patient Sub-Group | N | No (%) Positive* on EliA SmD^{P} | No (%) Positive* on predicate |
| Dermatomyositis | 4 | 0 (0%) | 0 (0%) |
| Polymyositis | 6 | 0 (0%) | 0 (0%) |
| Scleroderma | 30 | 1 (3%) | 1 (3%) |
| MCTD | 20 | 1 (5%) | 0 (0%) |
| RA | 50 | 1 (2%) | 2 (4%) |
| Hashimoto | 10 | 0 (0%) | 0 (0%) |
| Graves' Disease | 10 | 0 (0%) | 0 (0%) |
| Sjögren's Syndrome | 20 | 0 (0%) | 0 (0%) |
| Bacterial infections | 30 | 0 (0%) | 0 (0%) |
| Viral infections | 30 | 0 (0%) | 1 (3%) |
* Equivocal samples (2) were considered negative for this analysis
b. Other clinical supportive data:
Not applicable.
{12}
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Four hundred samples from healthy Caucasian individuals were measured on the Phadia 250 instrument. The results are presented in the following table:
| | U/mL |
| --- | --- |
| Mean | 1.4 |
| Median | 2.8 |
| Range | 0.1 – 9.1 |
| 95th percentile | 3.0 |
| 99th percentile | 6.7 |
The results were equally distributed and not dependent on age or gender. No samples were positive ( $>10\mathrm{U/mL}$ ) in this study, but 3 (0.75%) of the samples fell in the equivocal range. Expected values may vary depending on the population tested. A subset $(n=70)$ of the samples tested on the Phadia 100 gave similar results.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.